Allyson R.
Tesky
a,
McKenna
Hicks
b,
Sujan
Aryal
c,
Brendan
Jones
c,
Julia
Molitor
d and
Anupama B.
Kaul
*ce
aDepartment of Biomedical Engineering, Denton, TX 76203, USA
bDepartment of Computer Science and Engineering, Denton, TX 76203, USA
cDepartment of Electrical Engineering, Denton, TX 76203, USA. E-mail: anupama.kaul@unt.edu
dDepartment of Mechanical Engineering, Denton, TX 76203, USA
eDepartment of Materials Science and Engineering, University of North Texas, Denton, TX 76203, USA
First published on 25th February 2025
Dihydrolevoglucosenone, commonly known as Cyrene, is a renewable and fully biodegradable cellulose-waste derived, environmentally friendly solvent, presenting a non-toxic alternative to N-methyl-2-pyrrolidone (NMP). Currently, solution-based processing of graphene and other similar van der Waals solids favor toxic solvents such as NMP, limiting their use for biosensing. However, with the use of Cyrene, bio-compatible printable devices are possible, and studies have already demonstrated its use in temperature and other biosensing methods through screen-printing. Screen-printing unfortunately often requires masks that constrain the minimum acquirable feature size to be above hundreds of centimeters and wastes material, adding to process complexity and cost. Conversely, inkjet-printing is an attractive alternative for the maskless patterning of hierarchically assembled structures, with micron length scales attainable. Graphene's high conductivity positions it ideally for long-wear sensors such as dry electrodes or respiration monitors. Here, we demonstrate the potential of Cyrene-based graphene inks through few-layer inkjet printing on flexible substrates for the first time, to produce non-toxic conductors toward a strain-mediated mechanism for biosensing, used to detect bodily motion for wearable electronics. The challenges overcome in this study include engineering ink chemistry and printing parameters such that Cyrene's relatively high viscosity compared to typical inkjet solvents, still allows for droplet ejection in a conventional material printer, yielding well-resolved clean line-edges in contrast to other solvents that exhibit diffuse line-edges possibly from stray droplets and ink-splashing. Temperature-dependent transport measurements on the inkjet-printed Cyrene-based graphene films showed the conductivity to be largely temperature-invariant but at lower temperatures below 100 K, conductivity decreased, likely as a result of increased inter-membrane separation arising from thermal contraction. Additionally, temperature-dependent Raman spectroscopy showed the red-shift in the G-band, 2D-band and D-band peaks, as temperature increased. By validating flexion motion detection of the proximal interphalangeal joint demonstrated in this study, our work is the first of its kind to successfully additively manufacture inkjet-printed Cyrene-based graphene strain sensors on flexible substrates for bio-sensing and wearables.
While the function of the electrodes remains the same across signal and electrode types, the configuration of the electrodes and accompanying circuit will vary. Electromyogram (EMG) signals are one of the simplest to capture as they only require a pair of bipolar electrodes connected to a differential amplifier circuit to measure the voltage between each point of contact. Moreover, EMG signals are easy to capture and recognize with limited equipment compared to electrocardiograms (ECG) or electroencephalograms (EEG); the latter techniques require more care towards electrode placement and more sensitive circuits with higher noise rejection. Thus, this positions EMG ideally as a platform for the prototypical demonstration of experimental electrodes in the exploration of human–machine interfaces for prosthetic control.6,9 While EMG principles are at the heart of detecting muscle activation and intensity, prosthetic systems relying on broader sensing and detection mechanisms, including strain-based sensors, may offer unique capabilities and solutions than those enabled by EMG principles alone. When considering locomotion of the body and designing soft robotics, it is important to consider the deformation of these structures, whether it is a stimulus used to direct the mechanics, or control against environmental stimuli.3,13 As such, low-profile, flexible strain sensors may assist in fine-tuning movements or developing feedback systems in prosthetic structures.
