Transient responses of glass electrodes in dilute solutions and freshwaters

Abstract

Glass electrodes respond slowly in dilute, nearly neutral solutions (l= 0.1–1 mmol dm–3, pH 6–9) of either natural lake water or simple inorganic salts, taking ca. 10 min to equilibrate irrespective of whether the solution is flowing or static. Although buffer capacity and ionic strength influence the response, they are not critical factors. The transient effect is associated with low concentrations of cations. There was no great difference in response for the different cations tested, but the sluggishness appeared to increase in the order K+≈ Ca2+ < Na+≈ Mg2+ < NH4+. The time dependence of the response is consistent with it being due to a diffusional exchange process. Glass electrodes are inappropriate pH sensors for the determination of the pH of dilute neutral solutions by flow injection analysis.

Article information

Article type
Paper

Analyst, 1988,113, 1537-1540

Transient responses of glass electrodes in dilute solutions and freshwaters

W. Davison and T. R. Harbinson, Analyst, 1988, 113, 1537 DOI: 10.1039/AN9881301537

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements