A dual-mode vertical flow assay for species-specific identification and total bacteria load assessment from a single urine sample
Received
26th March 2026
, Accepted 25th May 2026
First published on 26th May 2026
Abstract
Rapid urine screening would benefit from simultaneous determination of pathogen identity and clinically relevant bacteria burden, yet most paper-based assays provide only one of these outputs. Here, we report a dual-mode vertical flow assay that combines species-specific HRP-antibody/TMB immunodetection with WST-8–mPMS metabolic quantification in spatially separated detection spots within an integrated vertical flow assay device. The device employs a vertically layered membrane stack with graded pore sizes to regulate capillary-driven flow and residence time, enabling efficient analyte transport and signal development. Quantitative analysis is achieved using smartphone-based RGB imaging converted to grayscale intensity. This platform enables simultaneous bacteria identification and quantification from a single urine sample, demonstrating threshold-oriented detection around the clinically relevant bacteriuria level of 105 CFU mL−1 in both tryptic soy broth and urine samples. This work demonstrates a dual-mode vertical flow assay capable of simultaneously identifying bacteria species and estimating bacteria burden, providing a practical strategy for rapid threshold-oriented urine screening at the point-of-care.
Introduction
VFAs are an emerging class of paper-based diagnostic platforms that direct samples vertically through stacked porous layers, where capillary forces and controlled pore gradients regulate fluid transport and reaction kinetics.1–6 Compared with LFAs, the architecture of VFAs provides faster flow rates, more uniform reagent–sample interactions, and higher compatibility with multiplexed layouts, while maintaining a simple, low-cost workflow suitable for POCT.2,7–14 This vertical configuration minimizes nonspecific carryover, stabilizes baseline signals, and promotes consistent colorimetric development that can be readily quantified using smartphones or portable readers, while also enabling multiplexed detection of multiple targets with high specificity.2–5,10–13,15–17
These characteristics make VFAs particularly attractive for infection screening, where rapid and quantitative results are required without the need for laboratory instruments.3,6,16 For example, in urine-based bacteria diagnostics, E. coli is the predominant causative agent of UTIs, accounting for approximately 75–85% of uncomplicated UTI cases, making it a clinically relevant primary target for species-specific detection.18,19 Conventional culture remains the gold standard but is too time-consuming, while molecular and immunoassay-based methods require specialized equipment and trained personnel.1,2,11,15,20–24 LFAs, although widely used in POCT, generally provide only qualitative results and are prone to background interference in complex matrices such as urine and whole blood.2,10,15,16,21,25 VFAs overcome these limitations by combining rapid capillary-driven flow with tunable residence time and enhanced analyte retention, providing a versatile and quantitative paper-based alternative for on-site microbial detection.3,6,16,26
Recent research on VFAs includes work in immunoassays, nucleic-acid detection, protein biomarkers, whole-blood testing, and AI-augmented optical readouts, collectively underscoring the versatility of the vertical format. In the field of pathogen detection, Kaur and Eltzov reported an HRP-antibody/TMB scheme augmented with a polyvinyl alcohol time barrier and lactose stabilization to improve the stability of antigen–antibody binding and extend reagent shelf-life, markedly lowering the LOD.16 In the field of nucleic acids, Tahmasebi et al. amplified the hepatitis B virus X gene by PCR and generated HRP chemiluminescence on a VFA, achieving sensitivity comparable to that of clinical real-time PCR with a simpler workflow.20 Regarding protein biomarkers, Wang et al. developed gold-nanoparticle-conjugated antibodies and built a smartphone-quantified VFA platform for acute kidney injury markers (neutrophil gelatinase-associated lipocalin, cystatin C, retinol-binding protein 4), achieving ∼1 ng mL−1 (single-analyte) and 5 ng mL−1 (multi-analyte) performance, demonstrating the advantages of vertical-flow multiplex detection.12 In another study, Lei et al. quantified interleukin-6 in serum within 15–20 min using optimized 0.2 μm pore size membranes and multiple control zones examined via image analysis to ensure accuracy, achieving an LOD of 3.2 pg mL−1 in serum.11 With respect to whole-blood compatibility, Serhan et al. separated plasma within a multilayer stack by leveraging pore-size differences, and measured total iron from finger-prick blood in ∼5 min via colorimetry, thus eliminating centrifugation.25
In addition to the abovementioned advances, progress in optical and data-science continues to enhance analytical sensitivity. Goncharov et al. introduced a paper-based fluorescent VFA for three cardiac injury markers (myoglobin, creatine kinase–muscle/brain type, heart-type fatty acid-binding protein), using deep learning to normalize spatial signals, which yielded strong correlation with ELISA and achieved LODs below 0.52 ng mL−1 for all three biomarkers.15 In another example of advanced optical integration, Chen et al. employed magnetic surface-enhanced raman scattering nanotags (Fe3O4@Au core–shell) for ultrasensitive multiplex detection of mycotoxins in complex food matrices, achieving signal enhancements of approximately three orders of magnitude compared with conventional gold-nanoparticle assays.27 Despite these advances, few VFAs are capable of providing both species identification and total bacteria load from urine within a single, low-complexity workflow explicitly aligned to a clinical threshold.
To address the growing demand for accessible, rapid, and specific detection of bacteria pathogens, we developed a VFA platform featuring multiple engineering innovations for simultaneous species-specific identification and quantitative POCT. The integrated device contains four spatially separated detection spots within the same VFA device: two spots for HRP-antibody/TMB-based species-specific immunodetection and two spots for WST-8–mPMS-based metabolic quantification, enabling parallel readouts from a single urine sample while minimizing chemical and optical interference between the two colorimetric reactions (Fig. 1A). First, we employed a simplified immunoassay using a single primary antibody directly conjugated with HRP, eliminating secondary antibodies to reduce steps, cost, and time (Fig. 2). This one-step detection approach offers a more practical and efficient solution for use in resource-limited settings. Second, the detection layer in our device used PT-R7 paper-based material, a high-capillarity conjugate-release matrix with interwoven cellulose–glass fibers that supports high loading, passive protein adsorption, uniform rehydration, and dry-storage stability.24,28–34 This material facilitates uniform fluid penetration, prolongs antigen–antibody interaction time, and enables efficient protein immobilization with controlled release after rehydration.24,28,30–32 Third, we integrated smartphone-based RGB colorimetry under standardized lighting, converting RGB values to grayscale intensities for quantitative evaluation without benchtop readers (Fig. 2). This approach enables instrument-free quantitative interpretation compatible with digital health workflows, supporting seamless integration with mobile-based data analysis and remote clinical decision-making, thereby enhancing the practicality of the platform for POCT and decentralized testing. Fourth, a funnel-like absorbent stack with a graded pore structure was placed beneath the detection plane to tune residence time, concentrate analytes at the reaction interface, and suppress downward diffusion of immunocomplexes. This configuration also facilitated the removal of unbound antibodies, thereby improving signal clarity and sensitivity (Fig. 1A–C).
 |
| | Fig. 1 Design and structure of the integrated dual-mode VFA device. (A) Schematic illustration of the integrated VFA device with spatially separated detection spots for the WST-8–mPMS metabolic assay and HRP-antibody/TMB immunoassay. The graded-pore paper stack guides vertical capillary flow through the device. (B) Top view of the assembled VFA device, including the plastic container, sample injection plate, and detection layers. (C) Exploded view of the device components, including absorbent layers, detection layer, filter layer, sample injection plate, and plastic container. | |
 |
| | Fig. 2 Workflow and detection principles of the dual-mode VFA. The device fabrication process includes partial hydrophobicity treatment, reagent immobilization, and device assembly. The platform contains two spatially separated detection modules: an HRP-antibody/TMB immunoassay for species-specific qualitative identification and a WST-8–mPMS metabolic assay for quantitative total bacteria load assessment. After sample loading, colorimetric signals are recorded by smartphone imaging, extracted as RGB values, and converted to grayscale intensity for data analysis. The figure was created using https://BioRender.com. | |
For total bacteria load assessment, the WST-8–mPMS metabolic readout was adapted from a viable microorganism detection strategy previously developed by our group.35 In this system, WST-8 serves as a tetrazolium salt that is reduced to a water-soluble formazan product, while mPMS functions as an intermediate electron mediator. EDTA is used to transiently increase bacteria cell wall permeability, allowing the reagent system to couple with intracellular metabolic activity. In metabolically active bacteria, endogenous dehydrogenase activity transfers electrons through mPMS to WST-8, resulting in formazan formation. Therefore, the colorimetric signal reflects viable bacterial metabolic activity and provides a species-independent readout of total bacteria load (Fig. 2). Finally, the HRP-antibody/TMB immunodetection module and the WST-8–mPMS metabolic module enable simultaneous species-specific identification and total bacteria load assessment from a single urine sample within one integrated VFA device (Fig. 1A–C and 2).
