Maria M. Morima,
Patrick Rupper
b,
Stefanie Altenried
c,
Adrian Meiera,
René M. Rossi
a,
Qun Ren
c,
Luciano F. Boesel
a and
Giorgia Giovannini
*a
aEmpa, Swiss Federal Laboratories for Materials Science and Technology, Laboratory for Biomimetic Membranes and Textiles, Lerchenfeldstrasse 5, CH-9014, St. Gallen, Switzerland. E-mail: giorgia.giovannini@marionegri.it
bEmpa, Swiss Federal Laboratories for Materials Science and Technology, Laboratory for Advanced Fibers, Lerchenfeldstrasse 5, CH-9014 St. Gallen, Switzerland
cEmpa, Swiss Federal Laboratories for Materials Science and Technology, Laboratory for Biointerfaces, Lerchenfeldstrasse 5, St. Gallen 9014, Switzerland
First published on 20th August 2025
Urinary tract infections (UTIs) are among the most common bacterial infections, affecting approximately 150 million people worldwide each year. Currently, diagnosis is often made using culture-based methods, which are time-consuming and therefore costly. Point-of-care (POC) devices have the potential to provide a rapid and accurate UTI diagnosis, thereby improving treatment efficacy. In this work, we developed a fast, specific, and accurate colorimetric sensor capable of indirectly detecting Enterococcus faecalis in urine samples by targeting its metabolite, L-lactate. The sensing probe consists of silica nanoparticles (SNP) loaded with MnO2 (MnSNP), functionalised on the surface with the enzyme lactate oxidase (Lac@MnSNP). The sensor enables both qualitative analysis – through a visible colour change – and quantitative analysis, using spectroscopy. The morphology and composition of the probe were characterised at each synthesis step, confirming that the incorporation of MnO2 into SNP and subsequent enzyme functionalisation did not alter nanoparticle morphology. Lac@MnSNP demonstrated responsiveness to L-lactate, showing a linear decrease in signal up to 50 μM of the analyte, with a limit of detection of 31 μM. The probe successfully detected E. faecalis in artificial urine medium and in complex samples at concentrations as low as 103 CFU mL−1 within 5 h. These results demonstrate the potential of this probe for fast, accurate, and lactate-specific diagnoses.
Clinical diagnosis of UTIs relies on urine culture, which enables the identification of the causative bacterial pathogen. However, as culture-based methods are often time-consuming, antimicrobial treatments are often prescribed without clear evidence of infection, contributing to the rise in antimicrobial resistance (AMR).3
Point-of-care (POC) devices offer substantial advantages over conventional diagnostic methods by enabling rapid and accurate diagnosis, which in turn supports timely and appropriate treatment. According to the World Health Organization (WHO), effective POC diagnostics should meet the ASSURED criteria: affordable, sensitive, specific, user-friendly, rapid and robust, equipment-free, and derivable to end-users.4 A widely used POC device commercially available for UTI diagnosis is the dipstick assay. Dipsticks enable rapid, routine qualitative analysis of urine samples, typically through colorimetric changes that indicate parameters such as pH.5 For more advanced, multisensing quantitative analysis, larger strip-based instruments – such as the Urisys 1100® Urine Analyzer or the LAURA® Semi-Automated Urine Strip Reader – are utilised. Despite the accessibility and ease of use, urine test strips have notable limitations in terms of detection range and sensitivity. They typically assess physical and chemical properties of urine (e.g. pH), which are not directly indicative of bacterial infection. As a result, they are more suitable for preliminary screening rather than definitive diagnosis.6 Furthermore, urinary composition can be influenced by metabolic, molecular, and genetic factors, potentially leading to misleading or inconclusive results7.
