Open Access Article
Luis Enrique Franco-Correa
a,
Luis A. Ortiz-Frade
b,
Alejandro Bravo Patiño
a,
Francisco Perez-Duran
a,
Fernando Calderón-Rico
a,
Alicia Gabriela Zamora-Avilés
a,
Martha Leticia Jiménez-González
b,
Gabriel Espinosa
c,
Daniel Durand-Herrera
a,
Ricarda Cortés-Vieyra
a and
Rosa Elvira Nuñez-Anita
*a
aFacultad de Medicina Veterinaria y Zootecnia, Posta Veterinaria y Zootecnia de la Universidad Michoacana de San Nicolás de Hidalgo, Universidad Michoacana de San Nicolás de Hidalgo, Edificio G. Km. 9.5 S/N, Carretera Morelia-Zinapécuaro, C.P. 58893, La Palma, Tarímbaro, Michoacán, Mexico. E-mail: rosa.anita@umich.mx; Fax: +52 (443) 295 8029; Tel: +52 (443) 295 8029
bCentro de Investigación y Desarrollo Tecnológico en Electroquímica S.C., Parque Tecnológico Querétaro S/N, 76700 Santiago de Querétaro, Querétaro, Mexico
cInstituto de Física y Matemáticas, Universidad Michoacana de San Nicolás de Hidalgo, Francisco J. Mújica S/N, Ciudad Universitaria, 58070 Morelia, Michoacán, Mexico
First published on 8th May 2026
Porcine reproductive and respiratory syndrome (PRRS) is one of the most economically devastating diseases that affect the swine industry worldwide. Although enzyme-linked immunosorbent assays (ELISAs) are the gold standard for analyzing antibodies against viruses, they have some limitations, such as sample transport and analysis times and cost. Rapid serological surveillance of PRRSV remains a critical challenge for swine health management. Immunosensors are feasible diagnostic alternatives, where antigen–antibody interactions can be immediately quantitatively determined. Electrochemical sensor operation is based on changes in interfacial electron transfer at the working electrode surface, which are monitored using electrochemical impedance spectroscopy (EIS). In this approach, a redox probe ([Fe(CN)6]3−/[Fe(CN)6]4−) is used to evaluate the charge-transfer resistance (Rct) at the electrode interface. Variations in analyte concentration lead to changes in the interfacial properties of the electrode, particularly after biomolecular interactions such as antibody binding. These interactions hinder electron transfer, resulting in an increase in Rct. The analytical signal is therefore based on the correlation between the change in Rct (ΔRct) and the concentration of the target analyte, enabling quantitative analysis through calibration plots. The goal was the establishment of a peptide based electrochemical immunosensor for quantifying antibodies against PRRSV. Gold electrode functionalization was characterized using EIS, the GP5-B peptide, and casein protein. Balb/c mice were immunized according to control, BSA-immunized, and GP5-B peptide-immunized groups. Control pigs were free from the porcine respiratory disease complex (PRDC) and nonvaccinated. Vaccinated pigs were injected using commercial Ingelvac vaccine according to the manufacturer's instructions, and naturally infected pigs were also included in the study. Antibodies were measured using ELISA assays and an immunosensor. The immunosensor recognized antigen–antibody interactions, with a limit of detection (LOD) of 6 ng mL−1, a repeatability of intra-assay CV of 5.05%, and a reproducibility inter-assay CV of 4.08% through batch to batch. Additionally, it achieved a sensitivity of 100%, and a specificity of at least 93%. Considering all the data, compared to immunoglobulin G detection via ELISAs, the proposed peptide-based electrochemical immunosensor is valuable for detecting antibodies against PRRSV. These results suggest the feasibility of biosensors for serological diagnosis in pigs.
PRRSV infection is commonly detected either directly by polymerase chain reaction (PCR)-based molecular assays or indirectly through serological techniques such as enzyme-linked immunosorbent assay (ELISAs). Although these methods are widely used, they often require specialized laboratory infrastructure, trained personnel, complex sample handling procedures, and long processing times, which may limit their applicability under field conditions.9,10 Despite advances in serological diagnostics, rapid, cost-effective point-of-care methods for PRRSV antibody detection remain limited.9
Electrochemical immunosensors have emerged as promising alternatives for pathogen detection due to their high sensitivity, relatively low cost, rapid response times, and compatibility with miniaturized and portable analytical devices.11 These characteristics make them particularly attractive for point-of-care testing (POCT) in veterinary diagnostics and livestock health monitoring.12,13 Several electrochemical immunosensing strategies have been reported for viral detection.