The combination of layered material engineering, simplified immunochemistry, mobile quantification, and device-level integration provides a novel and versatile platform with strong potential for practical use in clinical and environmental microbiological diagnostics. To evaluate its analytical performance, the assay was tested across bacteria concentrations centered around 105 CFU mL−1, corresponding to the clinically significant bacteriuria threshold commonly used for urinary tract infection screening.22–24 Here, 105 CFU mL−1 was used as a clinically relevant decision threshold for assay evaluation rather than as a formally calculated limit of detection. This approach ensured that device performance directly reflected screening and triage decision levels. We further demonstrated species-specific immunoassay detection in both TSB and urine, quantitative trends for total bacteria load across multiple species and matrices, and integrated a cartridge operation that enabled simultaneous metabolic and immunological readouts from a single sample. Rather than maximizing analytical sensitivity, this work introduces a clinically aligned sensing paradigm that simultaneously resolves pathogen identity and decision-level bacteria burden within a single assay.
Experimental section
Reagents and materials
E. coli DH5α (BCRC No. 51731) and S. aureus subsp. aureus (BCRC No. 12989) were obtained from the Bioresource Collection and Research Center (BCRC, Hsinchu, Taiwan). Paper-based substrates used for device fabrication and flow optimization included Whatman filter papers (Lot No. 1002-150, 1003-185, 1004-240, 1005-185, 1006-185, 1440-150/185, 1441-185, 1442-185, 1454-185, and 1542-185; Cytiva Life Sciences, Marlborough, MA, USA), ADVANTEC filter papers (Lot No. 5A, 5B, and 5C; Toyo Roshi Kaisha, Ltd., Tokyo, Japan), PT-R7 conjugate-release membranes (Model R-1343/1554, Lot No. 343438F; MDI Technologies, Ambala, Haryana, India), and blotting papers (Model FP-1, Catalog No. OWFP1, 20 × 20 cm, 100 sheets per box; Thermo Fisher Scientific, Waltham, MA, USA). The plastic container and sample injection plate were adapted from the Watson Co., Ltd. Plasma Separation Device (Cat. No. 176-600C, Tokyo, Japan).
HRP conjugation Kit, Lightning-Link® (ab102890), 3,3′,5,5′-tetramethylbenzidine (TMB) substrate solution (ab99995), and anti-E. coli LPS monoclonal antibody (ab35654, 100 μg) were purchased from Abcam (Cambridge, UK). Bovine serum albumin (BSA, ≥98%, Cat. No. A7906, CAS 9048-46-8), tryptic soy broth (TSB, NutriSelect® Basic, Cat. No. T8907-500G), and phosphate-buffered saline (PBS, Cat. No. P4417-50TAB) were obtained from Sigma-Aldrich (St. Louis, MO, USA). Tween 20 (Cat. No. TB0560, CAS 9005-64-5) was supplied by Bio Basic Inc. (Markham, ON, Canada). Ethylenediaminetetraacetic acid (EDTA, MW 292.25, CAS 60-00-4, Cat. No. 8991-01) was obtained from J. T. Baker (Phillipsburg, NJ, USA). WST-8 (2-[2-methoxy-4-nitrophenyl]-3-[4-nitrophenyl]-5-[2,4-disulfophenyl]-2H-tetrazolium sodium salt) and mPMS (1-methoxy-5-methylphenazinium methyl sulfate) were purchased from IMT Formosa New Materials Co., Ltd. (Kaohsiung, Taiwan).
Water-soluble red ink (LIFE No. 100, red, 10 mL, water-based refill ink for automatic stamps) was purchased from a local stationery store (Hsinchu, Taiwan). Scanning electron microscopy (SEM) images were obtained using a JEOL IT800 field-emission scanning electron microscope. Colorimetric signal analysis was performed using the Colorimeter App (iOS; free download: https://apps.apple.com/ee/app/colorimeter-app/id1542365656). Deionized water (18.2 MΩ·cm, ddH2O) was used in all experiments, and all reagents were of analytical grade unless otherwise stated.
Water-soluble red ink flow test
To identify an optimal paper-based material combination that ensures stable, rapid, and uniform fluid flow within the detection device, red ink was used as a preliminary fluid model to visualize capillary flow behavior and optimize layer configurations. Because the vertical flow system is primarily governed by capillary-driven liquid transport rather than particle-specific interactions, the dye solution provides a convenient approximation for evaluating flow distribution and residence time during device optimization.
The evaluated parameters included thickness (μm), pore size (μm), and wet burst strength (psi) (Table S1). Based on pore size and experimental requirements, all paper substrates were classified into three groups: group 1 (samples ①–③), group 2 (samples ④–⑥ and
), and group 3 (samples ⑦–⑩,
, and
). To assess the applicability of these materials in vertical flow detection devices, we conducted a series of water-soluble red ink flow tests designed to evaluate flow uniformity and stability. Water-soluble red ink was chosen because its high visibility allowed direct visualization of capillary transport through multilayer paper structures. The paper substrates were primarily sourced from two manufacturers: Whatman (samples ①–⑩) and ADVANTEC (samples
–
). Their specific models and physical properties were extracted from the manufacturers' official databases (Whatman: https://www.cytivalifesciences.com/en/us; ADVANTEC: https://www.advantec.co.jp/en/). These data served as the initial criteria for material selection in subsequent experiments aimed at optimizing device fabrication and performance.
For flow testing, Whatman substrates were first assembled in layered stacks according to the experimental design, with a blotting paper sheet placed at the bottom as the absorbent layer to capture excess fluid after applying 1 mL of water-soluble red ink. The stacking arrangements are detailed in Table S2. The flow behavior of the top layer was directly observed, while flow through the second and third layers was quantified using a timing method, recording the elapsed time until the ink reached the bottom blotting paper. These measurements enabled evaluation of both flow uniformity and transport efficiency. Subsequently, untested ADVANTEC substrates were integrated with the Whatman substrates to generate 11 new stacked combinations, organized by pore size, as shown in Table S3. Each combination was tested under identical conditions, with flow time, uniformity, and stability recorded as key metrics. The results of these experiments provided the basis for selecting the most suitable paper-based material combinations to support the development and optimization of the VFA device, as summarized in Tables S4 and S5.
Scanning electron microscopy
Scanning electron microscopy imaging was performed to verify bacterial transport and retention within the optimized paper-based vertical flow structure. Based on the optimized parameters obtained from the red ink assay, a layered configuration consisting of Whatman 1454–185 as the first layer and Whatman 1003-185 as the second layer was selected. S. aureus (108 CFU mL−1) was applied onto the stacked substrates, while ddH2O was used as the control. Bacteria penetration and retention in the lower layers were examined by SEM. Prior to imaging, samples were sputter-coated with an approximately 3 nm platinum layer to enhance conductivity. SEM analysis was performed at an accelerating voltage of 5 kV under high-vacuum conditions, and images were acquired at 10
000× and 20
000× magnifications to visualize bacteria cells and paper microstructures.
Device configuration and fabrication
The optimized paper substrates and PT-R7 membrane were cut into 1 cm × 1 cm squares using a paper cutter. To guide sample delivery toward predefined reaction zones, circular openings (0.6 cm in diameter) were created in laminating films using a hole puncher. These films were aligned with paper substrates and processed under heat-lamination conditions, yielding detection layers with guided hydrophilic reaction zones rather than fully enclosed microfluidic channels.
