Open Access Article
Julia Sophie Böke†
ace,
Cornelia Reuter†
*abc,
Mark Kielpinskia,
Anett Reichertab,
Iris Prinzd,
Daniel Hornerd,
Adrian Prinzd,
Fritz Bichlerd,
Rudolf Aichingerd,
Anton Rettenbacherd,
Georg Bauerd,
Maria Bauerd,
Susanne Pahlow
abc,
Karina Webera and
Thomas Henkel
*a
aLeibniz Institute of Photonic Technology, Member of the Research Alliance “Leibniz Health Technologies”, The Leibniz Centre for Photonics in Infection Research (LPI), Albert-Einstein-Straße 9, 07745 Jena, Germany. E-mail: cornelia.reuter@leibniz-ipht.de; thomas.henkel@leibniz-ipht.de
bInfectoGnostics Research Campus Jena, Center for Applied Research, Philosophenweg 7, 07743 Jena, Germany
cFriedrich Schiller University Jena, Institute of Physical Chemistry and Abbe Center of Photonics, Helmholtzweg 4, 07743 Jena, Germany
dSTRATEC Consumables GmbH, Sonystr. 20, 5081 Anif, Salzburg, Austria
eCluster of Excellence - Balance of the Microverse, Friedrich Schiller University Jena, 07743 Jena, Germany
First published on 22nd May 2026
Digital droplet-based bioassays (ddAssays) have been applied in numerous fields, though many existing microfluidic platforms are tailored to specific assay classes and require complex workflows involving multiple processing steps. Here, we present a fully integrated microfluidic platform for generic digital bioassays that combines droplet generation and storage for on-chip incubation and optical readout on a single disposable device. This microfluidic platform enables the parallel execution of test and control experiments, each using more than 200
000 monodisperse droplets of 35 pL. CFD simulations of the step emulsification unit, together with experimental characterization, confirm the predicted device performance. The device is prepared by variothermal compression molding. A dual-layer photoresist strategy was used to create the master for replicating the channel systems with two different channel heights and step transitions between them. The core functionality and the applicability of various optical modalities for image-based readout are demonstrated by exemplary biochemical and microbial growth reporter assays. The platform is compatible with automated image-based analysis for high-throughput digital assay readout.
Since the breakthrough of encapsulating deoxyribonucleic acid (DNA) in droplets using microfluidics,8,9 ddPCR platforms have been applied in numerous applications.10,11 Commercial state-of-the-art devices using ddPCR, as reviewed by Tan et al., are Naica by Stilla Technologies, the QX series and RainDrop, both by Bio-Rad Laboratories.12 Naica is prefilled with the continuous phase. Summarized, the QX series and Naica have similar capabilities, with around 25
000 droplets of approximately 500 pL partitioned volume per sample, leading to a droplet diameter of approximately 100 µm, and 20 µL total reaction volume. The RainDrop can produce up to 1 × 107 droplets, with a droplet diameter of 17 µm and respective volume of 5 pL, in a total reaction volume of up to 50 µL.12,13
The field of digital droplet-based ELISA (ddELISA) is less advanced than ddPCR, as it relies on the isolation of functionalized beads in picoliter droplets for protein quantification.14,15 Its complexity arises from the challenge of bead encapsulation and the inherently weak fluorescence signal of single-enzyme detection, which lacks PCR-like amplification.14
In droplet-based microfluidics,16 a diluted aqueous sample is partitioned through an often perfluorinated oil phase into droplets.9,17,18 This approach combines emulsion systems with precise liquid handling in continuous-flow microfluidics. The most common approaches for on-chip droplet generation include step emulsification (SE),19,20 and flow-focusing21 on nozzles,22,23 T-junctions24,25 or X-junctions.26,27
Kaminski et al. summarized the progression of droplet microfluidics, focusing on its relevance to microbiology. Ongoing developments aim to improve droplet stability and generation rates. Methods have been developed to support the growth of microorganisms with dissolved gases such as oxygen and CO2.6 Various microfluidic platforms exist and have been used to implement ddAssays at different levels of integration. Microchannels with prolonged residence time28 have been used for decades to enable flow-through incubation, e.g., in combination with thermocyclers for PCR.29 Schuler et al. developed a centrifugal microfluidic platform that combines droplet generation and PCR amplification within a single device architecture,30 while Schulz et al. presented a highly integrated digital PCR system that enables sample-to-answer nucleic acid detection using step emulsification and on-chip thermocycling.31 In parallel, several studies have focused on improving droplet generation throughput and monodispersity. Xu et al. parallelized 2000 nozzles for spontaneous droplet generation of pL-volume governed by a Laplace pressure difference at the SE unit.23 Wei et al. designed a microfluidic chip that employs step emulsification to directly generate droplets into the storage reservoir, producing droplets with diameters of 38.2 µm and 110.3 µm.13 Comparable work by Zhang et al. achieved high-throughput droplet formation using 50 parallel step emulsification units, generating uniform ∼175 µm droplets with a coefficient of variation (CV) of 3% at flow rates of 1.4–17.5 nL s−1.32 Lastly, the application of deep learning algorithms has shown promise in further lowering detection limits for rapid classification and analysis.33
Despite significant progress in droplet microfluidics, many existing systems are tailored to specific assay types, rely on complex workflows involving multiple devices or off-chip transfer steps between droplet generation, incubation, and readout, use unconventional microfluidic device footprints or require specialized instrumentation for their operation. This limits their flexibility and increases the risk of droplet loss or instability during handling. In addition, popular commercial platforms require prefilled microfluidic devices or specialized preparation procedures, which complicate operation and reduce compatibility with standard laboratory workflows.