Consequently, clinical practice continues to rely on laboratory-based techniques for bacterial identification and antimicrobial susceptibility testing. While these methods are accurate, they are time-consuming and delay the initiation of targeted treatment, thereby contributing to the spread of AMR.8
Given the substantial socioeconomic burden of UTIs and their contribution to the development of AMR, there is an urgent need to develop advanced POC diagnostic tools that enable rapid, accurate, and pathogen-specific detection to guide prompt and effective treatment. The POC device currently used, the dipstick, uses pH as a biomarker for the identification of UTI. However, pH is not necessarily correlated with infection. For a more accurate diagnosis, a POC device should be able to target bacterial-related biomarkers. Although UTIs can be caused by multiple bacterial species, E. faecalis is an important cause in certain patient groups, such as those with catheter-associated or complicated UTIs.2 E. faecalis is a commensal, L-lactate-producing bacterium commonly found in the gastrointestinal tract, but it can become opportunistic when it colonises other organs, leading to infections such as UTIs. Its ability to form biofilms make these infections particularly difficult to treat, especially in catheterised patients, often resulting in a chronic infection.8,9 Moreover, due to its ability to develop AMR and multidrug-resistance, E. faecalis is listed in the “WHO Bacterial Priority Pathogens List (2024)” and is recognised as the third leading cause of nosocomial infections.10–14
L-Lactate is a versatile biomarker widely used in prognostic, diagnostic, and monitoring tools across a range of clinical conditions (e.g. infectious diseases, sepsis, trauma, and perinatal conditions)15 and physiological states (e.g. monitoring athletic fatigue and workload).16,17 L-Lactate-based sensors are most commonly designed for health monitoring applications, being also designed for food applications. Several POC devices for lactate testing are already commercially available, and they typically rely on colorimetric pH-based detection, since lactate accumulation leads to acidification of physiological fluids.18,19 A more detailed review about lactate-based sensors can be found in ref. 15, 16, 20 and 21.
Colorimetric sensors offer several advantages for POC diagnostics, including a visible colour change that can be detected by the naked eye. These sensors are simple, specific, and sensitive, supporting the development of rapid, cost-effective, user-friendly, and affordable diagnostic tools that do not require complex or expensive equipment.22–26 Direct detection of L-lactate has also been achieved using a limited number of recently commercialised electrochemical devices.27,28 These systems, which directly target L-lactate via enzymatic reactions, are widely used in clinical settings due to their rapid turnaround time and ease of use. However, a key limitation is the relatively high limit of detection (LOD), typically in the millimolar (mM) range, which restricts their applicability in diagnostic contexts that require detection in the micromolar (μM) range.27,28
In this work, we developed a rapid and cost-efficient colour-based detection mechanism for the indirect identification of E. faecalis in urine samples, by targeting its metabolic by-product, L-lactate. The sensing probe (Lac@MnSNP) consists of silica nanoparticles loaded with manganese dioxide (MnSNP) and functionalised with the enzyme lactate oxidase. In the presence of L-lactate – produced by E. faecalis – a cascade of enzymatic reactions is triggered, leading to the accumulation of hydrogen peroxide (H2O2). This accumulation reduces MnO2 to MnO/Mn2+, resulting in a visible colour change in the suspension from brownish to colourless. In addition to qualitative detection by the naked eye, the reduction of MnO2 can be quantitatively monitored by measuring the decrease in absorbance at 400 nm. The probe was characterised at each stage of synthesis in terms of morphology, composition, and sensitivity to both L-lactate and E. faecalis. The sensor presented here offers a simple, user-friendly, and affordable POC diagnostic tool, with the potential to enable advanced, bacteria-specific detection for accurate diagnosis of UTIs caused by E. faecalis.
![]() | ||
Fig. 1 (A) Schematic representation of the detection mechanism of the colorimetric probe for lactate-producing bacteria, such as E. faecalis. (B) Synthetic procedure for the preparation of Lac@MnSNP. |
L-Lactate, released by the bacteria as a metabolic by-product of glucose fermentation, is oxidised by lactate oxidase on the surface of Lac@MnSNP. This enzymatic reaction produces H2O2, which subsequently reacts with the MnO2 embedded within the silica core. The reduction of MnO2 to MnO/Mn2+ results in the loss of the characteristic brown colour and the dissolution of the MnO2 particles. The visible discolouration of the suspension serves as an indirect indicator of the presence of E. faecalis in urine, suggesting a potential UTI.
The sensor is based on two essential components: (i) lactate oxidase, which serves as the sensing element by specifically recognising the target analyte L-lactate, a product of bacterial carbohydrate metabolism; and (ii) MnO2 nanoparticles, which function as the transducer element, converting the biochemical detection of L-lactate into a measurable optical signal.
SNPs were selected as the structural scaffold between sensing and transducer elements due to their favourable surface chemistry and optical inertness, properties that facilitate this effective integration. Specifically, SNPs enable the stable immobilisation of lactate oxidase on their surface, promoting the localised production of H2O2 in proximity to the MnO2 nanoparticles embedded within the silica core.
