A low-cost titanium suboxide pH sensor with competitive operational lifetime assessed with electrochemical impedance spectroscopy
Received
2nd December 2025
, Accepted 11th June 2026
First published on 11th June 2026
Abstract
Smart Bandages are an emerging field of wearable electronics. Biosensors provide a medium of communication between the wound site's chemical status and the clinicians. The pH is a critical biomarker, as it is a suitable indicator for bacterial infection. In this paper, a low-cost electrochemical impedance spectroscopy (EIS)-based titanium suboxide (TiOx) pH sensor with silver (Ag) electrodes is presented. A sensitivity of −27.8 Ω pH−1 is achieved in the pH 4–8 region of interest related to wound infection. The measured pH sensitivity, after a one-month stability study, is lost only 2.25% for the TiOx sensor - reaching the state-of-the-art - compared to the Polyaniline (PANI) sensor, which lost 10% of its pH sensitivity. To further assess stability, we evaluated sensor behaviour at the edge pH values. The TiOx sensor exhibits impedance increases of 5.5% and 4% at pH 4 and 8, respectively, which are approximately half of those reported for PANI devices (10% and 9% at pH 4 and 8, respectively). The stability experiments highlighted TiOx's chemical stability relative to PANI in both acidic and alkaline media. Further, the sensitivity of the TiOx sensors is recovered to 95.7% after 200 °C thermal annealing, thereby recalibrating the sensors and prolonging their lifetime whilst reducing waste. To better understand the sensing and degradation mechanisms, a model-free method is employed for both TiOx and PANI pH sensors. The reason is to demonstrate that pH sensitivity is related to the film surface (given the circuit complexity of the equivalent models in 2-electrode systems). This work is an important step for a potential translation to real-world wound monitoring, where long-term stability and low cost are critical for an accurate and efficient biomarker detection and sustainable well-being.
Introduction
Chronic wounds, defined by a failure to heal within 30 days,1,2 represent a significant and growing healthcare challenge. Smart bandages have been proposed as a solution for monitoring chronic wounds.3,4 Their role is to track wound conditions and enable the early detection of infections and/or delayed healing, making them preferable to conventional bandages. These biomedical platforms contribute to the field of wearable electronics5 by not only monitoring biomarkers (such as pH, temperature, glucose, etc.)6 at the wound site, but also enabling on-demand therapeutics.7 Specifically, a biosensor array is embedded to transduce data (such as the concentration of analytes and metabolites8) both with faradaic (charge transfer) or non-faradaic (adsorption)9 chemiresistive approaches. These electrochemical phenomena are translated to the healthcare providers or users as voltage or current10,11 signals, while also possessing the capability to provide targeted drug delivery. Biosensors for biomarker monitoring, embedded in smart bandages, provide several benefits versus conventional bandages, such as a decrease in the number of wound dressing changes,4 reduced healthcare costs and hospitalisation time.12 These benefits are critical for the long-term health and well-being, one of the Sustainable Development Goals adopted by the United Nations.13
Among the monitored biomarkers, pH is important for chronic wound management because an alkaline pH is a key indicator of bacterial infection14 at the wound surface. Also, an accurate pH measurement is essential for calibrating15 other enzymatic biosensors (e.g., for glucose and lactate). Organic semiconductors like polyaniline (PANI)16 are widely used for pH sensing due to their high proton sensitivity.17 However, PANI changes phase from Emeraldine salt to base, which leads to limited chemical stability18 in aqueous environments. The tendency of organic semiconductors to oxidise18 and swell19 causes performance degradation over periods ranging from hours20 up to 14 days.21–23 These stability periods make organic semiconductors inadequate for chronic wound monitoring, which requires stability over several weeks to months. Therefore, an accurate aftercare cannot be accomplished, as it takes one month to determine whether a wound is chronic or not.1
In contrast, transition metal oxides (TMOs) such as iridium suboxide (IrOx24–27), molybdenum suboxide (MoOx28), and titanium suboxide (TiOx)29,30 have demonstrated a long-term chemical stability in emerging applications such as electrocatalysis,31,32 memristors33 and solar cells.34,35 However, the existing stability studies on pH sensors have demonstrated 2%–10%16,23,36 degradation, but with some caveats related to the testing stability protocol and the real-world timeline for chronic wound monitoring (one-month lifetime). The reported stability tests are often too short (a few hours37,38 to two weeks23) or conducted at a single pH value (mostly around 7.4,23,36 837 or 5.516). Also, studies involving high-temperature in situ processes (e.g., continuous heating of the substrate at 800 °C for Al2O339) are incompatible with flexible, wearable substrates and IoT applications. Furthermore, other studies report substantial losses in sensitivity (∼20%40) or an increase (∼50%26) in the response time after four weeks. Hence, in this work, we report a more extensive chemical stability study that monitors a pH sensor: (a) at high and low pH values, (b) sensitivity over one month and (c) recovery after 3 months.
Most wearable pH sensors rely on potentiometric methods like open circuit potential (OCP),41 for detecting voltage changes for different pH solutions, which offers a non-invasive measurement (steady-state equilibrium, no current in the cell42). However, OCP provides limited insight into the interfacial kinetics at the electrode–electrolyte interface, which are related to the pH sensing (proton detection) mechanism.16,43 Electrochemical impedance spectroscopy (EIS)44 is a powerful technique that characterises these interfacial processes—double-layer charging,45 charge-transfer resistance,46 ionic motion.47 EIS operates with a minimal signal perturbation by applying a small AC voltage (typically 10 mV) across a frequency spectrum (Hz to MHz).48
While EIS is typically performed using a three-electrode setup, working electrode (WE), counter electrode (CE) and reference electrode (RE), a two-electrode system (WE, CE – such as in this paper, see Fig. 1) is preferred for wearable applications due to its simplicity and necessity for RE's calibration. This calibration process requires care with electrolyte management, junction maintenance, and cleaning, which is largely incompatible with flexible smart bandage designs. Previous attempts at two-electrode impedimetric pH sensing, achieving sensitivities of some kΩ pH−1
49–52 but have faced issues with structural stability (interdigitated electrodes structural instability after two weeks50,52). Also, they have tested at a limited pH range (pH 6.5–7.550), poor reproducibility (high impedance after 3–4 tests51), and a lack of validation regarding signal causality and selectivity. Moreover, a full frequency sweep requires complex hardware that increases the size and power consumption of a wearable device. Therefore, a pH sensor that leverages the analytical power of EIS while operating at a single optimal frequency would be advantageous. Single-frequency EIS provides an informative and accurate signal while minimising the hardware complexity.
 |
| | Fig. 1 The sensor structure will be utilised in this paper. The TiOx and PANI films were fabricated using e-beam and drop-casting, respectively. Concept: from pH concentration to a measurable electric signal (EIS signal, in this study). | |
Furthermore, advanced EIS data analysis techniques (except for the typical circuit equivalent fitting53), such as the Distribution of Relaxation Times (DRT),54,55 offer a model-free method to deconvolute contributing electrochemical processes. DRT can cross-validate sensing mechanisms, enhancing the reliability of the data interpretation. Also, DRT eliminates the bias that is typically introduced from circuit models that often cannot deconvolute processes with near-overlapping relaxation time values.56
The electrode-conducting material is a key factor in electrochemical biosensors because it contributes significantly to the sensor's final cost. The current state-of-the-art for biosensors often relies on microfabricated electrodes made from noble metals such as gold (Au), platinum (Pt), palladium (Pd).5,15,57 Exploring alternative, low-cost materials such as silver (Ag)58 or aluminium (Al)51 for electrodes and TMOs (such as TiOx) for sensing layers, that do not compromise the signal integrity, is necessary for sustainable and accessible wearable technology. The Ag can be fully recycled many times without any loss of quality59 and many research works highlighted its longevity.60 Hence, the environmental harm associated with its use can be reduced, while its circularity provides an opportunity to reuse Ag resources without extracting new minerals from the earth. On the other hand, TiOx films (not nanoparticle form) are a well-known material for their longevity,61 and their water insolubility61 minimises the risk of leakage into the sea and sewage water.
The Ag can be a potential for integration to commercialised smart bandages, if the surface corrosion or any Ag+ release to the wound can be minimised. In the literature, many biocompatible materials such as Nafion,62 PDMS, Parylene C15,63 and Ti alloys64 have been employed for encapsulation. Hence, both the sensor's environmental stability and the Ag+ release may be reduced. Furthermore, cytocompatibility tests such as MTT assay, in accordance with the ISO 10993-5, are employed to investigate any cytotoxicity from the Ag+ release.65 A careful experimental planning such as:66,67 (a) washing cells with PBS and afterwards adding the fresh media with MTT to reduce the chemical interaction during the assay test, (b) conduct control tests between cells, MTT and Ag+ (in pairs) to subtract the background readings, and (c) cross-validate the results with less Ag+ sensitive assays, such as ATP-based assays can increase the reliability of the cytocompatibility results regarding the integration of Ag electrodes to smart bandages.
