Issue 1201, 1976

The use of a spark as a sampling-nebulising device for solid samples in atomic-absorption, atomic-fluorescence and inductively coupled plasma emission spectrometry

Abstract

A conventional high-voltage spark, operating at 50 Hz, is used for sampling and nebulising material from solid conducting samples. Gas, fed through the spark chamber, transports the metal particles into a flame or plasma for analysis by atomic-absorption, atomic-fluorescence or plasma emission spectrometry. Calibration graphs for the copper present in aluminium alloys and iron in brass are presented.

Article information

Article type
Paper

Analyst, 1976,101, 265-271

The use of a spark as a sampling-nebulising device for solid samples in atomic-absorption, atomic-fluorescence and inductively coupled plasma emission spectrometry

H. G. C. Human, R. H. Scott, A. R. Oakes and C. D. West, Analyst, 1976, 101, 265 DOI: 10.1039/AN9760100265

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.