For the integrated dual-mode VFA, four independent detection spots were arranged within the same plastic container. Two spots were assigned to the HRP-antibody/TMB-based species-specific immunodetection module, while the other two spots were assigned to the WST-8–mPMS-based metabolic quantification module. These detection spots were spatially separated to minimize chemical and optical interference between the two colorimetric reactions while enabling parallel readouts from a single urine sample within one device. The sample injection plate contained guided flow channels to direct the sample and washing buffer into the predefined detection spots in a stable and controlled flow direction (Fig. 1A–C).
For the HRP-antibody/TMB immunodetection module, the vertical flow device was constructed with a layered configuration composed of a detection layer and a multi-layer absorbent layer. The detection layer consisted of a PT-R7 membrane patterned with hydrophilic channels to define the immunodetection spots for antigen recognition and signal generation. The absorbent layer was assembled as a five-layer structure arranged sequentially from top to bottom: one sheet of Whatman No. 1454 (pore size ≈ 22 μm), two sheets of Whatman No. 1003 (pore size ≈ 6 μm), and two sheets of blotting paper with high absorbency. This vertical pore gradient guided fluid transport through the stack. The assembled substrates were placed into a plastic container, with a sample injection plate mounted above the detection layer to facilitate sample loading. The overall configuration is shown in Fig. 1B and C. This configuration provided a controlled capillary pressure gradient, enabling reproducible downward flow and enhanced residence time at the detection interface.
For the WST-8–mPMS metabolic quantification module, a simplified two-layer format was used during module-level optimization and standalone performance evaluation, consisting of a filtration layer (Whatman 1454-185) on the top and a detection layer (Whatman 1003-185) on the bottom. The same WST-8–mPMS-based metabolic detection principle was then incorporated into the integrated dual-mode device as spatially separated metabolic detection spots. To improve handling and stability, the detection layer was supported by a transparent laminating film at its base, serving as a backing layer.
HRP-antibody conjugation
HRP-antibody conjugation was performed using an HRP conjugation kit. Prior to use, all reagents and consumables were equilibrated to room temperature (20–25 °C). Antibodies were purified and dissolved in PBS (pH 7.2–7.4). Buffers containing primary amines (e.g., glycine, ethanolamine) or thiol compounds (e.g., dithiothreitol, β-mercaptoethanol) were strictly avoided, as these interfere with the conjugation reaction.
For the conjugation reaction, 1 μL of modifier reagent was added per 10 μL of antibody solution, followed by gentle mixing to ensure homogeneity. The HRP conjugation mix vial was then opened, and the antibody–modifier mixture was pipetted directly onto the thawed HRP mix. The solution was gently mixed by pipetting up and down 1–2 times to ensure complete dissolution. The conjugation mixture was incubated at 25 °C for 3 h in the dark. Following incubation, 1 μL of quencher reagent was added per 10 μL of reaction mixture. The solution was gently mixed and allowed to stand for 30 min at room temperature to terminate the reaction and stabilize the conjugates. The resulting HRP-antibody conjugates required no further purification and were stored at 4 °C for up to 18 months. For long-term storage, conjugates were supplemented with 50% glycerol and stored at −20 °C in the dark. Repeated freeze–thaw cycles were avoided to preserve antibody activity. Successful conjugation was verified by visible color change and confirmed by comparing HRP activity between conjugated and unconjugated controls.
Antibody immobilization and blocking
The HRP-conjugated primary antibody was diluted 1
:
500 in PBS and subsequently mixed at a 1
:
1 (v/v) ratio with blocking buffer (PBS supplemented with 0.05% Tween 20 and 1% BSA). A 20 μL aliquot of the antibody–blocking buffer mixture was then dispensed onto the hydrophilic center of the detection layer corresponding to the HRP-antibody/TMB immunodetection spots. To ensure efficient adsorption and prevent enzyme inactivation, all procedures were performed within a biosafety cabinet, and the samples were incubated in the dark for 2 h. This treatment promoted stable antibody immobilization while minimizing nonspecific binding through BSA blocking, thereby supporting the sensitivity and specificity of subsequent immunoassays.
Sample loading and washing
To evaluate device sensitivity across different bacteria concentrations, E. coli suspensions were serially diluted in TSB and human urine to final concentrations of 108, 107, 106, 105, and 5 × 104 CFU mL−1. For each assay, the prepared sample was applied onto the top of the device through the sample injection plate, which guided liquid delivery toward the predefined detection spots. The sample then infiltrated naturally into the detection layer by capillary action, ensuring complete contact with the reaction zones for antigen–antibody binding. The devices were incubated at room temperature (22–25 °C) for 30 min without agitation to maintain consistency across tests. Following incubation, unbound antibodies and residual interfering substances were removed by a washing step. This washing step was required for the HRP-antibody/TMB immunodetection module to reduce background signals and improve species-specific signal clarity. The PBS containing 0.05% Tween 20 was dispensed dropwise onto the top of the device through the guided sample injection plate, enabling flow through the paper matrix via capillary action. The washing procedure was repeated twice to ensure thorough cleaning of the reaction zone and to reduce background signals. Care was taken to avoid rapid dispensing, which could disturb antigen–antibody complexes bound within the detection layer. Because liquid transport was guided by the sample injection plate and driven by the absorbent paper stack, no active pumping or external fluidic equipment was required.
Colorimetric reaction
Following the washing step, the detection layer was placed on a white background. A 10 μL aliquot of TMB substrate solution was directly added to the reaction zone. TMB color development began immediately after substrate addition, and images were acquired at a standardized endpoint of approximately 3 min to ensure consistent readout timing across all samples. The oxidized TMB product was observed as a blue coloration, reflecting HRP enzymatic activity and indicating antigen–antibody binding.
Image acquisition and data analysis
For both the HRP-antibody/TMB immunodetection module and the WST-8–mPMS metabolic module, colorimetric signals were quantified by smartphone-based RGB extraction, white-reference normalization, and grayscale conversion. After the colorimetric reaction, images of the detection zones were immediately captured using a smartphone. To ensure consistency, all images were taken under identical lighting and exposure settings to minimize variability across tests. A white reference background was used for normalization across images. Image analysis was performed using the Colorimeter App (iOS, free download: https://apps.apple.com/ee/app/colorimeter-app/id1542365656), which extracts real-time intensity values of red, green, and blue channels (Fig. 2).
For each reaction zone, RGB values were obtained by averaging a fixed circular ROI centered within the reaction spot to minimize edge effects and spatial heterogeneity. The same ROI size and placement rule were applied across all samples. To reduce the influence of illumination variation, camera settings, and strip-to-strip background differences, a white reference region on the same test strip/device was used as an internal reference for RGB normalization before grayscale conversion. Each RGB channel of the reaction zone was normalized against the corresponding white reference value according to R′ = R/(Rwhite/255), G′ = G/(Gwhite/255), and B′ = B/(Bwhite/255). The normalized RGB values were then converted into a single grayscale intensity using a weighted luminance formula: Grayscale = 0.229 × R′ + 0.587 × G′ + 0.114 × B′, reflecting the relative contributions of red, green, and blue channels to perceived brightness. This device-specific white-reference normalization was applied once for each device and was used to reduce variations caused by lighting conditions, smartphone imaging, and differences in background brightness among test strips. Normalized grayscale intensity was calculated as the difference between the background/reference region and the detection zone signals. This normalization strategy compensates for intrinsic sample coloration, including the yellowish background commonly observed in urine, by referencing each reaction zone to its local background.
Calibration curves were then generated by nonlinear regression using the Hill equation:
where
Y is the normalized grayscale,
X is the bacteria concentration,
Bmax is the maximum response,
Kd is the apparent dissociation constant, and
h is the Hill coefficient. This model provided robust fitting of the nonlinear dose–response relationship and enabled back-calculation of bacteria concentrations in unknown samples.
Total bacteria quantification (WST-8–mPMS assay)
An appropriate amount of WST-8 (2-[2-methoxy-4-nitrophenyl]-3-[4-nitrophenyl]-5-[2,4-disulfophenyl]-2H-tetrazolium, molecular weight 600.48) was weighed and dissolved in ddH2O to prepare solutions of various concentrations. Similarly, mPMS powder was weighed and dissolved in ddH2O at different concentrations. Additionally, EDTA was dissolved in ddH2O to prepare a 10−5 M solution. EDTA was included to transiently permeabilize bacteria cell walls, thereby enhancing the coupling between intracellular dehydrogenase activity and WST-8 reduction.
