Feasibility study of superoxide chemical ionization mass spectrometry (O2 CIMS) for real-time gas-phase measurements of per- and polyfluoroalkyl substances (PFAS)

Yufan Hu a, Joji W. Shermanb, Michael J. Daverna, Zane A. Alsebaia, Yoonsub Kimb, Yue Zhangc, Chenyang Bid, Barbara J. Turpinb and Jason D. Surratt*ab
aDepartment of Chemistry, College of Arts and Sciences, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA 27514. E-mail: surratt@unc.edu; Tel: +919-966-0470; Fax: +919-966-7911
bDepartment of Environmental Sciences and Engineering, Gillings School of Global Public Health, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA 27599
cDepartment of Atmospheric Science, College of Arts and Sciences, Texas A&M University, College Station, Texas, USA 77843
dAerodyne Research Inc., Billerica, Massachusetts 01821, USA

Received 23rd December 2025 , Accepted 4th May 2026

First published on 5th May 2026


Abstract

Per- and polyfluoroalkyl substances (PFAS) are highly persistent pollutants with known adverse impacts on environmental and public health. Traditional gas-phase PFAS detection approaches often involve labor-intensive sample collection and preparation, while offering low temporal resolution. Alternatively, chemical ionization mass spectrometry (CIMS) allows for real-time airborne PFAS detection at sub-pptv sensitivity. While reagent ion generation often requires hazardous chemicals (e.g., nitric acid, methyl iodide (I), and acetic anhydride), superoxide (O2) CIMS provides a safer alternative and is better suited for mobile platforms where ventilation, space, and weight are constrained. O2 CIMS has five main reagent ions (i.e., O2, (H2O)O2, CO3, (CO2)O2, and CO2(H2O)O2) and low-background mass spectra above m/z 200. Thus, it is well suited to the relatively high molecular weights of airborne PFAS. Mass calibration was performed with a 5[thin space (1/6-em)]:[thin space (1/6-em)]1 fluorotelomer alcohol (FTOH) permeation tube. Ionization was found to occur mainly through deprotonation or adduct formation. Calibrations for fourteen environmentally- and industrially-relevant PFAS compounds are presented, including FTOHs, fluorotelomer diols, fluorinated sulfonamides, epoxides, and glycol ethers. While perfluoroalkyl carboxylic acids (PFCAs) were not detected, O2 CIMS offered higher sensitivity and lower detection/quantification limits than I CIMS for FTOHs; however, it remains a complementary PFAS measurement technique to I CIMS. Moreover, it yielded distinct fingerprint signals for FTOHs, confirming compound identification. This study demonstrates the utility of O2 CIMS for real-time airborne PFAS analysis in commonly encountered environments by capturing 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH gaseous emissions from fast-food packaging at room temperature, underscoring its strong promise for future development and applications.


Introduction

Per- and polyfluoroalkyl substances (PFAS) are a class of synthetic chemicals that have been widely used in industry and consumer products since the mid-20th century.1 Their strong carbon–fluorine bonds give them unique water- and grease-resistant properties, making them highly valued for their durability and versatility.2–4 As a result, PFAS have been incorporated into a broad range of products, from everyday items to high-tech applications, including non-stick cookware, food packaging, water-repellent clothing, plumbing tape, refrigerants, firefighting foams, firefighter turnout gear, and semiconductor manufacturing.3,5–14 However, while these chemicals offer significant functional benefits, they are highly mobile and persistent in the environment.3,15 PFAS do not readily break down or biodegrade, leading to their accumulation in both living organisms and the environment over time.16–18 This persistence has earned them the nickname “forever chemicals”. Their widespread use and resistance to degradation have resulted in global contamination of air, soil, and water,19 raising serious concerns about their environmental and ecological impacts, including potential disruption of ecosystems and entry into the food chain.7,20,21 Studies have linked PFAS exposure to adverse human health effects, including suppressed immune response, detrimental lipid profiles, neurobehavioral defects, and certain cancers.22–25 Numerous research efforts are underway to address PFAS contamination through remediation strategies, including capture technologies and degradation methods.26–31

Current PFAS detection efforts are primarily focused on water and soil, not only because analytical techniques are relatively well-established, but also due to recent regulatory policies targeting water contamination,28,32–37 as well as air measurements of PFAS being challenging.38,39 However, investigating airborne PFAS is also important, as it helps explain the potential for atmospheric transport of PFAS to remote regions on Earth (e.g., Arctic),40,41 sheds light on their partitioning behavior in multiphase environments, and ultimately helps to resolve the sources of PFAS contamination in downwind private well water.42–47 PFAS in air presents unique challenges, particularly in effectively collecting and storing these dilute and often volatile samples.48 Traditional methods rely on sorbent materials or filters to collect samples over periods ranging from several hours to weeks, allowing PFAS concentrations to accumulate above instrumental detection limits.48–50 After collection, samples are typically spiked with internal standards to correct for extraction recovery. Extracted samples are then concentrated and reconstituted for analysis by gas chromatography (GC) or liquid chromatography (LC) coupled with mass spectrometry (MS).48,50–52 Though GC-/LC-MS provides two dimensions of compound confirmation (i.e., retention time and mass), these offline methods require the use of expensive mass-labelled internal standards of PFAS.48,50–52 While integrated samples have some advantages, their limitations also include significant time involved in sample processing and associated uncertainties. Moreover, the method only yields time-averaged PFAS concentrations over the sampling period with limited temporal resolution.

Chemical ionization mass spectrometry (CIMS) enables real-time gas-phase PFAS analysis using a soft ionization method.53 Originally developed in the 1960s and 1970s, CIMS has significantly advanced in recent years.54 Coupling chemical ionization (CI) with a time-of-flight (TOF) mass analyzer now allows for high-resolution MS detection and precise identification of complex compounds. Modern CIMS designs offer high sensitivity, enabling detection limits in the sub-pptv range, and are often compact and field-deployable, suitable for use on aircraft and mobile laboratories.55,56 To best capture a wide array of compounds, CIMS employs different reagent ions that select for particular chemical properties, such as certain levels of volatility and oxidation states of carbon.54,56 Reagent ions used in proton transfer reaction (PTR) systems, like H3O+ and NO+, are standard for real-time monitoring of volatile organic compounds (VOCs).57,58 Now it also has been used for volatile PFAS measurements.59 Other reagent ions, such as I and NO3, are commonly used for detecting oxygenated organics and highly oxidized organic gases.54–56,60

Iodide CIMS has previously been calibrated and applied to quantify oxygenated PFAS in the gas phase, including PFCAs from aqueous film-forming foam combustion,61 and capturing 6[thin space (1/6-em)]:[thin space (1/6-em)]2 fluorotelomer alcohol (FTOH) emissions from rain jackets, floor wax applications, and microwaving popcorn.62–65 However, there are notable limitations on using I CIMS. First of all, PFAS has a wide range of structures, but I is selective,54,66 so it does not work for all species. Moreover, when the high molecular weight of PFAS adducts with I, which is a large ion, the high adduct weight makes it hard to be revolved in a TOFMS. Lastly, the reagent ions are generated by introducing pure nitrogen gas through a methyl iodide permeation tube,62,67,68 which requires a robust ventilation system and secure exhaust handling for safe operation. These requirements limit their suitability for mobile laboratories, where space, weight, and ventilation capacity are constrained. Therefore, a safer alternative reagent ion that allows for detection of PFAS classes not previously detected by other CIMS methods and does not require the use of toxic chemicals (e.g., methyl iodide) is needed. Superoxide (O2), which can be generated more safely, has previously been used in detecting nitrates, sulfates, and highly oxygenated organic compounds, showing a non-selective but general reagent chemistry property.69,70 Based on these results, superoxide may also show strong potential for real-time PFAS measurements in air. As a result, in this study we systematically examine the potential for O2 CIMS to make real-time quantitative measurements of airborne PFAS.