In this work, we present a generic microfluidic platform for ddAssays that integrates droplet generation, incubation, storage, and optical readout on a single disposable microfluidic chip. The platform supports two assay units, each with more than 200
000 droplets, implemented on the footprint of a standard microscopy slide, enabling direct image-based readout with conventional microscopy systems. The droplets remain in the storage unit during the assay34 but can be released and completely recovered if needed. Initial filling with custom fluids and reagents can be easily done on-site as a part of the assay workflow. The droplet generator design is guided by rational design principles and validated by computational fluid dynamics (CFD) simulations35–38 and experimental characterization. The device is prepared by STRATEC Consumables GmbH using variothermal injection-compression molding, a technology established for high-volume replication of microstructured polymer components. In this paper, we demonstrate the operation of the microfluidic ddAssay chip in terms of initial filling with the continuous phase, sample droplet formation, incubation and on-chip readout. Application examples across digital droplet-based assay classes are presented and discussed, and compatibility with various optical readout modalities is demonstrated.
The unique selling point of the designed droplet generator is that droplet volume and generation rate can be easily predicted by simple geometrical rules. In the rational design phase, the geometric and control parameters are adjusted to the defined process parameters using mathematical models. Afterwards, it is validated by in silico experiments using CFD simulations. Therefore, this structure can be easily customized and integrated into microfluidic networks.
For the released droplet, a diameter of 38 µm (V = 28.7 pL) was specified in advance. Around 200
000 droplets should be generated in less than two minutes, which requires a droplet generation rate in the order of >1500–2000 droplets per second and a total flow rate of >50 nL s−1 for the dispersed phase fluid.
Within the head section, the accumulated liquid forms a disk-shaped body with half-circular side walls before being released into the droplet-collecting channel. The volume of this body can be calculated from its height H and diameter D using eqn (1), obtained by integrating the droplet shape in cylindrical coordinates.
![]() | (1) |
For a design volume of 25 pL and a height of 12 µm, the diameter evaluates to 54 µm at the center plane of the body. The head section can be constructed by fitting this body into the geometry in Fig. 2C.
| pSE unit = pHP + piface | (2) |
Pressure drop due to viscous losses is predicted using the Hagen–Poiseuille model with the hydraulic diameter approximation for channels with non-circular cross-section. The model eqn (3) already includes the simplifications for square channel cross-sections. The symbol µ represents the viscosity, L the length, Q the volume flow and H the width and height of the channel. The negative sign reflects the pressure drop p = pout − pin.
![]() | (3) |
To predict the interface-generated back pressure, the Laplace equation is used, where σ denotes the interface tension and R1 and R2 the vertical and horizontal radii of curvature of the liquid–liquid interface. The negative sign reflects the orientation of the curved interface relative to the channel direction from inlet to outlet.
![]() | (4) |
The merged models provide the pressure/throughput characteristics for either p or Q as the independent parameters.
