Open Access Article
Abdul Hoque
a,
Evan L. Pannkukbcd,
Jerome Lacombeae,
Sunil Bansalb,
Jerry Angdisenb,
Geraldine Vitryb,
Abel Encarnacion Hernandeza,
Nabhan M. Fakrudin
a,
Mathew Barretta,
James Heltona,
Jian Gu
a,
Albert J. Fornace Jrbcdf,
Evagelia C. Laiakisbcdf and
Frederic Zenhausern*aeg
aCenter for Applied NanoBiosience & Medicine, University of Arizona, Phoenix, Arizona 85004, USA. E-mail: fzenhaus@arizona.edu
bDepartment of Oncology, Lombardi Comprehensive Cancer Center, Georgetown University Medical Center, Washington, DC 20057, USA
cDepartment of Biochemistry and Molecular & Cellular Biology, Georgetown University Medical Center, Washington, DC 20057, USA
dCenter for Metabolomic Studies, Georgetown University, Washington, DC 20057, USA
eDepartment of Basic Medical Sciences, College of Medicine Phoenix, University of Arizona, 85004, USA
fDepartment of Radiation Medicine, Georgetown University Medical Center, Washington, DC 20057, USA
gHonor Health Research Institute, Scottsdale, Arizona 85260, USA
First published on 8th May 2026
Novel biodosimetry assays are needed for potential radiological incidents to rapidly assess radiation exposure and guide medical treatments. Mass spectrometry-based metabolomic analysis using a sorptive phase extraction is a rapid and efficient method for radiation-induced biomarkers in biofluids. Here, we developed a chemically functionalized polymeric Sorbent Phase Sorptive Extraction (SPSE) method. This method employs polymeric thin film sorbents with tailored organic functional groups, which directly bind radiation-responsive biomolecules and increase sample absorption capacity. This microporous membrane system enables rapid, high-sensitivity extraction of metabolites spanning a wide polarity range from urine, serum, and whole blood. We characterized the surface morphology, chemical functionality, and hydrophilicity of multiple sorbent-coated cellulose membranes, including plasma-functionalized nylon-6. Matrix interference was evaluated using untargeted metabolomics, and analytical performance was assessed using a targeted multiplex radiation biomarker panel. Urine, serum, and whole blood were collected from male and female C57BL/6 mice (9 weeks old) exposed to X-rays at 1 day (0, 2, 8, 13 Gy) and 7 days (0, 2, 8 Gy) post-irradiation. The membrane types preserve metabolite stability at room temperature for up to two weeks; however, nylon-6-based cellulose paper membranes exhibited the highest surface porosity, absorption capacity, and metabolite recovery. Classifier performance evaluated using receiver operating characteristic analysis demonstrated comparable sensitivity and specificity between SPSE and conventional dilute-and-shoot workflows. Collectively, these results support further development of polymeric sorbent coated paper and fabric-based substrates to increase throughput while eliminating cold-chain requirements. These environmentally conscious design features exemplify principles of green chemistry including lowering chemical waste and operational energy demand.
Collecting and preparing biological samples for the assessment of IR-induced biomarkers from such a large population would ideally incorporate a process that is fast, cost effective, and robust compared to classical sample preparation techniques.23 For this purpose, a pragmatic and easily accessible platform is required that ensure simplicity, scalability, and make it suitable for point-of-care (POC) applications, where only a minimal blood volume, such as a finger-prick sample, is needed for rapid biomarker extraction and stabilization. Such capability could enable high-throughput screening and triage of large populations following a radiological incident, thereby accelerating first-responder decision-making and improving the timeliness of medical countermeasures.24 The standard “dilute-and-shoot” sample preparation entails five primary pretreatment steps: solvent addition, mixing, centrifugation, protein precipitation, and aliquot transfer.25 Furthermore, blood is typically separated into either serum or plasma before sample preparation.26 These sample processing steps require use of large volume of hazardous organic solvents and are typically performed in refrigerated equipment which inherently introduces analytical variation, sample loss, possible human error, and time delays during this process.