Materials and methods

Section S1 of SI provides suggested safety precautions when working with PFAS.

O2 ionization

For the setup (Fig. 1), considering previous methods,70–72 compressed house air first passes through a Drierite Gas Purifier with 5A molecular sieves (Drierite 27068 L68GP Gas Purifier) to reduce humidity and remove impurities like CO2, followed by a high-efficiency particulate air (HEPA) filter to remove particles. The cleaned air then flows through a NRD P-2021 Polonium-210 (Po-210) source at 2 liters per minute (LPM) to generate superoxide (O2). These reagent ions are then transferred into the ion–molecule reactor (IMR), where they mix with gas-phase ambient samples at 80 mbar. Within the IMR, two dominant pathways are responsible for analyte ionization: deprotonation of the analyte (Reaction (R1)) and adduct formation with the reagent ion (Reaction (R2)). In some cases, analytes undergo additional adduct formation with water (H2O) or carbon dioxide (CO2), or they may fragment. Further details are provided in the Results and Discussion Section.
 
HX + O2 → X + HO2 (R1)
 
HX + O2 → HX(O2) (R2)

image file: d5an01352f-f1.tif
Fig. 1 O2 CIMS schematic diagram with CAMILS interfaced to inlet for PFAS calibrations. Reagent ions are generated using purified house air, and sampling flow uses room air, both at a flow rate of 2 LPM.

The resulting charged molecules pass through three differential vacuum chambers before entering the time-of-flight (TOF) mass analyzer for detection.55 This allows for accurate mass fittings and subsequent identification of molecular formulae.

PFAS calibrations

PFAS compounds for O2 CIMS calibration were selected based on their reported atmospheric and occupational relevances.73 Section S2 describes standard information and preparation protocol. To aid the introduction of PFAS standards into the O2 CIMS, the Calibration Apparatus for Manual Injection of Liquid Standards (CAMILS),62 a Restek sulfinert-coated Swagelok tee with GC septa used for liquid standard introduction for gas-phase PFAS analysis, was attached to the IMR as described previously by Davern et al.62

For each PFAS standard, 5, 10, 15, and 20 µL aliquots of 100 ng mL−1 and 1000 ng mL−1 solutions prepared in LC/MS grade methanol were injected in triplicate with a 25 µL Hamilton gastight syringe. More solvent selection information is discussed in the Results and Discussion Section (Fig. 3). Sufficient time was allotted between injections so that signals could return to the baseline. In some cases, to decrease this time, the temperature of the CAMILS increased up to 80 °C to encourage faster volatilization of the liquid standard. Each calibration was preceded by three 20 µL injections of methanol blanks to establish a background count for determining the sensitivity of calibrant ion peak m/z values. Limits of detection (LOD) and limits of quantitation (LOQ) were calculated as outlined in Bertram et al., using a signal-to-noise ratio of 3 and 10, respectively.55

Permeation tube for mass calibration

Given the lack of abundant signals above m/z 200, there arose a need to introduce small but consistent reference ions for mass calibration. Based on the lab inventory, compound volatility, high molecular weight (MW), and detectable structural analogs, we selected 5[thin space (1/6-em)]:[thin space (1/6-em)]1 FTOH (MW 300 g mol−1, and primarily detected by O2 CIMS as a superoxide adduct at m/z 332) to prepare a permeation tube for mass calibration. As it is less commonly reported than the more abundant n[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOHs (where n is an even number from 2 to 12), this compound is a suitable artificial standard, designed to minimize overlap with commonly occurring species found in ambient air.51,64,65,73,74 The 5[thin space (1/6-em)]:[thin space (1/6-em)]1 FTOH permeation tube is made by using a permeation tube kit (Owlstone Inc.), and 99.0% pure 5[thin space (1/6-em)]:[thin space (1/6-em)]1 FTOH (Flouryx Labs). As this is not for quantification purposes, the emission rate was not characterized. At room temperature (20 °C), with a room air flow rate of 2 LPM, it provides a stable, trace level of 5[thin space (1/6-em)]:[thin space (1/6-em)]1 FTOH, which forms different adduct clusters with the reagent ions (e.g., M + 32 and M + 94) allowing for its low-level continuous detection for mass calibration with little time required for equilibration. The O2 CIMS mass spectrum of 5[thin space (1/6-em)]:[thin space (1/6-em)]1 FTOH is provided in Section S3. The two 5[thin space (1/6-em)]:[thin space (1/6-em)]1 FTOH cluster peaks (i.e., m/z 332 and 394) alongside other known reagent ion peaks (i.e., m/z 32, 50, 60, 76 and 94, where details are provided in the Results and Discussion Section) were used as mass calibrants. The setup of the instrumentation is shown in Fig. 1, where the 5[thin space (1/6-em)]:[thin space (1/6-em)]1 FTOH permeation tube is placed upstream of the CAMILS to eliminate the contamination from standard injection in CAMILS. The impact of the permeation tube on reagent ion consumption is discussed in Section S4.

CIMS data collection and analysis

Data were collected using a high-resolution time-of-flight chemical ionization mass spectrometer (ToF-CIMS, Aerodyne Research Inc./TOFWERK AG), referred to as CIMS. It was characterized and described previously.55,62,67,75 Tofware Version 4.0.3 (Tofwerk AG) was used to analyze data. Instrument voltages, temperatures, and pressures were tuned to optimize sensitivity for I mode, which has been more well studied and characterized for PFAS air measurements. This was to prepare this method for CIMS operation with both I and O2 CIMS with rapid switching,71 get it ready for future field deployment, and enable cross-comparison of sensitivities. Mass spectra were collected with a range of 11–824 Da, at 1 Hz frequency and an extraction period of 50[thin space (1/6-em)]000 ns. More detailed instrument parameters including temperature, pressure and voltages can be found in Section S5.

Fast food package collection and storage

The fast-food paper wrap was obtained from a fast-food chain restaurant in Chapel Hill, NC, USA, and stored at room temperature and opened to ambient for a day before the analysis. After the perturbation experiment, the paper wrap was stored in a clear sandwich bag at room temperature. Half of it was then used for the GC-MS qualification.

Results and discussion

O2 CIMS mass spectrum, reagent ions, and normalization

Previously, O2 CIMS has been used to measure small molecules including methyl peroxide, hydrogen peroxide (H2O2),72 ozone (O3), and nitrogen dioxide (NO2).76 However, spectral complexity and interferences in the m/z 50–200 range make interpretation of small molecules more challenging. Once m/z values exceed 200, the O2 CIMS mass spectrum becomes clean, as shown in Fig. 2. This low-level mass spectral background is particularly advantageous for detecting high molecular weight compounds such as PFAS, and to the best of our knowledge, this capability has not previously been employed for this purpose, and all the previous O2 work only investigated compounds with m/z values <300.69,70,72,77
image file: d5an01352f-f2.tif
Fig. 2 General O2 CIMS mass spectrum with main reagent ions peaks (m/z 32, 50, 60, 76, and 94 corresponding to O2, (H2O)O2, CO3, (CO2)O2, CO2(H2O)O2, respectively) labeled black and trace environmental ions labeled blue.

To understand the ionization chemistry of O2 as a regent ion, it is important to confirm the identities of the low m/z peaks when no analytes are present (Fig. 2) through high-resolution peak fitting and introducing perturbations of suspected reagent compounds to the system. When no analytes are present, the O2 CIMS mass spectrum has major peaks at m/z values of 32, 50, 60, 76 and 94 (Fig. 2). Peak fitting and CO2 perturbations (Section S6) showed these correspond to O2, (H2O)O2, CO3, (CO2)O2, and CO2(H2O)O2, respectively, which is consistent with previous studies.72,76 These five peaks, which are produced from the primary reagent ion (O2) readily interacting with H2O and CO2 in ambient air, will henceforth be called the five main reagent ions and used for normalization due to their fluctuations as analytes enter the IMR and react with them. Rationale for the selection of the reagent ion pool in O2 CIMS is described in Section S7. Other ion peaks (labeled in light blue) are not considered as part of the main reagent ions because either they are present in trace amounts, or they are a constant background most of the time and do not readily react with analytes.