MnO2 was synthesised by reducing potassium permanganate (KMnO4) with ammonium hydroxide (NH4OH), as shown in eqn (1) (Fig. 1(B)-i). The formation of MnO2 molecules was monitored by measuring the absorbance spectrum of the reaction over time. As shown in Fig. S1, the characteristic absorbance peaks of KMnO4 at approximately 545 nm gradually disappeared, while a broad peak emerged around 350 nm, indicative of MnO2 formation. These spectral changes confirmed the presence of MnO2 after 40 minutes. The progression of the reaction was also visually evident from the colour change of the suspension – from purple (KMnO4) to brown (MnO2).
2KMnO4 + 2NH4OH → 2MnO2 + N2 + 2KOH + 4H2O | (1) |
The aqueous suspension of the synthesised MnO2 particles was employed as the water phase for the synthesis of core–shell silica nanoparticles using the microemulsion method (Fig. 1(B)-ii). Tetraethyl orthosilicate (TEOS) was used exclusively to form the silica core, while the shell was constructed using a combination of TEOS, a phosphate-containing silane – 3-(trihydroxysilyl)propyl methylphosphonate (THPMP) – and an amino-derived, (3-aminopropyl)trimethoxysilane (APTMS) silicate derivative. The amino groups present on the surface of the resulting MnSNP were subsequently utilised to covalently conjugate the enzyme lactate oxidase through succinic anhydride linker (Fig. 1(B)-iii). The extent of enzyme immobilisation on the nanoparticle surface was determined by quantifying the residual protein content in the supernatant following particle purification by centrifugation. The calculated functionalisation was 83%.
The synthesised MnO2, SNP, MnSNP, and Lac@MnSNP were morphologically characterised by dynamic light scattering (DLS) and transmission electron microscopy (TEM). The optimised synthesis of MnSNP produced a homogeneous suspension of nanoparticles with an average diameter of 158 ± 3 nm and a polydispersity index (PDI) of 0.17 ± 0.05 (Table 1). These results indicate that the incorporation of MnO2 did not significantly alter the morphology of SNPs, aside from a slight reduction in size compared to SNPs (184 ± 2 nm size, PDI: 0.15 ± 0.02). DLS analysis of MnO2 was inconclusive due to the small particle size and their tendency to form agglomerates, as confirmed by TEM imaging. TEM images of MnSNP demonstrated that the incorporation of MnO2 did not alter the overall morphology of the silica particles (Fig. 2). In the case of Lac@MnSNPs, DLS revealed a significant increase in hydrodynamic diameter compared to MnSNPs, with a measured size of 1438 ± 52 nm and a PDI of 0.30 ± 0.02. However, TEM analysis showed that the core particle size remained small, with MnSNPs and Lac@MnSNPs exhibiting diameters of 153 ± 1 nm and 105 ± 1 nm, respectively. It was observed that the functionalisation of MnSNPs to form Lac@MnSNPs resulted in an increase in surface charge from –34 ± 1 mV (MnSNP) to −13 ± 1 mV (Lac@MnSNP). As the zeta potential approaches neutrality, the electrostatic repulsive forces between particles are reduced, promoting nanoparticle agglomeration in suspension, as indicated by DLS measurements. Nevertheless, TEM analysis confirmed that the nanoparticles retained their individual morphology despite the tendency to aggregate.
DLS | TEM | |||
---|---|---|---|---|
Diameter (Z-average) [nm] | Polydispersity index (PDI) | Zeta potential (ζ) [mV] | Diameter (Z-average) [nm] | |
SNP | 184 ± 2 | 0.15 ± 0.02 | −25 ± 1 | 87 ± 1 |
MnSNP | 158 ± 3 | 0.17 ± 0.05 | −34 ± 1 | 153 ± 1 |
Lac@MnSNP | 1438 ± 52 | 0.30 ± 0.02 | −13 ± 1 | 105 ± 1 |
The elemental composition was examined using energy-dispersive X-ray (EDX) analysis. The detection of Mn in the MnSNP sample (Table 2) confirmed the incorporation of MnO2 particles in the silica nanoparticle matrix. In addition, following treatment with H2O2, the MnO2 particles were isolated via centrifugation, and EDX analysis revealed that the MnO2 particles were completely degraded into soluble Mn2+ cations, as no pellet was observed after centrifugation of the MnO2 sample. In contrast, EDX analysis of the pellet obtained after 2 hours of incubating MnSNP with 50 μM of H2O2 confirmed the degradation of MnO2 while indicating the stability of the silica matrix. The MnSNP retained their morphology after treatment, with occasional signs of Ostwald ripening (Fig. 3), a phenomenon that is typically observed in silica nanoparticles in aqueous environments,29,30 where there is a structural rearrangement of the silica matrix forming the particles, and it is not indicative of degradation. Lac@MnSNP exhibited signs of Ostwald ripening from the time of synthesis (Fig. 3 and Fig. S2).