Green chemistry-based sample preparation for biomolecules extraction considers minimizing hazardous solvent use and energy consumption, require minimal sample volume, and maximize sample throughput.27–29 The practical applicability, simplicity and economic feasibility also the important parameters for the development of green analytical sample preparation methods.30–32 Our approach mainly relies on the elimination of the cold chain logistics for sample storage, miniaturized extraction platforms, and energy efficient methods for high-throughput large-scale sample analysis. In addition, in the context of a mass-casualty radiological scenario, the usefulness of a metabolomic study is constrained by the sample timing, cold-chain logistics, standardization, and biological confounders etc.33 Complications that can arise from cold chain management have been recently documented from the delivery of SARS-CoV-2 vaccinations, the need and maintenance of laboratory freezers for storage and transport during an emergency situation posed a great challenge.34 Radiation responsive metabolites could be transient and require faster sampling. Therefore, the ideal approach would be to develop a high-throughput sample collection strategy, preferring dried blood spots or other biofluids in a membrane that can stabilize the markers during transportation. In situ sample preparation can save energy and time for such events that require a large number of sample collection in an emergency response.35 Developing an advanced sample preparation technique for metabolomic analysis that can eliminate cold-chain logistics, reduce hazardous solvent use, and facilitate large scale in situ sample collection is a need for a radiological emergency or similar mass-casualty events.36
Fabric Phase Sorptive Extraction (FPSE)37 is an established sample preparation technique for the sorptive extraction of different compounds from various sample types such as, environmental,38–40 pharmaceuticals,41,42 biological,43,44 textile materials,45,46 food47,48 and non-invasive exhaled breath aerosol sample etc.49 A wide range of compounds, such as drug molecules,50–52 polycyclic aromatic hydrocarbons (PAHs),53,54 and toxic heavy metals,55,56 have been successfully extracted from different sample types mentioned above using this technique. However, small molecule biomarkers extraction from whole blood or serum using FPSE sample preparation method is still limited.44,57,58 The selectivity and extraction efficiency depends on the surface chemistry or sample absorption properties of FPSE membrane. The polymeric sorbent improves surface porosity, chemical functionality and hydrophilic properties of the membrane that help to better permeate the sample matrix through the FPSE membrane and increase the extraction efficiency of target metabolites.58 Since polymeric sorbent improves the thermal, mechanical, and even chemical stability of the FPSE membrane, it is an excellent medium for long time sample storage without any preservatives. Therefore, the FPSE method offers many advantages in the sample preparation that include: (i) simpler workflow, (ii) refrigeration, (iii) small volume of elution solvent required,59 (iv) use of any solvent for back extraction, (v) eliminating/minimizing sample pre-treatment steps, (vi) minimal sample volume required,60 and (vii) maximize biomolecular extraction efficiency from whole blood.61 We aim to utilize the FPSE method as a field deployable format to capture and stabilize analytes at the point of collection at ambient temperature to establish a fast, scalable, and low-cost sample processing platform for metabolomic analysis.
In this study, we expanded some features and the utility of FPSE to increase sample absorption and aid in the user-interface and efficiency of biofluid preparation for metabolomics based biodosimetry. We used polymeric sorbent phase sorptive extraction (SPSE) technique that uses thin film of sorbent on the fabric/paper substrates using variable concentrations of polymeric coatings to develop a field-deployable efficient extraction platform. This SPSE membrane can stabilize metabolites, maximize extraction efficiency, and eliminate drawbacks associated with sample preparation for rapid biomarker assessment. In the unexpected nuclear or radiological events, such as nuclear accidents or act of terrorism,62 there is an urgent need to rapidly and accurately collect the samples from a large number of individuals who are exposed to radiation. Therefore, it is important to categorize those exposed individuals into different dose and exposure types for the treatment plans. We previously found cellulose based membranes provided increased temporal stabilization of a blood-based radiation biomarker panel.44 The present study reports improvements with the development of novel cellulose membranes with more hydrophobic polymeric sorbent coatings to test in our standardized biodosimetry pipeline with additional types of biofluids (urine, whole blood, and serum). We characterized surface morphology, functionality, and hydrophilicity of novel sorbents. The utility of these new materials in MS-based biodosimetry analysis was assessed by both untargeted and targeted metabolomics using the murine model. Lastly, we developed a high-throughput 12-well plates sample preparation device that was developed using SPSE membrane considering a radiological emergency or such mass-casualty scenario to collect large number of blood samples in the point of collection. Overall, this technology would contribute to address some post-logistical gaps of a radiological emergency response, including reducing the need for specialized equipment in sample shipping and storage.
Prior to applying sorbent materials on the commercial cellulose fabric substrate, a pretreatment/mercerization process was performed to remove all the residual chemicals such as additives, dye, and fillers from the substrate. A piece of fabric (3 × 3 inches) substrate was soaked into deionized water followed by treatment with a 1 M NaOH solution for 1 h with sonication. The fabric was then washed with deionized (DI) water multiple times followed by soaking again into 0.1 M HCl solution for 1 h under sonication. The acid-treated fabric was washed with copious amount of DI water multiple times and dried at 70 °C in an argon environment. The cleaned cellulose fabric substrate was then dipped coated with different concentrations of nylon-6 (0.5%, 1.0%, 1.5%) solution with 3 dips (30 s dip and 30 minutes dry after each dip at room temperature). The membrane was then annealed at 70 °C in an argon medium overnight to evaporate solvent completely. To improve the hydrophilic interactions of nylon-6 coated SPSE membrane, we applied 60 s of oxygen plasma on the surface using Harrick PlasmaFlo. The SPSE membrane was placed inside the plasma chamber and cleaned the chamber with ultrahigh purity nitrogen gas, and purge oxygen gas before final pump down. The membrane was treated with oxygen plasma with low pressure (300 mtorr) at 25 W in high frequency (10 MHz) with medium flow rate. The cellulose paper membrane was prepared using a similar protocol without pretreatment steps.
