Nanopipettes are a powerful solid-state nanopore platform with applications including single-particle characterization, surface mapping, and high-sensitivity analytical measurement in both bulk solutions and in localized spaces. While the capabilities of the platform have been aptly demonstrated, practical challenges associated with the nanopipette fabrication, filling, and surface modification remain. Progress and advancement to a high device readiness level are also hindered by low reproducibility and low throughput, as well as the consequent entry barrier for new research groups. This perspective provides the accumulated practical experience of our research group, forming a practical guide to overcoming the technical hurdles that are rarely discussed within the primary literature. We outline fabrication protocols, namely the importance of environmental conditions and instrument maintenance for enhanced reproducibility, as well as the desirable nanopipette geometries to facilitate ease of filling and surface modification. Filling procedures are discussed, and the interplay of surface wetting, capillary forces, and surface chemistry is used to understand how filling issues might be mitigated. Silanization of nanopipettes as a route for surface modification of the nanopipette interior is also discussed, with a particular focus on vapor-phase silanization. General design principles for surface grafting reactions are offered to maximize throughput and reproducibility. Throughout the perspective, the interconnected nature of the fabricated geometry, filling, and the modification pathways are highlighted, and this perspective is meant to facilitate the design of a holistic workflow where each stage is considered in tandem, facilitating higher throughput and more reproducible nanopipette research.
The desire for higher spatial precision, higher throughput, and the subsequent ability to use nanopores as scanning probes has led to the widespread adoption of nanopipettes.16–18 Nanopipettes consist of glass capillaries that taper to needle-like conical geometries terminating in a single, well-defined nanopore. Their rigid, needle-like geometry makes them ideally suited for scanning probe applications, such as Scanning Ion Conductance Microscopy (SICM),19–21 and enables their use as sensors capable of highly localized measurements through direct insertion into cells or other spatially constricted analyte environments.22–24 Although nanopipettes are uniquely suited for probe-based and localized measurement applications, nanopipettes retain full compatibility with bulk solution measurements and have found extensive use in analytical chemistry applications25 (Fig. 1).
The utility of nanopipettes stems from the unique electrokinetic behaviors that arise due to the nanoconfined ion transport at the tip region. In Coulter-type particle counting applications (resistive pulse sensing), the entry of particles into the nanoscale aperture displaces a significant fraction of the available ions, resulting in decreased pore conductance and a measurable ion current drop.26,27 In another analytical method, ion current rectification sensing exploits the overlapping electric double layers, which impose ion selectivity and generate a characteristic diode-like electrical response.28–31 The immobilization of charged analytes onto the nanopipette surface through analyte-specific probe molecules modulates the electric double layer and leads to a detectable change in the diode-like behavior, forming the basis for quantitative measurements.2 While the sensitivity and utility are enabled by these nanoconfined transport phenomena and the needle-like geometry, they also impose substantial practical challenges.
Despite the growing popularity of nanopipette-based research and the publication of many exciting applications and fundamental studies, the practical difficulty of working with nanopipettes remains a significant bottleneck for the field. Various practical challenges that are seldom discussed—including reproducible fabrication, pore blockage prevention, reliable filling procedures, and contamination control—hinder research efficiency in active laboratories and present a formidable entry barrier for new researchers. This perspective summarizes our accumulated experience in nanopipette fabrication, filling, and surface modification, providing detailed practical guidance for both established research groups as well as newcomers to the field. Through the inclusion of numerous troubleshooting approaches and technical recommendations, we aim to facilitate higher throughput experimental workflows and improved data quality across the nanopipette research community.
In our experience, two-line programs are the most reliable and reproducible for fabricating nanopipettes with a pore radius in the 20 nm to 150 nm range; however, two-line programs fabricating sub-10 nm pores have also been reported.32 In a two-line program, the first heating cycle melts the capillary, but the subsequent hard pull does not separate it into two nanopipettes. This first heating cycle's role is to shape the neck of the pipette. In the next heating cycle, the laser further melts and shapes the neck of the nanopipette, and then a hard pull separates the capillary to produce the nanopipettes. This two-line approach not only offers high reproducibility of the pore radius (as reproducible as 8% RSD), but also shorter and steeper neck tapers that facilitate subsequent filling. An example of such a program is provided in Table 1. This program can most easily be tuned to suit an individual instrument by adjusting the HEAT values until a heat-on time of approximately 4.6 seconds is obtained.
| Radius (nm) | Heat | Filament | Velocity | Delay | Pull | |
|---|---|---|---|---|---|---|
| ∼60 | Line 1 | 700 | 4 | 20 | 170 | 0 |
| Line 2 | 680 | 4 | 50 | 170 | 200 | |
| ∼6 | Line 1 | 750 | 4 | 40 | 135 | 180 |
| ∼250 | Line 1 | 575 | 3 | 60 | 128 | 100 |
One-line programs are most effective for fabricating nanopipettes with pore radii below 20 nm or above 150 nm. In our experience, one-line programs are less reproducible than two-line programs, yielding higher pore-to-pore variance (≈20% RSD). A one-line program shapes the neck of the nanopipette in a single heating event, typically producing longer and more gradual tapers that can make filling challenging. Table 1 also provides examples of one-line programs, and these can also be adjusted to an individual instrument by tuning the HEAT value until a consistent heat-on value is obtained.