For example, a gold immunochromatographic strip coupled with an electrochemical reader was used to quantify anti-Newcastle disease virus; other systems include a label-free sensor for infectious bursal disease virus (IBDV) with a limit of detection (LOD) of ∼14 particles per mL,14 and impedance-based platforms for avian influenza A (H5N1).15 In swine diagnostics, sensors have been implemented for H1N1 influenza,16 porcine circovirus 2 (PCV2),17 and pseudorabies virus (PRV) using magnetic beads to achieve a wide linear range.18
In recent years, peptide-based biosensors have attracted increasing attention as alternative to whole-protein recognition elements.13 Synthetic peptides offer several advantages, including high chemical stability, reproducible synthesis and lower production cost.19,20 Their relatively small size facilitates controlled immobilization on electrode surfaces and promotes well-defined molecular orientation, which can enhance electron transfer efficiency and improve sensor performance.13,21 We hypothesized that a synthetic peptide derived from the GP5 protein could serve as a stable and selective recognition element for electrochemical detection of anti-PRRSV antibodies, enabling rapid serodiagnosis compared with conventional immunoassays.
In this context, we comparatively evaluated the analytical performance of the GP5-derived synthetic peptide using both conventional ELISA and an electrochemical impedimetric immunosensor platform. The proposed sensor integrates a peptide-based recognition interface with a self-assembled monolayer-modified gold electrode to enable label-free detection of anti-PRRSV antibodies. This approach addresses the growing demand for rapid, portable, and cost-effective serological tools for veterinary disease surveillance. These electrochemical platforms may contribute to improved biosecurity strategies, rapid outbreak monitoring, and enhanced animal health management in the swine industry.22
To our knowledge, this is the first electrochemical immunosensor based on a GP5-derived peptide for PRRSV antibody detection, tested in both experimental murine immunization and vaccinated/naturally infected swine.
For the GP5-B-immunized group, a volume of 120 μL was administered to each animal; the mixture comprised 30 μL of 1 mg mL−1 GP5-B and 30 μL of 1 mg mL−1 maleimide (Imject™ maleimide-activated BSA, Thermo Fisher), conjugated according to the manufacturer's specifications, and 60 μL of complete Freund's adjuvant (Sigma-Aldrich), substituting incomplete Freund's adjuvant for the complete adjuvant in booster immunizations.
The mice were subcutaneously injected using an immobilizer and without anesthesia and were kept in a temperature-controlled environment at 22 °C ± 1 °C and 50% relative humidity in addition to a 12
:
12 h light–dark cycle. The mice were fed Purina-certified rodent chow and water on demand, and their body weights were recorded every 3 d. The procedures were approved by the local ethics committee, CICUMSNH-A103-FMVZ.
Blood samples were drawn from the immunized mice by puncturing the tail vein using 0.80 mm diameter needles, while the mice were confined using an immobilizer. Blood was collected using Pasteur pipettes moistened with 60 mM ethylenediaminetetraacetate (EDTA) at pH 8. Blood samples were centrifuged at 2500 rpm for 5 min to obtain serum, which was stored frozen at −20 °C. For immunization, the mice were held down by the skin of the back and tail for subcutaneous injection. The mice were followed up over time, with blood samples drawn 45 days after immunization.
Serum samples from an immunosensor activity assay were collected at 42 days post immunization (dpi) via jugular venipuncture from a group of piglets that received a single intramuscular dose of Ingelvac–PRRSV, MLV® (2 mL containing a 50% tissue culture infectious dose (TCID50) of 104.9 PRRSV at day 0), and negative control sera were obtained from age-matched, nonimmunized piglets (n = 6 per group); all the animals were housed and allocated under the previously approved protocol CICUMSNH-A101-FMVZ, following the previously described method.7
The piglets were housed following approved animal welfare protocols and randomly allocated into treatment groups using the GraphPad random number generator. For this work, serum samples were specifically collected from the group intramuscularly immunized using the commercial vaccine Ingelvac PRRS MLV (Boehringer Ingelheim). The piglets received a single 2 mL intramuscular dose (containing live attenuated PRRSV strain ATCC VR-2332, as specified by the manufacturer) at day 0. Serum was collected at 42 dpi via jugular venipuncture. Negative control sera were obtained from age-matched, nonimmunized piglets maintained under the same housing and management conditions (n = 6 per group).
Prior to self-assembling monolayers, the electrodes were washed with ethanol and distilled water. Initially, the working electrode was modified by adding 10 μL of a 50 mM thioglycolic acid (TGA) solution in ethanol, drying the electrode at room temperature, adding another 10 μL of the TGA solution, and redrying the electrode. The entire process required approximately 15 min. Then, the electrodes were washed with distilled water. The electrode was exposed twice to the thioglycolic acid (TGA) solution with an intermediate drying step to promote controlled and reproducible self-assembled monolayer formation on the gold surface. This stepwise deposition protocol was empirically optimized and monitored by electrochemical impedance spectroscopy (EIS).