Given that analyte signals are dependent on the availability of reagent ions at a given time, quantification requires normalization. In I CIMS, the sum of the signals of I and H2OI are used as an approximate measure of the total reagent ion signal, which is then normalized per million counts as a normalization factor for all analyte signals, a standard data processing step.62,66 This procedure accounts for both instrument variability and changing operating conditions. O2 CIMS has not previously been applied to PFAS measurements in air, there is no established uniform processing method. Whereas I CIMS analysis has established normalization, there is no such common practice for O2 CIMS. To address this, we normalized O2 CIMS analyte signals per million counts of the five main reagent ions described above to partially account for variations in humidity and CO2 (eqn (1)). While this approach reduces variability associated with changing environmental conditions, which is advantageous for indoor deployments where CO2 levels fluctuate with ventilation and occupancy,78 our results suggest that sensitivity variations associated with RH are not fully accounted for by this normalization, as discussed in Section S8.

 
image file: d5an01352f-t1.tif(1)

Solvent selection for calibration

Fig. 3 shows the main reagent ion fluctuations when a new solvent is introduced into the system. As PFAS calibration requires the selection of a suitable solvent, an ideal solvent must sufficiently dissolve the compound of interest, be volatile enough to carry standards to the CIMS for analysis, and have minimal impact on reagent ionization efficiency and distributions. We examined three candidate LC/MS grade solvents by injecting 20 µL of each (i.e., methanol, acetonitrile, and ethyl acetate) without PFAS into the CIMS.
image file: d5an01352f-f3.tif
Fig. 3 O2 CIMS time series of main reagent ion peak response to methanol (MeOH, yellow shading), acetonitrile (ACN, red shading), and ethyl acetate (EtOAc, blue shading) in chronological order.

As shown in Fig. 3, although methanol consistently caused a decrease in the five main reagent ion peaks due to formation of the methanol-superoxide adduct at m/z 64, it produced the smallest change in the major reagent ion (O2) and the decrease was uniform across all reagent ions. Acetonitrile caused larger, contrasting changes at m/z 32 (O2) and 50 ((H2O)O2), while ethyl acetate resulted in the greatest fluctuations, most notably demonstrated by a nearly 25% decrease in O2.

Based on these observations, methanol was selected as the solvent for blanks and PFAS standard preparation. This choice minimizes the impact on reagent ion distributions. Conveniently, most certified PFAS standards are supplied in methanol, simplifying laboratory preparation, and avoiding solvent incompatibility issues. While methanol yielded the most favorable results for calibration, acetonitrile remains a viable alternative, such as when standards are only available in acetonitrile or a PFAS has poor solubility in methanol.

PFAS compounds detected

O2 CIMS appears to be sensitive to oxygenated semi-volatile PFAS compounds, including FTOHs, sulfonamides, epoxides, fluorotelomer diols (FTdiOH), and glycol ethers. However, some notable classes of PFAS were observed to have low sensitivity or no distinguishable signal, including perfluorinated carboxylic acids (PFCAs), fluorotelomer acrylates (FTACs), and perfluorinated sulfonic acids (PFSAs). It should be noted that PFCAs have been previously detected by I CIMS.62,67,79 The 37 compounds tested for suitability with O2 CIMS and their detectability are summarized in Section S9 (Table S4).

While the multiple reagent ions of O2 CIMS can complicate mass spectra, they also offer distinct advantages. A common challenge in PFAS detection by CIMS is that FTOHs and PFCAs have similar masses (<0.05 Da), making it difficult for instruments with resolving power less than 10[thin space (1/6-em)]000 to confidently distinguish and separate FTOHs in complex matrices based solely on mass.64 As such, additional GC or LC separation is typically required for high-confidence confirmation of signal identification. O2 eliminates interference for FTOH measurements or otherwise enables deconvolution of I CIMS signals. With the potential to switch rapidly between I and O2 mode, it is possible to monitor both PFCAs and FTOHs. Moreover, O2 CIMS produces a characteristic series of peaks for n[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOHs (n = 4, 6, 8, 10) at M + 46, M + 60, M + 78, M + 92, and M + 94 (Fig. S4), in addition to the main superoxide cluster (M + 32). These “fingerprint” ion peaks can aid in verifying the presence of FTOHs, much how tandem MS fragment ions are used in signal interpretation. For example, as shown in Fig. 4, the mass spectrum of a 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH standard closely matched that of a floor wax sample previously confirmed to contain 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH by GC/MS and I CIMS,64 with identical fingerprint ion peaks observed. This finding suggests that, in some cases, ultra-high-resolution mass spectrometry or upstream GC/LC separation may not be necessary for compound identification when using O2 CIMS. While these fingerprint ion peaks are not used for quantitative purposes, they provide valuable complementary information for verification of PFAS compound identity as qualifying ions.


image file: d5an01352f-f4.tif
Fig. 4 Mass spectra obtained using O2 CIMS showing 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH fingerprint peaks comparison in (a) 1000 ng mL−1 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH in methanol headspace sniff test, (b) brand name floor wax headspace sniff test. The y-axis was scaled to highlight the isotopic peak (M + 1 O2 adduct at m/z 397) and other diagnostic peaks. The peak marked with a star at m/z 449 originates from an unidentified background component in the floor wax. With increasing 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH concentration, a corresponding increase in the relative intensity of the M + 94 peak at m/z 458 was observed, and the 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH dimer with superoxide cluster was also detected (m/z 760). This trend was similarly observed for other n[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOHs. The mass error for M + 32 at m/z 396 (C8H5F13O3) is 14.6 parts per million (ppm), and for M + 94 at m/z 458 (C9H7F13O6) is 31 ppm.

PFAS calibration results

All sensitivity calibration results are shown in Table 1, with comparisons to I CIMS addressed later (Table 2). Given the complexity of superoxide ionization, preliminary injections were performed to identify the major ion peaks used for calibration. Complete mass spectra of 14 individual PFAS compounds are provided in Section S10, along with detailed reasoning for calibration ion peaks selection in each figure caption. Calibrations were performed using methanol as the solvent at 50 ± 2% relative humidity (RH), where the humidity reflected ambient laboratory air (i.e., room air measured at that level) and no additional humidification of the carrier gas was applied, which is representative of typical daytime outdoor conditions in the southeastern United States, despite expected seasonal and diurnal variability.80,81 The sensitivities’ dependency on the RH is discussed in Section S8 with FTOHs.
Table 1 O2 CIMS calibration results
Compound CAS # Formula Unit m/z Calibrant ion peak (s) (m/z) Sensitivity (ncps pptv−1) R2 LOD (pptv) LOQ (pptv)
4[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH 2043-47-2 C6H5F9O 264 296 11.8 ± 0.3 0.9953 0.75 3.78
6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH 647-42-7 C8H5F13O 364 396 13.7 ± 0.4 0.9953 0.51 2.97
8[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH 678-39-7 C10H5F17O 464 496 9.7 ± 0.2 0.9970 0.80 4.75
10[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH 865-86-1 C12H5F21O 564 596 7.2 ± 0.2 0.9973 0.68 5.22
MeFOSA 31506-32-8 C9H4F17NO2S 513 512, 545 25.4 ± 0.5 0.9979 0.23 1.46
EtFOSA 4151-50-2 C10H6F17NO2S 527 526, 559 27.7 ± 0.4 0.9988 0.21 1.35
MeFOSE 24448-09-7 C11H8F17NO3S 557 589, 651 13.4 ± 0.4 0.9954 0.50 2.99
EtFOSE 1691-99-2 C12H10F17NO3S 571 603, 665 13.9 ± 0.5 0.9934 0.52 3.13
3-(Perfluorohexyl)propyl epoxide 38565-52-5 C9H5F13O 376 408 0.124 ± 0.002 0.9992 68.61 362.53
3-(Perfluorooctyl)propyl epoxide 38565-53-6 C11H5F17O 476 508 0.069 ± 0.001 0.9993 78.83 541.59
FBSA 30334-69-1 C4H2F9NO2S 299 298, 331, 359 48.3 ± 0.5 0.9993 0.38 1.55
1[thin space (1/6-em)]:[thin space (1/6-em)]8[thin space (1/6-em)]:[thin space (1/6-em)]1 FTdiOH 754-96-1 C10H6F16O2 462 461, 494, 556 21.3 ± 0.2 0.9995 0.64 3.06
Fluorinated triethylene glycol 129301-42-4 C6H6F8O4 294 293, 326. 388 22.9 ± 0.5 0.9975 0.12 1.38
Fluorinated tetraethylene glycol 330562-44-2 C8H6F12O5 410 307, 409, 442 41 ± 2 0.9920 0.07 0.72