MC and APTES grafted chitosan sorbents were applied in the same dip coating process with 1, 5, and 10 dips and annealed at 70 °C for 12 hours in an argon environment. All the membranes were then laser cut into 10 mm × 10 mm circular disc using versa laser for downstream biological sample processing. Fig. 1 shows the schematic illustration of sorptive phase extraction that indicates multi-steps process of SPSE membrane development and LC–MS/MS workflow for biomarker analysis. SPSE is a deployable technique which can be used for high throughput sample preparation in the radiological emergency with minimal sample volume (∼5 µL) with rapid, robust, and high trueness. The goal of these studies is to minimize logistical issues associated with the need for refrigerated equipment during sample delivery and storage, as was encountered during distribution of COVID-19 vaccinations.34
000×g, 4 °C). A 40 µL aliquot of the supernatant was recovered and diluted with 40 µL solvent (66% ACN whole blood and serum; 50% ACN urine) containing internal standards [final concentration (1 µM): carnitine (d3), TML (d9), sphinganine (d7); (2 µM): chlorpropamide, taurine (d4); (3 µM): hypoxanthine (d4), creatinine (d3), citrulline (d7), malic acid (d3), citric acid (d4)] and injected for LC–MS analysis. For comparison to biological samples, we deproteinated biofluids by mixing with cold solvent [whole blood or serum (5 µL, 66% ACN) or urine (10 µL, 50% ACN)], vortexed the samples, centrifuged for 10 min (10
000×g, 4 °C), and diluted a 40 µL aliquot of the supernatant with 40 µL solvent and internal standards as above. Optimization of extraction solvent and time was performed by applying a 10 µL standard solution [carnitine, taurine, TXB2, Sphingomyelin (d18:0/18:0), GlcCer (d18:124:1), Ceramide (d18:1/18:1), final concentration 1 µM] to 0% and 0.5% nylon-6 membranes, incubating them for 24 h at room temperature under vacuum, extracting with 50% ACN, 50% methanol, 40% ACN/40% methanol for 5, 10, and 30 min, and then comparing to a standard solution prepared by the above standard protocol.
Untargeted metabolomic profiling was used to test matrix interference and separation of biofluids from irradiated animals compared to sham-irradiated mice. SPSE processed samples were injected (2 µl) and analyzed by Waters Acquity Ultra Performance (UP) LC coupled to a Xevo® G2S quadrupole time-of-flight (QTOF) MS (Waters, Milford, MA, USA). The UPLC was equipped with an Acquity BEH C18 1.7 µm, 2.1 × 50 mm column kept at 40 °C for urine and 60 °C for whole blood and serum. Data were collected in both negative and positive electrospray ionization (ESI) data independent acquisition modes and were run with leucine enkephalin, 556.2771 [M + H]+ or 554.2615 [M − H−], used as LockSpray™. Mobile phase separation for urine was achieved using water/0.1% formic acid (FA) and ACN/FA. The flow rate was 0.5 ml min−1 with a gradient of 4.5 min 5% ACN/FA, 4.0 min 20% ACN/FA, 5.1 min 95% ACN/FA, and 1.9 min 5% ACN/FA. Mobile phase separation for whole blood and serum was achieved using water/FA, ACN/FA, isopropanol (IPA)/FA. The flow rate was 0.5 ml min−1 with a gradient of 4.5 min 2% ACN/FA 0% IPA/FA, 4.0 min 60% ACN/FA 0% IPA/FA, 1.5 min 98% ACN/FA 0% IPA/FA, 2.0 min 11.8% ACN/FA 88.2% IPA/FA, 0.5 min 50% ACN/FA 0% IPA/FA, and 1.0 min 2% ACN/FA 0% IPA/FA.
