Sodium is a preferred coolant in Generation IV nuclear reactors due to its excellent heat transfer and nuclear properties. However, decreased cooling efficiency, material corrosion and blockages can be caused by excessive amounts of impurities in the sodium. Conventional detection methods are unable to realize rapid, in situ, sensitive and accurate online sodium impurity monitoring owing to factors such as equipment size, service life and detection time. In this study, a laser-induced breakdown spectroscopy device with a gas protection component was proposed. This design not only prevents sodium aerosols or liquid sodium splashing from contaminating the optical lens, but also enables sensitive analysis of Fe and O in sodium, with limits of detection of 0.68 µg g−1 and 0.11 µg g−1 and RSDs of 7% and 4%, respectively. The results demonstrated that this method has potential application prospects in the nuclear industry and could provide a guarantee for the safe operation of sodium-cooled fast reactors.
Liquid sodium is ideal as a fast reactor coolant due to its excellent thermophysical properties, including a low neutron absorption cross-section, weak moderating ability, high thermal conductivity and a moderate coefficient of thermal expansion. However, there are significant technical challenges posed by the reactive chemistry of sodium, including corrosion and deposit clogging. Corrosion behavior is mainly influenced by the concentration of O impurities. The corrosion rate of the core shell and structural material and induced material embrittlement are accelerated by higher O concentration, which leads to degradation of the mechanical properties, even possibly leading to clogging of the coolant flow paths or failure of critical components.3–5 In addition, deposits could be formed by excessive Fe impurities in the sodium, which impede coolant flow. Additionally, dissolved alloying elements may migrate in the circuit and undergo neutron activation, increasing the risk of radioactive contamination.6,7 It is worth noting that fission products detected in the coolant are often a sign of a possible breakage of a fuel element.
In fast reactors, the detection methods of O and Fe impurities in liquid sodium are mainly electrochemical methods,8 vacuum distillation9 and vacuum distillation-atomic absorption spectroscopy.7 These methods are mature and have potential for field application.8 However, the electrochemical method still suffers from several critical limitations, such as a narrow applicable temperature range, insufficient selectivity, limited sensor lifetime and potential measurement interference. Vacuum distillation separates and enriches impurities to improve the limit of detection (LOD). Chun Xie summarized that this method has a minimum detection amount of 10 µg of O impurities with a relative standard deviation (RSD) of 9.0% when the O concentration in sodium is 10–100 µg g−1.10 Nevertheless, the contamination may be caused by the high-temperature distillation, and a long time is needed for sample processing, making it difficult to realize continuous online monitoring. Vacuum distillation-atomic absorption spectroscopy is mainly used for the detection of iron impurities. This method can accurately determine specific metal impurities, but the equipment is complex, and it can be applied to only some elements. Moreover, vacuum distillation-atomic absorption spectroscopy cannot be used for real-time monitoring. Therefore, the development of an easy maintenance, low-cost, fast and in situ monitoring device for Fe and O impurities in high-temperature liquid sodium is essential for the safe and efficient operation of a sodium-cooled fast reactor.
Laser-induced breakdown spectroscopy (LIBS) has been widely used in the past two decades because of its excellent applicability and environmental adaptability.11,12 It has demonstrated unique advantages in the field of nuclear energy.13,14 As a rapid multi-element analysis tool, LIBS has been successfully applied in various disciplines such as medical diagnosis, precision agriculture and geological exploration.15–17 The core principle of LIBS is to use high-energy laser pulses focused on the sample surface to form a transient plasma on a trace sample through inverse bremsstrahlung and multiphoton ionization effects. Characteristic radiation, allowing the observation of continuum spectra, ion spectral lines, atomic spectral lines and molecular bands, is released during plasma cooling. These optical signals are collected and analyzed by high-resolution spectrometers and intensified charge-coupled devices (ICCDs). In the nuclear industry, LIBS has successfully realized the detection of lead, indium and other metal impurities in liquid sodium, with a detection limit of 5 ppm.18 Hence, LIBS technology combines broad-spectrum analytical capability, elemental specificity, and in situ detection, making it ideal for real-time monitoring of the composition of nuclear reactor materials, especially in critical scenarios such as liquid metal coolant impurity monitoring.
In this work, a LIBS device with a gas protection component (GP-LIBS) was proposed. The GP component is effective in preventing contamination of the optical lens caused by sodium aerosols generated by high-temperature liquid sodium or by splashing of liquid sodium due to laser kinetic energy acting on the liquid sodium. Moreover, the sensitive detection and quantitative analysis of Fe and O impurities in high-temperature liquid sodium coolant were realized. The device provides a new monitoring strategy and equipment for the impurity elements in the sodium coolant in sodium-cooled fast reactors, which has potential application in the nuclear industry and provides a guarantee for the safe operation of nuclear power plants.
Fig. 1(b) demonstrates the structures of the sample chamber and the GP component. The top of the GP component is a circle to facilitate the penetration of the laser beam and the emitted light. In addition, the cylinder at the top was narrowed inward by 1 mm to form a gas duct at the point where the gas inlet was connected to the GP component when it was placed in the sample chamber. Argon gas entering the sample chamber through the gas inlet would move upward through this duct, due to the GP component being spaced a short distance from the top of the sample chamber, and it can effectively form a gas knife to block upwardly dispersing sodium aerosols formed by high-temperature liquid sodium or block sodium splattering caused by the kinetic energy of the laser beam by a transverse force. The bottom of the GP component is a semicircle and a semi-ellipse to minimize the bottom area in case the laser beam and the emitted light can pass through completely. The gas flow moved downward along the inner opening of the GP, creating a downward pressure that changes the direction of the diffusion of the sodium aerosol or sodium splash, causing it to be deposited on the inner sidewall of the GP. The bottom structure and the top duct of the GP form two protective mechanisms that effectively prevent sodium aerosol or liquid sodium splashing from reaching the optics, thereby improving the accuracy and sensitivity of detection.
The variation of the signal intensity and background value with gas flow rate is demonstrated in Fig. 3(a). The signal intensity and background value exhibited the same trends with the gas flow rate. When the gas flow rate was lower than 0.4 L min−1, the signal intensity and background value were weak. This is because aerosols may penetrate the gas curtain and attach to the lens when the kinetic energy of the gas stream is insufficient, because of the low gas flow rate to effectively deflect upward particles (especially high momentum particles).
Consequently, this led to a decrease in the signal intensity of the Fe impurity. When the gas flow rate was higher than 0.4 L min−1, the signal intensity and background value showed a gradual decrease. This is because when the gas flow rate increased, the refractive index of the optical path was not homogeneous due to the changes in the density of the gas flow, which interfered with the accuracy of the laser measurement, thus leading to a decrease in signal intensity. In addition, the faster velocity gas flow rate would blow the liquid sodium, making the surface of the liquid sodium deform, which caused the laser focus to be offset in the longitudinal direction. This resulted in only a portion of the laser energy being applied to the sample, reducing both the signal intensity and the background intensity. As shown in Fig. 3(a), the signal intensity reached a maximum when the gas flow rate was 0.4 L min−1. Although the background value was increased at this flow rate, affecting the quantitative analysis, the sensitivity was most important for the detection of impurities in the liquid sodium.
Meanwhile, the laser may act on the oxide on the surface of the liquid sodium. Simultaneously, the optimal flow rate was evaluated based on signal stabilities of Fe(II) (259.94 nm) and Na(II) (588.95 nm), with results shown in Fig. 3(b) and (c). The signal stabilities remained remarkably consistent across different flow rates, and the RSDs were all below 8%. It can be seen that the signal variations were not significant at different flow rates. This is because, despite varying gas flow rates, the plasma region remains consistently saturated with the effect of the GP component. Different gas flow rates only affect the signal intensity. In summary, 0.4 L min−1 was selected as the optimal gas flow rate.
The noise could be suppressed by adjusting the number of averaging times, which can reduce the limit of detection (LOD). The signal-to-noise ratio (SNR) as a function of the number of averaging times of the spectrometer is illustrated in Fig. 4(b), and the SNR was decreased when the number of averaging times was more than 10. This indicated that the random noise in this system was mainly suppressed by the spectrometer when the number of averaging times was lower than 10. As the number of averaging times increased, there was a noticeable decrease in the signal intensity of Fe(II) (259.94 nm). This may be attributed to the increased sampling number resulting from the number of averaging times, which caused deformation of the liquid sodium due to the number of the laser irradiations. As a consequence, some invalid spectra were captured because of the focal point shifts, which reduced the signal intensity by averaging, thus reducing the SNR. As indicated in Fig. S2, it can be observed that the SNR exhibits a gradual upward trend after removing invalid spectra, while the signal intensity continues to decrease. Considering that increasing the number of averaging times requires higher sampling numbers and longer sampling times, and that the SNR growth rate slows significantly above 10 averaging times despite removing invalid spectra, the number of averaging times was set to 10 for the subsequent experiment.
The would be generated by the laser-induced plasma in the early stage, which affected the background signal and thus the accuracy of the detection. The bremsstrahlung was mainly suppressed by the delay time, so the SBR was used as the evaluation index of this parameter. At shorter delay times, the higher intensity was mainly due to the presence of the background spectrum. While the signal intensity decreased with increasing delay times, the SBR increased and then decreased. This implied that when the delay times were large, not only was the background spectrum eliminated, but also a part of the signal value was filtered out. Consequently, the delay time was set to 3 µs.
The signal enhancement of the ICCD in the spectrometer was controlled by the gate width. As the gate width increased, both the SBR and signal intensity increased. Similar to the exposure time, as the gate width exceeds 1000 ns, changes in the SBR value become gradual while the signal intensity continues to increase. A gate width of 1000 ns was used for the subsequent analysis.
The laser frequency also impacted the signal emission intensity. As demonstrated in Fig. 4(e), when the laser frequency was low, a lower signal intensity resulted, leading to a reduction in the SNR values. This indicated that the signal intensity obtained at that frequency is not reliable. As the laser frequency increased, the thermal effect of the laser on the sample surface enhanced the signal intensity to some extent, thereby increasing the value of the SNR. When the laser frequency was higher than 4 Hz, the kinetic energy of the laser acting on the surface of the liquid sodium was too large, resulting in the GP component being unable to prevent the liquid sodium from splashing, thus contaminating the optical lens. In addition, with a higher laser frequency, a deformation of the sodium liquid surface occurred, and before the deformation could be reversed, the laser acted again on the sample surface, thus reducing the signal intensity. Therefore, the laser frequency was set to 4 Hz.
Additionally, to verify that GP effectively reduces lens contamination, measurements were conducted for 10 min on the same sample under both non-GP and GP conditions. The results are shown in Fig. 5(b) and (c). It can be observed that there were distinct white spots that appeared on the flange viewing window without the GP component. These spots result from liquid sodium splattering caused by laser interaction. In contrast, when the GP was present, no significant contamination was visible on the flange viewing window. Therefore, the GP component not only does not interfere with the transmission of emitted and collected light but also effectively blocks contamination caused by liquid sodium splatter.
, where σ denotes the spectral noise (standard deviation of the background) and k is the slope of the calibration curve. The results of the calibration curve of Fe and the signal intensity are shown in Fig. 6. It is worth mentioning that the LOD of the Fe impurity in liquid sodium was calculated to be 0.68 mg L−1, and all the RSDs of the signal intensity were lower than 7%.
![]() | ||
| Fig. 6 Calibration curve for Fe(ii) (left) and the corresponding spectra showing signal intensity at various concentrations (right). | ||
To exclude interference from the presence of oxygen in the detection environment, a blank sample, and samples with high and relatively low concentrations of O were analyzed. As demonstrated in Fig. 7, the emission peak for O(I) (777.14 nm) was absent for the blank sample. For liquid sodium samples with high and low concentrations of O impurities, there was a significant difference in the signal intensity for O(I) (777.14 nm) and the signal intensity showed the same trend as the concentration. Furthermore, the calibration curve for O and the signal intensity are shown in Fig. 8. It is worth noting that the LOD of the O impurity in liquid sodium was calculated to be 0.11 mg L−1, and all the RSDs of the signal intensity were lower than 4%. The results indicated that the GP-LIBS realized sensitive, stable, real-time monitoring of O and Fe impurities in liquid sodium.
![]() | ||
| Fig. 7 The emission peaks of blank liquid sodium, and high and low concentrations of the O impurity in liquid sodium. | ||
![]() | ||
| Fig. 8 Calibration curve for O(i) (left) and the corresponding spectra showing signal intensity at various concentrations (right). | ||