Table 2 Comparison with Davern et al. I CIMS calibrations
Compound name/abbreviation Sensitivity (ncps pptv−1) LOD (pptv)
This study Davern et al.62 This study Davern et al.62
ND: not detected; NA: not applicable.a Standard deviations were not reported in Davern et al.
4[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH 11.8 ± 0.3 4.3 ± 0.01 0.75 3.1
6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH 13.7 ± 0.4 6.1 ± 0.03 0.51 1.5
8[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH 9.7 ± 0.2 6.2 ± 0.06 0.8 2
10[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH 7.2 ± 0.2 5.7 ± 0.07 0.68 1.7
EtFOSA 27.7 ± 0.4 19.7a 0.21 0.2
1[thin space (1/6-em)]:[thin space (1/6-em)]8[thin space (1/6-em)]:[thin space (1/6-em)]1 FTdiOH 21.3 ± 0.2 46.9a 0.64 0.4
PFOA ND 12.4 ± 0.2 NA 0.7
PFBA ND 20.7 ± 0.3 NA 1.3
HFPO-DA NA 23.1 ± 0.3 NA 0.7
E2 NA 0.0047 NA 1700


In some cases, more than one m/z value was used for quantification. In those cases, the normalized signals of selected ion peaks were summed and plotted against time, and the integrated areas were used to construct calibration curves. Calibration curves are provided in Section S11. The slope of each 8-point calibration curve represents the instrument's sensitivity, reported as normalized counts per second per parts per trillion by volume (ncps pptv−1). R2 values greater than 0.99 were calculated for the orthogonal linear regressions of all 14 calibrations. Furthermore, all calibrations yielded relative standard deviations below 10%, indicating acceptable precision for quantitative analysis. An example calibration plot of FTOHs, which have been quantified through many gas-phase methods,62,67 is shown in Fig. 5.


image file: d5an01352f-f5.tif
Fig. 5 FTOHs calibration plot, where 4[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH (orange and square), 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH (blue and round), 8[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH (green and triangle), 4[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH (purple and rhombus) is labeled in the figure. Best-fit orthogonal linear regressions are shown with dashed lines. Each calibration point was measured in triplicate on the same day, with error bars (black) representing the standard deviations of replicate measurements. All reported sensitivities are normalized per million counts of reagent ion signal.

When compared to I CIMS calibrations obtained using the CAMILS method by Davern et al.62 (Table 2), O2 CIMS had higher sensitivity for five of the six compounds tested in both studies (e.g., 270%, 220%, 160%, 126%, and 140% the sensitivity of 4[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH, 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH, 8[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH, 10[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH, and EtFOSA, respectively). The FTOHs all had lower LODs (i.e., sub-pptv) using O2 CIMS, while the EtFOSA LOD was comparable. However, O2 CIMS was 45% as sensitive to 1[thin space (1/6-em)]:[thin space (1/6-em)]8[thin space (1/6-em)]:[thin space (1/6-em)]1 FTdiOH, and it failed to significantly detect any of the PFCAs (e.g., PFBA and PFOA) that Davern et al.62 measured using I CIMS, suggesting that O2 reagent ion chemistry has a lower affinity for PFCAs. Furthermore, despite the similar instrument settings and the use of CAMILS, the uncertainties for the sensitivity in this study were greater than in Davern et al.,62 likely due to the more complex and variable ionization pathways present in O2 reagent ion chemistry.

Real-time gas-phase FTOH emissions from fast-food packaging

When the food paper wrap was present, real-time 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH emissions from a consumer product at room temperature were detected. As shown in Fig. 6, even when the paper wrap was left undisturbed at 10 cm from the instrument inlet, the 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH concentration from food-paper wrap emissions was above the instrument's LOQ. Upon agitating the paper wrap by rubbing its surfaces together, higher 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH emissions were observed, with peak concentrations reaching up to 16 pptv. Once the paper wrap was removed from the vicinity of the inlet, the signal immediately returned to background levels, suggesting minimal measurement delays. This experiment highlights the potential for real-time PFAS air measurements and point-of-source determination using O2 CIMS, even for highly dynamic systems. Even though PFAS have previously been reported in food packaging and suggested to volatilize into indoor air,10 the use of traditional offline methods required extensive sample collection, preparation, and extraction. The real-time (online) approach presented herein requires no sample preparation while providing highly time-resolved measurements. Supporting results, including GC/EI-MS verification for 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH and perturbation fingerprints of 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH, are provided in Section S12.
image file: d5an01352f-f6.tif
Fig. 6 Time series of 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH perturbation (shown in red) from a fast-food packaging paper wrap. The results are smoothed for 3 seconds to reduce fluctuations. Instrument LOQ of 2.97 pptv is shown with a yellow, dashed line calculated based on calibration results.

Limitations & further characterization

There are some limitations when using the O2 CIMS method. Although it offers safer reagent ion generation through elimination of the need for methyl iodide compared to I CIMS, it still relies on a commercially available sealed Po-210 radioactive source. Its use requires compliance with laboratory safety standards.82 With a half-life of 138 days, the ion source must be replaced approximately once per year with notable decreases in absolute sensitivity happening before then. Another common challenge with general CIMS is the need for calibration prior to quantification, as well as the limited availability of standards for short-lived or otherwise highly reactive gas-phase analytes of interest. Thus, non-targeted PFAS analysis can be challenging,83 but a GC separation upstream can help rule out complexities. Future work will examine the coupling of GC to CIMS for non-targeted PFAS detection.

It is important to emphasize that the present work is intended as a feasibility study demonstrating the applicability of O2 CIMS to capture environmentally and industrially relevant classes of gas-phase PFAS like FTOHs, further analytical exploration is warranted. Due to the nature of this work being a feasibility study, future work should include systematic optimization of instrument voltages to balance declustering and analyte ion stability, incorporation of higher molecular weight calibration compounds to improve mass accuracy at high m/z, and further mechanistic characterization of ionization pathways alongside controlled studies of H2O and CO2 effects to establish a more rigorous normalization framework. Exploration of alternative ionization sources (e.g., VUV lamps)84 may also reduce radiation safety concerns and provide insight into differences in ionization behavior. In addition, expanding compound coverage and evaluating performance in complex matrices with coexisting VOCs will be important, potentially leveraging AI-assisted data analysis and spectral prediction.