![]() | (5) |
![]() | (6) |
The target system with parallelized step emulsification units is operated at a constant pressure drop between the sample loop and droplet collecting channel as discussed before. Onward, the specified pressure reflects the pressure settings, so that the minus sign disappears unless it is explicitly stated that it is the negative pressure drop. During droplet generation, the curvature of the liquid–liquid interface in the head section periodically changes, resulting in fluctuations in the flow rate.
These volume flow fluctuations are illustrated in Fig. 3 for the volume flow dependent on the diameter of the growing droplet disk in the head section. At low pressure settings of 10 mbar, the interface-generated back pressure is not compensated, and negative volume flow towards the inlet is observed. For pressure settings >20 mbar, volume flow is permanently directed to the outlet.
During droplet release, the volume accumulated in the head section is spilled into the droplet-collecting channel as a spherical droplet with a lower pressure drop across the interface compared to the disk-shaped volume part inside the head section. This causes the liquid to be expelled from the head section into the droplet. At the end of this process, the droplet is released by interface rupture. This requires that the volume flow through the unit be limited. For reliable operation, we expect that the volume flow ratio qratio through the SE unit at the time of droplet release (Fig. 2F) should be less than 1.25 times the volume flow when filling the head section (Fig. 2E). This value is defined in eqn (7) where R1 represents the radius of the vertical, R2 the radius of the horizontal curvature of the maximum volume body inside the neck section and R3 represents the radius of the droplet immediately before interface rupture.
![]() | (7) |
Substitution and simplification yield eqn (8),
![]() | (8) |
The pressure dependency of the qratio parameter is shown in Fig. 5 for the target-step emulsification unit. The value of qratio < 1.25, where the unit is expected to work in a reliable regime, is found at a pressure setting of 3000 Pa (30 mbar).
Now, the length of the neck channel is defined so that the volume flow at 3000 Pa matches the desired volume flow to achieve the desired droplet generation rate.
The maximum and minimum volume flow for the parallelized droplet generator with 92 SE unit elements, as a function of the neck channel length, are shown in Fig. 4. A volume flow of 80 nL s−1 can be achieved with a neck channel length of 1 mm.
![]() | ||
| Fig. 4 Minimum and maximum volume flow through the parallelized droplet generator with 92 SE unit elements dependent on the neck channel length. | ||
Simulations of the droplet generation process were performed using the interFoam solver of the OpenFoam CFD toolkit,35,36 version 2406. Post-processing and visualization were performed using ParaView,37 based on VTK.38 A zero-contact-angle boundary condition was applied by setting the contact angle between the continuous phase fluid and channel wall to 0°.46 Details are provided in the Materials and methods section. The template case for rerunning the simulation is provided in the digital supplement.
In the simulation, a pure pressure-driven flow was used, with fixed pressure constraints at the inlet and outlet faces, and a pressure drop of 3000 Pa between them. This corresponds to a single-step emulsification unit within the parallelized droplet generator, in which all 92 units were fed from a single sampling channel at (almost) identical pressure. Fluid properties and simulation parameters are summarized in Table 1. The throughput through the unit depends on the steepness of the pressure gradient, which is influenced by the Laplace pressure drop across the evolving interface of the formed droplets, as discussed in the Rational design – head section.
| Interface property (fluid/wall) | ||
|---|---|---|
| Contact angle | 0° | 180° |
| Boundary conditions (fluid/wall) | ||
|---|---|---|
| Outlet pressure | fixedValue 0 Pa | Uniform 0 Pa |
| Inlet pressure | fixedValue 0 Pa | Uniform 3000.0 Pa |
As long as the interface is within the neck channel section, where it has the smallest curvature, as shown in Fig. 6B, the minimum flow rate is observed. The situation reflects the initiation process, in which the continuous oil phase in the neck channel is successively substituted by the sample fluid.
![]() | ||
| Fig. 6 Visualization of CFD results‡ with fixed inlet pressure – constraint. (A) Domain overview for the single droplet generator unit showing the sample loop channel, the neck section and the head section with the step transition into the main channel (from left to right). (B–G) Geometry of the liquid–liquid interface at different time steps of the droplet generation cycle. Local interface velocity is visualized by the color map. The mean fluid velocity in the neck channel is shown in the bar chart, and the pressure drop along the x-axis (green line) is shown in the line chart insert of each subplot. | ||
The droplet generation cycle starts when the interface expands into the head section and forms a disc-shaped body. As shown in Fig. 6C and D, the horizontal radius of curvature increases, thereby reducing the pressure drop across the interface. Due to the fixed pressure settings at the inlet and outlet, the pressure drop along the neck channel increases, thereby increasing the transport velocity of the fluid. The situation reflects the stage at which the liquid forming the droplet accumulates in the head section.