H2O2 treat. | MnO2 | MnSNP | ||
---|---|---|---|---|
Before | After | Before | After | |
Mn | 33 ± 2 | — | 7 ± 2 | — |
Si | — | — | 28 ± 2 | 25 ± 2 |
O | 51 ± 4 | — | 54 ± 4 | 57 ± 1 |
C | 13 ± 0 | — | 9 ± 0 | 16 ± 1 |
Extra | 3 ± 1 | — | 2 ± 1 | 2 ± 0 |
Notably, after 7 days of storage in deionised (DI) water, MnSNP exhibited a similar appearance as Lac@MnSNP, which had also been functionalised and stored under the same conditions (Fig. S3). This observation suggests the presence of Ostwald ripening, which occurred predominantly in smaller particles that probably did not contain MnO2. Despite the structural rearrangement, both MnSNP and Lac@MnSNP retained their overall morphology and remained effective in detecting L-lactate, respectively. The stability of Lac@MnSNP upon to lactate was further confirmed by DLS analysis. Both MnSNP and Lac@MnSNP exhibited changes in hydrodynamic size upon interaction with the analyte (Table S2). Nevertheless, both nanoparticle types remained stable in the presence of the analyte, maintaining their morphology, as confirmed by TEM (Fig. 3).
In addition to the elemental analysis, X-ray photoelectron spectroscopy (XPS) was performed to further investigate the chemical composition of the probe and any changes induced by H2O2 treatment. Measurements taken at different positions across the samples indicated a homogenous elemental distribution. Furthermore, in repeated measurements, at the sample position, under the same experimental conditions, neither the X-ray radiation nor the electrons/ions from the charge neutraliser system had any detectable effect on the recorded spectra.
High-resolution XPS elemental scans were performed on MnO2, MnSNPs (treated and untreated with H2O2), and SNPs (Fig. 4). Trace amounts of potassium (∼5 at%) were detected (Fig. 4(A)), likely originating from either residual KOH formed as a by-product of KMnO4 reduction or incomplete reduction of KMnO4 (see eqn (1)). The MnO2 sample exhibited characteristic peaks for oxygen (Fig. 4(B)) and manganese (Fig. 4(C)), along with minor carbon contamination.
XPS analysis also confirmed that the chemical composition and structure of SNPs remained unchanged following H2O2 treatment (Table S1 and Fig. 4(B) and (D)). The silicon-to-oxygen (Si:
O) atomic ratio in MnSNP was 23
:
59 before treatment and 26
:
60 after treatment, with no significant changes in the spectral position or shape of the O 1s and Si 2p peaks. A minor nitrogen signal (<1 at%) was detected, confirming successful functionalisation of the SNP surface with amino groups from APTMS (Fig. 1(B)-iii). The elemental concentrations derived from the XPS survey scans (Fig. 4(F)) are summarised in Table S1. Reference spectra for various manganese compounds have been extensively studied in the literature.31,32 In this work, the binding energy maximum of the Mn 2p3/2 peak for MnO2 was observed at 642.5 eV (Fig. 4(C)), which is consistent with reported values – for example, Ilton et al. reported the Mn2p3/2 peak for MnO2 at approximately 642 eV.32 Additionally, the peak exhibited a tail extending towards lower binding energies (Fig. 4(E)), which can be attributed to the presence of Mn3+.31 No signals corresponding to metallic Mn or Mn2+ were detected, as these would appear at binding energies ≤641 eV.