For activating the –COOH groups of the formed SAM, 0.4 M N-hydroxysuccinimide (NHS) and 0.1 M 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide (EDC) were mixed with ethanol in a 1
:
1 ratio, and 10 μL of the mixture was added to the working electrode and allowed to react for 30 min to self-assemble the monolayer. Then, the electrode was washed with ethanol and stored in PBS at pH 7.2 until use within 5 h.
For peptide immobilization, NHS was also used to prepare amine-reactive esters of carboxylate groups. Carboxylates (–COOH) were reacted with NHS in EDC, producing esters, which could then react with primary amines (–NH2), forming amide crosslinks. Although NHS is not required for EDC reactions, it substantially improves the coupling efficiency. Both NHS and EDC are soluble in aqueous and organic solvents. However, NHS activation decreases the water solubility of the modified carboxylate molecule so that it dissolves in ethanol. Additionally, because NHS esters have a half-life of 4–5 h at pH 7.2, the peptides were immobilized in this step.
| Anti-PRRSV antibodies | |||
|---|---|---|---|
| + | − | ||
| Test | + | True positive (TP) | False positive (FP) |
| ELISA positive | ELISA negative | ||
| Immunosensor positive | Immunosensor positive | ||
| Test | − | False negative (FN) | True negative (TN) |
| ELISA positive | ELISA negative | ||
| Immunosensor negative | Immunosensor negative | ||
As shown in Fig. 1c, the mice immunized with GP5-B showed an increase in antibody response. At 15 dpi, the peptide-specific IgG level rose markedly above the baseline (0 dpi) and remained high at 30 and 45 dpi. In contrast, the control mice only maintained a background-level signal throughout. Statistical analysis confirmed that at 15, 30, and 45 dpi, IgG concentrations were significantly higher than preimmune levels (p < 0.01–0.0001), indicating that GP5-B elicits a strong humoral immune response. ELISA data reveal that GP5-B immunization produces high-titer, peptide-specific antibodies by 15–45 dpi, whereas preimmune and adjuvant-only sera antibody levels did not change (the data are not shown), suggesting the suitability of GP5-B as a recognized element for antibody detection.
As a result, the generation and validation of specific anti-GP5B antibodies in the murine model provided a controlled set of serum samples with known serological profiles. These characterized sera were subsequently used as analytical inputs to evaluate the performance of the peptide-based electrochemical immunosensor. Thus, the immunized mouse sera served as a reference to assessing the capability of the functionalized biosensor to detect specific antibodies under controlled conditions, prior to its application to swine samples.
GP5-B was used in the EIS-based immunosensor for detecting antigen-specific antibodies. The quantitative behaviors of the immunosensors were evaluated by measuring the charge-transfer resistances (Rct) of different concentrations of anti-GP5-B antibodies. For the immunosensor preparation, AuSPE electrodes were sequentially modified with a TGA self-assembled monolayer, activated with NHS/EDC chemistry, and functionalized with GP5-B peptide, and finally unspecific binding with casein was avoided. Casein was selected after preliminary optimization experiments, comparing with BSA or ovalbumin proteins.
Impedance spectra were generated for the surface-modified electrodes and are presented as Nyquist plots in Fig. 2a. For the AuSPE electrode, the Rct value is 62 Ω. Then, the SAM (Rct = 90 Ω) formed by the addition of TGA was attached to the AuSPE electrode, and GP5-B (Rct = 150 Ω) was immobilized on the activated electrode surface (TGA-NHS/EDC), increasing Rct.
The modified electrodes were incubated with casein, which further increased the charge-transfer resistance (Rct) to approximately 400 Ω. Consistent electrochemical responses were observed across 30 independently prepared electrodes. For repeatability assessment, ten independent batches were prepared, each consisting of three immunosensors. Measurements were performed in triplicate for each electrode. The intra-assay coefficient of variation (CV) was 5.05%, and the inter-assay CV was 4.08% (Fig. 2b); the data showed non-significant differences among batches. Sera from non-immunized mice, non-vaccinated pigs, and rabbits were used to compare the complexity of the blood serum. These samples showed impedance responses comparable to baseline values (Fig. 2c). In this case, the intra-assay CV was 7.31% and the inter-assay CV was 2.97%, supporting the analytical assessment.
The modified electrodes (TGA-NHS/EDC) were further incubated with 1 μL serum from a GP5-B-immunized mouse before calibration with the blank sample, and Rct increased (ΔRct = 600 Ω). Fig. 3a shows the Nyquist plots for the modified electrodes incubated with different serum antibody concentrations, with a linear range from 0 to 42 ng mL−1 after serum addition from GP5-B-immunized mice, while Fig. 3b shows the linear regression of the antibody concentration vs. ΔRct relationship, revealing a correlation coefficient of R2 = 0.9541, and compares the plots of the antibody concentration vs. ΔRct relationship obtained using both the immunosensors and ELISA (ΔOD450nm; R2 = 0.9636). Compared to the baseline EIS signal (Rct = 400 Ω at a serum antibody concentration of 0 ng mL−1), the EIS signal intensified to Rct = 600, 700, 850, and 900 Ω with increasing serum antibody concentration to 6, 18, 30, and 42 ng mL−1, respectively.
