Open Access Article
David T. Murphy
*a,
Balz S. Kamber
a,
Naoki Sugiyama
bc and
Fred Freyerb
aQueensland University of Technology, Australia. E-mail: david.murphy@qut.edu.au
bAgilent Technologies International Japan, Japan
cAgilent Technologies Australia, Australia
First published on 28th January 2026
Quadrupole inductively-coupled-plasma mass-spectrometers (Q-ICP-MS) are used to analyse enormous numbers of isotope ratios, most prominently 206Pb/238U for age determination. With the advent of reaction-cell equipped tandem Q-ICP-MS (Q-ICP-MS/MS), the scope for isotope ratio determination has grown, especially for β-decay isotope systems (e.g., 87Rb/87Sr). Reaction of unwanted interfering species (e.g. 204Hg on 204Pb) has also increased the feasibility of Pb-isotope ratio measurements by Q-ICP-MS. Unlike in sector-field MS isotope ratio analysis where mass bias is corrected via a known isotope ratio or with calibrated double or triple-spikes, Q-ICP-MS(/MS) analysts generally apply an ‘external correction’ with reference to interleaved calibration standard analyses. Despite this protocol, there remains inaccuracies in derived isotope ratios when compared to reference values which are inadequately explained. The aim of this study was to investigate the origin of the inaccuracy in Q-ICP-MS isotope analysis, which has so far received surprisingly little attention. We assessed whether improved detector dead time correction can achieve more accurate isotope results and explored fractionation effects arising from steering ion beams across complex Q-ICP-MS/MS paths. We investigated detector dead time as a function of Z in both single MS and mass-shifted MS/MS mode. We document how dead time varies over time as detectors and electronic components of the ICP-MS age. We show that session specific dead times yield substantially more accurate isotope ratios for mass shifted Sr isotope ratios than when applying a generic dead time. We also provide recommendations for how to incorporate session-specific dead time analyses into an isotope ratio run. Despite the much-improved quality of dead time and conventional mass bias corrected mass shifted Sr isotope ratios, small (up to 4‰) inaccuracies remain. The inaccuracy is systematic and specific for each session. Therefore, it can be corrected with a small bias correction relative to a CRM analysed throughout the session. The origin of the Sr isotope ratio inaccuracy after dead time and conventional mass bias correction remains speculative. Our exploratory analysis suggests that voltages on lenses tuning and directing the ion beams through the MS/MS may be a source of an additional isotope fractionation process (bias) that cannot be fully corrected with conventional mass bias procedures.
It has been recognised for some time that isotope analysis by Q-ICP-MS(/MS) can be inaccurate when using conventional mass bias correction via a known constant isotope ratio, instead requiring additional or alternative correction methods, for example via standard-sample-standard bracketing.3,6,7 This is particularly evident in the Q-ICP-MS/MS analysis of 87Sr/86Sr where Bolea-Fernandez and Balcaen3 proposed that the conventional IUPAC approach of using a known 88Sr/86Sr ratio for mass bias correction does not yield accurate 87Sr/86Sr instead requiring an additional correction (bias) factor.
All isotope analyses by mass spectrometry require correction of the raw data for well-known and understood phenomena including mass bias (e.g. ref. 8–11), removal of isobaric interferences12 and detector dead time.6,13–21 Both mass bias correction and correction of isobaric interferences have received a great deal of attention in the Q-ICP-MS community. By contrast, although detector dead time as a source of inaccuracy in Q-ICP-MS isotope analysis has been well documented,6,7,17,18,22,23 it has received surprisingly little attention relative to its importance in recent years until now (e.g. ref. 15 and 16). Considering session-specific detector dead time correction has not become standard in Q-ICP-MS isotope data reduction. This is at least in part because dead time is not an easily changeable parameter in most modern Q-ICP-MS software. Instead, in most laboratories, dead time is normally set upon installation or detector replacement and not subsequently changed,17 despite detector aging and the likelihood of being sensitive to laboratory and/or pre-amplifier electronics temperature.18
Here we investigated whether detector dead time impacts isotope ratios for elements over a range of atomic numbers (Z: Li, Sr, Ce, Pb & U) and across multiple Sr isotope analytical sessions. This allowed us to assess the suitability of an unchanged dead time set upon installation, to assess the constancy of dead time over most of the mass range of the periodic table and to document dead time fluctuations between different analytical sessions.
