Rapid bloodstream infection diagnosis is essential to enable timely targeted therapy, to reduce the risk of progression to sepsis, to limit unnecessary antimicrobial exposure, and to support effective antimicrobial stewardship. However, conventional culture-based workflows in routine practice often require 48–72 hours to deliver organism identification and antimicrobial susceptibility testing (AST) results. This review summarises rapid AST analytical technologies for bloodstream infections, including commercial systems and emerging phenotypic and molecular methods that aim to shorten time to actionable susceptibility information to less than 8 hours from a positive blood culture. Phenotypic platforms are to be preferred, and they are compared and reviewed according to their dominant analytical principle, while the role of genotypic assays for early identification and resistance inference is discussed as a complementary strategy. Across approaches, sample preparation, including bacterial enrichment and matrix removal, emerges as the dominant limitation, while speed and organism–antibiotic coverage shape clinical utility and implementation. Progress of rapid AST platforms toward routine adoption will require streamlined sample preparation, broader panels, and robust evidence of clinical benefit and cost-effectiveness.
939 in the US.1,2 Globally, BSI are associated with nearly 3 million deaths each year,3 with reported 30-day mortality rates varying from 17% to almost 40%, depending on the geographical region and the causative pathogen.4–6 Escherichia coli, Klebsiella pneumoniae, and Pseudomonas aeruginosa are among the most frequently isolated pathogens, while Staphylococcus aureus accounts for the highest number of fatal bloodstream infections.5–7 Importantly, although BSI reflect invasive infection and may lead to pathogen dissemination to distant sites, mortality is primarily driven by sepsis, a dysregulated host response to infection that results in life-threatening organ dysfunction,8 and its progression to septic shock.
Blood culture remains the reference standard for the diagnosis of bloodstream infections, bacteraemia and sepsis. The pathogen concentrations in whole blood are typically very low, often between 1 and 100 colony-forming units per mL (CFU mL−1),9 and these numbers are masked by abundant host cells. Automated blood culture systems allow pathogen detection and enrichment, typically requiring 10–19 hours to signal positivity, and up to 5 days to confirm a negative finding.10 Time to positivity is dependent on the growth rate of the microorganism. While blood culture enables reliable downstream identification and susceptibility testing, it introduces a significant delay between sampling and actionable results (Fig. 1), during which patients are commonly managed with empirical broad-spectrum therapy.
![]() | ||
| Fig. 1 Clinical workflow for identification and antimicrobial susceptibility testing of bloodstream infections. | ||
Following blood culture positivity, bacterial identification is typically performed using Matrix-Assisted Laser Desorption/Ionization Time of Flight Mass Spectrometry (MALDI-TOF-MS).11,12 This method has been shown to reduce turnaround time significantly, the time to effective antibiotic administration, and the length of hospitalisation compared to conventional phenotypic testing,13 and therefore represents a cost-effective solution in high-volume settings such as microbiological laboratories within secondary care centres. Despite its advantages, MALDI-TOF-MS presents some limitations in its ability to identify polymicrobial cultures and anaerobic bacteria, as well as showing limited identification below bacterial species level.14 Consequently, complementary approaches such as molecular methods and biochemical tests are still commonly employed in routine diagnostics, as well as in low-resource settings.
Antimicrobial Susceptibility Testing (AST) plays a crucial role in guiding clinicians on the appropriate antimicrobial treatment of bloodstream infections, improving patient outcomes and limiting the emergence of antimicrobial resistance. AST is traditionally performed using culture-based methods, such as disk diffusion and broth microdilution. While these tests are considered the gold standard for accuracy and reliability, they are labour-intensive and time-consuming, often requiring 24 hours or more after blood culture positivity to deliver results. In recent years, the European Committee on Antimicrobial Susceptibility Testing (EUCAST) and the Clinical and Laboratory Standards Institute (CLSI) introduced guidelines for performing rapid disk diffusion methods directly from positive blood cultures (PBC).15,16 These methods shorten the time to results to approximately 8 hours but are restricted to a limited number of bacterial species and antibiotics and often require multiple manual readings. Nevertheless, rapid disk diffusion methods rely on routinely available equipment and consumables and, given their low cost and simplicity, remain a useful option for low-resource settings.
As an alternative to manual methods, automated platforms such as the VITEK 2 (bioMérieux), BD Phoenix (Becton Dickinson), and Microscan WalkAway (Beckman Coulter) have dominated the AST technology landscape in routine clinical testing over the last few decades. These platforms rely on biochemical reactions (colorimetric or fluorescent) and turbidimetric measurements to identify a broad range of organisms and determine their antimicrobial susceptibility. They have shown good reliability, are easily integrated in the laboratory workflow, and although a standardised inoculum is usually needed, studies have shown that they can provide identification and susceptibility profiles directly from a positive blood culture for a range of organisms and antimicrobials.17–19 However, their time to results often ranges from 8 to 24 hours, which limits their utility in the management of bloodstream infections and sepsis, where early targeted intervention is critical.