To prepare the WST-8–mPMS detection reagent, WST-8 and mPMS solutions were mixed at a 1
:
1 volume ratio. This reagent, together with the EDTA solution, was drop-cast onto the surface of hydrophobically treated paper-based materials. The coated papers were then placed in a biosafety cabinet and air-dried overnight to ensure stable immobilization of the reagents. After complete drying, the treated paper substrates were sequentially stacked, layer by layer, according to the designed configuration to guarantee sufficient interlayer contact.
To evaluate device performance, bacteria suspensions at various concentrations were added to the top paper layer containing EDTA. The sample then passed through the filtration layer and reached the WST-8–mPMS-loaded detection layer by capillary flow. In metabolically active bacteria, endogenous dehydrogenase activity reduced WST-8 through mPMS-mediated electron transfer, producing a water-soluble orange formazan product. Color changes were monitored within 20 min after sample application to confirm the device's ability to detect total bacteria load.
Bacteria quantification and RGB value analysis
For total bacteria load detection, bacteria suspensions of E. coli and S. aureus with concentrations ranging from 108, 107, 106, 105, and 5 × 104 CFU mL−1 were applied to the device, and color changes in the detection layer were monitored. Quantitative analysis was performed by RGB extraction, white-reference normalization, and grayscale conversion using the same image-analysis procedure described above.
Results
Paper-based material selection by red ink flow test
To establish stable and uniform fluid flow, commercially available paper substrates from Whatman and ADVANTEC were systematically screened using a water-soluble red ink flow assay. Thirteen paper types were characterized by their physical properties, including thickness (150–390 μm), pore size (1–25 μm), and wet burst strength (0.22–7 psi) (Table S1). Based on pore size and flow behavior, these materials were grouped and vertically stacked into multi-layer configurations to evaluate their combined wicking performance (Tables S2 and S3).
Red ink flow tests revealed that combinations employing intermediate pore sizes in the upper layer (20–25 μm) coupled with smaller pores in the middle and lower layers (2–7 μm) produced the most uniform capillary-driven flow and stable transport dynamics. Among 27 initial Whatman combinations, configurations such as ③–⑥–⑧ and ③–⑥–⑦ demonstrated the shortest flow times (≈26–27 s) and superior flow uniformity, whereas assemblies composed of either overly large or overly dense papers led to uneven spreading or delayed transport (Table S4). Further optimization integrating ADVANTEC materials (samples
–
) confirmed that graded-pore stacking enhanced fluid distribution and minimized backflow. The combination ③–⑥–⑧, featuring a moderate top-layer pore size (22 μm) and progressively decreasing porosity in the lower layers (6 μm → 2.7 μm), yielded the best compromise between flow velocity and uniform signal formation (Table S5). The final design adopted this funnel-like absorbent stack with a graded pore structure, which balanced capillary pressure and analyte retention, ensuring efficient delivery toward the reaction interface while minimizing downward diffusion of immunocomplexes. These results established a reliable paper-based architecture with reproducible vertical flow dynamics, providing a robust foundation for subsequent immunoassay integration.
Species-specific bacteria detection in TSB
Following material and structural optimization, the HRP-antibody/TMB immunoassay in a VFA was evaluated using E. coli suspensions prepared in TSB. A fixed volume of 1 mL was applied through the sample injection plate to guide sample delivery into the predefined immunodetection spots (Fig. 1B and C). The sample injection plate guided the liquid into the predefined immunodetection spots, and TMB-based color development produced distinct blue signals corresponding to bacteria concentration. Quantitative analysis was performed by extracting RGB values, applying white-reference normalization, and converting the normalized RGB values into grayscale intensities (Fig. 3A, B and E). The colorimetric response changed in a concentration-dependent manner across the tested E. coli concentration range of 5 × 104–108 CFU mL−1, supporting threshold-oriented detection around the clinically relevant bacteriuria level. Hill equation fitting yielded the following model (Fig. 3A):
Grayscale = (84 271 × [E. coli]0.5730)/(K0.5730d + [E. coli]0.5730); R2 = 0.7422 |
Although model convergence of Kd was limited, a slightly stronger fit (Fig. 3B):
| Grayscale = 10.52 × [E. coli] − 28.50; R2 = 0.7718 |
CV analysis revealed fluctuations at higher concentrations, with the maximum CV reaching 20.15% at 106 CFU mL−1 (Fig. 3E). The lateral color spreading observed in the reaction zones (Fig. 3E) is attributed to capillary diffusion within the porous detection matrix rather than insufficient hydrophobic confinement by lamination. Non-uniform color development, particularly at low bacteria concentrations, reflects heterogeneous enzyme distribution and capillary transport within the porous matrix; however, averaging grayscale values over a fixed region of interest mitigated local variability and enabled reliable concentration-dependent trends. These results suggest that, while the assay delivered a consistent concentration-dependent response, performance at high bacteria loads could be further stabilized. To further validate the specificity of the E. coli immunoassay, S. aureus suspensions (5 × 104–108 CFU mL−1) were tested under the same immunoassay protocol but without integration into the plastic container system. No visible color development was observed across all concentrations (Table S6), supporting the initial specificity of the anti-E. coli immunoassay and minimal cross-reactivity against the tested Gram-positive non-target bacterium.
 |
| | Fig. 3 Species-specific HRP-antibody/TMB immunodetection of E. coli in TSB and urine using the integrated VFA device. (A and B) Calibration curves for E. coli detection in TSB fitted using (A) Hill and (B) linear regression models. (C and D) Calibration curves for E. coli detection in human urine fitted using (C) Hill and (D) linear regression models. (E) Representative colorimetric responses and grayscale quantification of E. coli at different concentrations in TSB and urine. Data are presented as mean ± SD (n = 5), with CV values shown for each concentration. Statistical significance is indicated as p < 0.05 (*), p < 0.01 (**), and p < 0.001 (***). | |
Species-specific bacteria detection in human urine
To evaluate clinical applicability, the same HRP-antibody/TMB immunodetection module was tested using untreated human urine spiked with E. coli suspensions (5 × 104–108 CFU mL−1). Distinct blue coloration and concentration-dependent contrast were maintained, demonstrating stable assay performance in complex matrices (Fig. 3C–E). Hill equation fitting yielded the following model (Fig. 3C):
Grayscale = (79 127 × [E. coli]0.3881)/(K0.3881d + [E. coli]0.3881); R2 = 0.8643 |
Although Kd convergence was limited, linear regression analysis provided an alternative fit (Fig. 3D):
| Grayscale = 18.85 × [E. coli] − 25.81; R2 = 0.8764 |
CV analysis demonstrated excellent reproducibility across concentrations, with CVs of 13.85%, 10.71%, 13.78%, 7.69%, and 7.11% for 5 × 104, 105, 106, 107, and 108 CFU mL−1, respectively (Fig. 3E). Despite the intrinsic coloration of urine, concentration-dependent signal trends and statistical discrimination were preserved after grayscale normalization, indicating minimal interference from sample background. Signal variation decreased at higher bacteria loads, suggesting improved saturation uniformity within the vertical flow structure. These findings indicate that the platform maintained threshold-oriented detectability in complex biological fluids, highlighting its potential applicability for UTI screening.
