Open Access Article
Sajed Mcheik
*,
Eric Da Silva and
Ana Pejović-Milić
Department of Physics, Faculty of Science, Toronto Metropolitan University, 350 Victoria Street, Toronto, ON M5B 2K3, Canada. E-mail: smcheik@torontomu.ca
First published on 11th June 2026
Portable X-ray fluorescence spectrometry (pXRF) has emerged as a promising technique for the in vivo quantification and monitoring of bone lead (Pb) through the detection of characteristic Pb L X-rays. Accurate calibration requires bone and soft-tissue surrogates that replicate photon attenuation and scattering at the measurement site. While significant efforts have focused on bone-mimicking materials, less attention has been given to soft-tissue equivalents. In this study, two soft-tissue mimicking materials—Lucite (PMMA) and paper (98% cellulose)—with thicknesses ranging from 0 to 7 mm were investigated and compared with porcine tissue, which closely approximates human soft tissue. The experimentally determined mass attenuation coefficients for Pb L X-rays at 10.5 keV were 5.22, 4.17, and 4.94 cm2 g−1 for Lucite, paper, and porcine tissue, respectively, indicating that paper provides attenuation properties comparable to those of soft tissue. Calibration curves were generated using six hydroxyapatite (HAp) bone phantoms doped with Pb at varying concentrations and measured for 180 s live time using a Niton XL5 spectrometer (Thermo Fisher Scientific, USA) operating at 40 kVp using an aluminum (Al) filter. At 2.00 mm thickness, the minimum detection limits (MDLs) were 2.2, 1.7, and 2.2 µg Pb per g Ca for Lucite, paper, and porcine tissue, respectively. At 5.00 mm, the MDLs increased to 10.6, 4.3, and 6.2 µg Pb per g Ca. Pb concentrations in three cadaveric tibiae overlaid with porcine tissue were quantified using direct calibration, adjusted coherent normalization, and Compton interpolation. No significant differences were observed among the methods (p > 0.05) using Lucite. Despite paper showing attenuation coefficients closer to porcine tissue, Lucite more accurately reproduced the combined attenuation and scattering behaviour of human soft tissue, supporting its use in in vivo pXRF calibration.
Lead (Pb) is one of the most extensively studied elements in human bone due to its toxicity, widespread environmental exposure, and significant adverse health effects.7 Human exposure to Pb and its associated health risks have been well documented in the literature and by the U.S. Centers for Disease Control and Prevention.8 Exposure to Pb has been linked to cardiovascular, neurological, and renal damage.8 Furthermore, even low-level exposure (<10 µg dL−1 in blood) has been associated with intellectual impairment in children9 and increased mortality risk.10,11 According to biokinetic models, approximately 94% of systemic Pb is stored in the skeleton,12 with half-lives in cortical bone (e.g., tibia) ranging from 7 to 26 years.13 This prolonged retention supports the use of in vivo bone Pb X-ray fluorescence (XRF) as a reliable method for assessing long-term exposure.1,13,14
Accurate calibration of pXRF systems requires that both bone and soft-tissue phantom materials replicate the attenuation and scattering properties of human tissues. For Pb L-line XRF systems, calibration is typically performed using bare bone phantoms, while the effects of overlying soft tissue are accounted for by measuring or estimating tissue thickness and applying attenuation corrections to the detected Pb signal. Bone phantoms used for calibration include plaster of Paris,15 bone meal,16 and hydroxyapatite (HAp).17,18
Lucite (PMMA) is widely used as a soft-tissue phantom material in XRF applications.3,4,15,16 Its composition—primarily carbon, hydrogen, and oxygen—with an effective atomic number (Zeff ≈ 6.6) closely approximates that of human soft tissue (ICRU-44) (Zeff ≈ 7.4),19 resulting in similar X-ray interaction characteristics. Additionally, Lucite is inexpensive, readily available, and easy to machine. However, Lucite has primarily been used to estimate soft-tissue thickness rather than to generate calibration curves across a range of tissue thicknesses. Nie et al.3 proposed estimating human soft-tissue thickness using the dependence of the Compton peak on Lucite thickness, assuming similar scattering behaviour between Lucite and biological tissue. This approach provides an alternative to ultrasound or other imaging modalities for determining overlying tissue thickness.20
Besides Lucite, paper (98% cellulose) has recently been proposed as a soft-tissue surrogate for bone strontium measurements based on K-shell X-ray emissions (Kα = 14.1 keV), which lie within a similar energy range to the Pb L X-rays (Lα = 10.5 keV and Lβ = 12.6 keV) used in pXRF measurements. Cellulose, composed primarily of carbon, hydrogen, and oxygen, has an effective atomic number (Zeff ≈ 6.9), which is comparable to that of Lucite but lower than that of ICRU-44 soft tissue. However, similarity in effective atomic number alone does not fully describe photon attenuation and scattering behaviour in heterogeneous materials.
