Magnetic resonance imaging (MRI) is a critical clinical diagnostic tool, but conventional 1H MRI suffers from background signal interference. 19F MRI offers a promising alternative with near-zero background noise, yet its progress depends on high-performance fluorinated probes. Existing probes often struggle to balance fluorine content, water solubility, and relaxation efficiency. To address this issue, we developed a metal–organic cage (MOC) as a platform for a novel dual-modal MRI nanoprobe. Using copper ions as paramagnetic nodes and fluorinated oligoethylene glycol as ligands, we synthesized a structurally precise and water-soluble nanoprobe, MOC-F. This nanoprobe not only significantly enhances the 1H relaxation rate through its paramagnetic copper centers, enabling high-contrast 1H MRI (r1 = 25.96 mM−1 s−1 at 298 K, 0.5 T), but also promotes an increased relaxation rate for 19F nuclei via paramagnetic relaxation enhancement (a T2/T1 ratio of 0.86 at 500 MHz), thereby facilitating efficient 19F MRI. Both in vitro and in vivo studies confirmed the outstanding dual-modal imaging performance and good biocompatibility of MOC-F. This work not only presents an innovative design strategy for developing highly sensitive 19F MRI probes but also highlights the considerable potential of MOCs in biomedical imaging, opening new avenues for the development of advanced smart MRI nanoprobes.
Unlike 1H MRI, the performance of 19F MRI depends critically on the use of exogenous fluorinated contrast agents. Existing 19F MRI probes are primarily categorized into five types—perfluorocarbons (PFCs), small fluorinated molecules (SFMs), fluorinated ionic liquids (FILs), fluorinated polymers, and inorganic fluoride nanoparticles (IFNPs)—each facing inherent trade-offs among fluorine content, water solubility, and relaxation efficiency.26 PFCs offer high fluorine content but suffer from spectral complexity and hydrophobicity requiring encapsulation.27 SFMs enable modular design, yet are limited by low fluorine content.28–31 FILs combine solubility with high fluorine content but exhibit rapid clearance and limited chemical diversity.32,33 Fluorinated polymers balance hydrophilicity and fluorine loading at the expense of chain rigidity and dispersibility,34–38 while IFNPs, despite their high fluorine density, demand specialized UTE sequences due to extremely short T2 relaxation.39 Collectively, existing 19F MRI probes are limited by trade-offs among sensitivity, design flexibility, water solubility, and imaging compatibility.40,41 There is a clear and urgent need for a new generation of 19F probes that combine high fluorine content, excellent biocompatibility, and good water solubility with the capacity for precise molecular-level design and regulation.
Recent advances in supramolecular chemistry, particularly metal–organic cages (MOCs), have opened new avenues for probe design.42 MOCs are discrete, precise nanostructures with tunable cavities formed via coordination-driven self-assembly.43–45 Their structural symmetry, functional versatility, and porosity enable diverse biomedical applications, yet current MOC-based imaging is largely limited to fluorescence, PET, and 1H MRI.46–48 Extending MOCs to emerging modalities therefore represents a promising frontier.
In response to current challenges and opportunities, we propose a MOC-based platform for constructing a novel 1H/19F dual-modal MRI probe. Using paramagnetic Cu2+ nodes49 and fluorinated oligoethylene glycol-modified isophthalic acid ligands, we assembled a water-soluble, structurally precise cuboctahedral nanoprobe (MOC-F) via coordination-driven self-assembly. This system preserves the structural advantages of MOCs while enabling synergistic enhancement of both 1H and 19F MR signals, as shown in Scheme 1. By integrating structural precision with functional tunability, MOC-F provides a versatile and efficient nanoplatform for high-contrast biological imaging and structure–performance research.