We show that with a session specific dead time, accuracy and precision of isotope analysis dramatically improve, and we present a straightforward method for obtaining a session specific dead time. However, minor inaccuracy remains in Q-ICP-MS isotope ratio data corrected with the IUPAC method and we explore potential origins of this inaccuracy.
![]() | (1) |
Russ and Bazan22 proposed a straightforward method for calculating τ by analysing 238U/235U at differing concentrations. They observed that the effect of dead time increased linearly with signal intensity, i.e. concentration of the analysed solution. Given that 238U is 138 times more abundant than 235U at a given concentration, the much larger 238U signal is proportionally more affected by dead time than 235U. This leads to a linear decrease in the observed dead time uncorrected 238U/235U with concentration. Russ and Bazan22 then applied eqn (1) to both the 238U and 235U observed count rates with a τ to achieve a constant 238U/235U with varying concentration. Of note, as mass bias occurs prior to and unrelated to detection in mass spectrometry it should be corrected for separately and after dead time correction.
This approach to deducing dead time can be generalised for any two masses, where the abundance of M1 is greater than M2, as shown in Fig. 1 and the relationship between the observed count rate and τ leads to the linear eqn (2).
![]() | (2) |
![]() | ||
| Fig. 1 Plot of the relationship between the ratio of masses M1 and M2 observed by a detector affected by dead time and the observed count rate of M1 where M1 is the more abundant mass. | ||
As the count rate approaches zero the effect of dead time becomes negligible and thus, in eqn (2), the y intercept (c) is the actual M1/M2 reaching the detector. Tau can be deduced by the reasonable assumption that a slope (m) of zero results from applying the ideal dead time. Thus, solving for τ the equation can be rearranged as follows:
![]() | (3) |
Eqn (1) and (2) can be used to calculate the best fit τ for an experiment provided it includes analysis of a suitable isotope pair analysed at different concentrations. When the actual M1/M2 reaching the detector has been deduced it is possible to use eqn (3) to calculate τ for individual analyses and thus assess whether τ has varied within an experimental run, as might occur if critical conditions, such as room temperature, change in the laboratory. Once the appropriate dead time has been applied to the data for all isotopes investigated, classic mass bias methods can then be applied to obtain the true isotope ratio, assuming there are no isobaric interferences.
More complex methods of τ correction have been proposed that combine the non-extending dead time (eqn (1)) and extending dead time (e.g. ref. 15, 16 and 18). However, for the purpose of this study we use the non-extending dead time eqn (1) initially and then assess whether a combined non-extending dead time and extending dead time correction is required to achieve improved accuracy and precision in the discussion.
Furthermore, we present a preliminary investigation of the effect on dead time with Z after reaction and mass shift during tandem Q-ICP-MS/MS by investigating mass shifted Sr, Ce and U isotope ratio data.
Analytical conditions including isotopes analysed and the duty cycle for each of the experiments are provided in SI Table 1. For the tandem Q-ICP-MS/MS, N2O was used as the reaction gas because of its efficient conversion of Sr+, Ce+ and U+ to SrO+, CeO+ and UO+. Note, the duty cycles for SrO+, CeO+ and UO+ had longer dwell times than those of the other analytical runs (SI Table 1) which gives rise to fewer data points in otherwise similar experimental runs.
Lithium, Ce and U solutions were prepared from ICP-MS 10 µg ml−1 single-component high purity standards from Choice Analytical; with catalogue part numbers and batch numbers provided in SI Table 1. Strontium and Pb solutions were prepared from CARF stock solutions of isotope standards, including National Institute of Standards and Technology (NIST) Standard Reference Materials (SRM) 987 (strontium carbonate) and 981 (equal-atom lead isotope standard).
For each experiment, 2 solutions of the target element were prepared by diluting the stock solution using double distilled HNO3 to achieve1 0.5–1.5 Mcps and2 6 Mcps on the target M1 isotope. The high concentration solution2 was continually added via a pipette into the low concentration solution1 to achieve a continual increase in Mcps over the course of the experiment.
In the right-hand side panels of Fig. 2, dead time corrected data are shown, where τ for each element was calculated using eqn (1) to achieve a slope as close to zero as possible in eqn (2). This can be calculated by plotting the slope from eqn (2) versus a range of τ, where the intercept of a regression through the resultant trend gives the ideal τ for the experiment. We then used the covariance of the dead time corrected data to calculate the τ uncertainty following Iso, OIML.24 The intercepts of the linear regressions through the τ corrected data are indistinguishable to those of the τ uncorrected data. There is a significant variation in τ with Z, with the calculated τ for Li of 7.8 ns lower than that for U (Fig. 3).
![]() | ||
| Fig. 3 Plot of mass number (Z) versus calculated τ for single Q-ICP-MS mode experiments (blue-coloured symbol) and mass shift Q-ICP-MS/MS mode (orange-coloured symbols). | ||
The linear regression through the τ values for the different isotope ratios investigated (Fig. 3) indicates a Z-dependent variation of 7.7 ns over the 0–270 amu mass range, slightly higher than that set upon machine installation, albeit with a high uncertainty on the fit.
The calculated τ values for both SrO and CeO are indistinguishable from the equivalent single Q-ICP-MS data. The UO τ is indistinguishable from those of SrO and CeO but slightly lower, outside uncertainty of the single Q-ICP-MS data for U isotopes (Fig. 3).
For NIST SRM 1570a ‘spinach leaves’ (87Sr/86Sr 0.70905 ± 0.00002;25) a 50 mg aliquot was dissolved in concentrated double-distilled HNO3, dried down, and the residue was dissolved in 2% HNO3 to make stock solutions that could be diluted for analysis.26
For the USGS-AGV2, USGS-BIR-1, USGS-W2a rock CRMs, 50 mg aliquots were digested in a milestone bench-top ultra-wave single reaction chamber microwave digestion system (Milestone Srl, Sorisole, Italy) following ‘Method guideline #1: silicate rocks without resistant minerals’27 to generate stock solutions.
All CRM dilutions were prepared from previously dissolved stock solutions and diluted to give signals between 0.5 and 4 Mcps on 88Sr16O. None of the solutions underwent Sr purification. Over the study period, the PA cross-over was set either to 1.5 or 5 Mcps. Only data analysed in pulse mode were included.
As most mass bias in a Q-ICP-MS occurs in the plasma torch, for two of the experiments run on the same day (320
707 – 8 mm & 320
707 – 6 mm) the distance of the torch to the cones was changed to deliberately vary the mass bias.
707 – 8 mm & 6 mm) the Mcps 88Sr16O = 0 intercepts are very different (88Sr16O/86Sr16O of 8.81 and 8.88 respectively). Despite this, the calculated dead times for the two experiments are within uncertainty. Since most of the experiments have different Mcps 88Sr16O = 0 intercept 88Sr16O/86Sr16O ratios, it is likely that dead time and mass bias are not related.
![]() | ||
| Fig. 6 Plots of mass bias corrected 87Sr16O/86Sr16O, using an internal correction derived from the 88Sr16O/86Sr16O, divided by accepted 87Sr/86Sr for the CRMs presented in Fig. 5 (same symbols). Panel (A) shows data that were dead time corrected with the fixed τ of 41.5 ns, as set upon installation. For samples with high Mcps for 88Sr16O (see Fig. 5) highly inaccurate results are obtained. Panel (B) shows the same data corrected with a session specific dead time (Fig. 5). | ||
However, in the experiments where Mcps 88Sr16O range from <0.5 to >2 the routine set dead time (τ = 41.5 ns) has a major effect on the mass bias corrected 87Sr16O/86Sr16O with percentage level deviation from the accepted isotope ratio. In the three experiments where different concentrations of NIST SRM 987 were analysed, yielding a range of 88Sr16O signal intensities of 0.3–2 Mcps, different mass bias corrected 87Sr16O/86Sr16O ratios were obtained. When using the installation set dead time (Fig. 6a) at low Sr concentrations, the sample's internal correction achieves a mass bias corrected 87Sr16O/86Sr16O relatively close to the accepted values. However, as Mcps 88Sr16O increases the mass bias correction leads to increasing over-correction of the 87Sr16O/86Sr16O that scales with Mcps 88Sr16O (Fig. 6a). The session specific dead time completely removed this effect with all concentration solutions having indistinguishable 87Sr16O/86Sr16O (Fig. 6b).
The dead time and mass bias corrected data presented in Fig. 6b include several experiments where all CRMs are within 2SE of the accepted 87Sr/86Sr values and others that are not. Despite this, in all experiments the ratio of corrected 87Sr16O/86Sr16O to accepted 87Sr/86Sr for each CRM is within uncertainty of each other. This means that for each of the experiments where the corrected 87Sr16O/86Sr16O ratios are not within uncertainty of the accepted 87Sr/86Sr, a systematic cause for inaccuracy is indicated. This systematic deviation allows for an external calibration, whereby one CRM (e.g. NIST SRM 987) is used to estimate the residual bias, which can then be applied to the remaining CRMs. With this final correction, the accuracy of CRMs run as unknowns can be estimated (Fig. 7).
A number of physical mechanisms that affect ion beam transmission have the potential to be responsible for residual inaccuracy; these include beam emittance and acceptance, beam matching, energy-dependent transmission and space-charge effects.28–30
One of the primary ways of assessing the effect of beam emittance and acceptance and beam matching while tuning a Q-ICP-MS is to investigate peak shapes. Therefore, we first explored peak shapes in Q-ICP-MS as a possible cause of inaccuracy, as it is one obvious difference from the high mass resolution MC-ICP-MS and TIMS, which exhibit flat topped peak shapes. In contrast, due to the lower abundance sensitivity of a Q-ICP-MS the peak shapes have more obvious asymmetry. In addition, during Q-ICP-MS routine auto-tuning, the peak shape is optimised for a wide range of Z, rather than the mass range measured in isotope ratio analysis. This global tuning can lead to asymmetrical peak shapes. In time resolved Q-ICP-MS isotope analysis, as was used in all our experiments, only one data point per peak is acquired. The exact position of Z for analysis is not determined by considering the peak top itself, rather as the half-way point of the shoulders of the peak. With a substantially asymmetric peak, the half-way point may not necessarily sit at the highest point on the peak. Therefore, we reasoned that if the peak shape is not carefully checked prior to an experiment, it could be possible that the analysis point sits slightly off-peak, which could lead to a systematic bias in the resultant isotope ratios. To explore this possibility, a series of experiments was run with deliberately varying the peak shape and symmetry and also varying the number of data points acquired per peak. The data showed no influence on isotope ratio accuracy and they are therefore not presented here.
It is worth noting that our deliberate variation of the peak shape and peak symmetry caused changes in beam emittance and acceptance and beam matching. Although not a systematic test of these factors, the lack of change in isotope ratio with peak shape strongly suggested that the residual inaccuracy is not dominantly caused by varying ion beam emittance and acceptance and beam matching.
Energy-dependent transmission was explored for MC-ICP-MS by Albarède and Albalat.28 They showed that isotope fractionation caused by energy-dependent transmission was predominantly due to the configuration of the ICP MS cones relative to the plasma torch, was mass dependent and could be corrected for. We argue that because the fundamental architecture of the plasma torch, cones and extraction ion optics is broadly comparable between MC- and Q-ICP-MS, energy-dependent transmission would also be mass dependent in a Q-ICP-MS. Our experiment of varying the torch distance confirms this. By varying the distance of the torch from the cones, ions are extracted from different regions of the plasma, which affects their kinetic energy distribution and thus energy-dependent transmission. As no variation was observed between the dead time and mass bias corrected data for these two experiments, we infer that energy-dependent transmission does not cause the additional bias required to correct the experimental data.
We next reviewed the analytical conditions of each run of the data presented here (SI Table 1) and by Zepeda and Kamber,31 who manipulated various tune parameters to minimise the BaO interference on Eu to achieve accurate Eu concentration analysis in high Ba samples. From the review, we hypothesised that the extent of ion beam deflection when directed through the offset between the reaction cell and quad 2 could impact accuracy. Therefore, we investigated whether varying the voltages on extract 2, plate deflect, and plate bias could help to elucidate upon the origin of the residual inaccuracy.
For these tests, all analyses were run on the same instrument with analytical conditions for each run presented in SI Table 1. Three sets of experiment were conducted using the analytical conditions of the 220
907 run that gave both precise and accurate dead time corrected and internally corrected 87Sr16O/86Sr16O (Fig. 6b). In each set, one single parameter (extract 2, plate deflect and plate bias) was varied.
For each experiment three different concentration solutions of NIST SRM 987 were run (15, 30 and 60 ppb), three times each so that dead time could be calculated independently for each run.
Varying extract 2 voltage produces a linear relationship with sensitivity and within-run precision, with the standard deviation of the corrected 87Sr16O/86Sr16O varying from 0.09% to 0.04% as the voltage is lowered from −240 to −160 V (Fig. 8). No variation in precision was observed when varying the voltage on plate deflect or plate bias (Fig. 8).
Neither extract 2 voltage nor plate deflect voltage showed a resolvable relationship with dead time and mass bias corrected 87Sr16O/86Sr16O. By contrast, a clear negative linear relationship is observed between plate bias voltage and accuracy (Fig. 8). The trend predicts that a plate bias voltage of close to 0 would yield accurate results. However, the signal sensitivity dramatically dropped above −21 V for plate bias and thus using 0 V was not viable.
The plate deflect and plate bias lenses control how the ion beam bends as it travels from the reaction cell into the second quadruple. This bend is required to stop non-charged particles formed in the reaction cell getting to the detector. It is possible that space charge effects, where a build-up of charge causes repulsion of similarly charged ions, could occur due to the high voltage applied to the plate bias.
Importantly, other types of mass spectrometers also employ lens sequences to direct ion beams, including multi-collector ICP-MS (MC-ICP-MS) with collision and reaction cells. It is possible that similar isotope ratio inaccuracies could occur in data derived from such instruments. For example, K isotope ratio measurements by reaction cell MC-ICP-MS recommend analysing solutions of constant K concentration, suggesting that variations in ion signal intensities can also cause inaccuracies in those instruments.32
Another possible cause of residual inaccuracy is where strongest space charge effect occurs in the cones between the plasma torch and the extraction optics.29,30 Yu and Hathorne30 demonstrated that plasma temperature influences mass bias and that low plasma temperatures lead to mass bias that cannot be internally corrected for. Further investigation of the space charge effect would require further experiments and is beyond the scope of this study.
![]() | (4) |
The formula can be used to assess the impact of extending dead time on an isotope ratio over a range of observed counts. The effect of correcting data that have been affected by non-extending and significantly extending dead times only with eqn (1) is shown in Fig. 9 for M1/M2 = 10 and τne = 40 ns. There is a clear increase in non-extending dead time corrected M1/M2 as the count rate increases, with high extending dead times having a greater deviation. Within the count range of interest to Q-ICP-MS (<2.5 Mpcs) only if the extending dead time is >25% of the non-extending dead time would you expect a resolvable (≥1‰) effect on isotope ratios analysed with Q-ICP-MS.
![]() | ||
| Fig. 9 Plot of the effect on an isotope ratio (M1/M2 = 10) where both non-extending dead time (40 ns) and extending dead time have a significant impact. The data were calculated with eqn (4) and only corrected for non-extending dead time (eqn (1)). | ||
We do not see a resolvable positive relationship between the count rate and non-extending dead time and mass bias corrected isotope ratios in our experiments. Therefore, we conclude that the extending dead time is <25% of the non-extending dead time, and thus not the cause of residual imprecision in corrected Q-ICP-MS isotope ratios and does not need to be corrected for. However, in higher precision instruments where an accuracy of ≪1‰ is required, extending dead time may be significant.
Our study indicates that there is a positive near-linear relationship between mass number (z) and dead time in single Q-ICP-MS mode, but not in reaction mode; further research is required to test this.
Investigating Sr isotope ratio experiments conducted episodically over 5 years (2020 to 2025) indicates that dead time for a single element can vary by several ns. It is likely that this range relates to ageing ion counters, ageing electronics and changing laboratory conditions such as temperature and humidity.
Session specific dead time and mass bias corrections yield reproducible isotope ratios over a range of element concentrations. However, the corrected isotope data can be systematically moderately inaccurate. Because the inaccuracy is consistent across all analysed CRMs, a simple final bias correction can be applied when one CRM is used as a calibration standard for the purpose of achieving accurate results for the other CRMs and the unknowns.
We started an investigation into the origin of the systematic residual inaccuracy and found that varying the voltage to one of the lenses that deflects the ion beam from the reaction cell through to the second quadrupole affected the degree of systematic inaccuracy. Thus, we conclude that there is an as yet unidentified process that is causing isotope fractionation in Q-ICP-MS/MS and potentially other mass spectrometers that require steering ion beams across complex paths.
We find no evidence for the requirement to correct for both non-extending and extending dead times for Q-ICP-MS isotope ratios measured at precisions of 1 to 2‰. However, for higher precision isotope analysis, extending dead time may be an important factor to obtain more accurate data.