Open Access Article
Alison M. Skelley
*ab,
Yasna Behmardia,
Luke F. Petersona,
David W. Inglis
c,
Mabel Shehadaa,
Laurissa Ouaguiaa,
Khushroo Gandhia,
Roberto Campos-Gonzáleza and
Tony Warda
aCurate Biosciences, USA
bZeon Specialty Materials, 25 Metro Drive, Suite 238, San Jose, CA 95110, USA. E-mail: alison.skelley@zeonsmi.com
cSchool of Engineering, Macquarie University, Sydney, NSW 2109, Australia
First published on 2nd April 2026
Autologous CAR-T cell therapy has demonstrated remarkable clinical efficacy in hematologic malignancies, yet its broader application remains limited by complex, labor-intensive cell therapy manufacturing and inconsistent product quality. We describe a novel microfluidic cell separation platform based on deterministic lateral displacement (DLD), integrated into a fully automated, closed-system instrument (Curate system). N = 150 leukopacks were processed at a flow rate of 400 mL h−1 with processed volumes up to 250 mL, white blood cell concentrations up to 168 M mL−1 and total white blood cell counts up to 24 billion white blood cells. Compared to Ficoll®-based density gradient centrifugation, microfluidic DLD processing yielded significantly higher leukocyte recovery (88% vs. 58%), superior platelet and red blood cell depletion, and reduced CD69+ T cell activation. Flow cytometric analysis revealed improved phenotypic preservation across key T cell subsets, including naïve and central memory populations. Cytokine profiling demonstrated enhanced washing efficiency, with markedly lower levels of biologic response modifiers. DLD-purified T cells exhibited enhanced expansion kinetics and greater yield, supporting improved manufacturing outcomes. These findings position microfluidic DLD-based processing as a clinically relevant, scalable alternative to conventional methods, with potential to improve consistency, potency, and accessibility of CAR-T therapies.
000.1–3 Despite therapies being available since 2017, and with 9 therapies currently on the market, fewer than 40
000 patients have been treated.4 One major hurdle is the very manual and lengthy cell therapy manufacturing process. Typical manufacturing protocols first perform a debulking step to purify white blood cells (WBCs) from an apheresis product, followed by T cell isolation using immunomagnetic beads. The T cells are then genetically engineered followed by expansion in culture to achieve the desired number of cells. Patient leukapheresis collections frequently contain the minimum number of T cells to create multiple therapeutic doses, but due to the inefficiencies of the manufacturing process, multiple days of ex vivo expansion are required to achieve the desired number of cells.5 Several steps in the process require manual manipulation of the sample, and cell therapy manufacturing facilities commonly do not have the capacity or staff to manufacture more than a few therapies at a time.6 As a result, strict evaluation criteria for patient eligibility and long wait times – upwards of 6 months – are a common reality.6 Biological variability in starting materials and complex manufacturing processes result in inconsistent therapeutic outcomes; furthermore, up to 4–7% of manufacturing processes fail and patients never receive their therapy.7 From a clinical perspective, cell therapy is often a therapy of last resort, and yet a significant portion of patients may require additional apheresis collections, bridging therapy, or because of a failed manufacturing process may not get their therapy at all.8,9
In typical CAR-T cell manufacturing, an initial debulking step is performed post-apheresis collection and prior to T cell selection and activation. Commonly used systems include density gradient centrifugation (Sepax™ C-Pro, Cytiva), counterflow centrifugation (Gibco™ CTS™ Rotea™, Thermo Fisher Scientific), and spinning membrane instruments (Lovo®, Fresenius Kabi). Although these instruments recover WBCs, they do so with variable cell losses, and with variable removal of erythrocytes and/or platelets. Contamination with platelets can alter the T cells and suppress proliferation during culture.10 It is also widely known that centrifugation induces stress and non-desirable cell to cell interactions.11–13 Additional treatment with ammonium chloride to lyse remaining erythrocytes can also negatively impact cell quality.13–15
Recent studies are linking cell quality with therapeutic outcomes. Lower levels of functional exhaustion markers in apheresis starting materials corresponded with a higher likelihood of therapeutic success.16 Longer ex vivo culture time was associated with differentiated T cells and depletion of stem cell memory (Tscm) and naïve phenotypes,17 potentially leading to poor efficacy due to T cell exhaustion and apoptosis.18 Higher percentages of central memory T cells (TCM)19,20 and lower levels of regulator T cells (TREG)21 resulted in greater in vivo persistence and greater anti-tumor activity. While alleviation of manufacturing hurdles and cost reduction is a primary focus, careful attention must be given to optimizing cell quality both into and out of the manufacturing process to achieve the best therapeutic outcome.