Graphene represents a suitable material for both strain sensors and low-profile dry electrodes. As a single sheet of carbon atoms arranged in a two-dimensional (2-D) honeycomb lattice, graphene has a high carrier mobility, strength, flexibility, biocompatibility, and versatility, making it an ideal material for nanoelectronics and more recently for biosensing.8,11,14–18 Furthermore, a wide variety of synthesis and manufacturing methods exist for the realization of graphene which include, micro-mechanical exfoliation, chemical vapor deposition (CVD), and liquid-phase exfoliation (LPE).14–16,19,20 Of these methods, LPE is of keen interest, especially for translational research, due to its potential scalability in industrial settings.16,19–21 Liquid-phase exfoliation is a method in which graphene is produced from dispersing graphite powder in a solvent and applying energy through ultrasonication or high-shear mixing to disrupt the inter-layer van der Waals bonds and produce a suspension of single-layer or few-layer membranes, known collectively as the ink.14,16,19 These inks may be produced with a variety of solvents, both aqueous and non-aqueous, but are most commonly made with N-methyl-2-pyrrolidone (NMP), dimethylformamide (DMF), and a combination of cyclohexanone and terpineol (C/T), with or without the binder ethyl cellulose (EC).12,14–24
Deposition methods for graphene inks vary as much as the manufacturing methods, including spray-coating,4,16 screen-printing,15,19,20 and inkjet-printing.15,25 Of these, the most common method is screen-printing due to its ease and ability to produce thick films in a single pass.20 However, inkjet-printing is a promising technique for high accuracy placement and is more suitable for rapid prototyping compared to either spray-coating or screen-printing, as it can produce complex, hierarchically-assembled structures without the need for masks or screens,14,26 though the former has a higher upfront cost than other LPE approaches.16,19 While graphene has shown promise towards wearable electronics and biomedical devices, traditional solvents used for graphene inks are known to be toxic, motivating a need for alternative solvents.
One of the most popular methods to circumvent the use of toxic solvents includes the suspension of hydrophilic graphene oxide (GO) flakes in non-toxic solvents that graphite is otherwise incompatible with, including water27,28 and alcohol,28,29 amongst others.30,31 After deposition, this suspension can be reduced via sonication or through thermal means,31 as well as electrochemical methods,29 yielding reduced graphene oxide (rGO). The properties can be enhanced further for bringing rGO closer to that of pristine graphene through the removal of oxygen-containing groups.29–31 However residual defects and functional groups in the rGO remain even after reduction, comprising its full potential,30 and thus the motivation to examine alternative solvents is still an active area of research.
Dihydrolevoglucosenone, commonly called Cyrene, is one such emergent alternative to DMF and NMP.16,19,22,32,33 As a biomass-derived, cellulose-based solvent produced by Circa since 2014,21 Cyrene is not only bio-degradable and renewable, but also non-toxic and capable of delivering higher concentrations of graphene within the ink dispersion compared to the similarly behaving dipolar aprotic solvent, NMP.19,22,32 Cyrene-based graphene ink has been demonstrated in both spray-coating16 and screen-printing,19,20 and while it has recently been used as an additive (4.3 wt%) for improving poly(3,4-ethylenedioxythiophene) and polystyrene sulphonate-based ink in inkjet printing,34 it has yet to be demonstrated as the primary solvent for inkjet-printed graphene. One of the reasons for this is its high viscosity,22 which makes it more difficult to print reliably, but this challenge can be overcome by engineering the printing parameters and printer settings, as we demonstrate here.