As soon as the interface passes the step in front of the head section, a spherical droplet starts growing into the droplet-collecting channel. Due to the smaller pressure drop across the droplet-shaped interface compared to the disc-shaped interface inside the head section, the liquid accumulated there spills into the growing droplet as visible in Fig. 6C–E. Due to the reduced interface-generated back pressure, the volume flow through the neck channel is also increasing. However, since this is limited by the throttling effect of the neck channel, the phase boundary within the neck section is narrowed, and the droplet formed in the droplet collecting channel breaks off as visible in Fig. 6F. After separation at the interface, the remaining volume in the head section snaps back and the formed droplet is released into the droplet collecting channel. Now that the droplet generation cycle is completed, the next cycle starts.
In the simulations, the droplet generation cycle takes 16.8 ms, corresponding to a droplet generation rate of 60 Hz. For the parallelized droplet generator with 92 droplet generator units, the simulations predict a total droplet generation rate of ∼4550 droplets per second and a sample volume flow rate of 140 nL s−1, which is sufficient to fill the chip with droplets in less than 1 minute.
This experimentally confirms the generation of a densely packed droplet suspension with a droplet volume of 35 pL at a total sample flow rate of 150 nL s−1 and a droplet generation rate of around 4300 droplets per second. CFD results and experimental data are in good agreement.
Minor variations in droplet diameter can occur during formation but remain sufficiently small for digital assay applications. Additionally, each droplet volume can be determined by image-based readout, allowing accurate quantification.
In summary, the droplet generator is confirmed to produce droplets of suitable size and droplet generation rate. The confirmed droplet generation rate and sample flow rate are three times higher than the initially drafted parameters and can be scaled down to a sample flow rate of 50 nL s−1 for the experiments.
Considering hexagonal dense packaging, it can hold up to 252
000 droplets with a diameter of 40 µm. In the case of cubic dense packaging, the number of droplets reduces to 204
000. If the optically measured droplet diameter falls below the channel height of 40 µm, the sphere volume equation can be used. Otherwise, the volume VRB of a rotational body with flat top and bottom face, as given in eqn (1), is recommended.
The droplet retainer unit is realized as a biconcave plateau structure with a center duct spanning the width of the main channel, with a plateau height of 28 µm, and a minimum duct width of 18 µm. The main channel height is 40 µm, and the main channel width is 1200 µm. Such a plateau can be passed by the continuous-phase fluid. At low flow rates, any droplet or air bubble will be retained, which corresponds to the target functionality. In certain scenarios, however, the droplets must also pass the barrier in a defined way. This applies when sample droplets need to be collected into an external reservoir, or when large air bubbles have to pass through the barrier during initial filling without leaving any air fragments behind. Both situations can be managed. Depending on the applied flow rate, droplet size and interface tension, droplets can pass through the duct or overcome the barrier across the full width of the main channel.
All targeted functionalities are experimentally confirmed and shown in Fig. 9A. A flow rate of 50 nL s−1 is used for the continuous phase fluid HFE7500 containing 2% of the surfactant dSurf. The air bubble passes through the barrier duct without fragmentation. Droplets with a diameter between 30 and 40 µm can be retained for droplet storage during incubation and readout at continuous-phase flow rates below 0.2 nL s−1 (Fig. 9C), which is below the parasitic flow rates induced by thermal or inertial effects during incubation. This flow rate can also be used to supply the growing cultures with gases dissolved in the continuous phase liquid. In this case, the complete liquid exchange can be achieved within 4.5 h.
Optionally, the chip can be detached from the fluid management system, and the fluid ports can be sealed with adhesive tape without disturbing the arrangement of the stored droplets. The droplet retainer contributes to this operation by retaining the droplets inside the storage and incubation unit. This makes the fluid management available for other tasks, e.g., next ddAssay platform filling, and improves the optical accessibility of the chip for optical readout using microscope objectives.
