Fluorimetric determination of aluminium with morin after extraction with isobutyl methyl ketone. Part I. Fluorescence of the aluminium-morin complex in an isobutyl methyl ketone-ethanol-water system
Abstract
A method for the fluorimetric determination of aluminium with morin (3,5,7,2′,4′-pentahydroxyflavone) in an isobutyl methyl ketone (IBMK)-ethanol-water system is proposed, with a 3-to 4-fold increase in sensitivity over the conventional method. The optimum conditions were found to be as follows: 40% of IBMK, 20% of an aqueous solution containing the aluminium and HCl-KCl to adjust the pH and 40% of methanol-ethanol-water (86 + 9 + 5) solution, which is necessary for the dispersion of the IBMK in water, and in which the reagent, morin, is dissolved. The optimum pH is 3.7–3.8 (HCl-KCl buffer) and the fluorescence is measured after 1 h at 495 nm, with excitation at 418 nm, at 25 °C.
Using a morin concentration of 0.005%, the calibration graph is linear for concentrations up to 40 p.p.b., and the limit of detection of the method is 0.2 p.p.b. A study of 17 interfering ions showed that the most important interferences are due to Be2+, Zn2+, Pb2+, Fe3+, Fe3+, Cu2+ and F–, and in most instances it was found that the tolerance limits are higher than with the conventional method for the determination of aluminium with morin.
Please wait while we load your content...