Open Access Article
Wanyan Wu†
ab,
Yan Guo†bc,
Lidan Dengb,
Wenwu Gongb,
Shunyu Huab,
Fen Liub and
Chang-ye Hui
*bc
aDepartment of Toxicology, School of Public Health, Southern Medical University, Guangzhou, China
bShenzhen Prevention and Treatment Center for Occupational Diseases, Shenzhen, China. E-mail: hcy_sypu@hotmail.com
cSchool of Public Health, Southern Medical University, Guangzhou, China
First published on 7th January 2026
We developed an enzyme-cascade-amplified whole-cell biosensor for sub-nanomolar methylmercury (MeHg) detection in environmental waters. By integrating a MerB-based organomercurial lyase with a water-soluble indigoidine reporter system in a decoupled genetic circuit, we achieved a limit of detection (LOD) of 0.04 nM and a linear range of 0.02–1.22 nM. Systematic comparison of intracellular versus surface-displayed MerB strategies revealed that intracellular expression (TOP10/pCon-IND-C-B) provided superior sensitivity. The biosensor demonstrated robust performance across tap water, surface water, and seawater matrices (R2 > 0.98) with negligible matrix effects, enabling direct colorimetric quantification of culture supernatants. Notably, metabolic burden from pigment synthesis dominated the dose-response profile rather than MeHg cytotoxicity. This biosensor offers a cost-effective alternative for screening MeHg in environmental water, with detection limits that far exceed WHO drinking water guidelines and a potential for high-throughput environmental monitoring.
Current methods for mercury speciation rely heavily on instrumental techniques such as liquid chromatography–inductively coupled plasma mass spectrometry.5,6 While sensitive, these approaches require expensive instrumentation, complex sample preparation (e.g., extraction, derivatization, separation), and specialized expertise, which limits their applicability for frequent, on-site environmental screening. In response, biosensors have emerged as promising alternatives, offering rapid, cost-effective, and field-deployable detection.7 Whole-cell biosensors, particularly those engineered around the natural mer operon, have been extensively explored for Hg(II) sensing.8 However, conventional reporters, such as fluorescence and bioluminescence, suffer from high detection thresholds, a lack of visual readout, and limited sensitivity for ultratrace analysis.9–11 Pigment-based reporting systems address these limitations through two key advantages: (i) enzymatic pigment synthesis provides cascade amplification, enhancing sensitivity; and (ii) colorimetric detection via visible-light spectrophotometry eliminates the need for sophisticated instrumentation while enabling high-throughput screening.12,13 Water-soluble pigments, such as the blue pigment indigoidine, further simplify analysis by allowing for the direct quantification of cell-free supernatants without the need for organic extraction.14–16
The mer operon encodes several proteins critical for mercury detoxification, including the organomercurial lyase MerB, which cleaves the carbon–Hg bond in MeHg to release Hg(II).17 Since the mer regulator MerR specifically responds to Hg(II), but not MeHg, the incorporation of MerB into biosensor circuits enables the indirect detection of organomercury species through this conversion step.3 Previous studies have leveraged MerA (mercuric reductase) and MerB to expand the detection spectrum and improve bioremediation capacity.18,19 Nevertheless, the integration of these modules with strong, water-soluble pigment reporters to enable convenient and effective detection remains inadequately explored.
Herein, we report an enzyme-cascade-amplified whole-cell biosensor that merges MerB-based MeHg cleavage with indigoidine synthesis for sub-nanomolar colorimetric detection. By optimizing a decoupled genetic circuit and incorporating the Hg(II) transporter MerC, we achieved a limit of detection (LOD) of 0.04 nM and a quantitative range of 0.02–1.22 nM. Critically, we systematically compared intracellular versus surface-displayed MerB strategies, revealing that intracellular expression provides superior sensitivity. The biosensor demonstrated negligible matrix effects in tap water, surface water, and seawater, validating its potential for practical environmental monitoring. This work advances the development of low-threshold tools for MeHg screening and highlights the importance of metabolic matching in engineering high-performance whole-cell biosensors.