For measurements based on low-energy X-ray emissions, attenuation by overlying soft tissue must be carefully accounted for. Ultrasound imaging is commonly used to determine soft-tissue thickness at the measurement site.20,21 Magnetic resonance (MR) and computed tomography (CT) imaging have also been explored; however, these techniques are less accessible, more costly, and, in the case of CT, contribute additional radiation dose.20
The minimum detection limit (MDL) is a key metric used to evaluate the performance of XRF measurements and their sensitivity for detecting elements of interest. In in vivo XRF systems, the MDL is typically calculated as follows:22
![]() | (1) |
![]() | (2) |
Although in vivo bone Pb L-XRF has been applied in recent years to monitor Pb concentrations in humans,1,3,13 limited attention has been given to incorporating soft-tissue materials directly into calibration phantoms. This is largely because existing approaches rely on measuring or estimating overlying soft-tissue thickness and applying attenuation corrections to recover the equivalent bare-bone signal. In this work, soft-tissue materials, specifically Lucite and paper, are investigated as human tissue surrogates for bone Pb pXRF measurements. Additionally, their impact on Pb quantification is evaluated using cadaveric tibiae overlaid with porcine tissue.
The linear attenuation coefficient of each soft-tissue mimicking material was determined by fitting the variation of the Pb signal (100 µg Pb per g Ca HAp phantom) as a function of soft-tissue thickness using the Beer–Lambert law. The experimental mass attenuation coefficient was then obtained by dividing the linear attenuation coefficient by the density of the corresponding material.
Bone Pb quantification in the postmortem tibiae was performed using three approaches: (1) direct quantification, (2) adjusted coherent normalization,28 and (3) Compton interpolation.7,15,29
Direct quantification is based on calibration curves obtained from HAp phantoms with varying Pb concentrations, each covered with a fixed thickness of soft-tissue material. This approach requires generating multiple calibration curves for both Lucite and paper across different thicknesses. During in vivo measurements, knowledge of the overlying tissue thickness is required to select the appropriate calibration curve. In this study, cadaveric tibiae were covered with known porcine tissue thicknesses, eliminating the need for thickness estimation.
Adjusted coherent normalization was adapted from Gevaert and Chettle.28 While Gevaert and Chettle applied this method for bone Sr quantification using paper as a soft-tissue surrogate, the present work extends the approach to bone Pb quantification. The procedure used to estimate Pb concentration in cadaveric tibiae overlaid with porcine tissue is summarized as follows:
(1) Plot the Compton peak area as a function of the thickness of paper and Lucite overlying HAp phantoms to determine the equivalent porcine tissue thickness.
(2) Calculate the net peak areas of Pb Lα, Pb Lβ, and coherent (Rayleigh) peaks from tibia measurements.
(3) Generate attenuation curves by measuring HAp phantoms with varying thicknesses of Lucite and paper, plotting Pb Lα, Pb Lβ, and coherent peak areas as functions of thickness.
(4) Use the equivalent porcine thickness (Step 1) and attenuation relationships (Step 3) to estimate unattenuated Pb Lα, Pb Lβ, and coherent peak areas.
(5) Construct calibration curves using bare HAp phantoms (no overlying soft tissue materials) by plotting Pb Lα and Pb Lβ peak areas normalized to the coherent peak area as functions of Pb concentration.
(6) Determine Pb concentration by applying the ratios of unattenuated Pb peaks to the coherent peak (Step 4) to the calibration curves (Step 5).
Compton interpolation7,15,29 assumes that the Compton peak (∼20.8 keV) is independent of Pb concentration. The method is summarized as follows:
(1) Plot Pb Lα and Pb Lβ peak areas as functions of soft-tissue thickness using a 100 µg Pb per g Ca HAp phantom for both Lucite and paper.
(2) Estimate the equivalent porcine tissue thickness using the Compton peak area, as described in Step 1 of the adjusted coherent normalization method.