In this paper, a prototype single-frequency EIS pH sensor based on TiOx as the pH-sensing material is presented, highlighting its greater time stability (31 day stability studies and impedance recovery after 3 months) compared with the state-of-the-art PANI, which is much more expensive than TiOx. The effect of the pH on the electrochemical cell was quantified by measuring the cell's impedance in the 100 Hz–1 MHz frequency range. The TiOx pH sensor lost only 2.25% of its sensitivity for one month, compared to 10% observed for the reference PANI devices. Additionally, the TiOx sensor's impedance increased 5.5% (pH 4) and 4% (pH 8), compared to 10% (pH 4) and 9% (pH 8) observed for the reference PANI pH sensors, showing TiOx's superior interfacial chemical stability to both alkaline and acidic media. Furthermore, it was possible to regenerate the sensor's sensitivity by up to 95.6%, after a simple thermal annealing at 200 °C for 2 hours. Thermal annealing enhances the lifetime of the TiOx sensor and reduces waste. The sensing mechanism has also been examined, and the contribution of the TiOx surface to pH sensing and degradation has been demonstrated. To further demonstrate the sensor's functionality, experiments have been conducted to highlight the reversibility and causality of the measuring signal, thereby enhancing its reliability. The results have also indicated a minimum Ag corrosion during the stability experiments, further highlighting the TiOx sensor's potential for low-cost applications. Finally, the sensing and degradation mechanisms have also been studied using a model-free approach to cross-validate our results and further demonstrate the potential of low-cost TiOx-based pH sensor platforms for real-world chronic wound monitoring, where long-term accuracy and robustness are critical.
Experimental
Materials
Silver wire was purchased from Goodfellows Advanced Materials in the UK. TiO2 pellets from Testbourne Ltd, UK. PANI Emeraldine Base, HCl, acetone and isopropyl alcohol were purchased from Sigma Aldrich, UK. Si/SiO2 wafers were purchased from Si-Mat Silicon Materials Technologies, Germany. LOR5A was purchased from Kayaki Advanced Materials, US. S1813. AZ726MIF and Hellmanex (TM) III were purchased from Merck, Germany. Microposit Remover 1165 (MR1165) was purchased from DuPont, US.
Fabrication of the Ag electrodes
Ag electrodes were fabricated using bi-layer lift-off photolithography. Si/SiO2 (300 nm thermal growth) wafers were cleaned by dipping in 1% v/v Hellmanex solution in deionised water, acetone and isopropyl alcohol for 10 minutes each. Next, the wafers were dried using an N2 gun and O2 plasma-treated (Diener Zepto-BRs 200, 10 minutes, 200W). LOR5A was then spin-coated (Ems 6000 photoresist spinner) at 500 rpm for 15 s, followed by 3500 rpm for 60 sec. A baking step on a hotplate (Stuart UC152D) was followed at 180 °C for 330 seconds. After cooling for 15 minutes to clean room temperature, a layer of S1813 positive photoresist was spin-coated (Ems 6000 photoresist spinner), using the same conditions as for LOR5A, and then baked (Ems 1000-1) at 115 °C for 15 minutes, and finally cooled down to clean room temperature for 15 minutes. The electrodes were patterned using a laser writer (Heidelberg Instruments Mikrotechnik GmbH CDL66+, 70 mW) and developed using AZ726MIF for 3 minutes. The wafer was inserted into an e-beam evaporator (Univex e-beam evaporator), and a thin film of Ti/Ag (15 nm/85 nm) was deposited at an evaporation rate of 0.2 Å s−1 whilst the wafer was rotated at 25 rpm. The film thickness was monitored in situ using a quartz-crystal microbalance sensor. The vacuum at the start of the evaporation process was 5 × 10−7 Torr. After the evaporation finished, lift-off was completed by dipping the wafer into MR1156 remover at 65 °C for 3 hours.
TiOx film deposition
The TiOx film was deposited using e-beam evaporation (Univex e-beam evaporator) from the TiO2 source (Fig. 1). The evaporation rate was kept at 0.2 Å s−1, and the wafer was rotated at 25 rpm. The vacuum at the start of the evaporation process was 6 × 10−7 Torr. The TiOx material was deposited only onto the working electrode area using laser-structured evaporation masks. For this purpose, a Kapton sheet with a thickness of 75 µm was placed on top of a vacuum table, and the required design was scribed with a picosecond laser from Coherent (HyperRapid NXT). The lasing was carried out with a laser burst frequency and speed of 500 kHz and 500 mm s−1, respectively. This resulted in a total power of 5.0 W cm2.
PANI film deposition
PANI emeraldine base (PANI-EB) was dissolved in DMSO at 0.2 wt%, similar to the method described in.58 The solution was stirred overnight. Following that, it was doped in a 2% wt 0.02 M HCl solution to induce PANI emeraldine salt (PANI-ES) conductive species and stirred overnight. 1 µL of the final solution was drop-casted onto the WE (Fig. 1), and an annealing step followed, at 60 °C for 2 hours.
Electrochemical characterisation
The pH sensors were tested using McIlvaine buffer solutions in the range 4–8 (each solution pH tested using a pH meter (Hannah Instruments Ltd.) and pH indicator strips (FisherBrand). PLA 3-D printed rings were placed onto the WE and CE area to keep the buffer solution in the WE-CE area. Using a micropipette, 33 µL of each buffer solution was placed onto the electrode area. EIS were obtained using Gamry1000, at 10 mV ac Voltage, 0 DC bias, and at frequencies from 100 Hz to 1 MHz with 10 points per decade and normal mode of denoising. Each measurement began after a 120-second open-circuit delay to allow system stabilisation. The pH sensors were connected to the Gamry1000 using test stand PCB spring-loaded clips. For the time stability, two different types of tests were performed: the first evaluates time stability by measuring the impedance at 100 kHz for pH 4 and pH 8 buffer solutions, for separate devices, and the second by measuring the sensitivity in the range pH 4–8, in a 31 day time window. The devices were kept in each experiment's pH conditions throughout the stability time. After every EIS measurement, the droplet was removed using absorbent wipes.68 The buffer solutions were replaced and tested every 15 days.
Thermal annealing treatment
The thermal annealing treatment (for sensitivity recovery) was carried out using a Lenton Furnace (Lenton Laboratory & Scientific Equipment PTY (Ltd)). The plateau temperature was 200 °C for 2 hours, with a temperature increase rate of 1 °C min−1.
Kelvin probe measurements (KP)
Kelvin Probe measurements were performed in air (KPTechnology Ltd, version KP020). The work function (WF) of the sample is calculated by measuring the contact potential difference (CPD), using the formulas below:
WF of the tip = WF of the test sample – CPD (test sample)
WF of the sample = WF of the tip + CPD (sample)
The test sample is the highly ordered pyrolytic graphite (HOPG) with WF = 4.48 eV.69
Ultraviolet-visible (UV-vis) – absorption spectroscopy
UV-Vis spectra (absorption) were obtained using the Varian Cary5000 UV-vis-NIR Spectrophotometer (from 300 nm to 800 nm).
X-Ray photoelectron spectroscopy (XPS)
XPS spectra were obtained using a Thermo Fisher Scientific Instrument K-alpha + spectrometer, which consists of a monochromatic Al K-α X-ray source. The XPS data was analysed using Avantage software.
Atomic force microscopy (AFM)
AFM images were obtained using Bruker Dimension Edge (tapping mode). The scanning area for each sample is 30 µm × 30 µm, and the data analysis was made using the Gwyddion software.
Contact angle
Contact angle measurements were obtained using the Drop Shape Analyser – DSA25, KRISS GmbH. The measurements were taken in the lab air atmosphere using distilled water.
Profilometry
The thickness measurements have been conducted using the Dektak XT (Bruker). The operating mode is Hills and Valleys.
Transmission electron microscopy (TEM)
TEM, High-Angle Annular Dark-Field, and Dark-Field Scanning Transmission Electron Microscopy (HAADF-STEM and DF-STEM, respectively) images were obtained using a Thermo-FEI Talos F200i.
Raman spectroscopy
Raman Spectroscopy measurements were obtained using Horiba Xplora Plus. Raman laser 532 nm and filter 25%.
Distribution of relaxation times (DRT) analysis
DRT analysis was performed in MATLAB R2023b using the DRTTools script.70 The regularisation parameter was set to 0.01, and the radial basis function was computed using the Full-Width-at-Half-Maximum (FWHM) with a control parameter of 0.3.
