Accurate quantification of nanoparticle concentration is important in a host of fields, particularly in nanomedicine, electronics, and catalysis. Microfluidic systems present an opportunity to develop low-cost tests for nanoparticle quantification but often suffer technical challenges related to small sample volumes and optical interference from materials used to construct the device. Here we introduce a microfluidic device that integrates an ultrathin silicon nitride nanoporous membrane (nanomembrane) with an on-chip pressure transducer, designed to precisely quantify nanoparticle concentrations within a microfluidic device using an electrical readout for quantification. As nanoparticles are captured by the membrane under pressure-driven flow, the pressure differential across it changes and is measured by an on-chip transducer. The pressure transducer utilizes a thin PDMS membrane that deflects under pressure to change the cross-section and ionic flow resistance of an adjacent channel. This enables the determination of nanoparticle concentration by analysis of the kinetics of trans-membrane pressure changes relative to particle blockage of the nanomembrane. We also propose a statistical model of partial blockage and particle caking in nanoporous membranes, which accounts for distributions in pore and particle sizes. This model provides a more detailed understanding of nanoparticle filtration behavior and the kinetics of nanopore blocking, enabling accurate concentration determination. Experimental validation of the model on the data acquired by the microfluidic device demonstrates a lower limit of detection on the order of 108 particles per mL, offering a versatile, non-optical approach for the in situ quantification of nanoparticles in a microfluidic device.
At the macroscale, optical methods such as UV-vis spectroscopy,7 turbidimetry,8 and dynamic light scattering (DLS)9 are among the more commonly used techniques to quantify the concentration of nanoparticles in solution. These approaches involve measuring the intensity of light upon absorption or scattering by a population of nanoparticles. However, they rely on knowledge of the extinction/scattering coefficient or employing a reference sample with a known concentration. Flow cytometry is one of the most commonly used techniques for detecting and quantifying particle concentrations, offering high sensitivity, precision, and rapid data acquisition, making it suitable for high-throughput analysis.10 However, flow cytometry has some limitations, including high costs for equipment, maintenance and reagents. Furthermore, the need for skilled operators, and challenges with analyzing data generated from submicron particles or those with low refractive indices also complicate quantification, especially within a microfluidic device where low sample volume, optical interference, and short optical path lengths becomes a significant challenge.11 Additionally, sample preparation requirements, such as labeling with fluorescent dyes, can introduce variability and complexity to the process. Fluorescent labeling is especially ineffective in microfluidic systems as the fluorescence output is determined by the number of particles that have been labeled, which presents a challenge at low concentrations and in small sample volumes. Laser-induced breakdown spectroscopy is another method that relies on measuring the plasma generation from nanoparticles in a suspension.12 It has a wide application for a variety of particles of different sizes, but a continuous wave laser system is required to ensure breakdown of nanoparticles in an aerosolized sample. Resistive-pulsed sensing,13 single particle inductively coupled plasma-mass spectrometry (spICP-MS),14 and light scattering particle counter provide concentration information based on the signal pulses from a sensor and a standard reference sample is usually required for calibration, unless sample volume and nanoparticle detection efficiency can be accurately determined. Lastly, while microscopy techniques enable direct visualization of nanoparticles on a surface, it is critical to ensure unbiased counting as the number of nanoparticles in one image may not represent the overall density of nanoparticles in solution.15
Nanofabricated porous membranes have shown great potential for the capture and size-selective analysis of nanoparticles.16,17 These membranes, characterized by their nanoscale pore sizes, are particularly useful in flux filtration and particle capture, enabling precise control over the separation of various components in a mixture.18 Ultrathin nanoporous silicon nitride (NPN) membranes have particular advantages for membrane-based ultrafiltration of solutions in microfluidic settings due to the tunable, macromolecular scale of both the pore sizes and thickness, the high pore density (>1010 pores per cm2), the chemically inert nature of silicon nitride, and the optical clarity of the membranes.18–20 The exceptional hydraulic permeability of NPN enables microfluidic flows to operate under manageable pressures, overcoming a major limitation of dramatically increased pressure requirements21 when incorporating porous media into microfluidic channels.
To enable the nanomembrane's use as in situ reporters of particle concentrations, deterministic models of pore blockage during pressure-driven particle capture are needed. Membrane fouling produces non-linear flow and pressure responses, as each capture event eliminates a path through the membrane, alters the local flow field, and reduces the overall hydraulic permeability.22,23 Fortunately, these complex mechanics can be interpreted from the well-established fluid mechanics of pore fouling during porous filtration. The membrane science literature teaches that kinetics are influenced by multiple mechanisms, such as pore narrowing, complete pore blockage, and particle caking.24 While traditional models of membrane filtration typically treat the membrane as a uniform permeable medium, NPN has the additional challenge of variations in pore sizes. A mathematical model specifically tailored to account for the unique properties of the microfluidic system is required to provide insight to the mechanisms of nanomembrane fouling as well as produce accurate predictions of sample concentration.
Here, we present a bespoke microfluidic device that integrates an NPN membrane with an on-chip pressure transducer, tailored for electrically quantifying the concentration of nanoparticles in a microfluidic volume. A sample containing nanoparticles is introduced into the microfluidic device using pressure-driven flow. The nanoparticles are captured by the integrated NPN membrane. Upon nanoparticle capture, the fluidic resistance of the NPN increases, provoking predictable changes in pressure within the microchannels. These intra-microchannel pressures are monitored using the custom on-chip pressure transducer. Relying on concepts from the field of membrane filtration science, tracking the kinetics of these pressure changes enables the determination of nanoparticle concentration. We show that the kinetics of nanoparticle capture by the embedded NPN is dominated by two processes: initial nanopore blockage and subsequent nanoparticle caking on the NPN surface. We also developed a computational (or in silico) model that accounts for these two mechanisms to predict the kinetics of the measured pressure changes within the device upon nanoparticle capture, providing an ability to quantify the sample's nanoparticle concentration in the ultra-low 800 fM to 80 pM range of concentrations. This microfluidic platform offers a complementary analytical technology capable of sub-picomolar nanoparticle concentration quantification.
Under the influence of pressure-driven flow and upon the introduction of a nanoparticle-containing sample, the pressure immediately below the NPN membrane varies as a function of the number of particles captured.25 These pressure changes are monitored using an integrated microfluidic pressure transducer. If the system is thought of as an analogous electric circuit,26 the principle used to measure the resistance at the NPN is that of the voltage divider.27 Placing the transducer pre vs. post NPN determines the direction of the measured pressure change as the NPN is blocked. Higher sensitivity will be achieved in whichever section of the channel has higher resistance. A fluidic design with higher resistance was placed post-NPN for simplicity of fabrication and operation compared to placing it pre-NPN. This ensures that the fluidic path pre-NPN could be kept short to avoid dead volume and particle settling while the post-NPN fluidic path could be lengthened to increase hydrodynamic resistance.
The transducer is composed of a thin (10–15 μm) polydimethylsiloxane (PDMS) membrane. When the intra-channel pressure below the NPN changes, the PDMS membrane deflects into a microfluidic transducer channel (TC) filled with an electrolyte solution (0.1 molar potassium chloride). The TC is kept at a constant pressure (6.9 kPa), causing all of the PDMS membrane deflection to be due to changes of pressure within the post-NPN microchannel. This deflection changes the cross-section of the transducer microchannel, which in turn changes the ionic resistance.28,29 Variations in the ionic current are monitored using a pair of electrodes to track the change in trans-membrane pressure during testing. The deflection of the pressure transducer's PDMS membrane can be visualized by incorporating coloured solutions into the TC, as shown in Fig. 2a and b. As pressure increases in the post-NPN microchannel, the red fluid in the TC is displaced by the deflected PDMS transducer membrane, which causes the fluid color to appear lighter.