Environmental implications and conclusions

The use of O2 as a reagent ion in CIMS demonstrates potential for real-time gas-phase measurement of PFAS in addition to other VOCs. Compared to I CIMS, O2 CIMS is safer and requires less ventilation, making it more suitable for field and mobile deployments, as well as space-constrained laboratory environments. It exhibits higher sensitivity and lower detection limits for FTOHs, making it well-suited for capturing transient gaseous FTOH emissions under ambient conditions. This capability highlights its potential for determining concentrations for a wide range of PFAS compounds relevant to occupational, residential, and ambient air environments. Real-time air measurements can also provide insight into PFAS sources and support the development of exposure mitigation strategies. Once calibrated, O2 CIMS can provide real-time quantitative PFAS measurements. For some PFAS compounds, distinct fingerprint ion peaks provide an additional dimension of confirmation, reducing reliance on offline separation methods such as GC-/LC-MS, thus saving time, cost, and analytical effort. Moreover, identical voltage and pressure settings shared between I and O2 CIMS parameters suggest the potential for rapid switching between these reagent ions, thereby facilitating deconvolution of CIMS signals and providing complementary dimensions of chemical information.

Author contributions

The manuscript was written through contributions of all authors. All authors have given approval to the final version of the manuscript. It is noted that Yufan Hu and Joji W. Sherman co-led this work in terms of the design and execution of the planned studies, and co-led the writing of this manuscript. As a result, they are co-first authors of this study.

Conflicts of interest

There are no conflicts to report.

Data availability

Raw and processed mass spectral data related to the method development, PFAS calibrations, and fast-food wrap will be deposited in the UNC Dataverse, a trustworthy, generalist data repository managed by the Research Data Management Core at the University of North Carolina at Chapel Hill. UNC Dataverse provides persistent identifiers, robust standardized metadata, and is committed to long-term preservation and access of research data. Data are published under a CC0 license by default with customizable terms of use as needed. Additionally, UNC Dataverse is routinely backed up and preserved on multiple geographically distributed servers and is a member of Data-PASS, a community committed to the sustainability and access of research data. Access to original data can be found at: https://doi.org/10.15139/S3/9SFZMS.

Supplementary information (SI): safety considerations, standards information and calibration preparation, a mass spectrum of 5[thin space (1/6-em)]:[thin space (1/6-em)]1 FTOH, permeation tube impacts on the reagent ions, O2 CIMS operating conditions and related parameters, confirmation of main reagent ion identities, rationale of reagent ions selection, sensitivities of FTOHs with different relative humidities (RH), a master list of calibration details, the mass spectra of all calibrated compounds along with discussions of the criteria used for selecting calibration ion peaks, calibration plots, and additional evidence of 6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH detection in fast-food wraps. See DOI: https://doi.org/10.1039/d5an01352f.

Acknowledgements

This project was primarily supported by the North Carolina Collaboratory at The University of North Carolina at Chapel Hill with funding appropriated by the North Carolina General Assembly. This publication was also developed under Assistance Agreement No. R840431 awarded by the U. S. Environmental Protection Agency to University of North Carolina at Chapel Hill. It has not been formally reviewed by EPA. The views expressed in this document are solely those of authors and do not necessarily reflect those of the Agency. EPA does not endorse any products or commercial services mentioned in this publication. This work was also funded in part by the National Institute for Occupational Safety and Health (5T42OH008673).