In this paper we have successfully used Cyrene-based graphene inks for inkjet-printed wearable biosensors with a focus on conductors which facilitate the development of experimental electrodes as well as on strain sensors, potentially useful in soft prostheses and respiration monitors.3,4 To establish a baseline for production of alternative biosensors, commercial electrodes were evaluated with a breadboard-built electromyogram for signal-to-noise ratio (SNR) determination—the details of which are available within the ESI† section. Prior to printing, characterization of our Cyrene ink and the ensuing films was conducted using Raman spectroscopy and temperature-dependent dc electronic transport measurements. This ink was then used to realize graphene-based strain sensors, which has been previously demonstrated on a variety of substrates including film, fiber, foam, and hydrogels,4 but not generally through inkjet printing. In the case of fiber-based strain sensors, conductive networks can either be created with wet-spinning or surface modification. These fiber-based graphene sensors can be worn directly on the body and even integrated into clothing, with some washability to allow for a breathable and reusable biosensor.4,35 However, biosensing with graphene is not limited to strain sensors alone, and it may also be used for electrochemical sensing of biomarkers such as glucose, folic acid, dopamine, or even cancer-related proteins.36 Other external sensors are also possible such as graphene-based electrodes, which can benefit similarly from flexible and reusable substrates. This work explores the additive manufacturing and inkjet printing of graphene with Cyrene on flexible polyimide (PI) substrates, where we have successfully demonstrated strain-dependent sensing with dry electrodes, overcoming some of the limitations of wet, gel-based electrodes commonly used in EMG.37 This was done through flexion motion detection of the proximal interphalangeal joint, where our work is the first of its kind to successfully additively manufacture inkjet-printed Cyrene-based graphene strain sensors on flexible substrates for bio-sensing and wearables. The environmentally friendly and non-toxic nature of this solvent has promise not only for wearables, but also in agricultural and food industries, where sensors need to be safe for potential contacts made to food supplies.
Though there are binder-free methods of formulating graphene inks,15 research with Cyrene is still in its infancy and has yet to demonstrate evidence towards one method or the other. Regardless, as binders can assist in maintaining graphene flake separation19 and have been reported in improving substrate adhesion,15 a compatible binder, cellulose acetate butyrate (CAB), was added to the Cyrene prior to ink formulation, detailed in Section 2.2. To characterize the long-range order and quality of resultant flakes, Raman spectroscopy of our graphene films was used here, similar to previous well-established studies.8,14,16,19–21,43–46 Raman spectroscopy data, conducted with a 532 nm excitation laser, are shown in Fig. 1(a) for the Cyrene-based graphene ink, and the results are compared to the spectra obtained using the conventional solvent C/T in a ratio of 70:
30 cyclohexanone
:
terpineol, 25 mg ml−1 graphite powder, and 20 mg ml−1 EC, with a laser wavelength of 633 nm. The inset in Fig. 1(a) depicts the graphene-Cyrene ink to be dark and optically absorbing. Provided there is no interfering photoluminescence, the Raman spectra for graphene with varying excitation lasers used in the visible region should yield largely similar results;47 though absolute intensities may vary with the lasers used, intensity ratios should reveal attributes of the actual material properties, independent of the laser. In this regard, the lower D/G intensity peak ratio, ∼0.34 for the Cyrene-graphene film, compared to that of the C/T synthesized graphene film where the ratio was calculated to be ∼0.40, indicates fewer defects in the former, as observed in prior studies as well.19,43,46 Additionally, the 2D/G-peak intensity ratio for the Cyrene ink was higher, ∼0.422 compared to the C/T ink, where it was ∼0.362. This is suggestive of Cyrene's greater effectiveness in exfoliating thinner membranes, although a ratio of two or three is representative of single-layer membranes and higher ratios may also be suggestive of doping.43Table 1 summarizes the parameters obtained using Raman analysis for both inks. Overall, our results corroborate the findings of a previous study19 that used a similar ink composition, both with and without CAB, which showed lower defects and no significant structural changes from the use of the binder.