For targeted analysis, SPSE processed samples were injected (5 µl) and analyzed with a Waters Acquity UPLC coupled to a Xevo® tandem quadrupole (TQ-S) MS operating in multiple reaction monitoring (MRM) mode (Waters, Milford, MA, USA). The UPLC was equipped with an Acquity BEH C18 1.7 µm, 2.1 × 50 mm column kept at 40 °C for urine and an Acquity CSH C18 1.7 µm, 2.1 × 50 mm column 60 °C for whole blood and serum kept at 60 °C. MRM transitions were developed using the “IntelliStart” feature in MassLynx (Table 1) using nominal mass with the MS/MS collision energies between 10–20 eV. Mobile phase separation for urine was achieved using water/1 mM ammonium formate and ACN/FA. The flow rate was 0.5 ml min−1 with a gradient of 4.5 min 5% ACN/FA, 4.0 min 20% ACN/FA, 1.1 min 95% ACN/FA, and 1.1 min 5% ACN/FA. Mobile phase separation for whole blood and serum was achieved using water/ACN (1
:
1) FA and 1 mM ammonium formate, and IPA/ACN (9
:
1) FA and 1 mM ammonium formate. The flow rate was 0.45 ml min−1 with a gradient of 5.0 min 40% IPA/ACN (9
:
1) FA and 1 mM ammonium formate, 1.0 min 100% IPA/ACN (9
:
1) FA and 1 mM ammonium formate.
| Metabolite | Mode | Parent ion (m/z) | Daughter ion (m/z) | Dwell time (sec) | Cone (eV) | Collision (eV) | R2 | LOD (nM) | LOQ (nM) | RT (min) | Biofluid |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Carnitine | ESI+ | 162.3 | 85.0 | 0.003 | 34 | 20 | 0.99 | 0.2 | 0.6 | 0.40 | Urine/whole blood/serum |
| Carnitine (d3) | ESI+ | 165.0 | 85.0 | 0.025 | 2 | 16 | — | — | — | 0.39 | — |
| Taurine | ESI− | 126.0 | 44.0 | 0.003 | 24 | 14 | 0.98 | 12.2 | 48.8 | 0.37 | Urine/whole blood/serum |
| Taurine (d4) | ESI− | 130.0 | 48.2 | 0.025 | 70 | 14 | — | — | — | 0.36 | — |
| Citrulline | ESI+ | 176.0 | 69.9 | 0.003 | 16 | 20 | 0.82 | 19.5 | 39.1 | 0.58 | Whole blood/serum |
| Citrulline (d7) | ESI+ | 183.1 | 77.0 | 0.003 | 28 | 20 | — | — | — | 0.58 | — |
| SM | ESI+ | 731.6 | 184.2 | 0.003 | 52 | 27 | 0.83 | 31.3 | 62.5 | 5.34 | Whole blood/serum |
| Sphinganine (d7) | ESI+ | 309.3 | 273.4 | 0.003 | 80 | 16 | — | — | — | 1.53 | — |
| Chlorpropamide | ESI− | 275.1 | 126.1 | 0.025 | 52 | 30 | — | — | — | 7.68 | Urine |
| Chlorpropamide | ESI+ | 277.1 | 111.1 | 0.025 | 32 | 36 | — | — | — | 0.77 | Whole blood/serum |
| Creatinine | ESI+ | 113.8 | 86.1 | 0.003 | 46 | 14 | 0.94 | 9.8 | 39.1 | 0.46 | Urine |
| Creatinine (d3) | ESI+ | 116.9 | 89.1 | 0.003 | 2 | 10 | — | — | — | 0.47 | — |
| TML | ESI+ | 183.1 | 84.0 | 0.003 | 40 | 20 | 0.92 | 0.6 | 2.4 | 0.42 | Urine |
| TML (d9) | ESI+ | 198.1 | 130.1 | 0.025 | 4 | 14 | — | — | — | 0.42 | — |
| Hypoxanthine | ESI+ | 137.1 | 109.9 | 0.003 | 44 | 20 | 0.96 | 2.4 | 39.1 | 0.46 | Urine |
| Hypoxanthine (d4) | ESI+ | 141.2 | 113.0 | 0.003 | 25 | 18 | — | — | — | 0.47 | — |
To test for recovery and stability of standards applied to membranes, we applied standards at a final working concentration of 5 µM and left at room temperature under vacuum. Samples were collected at 1 day for recovery and at 1, 5, and 14 days for stability. For untargeted metabolomics, we inspected the raw data files in MassLynx v.4.1 (Waters Corporation, Milford, MA) before data pre-processing with progenesis QI (Nonlinear Dynamics, Newcastle, UK), which was used for peak alignment using software chosen QC chromatogram, peak picking, and normalization (normalize to all compounds functions). The data matrix was imported into the online software MetaboAnalyst 5.0 (log transformed and pareto scaled) to generate PCA plots.64 For targeted metabolomics, data files were processed in the software TargetLynx. We prepared a standard solution of all metabolites using a 4-fold dilution representing the dynamic range of 0.0005–10 µM to construct 10-point calibration curves. Calibration curves were fitted using the nonlinear regression function without the Y intercept constrained to zero. Values were plotted in GraphPad Prism 9.2.0 and checked for outliers (ROUT Q = 1%) (GraphPad Software, La Jolla, CA, USA). Receiver operating characteristic (ROC) curves were generated in MetaboAnalyst 6.0 to calculate the area under the curve (AUC) as a measure of the sensitivity and specificity.65
O) functional group at 1636 cm−1 (stretching bend) and C–N stretching at 1535 cm−1, indicating the successful coating on cellulose substrate. Other N–H stretching bands from amino groups might be overlapped on O–H stretching at 3298 cm−1. The FTIR peak intensity (Fig. 2A and B) related to the C
O and C–N stretching band increased as the concentration of nylon-6 sorbent increased in both cellulose fabric and paper substrate. Additionally, the paper membrane showed a greater adsorption of nylon-6 on the surface compared to the fabric substrate because of the greater peak heights (Fig. 2B) observed for similar stretching bands (C