Novel microfluidic technologies are showing promise for improving the cell therapy manufacturing process. High-throughput purification of WBCs from blood and blood products has been demonstrated by numerous different technologies including deterministic lateral displacement (DLD),22 spiral inertial microfluidics,23 multi-channel inertial designs,24 inertial separation arrays25,26 and other passive, active, and hybrid approaches.27 Other technologies are focused on later stages of the manufacturing process, including T cell isolation, transfection and expansion.28,29 While these technologies have demonstrated greater efficacy than their non-microfluidic counterparts, and may be optimized for high throughput operation, very few, if any of the technologies have demonstrated full clinical-scale operation from leukopacks (apheresis collections), operating at hundreds of mL and manipulating billions of cells.
Previously, we described a microfluidic chip that used the principles of deterministic lateral displacement (DLD) to process blood and blood products for the production of CAR-T cells.30,31 This manually operated proof-of concept scale device demonstrated superior purification and phenotypic advantages in the microfluidic product compared to density gradient centrifugation. In this paper, we focus on the next generation of microfluidic DLD cell separation devices, optimized and scaled up to process full leukopacks (standard apheresis units with volumes of up to 250 mL) within an hour. We also describe the integration of these massively parallel microfluidic arrays into a novel single-use disposable cassette and its use on an easy-to-use automated prototype system. We characterize the performance of this system through head-to-head runs with density gradient centrifugation (one of the most widely used methods for debulking post-apheresis) and show the functional benefits of cells isolated using the microfluidic approach.
Running buffers were consisted of an albumin mixture in phosphate-buffered saline (PBS) or Plasma-Lyte®-A Injection, pH 7.4 (PLA, VWR cat. no. 80089-818). Concentrated bovine serum albumin (BSA) or human serum albumin (HSA, Grifols Albumin (Human) 25%, USP, Plasbumin®-25, cat. no. 13533-692-71) was injected into bags of sterile PLA or added to PBS and sterile filtered. The optimal final protein concentration was 2–4%. In some runs, cell culture media were used, such as TexMACS™ (Miltenyi Biotec) or CTS™ OpTmizer™ (Thermo Fisher Scientific). If reagents were prepared more than a day prior to run, they were sterile filtered (0.2 μm pore size) and then loaded into sterile bags. Sterile collection bags for the product and waste were connected to the cassette prior to use via tube welding.
A benzonase pre-treatment was performed prior to processing the samples. Benzonase nuclease HC (final concentration 50 units mL−1, Millipore, cat. no. 71205-3) and MgCl2 (final concentration 5 mM) were added to the leukopack followed by 1 h incubation at room temperature while rocking. Before processing, EDTA (final concentration 5 mM) was added followed by another 15 min incubation at room temperature.
All samples used in this study were purchased from commercial organizations that collect healthy donor blood and apheresis materials under the IRB-approved protocols in compliance with the U.S. Department of Health & Human Services regulations (45 CFR 46). The IRB protocols included documented informed consent and privacy and confidentiality protections in compliance with HIPAA and 21 CFR Part 50. All donors provided informed consent to the various commercial organizations for the collection and use of their blood for research purposes, and all samples were de-identified prior to purchase. This study is considered exempt from IRB oversight and approval because the use of de-identified samples does not constitute human-subjects research.
https://grants.nih.gov/grants/policy/hs/private-information-biospecimens-flowchart.pdf.
:
1 in cell culture media and layered on top of an equal volume of Ficoll®. The sample was centrifuged at 400× g for 35 min with no break. The PBMC layer was carefully removed using a pipet. The PBMC layer was washed via resuspending in 50-fold excess media, followed by pelleting by centrifugation at 400× g for 10 min (wash 1). The pellet was again resuspended in 50-fold excess media and then centrifuged at 200× g for 10 min (wash 2). The final pellet (PBMC) was resuspended in media to the original volume. This manual process took ∼1.5 h to process only 10–20 mL of the initial sample.