Total bacteria load assessment in TSB and human urine
The VFA design also allowed for implementation of alternative detection principles. In addition to immunoassay-based recognition, we developed a WST-8–mPMS metabolic VFA to quantify total bacteria load, thereby expanding the platform's applicability beyond species-specific detection. Both E. coli and S. aureus were tested using concentrations ranging from 5 × 104 to 108 CFU mL−1 in TSB and human urine. Linear regression analysis demonstrated that in TSB, E. coli exhibited a concentration-dependent metabolic response (Fig. 4A):
| Grayscale = −0.1259 × [E. coli] + 23.558; R2 = 0.84 |
In human urine, the response remained consistent, yielding (Fig. 4B):
| Grayscale = −4.263 × [E. coli] + 195.2; R2 = 0.82 |
For S. aureus, linear fitting also supported concentration-dependent total bacteria load estimation. In TSB, the relationship was (Fig. 4C):
| Grayscale = −4.883 × [S. aureus] + 205.3; R2 = 0.72 |
while in human urine, the corresponding fit was (Fig. 4D):
| Grayscale = −5.574 × [S. aureus] + 200.0; R2 = 0.78 |
These findings demonstrate that the WST-8–mPMS metabolic VFA supports total viable bacteria load assessment across different bacteria species and sample matrices. Importantly, this demonstrates the versatility of the VFA, which can be adapted for either metabolic or immunological detection. Building upon this flexibility, we further combined the two modules into an integrated platform capable of simultaneous species-specific and quantitative bacteria detection.
 |
| | Fig. 4 WST-8–mPMS metabolic quantification of total viable bacteria load across bacteria species and sample matrices. (A and B) Quantitative detection of E. coli using the WST-8–mPMS metabolic assay in (A) TSB and (B) human urine. (C and D) Quantitative detection of S. aureus using the WST-8–mPMS metabolic assay in (C) TSB and (D) human urine. Grayscale intensity decreased with increasing bacteria concentration, reflecting stronger WST-8 reduction and formazan formation at higher viable bacteria loads. Data are presented as mean ± SD (n = 5). Statistical significance is indicated as p < 0.05 (*), p < 0.01 (**), p < 0.001 (***) and p < 0.0001 (****). | |
Integrated species-specific and quantitative bacteria detection
To demonstrate multiplexing capability, an integrated platform combining total bacteria load assessment using the WST-8–mPMS metabolic assay and species-specific detection using the HRP-antibody/TMB immunoassay was evaluated in human urine. Each test used 1 mL of spiked E. coli suspension across five concentration levels (5 × 104–108 CFU mL−1). In this integrated device, the two readouts were generated in parallel from spatially separated detection spots within the same plastic container, using the same urine sample. For total bacteria load detection, grayscale values generally increased with concentration. Statistical analysis revealed highly significant differences between 5 × 104 and 107 CFU mL−1 (***, p < 0.001) and between 5 × 104 and 108 CFU mL−1 (**, p < 0.01), indicating that the WST-8–mPMS module enabled threshold-oriented total bacteria load assessment across a concentration range spanning the clinically relevant bacteriuria level of 105 CFU mL−1 (Fig. 5A). For species-specific detection, concentration-dependent signals were observed despite variability. A statistically significant difference between 5 × 104 and 108 CFU mL−1 (***, p < 0.001), demonstrated that the antibody-based immunoassay module could distinguish high bacteria load from low-level bacteria concentrations within the integrated device (Fig. 5B).
 |
| | Fig. 5 Integrated dual-mode detection of E. coli in urine using a single VFA device. (A) Total bacteria load assessment of E. coli in urine using the WST-8–mPMS metabolic module integrated in the plastic container. (B) Species-specific identification of E. coli in urine using the HRP-antibody/TMB immunoassay module integrated in the same device. (C) Representative photograph of the integrated plastic-container device used for parallel dual-mode detection from a single urine sample. Data are presented as mean ± SD (n = 5), with CV values shown. Statistical significance is indicated as p < 0.01 (**), and p < 0.001 (***). | |
Overall performance
Across all tested configurations, the VFA consistently demonstrated its ability to provide both qualitative and quantitative bacteria detection under simple operating conditions. The HRP-antibody/TMB immunoassay module enabled species-specific recognition of E. coli with clear visual signal discrimination in both TSB and urine. The WST-8–mPMS metabolic module further supported quantification of bacteria load, showing reliable performance for both E. coli and S. aureus across concentrations ranging from 5 × 104 to 108 CFU mL−1 in different sample matrices. In the integrated device, the two modules provided parallel readouts from a single urine sample using spatially separated detection spots within one plastic container. These findings validate the robustness and versatility of the VFA architecture, demonstrating its adaptability for metabolic or immunological detection and its integration potential for multiplex POCT.
Discussion
This study establishes a single-sample, dual-mode VFA that integrates species-specific immunodetection and total viable bacteria-load assessment within one paper-based device. The developed dual-mode VFA combines an HRP-antibody/TMB immunoassay with a WST-8–mPMS metabolic module, enabling complementary information on bacteria identity and overall burden to be obtained from a single urine sample. Importantly, the two readouts are generated from spatially separated detection spots within the same integrated device rather than from a single shared test zone, allowing parallel operation while minimizing chemical and optical interference between the two colorimetric reactions. Given that E. coli is the leading causative agent of UTIs, accounting for approximately 75–85% of uncomplicated UTI cases, it was selected as the primary target for evaluating species-specific identification in this study.18,19 Across TSB and urine matrices, the platform produced visually distinguishable and quantitatively interpretable responses, with statistically significant discrimination around the clinically relevant threshold of 105 CFU mL−1.22–24 The detection range was defined from 5 × 104 to 108 CFU mL−1, and 105 CFU mL−1 was used as a clinically relevant bacteriuria decision threshold for assay evaluation rather than as a formally calculated LOD. Rather than emphasizing maximum analytical sensitivity, this study highlights a clinically aligned sensing strategy for rapid POCT assessment of both pathogen identity and decision-level bacteria load.
Selection of PT-R7 as the detection layer
According to the red ink flow test results, initial trials using the Whatman 1454-185 paper (Experimental ID ③) produced uniformly intense color signals across all E. coli concentrations (Table S7). However, the absence of a concentration-dependent gradient indicated excessive background and limited quantitative resolution. To overcome this limitation, the Whatman substrate was replaced with PT-R7, a conjugate-release matrix commonly employed in LFA.24,31,32 When incorporated into the vertical-flow configuration, PT-R7 markedly improved signal discrimination among bacteria concentrations (Fig. S1), yielding a clear positive correlation fitted by both Hill and linear models (R2 = 0.7051 and 0.6136, respectively). This improvement can be attributed to the material's more uniform liquid transport and reduced reagent over-retention, which together lowered background coloration and improved sensitivity. PT-R7 was selected as the detection layer because its interwoven cellulose-glass-fiber mesh provides a high surface area and strong capillary action.24,28,30,31,34 In contrast to conventional nitrocellulose membranes, which exhibit rapid wicking and strong but often irreversible protein adsorption that can limit controlled reagent release and residence time in enzyme-based assays, PT-R7 enables more controlled fluid transport and reagent retention. The conjugate-release matrix structure supports high loading of antibodies, enzyme labels, and gold nanoparticles, while enabling uniform rehydration and stable signal development.28–30,33,36 The surface chemistry of PT-R7 enables passive adsorption through electrostatic, hydrogen-bonding, and hydrophobic interactions, eliminating the need for chemical coupling and simplifying fabrication while preserving biological activity.24,29,30,32,34 Upon rehydration, PT-R7 facilitates controlled release of pre-deposited reagents rather than permanent sequestration, resulting in consistent signal generation and reduced variability.30 In addition to its analytical advantages, PT-R7 offers excellent process compatibility. It endures dry storage, retains mechanical integrity during lamination, and readily accommodates hydrophilic channel patterning.28–30 These features mirror its established role in LFAs as a conjugate or release pad, where on-demand reagent delivery after rehydration is essential.24,28–33 Overall, the optimized capillary flow, enhanced protein adsorption, and manufacturing robustness of PT-R7 significantly improved signal clarity and reproducibility in the vertical-flow format.30 These characteristics confirm PT-R7 as a high-performance detection substrate that enhances both analytical performance and device manufacturability.
Vertical flow architecture and fluid control
The core design centers on the vertical flow immunoassay device. A PT-R7 detection layer was coupled to a five-layer absorbent stack arranged from top to bottom as one sheet of Whatman 1454 (pore size 22 μm, Experimental ID ③), two sheets of Whatman 1003 (pore size 6 μm, Experimental ID ⑥), and two sheets of high-absorbency blotting paper (Fig. 1C). This graded-pore structure regulated capillary flow and mass transport (Fig. 1A and C). As liquid moved from larger to smaller pores, flow velocity decreased and residence time at the detection interface increased, enhancing antigen–antibody binding and enzymatic turnover. Unlike conventional microfluidic devices that rely on rigid hydrophobic barriers to strictly confine lateral flow, the presented vertical flow architecture prioritized capillary-driven vertical transport within a porous matrix. Consequently, limited lateral diffusion within the detection layer is expected in this architecture. This design trade-off favors robustness, ease of fabrication, and tolerance to user handling, which are critical considerations for POCT applications.