Solvent used for graphene film | Peak positions (cm−1) | Ratios | |||
---|---|---|---|---|---|
D-Peak | G-Peak | 2D-Peak | I 2D/IG | I D/IG | |
C/T | 1335.1 | 1580.5 | 2674.7 | 0.36 | 0.41 |
Cyrene | 1350.9 | 1581.2 | 2710.0 | 0.42 | 0.34 |
Further characterization was performed on the Cyrene-based graphene through temperature-dependent Raman spectroscopy conducted from 77 to 800 K in atmospheric air, to decipher how the phonon spectra are modulated and possible interactions of phonons with defects, while at the same time shedding insights into their mechanical and thermal robustness in extreme environments. Being able to survive extremes in temperature may indicate potential use of our sensor in unusual thermal environments, such as deep space, hot planetary bodies, diving suits, and off-world machinery. To perform the temperature-dependent Raman measurements, the graphene ink was spin-coated onto an SiO2/Si substrate, where cooling was performed with liquid Nitrogen (LN2) down to 77 K, and the vibrational spectra were gathered in 50 K increments with heating through a heating stage up to 800 K. Four locations on the sample were selected for this measurement. Raman measurements were taken in each location by moving the laser spot to separate locations, before altering the temperature for the next set of temperature-dependent measurements. The sample was allowed to reach steady-state and equilibrate at its new temperature, prior to acquiring the data and this methodology was used from 77 K to 800 K in the rising cycle, and again from 800 K to 77 K in the temperature descending cycle. The data were averaged for the peak locations over the four sites during the upward-cycle and similarly for the downward-cycle. The difference in the Raman shift and the full-width at half-maximum (FWHM) of the peaks between the cooling or heating cycle appeared insignificant, indicating little hysteresis in the data—please see the ESI† section (Fig. S2 and S3) for further details. Fig. 1(b) displays the representative temperature-dependent Raman spectra, where both the intensity and the Raman shift are shown simultaneously. The evident red-shift in all the graphitic peaks as the temperature increased is seen here but visualized in more detail in Fig. 1(c)–(e). Moreover, one of the characteristic CAB peaks48 was also visible in the spectra of the graphene ink in the prior mentioned Cyrene/CAB study19 which appeared at 2937 cm−1. While this peak can still be observed in the temperature-dependent Raman characterization of our graphene films, the intensity of this peak at 2937 cm−1 is far lower than those attributed to the dominant graphitic G-, 2D- and D-peaks, as seen in Fig. 1(b). Additionally, when Raman characterization was conducted on the CAB film alone, as detailed in Fig. S5 of the ESI† section, no discernable peak-shifting is observed, despite its far reduced intensity, nor is there a marked change in intensity values as a function of temperature; this is unlike the observations made on the graphitic peaks. These point to the possible degradation and decomposition of the CAB during the film annealing process, as intended.
Though there appeared to be some nonlinear effects seen in the D- and 2D-peak locations with temperature, the trend nonetheless was similar for the red-shift observed. Although this is expected,49–52 as red-shifting is generally associated with material expansion upon heating, the FWHM, pictured in Fig. 1(f), was largely temperature invariant save for the 2D-peak, affirming the results observed in other studies49 which have seen a gradual increase in the FWHM of the 2D-peak above 300 K. The invariance we find here may be uniquely attributed to the use of Cyrene as the dispersing media, though more detailed experiments are necessary to more concretely validate this. With no noticeable difference in peak shape observed over the wide range of temperatures examined, from 77 K to 800 K, there appears to be little indication that mechanical robustness and thermal behavior through adhesion to the substrate is hindered in any noticeable way. This further suggests the promise of such emergent materials as the sensing media in extreme thermal environments.
Prior to proceeding with inkjet printing, further characterization of the ink was necessary to ensure it was appropriate and lent itself to printing due to concerns about its high surface tension and viscosity.22 Invoking the Reynolds (Re) and Weber (We) numbers, which are used to theorize the behavior of a fluid while printing, a parameter Z is introduced that is inversely related to the Ohnesorge (Oh) number.53–57 Together, these fluidic parameters are used to evaluate jettability according to,
![]() | (1) |
Low concentrations of binders are adequate to ensure ink stability, and concentrations higher than 1 mg mL−1 (CAB-to-Cyrene) can negatively affect dynamic or sheet resistance.19 For these reasons, CAB was added in increasing amounts, starting at low weight percentages and the resultant viscosity was then measured at ambient temperature (20 °C). Measurements for η, as in Fig. 2(b), were taken using a rheometer equipped with a small sample adapter, starting with pure Cyrene and progressively increasing the CAB content. In all recorded viscosity measurements, it took several seconds for the rheometer measurement to stabilize, evident from the inset data in Fig. 2(b), particularly for the more viscous formulations. As expected, however, higher concentrations of the binder also resulted in higher steady-state viscosity.