| Strains and plasmids | Genotypes or description | Ref. |
|---|---|---|
| E. coli strain | ||
| TOP10 | F− Φ80 lacZΔM15 ΔlacX74 recA1 | Invitrogen |
| Vectors | ||
| pCon-IND | pHg-IdgS-Sfp derivative with a constitutive PceuR promoter inserted in front of the merR gene | 21 |
| pCon-IND-C | pCon-IND derivative containing a constitutive PJ23119-controlled merC gene inserted as a SacI-HindIII fragment | 21 |
| pCon-IND-C-B | pCon-IND-C derivative containing a constitutive P637-controlled merB gene inserted as a HindIII-XhoI fragment | This study |
| pCon-IND-C-surB | pCon-IND-C derivative containing a constitutive P637-controlled inaPb-merB gene inserted as a HindIII-XhoI fragment | This study |
Safety Warning: methylmercury (MeHg) is an extremely potent neurotoxin that can be absorbed through inhalation, dermal contact, or ingestion.20 All manipulations must be performed in a fume hood, and personnel are required to wear nitrile gloves, safety goggles, and a lab coat. Waste solutions containing MeHg must be treated with sodium sulfide to generate HgS precipitate prior to disposal as hazardous waste. All experiments in this study were conducted in a laboratory meeting Biosafety Level 2 (BSL-2) standards.
To generate pCon-IND-C, a constitutive PJ23119-controlled merC gene (encoding the Hg(II) transporter MerC) was inserted downstream of the reporter cassette as a SacI-HindIII fragment.
For methylmercury sensing, two MerB expression modules were constructed at the HindIII-XhoI site downstream of the merC cassette: (i) an intracellularly expressed MerB under the constitutive P637 promoter (pCon-IND-C-B), and (ii) a surface-displayed MerB fused with the ice nucleation protein InaPb (pCon-IND-C-surB).
All biosensing plasmids were transformed into E. coli TOP10 via chemical transformation to generate whole-cell biosensors for subsequent characterization.
000 rpm for 10 min, and the supernatants were filtered through a 0.22 µm membrane filter (Millipore, Billerica, MA, USA) to remove particulates and microorganisms.
We prepared sensing culture systems by mixing four types of water with 10× LB broth, 10× NaCl-free LB broth, and sterile water in specific volume ratios, as previously described in Table 2.22,23
| Component | Deionized water | Tap water | Surface water | Seawater |
|---|---|---|---|---|
| Water samples | 90% | 90% | 90% | 50% |
| 10× LB | 10% | 10% | 10% | — |
| 10× NaCl-free LB | — | — | — | 10% |
| Sterile water | — | — | — | 40% |
Overnight cultures of TOP10/pCon-IND-C-B were inoculated at a 2% (v/v) concentration into four sensing-culture systems, each supplemented with 50 µg mL−1 of ampicillin. MeHg was introduced through a two-fold serial dilution to achieve final concentrations of 0, 0.02, 0.03, 0.06, 0.13, 0.25, 0.5, and 1 nM. Following incubation at 30 °C with shaking at 150 rpm for 4 hours, we measured bacterial density and pigment signals.
The sensing principle operates as follows: in the absence of Hg(II), dimeric MerR binds to Pmer and represses transcription of the downstream genes.24 Upon exposure to MeHg, either intracellularly expressed or surface-displayed MerB catalyzes the cleavage of MeHg to Hg(II), which binds MerR and converts it into a transcriptional activator, which activates Pmer, initiating transcription of the idgS-sfp cluster. The expressed Sfp phosphopantetheinyl transferase activates apo-IdgS to holo-IdgS, which catalyzes the condensation of two L-glutamine molecules to produce the water-soluble blue pigment indigoidine.25 This pigment exhibits a characteristic absorption maximum at 610 nm, enabling colorimetric quantification.21
We hypothesized that anchoring InaPb-MerB fusion protein on the outer membrane would facilitate extracellular cleavage of MeHg, thereby preventing the cytotoxicity associated with intracellular accumulation of MeHg and enabling detection at elevated concentrations.
Unlike Hg(II), which is actively imported via MerT/MerP and MerC transporters, MeHg uptake does not rely on mer operon-encoded transport systems.8,26 This is likely attributed to its physicochemical properties, including a small molecular size and higher lipophilicity, which enable effective membrane permeation without the need for transporter assistance.27 MeHg is sufficiently lipophilic to traverse the E. coli membrane via simple diffusion, particularly when complexed with monovalent anions (e.g., Cl−) present in LB medium.3
Since dimeric MerR exclusively responds to Hg(II), the observed pigment signal upon high-concentration MeHg exposure can be rationalized as follows. The native E. coli TOP10 strain lacks the merB gene, excluding the possibility of endogenous MeHg cleavage.3 First, trace impurities of inorganic Hg(II) may persist in the MeHg stock solution despite its high purity. Second, ambient light exposure during storage or incubation could induce spontaneous photodegradation of MeHg, generating trace Hg(II).2 These sources of Hg(II) likely account for the pigment response observed at concentrations greater than 312.5 nM, where cytotoxicity severely suppresses biosensor growth, leading to rapid signal attenuation and the unimodal response profile.4 The invariant 610 nm peak across all concentrations indicates that indigoidine maintains a stable chemical structure in the supernatant, which is advantageous for accurate colorimetric quantification.