An ideal AST platform would deliver actionable results within a clinically meaningful time window, enabling early therapy optimisation. The emphasis on speed is supported by sepsis/septic shock data showing that each hour of delay in effective antibiotic administration after the onset of hypotension is associated with a 7.6% decrease in survival.20 Beyond rapid turnaround, an optimal test would have high sensitivity at low pathogen concentration, cover a wide range of organisms and antimicrobials, and perform reliably in cases of polymicrobial infection. Equally important are ease of use (minimal handling, streamlined workflow, and compatibility with laboratory practice) and clear interpretability of the results. Finally, balancing initial costs with downstream savings from reduced broad-spectrum antibiotic use and shorter hospital stays would support widespread adoption.
Delayed initiation of effective antimicrobial therapy in bloodstream infection and sepsis has consistently been associated with increased mortality and poorer clinical outcomes.21,22 In addition, prolonged broad-spectrum antibiotic exposure is associated with increased selection pressure and the emergence of antimicrobial resistance (AMR). Recent global estimates suggest that bacterial AMR was directly responsible for 1.27 million deaths and associated with 4.95 million deaths in 2019.23 Reducing the time to effective antibiotic administration is therefore critical. Despite advances in automation and rapid identification, progress in the development of rapid AST platforms has been slower. Two main diagnostic strategies, phenotypic and genotypic AST, have been developed to guide antimicrobial therapy, each with distinct advantages and limitations (Fig. 2). This review summarises commercial and emerging AST technologies for bloodstream infections, describes their underlying principles (Fig. 3), and highlights key advantages, limitations, and implementation challenges.
![]() | ||
| Fig. 2 Key characteristics of phenotypic and genotypic antimicrobial susceptibility testing methods. S: Susceptible; I: Susceptible, Increased Exposure; R: Resistant. | ||
![]() | ||
| Fig. 4 PRISMA-style flow diagram of the search and study selection (adapted from Page et al.24). | ||
Diagnostic technologies were categorised into two groups: Commercial Platforms, defined as technologies that have received regulatory clearance or authorisation through the U.S. Food and Drug Administration or a CE mark for use in the European Economic Area; and Emerging Technologies, which include platforms currently in development or under clinical evaluation but not yet cleared for clinical use, proof-of-concept studies, and literature-described methodologies with preliminary validation.
Studies were excluded if they: addressed pathogen detection and identification only; were not designed as a rapid AST workflow (defined here as Time to Result (TTR) ≤8 hours from a positive blood culture); described protocols developed by clinical laboratories; evaluated fluorogenic probes as standalone assays; or discussed platforms that have been discontinued or withdrawn from the market. Literature articles evaluating lateral flow tests, addressing fungal or mycobacterial AST, or discussing AST or resistance marker detection only as a potential future extension were also excluded from this review.
| Platform | Company | Principle | Average TTR | Gram-positive panel | Ref. |
|---|---|---|---|---|---|
| Platforms providing Gram-positive testing are indicated in the table. NA: not available. | |||||
| FASTinov AST | FASTinov | Flow cytometry coupled with fluorescent probes to detect bacterial cell damage | 2 h | ✓ | 25 |
| QuickMIC | Gradientech | Real-time imaging of bacteria exposed to a linear antibiotic gradient | 3 h | NA | 26 |
| LifeScale | Affinity biosensors | Microfluidic sensor and mechanical resonator measure the mass of individual microbes | 4.5 h | NA | 27 |
| ALFREDAST 60 | Alifax | Broth microdilution coupled with light-scattering technology | 5 h | ✓ | 28 |
| Selux NGP system | Selux diagnostics | Fluorescence-based viability and surface area growth assay | 5.5 h | ✓ | 29 |
| VITEK REVEAL | bioMérieux | Broth microdilution coupled with volatile organic compounds detection sensors | 5.5 h | NA | 30 |
| dRAST | QuantaMatrix | Microbial growth monitored via time-lapse microscopy | 6 h | ✓ | 31 |
| ASTar | Q-linea | Broth microdilution coupled with time-lapse microscopy | 6 h | NA | 32 |
FASTinov (FASTinov SA) and QuickMIC (Gradientech AB) provide results within 2 to 3 hours, making them promising tools for early sepsis intervention. However, their integration into the laboratory workflow presents practical limitations, due to the need for labour-intensive sample pre-processing (FASTinov) and limited organism–antibiotic panel coverage (QuickMIC). Other systems, including dRAST (Quantamatrix Inc.) and ASTar (Q-linea AB) offer a more balanced solution, with an average TTR of 5–6 hours and comprehensive organism–antibiotic coverage.
