Open Access Article
John M. Branning† Jr.
ab,
Brianna M. Ruff
c,
Samuel C. Saccomano
c,
Avi I. Flamholz
d,
Dianne K. Newman
efg and
Kevin J. Cash
*ac
aQuantitative Biosciences and Engineering Program, Colorado School of Mines, Golden, CO, USA. E-mail: kcash@mines.edu
bThe MITRE Corporation, Bedford, Massachusetts, USA
cChemical and Biological Engineering Department, Colorado School of Mines, Golden, CO, USA
dLaboratory of Environmental Microbiology, The Rockefeller University, New York, NY 10065, USA
eDivision of Biology and Biological Engineering, California Institute of Technology, Pasadena, California, USA
fResnick Sustainability Institute, California Institute of Technology, Pasadena, California, USA
gDivision of Geological and Planetary Sciences, California Institute of Technology, Pasadena, California, USA
First published on 15th June 2026
We developed a fluorous-phase oxygen-sensitive nanosensor that mitigates quenching effects caused by redox-active microbial metabolites whose effective lipophilicity depends on local chemical conditions, notably pyocyanin. The design encapsulates the fluorinated near-infrared (NIR) oxygen-sensitive luminophore platinum(II) meso-tetra(pentafluorophenyl)porphine (PtTFPP) within a fluorous-phase matrix that restricts pyocyanin access to the dye relative to conventional non-fluorous polymer matrices, thereby reducing interference. Encapsulation within a fluorous polymer-based nanoparticle matrix maintains the dye at a constant loading in biological samples and avoids the need for complex synthetic approaches. The resulting fluorous-phase optical nanosensors exhibited consistent and reversible oxygen measurements across a wide concentration range, with pyocyanin-induced interference substantially attenuated relative to reference polymeric nanosensors. This design provides a framework for future research into fluorous nanosensing technologies and their application in diverse and complex environments.
Nanoparticle-based sensors (nanosensors) are an emerging technology designed to address many challenges associated with biological measurements.21–28 One class of nanosensors consists of a highly plasticized polymer matrix that encapsulates one or more luminescent indicator dyes while allowing molecular oxygen to permeate the sensor.26,29,30 Many potential indicators exhibit poor solubility in water, which can pose difficulties in biological applications.31 Encapsulating these indicators within a polymer matrix improves control over the sensor's response and facilitates their use across a diverse range of samples. Moreover, these nanoparticles maintain stability over extended periods, ranging from hours to days, and can be uniformly distributed throughout the sample to measure three-dimensional gradients when used with appropriate optical techniques.17 Notable applications of oxygen-sensitive nanosensors include measurements of intracellular oxygen in cancer cells,32–35 monitoring oxygen dynamics in plants,28,36 and oxygen detection in microfluidic systems37,38 and in vivo measurements within bacterial biofilms.18,39–42
However, recent studies have demonstrated that redox-active molecules secreted by bacteria can interfere with these sensors,43 complicating accurate oxygen measurements in systems where these molecules are present. Experiments with purified substances revealed that this quenching arises from interactions with redox-active small molecules such as phenazines and flavins, which are produced by a wide range of microbial species in a density-dependent manner and can reach high concentrations.44 Pyocyanin, a defining phenazine produced by Pseudomonas aeruginosa, has been identified as a particularly strong quencher in these sensors.43,45 Importantly, the extent to which pyocyanin can access and accumulate within hydrophobic sensor matrices depends on local chemical conditions, including redox state and pH,46 which modulate its effective lipophilicity. As a result, phenazines such as pyocyanin may partition into the hydrophobic polymeric core of conventional oxygen sensors, enabling proximity-dependent quenching of embedded indicator dyes and complicating data interpretation in chemically complex environments.
In optical nanosensors, the sensing phase typically consists of a polymer-plasticizer matrix that encapsulates the luminescent dyes, while a surfactant coating enables compatibility with aqueous environments (Fig. 1). The chemical composition of this matrix strongly influences sensor selectivity and dynamic range.47–51 Fluorinated compounds possess several distinctive physicochemical properties that are advantageous for nanosensor design.