:
1 beads per T cell for 1 h at 37 °C.33 After the incubation, the bead/cell slurry was placed in a magnet for 5 minutes and then the supernatant was discarded. The slurry was resuspended to a T cell concentration of
1 × 106 mL−1, or the desired concentration, in full media containing TexMACs™, 10% fetal bovine serum (FBS, Phoenix Scientific, PS-100) or 5% HSA, plus 5 ng mL−1 of IL-7 and IL-15 (Biolegend cat no. 777704 and 570308), and 1% penicillin/streptomycin (Thermo Fisher Scientific cat. no. 15140122). The cells were plated in 6-well GRex plates or 1 L G-Rex containers (Wilson Wolf Manufacturing cat. no. 80240 M and RU81100 respectively) and placed at 37 °C and 5% CO2 in a humidified incubator.
The array used a mirrored geometry to maximize the concentration of the collected product, with two arrays on either side of a central collection channel (also called a bypass channel, Fig. 1B). As cells were deflected by the arrays, they moved towards the central collection channel. Only cells and fluid in this central collection channel and immediately adjacent columns were collected as products, and the remainder of the fluid in the array went to waste. A small portion of the buffer was also rejected to waste to make sure that the collected product was free of diffusible species that may have crossed the boundary.
The design of the micropost array was optimized for throughput and capacity, anticipating the high concentrations of cells present in apheresis samples. To improve throughput per lane, we created an asymmetric DLD array35 with wider gaps perpendicular to the flow direction (also referred to as the lateral gap, or GL) and narrower gaps parallel to the flow direction (also called the downstream gap, or GD). Widening the lateral gaps between microposts reduced the fluidic resistance within the array and had the benefit of making the array more resistant to biofouling. We also reduced the spatially average shear rate on the cells by using microposts with sharp vertices facing into the gap22,36 (SI). To maintain the DC of ∼4 μm while avoiding unnecessarily small gaps, a challenge for high-volume manufacturing of microfluidic devices, we used microposts with a hexagonal cross-section (as-fabricated: 18.8 × 39.2 um). This created an asymmetric gap profile (thus minimizing the flow lane width in the minor flux direction) but did so by elongation as opposed to constriction (Fig. 1C). The downstream gap between faces of the hexagon posts (GD) was 16.2 μm, and the lateral gap between the corners of the hexagons (GL) was 21.8 um. These feature sizes were well within the manufacturability limits for compression molding as shown by the SEM photo of the mirrored array constructed with COP plastic (Fig. 1D).
The optimal array tilt angle was a tradeoff between gap size, array length and multi-lane manufacturability. Arrays with a smaller tilt angle (lower horizontal shift of microposts between rows and the net larger number of rows per repeating array unit) result in narrower flow lanes following the minor flux direction. This directly correlates with a smaller functional DC. However, smaller array tilt angles also require longer DLD arrays to obtain the same lateral displacement of cells. We wanted to fabricate numerous lanes on a 6″ wafer which limited the maximum array length, especially near the perimeter of the wafer. We also wanted to increase the redundancy of the array, creating an array with ∼2× the required length to transit from the negative to positive boundary. While this modification did increase fluidic resistance, there was a positive impact on net recovery when processing higher cell concentrations from apheresis materials, allowing extra time to collect cells that had been bumped off their trajectory due to cell–cell interactions in the array. A tilt slope ε of 1/34 radians was chosen as a compromise of maintaining larger gaps and enabling a 2× redundancy while also keeping the net array length under 42 mm.
We calculated the DC using Zeming et al.'s35 method for calculating an effective epsilon (εeffective) that factors in different gaps downstream (parallel) vs. lateral (perpendicular) to flow directions. The equation uses a ratio of the downstream (PD) vs. lateral pitch (PL), where the pitch is the sum of the gap plus post size (width or length) in that direction:
| PLateral = GapL + PostL |
(θ) = 1/34.
We then used the DC equation from Inglis et al.:34
| DC = 1.4 × GL × ε0.48 |
An additional feature was added at the inlet of each lane to help minimize the impact of any biofouling. A partial separator wall37 was added between the sample and buffer arrays, extending only 7.8 mm down from the array inlet (Fig. 2A). This distance was short enough relative to the full length (41.5 mm) to have no impact on cells transiting into the central collection channel. The separator wall compartmentalized any biofouling at the entrance to the array and prevented the sample from spilling over into the buffer stream and contaminating the product. At the terminus of the separator wall, guide posts were used to keep fluid streams organized as the arrays were merged into one larger array.