Based on Fig. 3, the lowest antibody concentration evaluated was 0.6 ng mL−1, which produced a reproducible and clearly impedimetric signal relative to the baseline. At this concentration, an approximate increase of 200 Ω in Rct was observed compared with the blank measurement, exceeding baseline variability confirming reliable signal discrimination. Therefore, the LOD of the proposed immunosensor was established at 6 ng mL−1, corresponding to the lowest concentration tested and consistent with the statistical 3σ criterion.24
The immunosensor's responses to antibodies in the GP5-B-immunized mice were compared with the responses to the control mice (e.g., the intact-animal control and maleimide-carrier/BSA-immunized groups). The results showed statistically significant differences between the GP5-B-immunized and both the control and BSA-immunized groups (Fig. 4a–d).
The immunosensor's responses to antibodies in the naturally infected and vaccinated swine were compared with the responses to the control swine (PRDC free and nonvaccinated animals). The results showed statistically significant between-group differences (Fig. 5a–e).
These results demonstrate that the GP5-B functionalized impedimetric immunosensor is capable of selectively and sensitively discriminating between sera containing GP5-B specific antibodies and control sera lacking antigen-specific immune responses. The significantly higher Rct values observed in GP5-B immunized mice, compared with the controls, confirm that the measured impedimetric charges arise from specific antigen–antibody interactions at the electrode surface. This clear separation between experimental and control groups validates the GP5-B peptide as an effective recognition element. Importantly, the robust and reproducible detection of peptide-specific antibodies in a controlled murine model provides strong experimental support for the proposed diagnostic applications, establishing a reliable proof of concept prior to translation to vaccinated and naturally infected swine samples.
To determine whether the test result was positive or negative, the cutoff point was analyzed using an ROC curve plotting Rct-derived sensitivity and specificity percentages for the control and experimental groups (n = 54), representing 100% for each group. For each possible cutoff point, the tabulated values between positive and negative test results are shown (p < 0.0001). A 95% confidence interval was obtained. The biosensor exhibited 100% sensitivity and 93.33% specificity, making it suitable for detecting antibodies (Fig. 6) and confirming the feasibility of practically applying the proposed GP5-B-probe-based immunosensor for detecting anti-PRRSV antibodies.
Overall, these results demonstrate that the proposed immunosensor effectively discriminates between PRRSV-seropositive and seronegative swine under biologically relevant conditions. The significantly higher ΔRct values observed in both naturally infected and vaccinated pigs, compared with PRDC-free and nonvaccinated controls, indicate that the biosensor reliably detects circulating anti-PRRSV antibodies generated by either natural infection or immunization. The consistent impedimetric responses across these clinically relevant groups highlight the robustness of the GP5-B peptide as a recognition element and underscore the diagnostic potential of the proposed platform for serological surveillance and herd-level monitoring of PRRSV exposure.
(ii) AuSPEs only required small samples, were less expensive than other conventional antibody detection methods, and provided immediate results.
(iii) The immunosensor calibration revealed analyte-dependent linearity and yielded an LOD of 6 ng mL−1.
(iv) The immunosensor selectively detected anti-PRRSV antibodies, without any interference from other proteins, compared with other animal species serum proteins.
(v) The biosensor exhibited 100% sensitivity and 93.3% specificity.
The proposed biosensor was applied to detect specific antibodies in a mouse model and pigs in farms, and this could represent a useful tool for detecting PRRSV antibodies and contributing to biosecurity and animal health.
Usually, peptide-based electrochemical biosensors integrate a specific peptide sequence as the main component, modified with a thiol for peptide immobilization. However, although immunosensors can be developed using peptide antigens or antibody biorecognition elements, they can also integrate antibody or antigen fractions.
Immunosensors are based on specific antigen–antibody interactions and can be a tool for quantitatively or qualitatively detecting antigen-specific antibodies.26
Although the preliminary results are promising, extensive field validation studies are required to confirm diagnostic performance across diverse epidemiological conditions. Likewise, the stability of the immunosensor over time could be evaluated for future research.
The proposed GP5-derived peptide immunosensor represents a promising alternative for rapid PRRSV antibody detection, combining analytical sensitivity, specificity, and operational simplicity, which may facilitate on-site serological surveillance in swine production systems.
| This journal is © The Royal Society of Chemistry 2026 |