![]() | ||
| Scheme 1 Structurally precise and water-soluble MOC-F nanoprobe with an enhanced 1H/19F dual-modal MRI signal. | ||
Fourier transform infrared (FT-IR) spectroscopy provided crucial evidence for successful coordination and fluorine source introduction. As shown in Fig. 1c, both L-F and MOC-F exhibit characteristic absorption peaks at 1723.4 cm−1 and 1734.6 cm−1 originating from the ester carbonyl group in the fluorinated structure, while this signal is completely absent in L or MOC, clearly demonstrating the successful introduction of the fluorine source. Furthermore, the characteristic stretching vibrations of the carboxyl group in ligands L and L-F occur at 1594.2 cm−1 and 1596.1 cm−1, respectively. Following the formation of the cage structure, MOC exhibits two sets of absorption bands at 1587.9 cm−1 and 1378.1 cm−1, while MOC-F shows two sets at 1588.8 cm−1 and 1377.3 cm−1. These are attributed to the stretching vibrations of the carboxylate ions, indicating coordination between the carboxyl group and the copper ion. Notably, both MOC and MOC-F exhibit characteristic Cu–O bond peaks at 487.8 cm−1 and 489.9 cm−1, respectively, providing further corroboration for the formation of the coordination structure.
Based on the confirmed molecular structure and the established tendency for isophthalic acid and copper ions to form coordination bonds within the Cu24L24 cuboctahedral skeleton, coupled with the high molecular weight characteristics of the ligands and MOCs, we systematically investigated the molecular weight, stability, homogeneity, and polydispersity of the assembly via size exclusion chromatography (SEC). Fig. 1d shows a single sharp peak for L and L-F at approximately 20.48 and 19.86 min, while peaks for MOC and MOC-F appeared at 18.43 and 18.05 min, respectively. The shorter elution times of MOC and MOC-F reflect their larger hydrodynamic volumes compared to the respective ligands. The molecular weights of MOC and MOC-F determined by SEC were 10
857 and 17
832 g mol−1, respectively, approximately matching the theoretical values. Additionally, all four samples exhibited extremely narrow polydispersity indices of 1.07, 1.06, 1.05, and 1.04, respectively, indicating low dispersity and structural stability. We further measured the molecular weight of MOC-F via matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry, observing a peak at 21
537 g mol−1 consistent with the theoretical value, thereby confirming the formation of the structurally expected MOC-F (Fig. 1e). Based on the molecular formula, each MOC-F contains 72 fluorine atoms residing in identical chemical environments, corresponding to a fluorine content of approximately 6.36 wt%. Detailed information regarding the comparative samples L-F and MOC is presented in Table S1.
Benefiting from the oligoethylene glycol (OEG) linkages between the fluorine moiety and the MOC core, MOC-F exhibits exceptional water solubility, reaching up to 50 mg mL−1. For subsequent bioapplication experiments, we employed dynamic light scattering (DLS) and zeta potential measurements to determine the size and surface charge of MOC-F. The results indicate that MOC-F exhibits a hydrated particle size of approximately 6.3 nm with a uniform distribution (PDI = 0.24) (Fig. 1f), carries a weak negative surface charge (−5.5 mV) (Fig. S18), and remains stable for 12 hours. To further evaluate its stability in physiological environments, we dissolved MOC-F in PBS buffer (pH 7.4) or fetal bovine serum (FBS) and retested after two days (Fig. S19). The DLS results showed that its size remained consistent with the aforementioned values, confirming the good structural integrity and stability under simulated physiological conditions. Finally, we visually characterized the nanoscale morphology of MOC-F using transmission electron microscopy (TEM) and atomic force microscopy (AFM). TEM imaging showed that MOC-F possesses a nearly spherical morphology with a uniform, ultrasmall size of approximately 7 nm (Fig. 1g). A representative AFM image (Fig. 1h) clearly reveals that the synthesized MOC-F consists of discrete, uniformly distributed nanoparticles. The average height of these particles was determined to be approximately 7.5 nm. These data align closely with the theoretical dimensions of Cu24L24 cuboctahedral MOC, not only experimentally confirming its successful synthesis and structural integrity but also providing crucial evidence for its excellent structural stability and narrow size distribution.