Both positive boundaries (the central collection channel and outer edge of the separator wall) and negative boundaries (the outer edge of the bulk array; the inner wall of the separator channel) were adjusted to maintain array functionality. These boundary modifications were performed according to the methods of Inglis et al., but with modifications to comply with minimum manufacturable gaps.38,39 Standard boundary correction at the negative edge requires gaps that are considerably smaller than the main array gaps. These could not be manufactured, so the shape of some boundary obstacles was changed (squared and elongated) to yield a longer and higher resistance flow path for the minor flux. At the positive boundary, the posts were also elongated to both deliver the major flux into the central collection channel and then also slowly allow flow to escape back into the array in subsequent rows, creating a DC that mimicked that of the bulk array (Fig. 2A, bottom). The boundaries were intentionally staggered such that row 1 for the negative boundary did not correspond with row 1 for the positive boundary. This prevented both major boundary adjustments from occurring in the same row.
Total system throughput and capacity were accomplished through massive parallelization, both in the number of lanes per layer and stacking multiple layers together. 34 lanes were fabricated in a single layer (Fig. 2B). Two of these layers were lidded and then laser welded to a central manifold layer containing a common set of manifold channels, one for each of the sample, buffer, product and waste. Two of these manifold units were then connected in parallel with 4 y-connectors, one for each manifold channel. The result was 136 lanes across 4 microfluidic layers operating in parallel, with only 2 inputs and 2 outputs from the cassette. Even fluid distribution was accomplished by keeping the manifold channels large with low fluidic resistance relative to the array lanes. In addition, identical path lengths were maintained for each lane by feeding the sample and buffer into the manifold ahead of lane 1 and collecting the product and waste from the manifold after lane 34. The throughput of the final geometry was 1000 mL h−1, with sample and buffer inputs comprising 400 mL h−1 and 600 mL h−1, respectively, and product and waste outputs generated at 300 mL h−1 and 700 mL h−1, respectively.
Each pump rotor featured dual sets of rollers to minimize pulsatile flow, critical for establishing stable fluid boundaries between the sample and co-flowing buffer in the microfluidic array. Briefly, a single fluid line on the cassette was split into two paths, travelled through peristaltic tubing over two sets of rollers 180 degrees out of phase, and then was recombined, delivering smoother flow. The microfluidic array separator wall described above also helped suppress flow oscillations, maintaining a stable boundary once the fluid streams were combined into the wider array.
The Curate system featured 2 modes of operation (Fig. 3B). In the standard mode of operation (separation and wash), white blood cells were deflected into running buffer and collected in the product bag. The output volume of the product bag was very similar to the input volume of the sample (some adjustment due to priming and flushing steps). An alternative mode of operation increased the concentration of white blood cells in the product by recirculating the collected product back through the array as the collection buffer. This was facilitated by a recirculation path on the cassette (dashed lines, Fig. 3A) and easily controlled by actuating just 2 valves open/closed. This mode of operation, titled separation, wash and concentration, required no additional processing time and the output concentration was adjusted based on the downstream assay requirements. Load cells on all bag hangers allowed real-time monitoring of total fluid volumes and flow rates. Final target volumes for separation, wash and concentration were achieved by triggering the recirculation step based on the load cell measurement for the product bag. Since the sample input volume was larger than the product output volume, there was a net decrease in product bag volume during recirculation. Thus, the optimal volume was achieved by actuating between run and recirculation modes, keeping the bag mass between an upper and a lower threshold. This flip/flop valve state continued until all the sample was processed, and enabled the same target output volume regardless of the input volume, without having any impact on processing time. An example of the sensor data for a recirculation run is shown in Fig. S3.