O, C–N) at 1636 cm−1 and 1535 cm−1, respectively. The diverse chemical functionality presents in nylon-6 coated cellulose membranes are key components that can chemically bind metabolites from different biofluids. The membrane that contains different oxygen-containing groups such as –OH, –COOH, C
O, primary and secondary amines (–N–H, –NH2), and long hydrocarbon chain (–CH2–) can interact with both polar and non-polar biomarkers.
MC and APTES grafted chitosan membranes were characterized by FTIR, which are presented in the SI section (Fig. S1) to confirm the successful coating on the cellulose substrate. FTIR spectra of MC-coated membrane (Fig. S1A) showed characteristic carbonyl (C
O) stretching band at ∼1713 cm−1 and C–N stretching at 1560 cm−1, indicating successful coating on the cellulose surface. The hydroxyl peak in the range of 3200 to 3450 cm−1 for MC-coated membrane showed deformation because of N–H stretching from amino group overlapping in the same region also confirm the MC sorbent present in the surface. APTES functionalized chitosan coating on the cellulose substrate was identified in Fig. S1B that indicated the C–N stretching peak at 1560 cm−1. The sorbent materials coated on both cellulose fabric and paper substrates using dip coating technique were measured with respect to concentrations of the polymeric solution. The bar chart presented in Fig. 2C (Fabric) and 2D (Paper) indicates the amount of nylon-6 sorbent loaded into the cellulose substrate. To develop a nanoscale coating on the surfaces of both fabric and paper substrates, we used three different concentrations (0.5, 1.0, or 2.0%) of nylon-6 sorbent solution. It was observed that increasing the concentration of nylon-6 increased the amount of sorbent material (Fig. 2C and D) after coating on the surface of both fabric and paper substrates. The maximum amount of sorbent (1.5 mg cm−2) was loaded into cellulose fabric substrates using 2.0% of nylon-6 solution. However, the cellulose paper substrate showed better performance in capturing more nylon-6 sorbent on the surface compared to cellulose fabric substrate. The maximum amount of 5.16 mg cm−2 sorbent was loaded into cellulose paper substrate using 2.0% nylon-6 solution, whereas fabric substrate absorbed 1.5 mg using a similar concentration.
Both cellulose fabric and paper substrates are made up of cellulose microfibrils that create interstices, voids, and nanoscale pores on the surface. Dip coating polymeric sorbent solution on the surface allows the polymeric materials to permeate through the wall of cellulose microfibrils, and upon annealing, nanoscale pores can form on the coated materials. The morphological characterization was conducted using SEM for both bare cellulose substrates and SPSE membranes to confirm the presence of coating materials and the surface morphology. The SEM images of uncoated cellulose fabric substrate (Fig. 3, A1 and A2) and 0.5% nylon-6 coated cellulose fabric membrane (Fig. 3, B1 and B2) revealed differences in their morphology. The high magnification SEM images (Fig. 3, A2 and B2) clearly indicate the presence of coating materials on the cellulose fabric membrane and heterogeneous nanoscale porosity on the surface. The surface nanopores can increase the sample absorption properties, which increase the physicochemical interaction of metabolites to the membrane surface that maximize biomarker extraction. The similar surface characterization of cellulose paper membrane presented in SI section (Fig. S2) before and after coating. Fig. S2A2 indicates the cellulose paper membrane that is coated with 0.5% nylon-6 polymer, which shows clear distinctions from the uncoated cellulose paper substrate (Fig. S2A1). Surface analysis of both paper and fabric membranes were performed by Brunauer–Emmett–Teller (BET) N2 adsorption–desorption studies. Adsorption–desorption isotherm and pore volume vs. pore diameter graphs of paper (Fig. S2B1 and S2B2) and fabric membranes (S2C1, S2C2) are presented in Fig. S2 in the SI section. It is observed that the uncoated cellulose paper substrate has surface area of 1.034 m2 g−1, pore volume is 0.006 cm3 g−1 and average pore diameter 26. 05 nm, whereas nylon-6 coated paper membrane showed surface area 1.304 m2 g−1, pore volume 0.009 cm3 g−1, and average pore diameter 27.6 nm, respectively. Therefore, 20.7% of surface area and 33.3% pore volume increased after nylon-6 polymeric coating of SPSE paper membrane. On the other hand, cellulose fabric substrate showed slight decrease in surface area (2.54%) and pore volume (16.63%) after sorbent coating which might be the predominance of macropores present on the surface.