Within a range of 0–96.5 kPa, the system demonstrated a linear response to pressure differential. The sensitivity of the transducer was determined to be −26 mV kPa−1 (Fig. 2c). The limit of detection (LOD) was calculated to be 0.6 kPa or 0.6% full scale. When the pressure differential between the post-NPN microchannel and TC exceeds 96.5 kPa, the system is expected to experience a non-linear voltage response, as the transducer's PDMS membrane completely collapses the TC.30 Because the Young's modulus of PDMS is thickness dependent for membranes thinner than 200 μm,31 thinner membranes are expected to lead to higher sensitivity to pressure changes, at the cost of a lower dynamic range. However, we avoided using thinner PDMS membranes to prevent micro-tears in the PDMS. The transducer's linear response and high dynamic range makes it suitable for measuring NPN blockage events within the proposed microfluidic device.
During testing, samples spiked with known concentrations of nanoparticles were passed through the microfluidic device. As nanoparticles were captured by the NPN membrane, the electrical signal measured by the transducer was recorded, allowing tracking of the hydrodynamic resistance across the NPN membrane. The first 180 seconds of the experiment consist of an initial water priming phase (using pressure-driven flow) and of an ITCC. At this stage, the initial wetting efficiency of the NPN was determined. We define wetting efficiency as the percentage of pores in the NPN which are fully accessible to sample flow at the start of the experiment. Wetting efficiency was calculated by dividing the expected hydrodynamic resistance of a theoretically ideal and 100% wetted membrane by the measured NPN flow resistance. Determination of the wetting efficiency of the NPN membrane enables its incorporation as an initial parameter within the fitting algorithm, enhancing the accuracy of subsequent particle concentration calculations. Between 180 s and 240 s of the experimental run the nanoparticle sample was loaded into the device. At time t = 240 s, the pre-NPN microchannel was pressurized to 75.8 kPa and the post-NPN microchannel pressure was set to 6.9 kPa to induce sample flow through the NPN, initiating the particle capture process. At set time points throughout the experiment (60 s, 270 s, 360 s, 660 s, 960 s, and 1260 s) ITCCs were performed to ensure reliable pressure measurements. Initial testing was performed using polystyrene nanoparticles with an average particle diameter of 92 ± 9 nm, at a concentration of 4.9 × 109 particles per mL (∼8 pM). These particles were selected to ensure that all particles were captured by the nanoporous membrane. NPN membranes with a preset pore size of 60 ± 20 nm were acquired from SiMPore Inc. The smallest particles commercially available that would have no particles flow through a pore given the variance in both pore diameters and particle sizes were selected for testing. As expected, particle capture increased the NPN resistance leading to a reduction in post-NPN microchannel pressure (Fig. 3). For ease of interpretation, time points associated with ITCC events were removed. Visible discontinuities in the signal are an artifact of the removal of these time points and the recalibration that occurs as a result of the ITCCs.
In an ideal scenario, each nanoparticle would fully restrict flow through a given pore. In the case that the membrane becomes saturated with particles, the measured pressure would reduce to 6.9 kPa, equal to the pressure applied to the post-NPN microchannel. However, under real world assay conditions, we expect that a particle will only partially block a given pore and lead to more complex dynamics. To determine the concentration of particles under these conditions, in silico modeling of the particle capture process was used to provide insight into the device's response, ultimately providing a method for determination of a sample containing an unknown nanoparticle concentration.