Accelerate Pheno (Accelerate Diagnostics, Inc.) was the only integrated system that provided both identification and AST results for a broad panel of organisms and antibiotics, reducing the overall time to results by approximately 40 h compared to traditional automated systems and culture-based methods.33,34 However, the platform has recently been withdrawn from the market following the manufacturer's liquidation.35
| Main method | Reference | Principle | Clinical sample typea | Average TTRb | Gram-positive | Gram-negative | Category agreementc | Maturity staged |
|---|---|---|---|---|---|---|---|---|
| a Clinical sample type: intended clinical sample matrix for application. WB: whole blood. PBC: positive blood culture.b Average time to results (TTR): time to results reported from the starting sample type used in the study.c Category agreement (CA): agreement with reference methods (broth microdilution or disk diffusion) or conventional automated systems. Only studies evaluating CA on the intended clinical sample type were included.n: indicates the total number of organisms–antibiotic combinations. ND: no data.d Maturity stage: validation stage reflects readiness for application in bloodstream infection workflows. Classification was guided by the intended clinical use of each technology. Studies targeting direct-from-blood detection were considered analytically validated when evaluated in whole blood at clinically relevant bacterial concentrations (≈1–100 CFU mL−1) and supported by quantitative performance metrics against a reference method. Studies intended for positive blood culture workflows but evaluated only using isolates grown in culture media, spiked whole blood samples, or lacking quantitative performance metrics, were classified as proof-of-concept. | ||||||||
| Spectroscopy-based technologies | Dekter et al., (2017)36 | Classical Raman spectroscopy measurements of bacterial metabolic activity. | PBC | 5 h | ✓ | ✓ | n = 13 | Proof-of-concept |
| 100% | ||||||||
| Zhu et al., (2022)37 | Raman microspectroscopy measurements of D2O metabolism. | PBC | 3 h | ✓ | ✓ | n = 86 | Analytical validation | |
| 93% | ||||||||
| Zhang et al., (2020)38 | Stimulated Raman scattering measurements of D2O metabolism. | PBC | 2.5 h | ✓ | ✓ | ND | Proof-of-concept | |
| Han et al., (2023)39 | Surface-Enhanced Raman Scattering measurements of bacterial metabolic activity. | PBC | 5 h | ✓ | ✓ | n = 401 | Analytical validation | |
| 96% | ||||||||
| Light-scattering technologies | Idelevich et al., (2018)40 | Forward-angle laser light-scattering and OD measurements of bacterial growth. | PBC | 5 h | ND | ✓ | n = 19 | Analytical validation |
| ≥90% | ||||||||
| Falconer et al., (2024)41 | Scattered-light integrated collection (SLIC) measurements of bacterial growth. | PBC | 2 h | ND | ✓ | n = 505 | Clinical validation | |
| 95.5% | ||||||||
| Cytometry-based technologies | Filbrun et al., (2022)42 | Sucrose separation of bacteria coupled with label-free flow cytometry. | PBC | 5–10 h | ✓ | ✓ | n = 49 | Analytical validation |
| 90% | ||||||||
| Edgar et al., (2021)43 | Acoustic signals generated by dye-labelled phage-bound bacteria measured via photoacoustic cytometry | PBC | 4 h | ✓ | ND | ND | Proof-of-concept | |
| Hornsey et al., (2026)44 | Electrical and morphological properties of bacteria measured via microfluidic impedance cytometry. | PBC | 4 h | ND | ✓ | n = 4,947 | Clinical validation | |
| 97.7% | ||||||||
| Microscopy-based technologies | Zhu et al., (2023)45 | Integrated microfluidic chip combining separator gel and AST chambers for optical imaging. | PBC | 3.5 h | ND | ✓ | n = 60 | Analytical validation |
| 93.3% | ||||||||
| Forsyth et al., (2021)46 | Dextran separation of bacteria coupled with optical imaging of single cells trapped in a microfluidic device. | WB | 2 h | ND | ✓ | ND | Proof-of-concept | |
| Kim et al., (2024)47 | Bacterial recovery via functionalised magnetic particles; multiplex DNA hybridisation for ID; optical imaging of bacteria immobilised in agarose matrix. | WB | 13 h | ✓ | ✓ | n = 98 | Analytical validation | |
| 94.9% | ||||||||
| Biosensor-based technologies | Sturm et al., (2024)48 | Metabolic activity-induced oscillations measured via nanomechanical sensors. | PBC | 4 h | ND | ✓ | n = 85 | Analytical validation |
| 97.6% | ||||||||
| Xu & Ren, (2023)49 | Wireless LC sensor detection of bacterial growth-induced permittivity changes. | PBC | 30 min | ✓ | ✓ | ND | Proof-of-concept | |
| Shi et al., (2018)50 | Immuno-electrochemical detection of bacterial growth using enzyme-labelled antibody sandwich complexes. | WB | 3.5 h | ND | ✓ | ND | Proof-of-concept | |