Results and discussion
TiOx films: materials characterisation
A Kelvin Probe set-up was used to measure the work function WF of TiOx films deposited on the Ag electrodes because the WF is related to its stoichiometry and the surface potential.71 The results are shown in Fig. 2a. The work function's mean value is 4.54 ± 0.01 eV, which is close to the work function values associated with the TiO2 anatase phase reported in the literature.71,72 However, the e-beam evaporation is a relatively low-energy deposition method, and combined with the fact that there is no substrate heating,73–75 this result indicates that the TiOx films are amorphous, which is beneficial for sensing because it suggests a larger specific surface area.76
 |
| | Fig. 2 Characterisation methods for the TiOx films. (a) Kelvin Probe (5 measurements), (b) Tauc Plot, (c) XPS in the area of Ti-related peaks and (d) XPS around the O2-related peaks. All characterisations lead to the conclusion that TiOx films are amorphous. | |
UV-Vis absorption measurements were used to determine the optical bandgap of the TiOx films, whilst the Tauc plot was used to extract the indirect allowed bandgap for the TiOx films. The results are presented in Fig. 2b. The measured optical bandgap of 3.33 eV is slightly higher than the anatase's theoretical value (3.2 eV);77 however, it matches the TiOx bandgap values documented in the literature.78,79 This deviation might be related to the amorphous nature or surface oxidation of the TiOx films. To further prove that the TiOx films are amorphous, TEM and Raman spectra are presented (see Fig. S1). It can be observed that the TiOx films do not have any obvious crystalline morphology, hence the characteristic Raman peaks either do not exist, or they are very broad with very small intensity.80,81
XPS measurements were performed to investigate the chemical structure of the TiOx films. The results for the Ti2p area and O1s are presented in Fig. 2c and d, respectively. It can be observed that two peaks at 458.6 (Full-width-at-half-maximum FWHM 1.3 eV) and 464.2 eV are dominant, which are related to 2p3/2 and 2p1/2 orbitals, respectively.82 Their binding energies and their difference (5.6 eV) – spin–orbit splitting between the doublet peaks – are characteristic of the Ti4+ oxidation state.83 The peaks around 470.8 and 476.5 eV are Ti2p satellite peaks from the TiOx films. In Fig. 2d, there are two peaks, one at 529.2 eV, which is related to the Ti-O bond (lattice oxygen) and another one at 531 eV, related to the –OH groups that are usually present in the TiOx surface.84
Sensor response: linearity, sensing mechanism insights and sensing layer thickness independence
The EIS results for the TiOx and PANI sensors (Fig. 1) in the pH range 4–8 are shown in Fig. 3a–d. The amplitude Bode plots (impedance vs. frequency) are presented to study the impedance response of each pH buffer solution on the sensors. The frequency of 100 kHz was selected to study impedance amplitude variations because it is relatively far from the effects of ionic phenomena,48 and it also falls within the plateau region48 for every pH in the range of 4–8. The TiOx pH sensors show a sensitivity of −27.8 Ω pH−1, while the PANI sensors report a sensitivity of −22.15 Ω pH−1, respectively (see Fig. 2b and 2d). This sensitivity difference can be correlated with the PANI's doping level, which is kept at 2% wt. The doping selection has been made during the annealing (see Methods) and the electrochemical characterisation steps. The reason is the minimisation of any HCl-related reaction issues on the Ag surface (conversion to AgCl).85 The 100 kHz frequency lies within the Bode plateau range, which corresponds to a resistive-double-layer phenomenon, as indicated by the semicircular region in the Nyquist plot (see Fig. S2). This implies that any ionic motion processes – reflected by the slope at low frequencies in Fig. 2b and the linear part of the Nyquist plot in Fig. S2 – have been saturated.86 Thus, the measured pH sensitivity at 100 kHz can be attributed directly to interfacial phenomena in TiOx films.
 |
| | Fig. 3 EIS measurements for both TiOx and PANI sensors in the pH range 4–8. (a) Bode plots for TiOx sensor, (b) TiOx sensor's impedance linear decrease (−27.8 Ω pH−1) at 100 kHz, (c) Bode plots for the PANI sensors and (d) PANI sensor's impedance linear decrease (−22.15 Ω pH−1) at 100 kHz. For subfigures (c) and (d), the black dots represent to the average values from n = 7 devices. | |
The TiOx pH interfacial sensing mechanism is based on the semiconductor–electrolyte interface model.87,88 In this model, the Fermi level of the sensing layer aligns with the electrochemical potential of the buffer solution, which creates a built-in (flat band) potential. However, because McIlvaine buffer is a non-redox electrolyte, the flat band potential is created by the protonation/deprotonation of the hydroxyl groups (–OH) onto the TiOx surface (rather than from the redox species in the traditional models89). According to the literature,90 the TiOx has two acid dissociation constants at pH ∼2–5 and 8–9, respectively (which determine the charge net dipole moment and the net surface charge on the TiOx surface), while the isoelectric point is around pH values 5.6–6.0. The pH range 5–8 lies between the two dissociation constants, hence there is linearity in the frequency range around 100 kHz. On the other hand, the small deviation of the pH 4 case from the linearity can be explained by the fact that this value is closer to a dissociation constant90 (see Fig. 3a). Finally, we conclude that the sensing mechanism of the TiOx film is based on double-layer changes due to protonation/deprotonation, which alter the surface potential of the TiOx film. Following, it can be concluded that studying the pH sensor in the semicircle frequency range, under this sensing mechanism, complies with the Nernst16 and Bergveld's43 surface sensing model. This means that measuring the TiOx impedance sensitivity at 100 kHz is related directly to the surface potential changes (mV pH−1) for different pH values.
Therefore, the –OH groups existence onto the TiOx surface needs to be proven for efficient pH sensing, according to the model described above. The contact angle of water onto the TiOx surface is found to be 60.2 degrees (see Fig. S3), which agrees with the literature.91–93 This contact angle value proves surface hydrophilicity94 and is related to the –OH groups on the TiOx surface.92 The contact angle (Fig. 2), along with the 531 eV peak from XPS measurements (Fig. 2d), proves the existence of- OH groups on the TiOx surface.
The PANI sensing mechanism, on the other hand, is based on the protonation of its polymeric chain,95 it also leads to limited chemical stability as cycling and environmental exposure lead to dopant leaching, oxidation, structural degradation, etc. As pH increases from 4 to 8, hydrogen doping of the conductive PANI-ES chains decreases, until PANI-ES transforms into non-conductive PANI-EB.96 In this work, HCl doping was used to keep the PANI chains doped; hence, as the H+ concentration decreases, when pH increases, the detection could still be feasible due to the doping-induced conductivity.
The next step is to establish the surface reversibility during pH measurements. The reversibility in the surface of the TiOx films implies that there are no surface chemistry changes in the McIlvaine buffer – TiOx interface. EIS measurements have been obtained in the pH range 4–8, but after every pH value testing, the pH EIS response has been captured at pH 7 (neutral pH value). The results are presented in Fig. S4. The linearity remained the same at −27.6 Ω pH−1, and the responses at 100 kHz for pH 7 after each buffer solution remained stable. These results indicate that the TiOx films undergo reversible surface hydrolysis, according to the protonation-sensing model described above.
EIS was also performed by swapping the Gamry WE and CE probes in the sensor platform to study the sensor's robustness and any Ag parasitic contribution. The results are presented in Fig. S5. The Bode plots and the sensitivity remain close to the results from Fig. 3a (median sensitivity is −27.33 Ω pH−1). This result indicates that, despite the Ag utilisation as a CE, the pH sensing is attributed to the TiOx film, and any Ag parasitic contribution is either minimal or negligible. This occurs because the TiOx has greater impedance than Ag, hence the TiOx is the dominant sensing-determining layer and is supported by the swapping electrodes experiment. Kramers-Kroning96 analysis has been executed for pH 4–8 EIS spectra. The results can be found in Fig. 4a. It can be seen that the data are four to five orders of magnitude less than the impedance values and are randomly distributed around zero, showing no upward or downward trend – strong evidence that the TiOx sensor is linear and stable.
 |
| | Fig. 4 (a) Kramers-Kroning analysis for the TiOx sensor for pH 4–8. No clear tendency signifies the success of this test. (b) The selectivity (20–25% difference from the pH 8 impedance) of the TiOx sensor against chemicals can be found in a wound site. For the subFig. (b), the colour dots correspond to the average values of n = 7 devices. | |
The selectivity of the TiOx sensor has also been examined against different chemical compounds that can be found in a wound site (i.e. CaCl2,97 KCl,98 NaCl,99 NH4Cl100) with results presented in Fig. 4b. At 100 kHz, their impedance response is markedly lower than that of the pH 8 McIlvaine buffer. This is because the different compounds are simple salt solutions, whereas the McIlvaine buffer contains multivalent ions. The resulting increase in the ionic strength reduces the plateau impedance, consistent with the accelerated phenomena on the TiOx interface.90 Their impedance is 20% (for the NaCl solution) to 25% (for the CaCl2 solution) lower than that of the pH 8 buffer solution. In other words, the TiOx pH sensor impedance response to the pH range 4–8 is highly separable from the impedance response of the other salt solutions. These results indicate a selectivity property of the TiOx films for pH biosensing. Given that ammonium-related substances are related to bacterial inflammation on the wound site (pH gets alkaline),100 the NH4Cl solution's impedance difference from the pH 8 buffer solution gives a potential for an accurate detection of bacterial infection at the wound site.