, corresponding to the percent reduction in flow rate through a given pore when it is partially blocked by a nanoparticle under constant pressure driven flow conditions. It is important to note that blocking efficiency is different than wetting efficiency. While wetting efficiency provides insight into the total number of pores initially available on the NPN membrane for blocking, blockage efficiency estimates the average reduction in flow each individual pore experiences when it becomes blocked by a nanoparticle of a given size. In the case of incomplete blockage, flow is not fully restricted through a given pore when a pressure gradient is applied across the NPN. Additional particles are still driven towards the NPN membrane, although at a reduced flow rate as a function of the given pore blocking efficiency. This phenomenon causes excess nanoparticles to settle on the surface of the NPN and begin stacking on top of each other in a process termed “particle caking” or “cake filtration” (Fig. 4b). As more particles are deposited on the cake layer, the hydrodynamic resistance through the particle caking layer increases.32,33
We developed a mathematical model to provide accurate simulations of the system's pressure response kinetics as the particles are captured by the NPN. The core assumption of the model is that the flow of particles through the system exactly follows the fluid flow. In the model, nanopores were considered as cylindrical channels of uniform length L, which allows for the expression of an individual pore's unblocked pore resistance Ri using the Hagen–Poiseuille equation:34
![]() | (1) |
As pores become blocked by nanoparticles, the resistance of a given pore increases towards infinity (in the case of perfect pore blockage by a particle). Due to variations in shapes and sizes of both the nanopores and nanoparticles, perfect blockage is rarely achieved and the resistance of a given pore, n, at a given time point t may instead be represented as Rp(n,t).
An NPN can be considered as a network of individual nanopores aligned in parallel with one another. The resistance of the NPN, (RNPN(t)), at a given time, t, can therefore be expressed as:
![]() | (2) |
![]() | (3) |
The cake height will increase over time as particles continue to be deposited, described by the function h(t). Particles will flow preferentially towards less blocked regions of the NPN. This affects the deposition density and layer permeability of the cake, which in turn affects the added resistance to the system. Lower sample flow occurs towards areas with high hydrodynamic resistance. Decreased flow causes a lower number of particles to settle on the NPN, therefore decreasing the deposition density and increasing layer permeability of the cake.36 The permeability can then be expressed as a function, k(z), of the vertical position in the caking layer z. Therefore Rc at any given time can instead be represented as:
![]() | (4) |
Using the mathematical model described above, the effects of blocking efficiency (Fig. 4c) and particle caking (Fig. 4d) on the pressure response of the device over time were simulated. Our model simulated the transducer's measured pressure after the introduction of a 5 × 109 particle per mL (∼8 pM) sample at the NPN membrane. It should be of note that our model considers the physical properties of the system including pore and particle sizes and variability as initial variables for the simulation. These variables are provided by the user and can be modified to fit the unique properties of the system. The model is therefore adaptable to a variety of different membrane and particle parameters. The simulation was considered at various blocking efficiencies over the range of 0–100%. The mixed model of pore blockage and particle caking was utilized to re-simulate the pressure kinetics given the same blocking efficiency values (Fig. 4d). When accounting for only blocking efficiency, a visible difference in pressure plateaus can be observed. Lower observed plateaus are related to increased resistance contributed by more efficiently blocked pores due to the decrease in flow. When a mixed model of pore blockage and caking is considered (Fig. 4d), the difference in plateaus at varying blockage efficiencies is significantly reduced compared to a pore blockage only model (Fig. 4c).
The kinetics of the pressure changes upon particle capture are associated with the sample's particle concentration. When simulating experiments running over 1000 seconds to mirror our experimental time frame, our model predicts the platform's dynamic range (the minimum and maximum detectable particle concentrations). The minimum detectable particle concentration was deemed to be the concentration at which any NPN blockage by particles that may have occurred provides a pressure output indistinguishable from the initial pressure over the given time frame. The maximum detectable particle concentration was deemed to be the concentration at which the complete NPN blockage induced pressure changes that occur so quickly that the kinetics of the system could not be determined due to the limitations caused by the temporal response of PDMS to changes in pressure.37
These minimum and maximum detectable concentrations are influenced by various system parameters, such as physical properties of the NPN membrane and/or set pressures pre and post chip. Moreover, the detection capabilities are also influenced by the dominant blockage mode of the system. A caking-dominant system will offer a larger dynamic range but will be less sensitive at low concentrations (Fig. 4e. i). A pore blockage-dominant system will allow for the detection at lower concentrations, but dynamic range will be reduced (Fig. 4e. ii).