Fitting of the Mn 2p3/2 peak, using literature-based multiplet splitting components for Mn4+ and Mn3+,32 revealed a mixture of Mn4+ (43%) and Mn3+ (57%) in the synthesised MnO2. In comparison, the commercially available MnO2 used as a reference contained only ∼30% Mn3+, in agreement with previous reports.32 The higher Mn3+ content in our synthesised material is attributed to the in-house production of MnO2 via the KMnO4 reduction method. Consistent with EDX results, XPS confirmed the presence of surface MnO2 following its incorporation into the SNPs. Fig. 4(E) shows a weak Mn signal (0.4 at%, Table S1) in MnSNPs. Moreover, the shape and the binding energy of the Mn 2p spectrum remained unchanged between the MnO2 and MnSNP samples (Fig. 4(E)), indicating that the chemical structure of MnO2 was preserved during incorporation. After treatment with H2O2, the Mn signal disappeared from the XPS spectra (Fig. 4(C) and (E)), consistent with the transformation of MnO2 into soluble Mn2+ and its subsequent dissolution, corroborating the EDX findings.
A stability test was conducted to compare the behaviour of MnSNPs (1 mg mL−1) and MnO2 in deionised water after 7 days of storage. MnO2 was tested at two different concentrations: 0.01 mg mL−1, which visually matched the colour of MnSNPs at 1 mg mL−1, and 0.025 mg mL−1, which exhibited similar absorbance values to MnSNPs at the same concentration. After 7 days, MnSNPs showed no change in the absorbance of the supernatant, whereas MnO2 samples exhibited a decrease in absorbance to 0.98 and 0.88 at 0.01 mg mL−1 and 0.025 mg mL−1, respectively, being the days of 0.025 mg mL−1 MnO2 statistically different (p-value < 0.05) (Fig. S3A). Additionally, MnO2 displayed a higher degree of agglomeration compared to MnSNPs during storage (Fig. S3B). Evidence of matrix rearrangement in MnSNPs was also observed on day 7 (Fig. S3C), likely due to Ostwald ripening – a common phenomenon in silica-based systems. Overall, the stability test confirmed that embedding MnO2 within the silica matrix not only prevents the spontaneous dissolution observed in free MnO2 but also significantly reduces particle agglomeration (Fig. S3B).
The selection of SNPs was guided by two main factors: (i) the ease with which their surface can be functionalised, enabling straightforward conjugation of the enzyme lactate oxidase; and (ii) the optical properties of silica. The optimal concentration of MnSNPs was determined by varying both amount of MnO2 loaded within the SNPs (Fig. S4A) and the concentration of MnSNPs used in the assay (Fig. S4B), in order to achieve suitable absorbance for detecting the desired range of L-lactate. This study confirmed that higher availability of MnO2 – whether through increased MnO2 loading or higher MnSNP concentration – resulted in greater H2O2 detection, but with reduced sensitivity. Specifically, we observed that using high concentrations of MnO2 particles diminished detection sensitivity due to a shielding effect, whereby small signal variations become undetectable. Based on NMR results, bacteria-producing lactate in urine samples is expected to range from 59 to 241 μg mL−1, corresponding to bacterial concentrations of 103 and 105 CFU mL−1, respectively.33 Therefore, we set a working concentration of MnSNPs at 2 mg mL−1, which corresponds to 26.9 μg mL−1 of MnO2 particles, as determined spectrophotometrically using a calibration curve. This concentration was chosen to enhance the sensitivity of the method.
The detection efficacy of the probe was initially evaluated by treating 2 mg mL−1 MnSNPs with varying concentrations of H2O2 ranging from 0 to 100 μM. An incubation time of 4 hours was selected, as it yielded a more stable signal (Fig. S5). Results are reported as relative absorbance, calculated as the ratio of the absorbance intensity at each H2O2 concentration to that of the untreated sample [In/I0]. The absorbance at 400 nm decreased linearly with increasing H2O2 concentration, reaching 0.78 relative absorbance at 50 μM of H2O2 (R2 = 0.94), and further declining slightly to 0.67 at 100 μM (Fig. 5(B)). The limit of detection (LOD) was determined from the linear portion of the response curve, yielding an LOD of 26 μM for H2O2. To validate the detection mechanism, 2 mg mL−1 of Lac@MnSNP was tested against L-lactate. A faster response was observed in the range of 50–100 μM L-lactate when the nanoparticles were functionalised with higher concentrations of lactate oxidase (Fig. S6). Two enzyme concentrations were evaluated: 50 μg mL−1 and 100 μg mL−1. After 2 h of incubation, the probe containing 50 μg mL−1 of enzyme exhibited a stabilised signal of 0.9 at 25 μM L-lactate. In contrast, the probe with 100 μg mL−1 of enzyme showed a linear decrease in absorbance, reaching 0.71 at 50 μM L-lactate (R2 = 0.81). Extending the incubation time to 4 h enhanced the responsiveness of the 50 μg mL−1 enzyme probe, which reached a plateau at 75 μM L-lactate with a signal decrease to 0.64. A similar trend was observed for the 100 μg mL−1 enzyme probe, which exhibited a linear decrease to 0.70 at 100 μM L-lactate (R2 = 0.94). Overall, the absorbance signal of Lac@MnSNP decreased with increasing L-lactate concentration, particularly up to 50 μM (Fig. 5(C)). At 100 μg mL−1 of enzyme concentration, the signal dropped to 0.70, with a calculated LOD of 31 μM (R2 = 0.94).