Compared with TOP10/pCon-IND-C, the intracellular expression of MerB in TOP10/pCon-IND-C-B shifted the pigment signal peak leftward to ∼2.44 nM, representing a two-order-of-magnitude improvement in detection limit relative to the parent sensor (peak at ∼312.5 nM). This demonstrates that intracellular MerB efficiently cleaves MeHg to Hg(II), which subsequently activates the pigment reporter system. Unlike TOP10/pCon-IND-C, where growth inhibition at concentrations greater than 19.6 nM resulted directly from MeHg cytotoxicity, the decrease in density in TOP10/pCon-IND-C-B at concentrations as low as 0.08 nM was mainly due to the metabolic burden from indigoidine synthesis.16 Possibly forming a “metabolic stress-pigment attenuation” peak. The incorporation of the intracellular cleavage module lowered the LOD to 0.04 nM. This sub-nanomolar detection limit, enabled by enzymatic signal amplification, is substantially superior to previously reported whole-cell biosensors for organomercury detection.3,4
Unlike TOP10/pCon-IND-C-B with intracellular MerB expression, TOP10/pCon-IND-C-surB employs surface-displayed InaPb-MerB fusion proteins to catalyze extracellular MeHg cleavage.28 Then, the resulting Hg(II) is transported into the cell via MerC to activate the pigment reporter system.26 However, compared with the intracellular strategy, surface display yielded lower pigment signal intensity and a modestly elevated LOD (0.08 nM versus 0.04 nM), failing to achieve the anticipated sensitization effect. This is likely because free diffusion of MeHg into cells provides direct access to intracellular MerB in the pCon-IND-C-B system. In contrast, exclusive reliance on surface-displayed MerB limits the amount of MeHg that can be processed, even with MerC-assisted Hg(II) transport.
The progressive genetic design (Fig. 5A) enabled direct comparison of how module incorporation affected the quantitative detection range (Fig. 5B). TOP10/pCon-IND-C (lacking MerB) showed a robust nonlinear relationship in the high-concentration range of 19.53–312.5 nM (R2 = 0.957), responding to trace Hg(II) impurities and photodegradation products. Both TOP10/pCon-IND-C-B and TOP10/pCon-IND-C-surB, equipped with intracellular and surface-displayed MerB, respectively, exhibited strong nonlinear relationships in the 0.02–1.22 nM range (R2 = 0.977 and 0.954) (Fig. 5B). Statistical significance (two-tailed t-test, P < 0.05) was achieved at concentrations of 19.53, 0.02, and 0.15 nM for TOP10/pCon-IND-C, TOP10/pCon-IND-C-B, and TOP10/pCon-IND-C-surB, respectively, yielding calculated LODs of 19.53, 0.04, and 0.08 nM.
As shown in Fig. 5, our progressive genetic optimization has significantly improved sensitivity. To provide context for our biosensing performance, we have included a comparative analysis with established analytical techniques in Table 3. While methods such as HPLC-ICP-MS achieve lower detection limits (0.003–0.028 nM), our platform's LOD of 0.04 nM is remarkably competitive considering its operational simplicity and cost-effectiveness. While individual sample detection requires longer incubation times than instrumental methods, the 96-well plate format allows for high-throughput parallel screening of numerous samples with minimal manual intervention. The low instrumental barrier and field-deployable nature make this platform especially suitable for large-scale preliminary screening and source tracking, acting as a practical complement to gold-standard instrumental techniques for analyzing trace MeHg in environmental monitoring.
| Analytical technique | LOD (nM) | Linear range (nM) | Advantages | Limitations | Ref. |
|---|---|---|---|---|---|
| a MSPE: magnetic solid-phase extraction; ICP-MS: inductively coupled plasma mass spectrometry; SPE: solid-phase extraction; EC-CVG-AFS: electrochemical cold vapor generation atomic fluorescence spectrometry. | |||||
| Whole-cell biosensor | 0.04 | 0.02–1.22 | Simple, low-cost, field-deployable | Longer detection time, cell culture required | This study |
| MSPE-HPLC-ICP-MS | 0.003 | 23.2–23,191 | High sensitivity, wide linear range | Expensive equipment, high operating costs | 29 |
| Online SPE-HPLC-ICP-MS | 0.028 | 0.23–232 | High automation, fast analysis | Complex system, high maintenance costs | 30 |
| Electrochemical sensor | 0.23 | 4.6–116 | Real-time monitoring, portable | Lower sensitivity | 31 |
| EC-CVG-AFS | 0.009 | 1–24.9 | Low detection limit | Cumbersome operation | 32 |
Collectively, TOP10/pCon-IND-C-B, which features intracellular MeHg cleavage, achieved an optimal balance between sensitivity and a linear quantitative range. Therefore, this sensor was selected for subsequent evaluation of trace MeHg detection in complex environmental water samples.