Because a low concentration of CAB is desirable for ensuring higher conductivity, the optimized wt% for CAB in Cyrene was below the 0.08 wt% mark, or roughly 1 mg mL−1 CAB for a volume of 10 mL Cyrene.21 The resulting Cyrene mixture had η ∼ 13–14 cP which is a higher η compared to commonly used inkjet solvents such as NMP, where η ∼ 1.7 cP,14 but it is still within the target range of 8–15 cP for good printing characteristics.24 Furthermore, inks between 2 to 50 cP are also eligible for printing but may require additional accommodations for even jetting.53 Once it was verified that Cyrene may be suitable for printing with the addition of CAB, an ink was developed via horn-tip sonication, the progress of which can be seen in Fig. 2(c). Here we clearly observe the final ink to be homogeneously black from the presence of graphene flakes and evenly dispersed without signs of sedimentation in the right-most vial. On the other hand, pure Cyrene is transparent (left-most vial) and the addition of CAB renders the solution darker in tint (middle-vial). Adding graphite powder to the ink slightly increases η but this change is largely negligible. More notable is the change in surface tension from the addition of the graphite powder and binder. This can be estimated easily via the capillary rise method, using the following,
![]() | (2) |
Jetting of the produced ink may be viewed in Fig. 2(d), where (i) shows a cross-sectional view of the nozzles prior to drop formation, and (ii)–(iv) illustrate the progression of these drops as they exit the nozzle towards the drop watcher pad. While these droplets have moderate uniformity among them, it can be noted that they are at a relatively higher v due to the length of their “tails”.40 Shorter tails with more noticeable droplet heads are preferred, but through our experiments this was not possible at all due to the need for higher jetting voltages given the high γ (72.5 dynes per cm45) of the Cyrene, and therefore a lower We. Though lower jetting voltages may be preferred, these however, did not produce droplets since there was insufficient energy for drop formation.40,41
High γ associated with low We make it difficult for the piezo-actuated nozzles to displace the air required for priming the jetting mechanism. To fulfill these priming requirements, typical values of γ range from 20 to 40 dynes per cm,53 but γ outside of window can be accounted for by changing ρ or v of the ink during formulation or through printer settings, respectively. The impact of η and γ on printed lines can be noted in Fig. 2(e) between printed C/T-based ink at (i) low, and (ii) high-magnifications, respectively, and similarly for printed Cyrene-based ink at (iii) low, and (iii) high-magnifications, respectively. The C/T ink has a lower η and γ, and there is evidence for splashing in (i), in contrast to Cyrene with a far cleaner edge profile where minimal stray droplets or splashing is evident in (iii). To determine if these defects are attributable to the Z-value discussed in eqn (1), it is worthwhile to estimate these values for the C/T and Cyrene-based inks.
Although the γ of our ink is unknown, using the γ values available for these solvents allows for a reasonable estimation of the Oh number, and therefore Z. The measured η for the solvent/binder concentrations used for these inks were 5.85 cP and 13.78 cP for C/T + EC and Cyrene + CAB, respectively. Including the approximate γ of 20.9 dynes per cm and 47.8 dynes per cm with the associated ρ of the solution and the nozzle width of 20 μm, the Z values are estimated to be ∼3.50 and 2.54, respectively. Interestingly, a splattering effect is visible on the edge of the C/T-based ink in (ii) due to the presence of satellite droplets despite 1 < Z = 4.27 < 10, compared to the Cyrene-based ink which is jetted relatively evenly on the SiO2/Si substrate, and also contains a Z-value within the optimal range.