Fluorous phases exhibit high oxygen solubility, enabling efficient oxygen transport to embedded dyes48,52–55 while their low affinity for polar and charged species limits the partitioning of many biological metabolites. This polarity mismatch reduces the accessibility of redox-active molecules, such as pyocyanin, to the sensing core, thereby mitigating metabolite-induced quenching. Fluorous polymers and surfactants also provide a favorable combination of permeability,56–58 selectivity,59–61 and chemical stability62–64 while remaining transparent through a wide ultraviolet, visible, and infrared spectral range, making them attractive candidates for nanosensor matrices.65,66
Fluorous materials have previously been used in ion-selective electrode sensors,48,67–70 fluorocarbon nanoemulsions for molecular imaging,71 and fluorous nanocomposite probes and sensors.65,72,73–75 Importantly, although elevated oxygen solubility supports efficient oxygen-dependent quenching, it does not inherently increase signal magnitude; rather, the primary advantage of fluorous phases lies in interference mitigation and environmental isolation.
Over the past decade, fluorinated variants of several dye and chromophore scaffolds, including coumarin, porphyrin, perylene, rhodamine, and BODIPY, have been developed.54,76 Fluorination of these structures often increases absorption coefficients and quantum yields, leading to advantageous properties such as red-shifted absorption maxima, enhanced brightness, and improved photostability.54 Platinum(II) meso-tetra(pentafluorophenyl)porphine (PtTFPP), a fluorinated variant of platinum(II) meso-tetraphenylporphine (PtTPP), is a luminescent dye that emits at λem = 650 nm when excited at λex = 405 nm and is reversibly quenched by molecular oxygen, thus making it well suited for optical measurements of oxygen in biological samples.77–79
We therefore hypothesized that incorporating fluorous polymers together with fluorinated surfactants would reduce the phase transfer of secreted bacterial metabolites into the nanosensor core, thereby minimizing their confounding effects on oxygen-sensitive optical nanosensors. In this design, the surfactant layer stabilizes the nanoparticle in aqueous environments, while the fluorous polymer matrix selectively promotes the diffusion of O2 into the sensing region, where it interacts with the fluorinated dye.
To prepare the nanosensors, 80 µL of PEG-750 (25 mg mL−1 in chloroform) was added to a 4-dram scintillation vial. The chloroform was evaporated under a gentle air stream, after which the PEG-750 residue was redissolved in 5 mL of PBS and sonicated with a probe-tip sonicator (Branson Ultrasonics, Brookfield, CT, USA) for 30 s at 20% amplitude. Subsequently, 125 µL of the optode solution was added, and the mixture was sonicated for another 3 minutes at 20% amplitude to form the nanosensors. The resulting nanosensor suspension was filtered through a 0.8 µm polyethersulfone membrane into a 5 mL glass vial and stored at room temperature protected from light.
To fabricate the fluorous nanosensor we first fabricated the fluorous optode formulation with the following sensor components. 15 milligrams of PVDF was weighed out into a 2 mL glass vial and combined with 33 µL of BEHS, followed by vortex mixing. 2.5 mg of PtTFPP was dissolved in 250 µL of THF and transferred to the PVDF slurry. The mixture was vortexed until the PVDF was completely dissolved. Subsequently, 375 µL of DCM was added to the vial containing the PVDF–dye mixture. The optode solution was then capped and stored at 4 °C until further use.
To prepare the nanosensors, 80 µL of PFPE-PEG (25 mg mL−1 in THF) was added to a 4-dram scintillation vial. The THF was evaporated under a gentle air stream, after which the PFPE-PEG residue was redissolved in 5 mL of PBS and sonicated with a probe-tip sonicator (Branson Ultrasonics, Brookfield, CT, USA) for 30 s at 20% amplitude. Subsequently, 125 µL of the optode solution was added, and the mixture was sonicated for another 3 minutes at 20% amplitude to form the nanosensors. The resulting nanosensor suspension was filtered through a 0.8 µm polyethersulfone membrane into a 5 mL glass vial and stored at room temperature protected from light.