| Metabolic markers quantification | Matsui et al., (2019)51 | Intracellular ATP levels measured via bioluminescence as a marker of metabolic activity. | PBC | 6 h | ✓ | ✓ | ND | Proof-of-concept |
| Rydzak et al., (2022)52 | Liquid chromatography-mass spectrometry quantification of multiple metabolic markers as a measure of bacterial growth. | PBC | 6 h | ✓ | ✓ | ND | Proof-of-concept | |
| Molecular markers quantification | Maxson et al., (2018)53 | Quantitative PCR measurements of 16S rRNA as a marker of bacterial growth. | PBC | 4 h | ✓ | ✓ | n = 432 | Analytical validation |
| 96.3% | ||||||||
| Young et al., (2024)54 | Multiplexed RNA hybridisation-based quantification of antibiotic-induced gene expression signatures | PBC | 4 h | ND | ✓ | n = 387 | Clinical validation | |
| 95% | ||||||||
Raman spectroscopy measures the inelastic scattering of photons, generating spectra that reflect certain kinds of vibrational modes of chemical bonds (functional groups) within a sample. While Raman spectroscopy can provide rapid AST directly from positive blood culture bottles within 5 hours,36 its low sensitivity and prolonged spectral acquisition limit throughput. Enhanced techniques, like Raman Microspectroscopy, Stimulated Raman Spectroscopy (SRS) and Surface-Enhanced Raman Scattering (SERS) have been investigated in an effort to address these limitations. Raman Microspectroscopy combines classical Raman measurements with microscopy to improve laser focusing and signal collection, with typical interrogation areas of ∼1 × 1 µm (the typical dimensions of an individual bacterium range from 1 to 2 μm, and these weigh ∼1 pg, and are therefore well within Raman spectroscopy's single cell analysis capabilities). Zhu et al.37 used Raman microspectroscopy coupled with the metabolic incorporation of deuterium oxide (D2O, heavy water) to determine the antimicrobial susceptibility of bacteria, and developed an automated system integrating sample preparation with spectra acquisition and analysis. In this approach, D2O served as a marker of metabolic activity, enabling detection of newly synthesised biomolecules through the incorporation of deuterium into C–D bonds. The resulting platform, named CAST-R (Clinical Antimicrobial Susceptibility Test Ramanometry), reduced the labour-intensive sample preparation, delivered results within 3 hours from PBC, and achieved a 93% CA when compared to broth microdilution. Similarly, Zhang et al.38 used D2O-probed SRS, a technique in which two synchronised laser beams are used to obtain a stronger Raman signal, to measure the MIC of common clinical isolates in 2.5 hours. The method was tested on spiked whole blood, demonstrating feasibility in this matrix, but validation on positive blood cultures remains to be demonstrated.
Whereas Raman Microspectroscopy and SRS improve signal collection through optical approaches, SERS addresses the sensitivity limitation by amplifying Raman signals using plasmonic nanostructured metallic surfaces.57 Incident light excites surface plasmons on these substrates, generating strong local electromagnetic fields that enhance Raman signals from nearby analytes. SERS allows the generation of AST results within 5 hours from a PBC and achieved high agreement (>95%) compared with conventional automated systems.39 Moreover, coupling this technique with microfluidics or deep learning analysis has been reported to improve accuracy, reproducibility, and turnaround times.58,59
More recently, scattered-light integrated collection (SLIC) was developed as an alternative light-scattering technique. SLIC employs an integrated detection space and a phase lock signal to collect and quantify the total scattering output while disregarding absorbed light. The method demonstrated high sensitivity, enabling the detection of bacterial concentrations as low as 100 CFU mL−1.62 A proof-of-concept study evaluated the ability of SLIC to perform rapid AST on Gram-negative bacteria isolated from PBC and achieved a 95.5% category agreement (CA) compared to VITEK 2, with minimal sample processing (dilution only) and an average TTR of 2 hours.41 While the authors reported a high rate of major and very major errors (4.1% and 7.1% respectively), the discrepancy may partly be attributable to the use of Tryptic Soy broth in place of the standard Mueller Hinton Broth (MHB) in their validation study.
In flow cytometry, heterogeneous cell populations are quantified and sorted based on light scattering and fluorescence signals as they flow through a laser-based detection system. Similarly to spectroscopy-based techniques, the effective implementation of flow-cytometry for rapid AST is dependent on the preparation of a pure bacterial suspension directly from blood cultures, as residual blood components can significantly interfere with signal detection.