The graded structure also selectively retains bound immunocomplexes within the upper layers while allowing unbound antibodies and small molecules to percolate downward, thereby reducing background and stabilizing the baseline. These features concentrate analytes at the reaction interface and promote uniform penetration across the detection zone, yielding reproducible transport and improved signal clarity. The washing step remains necessary for the HRP-antibody/TMB immunodetection module because it removes unbound HRP-conjugated antibodies and residual urinary matrix components, thereby reducing nonspecific background signals. Although washing steps increase operational complexity and may introduce user-dependent variation in POCT use, this issue was partially mitigated in the current prototype by incorporating guided flow channels into the sample injection plate. These channels direct both the sample and washing buffer into the predefined detection spots in a stable and controlled manner (Fig. 1B and C). This design reduces the need for users to manually target each reaction zone and helps lower dispensing-related errors. Nevertheless, further simplification of the washing process, such as preloaded washing buffer or absorbent-layer-controlled automatic washing, will be important for future product-oriented development.
SEM verification of bacterial transport in the WST-8–mPMS module
For the WST-8–mPMS metabolic readout, a simplified two-layer format was used. The top filter layer was Whatman 1454, which removed common matrix particulates such as cellular debris, exfoliated epithelial cells, mucus, colloidal protein aggregates, and amorphous urinary salts or crystals before the filtrate reached the detection layer, reducing fouling and variability while preserving rapid assay kinetics. To further verify bacteria transport through this two-layer configuration, SEM was performed after applying S. aureus (108 CFU mL−1) onto the optimized stack. Control samples without bacteria showed only the fibrous microstructure of the filter and detection layers, whereas bacteria-treated samples revealed S. aureus adhering to the fiber surfaces of both layers (Fig. S2). These results confirm that bacteria can traverse the upper filtration layer and reach the underlying detection layer under capillary-driven flow. This observation supports the functional design of the WST-8–mPMS metabolic module, which relies on controlled bacteria permeation to enable total-load quantification. In contrast, the species-specific immunoassay confines antigen–antibody binding within the top PT-R7 layer, whose tighter pore structure and immobilized antibodies prevent cellular penetration, thereby preserving assay specificity.
Species-specific immunoassay performance
The HRP-antibody/TMB readout showed clear concentration-dependent responses for E. coli in both TSB and urine. In TSB, model fits indicated moderate quantitative behavior (Hill R2 = 0.7422; linear R2 = 0.7718) with a maximum CV of 20.15% at 106 CFU mL−1 (Fig. 3A, B and E). In urine, correlation improved (Hill R2 = 0.8643; linear R2 = 0.8764), and visual discrimination remained robust across 5 × 104 to 108 CFU mL−1, supporting applicability to clinically relevant samples (Fig. 3C–E). Importantly, the intrinsic coloration of urine did not compromise assay performance, as background-referenced grayscale analysis effectively decoupled sample color from enzymatic signal intensity. This robustness is critical for POCT deployment, where sample appearance can vary substantially between individuals. Performance trends align with immunoassay architecture. The PT-R7 detection layer affords high surface area and uniform wicking, which supports dense antibody immobilization and homogeneous wetting of the reaction zone. The downstream pore gradient increases residence time at the interface and limits downward transport of unbound species, which reduces background and sharpens contrast (Fig. 1A).
The remaining signal variability is likely associated with non-uniform liquid distribution, spatially heterogeneous color development within the porous paper matrix, and reaction-zone geometry rather than the antibody–antigen reaction alone. Variability at high bacteria loads is attributable to transport limitations and interfacial crowding within the porous detection layer rather than to the antibody–antigen reaction. Such spatial heterogeneity is intrinsic to paper-based assays and underscores the importance of standardized image-based averaging rather than sampling for quantitative analysis. Rapid capillary advection can shorten residence time and create uneven infiltration, while dense fluxes of cells and immune complexes partially occlude pores and saturate local binding sites. These conditions promote rapid local substrate depletion, accumulation of colored products, and occasional spillover into adjacent zones, all of which amplify signal spread. Incremental improvements are expected from tighter control of dispensing volume and rate, modest enlargement or repositioning of reaction spots to balance lateral spreading with vertical flow, and small adjustments to wash stringency to stabilize the baseline. Overall, the species-specific immunoassay maintains quantifiable behavior in TSB and urine and provides visually discriminable readouts over a clinically meaningful range.
Quantitative total bacteria performance
The WST-8–mPMS metabolic readout enabled rapid quantification of total bacteria load and extended the platform beyond species specificity. Using the two-layer format for this assay, both E. coli and S. aureus were measured across 5 × 104 to 108 CFU mL−1 in TSB and urine, demonstrating cross-species and cross-matrix applicability. Linear models showed consistent concentration–signal relationships. For E. coli, R2 = 0.84 in TSB and R2 = 0.82 in urine (Fig. 4A and B). For S. aureus, R2 = 0.72 in TSB and R2 = 0.78 in urine (Fig. 4C and D). The grayscale response decreased monotonically with increasing bacteria concentration, reflecting stronger color development at higher metabolic activity and enabling straightforward standard-curve calibration with smartphone imaging. These measurements confirmed that the WST-8–mPMS module can provide a species-independent metabolic readout of total viable bacteria load. Practical advantages include minimal sample preparation, compatibility with complex matrices, and stable operation with a front-end filtration layer that reduces particulates and fouling before the reagent layer. Although these module-level measurements were performed outside the final integrated cartridge, they provided the analytical basis for incorporating the WST-8–mPMS assay into the spatially separated metabolic detection spots of the integrated dual-mode device.
Integrated plastic container for species-specific and quantitative urine testing
By integrating the HRP-antibody/TMB immunoassay and WST-8–mPMS metabolic modules within one plastic-container-based vertical flow device, the cartridge delivered species-specific recognition and total-load assessment from a single urine sample. Each plastic container contained four spatially separated reaction sites: two configured for HRP-antibody/TMB immunodetection and two for WST-8–mPMS metabolic detection. This spatial separation allowed the two colorimetric reactions to proceed in parallel while minimizing chemical and optical interference.
When urine samples were tested, both the metabolic and species-specific channels showed concentration-dependent responses across the tested range. For the WST-8–mPMS metabolic module, statistically significant differences were observed between 5 × 104 and 107 CFU mL−1 and between 5 × 104 and 108 CFU mL−1 (Fig. 5A). For the HRP-antibody/TMB immunodetection module, a significant difference was observed between 5 × 104 and 108 CFU mL−1 (Fig. 5B). These results support threshold-oriented assessment across a concentration range spanning the clinically relevant bacteriuria level of 105 CFU mL−1, rather than representing a formally calculated limit of detection.
Variability in the prototype arose mainly from fluid distribution, spatially heterogeneous color development, and spot geometry rather than from the sensing chemistries themselves. The highest CV, around 24.82%, occurred at 5 × 104 CFU mL−1. More refined hydrophilic patterning, improved control of dispensing volume and rate, and minor adjustments to washing and absorbent stacking are expected to reduce signal spread while preserving simultaneous readouts. Overall, the integrated cartridge demonstrates that species identity and total bacteria load can be obtained in parallel within a compact, single-sample workflow, supporting multiplex POCT screening. Unlike conventional ELISA, which requires multiple manual handling steps, trained personnel, and benchtop instrumentation, the presented VFA emphasizes operational simplicity, smartphone-based readout, and early result visibility, making it more suitable for decentralized and resource-limited settings.
Limitations and comparison to existing methods
This study establishes a modular dual-mode vertical flow architecture, while further refinement of fluid distribution and reaction-zone geometry should improve linearity and precision across the full concentration range. Variability observed in some configurations arose chiefly from non-uniform liquid distribution, spatially heterogeneous color development within the porous paper matrix, and reaction-zone geometry, rather than from the underlying immunochemical or metabolic reactions alone. Urine also introduced matrix effects that may compress the dynamic range. These limitations are tractable engineering targets: more uniform sample delivery, refined hydrophilic patterning, modest adjustments to wash conditions, and matrix-matched calibration can all reduce spread and stabilize the baseline.