References

  1. K. Prevedouros, I. T. Cousins, R. C. Buck and S. H. Korzeniowski, Sources, Fate and Transport of Perfluorocarboxylates, Environ. Sci. Technol., 2006, 40, 32–44 CrossRef CAS PubMed.
  2. J. Glüge, M. Scheringer, I. T. Cousins, J. C. DeWitt, G. Goldenman, D. Herzke, R. Lohmann, C. A. Ng, X. Trier and Z. Wang, An overview of the uses of per- and polyfluoroalkyl substances (PFAS), Environ. Sci. Process. Impacts, 2020, 22, 2345–2373 RSC.
  3. L. G. T. Gaines, Historical and current usage of per- and polyfluoroalkyl substances (PFAS): A literature review, Am. J. Ind. Med., 2023, 66, 353–378 CrossRef CAS PubMed.
  4. Z. Wang, J. C. DeWitt, C. P. Higgins and I. T. Cousins, A Never-Ending Story of Per- and Polyfluoroalkyl Substances (PFASs)?, Environ. Sci. Technol., 2017, 51, 2508–2518 CrossRef CAS PubMed.
  5. I. van der Veen, S. Schellenberger, A.-C. Hanning, A. Stare, J. de Boer, J. M. Weiss and P. E. G. Leonards, Fate of Per- and Polyfluoroalkyl Substances from Durable Water-Repellent Clothing during Use, Environ. Sci. Technol., 2022, 56, 5886–5897 CrossRef CAS PubMed.
  6. G. Glenn, R. Shogren, X. Jin, W. Orts, W. Hart-Cooper and L. Olson, Per- and polyfluoroalkyl substances and their alternatives in paper food packaging, Compr. Rev. Food Sci. Food Saf., 2021, 20, 2596–2625 CrossRef CAS PubMed.
  7. L. A. Schaider, S. A. Balan, A. Blum, D. Q. Andrews, M. J. Strynar, M. E. Dickinson, D. M. Lunderberg, J. R. Lang and G. F. Peaslee, Fluorinated Compounds in U.S. Fast Food Packaging, Environ. Sci. Technol. Lett., 2017, 4, 105–111 CrossRef CAS PubMed.
  8. M. Sajid and M. Ilyas, PTFE-coated non-stick cookware and toxicity concerns: a perspective, Environ. Sci. Pollut. Res., 2017, 24, 23436–23440 CrossRef CAS PubMed.
  9. G. Guerrero-Vacas, F. Comino and O. Rodríguez-Alabanda, Evaluation of the effectiveness and durability of commercial non-stick coatings, J. Food Eng., 2024, 370, 111959 CrossRef CAS.
  10. H. Schwartz-Narbonne, C. Xia, A. Shalin, H. D. Whitehead, D. Yang, G. F. Peaslee, Z. Wang, Y. Wu, H. Peng, A. Blum, M. Venier and M. L. Diamond, Per- and Polyfluoroalkyl Substances in Canadian Fast Food Packaging, Environ. Sci. Technol. Lett., 2023, 10, 343–349 CrossRef CAS PubMed.
  11. J. Glüge, K. Breuer, A. Hafner, C. Vering, D. Müller, I. T. Cousins, R. Lohmann, G. Goldenman and M. Scheringer, Finding non-fluorinated alternatives to fluorinated gases used as refrigerants,  10.1039/D4EM00444B.
  12. D. J. Seow, Fire Fighting Foams with Perfluorochemicals-Environmental Review.
  13. K. Pozo, L. B. Moreira, P. Karaskova, P. Přibylová, J. Klánová, M. U. de Carvalho, L. A. Maranho, D. M and A. de Souza, Using large amounts of firefighting foams releases per- and polyfluoroalkyl substances (PFAS) into estuarine environments: A baseline study in Latin America, Mar. Pollut. Bull., 2022, 182, 113938 CrossRef CAS PubMed.
  14. G. F. Peaslee, J. T. Wilkinson, S. R. McGuinness, M. Tighe, N. Caterisano, S. Lee, A. Gonzales, M. Roddy, S. Mills and K. Mitchell, Another Pathway for Firefighter Exposure to Per- and Polyfluoroalkyl Substances: Firefighter Textiles, Environ. Sci. Technol. Lett., 2020, 7, 594–599 CrossRef CAS.
  15. I. T. Cousins, J. C. DeWitt, J. Glüge, G. Goldenman, D. Herzke, R. Lohmann, M. Miller, C. A. Ng, M. Scheringer, L. Vierke and Z. Wang, Strategies for grouping per- and polyfluoroalkyl substances (PFAS) to protect human and environmental health,  10.1039/D0EM00147C.
  16. M. G. Evich, M. J. B. Davis, J. P. McCord, B. Acrey, J. A. Awkerman, D. R. U. Knappe, A. B. Lindstrom, T. F. Speth, C. Tebes-Stevens, M. J. Strynar, Z. Wang, E. J. Weber, W. M. Henderson and J. W. Washington, Per- and polyfluoroalkyl substances in the environment, Science, 2022, 375, eabg9065 CrossRef CAS PubMed.
  17. L. Zhao, L. Zhu, S. Zhao and X. Ma, Sequestration and bioavailability of perfluoroalkyl acids (PFAAs) in soils: Implications for their underestimated risk, Sci. Total Environ., 2016, 572, 169–176 CrossRef CAS PubMed.
  18. J. P. Giesy and K. Kannan, Global Distribution of Perfluorooctane Sulfonate in Wildlife, Environ. Sci. Technol., 2001, 35, 1339–1342 CrossRef CAS PubMed.
  19. P. Zareitalabad, J. Siemens, M. Hamer and W. Amelung, Perfluorooctanoic acid (PFOA) and perfluorooctanesulfonic acid (PFOS) in surface waters, sediments, soils and wastewater – A review on concentrations and distribution coefficients, Chemosphere, 2013, 91, 725–732 CrossRef CAS PubMed.
  20. A. O. De Silva, C. Spencer, B. F. Scott, S. Backus and D. C. G. Muir, Detection of a Cyclic Perfluorinated Acid, Perfluoroethylcyclohexane Sulfonate, in the Great Lakes of North America, Environ. Sci. Technol., 2011, 45, 8060–8066 CrossRef CAS PubMed.
  21. K. Y. Christensen, M. Raymond, M. Blackowicz, Y. Liu, B. A. Thompson, H. A. Anderson and M. Turyk, Perfluoroalkyl substances and fish consumption, Environ. Res., 2017, 154, 145–151 CrossRef CAS PubMed.
  22. P. Grandjean, E. W. Andersen, E. Budtz-Jørgensen, F. Nielsen, K. Mølbak, P. Weihe and C. Heilmann, Serum Vaccine Antibody Concentrations in Children Exposed to Perfluorinated Compounds, J. Am. Med. Assoc., 2012, 307, 391–397 CrossRef CAS PubMed.
  23. E. M. Sunderland, X. C. Hu, C. Dassuncao, A. K. Tokranov, C. C. Wagner and J. G. Allen, A review of the pathways of human exposure to poly- and perfluoroalkyl substances (PFASs) and present understanding of health effects, J. Expo. Sci. Environ. Epidemiol., 2019, 29, 131–147 CrossRef CAS PubMed.
  24. V. M. Vieira, K. Hoffman, H.-M. Shin, J. M. Weinberg, T. F. Webster and T. Fletcher, Perfluorooctanoic Acid Exposure and Cancer Outcomes in a Contaminated Community: A Geographic Analysis, Environ. Health Perspect., 2013, 121, 318–323 CrossRef PubMed.
  25. N. Johansson, A. Fredriksson and P. Eriksson, Neonatal exposure to perfluorooctane sulfonate (PFOS) and perfluorooctanoic acid (PFOA) causes neurobehavioural defects in adult mice, NeuroToxicology, 2008, 29, 160–169 CrossRef CAS PubMed.
  26. T. G. Ambaye, M. Vaccari, S. Prasad and S. Rtimi, Recent progress and challenges on the removal of per- and poly-fluoroalkyl substances (PFAS) from contaminated soil and water, Environ. Sci. Pollut. Res., 2022, 29, 58405–58428 CrossRef CAS PubMed.
  27. K. H. Kucharzyk, R. Darlington, M. Benotti, R. Deeb and E. Hawley, Novel treatment technologies for PFAS compounds: A critical review, J. Environ. Manage., 2017, 204, 757–764 CrossRef CAS PubMed.
  28. B. C. Crone, T. F. Speth, D. G. Wahman, S. J. Smith, G. Abulikemu, E. J. Kleiner and J. G. Pressman, Occurrence of per- and polyfluoroalkyl substances (PFAS) in source water and their treatment in drinking water, Crit. Rev. Environ. Sci. Technol., 2019, 49, 2359–2396 CrossRef PubMed.
  29. E. Kumarasamy, I. M. Manning, L. B. Collins, O. Coronell and F. A. Leibfarth, Ionic Fluorogels for Remediation of Per- and Polyfluorinated Alkyl Substances from Water, ACS Cent. Sci., 2020, 6, 487–492 CrossRef CAS PubMed.
  30. I. M. Manning, N. Guan Pin Chew, H. P. Macdonald, K. E. Miller, M. J. Strynar, O. Coronell and F. A. Leibfarth, Hydrolytically Stable Ionic Fluorogels for High-Performance Remediation of Per- and Polyfluoroalkyl Substances (PFAS) from Natural Water, Angew. Chem., 2022, 134, e202208150 CrossRef.