Nitrogen and air flow rates were regulated using two mass flow controllers (Alicat Scientific, Tucson, AZ, USA). The gas streams were then mixed in a 50 mL stainless-steel gas mixing chamber before entering the cuvette.79 The gas mixture was bubbled directly into the cuvette through the septum seal using a 15-gauge needle attached to the gas line. A second needle was inserted through the septum to allow excess gas to vent from the cuvette during bubbling.
The total gas flow rate was maintained at 5 mL min−1, while the relative flow rates of nitrogen and air were adjusted to achieve oxygen concentration between 0% and 21% in the gas phase. Measurements were obtained at 0%, 5%, 10%, 15% and 21% O2 in the gas phase, which was equivalent to 0, 1.58, 3.17, 4.75 and 6.65 mg L−1, respectively, confirmed with microelectrode measurements (UniSense, Aarhus, DNK). The initial equilibration period at 0% O2 was 45 min, while subsequent equilibration steps were 45 min.
A calibration curve was generated by measuring the intensity at λem = 650 nm (PtTFPP emission) at each oxygen concentration. The data were analyzed using the Stern–Volmer relationship, expressed as a ratio of the intensity in the absence quencher (I0, 0 mg L−1 O2) to the intensity at a given concentration (I). The Stern–Volmer constant (KSV) was determined by linear regression of the Stern–Volmer plot, as described by eqn (1) in the Results section.
Nanosensor reversibility was evaluated by alternating between pure nitrogen (deoxygenation) and air (21% O2) to determine whether the luminescence response changed as a function cycling. Each bubbling step was maintained for 45 min, between measurements, and the experiment was repeated for 5 cycles.
The pH response was assessed by diluting the sensor suspension 1
:
1 with potassium buffer solutions prepared from phosphate dibasic and potassium phosphate monobasic adjusted to different pH values.
The quantum yield82 of PtTFPP within the fluorous nanosensor (Φ = 0.070 deoxygenated, Fig. S1) is within the range of values reported for PtTFPP-based nanosensors77,83–88 (Table S1). The lifetime of the fluorous nanosensors measured using the Rapid Lifetime Determination (RLD) method,89,90 τ0 = 63–69 μs, is approximately 1.8× longer than the reference nanosensors31,48,54,91–95 (Table S2). The radiative rate constant kR = Φ/τ0 = 0.0011 μs−1 and total non-radiative rate constant kNR = (1 − Φ)/τ0 = 0.0135 μs−1 were estimated using the midpoint τ0 value.96,97
In luminescence quenching-based methods, the Stern–Volmer equation (eqn (1)) describes the relationship between the quencher concentration (here [O2]) and signal intensity.
![]() | (1) |
We created Stern–Volmer plots (Fig. 3) by measuring the luminescence intensity of PtTFPP as a function of [O2]. The Stern–Volmer constant (KSV = 0.55 ± 0.014 L mg−1) was determined by linear regression of the Stern–Volmer plot, as described by eqn (1). This value quantifies the degree to which the nanosensor-embedded PtTFPP is quenched by O2, with larger values indicating stronger quenching. Given this KSV, luminescence intensity is roughly fourfold higher in anoxic conditions (0%) than air-levels of O2 (21%).
To monitor oxygen concentrations in biological systems, sensors must be reversible, i.e. able to recover luminescence after transient O2 exposure. To test reversibility of the nanosensors, we measured the linearity of the Stern–Volmer response during deoxygenation and reoxygenation by alternately bubbling the same sample with nitrogen and air mixtures. The oxygen concentration was varied in 5% increments from 0% to 21%, and then back from 21% to 0%, to assess consistency in both directions. The signal was measured over two complete cycles between oxygenated and deoxygenated conditions. No statistical difference in KSV values was observed between the forward and reverse cycles (Fig. S2A), demonstrating that the nanosensor response is both reproducible and reversible. This behavior suggests that the curvature observed in (Fig. 3 and Fig. S2) does not arise from irreversible or static quenching effects. Instead, residual nonlinear response (Fig. S3) may reflect measurements acquired under non-equilibrium conditions during oxygen transitions.