Several limitations should also be noted. First, the washing step remains necessary for the HRP-antibody/TMB immunodetection module to remove unbound HRP-conjugated antibodies and reduce nonspecific background. Although the sample injection plate with guided flow channels reduces direct manual targeting of the reaction zones, washing still adds operational complexity and may introduce user-dependent variation. Second, the present study used S. aureus as a Gram-positive non-target control for initial specificity evaluation; therefore, the current data should be interpreted as proof-of-concept E. coli-specific detection rather than comprehensive specificity validation across a full uropathogen panel. Broader pathogen-panel testing would be required before practical diagnostic translation. Third, 105 CFU mL−1 was used as a clinically relevant decision threshold for assay evaluation, and a formal LOD was not determined in this pilot study. To better position the proposed sensing strategy within the context of existing bacteria detection approaches, a comparison of representative methods is summarized in Table 1.
Table 1 Comparison of representative bacteria detection methods and the proposed dual-mode VFA
| Method |
Species identification |
Total bacteria quantification |
Detection time |
Instrument |
| Culture |
V |
V |
24–72 h |
Incubator |
| PCR |
V |
× |
1–3 h |
Thermocycler |
| Immunoassay |
V |
Limited |
2–4 h |
Reader |
| Metabolic assay |
× |
V |
1–2 h |
Reader |
| This work (dual-mode VFA) |
V |
V |
35–40 min |
Smartphone |
Conventional bacteria detection techniques differ in their ability to provide species identification, bacteria load estimation, analysis time, and instrumentation requirements. While culture-based assays can provide both identification and quantification, they typically require long incubation times (24–72 h). Molecular techniques such as PCR offer rapid species identification but generally do not directly quantify total bacteria load. Compared with plate ELISA methodology, the described platform reduces steps, time, and instrument dependence while preserving species specificity.1,2,11,15,20,21 In contrast, metabolic assays estimate overall bacteria activity but lack species specificity. These limitations highlight the need for diagnostic platforms capable of simultaneously reporting both bacteria identity and bacteria load within a rapid and simple workflow.
In practical use, pre-fabricated and reagent-preloaded VFA cartridges could be taken directly from storage and used for testing, without on-site fabrication or assay preparation. Under this use scenario, both species-specific immunoassay and total bacteria load readouts are obtained within approximately 35–40 min from a single urine sample, enabling simultaneous identification and quantification in one test. Compared with culture plate methodology, this approach offers rapid readouts relevant for triage, albeit without the definitive organism growth information that culture provides. Relative to qPCR, this approach trades ultimate sensitivity and genotypic detail for simpler workflows and lower cost.20 Compared with conventional LFAs, the vertical architecture enables controlled residence time, analyte retention, and multiplexable layouts. Despite modest R2 values in select cases, the present results support threshold-oriented bacteria screening across a clinically relevant concentration range rather than definitive diagnostic confirmation.
Despite the widespread use of HRP-antibody/TMB colorimetric immunoassays in LFAs and VFAs, most reported systems are designed as single-mode assays that provide only species-specific binding readouts.3,10,14–16,31,32 In these platforms, performance improvements are typically pursued through antibody optimization, multilayer delays, or signal amplification strategies. In contrast, the present work does not aim to introduce a new HRP–TMB chemistry or a new WST-8–mPMS reaction. Instead, it integrates these established chemistries into a dual-mode vertical flow platform that simultaneously reports species identity and total bacteria load through spatially separated detection spots within one device. This system-level integration enables contextual interpretation at clinically relevant thresholds, which is not addressed by conventional single-mode HRP–TMB-based LFA or VFA.
Conclusions
In this study, we developed a dual-mode VFA that integrated species-specific immunodetection and quantitative total bacteria assessment within one integrated paper-based vertical flow device. By combining an HRP-antibody/TMB immunoassay with a WST-8–mPMS metabolic readout, the system enabled complementary immunological and metabolic information to be obtained from a single urine sample through spatially separated detection spots. The platform demonstrated threshold-oriented detection across 5 × 104 to 108 CFU mL−1, spanning the clinically relevant bacteriuria decision level of 105 CFU mL−1, in both TSB and human urine, supported by smartphone-based grayscale analysis without dedicated instrumentation. Initial species-specific detection of E. coli was demonstrated with minimal cross-reactivity against the tested Gram-positive non-target bacterium, S. aureus, while the metabolic module enabled species-agnostic quantification of total viable bacteria load across multiple matrices. These results support the feasibility of using the platform to simultaneously report target bacterial identity and total bacteria load within a single integrated test. The findings establish a practical framework for POCT bacteria screening and triage, where rapid assessment at clinically actionable thresholds can guide downstream confirmatory testing. While further engineering refinements and broader pathogen-panel validation may improve precision and scalability, the present work demonstrates that dual-mode bacteria detection can be achieved with minimal assay complexity, supporting decentralized and resource-limited diagnostic settings.
Institutional review board
Approval was obtained from the Institutional Review Board of National Cheng Kung University Hospital in Taiwan (NCKU IRB No. B-BR-109-056). Informed consent was obtained from the subjects involved in the study.
Author contributions
Shu-Yun Sheu: writing – review & editing, writing – original draft, methodology. Ching-Fen Shen: writing – review & editing, writing – original draft, conceptualization, supervision, methodology. Chao-Min Cheng: writing – review & editing, writing – original draft, supervision, conceptualization.
Conflicts of interest
The authors have no interests to declare.
Abbreviations
| BSA | Bovine serum albumin |
| CFU | Colony-forming unit |
| CV | Coefficient of variation |
| E. coli | Escherichia coli |
| EDTA | Ethylenediaminetetraacetic acid |
| ELISA | Enzyme-linked immunosorbent assay |
| HRP | Horseradish peroxidase |
| LFA | Lateral flow assay |
| LOD | Limit of detection |
| LPS | Lipopolysaccharide |
| mPMS | 1-Methoxy phenazine methosulfate |
| PBS | Phosphate-buffered saline |
| PCR | Polymerase chain reaction |
| POCT | Point-of-care testing |
| RGB | Red, green, and blue channel |
| ROI | Region of interest |
| S. aureus | Staphylococcus aureus |
| SEM | Scanning electron microscope |
| TMB | 3,3′,5,5′-Tetramethylbenzidine |
| TSB | Tryptic soy broth |
| UTI | Urinary tract infection |
| VFA | Vertical flow assay |
| WST-8 | Water-soluble tetrazolium salt-8 |
Data availability
The data and materials for this study are available from the corresponding authors upon reasonable request.
Supplementary information (SI) is available. See DOI: https://doi.org/10.1039/d6lc00265j.
Acknowledgements
This research was funded by Taiwan's National Science and Technology Council (113-2221-E-007-021-MY3, 114-2622-B-006-007 & 113-2622-E-007-019).
References
- S.-Y. Sheu, C.-M. Cheng and C.-F. Shen, Current status and future perspectives of vertical flow assays, Trends Biotechnol., 2026, 44, 40–52 Search PubMed.
- N. Jiang, R. Ahmed, M. Damayantharan, B. Ünal, H. Butt and A. K. Yetisen, Lateral and Vertical Flow Assays for Point-of-Care Diagnostics, Adv. Healthcare Mater., 2019, 8(14), 1900244 Search PubMed.
- P. Chen, M. Gates-Hollingsworth, S. Pandit, A. Park, D. Montgomery, D. AuCoin, J. Gu and F. Zenhausern, based Vertical Flow Immunoassay (VFI) for detection of bio-threat pathogens, Talanta, 2019, 191, 81–88 CrossRef CAS PubMed.
- B. Liu, Q. Liu, K. Zhong, W. Wu, S. Zheng, W. Yao, B. Gao and F. Sun, Enhanced SERS-based vertical flow assay for high sensitivity multiplex analysis of antibiotics, Microchem. J., 2024, 199, 110008 Search PubMed.
- M. Moumita, K. Shankar, P. Abhiman and B. Shamasundar, Development of a sandwich vertical flow immunogold assay for rapid detection of oxytetracycline residue in fish tissues, Food Chem., 2019, 270, 585–592 CrossRef CAS PubMed.