(3) Determine Pb concentration using the Compton interpolation equation:
![]() | (3) |
Statistical comparisons between the bare bone Pb measurements and the extrapolated Pb concentrations obtained using the three quantification methods were performed using Welch's t-test. Each measurement was based on three independent replicates (n = 3). Statistical analyses were performed at the 95% confidence level, and the results were considered statistically significant when p < 0.05.
| Material | Peak | Experimental (cm2 g−1) | Theoretical (cm2 g−1) | % Diff. | p-Value |
|---|---|---|---|---|---|
| Paper | Lα | 4.17 ± 0.20 | 4.22 | 9% | 0.06 |
| Lβ | 2.54 ± 0.30 | 2.48 | 8% | 0.30 | |
| Lucite | Lα | 5.22 ± 0.20 | 5.90 | 18% | 0.02* |
| Lβ | 3.41 ± 0.30 | 3.46 | 25% | 0.05* | |
| Porcine | Lα | 4.94 ± 0.20 | — | 7% | 0.10 |
| Lβ | 2.53 ± 0.50 | — | 9% | 0.47 | |
| Soft tissue (ICRU-44) | Lα | 4.62 | — | — | — |
| Lβ | 2.78 | — | — | — |
A suitable soft-tissue surrogate should exhibit attenuation and scattering properties comparable to those of human soft tissue. To evaluate the attenuation behaviour of Lucite and paper, their mass attenuation coefficients were experimentally determined and compared with those of porcine tissue and ICRU-44 reference human soft tissue.32
Table 1 summarizes the experimentally determined mass attenuation coefficients for Lucite, paper, and porcine tissue at the energies corresponding to the Pb Lα (10.5 keV) and Pb Lβ (12.6 keV) characteristic peaks. These values were obtained by dividing the linear attenuation coefficients—derived from the variation of Pb L peak areas as a function of material thickness (Fig. 1)—by the corresponding material densities (Table 2). The densities of all materials were calculated from their measured mass and volume.
![]() | ||
| Fig. 1 Pb Lα peak area as a function of soft-tissue thickness (cm) for Lucite, paper, and porcine materials. The uncertainties are smaller than the marker size and are not visible. | ||
| Material | ρ (g cm−3) |
|---|---|
| Paper | 0.82 |
| Lucite | 1.18 |
| Porcine | 1.19 |
When compared with the ICRU-44 reference human soft-tissue density of 1.06 g cm−3,32 the experimentally determined densities showed noticeable differences. This discrepancy is expected, as porcine tissue consists of a heterogeneous mixture of fat and muscle, whereas ICRU-44 soft tissue represents an idealized reference composition. These compositional differences likely account for the observed ∼12% deviation between the theoretical human soft-tissue density and the measured density of porcine tissue.
The experimental mass attenuation coefficients of paper and porcine tissue differ by less than 10% at these energies from the National Institute of Standards and Technology (NIST) ICRU-44 mass attenuation coefficients of human soft-tissue, also included in Table 1. In contrast, Lucite has a significantly higher experimentally determined mass attenuation coefficient than the theoretical ICRU-44 soft-tissue coefficient. Therefore, Lucite attenuated the bone Pb signal more than the porcine tissue, which is also experimentally observed in Fig. 1.
The peak areas of Pb Lα (10.5 keV) and Pb Lβ (12.6 keV) were determined for bare bone HAp phantoms with Pb concentrations ranging from 0 to 100 µg Pb per g Ca. Subsequently, the HAp phantoms were overlaid with Lucite, paper, or porcine tissue to construct calibration phantoms. All calibration curves showed that the Pb peak areas increased proportionally with Pb concentration, exhibiting strong linear relationships, as illustrated in Fig. 2.
To evaluate the sensitivity of the measurements, defined as the slope of the calibration curves (Table 3), it was observed that sensitivity decreased with increasing soft-tissue thickness. Specifically, the sensitivity decreased from 76 counts µg Pb per g Ca for the bare bone phantom to 8 counts µg Pb per g Ca for 4.60 mm of Lucite. Similarly, the sensitivity decreased to 18 and 12 counts µg Pb per g Ca for 5 mm thick paper and porcine tissue, respectively, overlying the HAp phantoms. These results demonstrate a significant reduction in sensitivity with increasing soft-tissue thickness (Fig. 2 and Table 3), which represents a key limitation of bone Pb L-XRF measurements using portable XRF systems. This observation is consistent with previous studies.2,7 Consequently, the reduction in system sensitivity with increasing overlying soft-tissue thickness may limit the applicability of this technique for detecting Pb concentrations near the detection limit.
| Material | Thickness (mm) | Sensitivity [counts/(µg Pb per g Ca)] | MDL [µg Pb per g Ca] |
|---|---|---|---|
| Lucite | |||
| 0.00 | 76 ± 1 | 1.0 | |
| 1.15 | 47 ± 2 | 1.7 | |
| 2.30 | 30 ± 1 | 2.4 | |
| 3.45 | 16 ± 1 | 4.9 | |
| 4.60 | 8 ± 1 | 9.7 | |
| 5.75 | 6 ± 2 | 12.5 | |
![]() |
|||
| Paper | |||
| 0.00 | 76 ± 1 | 1.0 | |
| 1.00 | 69 ± 1 | 1.3 | |
| 2.00 | 49 ± 1 | 1.7 | |
| 3.00 | 37 ± 1 | 2.2 | |
| 4.00 | 26 ± 1 | 3.0 | |
| 5.00 | 18 ± 1 | 4.3 | |
![]() |
|||
| Porcine | |||
| 0.00 | 76 ± 1 | 1.0 | |
| 1.25 | 48 ± 1 | 1.7 | |
| 3.00 | 26 ± 1 | 2.8 | |
| 4.00 | 17 ± 1 | 4.4 | |
| 5.00 | 12 ± 1 | 6.2 | |
In addition to sensitivity, the combined MDL from Pb Lα and Lβ peaks was calculated in the presence of tissue-mimicking materials and porcine tissue using eqn (1) and (2). As expected, and as shown in Table 3, the MDL increased with increasing soft-tissue thickness. To enable comparison at identical thicknesses, MDL values were extrapolated using fitted functions based on the data in Table 3. The extrapolated results (Table 4) indicate that, for a given thickness, the MDL is highest for Lucite, followed by porcine tissue, and lowest for paper.
| Thickness (mm) | MDL (µg Pb per g Ca) | ||
|---|---|---|---|
| Lucite | Paper | Porcine | |
| 0.00 | 1.0 | 1.0 | 1.0 |
| 1.00 | 1.6 | 1.3 | 1.5 |
| 2.00 | 2.2 | 1.7 | 2.2 |
| 3.00 | 3.9 | 2.2 | 2.8 |
| 4.00 | 7.2 | 3.0 | 4.4 |
| 5.00 | 10.6 | 4.3 | 6.2 |
| p-Value vs. porcine | 0.4610 | 0.4351 | — |
Statistical analysis was performed to assess differences in MDL across the tested materials. No statistically significant differences were observed for tissue thicknesses below 5 mm, indicating comparable detection performance among Lucite, paper, and porcine tissue within this range. However, at thicknesses of 5 mm and above, statistically significant differences in MDL were observed, suggesting that additional material-dependent factors influence system performance at greater thicknesses.
It is informative to compare the performance of the present system with earlier L-shell XRF implementations. In 1991, Rosen et al.34 reported a minimum detection limit of approximately 7 µ g Pb g−1 for an overlying soft-tissue thickness of 5 mm using an L-XRF system based on a silver anode X-ray source and a Si(Li) detector, with a measurement time of 16.5 min. In that work, soft-tissue thicknesses in the range of 3–8 mm were explicitly accounted for through attenuation corrections, highlighting the importance of tissue effects in in vivo measurements. In a more recent study by Specht et al.,15,35 an XL3t GOLDD+ portable XRF system (Thermo Fisher Scientific Inc., Billerica, MA, USA) was used with a 3 min measurement time, yielding a detection limit comparable to that of our system.
In comparison, the present study demonstrates substantially higher Pb peak areas under similar concentration ranges, despite a significantly shorter acquisition time of 180 s. This improvement can be attributed to advances in detector technology, particularly the use of silicon drift detectors (SDDs), as well as improved excitation and measurement conditions in modern pXRF systems. However, despite these improvements in signal acquisition, attenuation by overlying soft tissue remains a critical factor influencing both sensitivity and detection limits. The results presented here are therefore consistent with earlier findings, while extending them by systematically evaluating the impact of soft-tissue-equivalent materials on Pb quantification accuracy.
![]() | ||
| Fig. 3 Spectra of cadaver Tibia #1 covered with Lucite, paper and porcine tissues of similar thickness, and a zoom-in between 10 and 13 keV to show Pb Lα (10.5 keV) and Pb Lβ (12.6 keV). | ||
In addition to Pb, other elements present in the tibia, including Ca, Fe, Ni, and Zn, are also identified in Fig. 3.
The MDL values of the bone Pb pXRF measurements suggest that Lucite or paper, as the soft-tissue material, can be added to the bone phantoms to create phantoms for direct calibration. If bone-soft-tissue phantoms are created, human in vivo bone Pb quantification can be achieved once the overlying soft-tissue thickness at the subject's tibia is known by using the direct calibration curve generated with the equivalent soft-tissue materials thickness overlying the HAp bone phantom. This approach, however, would require multiple calibration curves since human tissue thicknesses can have many values, which is not practical. Therefore, other approaches to quantifying Pb in human bone have been proposed in the literature.
In our study, three quantification approaches—direct quantification, adjusted coherent normalization, and Compton interpolation—were applied. Example calculations for determining Pb concentrations in postmortem human Tibia #1 are presented to illustrate each method. The calibration equations of Pb Lβ are listed in Table 5, while the Pb Lβ, Compton and coherent peak areas are given in Table 6.
| Sample | Slope [counts/(µg Pb per g Ca)] | Intercept (counts) | R2 |
|---|---|---|---|
| Bare HAp | 76.0 ± 1.4 | 113 ± 120 | 0.999 |
| Paper 1.25 mm | 50.8 ± 1.4 | 89.8 ± 21.6 | 0.998 |
| Paper 3.00 mm | 32.0 ± 4.0 | 110 ± 310 | 0.977 |
| Paper 5.00 mm | 22.0 ± 4.0 | −270 ± 90 | 0.977 |
| Lucite 1.25 mm | 46.0 ± 4.0 | 0 ± 200 | 0.989 |
| Lucite 3.00 mm | 33.0 ± 9.0 | 200 ± 400 | 0.822 |
| Lucite 5.00 mm | 24.0 ± 4.0 | 180 ± 250 | 0.998 |
| Quantity | Counts |
|---|---|
| Lβ peak area | 3.07 × 103 |
| Lβ uncertainty | 3.80 × 102 |
| Compton peak area | 2.51 × 106 |
| Compton uncertainty | 2.21 × 104 |
| Coherent peak area | 3.79 × 105 |
| Coherent uncertainty | 1.38 × 104 |
The Pb Lβ calibration equation for 1.25 mm paper is given by y = (50.8 ± 1.4)x + (89.8 ± 21.6), where y is the Pb peak area (counts) and x is the Pb concentration (µg Pb per g Ca).
From Table 6, the measured peak area of Tibia #1 covered with 1.25 mm porcine tissue was y = (3.07 × 103 ± 3.80 × 102) counts, which corresponds to a bone Pb concentration of x = (58.7 ± 7.7) µg Pb per g Ca.
Using the direct quantification method based on Pb Lα and Pb Lβ, the Pb concentrations for the three cadaveric tibiae are shown in Table 7, where the reported bone Pb concentration is the inverse weighted average of the Lα and Lβ based estimates.
| Tibia | Soft tissue | Method | Calculated tibia [Pb] | Calculated SD | Bare tibia [Pb] | Bare SD | % Diff vs. bare | p-Value |
|---|---|---|---|---|---|---|---|---|
| 1.25 mm porcine | ||||||||
| #1 | Paper | Direct | 65.1 | 6.9 | 83.6 | 6.0 | −22.1 | 0.025* |
| #1 | Paper | Coherent norm. | 173.0 | 39.9 | 83.6 | 6.0 | +106.9 | 0.018* |
| #1 | Paper | Compton interp. | 186.2 | 34.2 | 83.6 | 6.0 | +122.8 | 0.007* |
| #1 | Lucite | Direct | 74.2 | 8.9 | 83.6 | 6.0 | −11.2 | 0.204 |
| #1 | Lucite | Coherent norm. | 89.7 | 10.5 | 83.6 | 6.0 | +7.3 | 0.432 |
| #1 | Lucite | Compton interp. | 79.7 | 7.7 | 83.6 | 6.0 | −4.7 | 0.527 |
| #2 | Paper | Direct | 55.8 | 2.3 | 65.2 | 5.3 | −14.4 | 0.048* |
| #2 | Paper | Coherent norm. | 127.9 | 24.1 | 65.2 | 5.3 | +96.2 | 0.012* |
| #2 | Paper | Compton interp. | 145.4 | 24.7 | 65.2 | 5.3 | +123.1 | 0.005* |
| #2 | Lucite | Direct | 65.8 | 7.3 | 65.2 | 5.3 | +0.9 | 0.914 |
| #2 | Lucite | Coherent norm. | 64.5 | 7.0 | 65.2 | 5.3 | −1.1 | 0.897 |
| #2 | Lucite | Compton interp. | 68.0 | 7.5 | 65.2 | 5.3 | +4.3 | 0.625 |
| #3 | Paper | Direct | 10.5 | 1.1 | 14.9 | 1.4 | −29.5 | 0.013* |
| #3 | Paper | Coherent norm. | 30.0 | 2.5 | 14.9 | 1.4 | +101.3 | 0.001* |
| #3 | Paper | Compton interp. | 33.5 | 3.0 | 14.9 | 1.4 | +124.8 | 0.001* |
| #3 | Lucite | Direct | 11.6 | 1.2 | 14.9 | 1.4 | −22.1 | 0.036* |
| #3 | Lucite | Coherent norm. | 30.0 | 7.5 | 14.9 | 1.4 | +101.3 | 0.027* |
| #3 | Lucite | Compton interp. | 33.5 | 8.0 | 14.9 | 1.4 | +124.8 | 0.017* |
![]() |
||||||||
| 3.00 mm porcine | ||||||||
| #1 | Paper | Direct | 62.7 | 6.2 | 83.6 | 6.0 | −25.0 | 0.014* |
| #1 | Paper | Coherent norm. | 127.6 | 22.8 | 83.6 | 6.0 | +52.6 | 0.032* |
| #1 | Paper | Compton interp. | 152.8 | 30.5 | 83.6 | 6.0 | +82.8 | 0.018* |
| #1 | Lucite | Direct | 63.6 | 22.2 | 83.6 | 6.0 | −23.9 | 0.206 |
| #1 | Lucite | Coherent norm. | 76.5 | 10.1 | 83.6 | 6.0 | −8.5 | 0.354 |
| #1 | Lucite | Compton interp. | 70.6 | 9.7 | 83.6 | 6.0 | −15.6 | 0.120 |
| #2 | Paper | Direct | 51.6 | 2.2 | 65.2 | 5.3 | −20.9 | 0.015* |
| #2 | Paper | Coherent norm. | 123.5 | 34.2 | 65.2 | 5.3 | +89.4 | 0.043* |
| #2 | Paper | Compton interp. | 140.5 | 46.2 | 65.2 | 5.3 | +115.6 | 0.049* |
| #2 | Lucite | Direct | 58.0 | 7.8 | 65.2 | 5.3 | −11.0 | 0.257 |
| #2 | Lucite | Coherent norm. | 56.0 | 18.1 | 65.2 | 5.3 | −14.1 | 0.446 |
| #2 | Lucite | Compton interp. | 64.9 | 19.1 | 65.2 | 5.3 | −0.5 | 0.980 |
| #3 | Paper | Direct | 11.9 | 1.2 | 14.9 | 1.4 | −20.1 | 0.048* |
| #3 | Paper | Coherent norm. | 36.6 | 12.5 | 14.9 | 1.4 | +145.6 | 0.040* |
| #3 | Paper | Compton interp. | 55.2 | 23.1 | 14.9 | 1.4 | +270.5 | 0.039* |
| #3 | Lucite | Direct | 11.5 | 1.3 | 14.9 | 1.4 | −22.8 | 0.037* |
| #3 | Lucite | Coherent norm. | 19.5 | 2.4 | 14.9 | 1.4 | +30.9 | 0.046* |
| #3 | Lucite | Compton interp. | 25.5 | 3.4 | 14.9 | 1.4 | +71.1 | 0.008* |
![]() |
||||||||
| 5.00 mm porcine | ||||||||
| #1 | Lucite | Direct | 87.8 | 41.7 | 83.6 | 6.0 | +5.0 | 0.184 |
| #1 | Lucite | Coherent norm. | 88.7 | 62.7 | 83.6 | 6.0 | +6.1 | 0.175 |
| #1 | Lucite | Compton interp. | 85.9 | 61.5 | 83.6 | 6.0 | +2.8 | 0.175 |
| #2 | Lucite | Direct | 70.2 | 16.2 | 65.2 | 5.3 | +7.7 | 0.330 |
| #2 | Lucite | Coherent norm. | 62.6 | 25.1 | 65.2 | 5.3 | −4.0 | 0.252 |
| #2 | Lucite | Compton interp. | 60.4 | 24.4 | 65.2 | 5.3 | −7.4 | 0.256 |
| #3 | Lucite | Direct | 37.7 | 2.5 | 14.9 | 1.4 | +153.0 | <0.001* |
| #3 | Lucite | Coherent norm. | 23.0 | 2.7 | 14.9 | 1.4 | +54.4 | 0.009* |
| #3 | Lucite | Compton interp. | 29.3 | 3.6 | 14.9 | 1.4 | +96.6 | 0.002* |
![]() | ||
| Fig. 6 Ratio of Pb Lβ to coherent peak area as a function of Pb concentration (µg Pb per g Ca) for bare bone phantoms. The solid line represents the calibration curve for the HAp bone phantom. | ||
Using the procedure described in Section 2.4, the Pb concentration in Tibia #1 was determined as follows:
(1) Using the Compton peak area as a function of paper and Lucite thickness (Fig. 4 and Table 6), the 1.25 mm porcine tissue with a Compton peak area of (2.51 × 106 ± 2.21 × 104) is equivalent to (1.65 ± 0.08) mm of Lucite and (5.69 ± 1.08) mm of paper.
(2) The Pb Lβ and Compton peak areas for Tibia #1 covered with 1.25 mm porcine tissue are listed in Table 6.
(3) The Pb Lβ peak area for 100 µg Pb per g Ca as a function of Lucite and paper thickness (Fig. 7) is given by:
| Lβ,Lucite = (7.68 × 103 ± 4.1 × 101) e−(0.406±0.003)x Lβ,paper = (7.16 × 103 ± 4.5 × 101) e−(0.208±0.006)x |
Similarly, the variation of the coherent peak area with material thickness (Fig. 5) is expressed as:
| Icoh,Lucite = (3.77 × 105) e0.0821x Icoh,paper = (3.27 × 105) e−0.0109x |
Using the equivalent thicknesses of Lucite and paper determined in step 1, the Pb Lβ and coherent peak areas corrected to the bare bone (using the Beer–Lambert attenuation law) are:
| Lβ,Lucite,bare = (4.88 ± 0.62) × 103 counts |
| Lβ,paper, bare = (1.00 ± 0.25) × 104 counts |
| Icoh,Lucite,bare = (3.47 ± 0.13) × 105 counts |
| Icoh,paper,bare = (3.57 ± 0.08) × 105 counts |
(4) From Fig. 6, the Pb Lβ-to-coherent peak ratio as a function of Pb concentration was used to establish the calibration equation for bare HAp:
Using this calibration, the estimated bone Pb concentrations for Tibia #1 were (75.1 ± 12.6) µg g−1 and (156.0 ± 44.0) µg g−1 for Lucite and paper, respectively. The results obtained using the adjusted coherent normalization method for all tibiae are summarized in Table 7. For each measurement condition, the reported Pb concentration corresponds to the inverse-variance weighted average of the independent estimates derived from the Pb Lα and Pb Lβ peaks.
(1) Using the Compton peak area measured for Tibia #1, a porcine soft-tissue thickness of 1.25 mm was determined to be equivalent to (1.65 ± 0.08) mm of Lucite and (5.69 ± 1.08) mm of paper, as obtained in Step 1 of the adjusted coherent normalization method.
(2) Based on the fitted relationships describing the variation of the Pb Lβ peak area as a function of Lucite and paper thickness (Fig. 7), the Pb Lβ peak areas corresponding to these equivalent Lucite and paper thicknesses were extrapolated to be (4268 ± 98) counts and (1955 ± 488) counts, respectively.
(3) Using eqn (3), the resulting Pb concentration for Tibia #1 was calculated to be (71.9 ± 9.1) µg Pb per g Ca when Lucite was used as the soft-tissue material and (157.0 ± 43.8) µg Pb per g Ca when paper was used.
The results of bone Pb quantification for the three human tibiae are summarized in Table 7. The reported concentrations correspond to the weighted average of the independent Pb estimates derived from the Pb Lα and Pb Lβ peaks. The table includes the Pb concentrations determined with the three quantification approaches and their associated p-values when compared to the bare tibia Pb concentration. For all three tibiae, the bone concentrations determined with different thicknesses of Lucite were statistically identical to the corresponding bare tibia Pb concentration for each quantification approach (p > 0.05), indicating that Lucite is a suitable soft-tissue phantom. In contrast, when paper was used as the soft-tissue material, the calculated Pb concentrations were consistently and significantly different from the corresponding bare tibia values at all tested thicknesses and for all three quantification methods (p < 0.05). Paper-based measurements were not available for the 5 mm porcine tissue, as the required paper thickness exceeded the practical thickness limit (greater than 6 mm), rendering reliable extrapolation of Pb concentration infeasible.
Furthermore, comparison of the direct, adjusted coherent normalization, and Compton interpolation methods across different tissue materials thicknesses for the higher-Pb tibiae (Tibia #1 and Tibia #2) using Lucite showed that the percent difference between the extrapolated Pb concentrations and the corresponding bare tibia values decreased with decreasing porcine thickness. This trend is attributed to reduced attenuation of the Pb signal at lower thicknesses, and thus no statistically significant differences were observed (p > 0.05). In contrast, for Tibia #3, which exhibited lower Pb concentrations, all three quantification methods yielded Pb values that were statistically different from the bare Tibia #3 Pb concentration. This finding indicates a limitation in the ability of these methods to accurately extrapolate Pb concentrations under the low Pb signal conditions, particularly when the bone Pb concentration approaches the minimum detection limit (MDL).
In contrast, analysis of Tibia #1 and Tibia #2 covered with varying Lucite thicknesses showed that all three quantification methods were able to estimate the tibial Pb concentrations with differences of less than 25% relative to the corresponding bare measurements. The largest percent differences were observed for Tibia #3, particularly when covered with a 5 mm porcine-equivalent thickness. These results indicate that the ability to quantify tibial Pb depends on both the overlying soft-tissue thickness and the present Pb concentration.
The results of this study indicate that, beyond elemental composition, material density is a key factor in determining the suitability of soft-tissue substitutes for in vivo bone Pb L-shell XRF measurements. Although the mass attenuation coefficient (µ/ρ) characterizes photon interactions on a per-mass basis, the experimentally relevant quantity is the linear attenuation coefficient, µ = ρ(µ/ρ), which depends directly on material density. The substantially lower density of paper (0.82 g cm−3) compared to Lucite (1.18 g cm−3) and porcine tissue (1.19 g cm−3) results in reduced attenuation and scattering for a given material thickness. In contrast, the similar densities of Lucite and porcine tissue produce more comparable photon interaction conditions. These findings suggest that the improved performance of Lucite arises predominately from its similarity in bulk density to biological soft tissue, in addition to its compositional characteristics.
The large discrepancies observed between quantification approaches when using paper as a soft-tissue substitute can likely be attributed to the combined effects of density mismatch and altered scattering conditions. Both the adjusted coherent normalization and Compton interpolation approaches assume a consistent relationship between scatter signal intensity and overlying mass thickness. These methods utilize the variation in Compton peak area with material thickness to estimate equivalent soft-tissue thickness, following the approach proposed by Nie et al.3 However, due to the lower density of paper and the resulting differences in photon scattering behaviour, this assumption is not maintained, leading to a systematic overestimation of the effective tissue thickness relative to Lucite and porcine tissue.
Furthermore, the results obtained for Tibia #3, which exhibited the lowest Pb concentration, indicate that the technique becomes increasingly sensitive to systematic errors at low Pb signal levels. The statistically significant differences observed across all quantification methods and tissue thicknesses suggest that small inaccuracies in attenuation correction, scatter normalization, and counting statistics have a proportionally larger impact near the detection limit. This finding highlights an important limitation of the Pb L-shell pXRF method, namely reduced reliability for quantifying low bone Pb concentrations, which is particularly relevant for in vivo applications involving populations with low lead exposure.
Overall, the results indicate that, under the tested experimental conditions, Lucite provides a more appropriate soft-tissue surrogate than paper. In addition, any of the three quantification methods can be used to estimate tibial Pb concentrations, provided that the limitations imposed by the limit of quantification are taken into account. These results are consistent with those reported by Specht et al., who used Lucite as a soft-tissue surrogate for calibrating pXRF systems in in vivo studies.7,15 In contrast, although paper has been suggested as a suitable soft-tissue surrogate for Sr measurements,28 our findings demonstrate that the use of paper results in statistically significant differences between the measured tibial Pb concentrations and the corresponding extrapolated values.
Although all measurements were performed using the same detection system, cadaver bones, and phantoms, and efforts were made to maintain consistent positioning and contact geometry, geometric effects may still have contributed to the observed differences and represent an inherent limitation of experimental pXRF measurements. Finally, although this study used a new and more powerful portable XRF spectrometer, the observed attenuation and scattering effects are not specific to portability of the XRF device. They arise from Pb L-shell X-ray energies, soft-tissue attenuation, detector sensitivity, and measurement geometry, and are, therefore, applicable to XRF-based bone Pb quantification based on the L-shell X-rays more broadly.
| This journal is © The Royal Society of Chemistry 2026 |