The C/T-based ink may have a high enough Re that despite an adequate Z-value, it has reached the “splashing region”.54,56 Adding a higher concentration of EC to the C/T ink to increase its viscosity could improve these results by reducing the Re. Certainly, Cyrene has a distinct advantage in this regard, as the amount of binder necessary is much lower compared to C/T, where the latter may require as much as 3 wt% EC to reach the recommended viscosity; this is exacerbated in the case of DMF and NMP, which may need up to 4 and 6 wt% EC, respectively, to reach this same range.54 A scanning electron microscope (SEM) image of the Cyrene-based graphene film on SiO2/Si, comprising of just three printed passes, is seen in Fig. 2(f), which further reveals its dense morphology, despite the relatively few passes used. In the case of C/T and other conventional solvents, typically 20–50 passes are needed, which not only increases device fabrication time, but also cost and printer instrumentation wear, challenges that are overcome here through our Cyrene-based ink and its unique attributes.
In addition to the Cyrene-based sample printed on SiO2/Si, another sample was printed on polyimide (PI) substrates using the same conditions. Polyimide is of interest due to both its flexibility, and thus potential for strain sensing, as well as its biocompatibility that may allow it to be used in bio-electrode fabrication.8,11,17,18 Both samples were evaluated using a Lakeshore Cryogenic probe stage for temperature-dependent dc transport measurements, as seen by the current–voltage (I–V) characteristics captured in Fig. 3(a). The data in Fig. 3(a)-(i) provides the I–V sweeps from −20 V to +20 V over a temperature range of 4 K to 350 K, with compliance set at ∼1 mA. Closer inspection into the differences in current from 18–20 V in Fig. 3(a)-(ii), shows current saturation for graphene on the SiO2/Si substrate at 350 K, but the film printed on PI had lower conductivity and did not reach current saturation at any temperature in the range tested. Here we tabulate the dynamic resistance, RDynamic, as the ratio of the instantaneous V to I, defined as RDynamic = dV/dI. The calculated RDynamic is illustrated in Fig. 3(b). It was notable that at very low temperatures (∼ 4 K), the RDynamic was much higher in both samples, but it stabilized at around 100 K on the SiO2/Si substrate, while this stabilization temperature was much lower, ∼10 K for the graphene on the PI substrate. The transport data is shown over a larger range of temperatures in the ESI† section, Fig. S4.
![]() | ||
Fig. 3 (a)-(i) The typical I–V characteristic of inkjet-printed Cyrene-graphene film measured in vacuum at four different temperatures on an SiO2/Si substrate (top) and on a PI film (bottom) as the substrate; magnified view is seen in (ii) from 18 V to 20 V. The current was higher for the graphene film printed on the SiO2/Si substrate compared to the film printed on a PI substrate. (b) Calculated RDynamic of graphene on SiO2/Si and PI substrates as a function of temperature, where the RDynamic decreased as temperature increased up to ∼ 10 K for the PI substrate, and up to ∼ 100 K for the SiO2/Si substrate. Beyond these temperatures, the RDynamic was largely temperature invariant. Error bars represent the 1-Sigma (1-σ) standard deviation of the resistance calculated across the entire voltage sweep in (a). (c) Process flow of the PVD deposited Ag electrodes on a PI substrate which serve as the contacts to the flexible strain-sensing devices: (i) bare PI film placed adjacent to the shadow mask; (ii) deposited Ag/Ti pads on PI; (iii) magnified view of a single Ag/Ti electrode pair used for printing and the strain-sensing experiments. (d) Printed graphene on Ag/Ti electrodes in (i) concave, (ii) relaxed, and (iii) convex configurations, respectively. (e) Temporal response of the current change in printed device, where increased current was observed during the concave configuration, while decreased current was observed in the convex configuration, at a bias of ∼5 V. (f) Overlap region viewed under an optical microscope between the Ag/Ti electrodes and the overlying printed graphene film at (i) high-magnification, and (ii) low-magnification, which shows the graphene film is largely unperturbed as a result of the instrumentation contacts placed there during electrical characterization of the devices. (g) Calculated RDynamic of the printed device with one measurement taken in all three configurations at varying time intervals, which revealed the RDynamic to be largely in-variant with time up to the 24 hours tested, unlike the somewhat time-dependent response measured equivalently for the wet-electrode devices tested in Fig. S1(e) (ESI†). (h) Single voltage sweep of the printed device during the static bending configurations displayed in (d). |
The behavior of LPE graphene trending towards lower resistances at higher temperatures has also been observed by others.59 Regardless, the samples were still conducting at extremely low temperatures, where RDynamic was ∼53.4 kΩ at 4 K on SiO2/Si substrates, which decreased to ∼20 kΩ at 300 K. The higher porosity of solution-exfoliated graphene may yield a greater flake-to-flake separation at lower temperatures due to thermally-induced mechanical contraction. This increases the energy barrier for carriers for intra-flake transport, and thus the RDynamic increases at low temperatures. This is a characteristic feature of conductors whose transport is governed by a defect-mediated hopping mechanism. Nevertheless, additional measurements should be done on multiple samples under the same conditions to allow comparison between substrates with significance.