The Stern–Volmer calibration was also performed for the reference polymeric nanosensors (Fig. S2B), revealing that the nanosensor systems do not exhibit equivalent intrinsic oxygen sensitivity. The observed difference in KSV does not confound interpretation of the pyocyanin interference experiments because all measurements were performed under fully deoxygenated conditions generated using excess glucose and glucose oxidase. Under these conditions, oxygen-dependent quenching is absent and both nanosensor systems operate at their respective I0 baseline, such that subsequent signal attenuation reflects exclusively pyocyanin-induced quenching.
Beyond reversibility, practical deployment in biological environments requires robustness to physiochemical fluctuations such as pH variation. The fluorous nanosensors exhibited stable emission across the investigated pH range (3–9.5) under atmospheric oxygen conditions (Fig. S4). Although one-way ANOVA detected statistically significant differences in luminescence intensity across pH 3–9.5 (F = 10.28, p = 0.00019), the overall variation was small relative to the mean signal (∼4% coefficient of variation) and showed no monotonic pH-dependent trend, indicating that the nanosensors maintain functional emission stability across a physiologically relevant pH range.
In addition to environmental stability, resistance to redox-active metabolite interference represents a central challenge for oxygen sensing in microbial systems. We therefore evaluated the ability of fluorous nanosensors to resist pyocyanin quenching by measuring the luminescence response of the sensor to pyocyanin in a deoxygenated buffer generated by the addition of excess glucose and glucose oxidase (see Methods).
As shown in Fig. 4 fluorous nanosensors were less affected by pyocyanin quenching than earlier nanosensor designs lacking fluorous phases. The deoxygenated response of oxygen-sensitive nanosensors in the presence of 100 μM pyocyanin reveals significant differences: reference polymeric nanosensors showed an 80% reduction in deoxygenated luminescence, whereas fluorous nanosensors exhibited only a 46% reduction (Fig. 4A). The fluorous nanosensors outperformed the traditional nanosensors across all measured pyocyanin concentrations (Fig. 4B).
Lifetime performance of the nanosensors over a 10 day period shows that the reference nanosensors’ luminescence remains significantly quenched, whereas the fluorous nanosensors continue to exhibit relatively higher residual luminescence levels. Functional lifetime measurements (Fig. 5) indicate that fluorous-phase nanosensors confer durable protection against pyocyanin over 10 days, with only a 1.3% decrease in luminescence intensity compared to a 16.8% decrease observed for reference polymeric nanosensors (Fig. S5). This resilience results in more reliable oxygen measurements in environments where organic quenchers like pyocyanin are present, supporting the use of fluorous nanosensors for applications requiring long-term and stable oxygen sensing in complex biological systems such as bacterial biofilms.
Pyocyanin is a redox-active phenazine that participates in coupled electron/proton transfer reactions under physiological conditions and can undergo oxygen-coupled redox cycling in enzymatic and biological environments.44,98–100 In glucose and glucose oxidase enzymatic reactions, pyocyanin acts as an artificial electron acceptor,101,102 where oxidized pyocyanin is reduced enzymatically and subsequently re-oxidized by transferring electrons to oxygen. This cycling can sustain a population of reduced pyocyanin under conditions of active enzymatic turnover, particularly in low-oxygen or metabolically active environments.