Lactate oxidase catalyses the conversion of L-lactate to pyruvate in a 1:
1 ratio, producing one molecule of H2O2 per L-lactate molecule oxidised. Both MnSNP and Lac@MnSNP exhibited a similar response, with a linear decrease in absorbance up to 50 μM of H2O2 and L-lactate, respectively. The corresponding LOD values were 26 μM for H2O2 and 31 μM for L-lactate. These results confirm the successful catalysis of L-lactate by Lac@MnSNP and its subsequent quantification via indirect detection of the generated H2O2. These results confirm the detection mechanism of the designed probe, in which the presence of L-lactate leads to a decrease in absorbance, indirectly indicating the presence of E. faecalis. MnSNP was tested in PBS at different pH (i.e. pH 5, 6, 7 and 8) to mimic the different conditions of urine. The result showed that pH does not interfere with the detection efficacy of the probe further confirming the suitability for the proposed application (Fig. S7).
Given that urine is a complex matrix containing various components that could interfere with the probe's performance, MnSNP and Lac@MnSNP were tested in an artificial urine medium (AUM) to assess potential matrix effects. The absorbance spectra of AUM was analysed to see if it could interfere with the probe (Fig. S8). AUM did not present an absorbance peak at 400 nm wavelength, indicating that the probe is suitable for measurements in AUM and urine samples (Fig. S8). MnSNP and Lac@MnSNP were evaluated in DI water and AUM, using H2O2 and L-lactate as analytes, respectively (Fig. 5(D) and (E)). MnSNP exhibited similar responsiveness in both media, with a linear decrease in absorbance to 0.78 at 50 μM H2O2 and a further slight decrease to 0.67 at 100 μM. However, the sensitivity of the probe was reduced in AUM, with a LOD of 58 μM (R2 = 0.70), compared to 26 μM in DI water.
Similarly, Lac@MnSNP showed comparable behaviour in both media, reaching a saturation point of 0.68 at 50 μM L-lactate in AUM, with an LOD of 47 μM (R2 = 0.86), compared to 31 μM in DI water. These findings indicate that the complex composition of AUM affects the sensitivity of the probe. Nevertheless, the probe remains suitable for the intended application, as bacteria are known to produce 59–241 μg mL−1 of L-lactate during UTIs,33 while normal urine does not typically contain L-lactate.34,35
As a proof-of-concept, the responsiveness of the probe was evaluated using the lactate-producing bacteria E. faecalis (Fig. 5(F)). The absorbance signal of 2 mg mL−1 Lac@MnSNP decreased linearly with increasing concentration of E. faecalis, reaching saturation at 103 CFU mL−1. Further increases to 104 and 105 CFU mL−1 resulted in higher absorbance signals, likely due to exceeding the probe's detection range. It is important to note that UTIs are typically diagnosed when bacterial concentrations of ≥105 CFU mL−1 for adults and 103–104 CFU mL−1 in children.43,44 Depending on the diagnostic requirements, using higher concentrations of the probe could enable detection of bacterial loads above 103 CFU mL−1. In contrast to Lac@MnSNP, MnSNP did not exhibit a decrease in signal upon exposure to E. faecalis (Fig. 5(F)), confirming that the probe responds to lactate produced by the bacteria, and it is not affected by bacteria-related molecules, as no signal variation was measured using the enzyme-free probe (MnSNP). These results demonstrate the potential of the probe to discriminate between bacterial species through the detection of lactate produced by bacteria at concentrations up to 103 CFU mL−1 (p-value < 0.05), thereby contributing to improved diagnostics. Fig. 5(F) confirms that the probe responds to lactate produced by the bacteria, as no signal variation was measured when using the enzyme-free probe (MnSNP) in the presence of E. faecalis. The relation between the bacterial concentration and L-lactate detected was calculated using a calibration curve of the probe, when tested in water (Fig. S9). Concentrations of L-lactate of 56 μM were calculated for bacterial concentrations up to 103 CFU mL−1. The calculated concentrations of L-lactate are above the LOD of the