Given the uncertainty of the pressure measurements, our simulated data, for our bespoke device, indicates a detectable minimum and maximum particle concentration between approximately 6 × 107 p mL−1 and 6 × 1011 p mL−1 (100 fM and 1 nM) (Fig. 4e. i) in a caking dominant system, while in a pore blockage dominant system, concentration can be measured between approximately 6 × 106 p mL−1 and 6 × 109 p mL−1 (10 fM and 10 pM) (Fig. 4e. ii). Experimentally, the dominance of pore blockage and/or caking is difficult to predict. Next, we will show that we can use this model to fit experimental data and the kinetics of particle capture to predict a given sample's particle concentration and the platform's dynamic range.
To evaluate the ability of our model to predict particle concentrations accurately and reproducibly, identical replicate experiments were performed four times. Polystyrene nanoparticle (92 ± 9 nm) samples at (4.9 ± 0.1) × 109 particles per mL (8.1 ± 0.2 pM) were introduced to the microfluidic device (inlet pressure: 75.8 kPa, outlet pressure: 6.8 kPa) while the transducer signal was recorded. An initial calibration using an ultrapure water sample was performed and the nanoparticle containing sample was introduced to the system at 240 s. The calibration events (ITCC, as previously described) were included periodically during the test to account for possible signal drift. These ITCC events were also modeled and are visible as sharp discontinuities in the solid line fits (Fig. 5). The time points immediately following the ITCC events were chosen to ensure the highest accuracy of the measured pressure. For each run, the selected points were fed to the analytical model to provide a particle concentration prediction. Nelder–Mead multivariable optimization was used to predict the particle concentration and NPN blocking efficiency that would best fit the pressure kinetics of the experimental results (Fig. 5).
The concentrations predicted by the analytical model for the experimental replicates were 3.6 × 109 particles per mL, 4.6 × 109 particles per mL, 5.4 × 109 particles per mL, and 8.4 × 109 particles per mL respectively. Interestingly, while the same sample concentration was introduced into the device, the capture kinetics are quite different from experiment to experiment. We postulate this is likely due to differences in the NPN wetting efficiency and subsequent particle blocking efficiency, rather than a difference in effective particle concentration. The predictive model was designed to account for differences in initial conditions of the test and still provide a prediction of the sample concentration.
The effects of wetting efficiency on system dynamics are illustrated in Fig. S4. Due to the small pore diameters, the surface tension at the site of the pore is very high, posing difficulty in displacing any air inside the pores and replacing it with liquid. This difficulty in removing excess air results in a decrease in total usable NPN surface and subsequently transmembrane pores. Furthermore, unintended particulate matter in solution, even in ultra-pure water, can block fluidic access to pores decreasing the functional wetting efficiency of the membrane. Although the NPN is treated with piranha solution to increase its hydrophilicity, and an initial wetting of the device is undertaken to maximize the % of pores accessible to flow, a large variance in chip-to-chip wetting efficiencies is still observed. Pending further optimization of membrane wetting, we show that NPN wetting efficiency needs to be accounted for as it affects the initial number of accessible pores and by extension, the initial hydrodynamic resistance of the NPN. This in turn has effects on the pressure response kinetics of the system.
