Effect of variation of temperature on second-derivative fluorescence spectra. Part II. Effect of experimental parameters on the observed temperature coefficients

Abstract

The effects of varying a number of experimental parameters on the temperature coefficients of the second-derivative fluorescence spectral amplitudes of chlorpromazine sulphoxide in pH 8 borate buffer have been investigated. The temperature coefficients were found to decrease with increasing concentration and with increasing scan speed, to increase with increasing excitation and emission slit widths and to vary widely with the wavelengths at which the measurements were made. These effects are discussed in relation to the observed or theoretical changes that occur in the fundamental spectra as a result of variation of the experimental parameters.

Article information

Article type
Paper

Analyst, 1988,113, 537-541

Effect of variation of temperature on second-derivative fluorescence spectra. Part II. Effect of experimental parameters on the observed temperature coefficients

A. G. Davidson and E. O. Fadiran, Analyst, 1988, 113, 537 DOI: 10.1039/AN9881300537

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements