Raman spectroscopy enables non-destructive chemical analysis and can be performed through-vial, eliminating air exposure. However, signal attenuation arises from refraction, absorption, and scattering within the sample, while distortions and spectral interference from the vial further degrade signal collection and data quality. This study investigates depth-dependent Raman intensity decay in polystyrene (PS) particle suspensions as a model for turbid liquids and proposes a framework of equations and approaches based on Mie scattering theory and the Beer–Lambert law to predict Raman intensity decay in a range of samples. Experimental validation was performed using aqueous suspensions containing 495 nm, 350 nm, and 200 nm PS particles using an 830 nm laser, with Raman spectra acquired at incremental depths. Peak areas corresponding to PS, water, and glass were recorded. Mie theory was used to model the UV-visible extinction response, yielding estimated absorbance and, combined with the intensity decay due to refraction established from non-turbid samples, Raman intensity decay lengths. Estimated decay lengths for PS and water showed good agreement with experimental values, typically within 20%. This framework enables rapid optimisation of focal depth and consistent comparison between samples and experiments for quantitative through-container Raman analysis in pharmaceuticals, materials science, and nanotechnology.
One of the key advantages of Raman spectroscopy is its ability to allow chemical detection and quantification of an analyte non-destructively. The configuration of the incident laser source and detection system allows measurements to be performed through containers such as glass vials, by directing an external laser to illuminate the sample within.4,5 These measurements support a wide range of applications – including in pharmaceuticals,6 forensics,7 and security8 – where preserving the sample or protecting against exposure is essential, either because only a small quantity of the analyte is available or because the toxicity of the sample is high or unknown. A common challenge for many of these applications is the rapid loss of Raman signal within the sample volume due to light absorption9 and elastic scattering by particles in the dispersion. This is more significant for larger particles and often constrains the location of the measurement to within few tens of microns of the sample container. However, the strong Raman signal from, for example, the glass container that can overwhelm the signal collected from within the vial and overlap with the analyte peaks of interest.10 Complex optical geometries of laser-illumination and photon-detection may be used such as spatially offset Raman,10,11 to minimize or eliminate the glass signal,10 in the conventional backscattering confocal Raman spectroscopy scheme, peak fitting methods are often necessary to quantify the Raman signal of interest from deeper into the vial where the signal overlaps with background signals from the vial itself in addition to the decrease in intensity due to the diffusion of photons in the medium.
Another challenge for such measurements is optimising the focusing parameters for layered (through-container) samples containing materials of differing refractive indices. Such samples will reduce Raman signal intensity and decrease axial resolution with increasing penetration depth, making it challenging for the Raman analyst to choose optimum conditions for reliable qualitative and quantitative measurements. To avoid signal attenuation, depth resolution should be maintained close to the nominal value. This can be achieved by avoiding laser refraction by using an immersion objective coupled with a fluid that matches the refractive index of the sample,12,13 or by analysing a serial cross section of the sample.14 Neither of which are practical for through-vial Raman measurements of a composite sample, therefore modelling and fitting methods can facilitate data interpretation in such measurements.
Over the past three decades, some effort has been devoted to interpreting Raman depth profiles in dry objective systems. For thin samples, Tabaksblat et al. employed a geometrical optics framework to characterise the intensity variation within transparent samples in the conjugate back-focal plane while ignoring diffraction effects,15 De Grauw et al. refined the model by considering a Gaussian beam profile instead of a spherical beam, leading to a Lorentzian Raman intensity axial distribution for a thin polystyrene film.16 For thicker planar interfaces, Everall developed a simple model to quantify the distortions encountered by the laser-illuminated spot as it undergoes refraction at the air/sample interface in accordance with Snell's law, demonstrating that the focal volume of illumination substantially exceeds the theoretical diffraction limit (several tens of micrometres) and the axial depth resolution may also degrade by several orders of magnitude relative to its nominal diffraction-limited value, with the extent of deterioration increasing with imaging depth.12,17,18 Baldwin and Batchelder's model incorporated a pinhole in the collection pathway to the model, which reduced the focal volume, and highlighted that collection efficiency markedly decreases with increasing focus depth.19 Further refinement by Sourisseau et al. using a vectorial electromagnetic framework, incorporated diffraction effects in their computational model – including off-axis incident intensity distributions, diffraction integrals for confocal aperture transmission, and phase aberrations – accurately predicting through-plane confocal Raman responses at varying focus depths.20 Others have experimentally validated and refined the models, Tomba et al. experimentally validated the simpler models by incorporating on-axis laser intensity profiles and renormalising the depth-resolution curves using independently measured collection efficiencies which significantly enhanced the agreement between experimental data and model predictions,21 they also proposed a regularised deconvolution approach to partition the sample thickness into multiple planes, reconstructing the overall emission intensity by combining the contribution from each.22 More recently, Chakraborty and Kahan extended the modelling and experimental validation of total intensity depth profiles in surface and volume emitting samples.23 Maier and Bohm presented a similar computational model that deconvoluted the system's point spread function from ideal sample profiles, accounting for refractive index mismatches and subsurface intensity loss, the model was able to accurately predict axial resolution and its decay with depth for layer thickness estimation when the sample's refractive index is known.24
Recently Giovannini et al. presented a comprehensive theoretical perspective on the challenges in modelling Raman data in complex environments, such as solutions, highlighting the importance of understanding solute–solvent interactions, phase space and molecular geometry.25 However, there is a lack of modelling validation studies focused on turbid colloidal systems containing micro- or nanoparticles.
In this paper, we use polystyrene (PS) nanoparticle colloids contained in a borosilicate glass vial – a typical container used in analytical sciences, to address the experimental and modelling challenges mentioned above through different approaches – (i) for experimental data, using direct least squares analysis to extract the PS, the container (glass) and the medium (water) signals at each incremental depth, and (ii) modelling the UV-visible and Raman intensity decay of PS particles dispersions to understand and ultimately predict the loss of Raman signal in turbid samples in a container. Modelled UV-visible and Raman data has been validated with experimental data. Successful modelling of Raman signal decay in turbid samples will enable analysts to adjust optical focusing to obtain more accurate quantitative measurements and to understand the limited sampling depth available in scattering samples.
For least squares analysis, individual components were extracted by performing depth scans of a water only sample in a vial using the same parameters described above. Individual glass components were extracted for each sample from within the depth series (around 140 µm within the glass wall). Individual PS components were arithmetically obtained by subtracting the isolated glass and water components. Least squares analysis was performed using a custom Python script, using Python 3.11.11 version and scipy.optimize module. Graphs were produced in Microsoft Excel and Origin 2025.
We find that the depth profile of the glass, taken from the intensity of the glass vial peak in the Raman spectra, can be fitted with a model function which considers the two interfaces of air–glass and glass–PS suspension. The approach is based upon the practical, empirical description of Maier and Böhm.24 Resolution functions at each interface are assumed to be described by a Lorentzian distribution and intensity decays by exponentials. We describe the shape of the intensity trace of glass by the combination of two cumulative Lorentzian resolution functions, where the first interface is modelled by one function representing increasing intensity, and the second interface modelled by another function which can be subtracted from the first to represent the decrease in intensity. The resulting intensity trace can be described by:
![]() | (1) |
Raman depth profiles of PS and water (experimental peak centred at 1002 and 1638 cm−1 respectively) were fitted with an exponential decay described by
![]() | (2) |
When interacting with incident light, PS particles are predominantly scattering in nature and exhibit low absorption at the Raman source and scattered wavelengths, and therefore the Raman decay length is primarily influenced by light penetration depth. Mie theory provides a comprehensive description of light scattering by spherical particles, offering exact solutions to Maxwell's equations in the context of particle sizes comparable to the wavelength of incident light.30,31
The Mie solution begins with the assumption of a plane-polarized electromagnetic wave incident upon a spherical particle with a well-defined diameter, D and complex refractive index, m(λ) = n(λ) + ik(λ), where λ is the wavelength of light, and n and k are the wavelength-dependent real and imaginary parts of the refractive index, respectively. The surrounding medium was assumed to have a refractive index, nmed, and the size parameter, x, is defined as:
![]() | (3) |
Given the size parameter and relative refractive index (mrel= mp/nmed), where mp is the complex refractive index of the particle, we used the widely available miepython module to calculate the Mie coefficients an and bn which arise from the series expansion of the scattered and internal fields in terms of spherical Bessel and Hankel functions.29 This ensures continuity of the tangential electric and magnetic fields that arise at the particle boundary as described by Maxwell's equations. Further details on miepython are provided in the materials section. From these coefficients, the wavelength-dependent Mie efficiencies were computed.
Extinction efficiency is calculated according to eqn (4):32,33
![]() | (4) |
describes the real part of the sum of the Mie coefficients, and N is the number of terms considered in the series. We note that Qext strongly depends on λ through the size parameter x and the Mie coefficients an and bn which are also a function of the size parameter, as well as through other parameters of the terms of the Bessel and Hankel functions.
These quantities could then be used to calculate model UV-visible extinction spectra using the Beer–Lambert law in its natural logarithm form, as in eqn (5), to calculate the effective absorbance, A, at a given wavelength. Similar works have been done elsewhere for nanoparticles of various materials and sizes.34–36
| A = 0.3411r2QextCL | (5) |
![]() | (6) |
Eqn (7) describes how light intensity decays exponentially as it passes through a material of thickness d, where ∧ is the characteristic decay length which represents the distance over which light intensity decreases by 1/e.
![]() | (7) |
![]() | (8) |
The overall Raman decay length was then estimated by combining the different sources of decay along the pathway, using the harmonic sum described in eqn (9). This is equivalent to the summation of effective cross sections for each of the processes:
![]() | (9) |
C and ring skeletal stretch respectively from PS.38,39 A peak at 1557 cm−1 is assigned to gaseous oxygen40 in atmospheric air within the instrument chamber. A near infrared 830 nm incident laser was chosen to avoid fluorescence signal originating from the borosilicate glass vial. These spectra demonstrate the overwhelming extent to which the signal from the container (glass) dominates the spectra – while the water and PS peaks are detected at depths of 100 µm (C) and 200 µm (D) below the inner glass surface, corresponding to their actual physical locations, glass signals appear throughout the analysis depth, including in regions where it is not physically present, such as in air (A) and solution (C, D). To address this issue and effectively separate the PS and water signals from the glass and from each other, direct least squares analysis was applied independently at each of the 231 depth increments. Raman spectra were acquired from 200 µm above the top wall of the glass vial to approximately 1600 µm into the vial. Fig. 3a shows one of the Raman spectra from the depth series measurement of 350 nm PS particles (2 g L−1), and the area of the characteristic PS (green), water (blue) and glass (red) peaks with maxima at 1002 cm−1, 1638 cm−1 and 1794 cm−1 respectively. Fig. 3b–d show the intensity of each component as a function of depth, where the glass–water interface is taken as the reference position.
PS and water signal areas reached a maximum at a depth of <100 µm inside the vial before decaying exponentially. The glass signal exhibited no noticeable loss of signal over the 500 µm wall thickness.
Fig. 4a shows the area of the PS peak centred at 1002 cm−1, at increasing depths for 495 nm, 350 nm and 200 nm (2 g L−1 and 4 g L−1) within the glass vial. Assuming that the deterioration of axial resolution due to refraction was similar for all samples, the decay of Raman signal with depth is due to absorption and scattering of incident and collected light along the optical path as size and concentration of particles increases. Doubling the particle mass concentration from 2 g L−1 to 4 g L−1 resulted in increases of 1.6-, 1.7-, and 1.5-fold in the maximum PS signal (expressed as area) for particles measuring 495 nm, 350 nm, and 200 nm, respectively. These increases were lower than expected, because the Raman signal is expected to scale linearly with concentration suggesting a two fold increase should have occurred. One possible explanation is sedimentation of particles, which may have reduced the effective PS signal.
Fig. 4b shows the area of the water peak centred at 1638 cm−1 for samples containing different size and concentrations of PS particles, and water in the absence of PS (in cyan). Larger and more concentrated particles showed an exponential loss of water signal due to scattering, while decay of water signal in samples with smaller 200 nm sized particles looked closer to the profile of water alone. Due to the small losses observed in the water signal for 200 nm size particles, water signal could also be used to normalise the PS signal to account for optical losses with depth and measurement conditions (Fig. 4c) to obtain consistent results at all depths. This approach has previously been used for signal correction in the relative quantification of polymeric particles in water.4,41 However, for larger particles, the signal collection for water was too poor to be used as a correction reference, indicating the importance of understanding particle size and the resulting turbidity dependent Raman signal decay behaviour.
In-depth experimental studies such as the one mentioned in this work that are crucial to devise quantitative studies are resource-intensive (>10 hours acquisition time for each sample), therefore a modelling approach would be valuable for Raman analysts to faster-optimize their measurements. A simple approach would be to measure the UV-visible extinction at the wavelengths used in the Raman experiment. However, this is not always feasible for the type of samples used by analysts where extraction and dilution can change the nature of the samples. In this case, the wavelengths used in Raman were outside the range of most UV-visible spectrometers and therefore an alternative approach was used. To model the Raman signal of transparent PS particles, the first step involved calculating their theoretical extinction which would enable the estimation of extinction behaviour of PS and water at their Raman emission wavelengths (905 nm for PS and 961 nm for water). The calculated extinction spectra were validated within the range of a UV-visible spectrometer. Fig. 5 shows the extinction spectra of PS particles experimentally measured using UV-visible spectroscopy (solid lines) and theoretically modelled using Mie theory (dashed lines). The experimental data is an average of three replicates. The results demonstrated a good agreement showing that Mie theory effectively describes the optical behaviour of these particles. The vertical lines display the incident laser wavelength (830 nm) and emission wavelengths of PS and water.