To assess the precision of the analytical model used to predict particle concentration, we employed a bootstrapping approach (S5: error analysis). Briefly, bootstrapping involves random resampling of the experimentally acquired data, fitting a new curve and generating a new concentration prediction for each resample.39 The resampling and refitting process is repeated 15
000 times, and the predicted concentrations were used to determine frequency and distribution of potential outcomes. The uncertainty determined via bootstrapping is representative of the error related to the multivariable fitting algorithm and highlights the variability in the robustness of the fitting algorithm from test to test. Some bootstrapping fits had a low variance in concentration while others were determined to have a wider distribution of predictions. The standard deviation of the bootstraps across experimental runs were 52%, 13%, 10%, and 60% of the mean predicted values. The average uncertainty of the bootstrapping analysis was about ±30% of the mean predicted concentration. The lowest standard deviation values observed were observed during experiments where the NPN membrane wetting efficiency was between 70–75%. Although our model was devised to account for a wide variety of initial variables, further testing is required to determine what role if any wetting efficiency plays in the reproducibility of predicting particle concentration using our multivariable analysis. Our data so far indicates that low wetting efficiency may contribute to higher uncertainty in predictions. We believe this is because of the reduced number of pores available to be blocked. When only a small percentage of pores are available, blocking a single pore contributes to a proportionately larger and more rapid change in pressure. This rapid change in system pressure may exceed the temporal resolution of our system and cause difficulty in producing an accurate and precise prediction. On the other hand, high variability in predicted values at high wetting efficiencies may be due to assumptions made about the physical properties of the membrane. As an example, in several NPN membranes we had tested, we observed the presence of microtears (i.e. defects). These defects in the membrane would permit nanoparticles to bypass capture, resulting in an underestimation of particle concentration. To mitigate this, we have implemented more detailed quality control measures in preparing our NPN membrane devices. Conversely, if pore dimensions exceed expected values, the total pore count may decrease while maintaining a high calculated wetting efficiency. Such an issue could increase the magnitude of a pressure drop as a result of any given pore being blocked and may lead to an overestimation of the sample concentration.
Experiments were also performed at three different input particle concentrations (4.9 × 108 particles per mL, 4.9 × 109 particles per mL, 4.9 × 1010 particles per mL or 800 fM, 8 pM, and 80 pM respectively) spanning three orders of magnitude to highlight the method's ability to predict particle concentration within the system's dynamic range. Our model predicted values of 7.33 × 108 particles per mL, 7.02 × 109 particles per mL, 3.75 × 1010 particles per mL or 1.2 pM, 11.7 pM, and 62.3 pM respectively (Fig. 6). Although it may seem as if the final plateau value is heavily correlated with the particle concentration, the entire time course of the test must be considered to properly determine concentration. While 0.08 pM (5 × 108 particles per mL) samples were tested, the algorithm was unable to accurately infer sample concentration (data not shown), highlighting a difference between the theoretical and experimentally validated predictive range of the system. At high concentrations, the NPN is blocked too quickly and the predictive algorithm accuracy decreased. At low concentrations, the system pressure changes minimally over time, and the algorithm is equally inaccurate. At this time, the model is only capable of accurately predicting concentrations in the 108–1010 particles per mL range. Further refinement of the predictive algorithm and a device redesign may help in reducing sampling uncertainty and extend the dynamic range of the system.
One technique capable of quantifying such low particle concentration is nanoparticle tracking analysis (NTA).40 While our technique is marginally less accurate, it operates in a larger dynamic range without increasing the minimum detectable concentration. Multi-angle dynamic light scattering (MADLS)41 is a commercially available technique capable of quantifying low nanoparticle concentration in solutions. The method we propose here is comparable in both accuracy and dynamic range to that of MADLS. Our proposed technique aims to act as a complementary tool to both NTA and MADLS, providing a secondary, non optical validation method for use within microfluidic systems.
Additional experiments were conducted for concentrations outside of the reported dynamic range. While our system was able to monitor the changes in pressure, the predictive model fitting failed to converge. The physics model and fitting algorithm may be evaluated and adjusted to increase both the accuracy and dynamic range of the predictive model moving forward. Our model currently considers all pores to be perfectly cylindrical and all particles to be perfectly spherical, but in reality, this is not the case. Better characterization of the NPN and the nanoparticles can help improve the simulations and the accuracy of the fit. Due to the disruptive nature of ITCC events on the flow conditions, the number of calibration events have been minimized, as they may contribute to inaccuracies in the predictive analysis. A less disruptive internal calibration cycle would allow for the acquisition of more data points within an experimental run, which in turn would increase the accuracy of the model fitting. Alternatively, investigation into the minimization of signal drift would mitigate requirements for ITCC events. Improving NPN wetting efficiency should additionally enhance pore accessibility, which in turn minimizes initial NPN resistance, ultimately leading to more pronounced pressure variations throughout the test. Matching the density of the suspension solution to the density of the nanoparticles may also reduce predictive error of the system. In a density matched solution, nanoparticles would remain suspended for prolonged periods of time, reducing the likelihood of local changes in sample concentration due to particles settling within the device.