Static and dynamic bending measurements were conducted using a Keithley 2400 source meter interfaced with LabView for data acquisition to record the current sweep over time, for a given bias voltage V. The printed graphene resistive network was bent in various directions, as shown in Fig. 3(d). A change in resistance was observed upon bending the device in the concave direction (printed lines in compression), as shown in Fig. 3(d)-(i), which resulted in a lower resistance than the device at rest, Fig. 3(d)-(ii), and a convex configuration, shown in Fig. 3(d)-(iii) (printed lines in tension), where a higher resistance was seen compared to the device in the relaxed state. The compressed case likely results in reduced membrane-to-membrane separation between the graphene flakes, thereby reducing the resistance, or equivalently increasing conductivity.3 The corresponding time-dependent current response for the three cases, concave, relaxed, and convex, are seen in Fig. 3(e), where the data show the higher current levels for the concave case (film in compression) compared to the convex case (film in tension). The interface between the Ag/Ti and the printed lines after bending can be observed in Fig. 3(f), where no visible damage is observed to the printed graphene film as a result of the instrumentation contacts placed there during the electrical characterization of the devices, validated through the low and high-magnification optical microscopy images in (i) and (ii). Next, the device was left attached to the measurement system and allowed to sit in ambient conditions and measurements were taken at equal intervals, analogous to the 24 hour electrode test conducted in Fig. S1(d) in the ESI† section.
The long-term stability of these devices is first suggested by the data in Fig. 3(g), which indicates they are relatively uniform under similar conditions, as opposed to the gel electrodes’ slightly bimodal response over time, detailed in the ESI† section. Effects seen during the dynamic bending tests are corroborated with the static I–V response captured in Fig. 3(h) corresponding to the bending configurations shown in Fig. 3(d). Though the response agrees with the dynamic results of Fig. 3(e), it is interesting to note that the current during the relaxed state appeared to depend on its previous state, as the current immediately after the concave state was higher than that after the convex state; this may imply there is a mild memory effect, possibly arising from the slightly porous nature of the films, where the spacing between the randomly oriented membranes may not be exactly the same as the initial state. Since these devices were manually bent, there may be small variations in user technique such as dissimilar bending angles, minor twisting of the device, or variations in the amount of time the device spent in each position throughout the 24 hours that could artificially create differences in the RDynamic but for the purposes of this preliminary demonstration study, the data suggest these electrodes to show a fairly stable response with time.
After the strain sensing response was verified for our Cyrene-graphene ink-jet printed film, a 50 mm long conducting line was printed on a 60 mm strip of PI film without metal conducting pads for ease of wearability of the electrodes. Here we validate flexion motion detection of the proximal interphalangeal joint, using additively manufactured inkjet-printed Cyrene-based graphene strain sensors on flexible substrates, a first such study. This extended PI strain sensor was formed using the same printer settings as the 15 × 15 mm2 electrode pads, and the printed lines here were 2 mm in width. Current measurements were taken along the length of the extended substrate at varying probe distances from 5 mm up to 50 mm, where once again the RDynamic, was calculated for each of the spatially spaced points used for the probing measurements, as shown by the data in Fig. 4(a). Unsurprisingly, the RDynamic demonstrated a largely linear response in accordance with Ohm's Law. This substrate could also be bent in similar configurations to the sensor pertaining to Fig. 4. Specifically, Fig. 4(b)-(i)–(iii) correspond one-to-one, with the data obtained in Fig. 3(d)-(i)–(iii), for concave (film in compression), relaxed, and convex (film in tension), respectively.