The phosphorescent dye PtTFPP emits from a long-lived triplet excited state that is efficiently quenched by molecular oxygen through collisional interactions, but can also undergo additional nonradiative decay pathways in the presence of redox-active species.97 In conventional polymeric nanosensors, weakly polar molecules such as pyocyanin can partition into the hydrophobic polymer matrix, enabling proximity-dependent quenching of the embedded dye.43 Fluorous polymers, in contrast, are characterized by low polarizability and reduced solubility for many organic solutes. The dominant non-radiative decay in the fluorous formulation (Φ, kNR ≫ kR) is consistent with published values for PtTFPP embedded in nanoparticle matrices, confirming that the fluorous matrix does not enhance dye brightness relative to the reference formulation.
Consistent with these properties, our data show that pyocyanin-induced quenching is substantially attenuated in fluorous nanosensors, indicating reduced molecular partitioning into the sensing phase and confirming that mitigation of oxygen-independent interference is the principal functional advantage of the fluorous matrix in biologically complex environments. However, the fluorous matrix does not provide complete exclusion, and measurable quenching persists under conditions that sustain redox cycling.
Importantly, expanded formulation screening revealed that fluorination of the polymer matrix alone is insufficient to guarantee improved resistance to pyocyanin interference. Several fluorinated formulations (Fig. S12–S15 and Table S6) exhibited pyocyanin-induced quenching comparable to or greater than conventional PVC-based reference systems. Likewise, incorporation of a non-fluorinated luminophore into a conventional PVC matrix did not improve resistance to pyocyanin quenching (Table S6). Collectively, these observations indicate that nanosensor performance depends on coupled interactions among polymer composition, luminophore physicochemical properties, nanosensor internal organization, and fabrication methodology.
Taken together, our results support a mechanistic framework in which pyocyanin-mediated quenching is governed by coupled redox-state and partitioning effects influenced by nanosensor microstructure and dye localization within the particle interior. The redox state of pyocyanin modulates its effective lipophilicity, with the reduced form exhibiting greater affinity for hydrophobic environments.46 This increased lipophilicity enhances partitioning into or near the nanosensor matrix, enabling proximity-dependent interactions with the luminophore. Under conditions that support sustained redox cycling, such as enzymatic reactions or biological systems, a persistent population of reduced pyocyanin can be maintained, increasing its effective local concentration and strengthening quenching. In contrast, under purely abiotic, oxygenated conditions, pyocyanin remains predominantly oxidized and exhibits reduced partitioning into the sensor matrix, resulting in weaker and less reproducible quenching.
Consistent with this framework, oxidized pyocyanin under abiotic, oxygenated conditions produced minimal and irreproducible quenching (see Fig. S6–S9), whereas stronger and more reproducible quenching was observed in glucose–glucose oxidase systems (Fig. 4 and 5). Although pH can influence pyocyanin protonation45 and, together with redox state, alter its effective lipophilicity,46 the pH range examined here did not measurably suppress quenching in conventional polymeric nanosensors (see Fig. S8). Instead, the presence of sustained redox cycling appears to be the dominant factor governing interference under the conditions tested. These observations indicate that pyocyanin interference is highly context dependent and arises from the interplay between solution-phase redox processes and matrix-level partitioning behavior.
The discrepancy between the weak abiotic quenching observed in this study and stronger abiotic effects reported previously suggests that pyocyanin-mediated interference is highly sensitive to local chemical conditions that control its redox state and effective partitioning. Even nominally abiotic systems may differ in oxygen availability, reducing capacity, and protonation conditions, leading to variations in the fraction of reduced pyocyanin and, consequently, its accessibility to the nanosensor matrix. These differences likely account for the variability in reported quenching behavior across studies.
While the present results are consistent with redox-state–dependent partitioning, this study does not directly quantify pyocyanin uptake into the fluorous phase nor independently measure the redox-state distribution within the nanosensor microenvironment. Consequently, we cannot definitively distinguish whether residual quenching arises primarily from limited metabolite penetration into the fluorous matrix, interfacial interactions near the nanosensor surface, or redox-mediated processes in the surrounding solution. Resolving these contributions will require future studies incorporating direct permeability measurements and controlled modulation of pyocyanin redox state.