We previously demonstrated that biomarker release was related to its physicochemical adhesion, with higher efficiency observed for hydrophilic substrates (cellulose) compared to hydrophobic substrates (fiberglass).44 Therefore, to increase substrate adhesion, we applied 60 s oxygen plasma to activate oxygen-containing groups on the surface membrane. The plasma-activated membranes were studied by X-ray photoelectron spectroscopy (XPS) to confirm the presence of oxygen-containing groups on the surface. Though cellulose substrate itself contains hydroxyl groups and surface carboxylic groups, plasma treatment on the nylon-6 coated membrane increases the hydroxyl (OH) and carboxylic (COOH) groups that improve the hydrophilic properties, resulting in better sample absorption. XPS survey spectra are presented in the SI section (Fig. S3) to compare the carbon (C) and oxygen (O) ratios before and after plasma treatment. The intensity of the oxygen (O 1s) peak increased, and the carbon (C 1s) peak (Fig. S3) decreased after the oxygen plasma exposure, suggesting the presence of more oxygen species in the nylon-6 polymer chain (hydrocarbon).66 The oxygen content (atomic percentage) increased 35.29% after oxygen plasma on the surface of the membrane, whereas carbon content reduced 12.1% indicating the formation of oxygen-containing functional groups such as –OH, –COOH, and C
O etc. These newly formed C–OH, C–COOH, C–C
O groups improved the hydrophilic properties of the nylon-6 polymer-coated cellulose fabric membrane to facilitate the absorption biofluids. Fig. 4(A and B) shows the deconvolution of carbon (C 1s) core level spectra to be characterized further, specifying the different oxygen-containing functional groups on the membrane surface before and after plasma treatment activation. The XPS core-level carbon (C 1s) spectra of untreated nylon-6 membrane (Fig. 4A, left) indicate higher peak (C–C) intensity at 284.8 eV binding energy compared to the oxygen-containing carbon–oxygen peaks C–OH (286.49 eV), C
O (287.65 eV), and C–COOH (288.86 eV). On the other hand, the plasma-treated membrane (Fig. 4B, right) showed lower intensity in the C 1s peak (C–C) than oxygen-containing C–OH peak, confirming newly generated hydroxyl group on the surface along with carboxylic, and carbonyl groups upon plasma functionalization. Increasing polar functionalities on the membrane surface using the plasma-generated oxygen radicals improves the hydrophilic properties as well as increases the chemical interaction with the targeted metabolites from biofluids.
The hydrophilicity of the SPSE membrane was evaluated by water contact angle measurement, as presented in Fig. 4C. Hydrophilic materials show water contact less than 90° while hydrophobic materials show greater than 90°. The other factors that can change the water contact angle of a material are surface roughness, heterogeneity, and humidity, etc. The water contact angle of uncoated cellulose fabric substrate, 0.5% nylon-6 coated membrane without plasma treatment, and 0.5% nylon-6 coated membrane after 60 s oxygen plasma functionalization are 119.1°, 111.3°, and 59.5°, are shown in Fig. 4C. The plasma-treated membrane exhibited hydrophilic properties, as indicated by a water contact angle of ≤90°, compared to both the bare cellulose substrate and the untreated nylon-6 coated membrane. We also tested the membrane by studying the absorption capacity of biological samples (whole blood and serum), as presented in Fig. 4D. All plasma-treated nylon-6 coated fabric membranes showed excellent absorption capacity for blood and serum compared to the uncoated cellulose substrate. Similar sample studies were performed for cellulose paper membranes, which are presented in the SI section (Fig. S4). All the paper membranes (coated and uncoated) showed outstanding absorption properties for all the biofluids such as blood, urine, and serum samples (Fig. S4). Uncoated cellulose paper substrate showed better sample absorptivity compared to cellulose fabric, which might be because of the different arrangement of cellulose microfibrils in the substrate. In cellulose paper substrates, fibers are thin and loosely oriented, whereas in the fabric or cloth substrates, cellulose fibers are tight and thick. Both fabric and paper membranes (plasma-treated) absorbed whole blood, serum, and urine samples within 30 seconds of its exposure.