Injection compression molding technology47 was originally developed for the mass production of optical data storage media such as CDs, DVDs, and Blu-ray discs, which require the precise replication of sub-micron surface relief structures. The process employs interchangeable master plates carrying the surface microstructures, allowing rapid adaptation of the mold insert. The molding cycle comprises injection, compression, and demolding steps. During injection, the polymer melt is introduced into the mold at a temperature above its melting point and subsequently cooled to a temperature between the melting and softening points. In the compression step, the polymer is pressed against the master plate to accurately replicate the surface microstructures and to define the final thickness of the workpiece. After demolding, the replicated disc is available for post-processing and finishing.
For the fabrication of microfluidic devices, the channel geometries must be implemented on the master plate as surface-relief structures with a small draft angle to enable reliable demolding. The realization of channel sections with different heights on a single substrate requires master plates featuring multiple structure depths, which significantly complicates master fabrication. During post-processing, fluidic access holes are introduced, the microchannels are sealed by bonding to a cover layer, and the individual chips are subsequently separated from the molded disc.
The presented microfluidic device was prepared from the cyclic olefin polymers, COP1420R (molded part) and COC6013M (cover film). A dual-layer photoresist strategy, based on sequential lithographic processing, is used to generate channel geometries and microfluidic operation units, followed by an electroplating process to convert them into a nickel master plate.
Achieving accurate alignment between the two layers, controlling the photoresist thicknesses, and incorporating an appropriate draft angle (typically < 2°) are critical process considerations to ensure successful demolding in subsequent manufacturing steps.
This dual-layer resist strategy enables the precise generation and reliable replication of the required geometries.48 High replication accuracy and excellent dimensional agreement with the design were achieved even for the challenging geometries of the parallelized droplet generator (Fig. 10) and the droplet retainer structures.
The variothermal molding process significantly enhances replication quality by suppressing premature solidification and uncontrolled polymer relaxation during cooling, which is critical for preserving fine microstructural features with high aspect ratios and for minimizing defects such as weld lines, flow marks, and sink marks.
About 1000 devices have been prepared with uniform quality within multiple batches. Batch-to-batch variability is influenced by several factors, including material batch variations, material pretreatment, and, most critically, the condition and quality of the mold stamper. While the variothermal process primarily improves the molding and demolding of fine microstructures, careful control of material properties and stamper integrity was maintained to minimize inter-batch variability.
A digital helicase-dependent amplification (HDA)50,51 assay was performed for the quantification of A. baumannii DSM30007 DNA. Droplets were created from a total volume of a nucleic acid amplification reaction mixture of 15 µL, which was loaded into the interconnection tubing of the sample syringe and attached to the short tubing at the chip mount using polydimethylsiloxane (PDMS) sleeves.
A second syringe filled with Novec HFE7500 was attached in the same way to the mineral oil inlet tubing of the chip mount. The outlet capillary of the sampling was opened. At this point, the initial filling procedure is started at a flow rate of 50 nL s−1 until the oil phase reaches the outlet capillary of the sampling loop.
This procedure takes about 3–4 minutes, depending on the length of the interconnecting capillary. Continuous-phase fluid flow continues during the subsequent operations until the droplet generation and loading process is finished. Subsequently, the sampling fluid is forwarded at a flow rate of 50 nL s−1 into the sampling loop until the preceding air has been completely passed through the sampling loop and the sample fluid has reached the open end of the capillary. The sample flow is stopped, and the sampling loop outlet is sealed by sticking a pin needle into the capillary. Following that, the system is ready to generate and load the droplets into the store. Droplets are generated at a flow rate of 150 nL s−1 for the sample fluid and 50 nL s−1 for the continuous oil phase. Under these conditions, the droplet generation and loading procedure is completed after 62 seconds, when the first droplets arrive and pass the outlet droplet retainer structure of the chip, and a total sample volume of 9.34 µL is converted into droplets. Fluid connections can be detached, the capillaries are closed with needles, and the device is incubated at 65 °C for one hour to complete the HDA. The droplet arrangements are subsequently read out using fluorescence microscopy (Fig. 11A and B).