. However, during MRI signal acquisition, the SNR is also influenced by the transverse relaxation time T2. Specifically, the average signal amplitude (Savg) obtained per acquisition, and consequently the overall SNR, is positively correlated with T2, i.e., SNR ∝ e−1/T2.53 Research has demonstrated that introducing chemical reagents to accelerate relaxation processes can simultaneously shorten both T1 and T2, though the extent of shortening may not be consistent between the two. If T2 is excessively shortened, the resulting decrease in Savg may outweigh the gain from increasing the number of average repetitions (NEX), leading to signal “quenching”. Conversely, if T2 is moderately reduced and the increase in NEX compensates for the loss in Savg, signal enhancement may be achieved. Generally, a T2/T1 ratio greater than 0.4 is considered a practical criterion for effective relaxation sensitization.54 Paramagnetic metal ions are frequently employed as contrast enhancers, with copper ions exhibiting particularly strong paramagnetic properties. This constitutes one rationale for selecting copper as the metallic node within the MOC structure in this study. We hypothesise that the copper ion at the MOC centre modulates the relaxation times of surrounding 1H nuclei of water molecules and the 19F nuclei at the MOC periphery through interactions with both the inner and outer spheres. This synergistically enhances 1H/19F magnetic resonance signals (Fig. 2a). To preliminarily evaluate the effect of Cu2+ on 19F relaxation, we compared the 19F NMR spectra of the L-F (without Cu2+) and the assembled MOC-F (with Cu2+). As anticipated, the 19F NMR spectrum of MOC-F exhibits a marked broadening of the chemical shift lines, confirming that the potential interaction between the Cu2+ ion and the 19F nucleus results in accelerated relaxation (Fig. 2b).
![]() | ||
| Fig. 2 (a) Relaxation modulation of paramagnetic ions on fluorinated moieties and water molecules. (b) 19F NMR spectra of the free L-F ligand and hinged L-F on MOC-F containing Cu2+ ions. | ||
Therefore, we investigated the performance of L-F, MOC, and MOC-F as potential 1H MRI contrast agents at 0.5 T. We first measured the relaxation times of the three samples at different concentrations. As shown in Table S2, the T1 and T2 of the 1H nuclei in L-F are in the thousands of milliseconds range, while the T1 and T2 of the 1H nuclei in both MOC samples are shortened to varying degrees. The shortened T1 and T2 of 1H in both MOC and MOC-F are attributed to the presence of paramagnetic Cu2+ within MOCs. With increasing sample concentration, the 1H T1 and T2 of MOC and MOC-F samples shortened significantly, while those of the L-F sample showed no noticeable change (Fig. S20). Simultaneously, the introduction of the paramagnetic metal ion Cu2+ into the probe consistently maintained the T2/T1 ratio within an appropriate range (>0.66), confirming the successful sensitization of this probe class. Since this probe exhibits a marked shortening of longitudinal relaxation time due to the presence of Cu2+, which maintains the T2/T1 ratio within an optimal range to enhance the SNR, it can be classified as a T1-weighted probe. Plotting the reciprocal of the relaxation time versus concentration yields longitudinal and transverse relaxation efficiencies of 26.13 and 33.68 mM−1 s−1 for MOC, respectively, while MOC-F exhibits longitudinal and transverse relaxation efficiencies of 25.96 and 34.18 mM−1 s−1, respectively, markedly higher than those of L-F (Fig. 3a). These results indicate that the 1H nuclear relaxation processes in both MOC samples are significantly accelerated. This acceleration stems from the pronounced paramagnetic relaxation enhancement (PRE) effect imparted by the paramagnetic Cu2+ ions. This effect also lays the foundation for enhancing the SNR and image quality in MRI using such probes.