probe (31 μM), showing the potential of the probe in detecting bacterial load until 103 CFU mL−1. The probe was also tested against the non-lactate-producing bacteria Staphylococcus aureus (S. aureus) (Fig. S10). In contrast to E. faecalis, Lac@MnSNP did not exhibit a decrease in signal upon exposure to S. aureus. These results demonstrate the potential of the probe to discriminate between bacterial species through the detection of lactate produced by bacteria. The proposed probe was compared with various L-lactate probes from the literature (Table 3) based on sensor type, incubation time, LOD, sample type, and application. As it is possible to observe, L-lactate-based sensors can be from different types of sensors, electrochemical,36–38 fluorescent,39–41 and colorimetric,42 as is the case of this work presented, being colorimetric the least common type. Lac@MnSNP is composed of SNP and MnO2 which are considered non expensive materials and is produced by a simple reverse microemulsion method. The majority of the sensors found in the literature use expensive materials such as multi-walled carbon nanotubes or metal–organic frameworks38,39,41 or use complex synthesis methods,36,41,42 turning the fabrication process more complex and less cost-effective. When comparing electrochemical and optical sensors, each type has its own advantages and disadvantages, electrochemical sensors may present a lower LOD, however, they also present more costly and complex fabrication processes.36,38,45 When comparing fluorescent and colorimetric sensor probes, both solutions can be quantitative and qualitative, however, fluorescent sensors need the use of an external light source such as an UV lamp. Regarding the health applications, L-lactate is used as a biomarker for monitoring the oxygenation levels of tissue,38 infections37 and other diseases such as diabetes or liver diseases,36,42 being sweat the most common sample used.38,39,41 In addition to this work, a colorimetric sensor that uses urine as a sample to determine the presence of a liver disease was found, presenting a LOD of 0.63 mM, roughly 20 times higher than the LOD of this work (31 μM).42 The literature includes an L-lactate based sensor for the detection of infection in wounds post-surgery. The authors propose a wireless implantable electrochemical sensor for the identification of infection in orthopeadics, presenting an incubation time of 5 h and a LOD of 66 μM,37 contrary to Lac@MnSNP, which presented a similar incubation time but a LOD of 31 μM. When comparing with the different sensors found in the literature, Lac@MnSNP presents a more cost-effective solution that has a LOD in the working range of the application. Regarding the sensors that use urine as a sample and sensors for infection-based application, Lac@MnSNP is one of the few examples of lactate-sensitive probe for the diagnosis of UTI in urine samples,37,42 which is user friendly, cost-effective and presents a qualitative and quantitative solution with a low LOD and simple readout.
Type of sensor | Incubation time | LODL-lactate | Type of sample | Application | Ref. |
---|---|---|---|---|---|
Electrochemical | 1200 s | 1.12 × 10−1 fM | Urine and blood | Diabetes | 36 |
Electrochemical | 5 h | 66 μM | Tissue | Infection in wounds post-surgery | 37 |
Electrochemical | 5 s | 55 pM | Sweat | Tissue oxygenation | 38 |
Fluorescence | 5 min | 0.52 mM | Sweat | Health monitoring | 39 |
Fluorescence | 20 min | 30 μM | Serum | POC chiral analysis in complexed biological samples | 40 |
Fluorescence | N/A | 0.091 μM | Milk and Sweat | Sports medicine, clinical environments, food industry | 41 |
Colorimetric | N/A | 0.63 mM | Urine | Liver disease | 42 |
Colorimetric | 4 h | 31 μM | Urine | UTI | This work |
Powder samples were pressed onto indium foil to create a flat, contiguous surface and mounted onto a stainless steel holder using double-sided adhesive tape. Minimal signal was detected from the indium substrate (<0.1 at%). Survey scans (0–1100 eV) were acquired with a step size of 0.8 eV, an acquisition time of 160 ms per data point, and an analyser pass energy of 187.85 eV.