The key separation metrics of %WBC recovery and % process efficiency (product vs. waste) were evaluated for dependency on input concentration or on total cells processed. Surveying the N = 150 samples, leukopack concentrations up to 168 M mL−1 (average of 56 ± 24 × 106 M mL−1), total WBC cellularity up to 24 billion leukocytes (average 8.5 ± 3.8 billion leukocytes) and volumes up to 245 mL (average of 150 ± 50 mL) were processed without dilution or modification of the protocol. While a slight decrease in %WBC recovery was observed with both increasing input concentrations and increasing input total WBCs processed, the % process efficiency showed no impact with very few WBCs in the waste fraction (t-test, deviation from 0 slope not significant). This indicates that with higher concentrations and/or total WBCs processed, there were more cells lost to the system, likely trapped in the upstream filter due to increased cell clumping that occurs at higher cell counts. Materials that passed the upstream filter were still separated with equivalent efficiency compared to lower concentrations and counts.
During the run phase, the in-line pressure detectors upstream of the DLD were used to monitor the run progress for any biofouling. If the pressure rose above 150 kPa then the system automatically scaled the total throughput from 1000 mL h−1 to 600 mL h−1. If the pressure continued to rise and the next threshold of 170 kPa was passed, the system aborted the run and notified the user while flushing any sample in the cassette back into the sample bag. This remaining sample could then be loaded on a new cassette, if so desired by the operator. Of the 150 samples loaded, only 9 runs were auto scaled by the instrument, and of those runs, 7 ran to completion at the scaled flow rate of 240 mL h−1. The other 2 runs still processed >85% of the sample before aborting and recovered >80% of the total white blood cells (WBCs) loaded. The performance metrics from those 9 runs are included in the N = 150 data set and details of a failed run, including sensor data, are described in the SI.
:
cell interaction and tonic signaling, N = 21 paired products were phenotyped by flow cytometry within two hours of processing. Of the N = 21 samples, N = 5 DLD samples were processed using pressure-driven flow to confirm that peristaltic pumping had no impact on activation in the microfluidic products. CD69+ was chosen as a marker to indicate activation levels in the cell population. Basal product activation state was measured prior to any cell engineering. As shown in Fig. 7A, the Ficoll® prepared cells had on average ∼2× higher percentage of CD69+ T cells (p = 0.0018). The Ficoll® prepared cells also had more variability across the N = 21 samples, with one product showing >25% of the CD3+ cells that were also CD69+. No differences were observed between DLD products collected via peristaltic and pressure driven flows (SI).
A randomly-selected smaller subset of samples (paired, N = 8) were further phenotyped to investigate CD69+ expression within the T cell effector differentiation pathway subsets. The 4 major subtypes are naïve T cells (TN, CCR7+CD95-CD45RA+), central memory T cells (TCM, CCR7+CD95+CD45RA-CD62L+), effector memory T cells (TEM, CCR7-CD95+CD45RA-CD62L-), and terminally-differentiated effector memory T cells (TEMRA, CCR7-CD95+CD45RA+CD62L-).40 Each of these subtypes play a different role when engineered with CAR-T, ranging from short-lived but potent effector function (TEM) to long-lived in vivo expansion and proliferation (TCM).41 The %CD69+ was determined within each subtype and the plot is shown as a ratio of DLD-prepared cells vs. Ficoll®-prepared cells. Equivalent %CD69+ activation is indicated by the red dotted line (value = 1). In all subtypes analyzed, a similar pattern was observed. The Ficoll® prepared CD4+ and CD8+ cells on average showed higher %CD69+ activation, resulting in mean ratio values <1. Using a one-sample, two-tailed t-test with a significance of 0.05, the individual ratios were compared to a theoretical value of 1. The CD4+ TEMRA, CD8+ TN and CD8+ TCM samples showed a ratio significantly different than 1, indicating that the microfluidic DLD cells had lower activation within those subsets.
The input material also contained alarming levels of some cytokines and chemokines, emphasizing why an efficient, fast wash is a critical first step. The highly potent proinflammatory chemokine RANTES, produced by activated T cells and platelets, was found in the source apheresis material at a level almost 2000-fold higher than the accepted biologically active level. Following processing, Ficoll® with the manufacturer recommended two wash steps eliminated ∼75% as compared to the microfluidic DLD processing, which eliminated ∼95% of RANTES. While this analysis is only from a small number of samples, it highlights how the different cellular purification efficiencies combined with the degree of soluble factor elimination of potent biological response modifiers are potentially impacting activation, differentiation and response of T cells as well as other cells present.