Droplet dense packaging is preserved during the amplification and incubation procedure. Fluorescent positive droplets and non-fluorescent negative droplets are clearly discriminated and can be applied to the subsequent data analysis for counting the number of positive and negative droplets.52
Droplets were generated from an initial bacterial suspension (300 µL), which was loaded into the interconnection tubing of the sample syringe and connected to the short tubing at the chip mount utilizing PDMS sleeves. The flow rates are adjusted to ensure stable droplet formation. During droplet generation, the bacterial suspension and the continuous phase are supplied at flow rates of 150 nL s−1 and 50 nL s−1, respectively. The droplets are collected in the droplet storage, and the chip is transferred to an incubator set at 37 °C for 6–20 hours to allow proliferation. After incubation, the droplets are analyzed using fluorescence microscopy to assess cell viability and growth. The readout of one representative microfluidic chip, as well as the associated imaging overlay, is presented in Fig. 11C–E. Successful growth of E. coli Nissle 1917 within the droplets was observed after the incubation period, as evidenced by a progressive increase in cell density and corresponding fluorescence signal. Brightfield imaging confirms the presence and proliferation of bacterial colonies, while fluorescence imaging reveals distinct signal development, indicating active expression of the fluorescent dye. The uniformity of growth across droplets demonstrated consistent encapsulation and stable culture conditions within the microfluidic system. Fluorescent-positive droplets and non-fluorescent-negative droplets are clearly distinguished and can be applied to the subsequent data analysis. The droplets remained stable throughout the experiment and maintained large size uniformity, without droplet merging observed over 24 hours, ensuring consistent microenvironments for bacterial cultivation.
The full case template for running the simulation is provided in the digital supplement under the file name <stepEmuV3ScaledFixedInletPressureTemplate>. Dimensions and parameters have been scaled to the base units of 1.0 mm for length, 1.0 s for time, and 1.0 mg for mass to ensure numerical stability in the CFD simulations. All parameters and fluid properties are given in the ParametersAndScaling spreadsheet table in the ParametersAndScaling subfolder of the case template. A hexagonal volume mesh has been generated from a surface triangulated mesh of the SE unit using the snappyHexmesh utility with a uniform cell size in x, y, z of 0.86, 0.86, 0.86 µm3.
The outlet patch was set to the ‘inletOutlet’ type for velocity, fixedValue’ type with p = 0 Pa for pressure and ‘inletOutlet’ type for the phase fraction alpha.water.
The inlet patch was set to the ‘pressureInletOutletVelocity’ type for velocity, ‘fixedValue’ type with p = 3000 Pa for pressure and ‘inletOutlet’ type for the phase fraction alpha.water.
Channel walls have been set to ‘notSlip’ type for velocity, fixedFluxPressure’ type for pressure and ‘constantAlphaContactAngle’ with theta0 = 180.0°.
The fluid properties of water and Novec HFE7500, with an interfacial tension of 5.5 × 10−3 N m−1, have been used for both the continuous and dispersed phases.
Dynamic time-step sizing was activated, with the Courant number limited to maxCo = 0.05. The neck channel was initially filled with the dispersed-phase fluid using the setFields utility of the OpenFoam Toolkit.
The shape of the liquid/liquid interface was initially allowed to converge using manual control and increasing the time step size, starting with deltaT = 1 × 10 s. The simulation was allowed to run for 23 ms, which was sufficient to model the creation and release of a single droplet.
Finally, the simulation results were converted to the generic, long-term-supported VTK data format using the toolkit's ‘foamToVTK’ utility and saved in the ‘VTK_RESULTS’ subfolder. All parts of the simulation can be controlled by starting the related run script.
Details are provided in the Readme.txt file in the root folder of the case template. For rerunning the postprocessing with ParaView, the respective pvsm state files from the VTK_RESULTS_ALL or VTK_RESULTS_SUBSET folder can be loaded into the ParaView app, where the VTK_RESULTS_SUBSETS folder contains all the data for reproducing Fig. 6.
Initially, a 12 µm-thick first lithographic layer of the negative photoresist SU-8 3010 was spin-coated, soft-baked, exposed, and post-baked (Fig. 12A), but not developed. Following the post-exposure bake, a second photoresist layer with a thickness of 29 µm was spin-coated on top of the first layer and subsequently exposed with an alignment accuracy of approximately 2 µm relative to the initial pattern (Fig. 12B and C). After exposure, a development step using PGMEA solvent was carried out to selectively remove the unexposed regions of both resist layers, thereby generating a dual-level lithographic structure (Fig. 12D).
To produce the injection-molding insert, the two-level lithographic master was replicated in a nickel mold via electroplating of around 300 µm thickness (Fig. 12E). After nickel deposition, the backside of the mold insert was polished to achieve a flat surface while preserving the integrity of the microstructures. The final polished nickel insert then served as the mold insert for the injection molding process (Fig. 12F).