Furthermore, EIS was performed for four different TiOx thicknesses (10, 25, 40 and 90 nm-please see Fig. S6) to prove experimentally that the frequency range chosen for sensitivity measurements is related to interfacial phenomena only. The results in Fig. 5a and b show that for either the pH upward (from 4 to 8) or downward testing (from 8 to 4), for all the TiOx film thicknesses, the pH sensitivity is stable, proving the interfacial nature of the TiOx films' pH sensing, in the plateau range's 100 kHz frequency. These results indicate that the surface hydrolysis is reversible and thus, the sensitivity is minimally affected by the TiOx thickness. Additionally, while the EIS ionic component exhibited linear behaviour, its sensitivity (slope) depended on the TiOx thickness (see Fig. S7). This observation can be attributed to the fact that, in these low frequencies, the double layer and the surface reactions described above do not produce any impedance signal. This is because electron kinetics and double layer are fast processes, taking µs to ms, hence their effect on the impedance signal is visible in higher frequencies. Hence, any contribution from the film resistance becomes the dominant factor since the proton-transfer effects are present in the transport-limiting regime. Consequently, the rate-determining factor shifts from the interfacial double-layer to the ionic diffusion to the electrode and film resistance, leading to a variable sensitivity in this frequency range.
 |
| | Fig. 5 The sensitivity of the pH sensors across different TiOx thicknesses when (a) pH tested from 4 to 8 and (b) pH tested from 8 to 4, in the same device. The sensitivity remained constant across different TiOx films’ thicknesses. For both the subfigures, the boxplots correspond to the measurements of n = 7 devices. | |
pH detection at a 0.5-fold diluted McIlvaine buffer solution
The EIS response of the TiOx sensors has been examined under 0.5-fold diluted McIlvaine buffer solutions. The reason behind this experiment is to test the TiOx sensor's pH sensitivity limits (proton detection). The results are presented in Fig. 6a. It can be observed that the sensitivity retained 96.9% of its value relative to the pristine buffer solution (see Fig. 3a). This indicates that the sensor maintained its interfacial pH sensitivity with lower H+ concentration. Hence, the TiOx sensor has the capability of accurate pH detection in lower concentration aquatic media. The sensitivity's consistency under 0.5-fold dilution demonstrates that the TiOx sensors have a potential for operating in environments with minimum proton concentrations.
 |
| | Fig. 6 (a) The sensitivity of the TiOx sensors for the pristine vs. 0.5-fold diluted McIlvaine buffer solution. The TiOx sensors maintained 96.9% of their sensitivity vs. the pristine buffer solutions. (b) The sensitivity of the TiOx fresh vs. thermal annealed pH sensors. A 95.7% sensitivity recovered after the thermal annealing treatment. For both the subfigures, the boxplots correspond to the measurements of n = 7 devices. | |
Recyclability – recovery after thermal annealing
The sensor's response was examined after a period of approximately 3 months (stored in cleanroom conditions) after fabrication to investigate its long-term functionality (see Fig. 6b). Given the chemical degradation and the one-month timeframe used to classify a wound as chronic, it is pertinent to investigate whether the TiOx sensor can recover its sensitivity after long periods (2–3 months). After three months of fabrication, there was no linearity in the pH range 4–8. After thermal annealing at 200 °C for 2 hours, the sensor sensitivity has been retained at 95.7% (−27.1 Ω pH−1 median value) of the sensitivity of the fresh TiOx devices (−28.3 Ω pH−1 median value). Before and after annealing, AFM was used to investigate any morphological surface modifications to the Ag thin film electrodes, with the results shown in Fig. S8. The surface roughness increased from 2 nm to 6.4 nm RMS without any indication of island formation. At 200 °C, the Ag tends to form larger grains, resulting in increased surface roughness.101 This demonstrates that the TiOx sensor can have an extended lifetime, which reduces the waste and maintenance costs.
Time stability
The performance of the TiOx and PANI pH sensors was tested for 31 days (approximately the time required to determine whether a wound is chronic). In the first test type of stability (see Methods - impedance value vs. time for pH 4 and 8, respectively), the impedance at 100 kHz was tested using only pH 4 and pH 8 buffer solutions every time, for both TiOx and PANI sensors. Regarding the pH 4-related time stability (see Fig. 7a), after 31 days, TiOx sensors showed an impedance increase of 5.65% (from 424.2 Ω to 448.2 Ω median values) while the PANI impedance increased 10.55% (from 393.2 Ω to 434.7 Ω median values). PANI chain is more prone to oxidation than TiOx, which is already in an oxidised state, as a metal oxide, which leads to chemical degradation, after consistent exposure to pH 4.
 |
| | Fig. 7 The one-month time stability results for both the TiOx and PANI sensors. (a) Measuring Impedance at 100 kHz with pH 4 buffer solution only. TiOx sensors report 5.65% increase while the PANI 10.55%. (b) Measuring Impedance at 100 kHz with pH 8 buffer solution only. TiOx sensors report 3.8% increase while the PANI 9.46%. (c) Measuring the sensitivity over time. The TiOx sensors lost only 2.25% of their sensitivity while the PANI lost 10% of its sensitivity. These percentages are calculated using the median values from the boxplots. For all the subfigures, the boxplots correspond to the measurements of n = 5 devices. | |
For pH 8, the time stability results showed that the impedance of the TiOx sensors increased (see Fig. 7b) by 3.8% (from 316.3 Ω to 328.4 Ω median values) while the impedance of the PANI sensors increased by 9.46% (from 305.4 Ω to 334.3 Ω median values). The time stability at pH 8 of TiOx sensors is expected to be higher than that of PANI at pH 8, since PANI undergoes a transition from the ES to EB form. At this stage, its conductivity can only be maintained through HCl doping.102 The consistent exposure of PANI at pH 8 probably leads to a partial degree of dedoping near the surface. Generally, lower degradation in the pH 8 experiments relative to the pH 4 experiments is expected because higher acidity leads to a higher concentration of H +, which promotes irreversible hydrolysis, backbone degradation (especially for PANI), and dissolution of atoms (especially for amorphous TiOx). The results for the second test of stability (see Methods - impedance slope at 100 kHz vs. time) are presented in Fig. 7c. After 31 days, the TiOx pH sensors lost only 2.25% of their sensitivity (from −27.45 Ω pH−1 to −26.86 Ω pH−1 median values) while the PANI pH sensors lost ∼10% of their sensitivity (from −22.21 Ω pH−1 to −20.05 Ω pH−1 median values). The reason behind this difference, except for the HCl partial dedoping, is that PANI is also more prone to degradation in air and humidity than TiOx, as an organic polymer.102 Finally, statistical tests, at 95% significance level, were performed on the datasets above (please see Statistical Analysis Section in the SI). The null hypothesis testing (2-sample t-test) will provide information for the statistical significance between the mean values of every dataset, or if any variations arise from random noise. Rejection of the null hypothesis (h = 1, p < 0.05) will be used to confirm statistically significant differences in the EIS pH sensor responses. The EIS data from Fig. 3b and d are statistically significant, hence the EIS data are realistic, highlighting the interfacial capacitive-resistive phenomena on the sensors’ surfaces. The data from Fig. 5a, b and 6a show little statistical significance, which is beneficial for our sensor because they cross-validate that the sensitivity of the TiOx film is thickness-independent. The dilution of the buffer solution causes a minimum effect on the TiOx sensor performance, as well. The data from recovery after annealing and 31 day stability (Fig. 6b and 7a–c) show statistical significance at 95% level. This means that these EIS data are reliable (not statistical noise) and cross-validate the greater TiOx pH sensor time stability performance versus the PANI pH sensors.