The commercial platform FASTinov (Portugal) employs fluorescence flow-cytometry to determine the susceptibility profile of a broad range of Gram-positive and Gram-negative bacteria in approximately 2 hours, and achieved >95% agreement when compared to disk diffusion.66 However, fluorescent dye uptake can vary in antibiotic-stressed bacteria. To address this, Filbrun et al.42 developed a label-free approach, which included bacterial separation via centrifugation through a sucrose cushion, and MIC determination based on both count rates and morphological changes. The method produced MIC values rather than susceptibility categories, but TTR was significantly longer (≈5 to 10 hours depending on the species), and accuracy was lower (≈73% vs. broth microdilution), though category agreement remained close to 90%.
Alternative cytometry methods have been explored to overcome the limitations imposed by optical scatter and uptake of fluorescent dyes in complex biological specimens. In photoacoustic flow cytometry, bacteria bound to fluorescently labelled bacteriophages pass through a pulsed laser light, where the dye absorbs energy and generates a transient acoustic wave via thermoelastic expansion. This method was used to detect and differentiate daptomycin-resistant from-sensitive S. aureus strains in under 4 hours and identified heterogenous resistance within clonal populations.43 While the approach could be applied to other bacterial species and multiple antibiotics, its reliance on fluorescently labelled phages adds complexity and cost.
Unlike optical or photoacoustic methods, impedance cytometry offers a label-free assessment of antimicrobial susceptibility. By measuring the electrical properties of cells (resistance and capacitance) as they flow through a multi-frequency electric field, impedance cytometry provides information on cell size, morphology and membrane properties. This principle underlies the iFAST (impedance-based Fast Antimicrobial Susceptibility Test) platform, which uses microfluidic single-cell impedance measurements to detect the phenotypic responses of bacteria to antibiotics in 3 to 5 hours.67,68 Preliminary multi-site clinical evaluation data suggest that this approach may achieve high agreement with reference methods when applied to bloodstream infection isolates.44
Effective microscopy analysis relies on stable bacteria immobilisation and reduced background interference. Although separation of bacteria from blood components improves signal clarity, a pure bacterial pellet is not always required, as computational methods can distinguish bacteria from complex matrices. However, this is dependent on the platform: a higher degree of bacterial separation may be necessary for systems incorporating microfluidic confinement, such as single-cell trapping, where residual blood components can interfere with cell capture and imaging. For instance, Zhu et al.45 developed a bacteria separation integrated AST (BSI-AST) chip enabling rapid phenotypic AST directly from PBC in 3.5 hours. The microfluidic platform incorporated a separator gel for bacterial extraction and multiple antibiotic-preloaded downstream chambers, thus providing integrated sample preparation. Centrifugation and vacuum-assisted fluid handling were used to concentrate and distribute cells for optical imaging, steps which could potentially be automated. However, the method was not applicable to Gram-positive bacteria, and the impact of bacterial motility on imaging performance was not addressed.
Similarly, a proof-of-concept study combined bacterial separation via dextran sedimentation with microfluidic single-cell trapping, allowing individual cells to be spatially confined within narrow channels and imaged under antibiotic exposure. The workflow was completed directly on whole blood within 2 hours, significantly shortening the time to results.46 However, MIC determination was limited to E. coli with a single antibiotic (ampicillin), and efficient bacterial recovery was demonstrated only at relatively high concentrations (10–100 CFU mL−1), despite evidence that pathogen concentration in BSI commonly fall within the range of 0.01–10 CFU mL−1.69
More recently, a culture-independent platform, named uRAST (Ultra-Rapid AST), was announced by QuantaMatrix (based in South Korea).70 The system employs magnetic particles coated with synthetic beta-2-glycoprotein I (sβ2GPI) peptides, short fragments of an innate immune response protein capable of binding to common pathogen-associated molecular patterns, to recover bacteria directly from whole blood at clinically relevant concentrations (≈1–10 CFU mL−1). A multiplex DNA hybridisation assay, termed quick-mapping ID (QmapID), is subsequently used for pathogen identification, while phenotypic AST is performed on bacteria immobilised within an agarose matrix. This study demonstrated clinical feasibility for pathogen identification directly from blood and reported an average TTR of 13 hours from initial blood processing to susceptibility results. However, although the QmapID assay was evaluated using patient samples, AST performance metrics were derived retrospectively from spiked blood samples, showing 94.9% CA with the broth microdilution method.47 Prospective clinical AST validation remains therefore to be demonstrated.
Nanomechanical sensors include a range of mechanical transducers, including cantilever beams and other resonator geometries. These sensors can be functionalised to promote bacterial attachment and operated in either static mode, where surface deflection is used to measure the oscillatory nanomotion of metabolically active bacteria (growth-independent AST), or dynamic mode, where resonant-frequency shifts are monitored to quantify mass changes associated with bacterial growth (growth-dependent AST).71 In practice, these systems require the preparation of a sufficiently clean and standardised bacterial suspension, typically achieved through selective blood cells lysis or centrifugation followed by dilution, as residual blood components and uncontrolled inoculum levels can interfere with sensor performance.