  31. K. Weitz, D. Kantner, A. Kessler, H. Key, J. Larson, W. Bodnar, S. Parvathikar, L. Davis, N. Robey, P. Taylor, F. De la Cruz, T. Tolaymat, N. Weber, W. Linak, J. Krug and L. Phelps, Review of per- and poly-fluoroalkyl treatment in combustion-based thermal waste systems in the United States, Sci. Total Environ., 2024, 932, 172658 CrossRef CAS PubMed.
  32. G. R. Johnson, PFAS in soil and groundwater following historical land application of biosolids, Water Res., 2022, 211, 118035 CrossRef CAS PubMed.
  33. Y. Hu, H. Chen, Y. Chen, Y. Wang, Y. Luo, L. Sang, T. Jin and S. Wu, Perfluoroalkyl acids (PFAAs) and their precursors in sediments and adjacent riparian soils from the Three Gorges Reservoir, China: Contamination characteristics, source apportionment and ecological risks, Environ. Res., 2025, 274, 121202 CrossRef CAS PubMed.
  34. O. US EPA, Per- and Polyfluoroalkyl Substances (PFAS), https://www.epa.gov/sdwa/and-polyfluoroalkyl-substances-pfas, (accessed 23 August 2025).
  35. T. Teymoorian, G. Munoz, S. Vo Duy, J. Liu and S. Sauvé, Tracking PFAS in Drinking Water: A Review of Analytical Methods and Worldwide Occurrence Trends in Tap Water and Bottled Water, ACS ES&T Water, 2023, 3, 246–261 Search PubMed.
  36. S. J. Chow, N. Ojeda, J. G. Jacangelo and K. J. Schwab, Detection of ultrashort-chain and other per- and polyfluoroalkyl substances (PFAS) in U.S. bottled water, Water Res., 2021, 201, 117292 CrossRef CAS PubMed.
  37. C. Gremmel, T. Frömel and T. P. Knepper, HPLC–MS/MS methods for the determination of 52 perfluoroalkyl and polyfluoroalkyl substances in aqueous samples, Anal. Bioanal. Chem., 2017, 409, 1643–1655 CrossRef CAS PubMed.
  38. X. Li, W. Li, Z. Wang, X. Wang, Y. Cai and Y. Shi, Atmospheric Emission of Per- and Polyfluoroalkyl Substances (PFAS) from a Fluoropolymer Manufacturing Facility: Focus on Emerging PFAS and the Potential Contribution of Condensable PFAS on their Atmospheric Partitioning, Environ. Sci. Technol., 2025, 59, 9709–9720 CrossRef CAS PubMed.
  39. W.-L. Li and K. Kannan, Determination of Legacy and Emerging Per- and Polyfluoroalkyl Substances (PFAS) in Indoor and Outdoor Air, ACS ES&T Air, 2024, 1, 1147–1155 Search PubMed.
  40. D. Persaud, A. S. Criscitiello, C. Spencer, I. Lehnherr, D. C. G. Muir, A. O. D. Silva and C. J. Young, A 50 years record for perfluoroalkyl acids in the high arctic: implications for global and local transport, Environ. Sci. Process. Impacts, 2024, 26, 1543–1555 RSC.
  41. J. J. MacInnis, K. French, D. C. G. Muir, C. Spencer, A. Criscitiello, A. O. D. Silva and C. J. Young, Emerging investigator series: a 14-year depositional ice record of perfluoroalkyl substances in the High Arctic, Environ. Sci. Process. Impacts, 2017, 19, 22–30 RSC.
  42. F. Wania, A Global Mass Balance Analysis of the Source of Perfluorocarboxylic Acids in the Arctic Ocean, Environ. Sci. Technol., 2007, 41, 4529–4535 CrossRef CAS PubMed.
  43. R. Holland, M. A. H. Khan, R. Chhantyal-Pun, A. J. Orr-Ewing, C. J. Percival, C. A. Taatjes and D. E. Shallcross, Investigating the Atmospheric Sources and Sinks of Perfluorooctanoic Acid Using a Global Chemistry Transport Model, Atmosphere, 2020, 11, 407 CrossRef CAS.
  44. E. L. D'Ambro, H. O. T. Pye, J. O. Bash, J. Bowyer, C. Allen, C. Efstathiou, R. C. Gilliam, L. Reynolds, K. Talgo and B. N. Murphy, Characterizing the Air Emissions, Transport, and Deposition of Per- and Polyfluoroalkyl Substances from a Fluoropolymer Manufacturing Facility, Environ. Sci. Technol., 2021, 55, 862–870 CrossRef PubMed.
  45. J. Garnett, C. Halsall, H. Winton, H. Joerss, R. Mulvaney, R. Ebinghaus, M. Frey, A. Jones, A. Leeson and P. Wynn, Increasing Accumulation of Perfluorocarboxylate Contaminants Revealed in an Antarctic Firn Core (1958–2017), Environ. Sci. Technol., 2022, 56, 11246–11255 CrossRef CAS PubMed.
  46. J. Roostaei, S. Colley, R. Mulhern, A. A. May and J. M. Gibson, Predicting the risk of GenX contamination in private well water using a machine-learned Bayesian network model, J. Hazard. Mater., 2021, 411, 125075 CrossRef CAS PubMed.
  47. J. E. Galloway, A. V. P. Moreno, A. B. Lindstrom, M. J. Strynar, S. Newton, A. A. May and L. K. Weavers, Evidence of Air Dispersion: HFPO–DA and PFOA in Ohio and West Virginia Surface Water and Soil near a Fluoropolymer Production Facility, Environ. Sci. Technol., 2020, 54, 7175–7184 CrossRef CAS PubMed.
  48. R. Wu, H. Lin, E. Yamazaki, S. Taniyasu, M. Sörengård, L. Ahrens, P. K. S. Lam, H. Eun and N. Yamashita, Simultaneous analysis of neutral and ionizable per- and polyfluoroalkyl substances in air, Chemosphere, 2021, 280, 130607 CrossRef CAS PubMed.
  49. H. P. H. Arp and K.-U. Goss, Irreversible sorption of trace concentrations of perfluorocarboxylic acids to fiber filters used for air sampling, Atmos. Environ., 2008, 42, 6869–6872 CrossRef CAS.
  50. J. L. Barber, U. Berger, C. Chaemfa, S. Huber, A. Jahnke, C. Temme and K. C. Jones, Analysis of per- and polyfluorinated alkyl substances in air samples from Northwest Europe, J. Environ. Monit., 2007, 9, 530–541 RSC.
  51. N. Y. Chang, C. M. A. Eichler, D. E. Amparo, J. Zhou, K. Baumann, E. A. C. Hubal, J. D. Surratt, G. C. Morrison and B. J. Turpin, Indoor air concentrations of PM2.5 quartz fiber filter-collected ionic PFAS and emissions to outdoor air: findings from the IPA campaign, Environ. Sci. Process. Impacts, 2025, 27, 1603–1618 RSC.
  52. S. Mok, S. Lee, Y. Choi, J. Jeon, Y. H. Kim and H.-B. Moon, Target and non-target analyses of neutral per- and polyfluoroalkyl substances from fluorochemical industries using GC-MS/MS and GC-TOF: Insights on their environmental fate, Environ. Int., 2023, 182, 108311 CrossRef CAS PubMed.
  53. W. Zhang, L. Xu and H. Zhang, Recent advances in mass spectrometry techniques for atmospheric chemistry research on molecular-level, Mass Spectrom. Rev., 2024, 43, 1091–1134 CrossRef CAS PubMed.
  54. L. G. Huey, Measurement of trace atmospheric species by chemical ionization mass spectrometry: Speciation of reactive nitrogen and future directions, Mass Spectrom. Rev., 2007, 26, 166–184 CrossRef CAS PubMed.
  55. T. H. Bertram, J. R. Kimmel, T. A. Crisp, O. S. Ryder, R. L. N. Yatavelli, J. A. Thornton, M. J. Cubison, M. Gonin and D. R. Worsnop, A field-deployable, chemical ionization time-of-flight mass spectrometer, Atmos. Meas. Tech., 2011, 4, 1471–1479 CrossRef CAS.
  56. Vocus - Aerodyne, https://aerodyne.com/gas-phase-analyzers/vocus/, (accessed 17 February 2025).
  57. A. R. Koss, C. Warneke, B. Yuan, M. M. Coggon, P. R. Veres and J. A. de Gouw, Evaluation of NO+ reagent ion chemistry for online measurements of atmospheric volatile organic compounds, Atmos. Meas. Tech., 2016, 9, 2909–2925 CrossRef CAS.
  58. B. Yuan, A. Koss, C. Warneke, J. B. Gilman, B. M. Lerner, H. Stark and J. A. de Gouw, A high-resolution time-of-flight chemical ionization mass spectrometer utilizing hydronium ions (H3O+ ToF-CIMS) for measurements of volatile organic compounds in the atmosphere, Atmos. Meas. Tech., 2016, 9, 2735–2752 CrossRef CAS.
  59. S. Gagan, M. Olin, A. J. Dodero, S. Gopalakrishnan, S. Niu, M. J. Davern, B. J. Turpin, J. D. Surratt and Y. Zhang, Real-Time Identification and Quantification of Per- and Polyfluoroalkyl Substances Using High-Resolution Time-of-Flight Chemical Ionization Mass Spectrometry with Positive Reagent Ions, Anal. Chem., 2026, 98, 124–133 CrossRef CAS PubMed.