We used XPS to investigate potential chemical structural changes at the TiOx film surface after 31 days of stability (continuous exposure to pH 4 and pH 8 buffer solutions). The results for the Ti2p and O1s areas are presented in Fig. 8. In Fig. 8a and c, the Ti2p peak positions are for pH 4, at 458.8 eV (FWHM 1.3 eV) and 464.5 eV and for pH 8, at 458.7 eV (FWHM 1.3 eV) and 464 eV. The Ti2p3/2 peak position and FWHM remain relatively stable, close to those reported for the fresh device (Fig. 2c), indicating that any change in the Ti oxidation state (such as oxygen deficiency) is negligible. In Fig. 8b and d, a shift in the O1s peak can be detected from 529.2 eV to 531.3 eV. This can be attributed to the fact that after one-month experiments, there is water adsorbed onto the TiOx surface (hydroxylated surface) because McIlvaine buffer is a water-based buffer solution, but the stoichiometry is maintained. The observed oxygen peak shift can explain the recyclability in Fig. 6b. The identified hydroxylated species and residual buffer compounds adsorbed on the TiOx surface, following the pH measurements and ageing. The efficacy of the thermal annealing treatment for the regeneration is well-documented, as it removes such humidity-related contaminants from metal oxide surfaces.103–105
 |
| | Fig. 8 XPS measurements, for the TiOx sensors, after the one-month stability at measuring impedance only with (a) and (b) pH4 and (c) and (d) pH 8 buffer solution, respectively. The Ti 2p3/2 (∼458.8 eV) peak position remain constant while the –OH related peak (∼531 eV) increased vs. the one in Fig. 2. This means that the stichoimetry of the film mentained but water chemisorbed on the surface after the stability experiments. | |
The series resistance after 31 days has been studied as well (the real axis’ shift at the Nyquist plot) to investigate any degradation of the Ag CE, which can distort the impedance signal. For the 2-electrode TiOx sensor, the Nyquist plot's real axis shift can be correlated with the electrodes' (series) resistance and thus, its degradation106 when it is increased. In EIS measurements, the series resistance can be found from the curve shift on the real axis. The results for the first type of time stability, for both TiOx and PANI pH sensors, are presented in Fig. 9 and Fig. S9. For the TiOx sensor, the series resistance increase is 2.3% (from 101.84 Ω to 104.2 Ω) for the pH 4 case, and 1.7% (from 109.54 Ω to 111.44 Ω) for the pH 8 case, respectively. For the PANI sensor, the series resistance increase is 1.94% (from 101.6 Ω to 103.58 Ω) for the pH 4 case, and 1.5% (from 108.96 Ω to 110.66 Ω) for the pH 8 case, respectively. In Fig. S9, the Nyquist plots for the pH 4 and pH 8 diluted buffer solutions are presented. The shift on the real axis is 100 Ω and 109 Ω for the pH 4 and pH 8, respectively. These results indicate that the Ag electrode undergoes minimal corrosion during EIS measurements. Thus, the sensor's capability for long-term stability assessment, using Ag as an electrode material, has been demonstrated. This can be attributed to the low-voltage, low-current nature of EIS, which generates only a weak electric field in the electrochemical cell. As a result, the ions lack sufficient energy to undergo parasitic reactions with either the sensing films or the metal electrodes. Finally, the above results (Fig. 9), along with the electrode swapping and the Kramers-Kroning analysis (Fig. 4), indicate that Ag can be used as an electrode material for pH sensing, further highlighting the TiOx sensor as a low-cost, sustainable pH sensor.
 |
| | Fig. 9 Electrode's degradation for pH 4 and pH 8- stability experiments, for (a) the TiOx sensors and (b) the PANI sensors. An Ag electrode degradation less than 2.3% is reported for all cases. | |
DRT analysis
The DRT analysis is an established model-free method used frequently to study electrochemical systems and analyse EIS data without any use of equivalent circuits and a priori knowledge of the processes that take place. Especially in this paper, where a 2-electrode system is used, the equivalent circuit more complex than the Randles circuit. In this paper, DRT is utilised to validate that the pH sensor's response is based on the interface phenomena.
The results for both TiOx and PANI are presented in Fig. S10. Both figures exhibit two characteristic peaks in the distribution: one around 10−6 s (fast process) and one at 5 × 10−3 sec (slow process). These peaks are related to resistive-capacitive phenomena and diffusion processes, respectively.55 Briefly, these figures show the contribution of each electrochemical process's impedance to the total cell impedance. The intensity of the fast process (<500 lnΩ) is always lower than that of the slow process (>1500 lnΩ). This observation can be correlated with Bode plots’ frequency range 100 Hz–1 MHz (see Fig. 2) and the curve's downward bending (due to the pole) after 100 kHz. Thus, this downward bending in the Bode plots exists in a small frequency range (>100 kHz to 1 MHz) relative to the whole frequency range (100 Hz to 1 MHz). The impedance in this area is almost one order of magnitude lower than the impedance in the diffusion section of the spectra (<∼10 kHz). Moreover, the diffusion-related area has two overlapping peaks, probably because the extrapolation or overlapped ionic processes cannot be detected in the Nyquist plot (because the slope at the Nyquist plot remains constant, see Fig. S2).
In Fig. 10a and b, the peak values in the area related to the resistive-capacitive phenomena are shown. Both TiOx and PANI show a linear decrease (−49.7 lnΩ pH−1 and –46.5 lnΩ pH−1 for the TiOx and PANI pH sensors, respectively) in peak height. This linearity is a strong indicator that the pH sensing, both for the TiOx and PANI, is an interfacial phenomenon related to resistive-capacitive phenomena. This observation agrees with the literature on pH models for Transition Metal Oxides and surface potential, as described in a previous section. Overall, with no a priori assumptions, DRT analysis shows a linear response in the relaxation time area related to interfacial phenomena, validating the analysis in the section above and the selection of 100 kHz for studying the sensor sensitivity. An equivalent circuit (transfer function) analysis has been executed to interpret the DRT analysis data (please see Fig. S11, Equivalent circuit analysis section – SI). Two district time-constant values have been derived and can be related to diffusion and interfacial phenomena in the pH range 4–8. DRT and equivalent circuit analysis are partially complementary; however, DRT's continuous γ function provides a more detailed presentation of (overlapping) relaxation processes and their characteristic time constants (Fig. 10 and Fig. S10).
 |
| | Fig. 10 DRT data in the region related to the resistive-capacitive phenomena for (a) the TiOx and (b) PANI pH sensors. The peak values show a linear decreasing trend, as pH increases, which validates the interfacial behaviour of the pH sensing. Insets: the DRT curves for each pH sensor in the area related to the resistive-capacitive phenomena. | |
Furthermore, DRT analysis was used on the EIS results from the first type of stability experiments to identify any changes in the distribution of relaxation times. The results are shown in Fig. 11a–d for TiOx and PANI, respectively (please find the equivalent circuit analysis and the time-constant calculation in the relevant section of the SI). For both TiOx and PANI, it can be observed that the diffusion-related area is increasing (its relaxation time becomes more dominant) after degradation. This can be explained by the fact that the interface is degraded (water chemisorption, see Fig. 8), hence the interaction (sensing mechanism) between the buffer solution and the –OH groups of the TiOx films is diminished. To cross-validate these findings, the Bode plots for both pH 4 and 8 are presented in Fig. S10. The correlation between these results (Fig. 11 and Fig. S12) has also been seen in another TMO-based work.40 For TiOx, for both pH 4 and 8, the low-frequency line gradually shifts progressively toward higher frequencies. This translates to the fact that diffusion becomes the dominant process in this frequency spectrum.106 This, in turn, has a substantial impact on the interfacial phenomena such as the double layer and charge transfer (which, in our study, affect the sensitivity of the pH sensor). For PANI, for both pH 4 and 8, there is again a low-frequency line shift towards the high frequencies. It can be noticed that for pH 4, this shift is almost one order of magnitude, marking the reason for such a high peak increase in Fig. 11c and its impact on the PANI surface and its corresponding drop in sensitivity (Fig. 7c).
 |
| | Fig. 11 DRT analysis for the pH 4 and pH 8 degradation studies for (a–b) TiOx and (c–d) PANI pH sensors, respectively. An increase over time in the area of ion-related motion phenomena is observed due to surface contamination. | |
Moreover, the effect of the adsorbed water (to explain the observations from Fig. 11) on the impedance of the TiOx film, at 100 kHz, has been examined. For this purpose, the HAADF-STEM and DF-STEM of the TiOx film's surface are provided in Fig. S13. These images show non-uniform densities of the order of nanometres (where HAADF contrast is proportional to density thickness and Z2, and Z is taken constant). Adsorbed water can affect the porosity of the TiOx film and potentially influence the interfacial sensing phenomenon. Here, as porosity, we define the defect states in the amorphous TiOx film, ranging from oxygen vacancies as trap states - their potential contribution to the interfacial capacitive-resistive phenomena - (see Fig. 3, 10, 11 and Fig. S10) to nano-metre scale non-uniformities (STEM images, Fig. S13), and to adsorbed water species (see XPS data, Fig. 8b,d). Hence, the approach of the constant-phase element (CPE)-like phase parameter α is used.107,108 The results are presented in Fig. S14. The position and the magnitude of α remain nearly unchanged across the sensor states (fresh, degraded, and annealed). The EIS 100 kHz working frequency lies close to the minimum for both pH 4 and 8, where the slope is minimal. This means that the TiOx pH sensor response at 100 kHz is relatively insensitive to interfacial variations, including effects on film porosity. Consequently, even though adsorbed species affect diffusion-related processes (Fig. 11), their effect on the sensitivity remains limited. This interpretation is consistent with our experimental observations, including the DRT analysis (Fig. 11a, b), and the sensitivity retention and recovery values (97.75% and 95.7%, respectively; Fig. 7c and 6b).