The four fluid ports on each side are drilled with a diameter of 0.6 mm. They are arranged with a line spacing of 3 mm, a hole spacing of 4.5 mm, and an offset of 2.25 mm.
Solvent-assisted thermal bonding is a proprietary internal process of STRATEC Consumables GmbH. It is a technique that does not require additional materials like adhesive layers. The principle works like this: the COP is activated by a solvent on the surface, and both parts are bonded by decent heat and pressure (typical ranges for such bonding are 60–90 °C) in a custom-specific bonding setup. Due to the activation, only the upper surface becomes soft, making it possible to bond channels with very high precision. Global flatness <100 µm could be achieved after bonding, without leakage or crosstalk of channels. A channel height precision with a tolerance of around 1 µm was achieved.
![]() | ||
| Fig. 13 The fluid interconnecting block on the left side connects PTFE tubings to the chip. The block on the right shows the inner compartments and sealings with flanged tubings and positioning pins. | ||
The fluidic connector consists of three main parts: the upper part with through holes for the tubing/screws and a mold for a rubber seal. The lower counterpart generates the clamp force. Additionally, screws and pins are inserted to clamp the assembly and to adjust the parallel alignment of the upper and lower parts.
The mounting is guided by a recess for inserting the chip and two pins. Sealing is achieved by deforming the seal as the connector closes, generating a seal-to-chip and seal-to-tubing contact.
The PTFE tubings, size OD 1.59 mm/ID 0.5 mm (BOLA, Bohlender GmbH, Grünsfeld, Germany), were flanged with the easy-flange tool (VICI Jour, Schenkon, Switzerland). The rubber seal is made by compression molding of HYPEX HX61 MB (TKV Thermoform GmbH, Pößneck, Germany). The upper part is made of aluminum, whereas the lower part is made of stainless steel.
Brightfield and fluorescence images for assay data acquisition are obtained using either a 10×/NA 0.5 objective (Zeiss, Oberkochen, Germany) or 10×/NA 0.45 and 40×/NA 0.95 objectives (KEYENCE Deutschland GmbH, Neu-Isenburg, Germany).
For the continuous phase fluid, Novec HFE7500 (3M, St. Paul, Minnesota, USA) has been used in combination with the surfactant dSurf (Fluigent, Le Kremlin-Bicêtre, France) at a working concentration of 2% (w/v).
First, the rational design of the droplet generator is confirmed by in silico evaluation via CFD simulations. The droplet generator is designed to hold a droplet volume of 25 pL, resulting in an experimentally confirmed droplet volume of 35 pL at a total sample flow rate of 150 nL s−1 and a droplet generation rate of around 4300 droplets per second. Considering hexagonal dense packing, the ddAssay can hold up to 252
000 droplets with a diameter of 40 µm and can be filled in less than two minutes. Thus, rational design, CFD simulations and experimental data demonstrate good agreement.
Second, the microfluidic device is fabricated as a single-use device according to the design parameters by injection-compression molding, using a two-layer resist strategy for master preparation. Two independent assay structures for the parallel execution of test and control experiments are realized on the footprint of a standard microscope slide. Approximately 1000 microfluidic devices were produced with uniform quality across multiple batches, highlighting the quality and robustness of the fabrication process and the well-thought-out design of the microfluidic platform.
As the final proof, we demonstrate two bioassays that highlight the versatility of the presented generic digital droplet-based bioassay platform. A digital helicase-dependent amplification (HDA) assay for Acinetobacter baumannii DSM30007 DNA confirms reliable droplet generation, storage, incubation, and fluorescence-based readout, validating the platform for digital nucleic acid amplification assays. In addition, a droplet-based single-cell growth assay using Escherichia coli Nissle 1917 enables monitoring of bacterial viability and growth dynamics at the single-droplet level, demonstrating the platform's capability for quantitative microbiological assays and its potential applicability in microbiome research.
The high droplet throughput of the presented platform further motivates the use of artificial intelligence-based image analysis for automated process control, multiplex droplet identification, and scalable downstream data processing in future digital bioassay applications.
The data supporting this article have been included as part of the supplementary information (SI). Supplementary information: CFD template case in S1 and videos supporting the experimental details S2. See DOI: https://doi.org/10.1039/d6ra02395a.
Footnotes |
| † J. S. B. and C. R. contributed equally. |
| ‡ Provided in the digital supplement S1. |
| This journal is © The Royal Society of Chemistry 2026 |