| 0% Nylon | 0.5% Nylon | |||||
|---|---|---|---|---|---|---|
| 50% ACN | 50% MeOH | 40% ACN/40% MeOH | 50% ACN | 50% MeOH | 40% ACN/40% MeOH | |
| Carnitine | 86 ± 4 | 84 ± 0 | 49 ± 2 | 73 ± 3 | 70 ± 5 | 43 ± 1 |
| Taurine | 71 ± 9 | 71 ± 2 | 68 ± 10 | 63 ± 2 | 65 ± 7 | 62 ± 4 |
| TXB2 | 109 ± 11 | 116 ± 14 | 105 ± 15 | 122 ± 20 | 98 ± 16 | 102 ± 14 |
| Sphingomyelin (d18:0/18:0) | 21 ± 3 | 0 ± 0 | 203 ± 24 | 14 ± 1 | 0 ± 0 | 178 ± 28 |
| GlcCer (d18:1/24:1) | 18 ± 3 | 1 ± 0 | 443 ± 20 | 13 ± 4 | 1 ± 0 | 335 ± 42 |
| Ceramide (d18:1/18:1) | 11 ± 5 | 2 ± 1 | 180 ± 18 | 7 ± 3 | 2 ± 0 | 136 ± 33 |
| 5 min | 10 min | 30 min | 5 min | 10 min | 30 min | |
|---|---|---|---|---|---|---|
| Carnitine | 102 ± 1 | 95 ± 6 | 80 ± 2 | 67 ± 6 | 68 ± 1 | 68 ± 2 |
| Taurine | 82 ± 3 | 66 ± 15 | 59 ± 3 | 67 ± 2 | 67 ± 5 | 62 ± 1 |
| TXB2 | 128 ± 4 | 112 ± 10 | 88 ± 21 | 103 ± 6 | 105 ± 4 | 98 ± 9 |
| Sphingomyelin (d18:0/18:0) | 87 ± 1 | 81 ± 8 | 73 ± 14 | 94 ± 6 | 78 ± 10 | 94 ± 6 |
| GlcCer (d18:1/24:1) | 47 ± 4 | 61 ± 9 | 59 ± 19 | 47 ± 4 | 46 ± 8 | 64 ± 7 |
| Ceramide (d18:1/18:1) | 64 ± 2 | 64 ± 11 | 68 ± 11 | 62 ± 6 | 64 ± 5 | 107 ± 4 |
Metabolite stability was tested using spiked standards on paper and fabric cellulose membranes with 0, 0.5, 1, and 2% nylon-6 coatings as a measure of the percent recovery at 1, 5, and 14 days at room temperature compared to a freshly prepared standard solution (Fig. 5). Consistent with our previous recovery measurements, we found >50% metabolite recovery for the nylon-6 membranes. As they maintained consistent stability over a 14-days period at room temperature this demonstrates that SPSE membranes can preserve small molecule radiation biofluid signatures and be useful in a radiological emergency scenario. Interestingly, higher recovery was observed for hypoxanthine compared to the freshly prepared standard solution that indicates removal of some potential matrix effects. Overall, similar performance is maintained irrespectively of paper or fabric substrates or percentage of nylon-6 coating application, further testing was restricted to 0.5% nylon-6 coated paper membrane since it showed high surface porosity and better sorptive properties compared to fabric membrane.
Untargeted metabolomics revealed minimal baseline noise is introduced when incorporating paper or fabric cellulose and nylon-6 membranes into standardized metabolomics sample preparation. Increased signal was observed in both ESI+ and ESI− between 2.2–3.5 min in whole blood and serum and 3.5–6.5 min in urine (Fig. 6), primarily for membranes with the nylon-6 sorbent (Fig. S5 and 6 shows chromatograms for the 13 min run for a whole blood sample). To determine how these minor artifacts could skew the spatial groups, we evaluated the untargeted sham and 8 Gy irradiated data sets using a PCA visualization tool. For serum, the first two vectors from a PCA plot accounted for 56.2% of the variation, while the first two vectors for urine accounted for 67.0% of the variation. Looking at the sham-irradiated samples in the PCA plot (Fig. 7), we observed that the membrane materials do shift the spatial distribution due to the presence of these artifacts. However, greater shifts in the radiation signature provide excellent separation at 1 day after 8 Gy IR exposure in the murine model.