One example of a dynamic, growth-dependent approach is the commercial LifeScale platform (Affinity Biosensors), which uses microchannel resonators to measure the mass and number of individual cells, and provide MIC results within 4–5 hours.27 However, growth-independent nanomechanical sensing has the potential to provide even faster results. In this context, Resistell (Switzerland) is developing Phenotech Multistar, a system that combines cantilever-based nanomotion detection with machine learning signal-analysis to detect and classify different bacterial susceptibility phenotypes in 2 to 4 hours.48 Whilst initial results are positive, the current scope of nanomechanical sensing AST remains narrow: studies are primarily focused on Enterobacterales, whereas certain Gram-positive organisms present additional challenges due to their tendency to form aggregates in blood, considerably complicating surface attachment and signal interpretation. Measurements are also sensitive to environmental factors, such as temperature, vibration, and fluid viscosity, and signal interpretation often requires advanced data analysis.
The passive electrical properties of bacteria differ substantially from those of aqueous media.72–75 Inductor-capacitor (LC) sensors are an alternative class of biosensors that use electrical resonance to detect changes in their environment. As bacteria grow and respond to antibiotics, their intrinsic dielectric properties alter the effective capacitance of the sensor, producing a measurable shift in resonance frequency.
In a recent study, wireless LC sensors were placed in bacterial cultures with or without antibiotics to produce MIC results in as little as 30 minutes.49 The technology did not require surface functionalisation and can be readily integrated into standard multi-well formats. However, measurements were performed in low-salt Lysogen Broth rather than conventional MHB to minimise background conductivity and optimise sensitivity, which may complicate direct translation to guideline-compliant AST conditions.
Beyond mechanical and resonance-based biosensors, a recent study developed an immuno-electrochemical sensing platform capable of detecting fewer than 10 CFU mL−1 directly from blood in under 4 hours. The platform uses a working electrode functionalised with species-specific antibodies and enzyme-labelled antibodies. Formation of the immune sandwich complex positions the redox enzyme within the electric field at the electrode interface, generating a measurable electrochemical signal that increases with bacterial growth.50 Notably, the assay operates directly on unprocessed whole blood, simplifying the workflow and reducing sample preparation. However, the use of small sample volumes (∼200 µL) at ultra-low bacterial concentrations means that bacterial cells may not always be present in the sample, a limitation that is not fully addressed.
Adenosine triphosphate (ATP) bioluminescence has been proposed as a possible marker for the detection of antibiotic resistance, due to the positive correlation between bacterial counts and ATP level.76 A recent study investigated a rapid ATP bioluminescence-based AST workflow performed on positive blood cultures to detect levofloxacin resistance. The protocol involved centrifugation through a serum separator gel to remove blood components, followed by sample dilution and treatment to reduce extracellular ATP prior to antibiotic exposure. Intracellular ATP was then released and converted into a luminescent signal via a luciferase-based reaction, quantified using a luminometer. When compared with broth microdilution, ATP-derived MICs showed only partial concordance, with several samples showing higher MIC values. The authors attributed this discrepancy to antibiotic-induced bacterial filamentation, which increases cellular ATP content despite inhibited cell division.51
Overall, these findings indicate that reliance on a single metabolic maker may limit the reliability of the method for routine clinical implementation. In contrast, multi-marker strategies may improve the robustness of metabolic phenotypic susceptibility testing.
One such approach was described by Rydzak et al.,52 who identified a panel of biomarkers capable of differentiating seven of the most common pathogens causing BSI, along with a set of metabolic flux markers that can be measured to infer antimicrobial susceptibility. Metabolic markers were measured via Liquid Chromatography-Mass Spectrometry, and the assay achieved a 95.2% agreement compared with the VITEK2 platform. While this strategy shows strong potential, advances in this field should prioritize reducing analytical complexity to enable clinical scalability.
The 16S ribosomal RNA (rRNA), a component of the small subunit (30S) of bacterial ribosomes, is a common target in PCR diagnostic assays for the detection and identification of bacterial species, particularly in culture-negative samples.77–79 As rRNA is rapidly produced in metabolically active bacteria, changes in its concentration following antibiotic exposure can serve as an early indicator of antimicrobial susceptibility. Following this principle, Maxson et al.53 developed the RAMAST (RApid Molecular Antibiotic Susceptibility Testing) assay, which used quantitative PCR (qPCR) to measure changes in 16S rRNA levels. The assay achieved 96.3% CA with broth microdilution and produced results in under 4 hours. However, the authors reported relatively high major error and very major error rates, which were attributed to operator errors, sample processing issues, or incorrect qPCR readings. Variations in the culture media used may also have contributed to the observed discrepancies.