  60. M. Riva, P. Rantala, J. E. Krechmer, O. Peräkylä, Y. Zhang, L. Heikkinen, O. Garmash, C. Yan, M. Kulmala, D. Worsnop and M. Ehn, Evaluating the performance of five different chemical ionization techniques for detecting gaseous oxygenated organic species, Atmos. Meas. Tech., 2019, 12, 2403–2421 CrossRef CAS.
  61. J. M. Mattila, J. D. Krug, W. R. Roberson, R. P. Burnette, S. McDonald, L. Virtaranta, J. H. Offenberg and W. P. Linak, Characterizing Volatile Emissions and Combustion Byproducts from Aqueous Film-Forming Foams Using Online Chemical Ionization Mass Spectrometry, Environ. Sci. Technol., 2024, 58, 3942–3952 CrossRef CAS PubMed.
  62. M. J. Davern, G. V. West, C. M. A. Eichler, B. J. Turpin, Y. Zhang and J. D. Surratt, External liquid calibration method for iodide chemical ionization mass spectrometry enables quantification of gas-phase per- and polyfluoroalkyl substances (PFAS) dynamics in indoor air, Analyst, 2024, 149, 3405–3415 RSC.
  63. J. Zhou, K. Baumann, N. Chang, G. Morrison, W. Bodnar, Z. Zhang, J. M. Atkin, J. D. Surratt and B. J. Turpin, Per- and polyfluoroalkyl substances (PFASs) in airborne particulate matter (PM2.0) emitted during floor waxing: A pilot study, Atmos. Environ., 2022, 268, 118845 CrossRef CAS PubMed.
  64. M. J. Davern, Y. Hu, G. V. West, Y. Kim, M. H. Francini, G. C. Morrison, Y. Zhang, B. J. Turpin and J. D. Surratt, Online Iodide Chemical Ionization Mass Spectrometry (I-CIMS) Enables Occupational Inhalation Exposure Assessment of 6[thin space (1/6-em)]:[thin space (1/6-em)]2 Fluorotelomer Alcohol (6[thin space (1/6-em)]:[thin space (1/6-em)]2 FTOH) Emitted to Air during Floor Waxing, Environ. Sci. Technol. Lett., 2025, 12, 607–612 CrossRef CAS.
  65. C. M. A. Eichler, M. J. Davern, J. D. Surratt, G. C. Morrison and B. J. Turpin, Fluorotelomer alcohol (FTOH) emission rates from new and old rain jackets to air determined by iodide high-resolution chemical ionization mass spectrometry, Indoor Environ., 2024, 1, 100055 CrossRef.
  66. T. H. Bertram, J. R. Kimmel, T. A. Crisp, O. S. Ryder, R. L. N. Yatavelli, J. A. Thornton, M. J. Cubison, M. Gonin and D. R. Worsnop, A field-deployable, chemical ionization time-of-flight mass spectrometer, Atmos. Meas. Tech., 2011, 4, 1471–1479 CrossRef CAS.
  67. T. P. Riedel, J. R. Lang, M. J. Strynar, A. B. Lindstrom and J. H. Offenberg, Gas-Phase Detection of Fluorotelomer Alcohols and Other Oxygenated Per- and Polyfluoroalkyl Substances by Chemical Ionization Mass Spectrometry, Environ. Sci. Technol. Lett., 2019, 6, 289–293 CrossRef CAS PubMed.
  68. C. Ye, B. Yuan, Y. Lin, Z. Wang, W. Hu, T. Li, W. Chen, C. Wu, C. Wang, S. Huang, J. Qi, B. Wang, C. Wang, W. Song, X. Wang, E. Zheng, J. E. Krechmer, P. Ye, Z. Zhang, X. Wang, D. R. Worsnop and M. Shao, Chemical characterization of oxygenated organic compounds in the gas phase and particle phase using iodide CIMS with FIGAERO in urban air, Atmos. Chem. Phys., 2021, 21, 8455–8478 CrossRef CAS.
  69. X. Li, Y. Li, M. J. Lawler, J. Hao, J. N. Smith and J. Jiang, Composition of Ultrafine Particles in Urban Beijing: Measurement Using a Thermal Desorption Chemical Ionization Mass Spectrometer, Environ. Sci. Technol., 2021, 55, 2859–2868 CrossRef CAS PubMed.
  70. G. A. Novak, M. P. Vermeuel and T. H. Bertram, Simultaneous detection of ozone and nitrogen dioxide by oxygen anion chemical ionization mass spectrometry: a fast-time-response sensor suitable for eddy covariance measurements, Atmos. Meas. Tech., 2020, 13, 1887–1907 CrossRef CAS.
  71. M. P. Vermeuel, G. A. Novak, H. D. Alwe, D. D. Hughes, R. Kaleel, A. F. Dickens, D. Kenski, A. C. Czarnetzki, E. A. Stone, C. O. Stanier, R. B. Pierce, D. B. Millet and T. H. Bertram, Sensitivity of Ozone Production to NOx and VOC Along the Lake Michigan Coastline, J. Geophys. Res.: Atmos., 2019, 124, 10989–11006 CrossRef CAS.
  72. D. W. O'Sullivan, I. K. C. Silwal, A. S. McNeill, V. Treadaway and B. G. Heikes, Quantification of gas phase hydrogen peroxide and methyl peroxide in ambient air: Using atmospheric pressure chemical ionization mass spectrometry with O2−, and O2−(CO2) reagent ions, Int. J. Mass Spectrom., 2018, 424, 16–26 CrossRef.
  73. F. Heydebreck, J. Tang, Z. Xie and R. Ebinghaus, Emissions of Per- and Polyfluoroalkyl Substances in a Textile Manufacturing Plant in China and Their Relevance for Workers’ Exposure, Environ. Sci. Technol., 2016, 50, 10386–10396 CrossRef CAS PubMed.
  74. J. N. Rewerts, J. T. Morré, S. L. Massey Simonich and J. A. Field, In-Vial Extraction Large Volume Gas Chromatography Mass Spectrometry for Analysis of Volatile PFASs on Papers and Textiles, Environ. Sci. Technol., 2018, 52, 10609–10616 CrossRef CAS PubMed.
  75. B. H. Lee, F. D. Lopez-Hilfiker, C. Mohr, T. Kurtén, D. R. Worsnop and J. A. Thornton, An Iodide-Adduct High-Resolution Time-of-Flight Chemical-Ionization Mass Spectrometer: Application to Atmospheric Inorganic and Organic Compounds, Environ. Sci. Technol., 2014, 48, 6309–6317 CrossRef CAS PubMed.
  76. G. A. Novak, M. P. Vermeuel and T. H. Bertram, Simultaneous detection of ozone and nitrogen dioxide by oxygen anion chemical ionization mass spectrometry: a fast-time-response sensor suitable for eddy covariance measurements, Atmos. Meas. Tech., 2020, 13, 1887–1907 CrossRef CAS.
  77. L. C. Chen, Z. Yu and K. Hiraoka, Vapor phase detection of hydrogen peroxide with ambient sampling chemi/chemical ionization mass spectrometry, Anal. Methods, 2010, 2, 897–900 RSC.
  78. D. Bastien, D. Licina, L. Bourikas, S. Crosby, S. Gauthier, I. Mino-Rodriguez and C. Piselli, The impact of real-time carbon dioxide awareness on occupant behavior and ventilation rates in student dwellings, Energy Build., 2024, 310, 114132 CrossRef.
  79. C. J. Young, S. Joudan, Y. Tao, J. J. B. Wentzell and J. Liggio, High Time Resolution Ambient Observations of Gas-Phase Perfluoroalkyl Carboxylic Acids: Implications for Atmospheric Sources, Environ. Sci. Technol. Lett., 2024, 11, 1348–1354 CrossRef CAS.
  80. D. J. Gaffen and R. J. Ross, Climatology and Trends of U.S. Surface Humidity and Temperature.
  81. L. Jia, T. L. Delworth, X. Yang, W. Cooke, N. C. Johnson, L. Zhang, Y. Joh, F. Lu and C. McHugh, Seasonal predictions of summer compound humid heat extremes in the southeastern United States driven by sea surface temperatures, npj Clim. Atmos. Sci., 2024, 7, 180 CrossRef.
  82. Y. R. Lee, Y. Ji, D. J. Tanner and L. G. Huey, A low-activity ion source for measurement of atmospheric gases by chemical ionization mass spectrometry, Atmos. Meas. Tech., 2020, 13, 2473–2480 CrossRef CAS.
  83. M. Heinritzi, M. Simon, G. Steiner, A. C. Wagner, A. Kürten, A. Hansel and J. Curtius, Characterization of the mass-dependent transmission efficiency of a CIMS, Atmos. Meas. Tech., 2016, 9, 1449–1460 CrossRef CAS.
  84. Y. Ji, L. G. Huey, D. J. Tanner, Y. R. Lee, P. R. Veres, J. A. Neuman, Y. Wang and X. Wang, A vacuum ultraviolet ion source (VUV-IS) for iodide–chemical ionization mass spectrometry: a substitute for radioactive ion sources, Atmos. Meas. Tech., 2020, 13, 3683–3696 CrossRef CAS.

Footnote

Both authors contributed equally to the work and are noted as co-first authors.

This journal is © The Royal Society of Chemistry 2026
Click here to see how this site uses Cookies. View our privacy policy here.