- S. Kim, Y. Hao, E. A. Miller, D. M. Tay, E. Yee, P. Kongsuphol, H. Jia, M. McBee, P. R. Preiser and H. D. Sikes, Vertical flow cellulose-based assays for SARS-CoV-2 antibody detection in human serum, ACS Sens., 2021, 6(5), 1891–1898 CrossRef CAS PubMed.
- R. Lei, D. Wang, H. Arain and C. Mohan, Design of gold nanoparticle vertical flow assays for point-of-care testing, Diagnostics, 2022, 12(5), 1107 CrossRef CAS PubMed.
- H.-A. Joung, Z. S. Ballard, A. Ma, D. K. Tseng, H. Teshome, S. Burakowski, O. B. Garner, D. Di Carlo and A. Ozcan, based multiplexed vertical flow assay for point-of-care testing, Lab Chip, 2019, 19(6), 1027–1034 Search PubMed.
- J. L. Vaitukaitis, G. D. Braunstein and G. T. Ross, A radioimmunoassay which specifically measures human chorionic gonadotropin in the presence of human luteinizing hormone, Am. J. Obstet. Gynecol., 1972, 113(6), 751–758 CrossRef CAS PubMed.
- J. Park and J.-K. Park, Pressed region integrated 3D paper-based microfluidic device that enables vertical flow multistep assays for the detection of C-reactive protein based on programmed reagent loading, Sens. Actuators, B, 2017, 246, 1049–1055 Search PubMed.
- R. Lei, H. Arain, M. Obaid, N. Sabhnani and C. Mohan, Ultra-sensitive and semi-quantitative vertical flow assay for the rapid detection of interleukin-6 in inflammatory diseases, Biosensors, 2022, 12(9), 756 CrossRef CAS PubMed.
- J. Wang, K. Mu, H. Wei, H. Chen, Y. Wang, W. Zhang and Z. Rong, based multiplex colorimetric vertical flow assay with smartphone readout for point-of-care detection of acute kidney injury biomarkers, Sens. Actuators, B, 2023, 390, 134029 Search PubMed.
- O. Clarke, B. Goodall, H. Hui, N. Vats and C. Brosseau, Development of a SERS-based rapid vertical flow assay for point-of-care diagnostics, Anal. Chem., 2017, 89(3), 1405–1410 CrossRef CAS PubMed.
- D. M. Tay, S. Kim, Y. Hao, E. H. Yee, H. Jia, S. M. Vleck, M. Chilekwa, J. Voldman and H. D. Sikes, Accelerating the optimization of vertical flow assay performance guided by a rational systematic model-based approach, Biosens. Bioelectron., 2023, 222, 114977 CrossRef CAS PubMed.
- A. Goncharov, H. A. Joung, R. Ghosh, G. R. Han, Z. S. Ballard, Q. Maloney, A. Bell, C. T. Z. Aung, O. B. Garner and D. D. Carlo, Deep Learning-Enabled
Multiplexed Point-of-Care Sensor using a Paper-Based Fluorescence Vertical Flow Assay, Small, 2023, 19(51), 2300617 CrossRef CAS.
- M. Kaur and E. Eltzov, Optimizing effective parameters to enhance the sensitivity of vertical flow assay for detection of Escherichia coli, Biosensors, 2022, 12(2), 63 CrossRef CAS PubMed.
- Z. Gu, H. Chang, G. Yang, B. Xu, B. Miao and J. Li, An integrated electronic tag-based vertical flow assay (e-VFA) with micro-sieve and AlGaN/GaN HEMT sensors for multi-target detection in actual saliva, Analyst, 2024, 149(16), 4267–4275 Search PubMed.
- M. T. Nüesch-Inderbinen, M. Baschera, K. Zurfluh, H. Hächler, H. Nüesch and R. Stephan, Clonal diversity, virulence potential and antimicrobial resistance of Escherichia coli causing community acquired urinary tract infection in Switzerland, Front. Microbiol., 2017, 8, 2334 Search PubMed.
- B. Foxman, The epidemiology of urinary tract infection, Nat. Rev. Urol., 2010, 7(12), 653–660 Search PubMed.
- M. Tahmasebi, T. Bamdad, W. E. Svendsen and M. Forouzandeh-Moghadam, An enzymatic nucleic acid vertical flow assay, Anal. Bioanal. Chem., 2022, 414(12), 3605–3615 Search PubMed.
- R. Chen, X. Du, Y. Cui, X. Zhang, Q. Ge, J. Dong and X. Zhao, Vertical flow assay for inflammatory biomarkers based on nanofluidic channel array and SERS nanotags, Small, 2020, 16(32), 2002801 Search PubMed.
- E. H. Kass, Asymptomatic infections of the urinary tract, J. Urol., 2002, 167(2), 1016–1020 CrossRef PubMed.
- M. L. Wilson and L. Gaido, Laboratory diagnosis of urinary tract infections in adult patients, Clinical infectious diseases, 2004, 38(8), 1150–1158 Search PubMed.
- G. Schmiemann, E. Kniehl, K. Gebhardt, M. M. Matejczyk and E. Hummers-Pradier, The diagnosis of urinary tract infection: a systematic review, Dtsch. Arztebl. Int., 2010, 107(21), 361 Search PubMed.
- M. Serhan, D. Jackemeyer, K. Abi Karam, K. Chakravadhanula, M. Sprowls, P. Cay-Durgun and E. Forzani, A novel vertical flow assay for point of care measurement of iron from whole blood, Analyst, 2021, 146(5), 1633–1641 Search PubMed.
- J. Li and J. Macdonald, Multiplexed lateral flow biosensors: Technological advances for radically improving point-of-care diagnoses, Biosens. Bioelectron., 2016, 83, 177–192 CrossRef CAS PubMed.
- R. Chen, J. Wu, H. Wang, M. S. Nisar, Y. Li, Y. Chen, Y. Mao, X. Nan, F. Zhang and L. Yang, Fe3O4@ Au magnetic SERS nanotags-based vertical flow immunoassay for simultaneous detection of fumonisin B1, aflatoxin B1 and deoxnivalenol, Anal. Chim. Acta, 2025, 1348, 343837 Search PubMed.
- V. Borse and R. Srivastava, Process parameter optimization for lateral flow immunosensing, Mater. Sci. Energy Technol., 2019, 2(3), 434–441 Search PubMed.
- M. Sajid, A.-N. Kawde and M. Daud, Designs, formats and applications of lateral flow assay: A literature review, J. Saudi Chem. Soc., 2015, 19(6), 689–705 CrossRef.
- N. Abu, N. Mohd Bakhori and R. H. Shueb, Lateral flow assay for hepatitis B detection: a review of current and new assays, Micromachines, 2023, 14(6), 1239 CrossRef PubMed.
- K. M. Koczula and A. Gallotta, Lateral flow assays, Essays Biochem., 2016, 60(1), 111–120 Search PubMed.
- B. O'Farrell, Lateral Flow Technology for Field-Based Applications—Basics and Advanced Developments, Top. Companion Anim. Med., 2015, 30, 139–147 CrossRef PubMed.
- N. Alam, L. Tong, Z. He, R. Tang, L. Ahsan and Y. Ni, Improving the sensitivity of cellulose fiber-based lateral flow assay by incorporating a water-dissolvable polyvinyl alcohol dam, Cellulose, 2021, 28(13), 8641–8651 CrossRef CAS PubMed.
- T. Tezcan, M. Calimci, I. H. Boyaci and U. Tamer, UV-ozone treated glass fiber based lateral flow DNA extraction platform integrated with LAMP for rapid detection of pathogen bacteria in whole blood, Microchem. J., 2024, 206, 111487 CrossRef CAS.
- C.-H. Chen, Y.-H. Liao, M. Muljadi, T.-T. Lu and C.-M. Cheng, Potential Application of the WST-8-mPMS Assay for Rapid Viable Microorganism Detection, Pathogens, 2023, 12(2), 343 CrossRef CAS PubMed.
- C. Parolo, A. Sena-Torralba, J. F. Bergua, E. Calucho, C. Fuentes-Chust, L. Hu, L. Rivas, R. Álvarez-Diduk, E. P. Nguyen and S. Cinti, Tutorial: design and fabrication of nanoparticle-based lateral-flow immunoassays, Nat. Protoc., 2020, 15(12), 3788–3816 Search PubMed.
|