Finally, after presenting the stability results (Fig. 7) and validating them with DRT analysis (Fig. 11), it can be concluded that the proposed low-cost TiOx pH sensor has a competitive operational lifetime compared with the high-cost pH sensors reported in the literature (please see Table S1). Also, the stability protocol presented in this paper examines both sensitivity and edge values over time. This gives a broader knowledge of the pH sensor lifetime, and thus, its translation potential to a smart bandage.
Conclusion
In this paper, an impedimetric TiOx pH sensor has been presented. The sensor exhibited linearity of –27.8 Ω pH−1 across the clinically relevant pH range of 4–8, which is critical for chronic wound monitoring. The TiOx pH sensor's time (chemical) stability has been examined versus PANI. After one month of operation—the period for defining wound chronicity—the TiOx sensor lost only 2.25% of its initial sensitivity. In contrast, the state-of-the-art material PANI showed a 10% loss in sensitivity under the same conditions. In pH 4 and 8 studies, the PANI sensor showed an impedance increase (10% and 9% increase for pH 4 and 8, respectively), almost double that of the TiOx sensor (5.5% and 4% increase for pH 4 and 8, respectively). The stability studies highlight TiOx's chemical interfacial stability after continuous exposure to acidic and alkaline media, compared with PANI. Analysis from both Bode and Nyquist plots revealed the interfacial nature of the pH sensing, as well as its reversibility and recyclability. The TiOx pH sensor recovered 95.7% of its sensitivity after 200 °C thermal annealing, further demonstrating its robustness. These results underscore the greater demonstrated chemical stability of the lower-cost TiOx compared to PANI. Minimal Ag corrosion and interference were observed in the pH sensing, further indicating the TiOx pH sensor's suitability in low-cost and sustainable sensing applications, whereas state-of-the-art PANI-based pH sensors cannot be utilised because of their cost. Finally, DRT analysis has been utilised as a model-free approach to cross-validate the interfacial nature of sensing and the degradation mechanism, both for TiOx and PANI films. In this work, photolithography and e-beam evaporation were used to fabricate the TiOx pH sensor prototype. Fabrication can also be achieved using solution-processed methods (such as spray coating for the TiOx) and printed electronics methods (such as inkjet or screen printing for the Ag electrodes). The future direction of this work is to test the prototype TiOx sensor's performance in complex biological environments with bacterial infection and to integrate it into a smart bandage. Towards clinical translation, the prototype TiOx pH sensor should be tested using cytotoxicity assays, such as the MTT assay, to evaluate the sensor's biocompatibility. This work provides an important demonstration towards long-term and low-cost pH biosensing. Hence, a pathway to more robust biosensor data is an optimistic indicator for the deployment of smart bandages to real-world applications.
Author contributions
Conceptualisation: K. A., A. P. and G. K. Discussion, literature analysis: K. A., A. P., G. K, S. R. P. S. and V. S. K. A. designed, fabricated, characterised the devices, conducted EIS experiments, DRT analysis and analysed the data. A. P. K. A. conducted UV-vis and contact angle measurements. A. P. conducted AFM measurements. L. A. conducted Kelvin Probe measurements. X. P. prepared the schematics. D. I. K. prepared the evaporation masks. S. J. H. conducted XPS measurements and the first peak analysis. K. A. conducted the Raman spectroscopy measurements. V. S. executed the TEM, HAADF-STEM, and DF-STEM imaging. First Draft writing: K. A. Review and editing: A. P., G. K, S. R. P. S. and V. S. supervision: V. S. and S. R. P. S. Funding acquisition: V. S. and S. R. P. S. All authors have read and agreed to the submission of the manuscript.
Conflicts of interest
There are no conflicts to declare.
Data availability
The data supporting this article have been included as part of the supplementary information (SI). Supplementary information is available. See DOI: https://doi.org/10.1039/d5tb02697k.
Acknowledgements
For open access purposes, the authors have applied a Creative Commons attribution license (CC BY) to any Author-Accepted Manuscript version arising from this submission. The authors would like to extend their appreciation to the University of Surrey for the doctoral scholarship of K.A.
References
- R. G. Frykberg and J. Banks, Adv. Wound Care, 2015, 4(9), 560–582 CrossRef PubMed.
- A. K. S. Iversen, M. Lichtenberg, B. G. Fritz, I. D.-P. Cort, D. F. Al-Zoubaidi, H. Gottlieb, K. Kirketerp-Møller, T. Bjarnsholt and T. H. Jakobsen, BMJ Open, 2024, 14, e084081 CrossRef PubMed.
- A. Horta-Velázquez, J. D. Mota-Morales and E. Morales-Narváez, Int. J. Biol. Macromol., 2024, 254(1), 127737 CrossRef PubMed.
- H. Derakhshandeh, S. S. Kashaf, F. Aghabaglou, I. O. Ghanavati and A. Tamayol, Trends Biotechnol., 2018, 36(12), 1259–1274 CrossRef CAS PubMed.
- N. Brasier, J. Wang, W. Gao, J. R. Sempionatto, C. Dincer, H. C. Ates, F. Güder, S. Olenik, I. Schauwecker, D. Schaffarczyk, E. Vayena, N. Ritz, M. Weisser, S. Mtenga, R. Ghaffari, J. A. Rogers and J. Goldhahn, Nature, 2024, 636, 57–68 CrossRef CAS PubMed.
- S. Cheng, Z. Gu, L. Zhou, M. Hao, H. An, K. Song, X. Wu, K. Zhang, Z. Zhao, Y. Dong and Y. Wen, Front. Bioeng. Biotechnol., 2021, 9, 765987 CrossRef PubMed.
- C. Yang, C. Yang, Y. Chen, J. Liu, Z. Liu and H.-J. Chen, J. Sci.: Adv. Mater. Devices, 2023, 8(4), 100619 CAS.
- C. Wang, E. Shirzaei Sani, C. D. Shih, C. T. Lim, J. Wang, D. G. Armstrong and W. Gao, Nat. Rev. Mater., 2024, 9, 550–566 CrossRef PubMed.
- F. Wang, Y. Xie, W. Zhu, T. Wei, F. Wang, Y. Xie, W. Zhu and T. Wei, Chemosensors, 2023, 11(9), 481 CrossRef CAS.
- A. Barhoum, S. Hamimed, H. Slimi, A. Othmani, F. M. Abdel-Haleem and M. Bechelany, Trends Environ. Anal. Chem., 2023, 38, e00199 CrossRef CAS.
- J. Wu, H. Liu, W. Chen, B. Ma, H. Ju, J. Wu, H. Liu, W. Chen, B. Ma and H. Ju, Nat. Rev. Bioeng., 2023, 1, 346–360 CrossRef CAS PubMed.
- N. Mehmood, A. Hariz, R. Fitridge and N. H. Voelcker, J. Biomed. Mater. Res., Part B, 2014, 102(4), 885–895 CrossRef PubMed.
- C. Tate, R. Wang, S. Akaraci, C. Burns, L. Garcia, M. Clarke and R. Hunter, Cities, 2024, 145, 104706 CrossRef.
- L. A. Schneider, A. Korber, S. Grabbe, J. Dissemond, L. A. Schneider, A. Korber, S. Grabbe and J. Dissemond, Arch. Dermatol. Res., 2006, 298(9), 413–420 CrossRef PubMed.
- E. Shirzaei Sani, C. Xu, C. Wang, Y. Song, J. Min, J. Tu, S. A. Solomon, J. Li, J. L. Banks, D. G. Armstrong and W. Gao, Sci. Adv., 2023, 9(12), eadf7388 CrossRef PubMed.
- H. J. Park, J. H. Yoon, K. G. Lee and B. G. Choi, Nano Convergence, 2019, 6, 9 CrossRef PubMed.
- G. P. Neupane, W. Ma, T. Yildirim, Y. Tang, L. Zhang and Y. Lu, Nano Mater. Sci., 2019, 1(4), 246–259 CrossRef.
- Z. Wang, X. Wu, S. Zhang, S. Yang, P. Gao, P. Huang, Y. Xiao, X. Shen, X. Yao, D. Zeng, J. Jie, Y. Zhou, F. Yang, R. Li and W. Hu, Proc. Natl. Acad. Sci. U. S. A., 2025, 122(14), e2419673122 CrossRef CAS PubMed.