This elongated resistive network allowed the strain sensor to be worn on the finger, where flexible 30-gauge wires were attached to either end of the printed graphene with double-sided carbon tape and then further adhered using insulating electrical tape. It was then fastened to the finger with pliant Teflon tape. Measurements of the finger in its straight position (Fig. 4(c)) or bent (Fig. 4(d)), correspond equivalently to the strain sensor in its relaxed and convex (film in tension) configurations, respectively. The device further validated previous findings during the static bending tests conducted from 0 V to 10 V, which yielded a decrease in current when the finger was bent, as shown in Fig. 4(e).
These effects were more pronounced through our manual time-dependent dynamic tests, represented by the data in Fig. 4(f), where the current drops during bending. Incidentally, these dynamic tests resulted in a more pronounced drop in current than was evident through the static bending tests in Fig. 4(e).
In order to develop a more robust system to gauge the sensor's ruggedness and repeatability, a Cycling Automator was designed, pictured schematically in Fig. 4(g). This instrumentation utilizes an Arduino UNO microcontroller with a VL6180 Time-of-Flight sensor for measurement of the bending angle, and a servo motor to actuate the bending mechanism, where the circuit board connections are shown on the right. This system enabled repeated bending cycles of the strain sensor to be conducted over the course of minutes to hours, keeping the bending conditions identical, while utilizing a 30-gauge flexible wire and light-weight carbon-tape to form an intimate contact, and to overcome the limitations posed by the use of rigid and heavy alligator clips which may interfere with the sample's intrinsic strain-induced sensing mechanism. One cycle of the actuation is defined by a period that starts when the assembly begins to move together, includes a momentary pause, and then moves back to rest, with the strain sensor restored to a neutral position. These cycles may be repeated as many times as desired, and as such, 100 convex and concave bending cycles were performed on a truncated strain sensor at a maximum 90° bending angle, demonstrated by the envelope of cyclical switching data in Fig. 4(h).
It is interesting to note that while the sensor returned to the same level during the resting portion of the cycles, the magnitude of these levels depends on which configuration the sensor was being bent in. As can be seen more clearly in the inset of Fig. 4(h), when the sensor was in the concave position, corresponding to compression, the current would increase relative to the sensor's relaxed position. However, in the convex position, corresponding to tension of the graphene resistive network, the current would decrease relative to relaxation. To ensure that the results of these tests were a product of the graphene resistive network, the mechanism was also tested with a reference strip of the bare PI substrate, without any film deposited. Although there was a small amount of current present during movement for the bare PI substrate (Fig. 4(i)), the peak-to-peak amplitude of this current was negligible (few hundred nanoamps), and there was no noticeable pattern when there was no connection at all. While these results further corroborate our previous findings through manual testing, we also demonstrate the robustness of our sensors through repeated bending cycles performed with our Automator, which should position these devices well for further translational applications towards sensor arrays and networks for human–machine interfaces.
Although the cycling automator marks an improvement over manual bending, additional measures can be adopted in future tests to make the contacts more robust for even longer duration testing. Due to their heavy weight and rigidity, alligator clips present problems either by slipping or by creating localized areas of compression and tension when the device is bent in the opposite configuration. Additionally, the large effective area and hardness of the clips may lead to damage at the contacts for longer duration testing, though we did not observe this over the 100 cycles tested in our current experiments. Alternative contact strategies may include the use printed circuit board (PCB) spring contacts—although many are not designed for substrates as thin as PI and may need spacers to allow a connection. Other alternatives may include the adaptation of a flexible printed circuit connector that would traditionally also be PCB-mounted.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d4tc04049j |
This journal is © The Royal Society of Chemistry 2025 |