![]() | ||
| Fig. 2 The improvement in reproducibility of the rectification ratio (|I(−0.4 V)/I(+0.4 V)|) as the pulling session continues. The moving standard deviation and average are calculated with a 6 pull window size (i.e., pull number 9 uses pull numbers 9–15). Three pull sessions of 15 pulls are used. The 6 nm program is used (Table 1). | ||
We have found that environmental factors exert a high influence on the pulling process, and fluctuations in the ambient temperature and humidity can greatly affect geometric reproducibility (especially between pulls at different times of the day, different days, and seasons). Consistent timing of fabrication, such as consistently fabricating nanopipettes only in the early morning, can help minimize the impact of these fluctuations.
The most straightforward approach to filling a nanopipette is backfilling, where a needle, such as MicroFil Flexible needle (World Precision Instruments, MF34G-5), is used to fill the solution into the interior of the nanopipette. Counterintuitively, optimal filling is achieved not by placing the solution directly into the taper of the nanopipette, but rather into the non-melted capillary section approximately 0.5 cm away from the onset of the taper (Fig. 3A). This will result in surface wetting and capillary forces drawing the solution into the narrow-tapered region, filling the nanopipette up from the tip-end in approximately 5–20 minutes, depending on taper geometry (Fig. 3B and C). Filling in this way typically leaves a single, large air bubble at the start of the taper that can be easily removed with a syringe (Fig. 3D). Conversely, direct injection into the taper often results in numerous small air bubbles trapped inside the narrow-tapered neck that are extremely difficult or impossible to remove with a syringe. An alternative approach is to mount the partially filled nanopipette into a 3D-printed centrifuge holder and use the centrifugal force to aid wetting and nanopipette filling.33 These filling methods work especially well if the capillary is filamented, but can also work with non-filamented capillaries (e.g., Sutter Instruments #100-70-7.5), depending on the geometry, as well as on the surface and the solution properties.
Although non-filamented capillaries can often be simply filled, sometimes the choice of solvent, electrolyte or the nanopipette geometry results in failure of the simple backfilling method described above. In this case, we have found the hot-plate method highly successful.34 In the hot-plate method, the solution is placed as close to the tapered neck as possible, and the nanopipettes are placed on a hot plate with their tips extending beyond the heating block (Fig. 4). The hot plate is set a few degrees below the solution's boiling point (i.e., 90 °C for aqueous solutions). As the tip extends beyond the heating block, and its surface-area-to-volume ratio increases as it tapers down into a nanopore, a temperature gradient is established along the length of the nanopipette, with the furthest and narrowest part of the tip being the coolest. This temperature gradient drives the repeated evaporation of solutions in the wider and hotter parts of the nanopipette, and condensation in the narrower and colder parts, gradually transporting solution towards the nanoscale aperture over a period of 30–60 minutes. Once the nanopipettes are completely filled, the electrolyte solution within the non-tapered regions of the nanopipette is replaced with fresh solution. This helps mitigate the issue of an evaporation-caused increase in the electrolyte concentration, especially in more volatile non-aqueous solvents. The hot-plate method is most effective for non-filamented capillaries, as the filament can disturb the evaporation and condensation, leading to small air bubbles trapped in the taper.
Persistent trapped air bubbles that cannot be resolved can be indicative of more fundamental problems associated with the nanopipette geometry. Tuning of the surface hydrophobicity/hydrophilicity (via surface functionalization), or adjusting the taper so that it is shorter and steeper, may be needed to facilitate good filling and mitigate the problem of stuck air bubbles.
Ethanol-based silanization suffers from substantial oligomerization/polymerization if anhydrous conditions cannot be maintained,51,52 which, in nanopipettes is difficult. While the oligomeric and polymeric by-products that form readily in ethanolic solutions can be easily removed in macroscale substrates by rinsing with acetic acid or methanol, the confined internal geometry of nanopipettes renders this cleaning process ineffective, and cleaning is limited to that undertaken via the diffusion of the undesired species out of the nanopipette. Consequently, we have found that ethanol-based silanization often yields sub-optimal results characterized by high device-to-device variance, nanopipette blockage, transient desorption of weakly attached silanes during measurements, and mechanical failure including cracking or complete breakage during drying procedures.50 These problems persist even in relatively large nanopipettes with diameters approaching 500 nm, highlighting the fundamental incompatibility between solution-phase silanization (which is widely and successfully utilized on macro-scale substrates) and nanopipette geometries.
