Optimization of the mass scanning rate for the determination of lead isotope ratios using an inductively coupled plasma mass spectrometer

Abstract

High-speed motion pictures of a plasma and the noise–amplitude spectra of the emission signals enabled clarification that the inductively coupled plasma coupled with a mass spectrometer had a noise component at a frequency of 380 Hz under the experimental conditions used. In order to improve the precision of lead isotope ratio measurements the mass scanning rate was optimized. With an increase in the mass scanning rate, the precision of lead isotope ratio measurements was improved because the low frequency noises could be eliminated. However, the quadrupole mass analyser could not be operated at a speed that was sufficient to smooth out the audio frequency noise.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1991,6, 199-203

Optimization of the mass scanning rate for the determination of lead isotope ratios using an inductively coupled plasma mass spectrometer

N. Furuta, J. Anal. At. Spectrom., 1991, 6, 199 DOI: 10.1039/JA9910600199

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements