Open Access Article
Massimo Onor
,
Beatrice Campanella
*,
Emanuela Pitzalis
and
Alessandro D'Ulivo
Institute of Chemistry of Organometallic Compounds, National Research Council, via Giuseppe Moruzzi 1, 56124, Pisa, Italy. E-mail: beatrice.campanella@cnr.it
First published on 11th December 2025
The aim of this investigation was to increase the tolerance limit of plumbane generation towards several transition metals by exploiting the hydrolysis products of borohydride – the hydridoboron intermediates (BH) – as a derivatization reagent instead of borohydride, alone or together with chemical masking agents (oxalic acid and thiourea). The BH intermediates were produced online at 0.1 mol L−1 HCl sample acidity in the presence of K3[Fe(CN)6] (0.1 or 1% m/v) and by varying the hydrolysis coil volume (0–2000 µL) before mixing with the analyte. The simple use of BH intermediates alone obtained with 1500–2000 µL hydrolysis coil volume allowed the achievement of tolerance limits of 100 mg L−1 for Ni(II), Co(II), and Fe(III) and 105 mg L−1 for a mixture of Cr(III)/V(IV)/Mo(VI) (75 + 15 + 15 mg L−1). The tolerance to the strong interfering action of Cu(II) was considerably improved but limited to 2 mg L−1. To mitigate the interference caused by 1000 mg L−1 Fe(III), BH intermediates were used in combination with thiourea and oxalic acid. However, this approach did not enhance the tolerance to Cu(II) (2 mg L−1), even in the presence of the masking agents. The established conditions allowed for the accurate determination of Pb in the complex matrix of SRM 663 (Cr–V steel) after sample dissolution with inorganic acids at high temperature.
Within the family of hydride-forming elements, lead represents a singular case, as CVG of plumbane cannot be achieved without suitable additives – the most popular and effective being potassium hexacyanoferrate (III), K3[Fe(CN)6] (FeC), which ensures generation yields >95%.24 CVG of plumbane is affected by severe interferences, particularly those arising from transition metals. These are difficult to control because most masking agents which are effective for other hydride-forming elements fail or are much less effective in plumbane generation.25 This may be due to the more complex formation mechanism of plumbane which, as recently demonstrated, involves the formation of intermediate non-volatile hydrides.26 Atomization interferences in CVG of plumbane are not a problem in diffusion flame atomizers, such as the one used in the present study, as recently reported by Vlčková et al.27
Recent papers have reported detailed studies on the control of interferences in CVG of plumbane in acidic FeC reaction media. Deng et al.,28 using CVG-ICPOES and FeC in HCl media in the presence of hexacyanoferrate (II), K4[Fe(CN)6], observed no signal depression for an interferent concentration of 0.4 mg L−1, except for Cu, whose tolerance limit was below 0.4 mg L−1. Beltrán et al.29 using FI-CVG-AFS and FeC in HCl media, investigated the effect of KI, L-cysteine and 1,10-phenantroline on the interference caused by 0.1–1.0 mg L−1 of As, Cu, Cr and Fe. The best control was achieved by KI, which improved the tolerance limit up to 0.3, 0.6 and 0.3 mg L−1 for As, Cu and Cr, respectively. In the absence of masking agents, Fe did not interfere up to 1.0 mg L−1. Schlotthauer et al.,30 using CVG-AAS in nitric acid plus borate buffer media and FeC, developed a CVG method for lead determination in natural water samples. They investigated the interference of hydride/cold vapor forming elements (As, Cd, Hg, Sb, Sn and Se) in the range from 0.005 to 0.05 mg L−1, of transition metals (Ag, Co, Cr, Cu, Mo and Ni) in the range from 0.1 to 1.0 mg L−1, and of Ba, Al, Fe and Mn at 10 mg L−1. The most critical elements were Cu, Ag, Sb and Sn, for which tolerance limits (±10% S/S0) were around 0.1, 0.005, 0.05 and 0.02 mg L−1, respectively. Among the tested masking agents – KI, L-cysteine, ascorbic acid and KSCN – only KSCN was effective, improving tolerance limits, particularly for Cu, up to 2.0 mg L−1. The methods developed by Beltran et al.29 and Schlotthauer et al.30 were applied to natural water samples. The interference effect caused by high Co concentration was reported by Zheng et al.31 using CVG-AFS: Co(II) concentrations of 10 and 100 mg L−1 caused 10% and 100% signal depression, respectively. The use of a mixture of KSCN, oxalic acid and 1,10-phenantroline dramatically improved the tolerance up to 1.4 g L−1 Co. Chuachuad and Tyson,32 using THB immobilized on an anion-exchange resin and CVG-AAS, investigated the interference of several elements in the 0–100 mg L−1 range. For the transition metals tested, the approximate tolerance limits (mg L−1) were reported as follows: Cr(VI) 100; V(V) < 10; Cu(II) and Fe(III) 1; Co(II), Ni(II), Zn(II) and Mn(II) < 1; Ag(I) < 0.1.
Recently, we reported the possibility of controlling interferences in CVG of stibane and bismuthane using the hydrolysis products of THB – without the use of any masking agents – by means of a simple reactor that allows the online generation of hydridoboron intermediates (BH).33 It therefore seemed worthwhile to verify whether the use of BH intermediates could also be useful in controlling interferences in the CVG of plumbane, thereby enabling its application to samples of complex composition.
Working solutions of THB were freshly prepared and stabilized with NaOH (0.1 mol L−1, unless otherwise specified). The solutions were clear and colourless and did not require filtration. Working solutions of Pb(II) were prepared by serial dilution of the commercial stock solution (1000 µg mL−1 in 2% HNO3, Fluka) with HCl (final concentration 5 × 10−3 mol L−1). Stock solutions of transition metals at 1000 mg L−1 in water were prepared from the corresponding chlorides. For Cr, V and Mo mixed solutions in water were prepared at 7.5–1.5–1.5 and at 75–15–15 mg L−1, respectively. Working solutions of K3[Fe(CN)6] were prepared by dilution in 0.1 mol L−1 HCl of a stock solution at 10% (v/v) in water, stored at 4 °C in a dark bottle.
A Standard Reference Material (SRM) 663 (chromium–vanadium steel) supplied by the National Institute of Standards and Technology (NIST), was chosen as a material of certain compositional complexity. Three independent samples were heated on a hot plate at 70 °C for three hours in 2 mL of HCl 5 M; then 0.25 mL of 65% HNO3 was added and the heating was continued until completed sample dissolution. The resulting solutions were appropriately diluted for the analysis. All serial dilutions of the solutions, including dissolved samples and Pb(II) solutions, were made by weighing rather than by volumetric dilution.
Ultra-pure water purified by using a Purelab Pro (USF, ρ = 18.2 MΩ cm−1) system was used in all the experiments.
Flow injection CVG (FI-CVG) experiments were carried out using a HPLC pump (Dionex IP20) with water as the carrier and working at 4 mL min−1, and a six-way injection valve (Rheodyne, RH7725I) fitted with a 200 µL sample coil for Pb introduction. Ismatec Tygon microtubing of appropriate diameters was used with peristaltic pumps (Cole-Parmer Masterflex; Ismatec Reglo) for propelling borohydride and ferricyanide, and for the removal of waste solutions from the gas–liquid separator (GLS). Unless otherwise specified, reagent flow rates were 4 mL min−1 for ferricyanide and 2 mL min−1 for THB solutions. The gas–liquid mixture leaving the chemifold was directly delivered to the GLS (60 mm long, 10 mm i.d., borosilicate glass) through a short transfer channel (50 µL volume), which was obtained by connecting the mixing block to the GLS inlet with the aid of PTFE fittings (Bola, Germany). Both coils L1 and L2 were kept low in volume in order to reduce the contact time of plumbane with the solution. The gaseous products leaving the GLS were delivered to the atomizer by a PFA tube (30 cm long, 3 mm o.d., 1 mm i.d.).
Depending on experiments, different chemifolds were assembled in such a way as to be able to vary the mixing sequence and the reaction times of the reagents (see Fig. 1). This was accomplished using various combinations of T-junctions (3-way mixing block, Kel-F, 0.8 mm i.d. channels, Ismatec) and mixing/reaction coils (Teflon PFA, 0.5–0.8 mm i.d.) of various volumes (from 0.1 to 2 mL).
A laboratory assembled atomic fluorescence spectrometer (AFS) was employed for the experiments. The AFS apparatus was described elsewhere.34 A miniature Ar–H2 diffusion flame (130 mL min−1 Ar, 80 mL min−1 H2) supported on an 8 mm quartz tube was used as an atomizer for AFS measurements. A commercially available radiation source (Pb electrodeless discharge lamp – EDL System 2 by PerkinElmer working at 400 mA and modulated at 500 Hz) was used.
![]() | ||
| Fig. 2 FI-CVG-AFS experiments for 100 µg L−1 Pb using 0.1% or 1% FeC and the mixing sequences A (L0 = 0 µL) and B (L0 = 100–2000 µL) reported in Fig. 1. Reference signal S0 obtained with the configuration having L0 = 0. The confidence interval was calculated at 95% from the mean of 3 replicates. | ||
![]() | ||
| Fig. 3 Relative AFS signal for 100 µg L−1 Pb as a function of the hydrolysis coil volume (L0). Curves a and b show measurements in the presence of interfering metals at two FeC concentrations: 0.1% (curve a) and 1.0% (curve b). The interfering metals were present at 100 mg L−1 Co and Ni, 1.0 mg L−1 Cu, and 105 mg L−1 total Cr (75) + Mo (15) + V (15). Curve c corresponds to 0.1% FeC with 10.5 mg L−1 total concentrations of Cr (7.5), Mo (1.5), and V (1.5). All measurements were obtained with setup B (Fig. 1), except those at L0 = 0 mL, which were obtained with setup A (Fig. 1). Confidence intervals were calculated at 95% with n = 3. | ||
The ability of BH intermediates to alleviate interferences is evident at 0.1% FeC (curve a and c). A hydrolysis coil of 0.1 mL appears indeed sufficient to overcome the interferences from 100 mg L−1 Co and 7.5–1.5–1.5 mg L−1 Cr/V/Mo. Nevertheless, these conditions are not able to properly control the interference of 100 mg L−1 Ni (S/S0 = 48%), 75–15–15 mg L−1 Cr/V/Mo (S/S0 = 10%, see curve a in Fig. 3), and 1 mg L−1 Cu (S/S0 = 7%). Increasing the volume of the hydrolysis coil up to 1 mL resulted in some improvement of S/S0 only for Cu (from 7% at L0 = 0 to 25% at L0 = 1 mL). Thus, with the exception of 100 mg L−1 Co and 7.5–1.5–1.5 mg L−1 Cr/V/Mo, increasing the hydrolysis coil volume does not appear to reduce interference to a level suitable for analytical purposes. These conditions were achieved increasing the concentration of FeC to 1%. (Fig. 3, curve b for Co, Ni, and 75–15–15 mg L−1 Cr–Mo–V), with the only exception of Cu where signal depression still persists (Fig. 3, curve b), despite its concentration being two orders of magnitude lower than that of the other metals.
Using 0.1% FeC, and THB (L0 = 0 mL) the interference of 10 mg L−1 Fe(III) is moderate (S/S0 = 80) but it is severe at 100 mg L−1 (S/S0 = 3%) (data not shown for brevity). The use of BH intermediates is able to completely remove Fe(III) interference at 10 mg L−1 (Fig. 4, curve a) but not that at 100 mg L−1 Fe(III), regardless of hydrolysis coil length. Interestingly, increasing FeC to 1% removed the interference of 100 mg L−1 Fe(III) using THB (L0 = 0 mL) and produced relative signal enhancement (120 ≤ S/S0 ≤ 134%) using BH intermediates (L0 = 100–1000 mL) (Fig. 4, curve b). Under the same conditions (1% FeC), 1000 mg L−1 Fe(III) produced severe depression of the signal using either THB or BH intermediates (Fig. 4, curve c).
![]() | ||
| Fig. 4 Relative AFS signal for 100 µg L−1 Pb as a function of the volume of the hydrolysis coil, L0, in the presence of Fe(III). (a) 100 mg L−1 Fe(III) – 0.1% FeC; (b) 100 mg L−1 Fe(III) – 1.0% FeC; (c) 1000 mg L−1 Fe(III) – 1.0% FeC. Measurements obtained with setup B (Fig. 1), except those at L0 = 0 mL obtained with setup A (Fig. 1). Confidence intervals were calculated at 95% with n = 3. | ||
From Fig. 3 and 4, the role played by BH intermediates in improving interference control is clear at low FeC concentrations. At high FeC concentrations, the role of the BH intermediates is much less pronounced because FeC starts behaving like a masking agent, playing a decisive role in eliminating or alleviating interferences.
Using 1% FeC and increasing the hydrolysis loop volume up to 2000 µL lead to better control of Cu interference, up to 2 mg L−1, while maintaining good control of interference from other elements up to 100 mg L−1 (see Table 1).
| L0 (µL) | S/S0b (%) | |||||||
|---|---|---|---|---|---|---|---|---|
| Ni(II)c | Co(II)c | Fe(III)c | Cr(III)–V(IV)–Mo(VI)c | Cu(II)c | ||||
| 100 | 100 | 100 | 1000 | 75–15–15 | 1.0 | 2.0 | 5.0 | |
| a 1% FeC.b Max RSD ≤ 5%, n = 3.c Concentration in mg L−1. | ||||||||
| 0 | 94 | 95 | 104 | 7 | 94 | 54 | 0 | 0 |
| 100 | 112 | 97 | 134 | 17 | 99 | 63 | 0 | 0 |
| 200 | 110 | 89 | 124 | 14 | 97 | 56 | 0 | 0 |
| 500 | 107 | 91 | 126 | 14 | 95 | 72 | 0 | 0 |
| 1000 | 105 | 88 | 120 | 11 | 87 | 62 | 0 | 0 |
| 1500 | 104 | 89 | 120 | 11 | 91 | 97 | 84 | 28 |
| 2000 | 103 | 84 | 87 | 11 | 97 | 97 | 94 | 40 |
These conditions (i.e., 1% FeC and L0 = 1500–2000 µL) ensure good interference control without using any additional reagents. Particularly, the tolerance toward Cu(II) interference is improved dramatically while the other interferences are not a problem using either THB (setup A, Fig. 1) or BH intermediates (setup B, Fig. 1) (Table 1, L0 = 0). The surprisingly high tolerance of the system towards interferences which are observed in this work using THB under analytical conditions (Table 1, L0 = 0) appears to be much better than that reported in the literature,29–32 as already illustrated in the Introduction. Even though the comparison of performances obtained with different chemical conditions and reaction setups is a difficult task in CVG, we believe that in the present apparatus, the use of flow injection (200 µL sample volume) and the low volumes of the chemifold setup (L0 = 0, and L1 and L2 = 50 µL) favored a fast separation of plumbane from the liquid phase, greatly improving interference control, as previously demonstrated by Ding and Sturgeon.35
Preliminary tests were performed on thiocyanate, thiourea, 1,10-phenantroline, and pyrocatechol (initial concentration set at 0.1 M), with 1% FeC. In the presence of 1 mg L−1 Cu, thiourea increased S/S0 from 52% to 77% using THB (Fig. 1, setup A, L0 = 0 mL), while 100% recovery was obtained with L0 = 1 mL. Even with L0 = 0.1 mL, S/S0 increased from 69% to 92%. By contrast, thiocyanate seemed to further depress the Pb AFS signal, with S/S0 equal to 42% (configuration A, L0 = 0 µL) and 46% (configuration B, L0 = 0.1 mL). 1,10-Phenantroline was less effective than thiourea at the same concentration (76% vs. 92%, 1 mg L−1 Cu and L0 = 0.1 mL), and the same applies to pyrocatechol (72%, 1 mg L−1 Cu and L0 = 0.1 mL).
Both thiocyanate and thiourea were ineffective in mitigating 1000 mg L−1 Fe(III) interference, regardless of their concentration (0.1 or 0.5 M). More efficient was the action of 0.1 M oxalic acid, if a comparison is made against the results reported in Table 1, for 1000 mg L−1 Fe(III), using both THB or BH intermediates. In the presence of oxalic acid S/S0 increased from 7% up to 80% using THB (L0 = 0 mL) and, using BH intermediates, from 17% up to 86% (L0 = 0.1 mL), from 14 up to 97% (L0 = 0.5 mL) and from 11 up to 90% (L0 = 1 mL). We also verified that the addition of FeC in water after THB hydrolysis with HCl, in order to avoid FeC reduction from THB, was not helpful in managing copper interference.
The combined effect of 0.1 M oxalic acid and 0.1 M thiourea was then evaluated more systematically using 1% or 2% FeC and in the presence of 1000 mg L−1 Fe(III) or 1 mg L−1 Cu (Tables 2 and 3, respectively). It should be underlined that the masking agent must be added to the sample solution containing both Pb and the interferences, in the following order: first oxalic acid and then thiourea, to reduce and complex iron (the colour changes from dark yellow to light green).
| L0 (µL) | S/S0b (%) [2 mg L−1 Cu(II)] | ||
|---|---|---|---|
| 1% FeC | TU–Ox–1% FeC | TU–Ox–2% FeC | |
| 0 | 0 | 15 | 41 |
| 500 | 0 | 28 | 60 |
| 1000 | 0 | 39 | 60 |
| 1500 | 84 | 49 | 63 |
| 2000 | 94 | 62 | 2 |
| L0 (µL) | S/S0 (%)b [5 mg L−1 Cu(II)] | ||
|---|---|---|---|
| 1% FeC | TU–Ox–1% FeC | TU–Ox–2% FeC | |
| 0 | 0 | 0 | 7 |
| 500 | 0 | 7 | 15 |
| 1000 | 0 | 9 | 16 |
| 1500 | 28 | 10 | 18 |
| 2000 | 4 | 12 | 18 |
From Table 2 it may be concluded that the presence of both masking agents is effective in controlling 1000 mg L−1 Fe(III) interference. Indeed, there is a wide choice of conditions allowing the removal of this interference; it can be controlled using either THB or BH intermediates (with almost any hydrolysis coil volume) in the presence of thiourea + oxalic acid, and using either 1 or 2% FeC.
As shown by Table 3, the relative signal of Pb, with 1 mg L−1 Cu and both additives, was lower than that obtained with thiourea alone, but it approached almost 100% when working with 2% FeC and L0 = 0.5 mL.
These conditions, while ensuring good control of both 1000 mg L−1 Fe(III) and 1.0 mg L−1 Cu(II) interferences (Tables 2 and 3), also maintained effective control of the interference from the other investigated metals: relative signals, S/S0 (%), of 103, 107 and 113 were obtained in the presence of 100 mg L−1 Ni(II), 100 mg L−1 Co(II) and a mixture of Cr(III)–V(IV)–Mo(VI) (75 + 15 + 15 mg L−1), respectively.
| Setupb | Sample | Slope (a.u.) | R2 | Slope ratioc |
|---|---|---|---|---|
| a Six lead concentrations in the range from 0 to 0.2 mg L−1.b See Fig. 1.c Slope normalized to that of setup A–H2O. | ||||
| A | H2O | 14.06 ± 0.09 | 0.9993 | 1.00 |
| A | CASS-4 | 13.53 ± 0.16 | 0.9973 | 0.96 |
| B (L0 = 2 mL) | H2O | 11.66 ± 0.01 | 0.9987 | 0.86 |
| B (L0 = 2 mL) | CASS-4 | 14.48 ± 0.07 | 0.9995 | 1.03 |
The loss of sensitivity observed for setup B–H2O with respect to setup A–H2O is in agreement with the signal loss reported in Fig. 2, and it indicates that plumbane is produced less efficiently, either due to BH intermediates or their concentration. In contrast, the increase in sensitivity for setup B with CASS-4 compared to pure H2O (about 20%) is a clear indication that plumbane formation is improved when the BH intermediates are formed in this type of matrix. This precludes the use of external calibration in pure water for lead determination in seawater. To overcome the limitations of standard addition, we propose better matching of the standard and sample matrices, for example by preparing standard solutions with higher chloride concentrations and by improving control of acidity.
The best control of interferences is achieved by using 1.0% instead of 0.1% FeC. This effect can be attributed to the fact that FeC at higher concentrations, in addition to promoting plumbane formation, also starts to act as a masking agent itself. The use of a long hydrolysis time (L0 = 2.0 mL) produces BH intermediates which are able to significantly improve the tolerance to Cu(II) interference up to 2.0 mg L−1. Under the same conditions, good interference control of 100 mg L−1 of each Co(II), Ni(II), Fe(III) and Cr(III)–V(IV)–Mo(VI) (75 + 15 + 15 mg L−1) is maintained compared to THB reduction (L0 = 0 mL). Therefore, the use of BH intermediates ensures good interference control without the use of additional masking agents.
For the stainless steel sample the high iron content required the use of additional masking agents (oxalic acid + thiourea) and 2% FeC with BH intermediates (L0 = 0.5 mL). The application to lead determination in stainless steel SRM663 gave results in good agreement with the certified value only using THB hydrolysis, while inaccurate results were obtained using THB (L0 = 0 mL).
As final comments, it must be observed that BH intermediates seem to be less effective in interference control during plumbane generation compared to stibane and bismuthane generation.33 This effect could be related to the complex mechanism of formation of plumbane,26 which is probably more vulnerable to the presence of interfering species.