Based on the above results, we further conducted in vitro 1H MRI imaging experiments on aqueous samples of L-F, MOC, and MOC-F to visually evaluate their imaging performance. As shown in Fig. 3b, the imaging results clearly demonstrate that both MOCs (MOC and MOC-F) produce MRI with a high SNR and uniform signal distribution. Their signal intensity exhibits significant dependence on probe concentration—with a regular increase in signal intensity as the probe concentration increases. This phenomenon aligns with typical relaxation-enhanced mechanisms, further confirming the crucial role of paramagnetic components in enhancing imaging contrast. In contrast, under identical imaging conditions, the L-F aqueous solution without paramagnetic metals produced only an extremely weak signal, with imaging performance significantly inferior to MOC and MOC-F. This intergroup discrepancy powerfully demonstrates that the paramagnetic centers provided by Cu2+ ions in the structure are the fundamental cause of effective MRI signal generation. The significant relaxation enhancement effect they induce not only substantially increases imaging sensitivity but also enables MOC and MOC-F to exhibit outstanding signal response characteristics in aqueous environments, thereby revealing their immense potential as high-performance 1H MRI probes.
Similarly, to investigate the feasibility of MOC-F for 19F MRI, we compared the 19F nuclear relaxation behavior of L-F and MOC-F. The results indicate that both 19F nuclear T1 and T2 of MOC-F in aqueous solution decrease significantly with increasing concentration (Fig. S21), with a T2/T1 ratio exceeding 0.86 (Table S3 and Fig. S22–24). This classification confirms its status as a T1-weighted 19F MRI probe. In contrast, the relaxation times of L-F exhibit no apparent concentration dependence. Furthermore, quantitative calculations revealed longitudinal and transverse relaxation efficiencies of 0.261 and 0.288 mM−1 s−1 for MOC-F, respectively (Fig. 3c). Compared to the ligand (Fig. S22 and Table S4), these relaxation efficiencies exhibited an order-of-magnitude enhancement, highlighting the crucial role of metal coordination in boosting fluorine nuclear relaxation efficiencies.
During in vitro 19F MR imaging experiments, we conducted comparative studies using fluorine atom concentration as the standard. Since MOC-F contains 72 fluorine atoms and L-F contains 3 fluorine atoms, the molecular concentration of L-F is three times that of MOC-F at the same fluorine atom concentration. In vitro 19F MRI imaging results further confirmed the critical role of the metal center; in the control group without Cu2+, no significant 19F MRI signal was detected in the L-F solution (Fig. 3d). In contrast, at the same fluorine atom concentration, MOC-F exhibited significantly enhanced imaging performance, with signal intensity increasing regularly with the sample concentration gradient, demonstrating clear concentration dependence. Based on this dose–response relationship, the detection limit of MOC-F for 19F nuclei under the experimental conditions was determined to be below 10.8 mM. These results conclusively demonstrate that MOC-F is a high-performance 19F MRI contrast agent. Combined with the previously described 1H MRI imaging performance, it is confirmed that MOC-F successfully achieves 1H/19F dual-modal MRI in vitro experiments. Furthermore, this nanoprobe exhibits excellent water solubility and high relaxation efficiency. These characteristics collectively provide a solid foundation for its further application in biomedical imaging.