The immiscibility between fluorous and conventional organic phases presents practical challenges for nanosensor fabrication,103 as many optode components are traditionally processed in organic solvents. Consistent with this limitation, fabrication of fluorous-phase nanosensors using conventional organic solvents resulted in pronounced batch-to-batch variability in sensor response (Fig. S10A–C). Hierarchical analysis of the coefficient of variation (Tables S4, S5 and Fig. S10A–D) indicates that variability between independently prepared batches exceeds the variability observed within individual nanosensor solutions. This pattern suggests that differences arising during nanosensor preparation contribute more strongly to signal dispersion than measurement-level noise. Formulation-dependent differences in pyocyanin resistance further indicate that fabrication conditions influence not only reproducibility, but also nanosensor transport properties and accessibility of the embedded luminophore to redox-active interferents. Dynamic light scattering (DLS) measurements further revealed moderate polydispersity, with average hydrodynamic diameter of 212 nm with a relatively broad size distribution (Fig. S11), consistent with non-uniform nanosensor formation. Although intra-batch precision remains high, these findings highlight batch reproducibility as a key practical consideration for translational applications and emphasize the importance of batch-specific calibration in biological measurements.
Systematic variation of fluorous polymers, solvents, and oxygen-sensitive dyes revealed substantial formulation-dependent differences in pyocyanin interference behavior (see Fig. S12–S15), suggesting that the current system may be limited by fundamental transport or interfacial constraints rather than specific material combinations. Nevertheless, fluorous nanosensors demonstrated reduced susceptibility to metabolite interference while maintaining robust oxygen sensitivity, indicating their potential utility in biofilm-relevant chemical conditions.
Although fluorous materials do not fully eliminate pyocyanin-induced quenching, they provide a meaningful reduction in interference and are expected to improve measurement robustness in biofilm-relevant conditions. Future work should focus on decoupling redox-state–dependent partitioning from solution-phase quenching mechanisms, for example through controlled electrochemical modulation of pyocyanin redox state. From an engineering perspective, the development of fully fluorous-processable nanosensor components and surfactants may enable fabrication in fluorous solvents, improving structural uniformity and reducing batch variability. Additionally, distinguishing whether residual signal loss arises from matrix partitioning, interfacial accumulation, or direct redox quenching will guide the rational design of nanosensor systems with enhanced selectivity in metabolically active environments.
Our results further indicate that pyocyanin-mediated quenching is strongly dependent on chemical context, particularly conditions that sustain reduced, more lipophilic pyocyanin. As such, interference is expected to be amplified in metabolically active or redox-cycling environments, underscoring the importance of considering redox state when evaluating nanosensor performance. Accordingly, this work should be interpreted as demonstrating that fluorous-phase encapsulation represents a mechanistically grounded strategy for improving the resistance to metabolite interference, rather than a complete solution to this challenge in complex microbial systems.
Importantly, pyocyanin resistance emerged as a formulation-specific property governed by coupled effects of polymer chemistry, luminophore identity, nanosensor microstructure, and fabrication conditions. These findings define the current scope of applicability for fluorous encapsulation strategies and highlight the need for formulation-specific validation when extending this approach to other sensing systems.
This work contributes to the broader understanding of fluorous materials in optical nanosensing and establishes a framework for rational sensor design based on controlling molecular partitioning and access to the sensing domain. While further validation in fully biological and in situ systems is necessary, these results highlight the potential of fluorous encapsulation to improve robustness in biofilm-relevant environments. Future studies should quantify performance limits under dynamic biological conditions, evaluate long-term biofouling behavior, and determine the physicochemical criteria governing successful implementation of fluorous encapsulation across different fluorophores, matrices, and analyte systems.
Footnote |
| † The author's affiliation with The MITRE Corporation is for identification purposes only and is not intended to convey or imply MITRE's concurrence with, or support for, the positions, opinions, or viewpoints expressed by the author. Approved for Public Release, Distribution Unlimited. Public Release Case Number 24-3656. |