- C. G. Tang, K. Hou and W. L. Leong, Chem. Mater., 2023, 36(1), 28–53 CrossRef.
- M. S. Hossain, N. Padmanathan, M. M. R. Badal, K. M. Razeeb and M. Jamal, ACS Omega, 2024, 9(38), 40122–40133 CAS.
- J. Min, S. Demchyshyn, J. R. Sempionatto, Y. Song, B. Hailegnaw, C. Xu, Y. Yang, S. Solomon, C. Putz, L. E. Lehner, J. F. Schwarz, C. Schwarzinger, M. C. Scharber, E. Shirzaei Sani, M. Kaltenbrunner and W. Gao, Nat. Electron., 2023, 6(8), 630–641 CrossRef CAS PubMed.
- X. Zhu, H. Sun, B. Yu, L. Xu, H. Xiao, Z. Fu, T. Gao and X. Yang, RSC Adv., 2024, 14(8), 5627–5637 RSC.
- J. Lee, L. M. Liberty, I. Soltis, K. Kwon, D. Chong, Y. Kwon and W.-H. Yeo, ACS Appl. Mater. Interfaces, 2025, 17(25), 36345–36355 CrossRef CAS PubMed.
- Y. Ding, W. Liu, Z. Xu and Z. Duan, J. Mater. Chem., 2024, 12(31), 20317–20326 RSC.
- R. V. Mom, L. J. Falling, O. Kasian, G. Algara-Siller, D. Teschner, R. H. Crabtree, A. Knop-Gericke, K. J. J. Mayrhofer, J.-J. Velasco-Vélez and T. E. Jones, ACS Catal., 2022, 12(9), 5174–5184 CrossRef CAS.
- I. A. Ges, B. L. Ivanov, D. K. Schaffer, E. A. Lima, A. A. Werdich and F. J. Baudenbacher, Biosens. Bioelectron., 2005, 21(2), 248–256 CrossRef CAS PubMed.
- J. Chu, Y. Zhao, S.-H. Li, H.-Q. Yu, G. Liu and Y.-C. Tian, Electrochim. Acta, 2015, 152, 6–12 CrossRef CAS.
- O. Concepción and O. D. Melo, J. Phys.: Condens. Matter, 2023, 35, 143002 CrossRef PubMed.
- Y. Bai, I. Moran-Sero, F. De Angelis, J. Bisquert and P. Wang, Chem. Rev., 2014, 114(19), 10095–10130 CrossRef CAS PubMed.
- U. Diebold, Surf. Sci. Rep., 2003, 48(5–8), 229–253 Search PubMed.
- J. Wang, J. Xiong, H. Guo and Y. Wei, ChemCatChem, 2024, 16(20), e202400707 CrossRef CAS.
- V. S. Saji and C.-W. Lee, ChemSusChem, 2012, 5(7), 1146–1161 CrossRef CAS PubMed.
- M. Solanki, U. Parihar, K. Patel, V. Jain, S. S. Sharma and J. Ray, Chem. Phys. Impact, 2025, 10, 100781 CrossRef CAS.
- A. A. Shaikh, J. Bhattacharjee, P. Datta and S. Roy, Sustainable Chem. Environ., 2024, 7, 100125 CrossRef.
- A. Panagiotopoulos, G. Kakavelakis, K. Almpanidis, L. Askew, D. I. Kutsarov and S. R. P. Silva, J. Mater. Chem. A, 2025, 13(12), 8774–8789 RSC.
- J. Lee, I. Soltis, S. A. Tillery, S. H. Lee, H. Kim and W.-H. Yeo, Biosens. Bioelectron., 2024, 254, 116223 CrossRef CAS PubMed.
- E. S. Yoon, H. J. Park, M. S. Kil, J. Kim, K. G. Lee and B. G. Choi, Bull. Korean Chem. Soc., 2023, 44(6), 528–535 CrossRef CAS.
- A. K. Pathak, V. Bhardwaj, R. K. Gangwar, M. De and V. K. Singh, Opt. Commun., 2017, 386, 43–48 CrossRef CAS.
- M. J. Schöning, D. Tsarouchas, L. Beckers, J. Schubert, W. Zander, P. Kordoš and H. Lüth, Sens. Actuators, B, 1996, 35(1–3), 228–233 CrossRef.
- M. Zea, A. Moya, M. Fritsch, E. Ramon, R. Villa and G. Gabriel, ACS Appl. Mater. Interfaces, 2019, 11(16), 15160–15169 CrossRef CAS PubMed.
- J. Wang, C. Wu, N. Hu, J. Zhou, L. Du, P. Wang, J. Wang, C. Wu, N. Hu, J. Zhou, L. Du and P. Wang, Biosensors, 2012, 2(2), 127–170 CrossRef CAS PubMed.
- T. Judit, S. Abdul and V. Gyöngyi, Fundam. Nanopart., 2018, 521–551 CAS.
- P. Bergveld, Sens. Actuators, B, 2003, 88(1), 1–20 CrossRef CAS.
- S. Wang, J. Zhang, O. Gharbi, V. Vivier, M. Gao and M. Orazem, Nat. Rev. Methods Primers, 2021, 1, 41 CrossRef CAS.
- J. Wu, Chem. Rev., 2022, 122(12), 10821–10859 CrossRef CAS PubMed.
- S. Åberg, J. Electroanal. Chem., 1997, 439(1), 63–71 CrossRef.
- H. C. Lim, J.-E. Park, J.-I. Hong and I.-S. Shin, Bull. Korean Chem. Soc., 2020, 41(3), 362–365 CrossRef CAS.
- H. S. Magar, R. Y. A. Hassan, A. Mulchandani, H. S. Magar, R. Y. A. Hassan and A. Mulchandani, Sensors, 2021, 21(19), 6578 CrossRef CAS PubMed.
- A. M. Manoj and L. R. Viannie, Electroanalysis, 2023, 35(11), e202300156 CrossRef.
- A. Bhat, J. M. Amanor-Boadu and A. Guiseppi-Elie, ACS Sensors, 2020, 5(2), 500–509 CrossRef CAS PubMed.
- P. Awasthi, R. Mukherjee, S. P. O. Kare and S. Das, RSC Adv., 2016, 6(104), 102088–102095 RSC.
- A. K. Aliyana, P. Ganguly, A. Beniwal, S. K. N. Kumar and R. Dahiya, IEEE Sens. J., 2022, 22(21), 21049–21056 CAS.
- K. Ariyoshi, Z. Siroma, A. Mineshige, M. Takeno, T. Fukutsuka, T. Abe and S. Uchida, Electrochemistry, 2022, 90(10), 102007 CrossRef CAS.
- Z. Wang, Nat. Rev. Clean Technol., 2025, 1(7), 453 CrossRef.
- C. Plank, T. Rüther, L. Jahn, M. Schamel, J. P. Schmidt, F. Ciucci and M. A. Danzer, J. Power Sources, 2024, 594, 233845 CrossRef CAS.
- E. v Hauff, J. Phys. Chem. C., 2019, 123(18), 11329–11346 CrossRef.
- H. C. Ates, P. Q. Nguyen, L. Gonzalez-Macia, E. Morales-Narváez, F. Güder, J. J. Collins and C. Dincer, Nat. Rev. Mater., 2022, 7(11), 887–907 CrossRef PubMed.
- S. NajafiKhoshnoo, T. Kim, J. A. Tavares-Negrete, X. Pei, P. Das, S. W. Lee, J. Rajendran and R. Esfandyarpour, Adv. Mater. Technol., 2023, 8(8), 2201655 CrossRef CAS.
- P. Dias, S. Javimczik, M. Benevit, H. Veit and A. M. Bernardes, Waste Manage., 2016, 57, 220–225 CrossRef CAS PubMed.
- V. V. Pinto, M. J. Ferreira, R. Silva, H. A. Santos, F. Silva and C. M. Pereira, Colloids Surf., A, 2010, 364(1–3), 19–25 CrossRef CAS.
- P. Hájková, J. Matoušek and P. Antoš, Appl. Catal., B, 2014, 160–161, 51–56 CrossRef.
- G. F. Huldin, J. Huang, J. Reitemeier and K. X. Fu, Faraday Discuss., 2025, 257, 316–332 RSC.
- P. Mostafalu, A. Tamayol, R. Rahimi, M. Ochoa, A. Khalilpour, G. Kiaee, I. K. Yazdi, S. Bagherifard, M. R. Dokmeci, B. Ziaie, S. R. Sonkusale and A. Khademhosseini, Small, 2018, 14(33), 1703509 CrossRef PubMed.
- E. Marin and A. Lanzutti, Materials, 2024, 17(1), 114 CrossRef CAS PubMed.