Rather than relying on a single target, the analysis of antibiotic-induced transcriptional signatures across multiple genes can be employed to assess antimicrobial susceptibility. The GoPhAST-R (Genotypic and Phenotypic AST through RNA detection) platform represents an example of this strategy. Following blood culture positivity, bacterial cells are diluted, exposed to antibiotics, and then lysed. RNA transcripts are quantified using a multiplexed hybridisation-based assay (NanoString), enabling the quantification of selected mRNA targets without nucleic acid amplification. Susceptibility is determined by comparing the transcriptional response signatures between treated and untreated samples, and key resistance genes are also reported.80 In a recent clinical study including Gram-negative organisms, the platform reported 95% CA compared to conventional automated systems. For most antibiotics, transcriptional changes were detectable within 4 hours, although longer exposure (up to 16 hours) was required for slower-acting drugs or specific organism–antibiotic combinations.54
| Platform | Panel/assay | Company | Principle | Average TTR | Regulatory status | Ref. |
|---|---|---|---|---|---|---|
| Biofire film array | BCID2 | bioMérieux | Nested multiplex PCR with amplicon-specific DNA melt analysis | 1 h | CE-IVD | 81 |
| FDA 510(k) | ||||||
| Verigene | BC-GP | DiaSorin molecular | Multiplex microarray-based PCR with gold nanoparticles-labelled probes and silver-mediated signal amplification | 2.5 h | CE-IVD | 82 and 83 |
| BC GN | FDA 510(k) | |||||
| ePlex | BCID-GP | Roche diagnostics | Multiplex microarray-based PCR with ferrocene-labelled probes and electrochemical signal detection | 1.5 h | CE-IVD | 84 |
| BCID-GN | FDA 510(k) | |||||
| Unyvero | BCU | Curetis | Multiplex PCR with amplicon detection via hybridisation on microarray probe membranes | 5 h | CE-IVD | 85 and 86 |
| Molecular mouse | Sepsis panel | Alifax | Multiplex real-time PCR | 1 h | CE-IVD | 87 |
| eazyplex | BloodScreen GP | Amplex diagnostics | Loop-mediated isothermal amplification with real-time fluorescence detection | 20 min | CE-IVD | 88 |
| BloodScreen GN | ||||||
| iC-System | iC-GPC assay | iCubate | Amplicon-rescue multiplex PCR with microarray hybridisation | 4 h | CE-IVD | 89 |
| iC-GN assay | FDA 510(k) |
Genotypic platforms offer short turnaround times and automated, standardised workflows which can be easily integrated into routine clinical practice. Nevertheless, they cannot replace phenotypic susceptibility testing, as the presence of known resistance genes does not necessarily indicate phenotypic resistance, and vice versa. Phenotypic expression may be modulated by the bacterial strain's genetic background and environmental conditions,90 and targeted panels may fail to detect resistance mediated by mechanisms not captured by the assay, including porin loss, efflux upregulation, and target-site alterations. Despite these limitations, evidence suggests that these tests can improve patient outcomes by reducing time to effective therapy and hospital length of stay, particularly when implemented together with antibiotic stewardship programmes.91,92
| Platform/test | Company | Principle | Clinical sample typea | Average TTRb | Reference |
|---|---|---|---|---|---|
| a Clinical sample type: intended clinical sample matrix for application.PBC: positive blood culture.b Average time to results (TTR): time to results reported from the starting sample type used in the study. NA: Not Available. | |||||
| QIAstat-Dx | Qiagen | Multiplex real-time PCR with fluorescence-based detection of amplification signals | PBC | NA | Ng et al., (2024)98 |
| XGEN multi sepsis flow chip | Mobius life science | Multiplex microarray-based PCR with biotinylated primers and colorimetric immunoenzymatic signal detection | PBC | 3 h | Maluf et al., (2023)99 |
| Omix-AMP | Omix labs | Loop-mediated isothermal amplification with real-time fluorescence detection | PBC | 4 h | Maheshwarappa et al., (2021)100 |
| Reference | Principle | Clinical sample typea | Average TTRb | Maturity stagec |
|---|---|---|---|---|
| a Clinical sample type: intended clinical sample matrix for application.WB: whole blood.b Average time to results (TTR): time to results reported from the starting sample type used in the study.c Maturity stage: validation stage reflects readiness for application in bloodstream infection workflows.d Clinical validation status based on information reported on the developer website. | ||||
| Deepull diagnostics – UIICORE101 | Multiplex real-time PCR on total microbial DNA extracted from whole blood | WB | 1 h | Clinical validationd |
| Hoj et al., (2021)94 | Bacterial separation via a spinning hollow disk coupled with Pentaplex PCR | WB | 2 h | Analytical validation |
| Ganguli et al., (2022)95 | Biphasic blood drying method coupled with loop-mediated isothermal amplification | WB | 2.5 h | Clinical validation |
| Abram et al., (2020)96 | Droplet digital PCR in a microfluidic format with fluorescence-based detection of amplification signals | WB | 1 h | Analytical validation |