We recommend the silanization of nanopipettes in the vapor-phase, which can circumvent many of the limitations inherent to solution-based approaches.50 In vapour-phase silanization, the pure liquid silane is placed at the bottom of a desiccator, and the nanopipettes are positioned on the desiccator plate above the liquid (Fig. 5A). Silane vaporization is achieved through the combined application of heating via a water bath and through the application of a vacuum using an in-house vacuum tap. Through a 60-minute exposure period, the vaporized silane molecules adsorb onto the nanopipettes, which are transferred to an oven for a 30-minute bake at 110 °C to complete the covalent surface condensation reaction.
Vapor-phase silanization offers several advantages for nanopipette modification. With the absence of backfilling, accidental breakages associated with solution removal, rinsing and handling are minimized. The purity and water content of the silane reagent are also of less importance, since the vapor-phase method naturally selects against oligomerized and polymerized species, as their higher molecular weights and lower vapor pressures limit their evaporation. This selectivity for monomeric silane results in more uniform silane layers that not only provide superior reproducibility (Fig. 5B) in onward applications but also favorable surface wetting properties that facilitate nanopipette filling. Vapor-phase silanization additionally enables high-throughput nanopipette silanization, with standard desiccators capable of accommodating approximately 80 nanopipettes for simultaneous silanization. The absence of the reactant solution also means that subsequent modification steps can be carried out immediately without the need to clean and dry the nanopipette.
The overarching principle governing the surface modification of nanopipettes is to minimize the amount of cleaning, drying, and solvent exchange. In the nanoconfined tip of a nanopipette, complete removal of unreacted starter reagents and side products is difficult, and the need for solvent exchange and drying steps, as well as the handling associated with these, can result in acute failures, low reproducibility, and low throughput. When designing modification strategies, reaction sequences should be designed to minimize/eliminate the need for cleaning, drying, and changing of solvents. Functionality should be imparted in as few modification steps as possible, with single step grafting representing the ideal scenario. If there is a requirement for complex functionality built in a multi-step assembly, it is strongly preferable to synthesize the complete functional molecule in bulk solution and then immobilize it onto the nanopipette surface in a single grafting step. This approach circumvents the significant challenges associated with multi-step synthesis within the confined nanopipette geometry while maintaining reaction control and enabling proper purification of synthetic intermediates. If multiple reaction steps must be carried out within the nanopipette itself, each step should ideally be performed in the same solvent system, and most preferably in the same solvent that will be used for subsequent electrochemical measurements. This, again, mitigates the need to exchange or remove the previous solvent so that it can be replaced with the solvent of the next reaction step.
When solvent changes are unavoidable, miscible solvents should be employed to enable smooth transitions between reaction steps without the formation of phase boundaries that can trap air bubbles or cause precipitation. Solvent exchange of miscible solvents is generally preferable to oven drying and re-filling, as the latter is more likely to lead to cracking, blockage, or precipitation. The hydrophobicity of intermediate and final surfaces must also be carefully considered throughout the modification process, as surfaces that are too hydrophobic can render nanopipettes completely unfillable with aqueous solutions, effectively terminating the synthetic pathway.
In situations where the ideal single-step, single-solvent approach cannot be implemented and multiple reaction steps, cleaning cycles, and solvent exchanges become necessary, the dip method offers a practical alternative to conventional backfilling procedures. In this approach, modification solutions are not backfilled into the nanopipette; instead, an empty nanopipette, or one filled with a compatible solvent, is simply dipped into the modification solution. The modification solution penetrates only as far as capillary forces or diffusion allow, creating a smaller reaction volume that can be more efficiently cleaned post-reaction through simple dipping in cleaning solvents. With a smaller reaction volume, solvent exchange is also faster and more thorough. We wish to highlight that the dipping method creates a transition zone between modified and unmodified surfaces within the nanopipette, and this can sometimes cause unusual ion transport behavior. This typically does not compromise the functionality of the nanopipette, as most applications exploit changes in transport properties.