This proof-of-concept device paves the way for designing more efficient filtration systems used in various applications, from biomedical devices to environmental monitoring. Future work will focus on exploring the effects of varying particle types and sizes, to explore the effects of different nanoparticle characteristics and operational conditions on capture efficiency and caking dynamics. By enhancing the understanding of these dynamics, advanced microfluidic platforms capable of in situ assessments of nanoparticle concentrations can be developed.
:
1 base to cross linker ratio, then poured onto the 3D-printed resin molds and degassed for 10 minutes. The molds were then placed in an oven at 70 °C for 4 hours. The cured PDMS was subsequently removed from the molds using a scalpel, and holes were punched in the PDMS using biopsy punches to provide fluidic and electrical access to the microchannels (0.75 mm in diameter for fluidic ports, and 1.25 mm in diameter for electrodes). PDMS mixtures for the transducer layer were degassed, then spun at a thickness of ∼105 μm, and baked in the oven at 70 °C for 40 minutes.The LabVIEW program sets the regulator's output throughout the test, processes the signal using smoothing and low pass filtering to reduce the system's noise, and records the signal for analysis. During testing, a 10 psi pressure differential is applied across the NPN to push the sample through the membrane, and the transducer microchannel is pressurized to 1 psi. The program additionally includes ITCC “calibration events” at set times throughout the test (60 s, 270 s, 360 s, 660 s, 960 s, 1260 s) to ensure accurate measurement of the pressure, by reducing uncertainty related to noise and electrical baseline shifts that may occur during the test. During these calibration events, the entire testing channel was set to 1 psi for 30s to establish a baseline voltage, then the pressure was changed to 2 psi, and the voltage was measured for 30 s to establish a trend between the measured voltage and pressure. The test was run for 30 minutes to allow for maximal saturation of the NPN with the polystyrene nanoparticles.
The membrane characteristics were provided by the manufacturer to be L = 100 nm, A = 0.7 mm2, ϕ = 0.104,
= 58.9 nm, σx = 18 nm, where L is the NPN thickness, A is the NPN surface area, ϕ is NPN porosity,
is the average pore diameter, and σx is the standard deviation of pore diameters. The manufacturer's given values were implemented in the model, along with the nanoparticles' specifications (ȳ = 92 nm, σy = 5 nm) where ȳ is the average nanoparticle diameter and σy is the standard deviation of the nanoparticles' diameter. Given that the nanoparticles sample was prepared in ultrapure water, the physical properties of water at 25 °C (ρ = 997 kg m−3, μ = 8.9 × 10−4 Pa s)51 were included.
The inlet and outlet tubing resistances were determined experimentally using a simple flow rate test, where water was flowed through the tubing at multiple set pressures and collected in a sample vial, and the difference in mass of the collection vial before and after the experiment gave a measure of the flowed fluid volume, which was used to determine the flow rate for the set duration. The resistance values were found to be Rin = 6.6 × 1011 Pa s m−3, Rout = 8.9 × 1012 Pa s m−3.
Fitting of the experimental data was performed using the Nelder–Mead method52 for optimizing multivariable functions. For a given concentration and blocking efficiency multiplier, the simulation was run and then compared to the experimental data to calculate a coefficient of determination, represented as the function R2. The Nelder–Mead method minimizes the function, so to maximize R2, the minimization was performed on the opposite function −R2. When the algorithm converges towards an optimal value, the coefficient of determination should generally be above 0.99. Further details on the modeling parameters are provided in the SI.