Lastly, we utilized a targeted metabolomic assay to measure changes in biofluid metabolite concentrations and used ROC curve analysis to measure the area under the curve (AUC) and compare sensitivity and specificity between sample preparation methods. Urine, whole blood, and serum were collected from mice irradiated at 0, 2, 8, and 13 Gy at day 1 and 7 (0, 2, and 8 Gy). Nearly identical metabolite recovery and IR responses were observed in urine samples compared to urine applied to membranes (Fig. S7). For whole blood and serum, similar trends were observed for polar metabolites; however, wider variation was observed in sphingomyelin (SM) levels after membrane application indicating that extraction may vary among metabolite polarity. To test the sensitivity and specificity we combined a metabolite panel from eight metabolites found in urine (carnitine, TML, taurine, hypoxanthine) and blood (carnitine, citrulline, taurine, SM). The 0 and 2 Gy cohorts were combined as these groups both represent individuals needing minimal medical attention due to radiation injury when compared to an 8 or 13 Gy dose, which would represent an equivalent dose in humans to elicit hematopoietic or gastrointestinal syndrome. There was slightly lower performance for the membrane preparation protocol (AUC = 0.86) compared to the standard protocol (AUC = 0.94) when comparing the 8 Gy cohort to the combined 0 and 2 Gy cohorts at 1 day (Fig. 8). However, we found similar performance for both protocols when comparing the 13 Gy cohort to the combined 0, 2, and 8 Gy cohorts at 1 day (AUC = 0.75) and the 8 Gy cohort to the combined 0 and 2 Gy cohorts at 7 days (AUC = 0.97 and 0.94) (Fig. 8). With serum and urine biomarkers combined, slightly lower performance (AUC < 0.80) was observed for the 13 Gy cohort compared to the combined 0, 2, and 8 Gy cohorts at 1 day, but good to excellent results (AUC > 0.80) were achieved for the other groups and were comparable between our standard protocol and the membrane protocol (Fig. S8). It should be noted that there were sex specific differences in ROC performance, primarily with lower AUC values observed for the male 7 days 8 Gy vs. 0 and 2 Gy group and the female 1 day 8 Gy vs. 0 and 2 Gy group (Fig. S9).
A high-throughput point of care blood collection device was developed, assembling the SPSE membrane in a 12-well plate configuration (Fig. 9A). This point of care device can be used to collect samples (e.g., whole blood) using finger stick and stabilize biomarkers at room temperature for up to two weeks. A multi-configuration, material-type customizable transwell insert was designed with a microcapillary (Fig. 9B) at the center to transport a blood drop from a finger to the SPSE membrane that is assembled downstream (Fig. 9C). An orthographic section-rendering of the microcapillary in the transwell is presented in red that showcases the travel path and configuration (Fig. 9D). The design of the transwell strip was developed to self-collect the biofluid sample into the SPSE membrane while facilitating biomarker extraction directly in the same well plate device with solvent wash performed by automated instrumentation available in a standard clinical laboratory. All configurations of the insert have been designed for fabrication using injection molding for mass production or in the case of this study Stereolithography (SLA) resin printed using an autoclavable biocompatible Surgical Guide Resin (Form3B+, 100 µm layer thickness; FormLabs, MA, USA) and the capillary was chemically modified using UV-enhanced physisorption of heparin to make the surface more hydrophilic and prevent coagulation of the samples. Additionally, at the apical head of the capillary, the surface was modified using silane chemistry to form a hydrophobic inert coating, facilitating easy uptake of the sample into the microcapillary. Sorptive phase extraction membranes have been used previously as a biofluid sampler (dried blood spot card) to collect and prepare whole blood samples.67
The transwell has a secondary opening that can be used to collect the eluted biomarker solution for mass spectroscopy analysis. An image of 12 well plate sample collection device with 4× of the individual transwell is presented in Fig. 9E. The transwell can be used to collect any biofluid samples (whole blood, serum, urine) through the microcapillary into the SPSE membrane for metabolomic analysis. A blood-drop was collected using a lancet from a finger and extended to the capillary opening to allow blood to be drawn by the capillary force and absorbed into the membrane (Fig. 9F). The developed transwell insert showed excellent performance to deliver the blood sample into the membrane and an image of membranes before and after blood absorption is presented in Fig. 9G, indicating the collection of blood through the microcapillary. Our goal is to upgrade this device up to 96 well plate which can even faster the sample collection process after a radiological event.
Supplementary information (SI): FT-IR spectra; SEM images; XPS spectra; biofluid absorption; biomarker recovery; ESI chromatogram; and ROC curves, etc. See DOI: https://doi.org/10.1039/d6ra00832a.