High-resolution elemental spectra were collected for C 1s and K 2p (278–298 eV), O 1s (523–543 eV), Si 2p (94–114 eV), and Mn 2p (630–665 eV), using a step size of 0.125 eV, analyser pass energy of 29.35 eV, and acquisition times of 1.92 s (C, O, Si) and 4.8 s (Mn) time per data point. The energy resolution (FWHM, full width at half maximum height) was 2.2 eV for survey scans and 0.7 eV for high-resolution scans, as determined from the Ag 3d5/2 photoemission line. Total acquisition times were approximately 4 min for survey scans and 40 min for the five combined high-resolution scans.
Randomly selected spots on the samples were analysed using a micro-focused X-ray beam (100 μm diameter, 25 W at 15 kV). The 180° spherical capacitor energy analyser was operated in the fixed analyser transmission (FAT) mode. To compensate for potential sample charging, dual beam charge neutralisation was employed using a flux of low-energy electrons (1.3 eV) combined with very low-energy Ar+ ions (10 eV).
The binding energy of the C 1s peak was referenced to 284.8 eV, corresponding to aliphatic carbon (C–C) from adventitious carbon contamination.31 Atomic concentrations were calculated using CasaXPS software (version 2.3.16, Casa Software Ltd, Teignmouth, UK). Spectral fitting was performed using a mixed Gaussian–Lorentzian product function (70% Gaussian, 30% Lorentzian) to curve fit the XPS spectra (least-squares fitting routine) for determining the chemical bonding states of the various elements. A Shirley-type background was subtracted from the XPS peak areas.
Quantification was carried out using tabulated PHI sensitivity factors,46 corrected for the instrument's transmission function and analyser asymmetry (accounting for the different angle between the X-ray source and analyser). Further detail on the XPS system and methodology have been previously published.47
Subsequently, 50 μL of TEOS was added, followed by 40 μL of THPMP after 20 minutes, and 10 μL of APTES after 5 minutes. The mixture was left for 24 hours.
Purification was carried out via three cycles of centrifugation at 15294 × g for 8 minutes, followed by redispersion of the pellet using a probe sonicator (10% amplitude, 3 seconds). The final MnSNPs were suspended in deionised water (DI water) and stored at 4 °C.
As a control, pristine SNPs were synthesised using the same procedure, substituting the MnO2 suspension with 0.48 mL of DI water during the microemulsion formulation step.
The resulting functionalised particles were incubated with varying concentrations of lactate oxidase (50 and 100 μg mL−1) for 12 hours at room temperature under shaking at 41 × g. After incubation, the particles were collected by centrifugation (6797 × g, 8 minutes) and gently redispersed in Milli-Q water using a vortex mixer. The resulting Lac@MnSNPs were either used immediately for assay or stored at 4 °C for a maximum of 24 hours.
On the respective day of analysis, the samples were centrifuged at 6797 × g for 8 minutes. The absorbance at 400 nm was measured for both the supernatant and resuspended pellet to evaluate the stability and retention of optical properties over time.
Absorbance spectra were recorded after 2 h and 4 h of incubation using a microplate reader. Prior to measurement, samples were gently mixed using a pipette. Absorbance was measured at 400 nm.
In a 96-well plate, 50 μL of Lac@MnSNP and 50 μL L-lactate were added in each well, resulting in final concentrations of 2 mg mL−1 for Lac@MnSNP and 3.13, 6.25, 12.5, 25, 50, 75, and 100 μM for L-lactate. Control wells were prepared by mixing Lac@MnSNP with 50 μL of Mili-Q water.
Absorbance spectra were recorded after 2 and 4 h of incubation using a microplate reader. Prior to measurement, samples were gently mixed using a pipette. Absorbance was measured at 400 nm.
MnSNP and Lac@MnSNP stock solutions (4 mg mL−1) were prepared in the same HEPES + glucose medium. In a 96-well plate, 50 μL of nanoparticle solution and 50 μL of diluted bacterial suspension were combined, resulting in a final nanoparticle concentration of 2 mg mL−1. Control wells include bacteria alone and nanoparticles alone in HEPES + glucose. Bacteria growth and MnO2 reduction were monitored by measuring absorbance at 600 nm and 400 nm, respectively, using a microplate reader.49 Measurements were taken hourly over an 8-hour period, with gentle mixing before each read to resuspend aggregates and ensure accurate readings.
The limit of detection (LOD) was determined from the linear region of the calibration curves using a linear region model. LOD was calculated using the formula: 3.3σ/slope, where σ represents the standard deviation of the blank measurements.
This journal is © The Royal Society of Chemistry 2025 |