Fig. 6 shows representative results of 350 nm (4 g L−1) PS particles in water obtained by fitting of experimental Raman spectra with individual components using least squares analysis as described in the methods (dots), and the intensity depth profiles fitted with curves described in the modelling section (in solid line) for glass (Fig. 6a), PS (Fig. 6b) and water (Fig. 6c). The residual distribution can be found in supplementary information and indicated a good fit between the two data sets.
The Raman decay lengths were obtained by fitting the experimentally obtained intensities with curves described in the modelling section. The resulting values are compared to theoretically estimated decay lengths calculated according to eqn (9).
Table 1 shows the estimated and experimentally obtained Raman decay lengths of the different size and concentrations of PS nanoparticles. These lengths describe the effective distance from the glass–water interface over which PS and water Raman signal can be collected with high confidence. Experimental values have an estimated uncertainty of around 10%, arising from the fitting uncertainty, instrumental noise and sample dependent effects. A good agreement was observed between the estimated and experimental decay length data sets with a typical agreement within ∼20%.
| Diameter (nm) | Concentration (g L−1) | Estimated decay length (µm) | Experimental decay length (µm) | Experimental/estimated decay length | |||
|---|---|---|---|---|---|---|---|
| PS | Water | PS | Water | PS | Water | ||
| 200 | 2 | 576 | 597 | 483 | 561 | 0.84 | 0.94 |
| 350 | 2 | 308 | 322 | 281 | 330 | 0.91 | 1.03 |
| 495 | 2 | 202 | 214 | 184 | 167 | 0.91 | 0.78 |
| 200 | 4 | 399 | 420 | 294 | 330 | 0.74 | 0.79 |
| 350 | 4 | 181 | 191 | 159 | 167 | 0.88 | 0.88 |
| 495 | 4 | 112 | 119 | 114 | 112 | 1.02 | 0.94 |
The loss of Raman signal from PS with depth was both size and concentration dependent – the experimental decay lengths decreased as the particle size or concentration increased. The overall decrease in Raman intensity with increasing particle size has been previously documented in the literature,42,43 with some exceptions.43 In addition to the previously discussed axial resolution worsening with depth which may contribute to signal loss, another possible effect is the poorer signal collection compounded by the increased extinction as PS solution gets visibly more turbid when the particle size or concentration increases.
By understanding the distance over which the scattered light significantly loses intensity within the vial, Raman spectroscopy instrument users designing their particle detection and quantification experiment can determine the appropriate acquisition depth to maximize the signal of interest. This approach establishes a framework that enable to compare results and predict different samples and experiments. For example, application of the present approach enabled to predict Raman decay lengths of 711 µm for PS and 730 µm for water in suspensions containing 125 nm PS particles, consistent with the ∼700 µm upper limit for reproducible measurements experimentally reported by Legge et al.4 When sample dilution is not possible, Raman analysts can use this modelling approach to improve the data interpretation of depth-dependent Raman measurements in through-container measurements.
Overall, this approach provides a rapid way to predict how Raman signals weaken as they penetrate deeper into a sample by incorporating both confocal resolution effects and intensity decay due to scattering losses. Analysts can apply this knowledge to optimize their experimental parameters, ensuring optimum signal collection and awareness of the confounding factors such as background interference or signal loss due to optical limitations.
Limitations of this approach include prior knowledge about the refractive index, size and mass concentration of the particles. The size of the particles can be easily measured by methods such as dynamic light scattering or particle tracking analysis. The mass concentration can be measured by performing titration measurements to calibrate the Raman signal. Mie scattering also has limitations that affect its accuracy in real-world applications such as disregarding multiple scattering events, and assumptions including particles being perfect spheres, monodisperse and optically uniform, and the surrounding medium being non-absorbing and homogeneous, which may not reflect the actual conditions in a sample. For higher confidence, a greater number of experimental and theoretical replicates must be generated, and future work should extend these models to account for absorbing particles and other non-geometrical structural complexities, further enhancing the predictive capabilities and broadening the potential applications of this technique in diverse analytical scenarios. For example, Lu et al. have developed a method for correcting the self-absorption effect, such that the non-attenuated Raman spectrum can be recovered.9