Here, we described a high throughput microfluidic DLD cell separation system as a method to eliminate some of the inconsistencies and impurities while minimizing losses during the earliest stages of CAR-T cell manufacturing. Inherent to DLD, size-based processing is an exceptionally efficient washing process. The single-pass separation–wash process eliminated the cell losses and forced cell–cell interactions from repeated centrifugation, and minimized non-desirable activations coming from a variety of soluble factors.
The device presented here achieved the highest WBC processivity of any microfluidic device to date. Non-equilibrium inertial arrays come close to our throughputs, showing high volumetric throughput (400 mL in less than 3 h, net ∼0.2–0.5 million WBCs per s),25 but similar devices for isolating CTCs from leukopacks have not yet demonstrated equivalent per cell metrics, with volumetric processivity (∼75 mL h−1) capacity (65 mL and 5 billion WBCs)26,47 and net WBC processivity of ∼1.5 million WBCs per s. Here, we demonstrated processing leukopacks up to 245 mL with up to 24 billion WBCs and at rates of up to 19 million WBCs per s. This is 5× more WBCs and a >10× improvement in the sorting rate, achieving the end goal of full-scale processing for cell therapy applications in under an hour.
Numerous factors contributed to the consistent processing performance. Proper handling and treatment of the sample was critical: keeping the sample on a rocker at 4 °C overnight minimized platelet activity and prevented cell clumps from forming. Pre-treatment with benzonase ensured that DNA and RNA from any dead cells were degraded, minimizing the blockages and biofouling. The running buffer contained protein (BSA/HSA) which blocked the COP surface during the priming steps, helping to prevent biofouling. The upstream 20 μm filter on the cassette also played a role, removing large cell clumps before they entered the microfluidic device, protecting the micropost arrays. Most importantly, the DLD contained key features designed specifically for biologically complex, high cellularity concentrations. The array was constructed with larger gaps to increase processivity and minimize biofouling. Elongated hexagon microposts both enabled an asymmetric array geometry and maintained larger gaps for easier manufacturing. The extra array length (2× redundancy) maintained high recoveries even with very high input concentrations. Finally, the separator walls compartmentalized any biofouling, preventing product contamination.
In comparison with standard centrifugation methods, microfluidic DLD cell separation is faster and superior in terms of platelet and red blood cell reduction while requiring less manual manipulation. Efficient platelet removal is the key for CAR-T cell manufacturing where processes may induce changes in T cells and compromise their proper development.10 The %PLT removal and CD69+ activation data shown here supports that trend with ∼2× lower CD69+ activation in the DLD isolated T cells. A smaller set of samples analyzed for T cell subtypes also demonstrated a lower activation state in the DLD-prepared cells, with significantly less activation vs. Ficoll® demonstrated for CD4+ TEMRA, CD8+ TN and CD8+ TCM cells.
Furthermore, the efficient microfluidic DLD wash process, in a small-scale study, resulted in lower levels of cytokines that negatively impact CAR-T therapies in the product. These cytokines have a range of known effects from inducing T-reg differentiation to suppressing T cell expansion and proper generation of CAR-T (Table S1 and SI data file). The reason for the lower levels of cytokines in the DLD-processed samples may be attributed to different sources. Some cytokines such as RANTES were observed in very high concentrations in the input material. RANTES is produced primarily by activated leukocytes (including T cells) during an immune response. The high level in the input material indicates that the apheresis process is causing stress. DLD processing resulted in a more complete washing of the sample, removing ∼95% vs. only 75% for Ficoll®. However, with other cytokines, Ficoll® washing was able to achieve higher wash efficiency (Table S1). Thus, the incomplete removal of RANTES here indicates that it is still actively generated in the Ficoll® sample, supporting the hypothesis that DLD-processing is a less stressful process. Alternatively, some cytokines such as PDGF-AA and BB were at higher levels in the Ficoll® products than in the input material. These cytokines are primarily platelet-derived, which indicates that the greater removal of platelets in the DLD process is playing a role, and that the platelets in the Ficoll® sample are actively releasing cytokines. Finally, considering the panel as a whole, Ficoll® was unable to achieve the wash efficiency obtained by the DLD process, with >2× or higher levels still present. This could be attributed to active, low-level generation of cytokines in the Ficoll® sample or it could indicate a greater removal of soluble, diffusible species through microfluidic processing. Further work is required to conclusively prove any of the above hypotheses and determine the exact cause of the lower cytokines in the DLD product. Regardless, the DLD-processed cells resulted in lower levels of numerous harmful cytokines, which is particularly relevant for minimizing inflammatory responses and more severe cytokine release syndrome for shortened CAR-T processes with minimal (or no) expansion times.