In the case that a silane layer has been deposited prior to further modification (either through solution- or vapor-phase deposition), the temporal aspects of the post-silanization chemistry should also be considered. The reaction immediately succeeding silanization should be fast, as grafting a molecule onto the silane layer stabilizes it and prevents the silane layer's self-hydrolysis and desorption. When the modification reaction proceeds slowly, the silane layer is afforded sufficient time to partially detach from the glass surface, compromising the reproducibility of the overall modification process.50
The practical implementation of these principles can be illustrated through a specific example developed within our laboratory. For nanopipette sensors designed to detect trace metal contaminants,53 we synthesize the analyte-specific probe (a silyl cyclam) in bulk solution, then dissolve the pre-assembled probe in acetonitrile—the same solvent used for subsequent electrochemical characterization. This modification pathway succeeds because it employs a single grafting reaction, avoids drying and refilling procedures, and maintains solvent consistency throughout the process, thereby preventing precipitation of unreacted species.
These considerations collectively form a framework for the design of nanopipette surface modification strategies. While specific applications will dictate the exact surface grafting reaction scheme required, adherence to these principles significantly improves the likelihood of developing robust, reproducible modification protocols that can be successfully implemented across multiple nanopipettes with a high throughput (Fig. 6).
![]() | ||
| Fig. 6 A flowchart that illustrates the main design choices that will lead to more or less reliable and reproducible surface modification schemes. | ||
The nanopipette pore radius can be electrochemically characterized by measuring the conductance in a high electrolyte concentration solution (i.e., 100 mM potassium chloride) where the electric double layers collapse, and the electrochemical response is Ohmic rather than diode-like. The pore conductance is related to the pore radius:54
![]() | (1) |
As eqn (1) is only reliable when surface conductance is minimized due to the electric double layer collapse, sizing sub-10 nm nanopipettes can be more challenging. If the electrolyte concentration cannot be further increased, electron microscopy, such as Scanning Transmission Electron Microscopy, remains an alternative. However, even when the electric double layers are overlapping at lower electrolyte concentration, the pore conductance can still be used to screen the nanopipettes for outliers in fabrication and in surface modification. A cracked or broken nanopipette will show higher conductance, while a dewetted/blocked nanopipette will show a lower conductance than average. These outliers can be discarded.
In addition to the pore conductance, the shape of the current–voltage curve should also be used to screen outlier nanopipettes, as well as to diagnose issues. If the pore conductance is normal, but very high rectification is unexpectedly observed, the cause is often chemical contamination, to which nanopipettes, especially unmodified nanopipettes, are extremely sensitive (Fig. 7A). In our experience, cleanliness is incredibly important, and even briefly storing unmodified nanopipettes in a box that has previously stored silanized nanopipettes, or using a contaminated electrode, can cause irreversible contamination. Some current–voltage curves that show a higher conductance, but not so high as to indicate a completely broken off tip, as well as unusual rectification, we associate with the cracking or chipping of the nanopipette (Fig. 7B). A current–voltage curve in which the pore conductance is significantly lower than expected, but rectification is still observed, can be indicative of partial blockage/dewetting, commonly caused by the accumulation or precipitation of surface modifications and reagents (Fig. 7C). This is often observed after a nanopipette is modified, oven-dried, and refilled. Fig. 7D illustrates the loss or shifting of a glob of surface functionality, which suddenly and irreversibly changes the current–voltage curve mid-scan. Current–voltage curves where the pore conductance is low, and there is a sudden increase or decrease in current magnitude at moderate to high applied voltage, are usually indicative of an air bubble in the nanoscale aperture, which moves as the potential is applied (Fig. 7E). These bubbles can sometimes be removed by the application of a large applied voltage while cycling at a fast scan rate. Lastly, current–voltage curves showing high hysteresis are characteristic of soft and flexible surface modifications on the surface (Fig. 7F). Although this is not a fault that should be specifically avoided, it may be undesirable depending on the application. A thorough understanding of the shapes of the current–voltage curves is, however, crucial towards the tuning and optimization of nanopipettes at each of the stages discussed in this perspective, including fabrication, filling, and surface modification.
The reproducible fabrication of nanopipettes requires environmental control, proper instrument maintenance, and program optimization. While reproducible surface modification strategies should prioritize the minimization of cleaning, drying, and solvent change steps, with single-step grafting reactions maximizing reproducibility and throughput.
Most importantly, these practical challenges are interconnected, with fabrication parameters influencing the filling, the required surface modification scheme dictating handling requirements, and subsequent filling challenges having to be traced back to geometric and chemical issues established at earlier stages. Successfully working with nanopipettes requires a holistic approach where the design of the entire workflow, from initial fabrication through to final measurement, is considered together. This is facilitated by some of the diagnostic current–voltage curves we have offered, which can identify issues associated with each stage of working with nanopipettes.
Although individual laboratories will require specific modifications or alternative approaches for their particular applications and environmental conditions, the general strategies outlined in this perspective are universally applicable and can guide the development of a workflow, enabling researchers, established and new alike, to advance the field of nanopipettes with higher throughput and more reliable measurements.