To further evaluate the MRI performance of this nanoprobe in biological systems, in vivo imaging experiments were conducted in mice. All animal procedures were performed in accordance with the Guide for the Care and Use of Laboratory Animals (National Institutes of Health, USA) and were approved by the Animal Ethics Committee of Shanghai Jiao Tong University (No. 202601056). Prior to in vivo experiments, the cellular biocompatibility of both MOC and MOC-F was evaluated. The methylthiazol-2-yl-tetrazolium (MTT) assay demonstrated that HK-2 cells essentially maintained a viability rate exceeding 80% after 24 h of incubation, even at high probe concentrations, indicating negligible cytotoxicity (Fig. 4a). To assess potential long-term cytotoxicity, we continuously monitored HK-2 cell viability at 36 or 48 h post-probe treatment via MTT assay. The results showed that HK-2 viability remained consistently around 80% (Fig. S25), indicating no significant long-term toxicity from the probe. To evaluate the probe's labeling efficiency and cellular uptake rate, we quantitatively measured intracellular copper content using inductively coupled plasma optical emission spectroscopy (ICP-OES) to determine cellular uptake of MOC and MOC-F. The results showed a linear increase in probe uptake over time until removal at 12 hours, followed by a gradual decline in concentration (Fig. 4b and c). For MOC-F, each cell contained approximately 2.7 × 1012 19F nuclei after 12 h incubation, and this level remained at 2.3 × 1012 19F nuclei per cell 8 hours after the medium replacement. This indicates sufficiently prolonged intracellular retention of the probe, which is crucial for future MRI-based long-term in vivo cell tracking. To evaluate the biocompatibility of MOC and MOC-F in the major mouse organs including lung, liver, spleen, kidneys, and heart, mice received intravenous injections of MOC or MOC-F solutions (the concentration of MOC is 1.2 mM, corresponding to 22.2 mg mL−1, and the concentration of MOC-F is also 1.2 mM, corresponding to 25.8 mg mL−1), with samples collected two days post-injection. Tissue sections were examined after hematoxylin and eosin (H&E) staining. As shown in Fig. 4d, all organs in the three groups exhibited structurally uniform histomorphology without obvious damage. This multi-level experimental evidence, ranging from in vitro cells to in vivo tissues, collectively supports the excellent biosafety of MOC and MOC-F probes under the described experimental protocols, thereby enabling subsequent in vivo imaging studies.
To systematically evaluate the in vivo imaging efficacy of the developed nanoprobe, we conducted MRI experiments using a C57BL/6 mouse model. A 200 μL PBS solution of the MOC-F probe (1.2 mM, equivalent to 25.8 mg mL−1, corresponding to 86.4 mM 19F) was subcutaneously injected into the right hindlimb of the mice, while the left hindlimb received 200 μL PBS solution of MOC (1.2 mM, equivalent to 22.2 mg mL−1) as the control group (Fig. 5a). Subsequently, 1H and 19F MRI scans were performed. The results demonstrated significant 1H MRI signal enhancement at both injection sites, indicating that the probe effectively enhances water proton relaxation in vivo. Furthermore, only the MOC-F-injected right hindlimb exhibited an intense 19F MRI signal (Fig. 5b), confirming MOC-F retains 19F imaging capability within living tissue. These findings confirm the success of our research strategy: Cu2+ simultaneously modulates 1H and 19F relaxation processes to enhance dual-modal MRI signals. Moreover, the MOC-F probe demonstrates excellent imaging performance in vivo, combining high 1H MRI resolution with high 19F MRI specificity. Furthermore, the probe exhibits good biocompatibility and stability in vivo, with sufficient signal intensity for clear imaging.
The successful development of MOC-F not only provides a novel strategy to overcome the sensitivity limitations of conventional 19F MRI probes but also highlights the immense potential of metal–organic cages in biomedical applications. Its excellent compatibility with clinical MRI equipment lays a solid foundation for practical translation. Looking ahead, this study pioneers a new pathway for precisely regulating nanoprobe performance and investigating structure–property relationships through supramolecular design. Leveraging the platform's modularity, targeted functionality can be introduced by modifying the organic ligands—for instance, conjugating peptides or small molecules (such as RGD peptides or folate) to the cage periphery without disrupting its core architecture. Alternatively, biotin–streptavidin interactions or click chemistry can be employed to link antibodies or aptamers, enabling active targeting of specific disease biomarkers. Further optimization of the probe's targeting capabilities and pharmacokinetic properties holds promise for significant contributions to precision disease diagnosis and therapeutic monitoring, ultimately accelerating the clinical translation of imaging probes.