Fig. 6C demonstrates that TOP10/pCon-IND-C-B maintained a consistent quantitative detection range (0.0625–1 nM) across all water matrices, including seawater, with highly overlapping response curves (R2 > 0.98). Dissolved organic matter in freshwater and high concentrations of inorganic ions in seawater did not significantly interfere with MeHg recognition, Hg(II) transport, or indigoidine synthesis, indicating negligible matrix effects, which were also observed in whole-cell sensing of other toxic metals.33,34 Visually discernible blue color intensification was observed in all water samples with increasing MeHg concentration (Fig. 6D and E).
The selectivity of this biosensing platform is primarily determined by the high specificity of dimeric MerR for Hg(II), which has been well documented for its ability to distinguish between other metal ions, such as Cd(II), Pb(II), Zn(II), and Cu(II).8,18,19 This molecular recognition specificity ensures minimal interference from common cations in environmental waters. Since MerB catalyzes the cleavage of the C–Hg bond in various organomercurials,3 our biosensor theoretically functions as a broad-spectrum detector for organomercury species beyond MeHg, including ethylmercury and phenylmercury. The primary determinant of response magnitude is the bioavailability of these organomercurials rather than MerB substrate specificity. Consequently, interference from dissolved organic matter or competing ligands is expected to be limited to their effect on MeHg uptake efficiency. In contrast, direct interference with the Hg(II)-MerR binding event remains negligible. This inherent selectivity profile makes the platform particularly robust for screening mercury species in complex matrices.
While the WHO drinking water guideline for MeHg is set at 30 nM, the sub-nanomolar detection range of our biosensor (0.02–1.22 nM) is highly relevant for environmental monitoring because ambient MeHg concentrations in most natural water bodies are significantly lower than regulatory thresholds, and MeHg in uncontaminated or lightly contaminated waters is far below the 30 nM safety limit.1,2 Moreover, MeHg exhibits strong bioaccumulation and biomagnification, whereby even picomolar water concentrations can amplify through food webs to reach toxic levels in higher trophic organisms including humans, making monitoring far below safety limits essential for understanding exposure risks and implementing preventive measures.35 Sub-nanomolar sensitivity also enables early pollution detection and source tracking, allowing identification of emerging contamination sources before they escalate to hazardous levels, which is particularly crucial for industrial discharge monitoring and watershed protection.
Metabolic burden from indigoidine synthesis might impose a significant growth penalty, limiting the quantitative range.21 However, for high-concentration samples from pollution events, simple 1
:
10 or 1
:
100 dilution brings them into the linear range, an operation far simpler than the extraction/derivatization required by traditional methods and still compliant with green analytical chemistry principles.36 In contrast, an excessively wide dynamic range would sacrifice sensitivity and resolution at the low-concentration end, hindering early detection of trace contamination. We acknowledge that this range has limitations for compliance testing, but in most environmental monitoring scenarios, a narrow yet sensitive range is more valuable than a wide but insensitive one.
The observed trade-off between bacterial growth inhibition and pigment signal response is a common phenomenon in whole-cell biosensing, particularly when using fluorescent or chromoprotein reporters.37,38 In such systems, increasing analyte concentrations often simultaneously induce reporter synthesis and cellular toxicity, leading to signal attenuation at high doses. It typically necessitates calculation of normalized responses (e.g., relative fluorescence units divided by OD600), which complicates data processing and introduces additional variability from dual-parameter measurements.4 In our indigoidine-based system, while both metabolic burden from pigment synthesis and MeHg cytotoxicity contribute to growth inhibition, the dose-response relationship remains quantitatively reliable within the ascending phase of the parabolic curve (0.02–1.22 nM). Within this sub-nanomolar range, the decrease in OD600 is minimal (<15%) and correlates linearly with MeHg concentration, allowing direct use of A610 values without normalization. This simplification enhances data analysis and lowers experimental uncertainty, strengthening the biosensing platform for field-deployable screening applications.
Footnote |
| † Wanyan Wu, and Yan Guo contributed equally to this work. |
| This journal is © The Royal Society of Chemistry 2026 |