While encouraging, the methods above targeted a very limited number of pathogens and resistance markers. As efforts shift toward multiplex detection, established technologies are being adapted into new assay platforms. For instance, droplet digital PCR (ddPCR) has been employed for whole blood testing in a high-throughput microfluidic format. In this approach, inhibitor-resistant PCR reagents are mixed with whole blood and partitioned into oil-in-water droplets via flow-focusing microfluidics. Following thermal cycling, the signal from fluorescent droplets containing the target gene is quantified, enabling results within 1 hour, with a reported detection limit of 10 CFU mL−1.96 However, a clinical validation study evaluating ddPCR performance against blood culture reported concordant results in 63.9% of the cases. Among discordant results, 12.9% were deemed presumptive false positives, whilst the remainder were classified as either probable or possible BSI cases.97
Direct-from-blood assays that omit bacterial separation inherently measure both DNA derived from intact bacterial cells and microbial cell-free DNA (cfDNA). Whilst total microbial DNA is being employed for the development of new molecular assays, such as Deepull Diagnostics’ UllCORE BSI test,101 the diagnostic value of cfDNA remains debated.102–104 Cell-free DNA does not necessarily indicate a viable bloodstream infection, as it may reflect background DNAemia, colonisation, or contamination. Nevertheless, it has been proposed as a complementary marker in culture-negative cases, in patients with mild or non-specific symptoms, or for the detection of polymicrobial infections and slow-growing organisms.105,106 Sequencing-based approaches using cfDNA or total microbial DNA have been widely explored as a diagnostic tool for early infection identification, and in some cases, detection of resistance genes.107–110 However, these methods present several limitations, including long turnaround times (>8 hours), complex workflows, high costs, and uncertain clinical impact.
A consistent theme emerging from this review is that sample preparation remains a key limitation across platforms. Most rapid technologies require the effective separation of bacteria from blood to reduce signal interference. Different strategies, including centrifugation, selective lysis, dilution, affinity capture, and density-based separation (e.g. sucrose cushioning or dextran sedimentation), present distinct trade-offs in sensitivity, turnaround time, automation potential, and workflow complexity. For instance, dilution-based workflows simplify processing but reduce the bacterial concentration, potentially increasing detection limits and time to results. As such, they are most useful in technologies requiring a positive blood culture, as well as in those which do not require a purified bacterial pellet. In contrast, some separation strategies can improve detection limits and provide a clean bacterial pellet but introduce complexity and additional handlings steps. The optimisation, standardisation, and automation of sample preparation workflows will therefore be critical to facilitate large-scale routine implementation.
Emerging platforms would also benefit from balancing speed with wider organism–antibiotic coverage. Given the high costs associated with these systems, platforms offering moderate turnaround times, but comprehensive coverage are likely to offer greater clinical utility than ultra rapid systems (TTR < 3 hours) with limited panels. While time to result is critical, particularly in light of data suggesting that each hour of delay in effective therapy is associated with a decrease in survival, speed alone is insufficient if the causative pathogen or resistance determinant falls outside the test's scope. Broader coverage, even at the expense of slightly longer turnaround times, may therefore translate into improved outcomes across a wider range of clinical scenarios.
Collectively, these findings suggest that the primary barrier to widespread adoption is not the ability to generate rapid signals, but the challenge of integrating streamlined sample processing with broad pathogen–antimicrobial coverage.
In recent years, direct-from-blood testing has become a central focus, particularly for molecular approaches. As detailed in the review, most phenotypic AST approaches require sufficient, and often substantial, numbers of viable organisms to generate robust signals, making broad direct-from-blood testing significantly more challenging. In contrast, molecular tests do not require organism preservation and overcome low pathogen concentrations through nucleic acid amplification and are therefore more likely to be integrated into fully automated, whole blood testing systems in the near future.
While molecular approaches cannot replace AST testing, rapid identification and resistance marker detection have the potential to support earlier therapeutic optimisation and reduce unnecessary broad-spectrum antibiotic use. In this context, although sequencing-based approaches are not yet routinely compatible with the rapid turnaround times required for early clinical decision-making, ongoing technological advances, including improvements in sequencing chemistry, library preparation, and bioinformatic pipelines, may allow them to achieve comparable clinical impact. When combined with established resistance databases such as the Comprehensive Antibiotic Resistance Database (CARD), sequencing-based strategies could expand resistance detection beyond predefined molecular panels, ultimately providing more comprehensive resistance profiling.
Comparative clinical trials and cost-effectiveness analyses will be essential to determine which technologies provide sustainable clinical benefits in terms of costs and patient outcomes. Ultimately, the success of next-generation diagnostics will depend not only on technical innovation but also on their capacity to easily integrate into existing clinical workflows and meaningfully influence therapeutic-decision making.