While manual Ficoll® density gradient centrifugation is common in academic labs, other full-scale instrumentation is typically used for industrial-scale cell therapy manufacturing workflows. Systems such as Sepax™ C-Pro, (Cytiva), Gibco™ CTS™ Rotea™ (Thermo Fisher Scientific), and Lovo® (Fresenius Kabi) provide GMP-compliant, commercially available solutions. Table 1 compares the performance of those systems using the manufacturer's specifications and application notes to the performance of microfluidic DLD processing established here. Microfluidic processing matches or exceeds the performance of these systems when looking at the common comparison metrics, without exposing cells to high centrifugal forces or lysis chemicals. This performance is achieved using the same protocol for all samples processed, and without requiring any process optimization by the user.
| Device name | Sepax™ C-Pro (Cytiva)48–50 | CTS™ Rotea™ (Thermo Fisher)51 | Lovo® (Fresenius Kabi)52,53 | Curate |
|---|---|---|---|---|
| Technology | Centrifugation (rotating syringe) | Counterflow centrifugation (CFC) | Spinning membrane filtration | Microfluidic DLD separation |
| WBC recovery | >80% | >90% | 97.2% (N = 5) | 87 ± 6% |
| Platelet depletion | >80% (with 2 wash cycles) | High (exact value not stated in specifications) | 98.4% (N = 5) | 99 ± 1% |
| RBC depletion | >99% possible but with only >52% WBC recovery49 | Partial; complete removal requires lysis buffer | ∼0% (limited by 4 μm pore size), requires lysis buffer | 91 ± 7% |
| Processing time | 30–120 min | 25–60 min | <7–9 min (per 200 mL) | ∼60 min |
| Challenges | Tradeoff between RBC removal and WBC recovery | -Requires significant process optimization by the user | RBC removal requires lysis buffer | |
| Pellet cells (high centrifugation forces) | -Requires lysis buffer for complete RBC removal |
Small scale experiments have confirmed this performance relative to full-scale systems. In a direct comparison to Rotea™ (using an RBC lysis protocol), microfluidic processing showed overall higher WBC recovery and similar purity.15 Interestingly, after freezing and thawing, the cells prepared by the Curate system had higher yields of PBMCs and CD3 Pan-T cells after 24 and 48 h of culture. The study also confirmed the deleterious impact of ammonium chloride treatment for lysis, with the T cells processed by Rotea™ showing decreased metabolic fitness.
A second study compared microfluidic processing to Sepax™, across 6 apheresis samples.54 Microfluidic processing on average achieved 78.7 ± 2.7% recovery of total nucleated cells, while Sepax only recovered on average 30%, a full 48% lower. When comparing T-cell recovery, the difference was even more significant, with microfluidic processing recovering 77% more CD3+ T cells. Furthermore, the difference in platelet removal between the two techniques was similar to that observed with our manual Ficoll® comparison, with microfluidic processing removing 99.7 ± 0.1% and Sepax™ removing 90.9 ± 0.6%. Similar T cell differentiation post-isolation was observed on both platforms. This study demonstrates that while current commercial systems may report higher performance in the manufacturer's specifications, the actual performance achieved by 3rd party users may be significantly different. Microfluidic processing, on the other hand, achieved the same performance, within error, as demonstrated in our large scale study.
| DLD | Deterministic lateral displacement |
| DCS | Deterministic cell separation |
| PBS | Phosphate buffered saline |
| PLA | Plasmalyte-A |
| LKP | Leukopack, standard apheresis unit |
| WBC | White blood cell |
| RBC | Red blood cell |
| PLT | Platelet |
| BSA | Bovine serum albumin |
| HSA | Human serum albumin |
| ACD-A | Anticoagulant citrate dextrose solution, solution A |
| COP | Cyclo olefin polymer |
Supplementary information (SI) is available. See DOI: https://doi.org/10.1039/d6lc00072j.
| This journal is © The Royal Society of Chemistry 2026 |