- M. Akter, M. T. Sikder, M. M. Rahman, A. K. M. A. Ullah, K. F. B. Hossain, S. Banik, T. Hosokawa, T. Saito and M. Kurasaki, J. Adv. Res., 2018, 9, 1–16 CrossRef CAS PubMed.
- D. F. Mello, R. Trevisan, N. Rivera, N. K. Geitner, R. T. D. Giulio, M. R. Wiesner, H. Hsu-Kim and J. N. Meyer, Chem.-Biol. Interact., 2020, 315, 108868 CrossRef CAS PubMed.
- M. Ghasemi, T. Turnbull, S. Sebastian, I. Kempson, M. Ghasemi, T. Turnbull, S. Sebastian and I. Kempson, Int. J. Mol. Sci., 2021, 22(23), 12827 CrossRef CAS PubMed.
- R. E. Smith, S. Totti, D. Reid, S. M. Hingley-Wilson, E. Velliou, P. Campagnolo, N. I. Ward, J. R. Varcoe and C. Crean, Mater. Adv., 2024, 5(6), 2306–2315 RSC.
- C. Melios, A. Centeno, A. Zurutuza, V. Panchal, C. E. Giusca, S. Spencer, S. R. P. Silva and O. Kazakova, Carbon, 2016, 103, 273–280 CrossRef CAS.
- T. H. Wan, M. Saccoccio, C. Chen and F. Ciucci, Electrochim. Acta, 2015, 184, 483–499 CrossRef CAS.
- V. Mansfeldova, M. Zlamalova, H. Tarabkova, P. Janda, M. Vorokhta, L. Piliai and L. Kavan, J. Phys. Chem. C., 2021, 125(3), 1902–1912 CrossRef CAS.
- S. Kashiwaya, J. Morasch, V. Streibel, T. Toupance, W. Jaegermann and A. Klein, Surfaces, 2018, 1(1), 73–89 CrossRef.
- N. Martin, C. Rousselot, D. Rondot, F. Palmino and R. Mercier, Thin Solid Films, 1997, 300(1–2), 113–121 CrossRef CAS.
- A. S. Bakri, M. Z. Sahdan, F. Adriyanto, N. A. Raship, N. D. M. Said, S. A. Abdullah and M. S. Rahim, AIP Conf. Proc., 2017, 1788(1), 030030 CrossRef.
- Y. M. Abdulraheem, S. Ghoraishi, L. Arockia-Thai, S. K. Zachariah and M. Ghannam, Adv. Mater. Sci. Eng., 2013, 574738, 10 Search PubMed.
- T. Jia, J. Zhang, J. Wu, D. Wang, Q. Liu, Y. Qi, B. Hu, P. He, W. Pan and X. Qi, Mater. Lett., 2020, 265, 127465 CrossRef CAS.
- D. A. H. Hanaor and C. C. Sorrell, J. Mater. Sci., 2010, 46(4), 855–874 CrossRef.
- D. A. K. Ramadhani, N. Sholeha, N. N. A. Khusna, M. Diantoro, A. N. Afandi, Z. Osman and H. Pujiarti, Mater. Sci. Energy Technol., 2024, 7, 274–281 CAS.
- M. B. Poudel, C. Yu and H. J. Kim, Catalysts, 2020, 10(5), 546 CrossRef CAS.
- Q. Zhang, L. S. Ma, M. H. Shao, J. Z. Huang, M. Ding, X. L. Deng, X. Q. Wei and X. J. Xu, J. Nanomater., 2014, 14, 831752 CrossRef.
- M. A. Alsaiari, N. A. Alhemiary, A. Umar and B. E. Hayden, Ceram. Int., 2020, 46, 16310–16320 CrossRef CAS.
- L. Zhu, Q. Lu, L. Lv, Y. Wang, Y. Hu, Z. Deng, Z. Lou, Y. Hou and F. Teng, RSC Adv., 2017, 7(33), 20084–20092 RSC.
- N. Kruse and S. Chenakin, Appl. Catal., A, 2011, 391(1–2), 367–376 CrossRef CAS.
- G. K. Q. Ganharul, A. Tofanello, A. Bonadio, A. L. M. Freitas, M. T. Escote, A. S. Polo, I. L. Nantes-Cardoso and J. A. Souza, J. Mater. Sci., 2022, 57, 17363–17379 CrossRef CAS.
- D. Depla, Coatings, 2022, 12(12), 1915 CrossRef CAS.
- M. Saghafi, S. Chinnathambi and S. G. Lemay, Curr. Opin. Colloid Interface Sci., 2023, 63, 101654 CrossRef CAS.
- A. Hankin, F. E. Bedoya-Lora, J. C. Alexander, A. Regoutz and G. H. Kelsall, J. Mater. Chem. A, 2019, 7(45), 26162–26176 RSC.
- V. M. Arutyunyan, Phys.-Usp., 1989, 32(6), 521 CrossRef.
- S. Al-Hilli and M. Willander, Sensors, 2009, 9(9), 7445–7480 CrossRef CAS PubMed.
- M. Bischoff, N. Y. Kim, J. B. Joo and A. Marchioro, J. Phys. Chem. C., 2022, 13(37), 8677–8683 CAS.
- Y. Li, B. Xia and B. Jiang, J. Sol-Gel Sci. Technol., 2018, 87(1), 50–58 CrossRef CAS.
- E. Kuźmicz-Mirosław, M. Kuśmierz, K. Terpiłowski, M. Śmietana, M. Barczak and M. Staniszewska, Materials, 2022, 15(12), 4113 CrossRef PubMed.
- A. Borrás, C. López, V. Rico, F. Gracia, A. R. González-Elipe, E. Richter, G. Battiston, R. Gerbasi, N. McSporran, G. Sauthier, A. E. György and A. Figueras, J. Phys. Chem. C, 2007, 111(4), 1801–1808 CrossRef.
- K.-Y. Law, J. Phys. Chem. Lett., 2014, 5(4), 686–688 CrossRef CAS PubMed.
- A. Samadi, M. Xie, J. Li, H. Shon, C. Zheng and S. Zhao, Chem. Eng. J., 2021, 418, 129425 CrossRef CAS.
- M. Beygisangchin, S. A. Rashid, S. Shafie, A. R. Sadrolhosseini and H. N. Lim, Polymers, 2021, 13(12), 2003 CrossRef CAS PubMed.
- T. Subramaniam, M. B. Fauzi, Y. Lokanathan and J. X. Law, Int. J. Mol. Sci., 2021, 22(12), 6486 CrossRef CAS PubMed.
- J. H. Shim, J. W. Lim, B. K. Kim, S. J. Park, S. W. Kim and T. H. Choi, Archives Plastic Surgery, 2015, 42(1), 11–19 CrossRef PubMed.
- M. Holman, J. Wound Care., 2023, 32(8), 507–512 CrossRef PubMed.
- P. L. Tran, A. N. Hamood, A. De Souza, G. Schultz, B. Liesenfeld, D. Mehta and T. W. Reid, Wound Repair Regeneration: official publication of the Wound Healing Society [and] the European Tissue Repair Society, Wound Repair and Regeneration, 2015, 23(1), 74–81 CrossRef PubMed.
- J. Lv, F. Lai, L. Lin, Y. Lin, Z. Huang and R. Chen, Appl. Surf. Sci., 2007, 253(17), 7036–7040 CrossRef CAS.
- M. Beygisangchin, S. A. Rashid, S. Shafie, A. R. Sadrolhosseini and H. N. Lim, Polymers, 2021, 13(12), 2003 CrossRef CAS PubMed.
- V. T. Lukong, K. Ukoba, K. O. Yoro and T. C. Jen, Heliyon, 2022, 8(5), e09460 CrossRef CAS PubMed.
- S. Saqlain, B. J. Cha, S. Y. Kim, J. Y. Sung, M. C. Choi, H. O. Seo and Y. D. Kim, Mater. Today Commun., 2021, 26, 102119 CrossRef CAS.
- C.-M. Chen, Y.-C. Hsu and S.-J. Cherng, J. Alloys Compd., 2011, 509(3), 872–877 CrossRef CAS.
- A. C. Lazanas and M. I. Prodromidis, Acs Meas. Sci. Au, 2023, 3(3), 162–193 CrossRef CAS PubMed.
- B. Hirschorn, M. E. Orazem, B. Tribollet, V. Vivier, I. Frateur and M. Musiani, Electrochim. Acta, 2010, 55(21), 6218–6227 CrossRef CAS.
- P. Córdoba-Torres, T. J. Mesquita and R. P. Nogueira, Electrochim. Acta, 2013, 92, 323–334 CrossRef.
|
| This journal is © The Royal Society of Chemistry 2026 |
Click here to see how this site uses Cookies. View our privacy policy here.