A highly accurate and selective non-enzymatic glucose sensor based on a Fe2O3/CuFe2O4/graphene nanoplatelet nanocomposite
Received
7th October 2025
, Accepted 21st October 2025
First published on 29th October 2025
Abstract
In this study, a sensitive and accurate non-enzymatic glucose sensor based on a Fe2O3/CuFe2O4/graphene nanoplatelet (GNPs) nanocomposite was performed. The nanocomposite was synthesized via a facile hydrothermal method and characterized using FTIR, XRD, SEM, Raman, XPS, cyclic voltammetry (CV) and electrochemical impedance spectroscopy techniques. The integration of Fe2O3 and CuFe2O4 with graphene nanoplatelets (GNPs) provides a synergistic effect, enhanced surface area, high electrochemical conductivity and improved catalytic activity toward glucose oxidation. Electrochemical measurements using CV and chronoamperometry demonstrated excellent sensing performance, with high sensitivity of 62.4 μA mM−1 cm−2, a lower of detection limit of 0.049 μM (S/N = 3) and a widely linear detection range from 5–75 μM and 75–13
000 μM. The sensor exhibited outstanding selectivity against common interfering species (e.g., dopamine, uric acid and ascorbic acid), along with good stability and reproducibility. These findings suggest that the Fe2O3/CuFe2O4/GNPs nanocomposite is a promising applicant for glucose non-enzymatic sensing in clinical and biomedical applications.
1. Introduction
Glucose monitoring has a critical role in the management of diabetes mellitus, a chronic metabolic disorder affecting millions worldwide. Accurate and reliable glucose levels detection is essential for diagnosis and daily glucose control. Among various glucose sensing technologies, electrochemical sensing63,65,73,77 has gained significant consideration due to its rapid response,66,67 ease of operation,39,68 high sensitivity62,67 and low cost.61,64 Traditionally, enzymatic glucose sensors based on glucose oxidase (GOx) have dominated the market. However, these sensors suffer from several limitations, including poor stability, high production cost, and susceptibility to different conditions such as pH, temperature, and humidity.64 In recent years, non-enzymatic electrochemical glucose sensors have emerged as promising alternatives. These sensors depend on direct glucose oxidation on the surface of the electrode, offering enhanced stability and reproducibility without the need for biological components. The performance of such sensors is dependent on the electrocatalytic properties of the electrode materials.
Transition metal oxides and spinel ferrites have attracted significant interest due to their brilliant redox activity,27,69 chemical stability,70 and catalytic potential.71,72 Among them, Fe2O3 and CuFe2O4 are particularly appealing owing to their abundance, environmental friendliness, and strong catalytic behavior toward glucose oxidation. However, their inherent low electrical conductivity limits their standalone performance. To address this challenge, incorporating conductive carbon-based materials such as graphene can significantly enhance electron transport and increase active surface area. CuFe2O4, a transition metal oxide, has garnered significant attention as an electrocatalyst in the development of electrochemical sensors, particularly for non-enzymatic glucose detection, owing to its excellent structural tunability, electrocatalytic activity, and compatibility with a wide range of nanocomposite configurations. CuFe2O4 employed in sensing other biomolecules. For example, Atacan 5 developed a glassy carbon electrode modified with a CuFe2O4/rGO decorated with gold nanoparticles, which approved a good sensitivity and a broad linear range for detecting L-cysteine hematite (α-Fe2O3) is extensively utilized in sensor development due to its favorable semiconducting characteristics, excellent chemical stability, and notable catalytic activity. Its natural abundance and environmental friendliness further enhance its suitability, particularly for applications in medical diagnostics and environmental monitoring.108 In non-enzymatic glucose detection, Fe2O3 offers a stable, low-toxicity alternative to enzyme-based sensors. Nanostructures like porous nanorods and nanoparticles, particularly when combined with materials such as rGO or g-C3N4, enhance performance by increasing surface area and improving electron transfer. These composites allow for sensitive glucose detection at physiological pH, even in complex samples.41,106,115 Electrochemical methods like CV and DPV are used to fine-tune these sensors, supporting their potential for real-world, portable applications.18,94 Meanwhile, graphene nanoplatelets (GNPs) have emerged as key electrode modifiers in sensor for glucose detection, primarily due to their highly surface area and excellent electron transport properties. These features significantly improve glucose adsorption and electrocatalytic oxidation efficiency.28,49,89 π–π interactions between glucose molecules and GNPs enhance molecular adsorption at the electrode surface, leading to lower detection limits and improved sensitivity.89 Graphene-based nanocomposites offer several advantages for enzyme-free glucose detection in alkaline media, driven by graphene's exceptional electrical conductivity and physicochemical stability.55 Strategic incorporation of graphene with metal oxides or other conductive nanomaterials can further amplify sensor performance. Notable examples include graphene–CuO, graphene sheet/graphene nanoribbon/nickel nanoparticles (GS/GNR/Ni), and graphene–ZnO–SnO2 nanocomposites, all of which leverage synergistic effects to boost electrocatalytic activity and electron transfer rates.31,47,48 Furthermore, graphene derivatives like GNPs are known to enhance sensor responsiveness by promoting fast electron mobility and efficient glucose molecule adsorption, thus significantly improving detection thresholds and response times.10,22 This study highlights the potential of Fe2O3/CuFe2O4/GNPs nanocomposites as efficient and reliable electrode materials for non-enzymatic glucose sensing applications. The synergistic interaction between Fe2O3 and CuFe2O4, combined with the exceptional conductivity and large surface area of GNPs, results in a highly sensitive and stable electrode material. The nanocomposite was synthesized via a simple hydrothermal route and thoroughly characterized using structural and electrochemical techniques. The resulting sensor exhibits excellent electrocatalytic performance, with brilliant sensitivity, lower detection limit, widely linear range and good selectivity against common interfering species.
2. Experimental
2.1. Materials and methods
Monohydrate of α-D-lactose (Lac), D-(+)-mannose (Man), D-(+)-galactose (Gal) and D-(+)-glucose were brought from Fluka (USA). Fructose, maltose, sucrose, urea, uric acid (UA), L-ascorbic acid (AA) were obtained from Sigma (USA). Monopotassium phosphate (KH2PO4), sodium citrate (Na3C6H5O7·2H2O), sodium chloride (NaCl), sodium hydroxide (NaOH), copper(II) chloride dihydrate, potassium permanganate (KMnO4) (Panreac), potassium ferricyanide (K3[Fe (CN)6]), FeCl3, (CuCl2·2H2O), and potassium ferrocyanide (K4[Fe(CN)6]·3H2O), graphite powder 20 < μm were obtained from Sigma Aldrich. Graphene nanoplatelets 5 microns wide × 6–8 nm thick × was purchased Strem chemicals.
2.2. Apparatus and instrumentation
Fourier-transform infrared spectroscopy (FTIR) using a PerkinElmer instrument with a range of 4000–400 cm−1 used for the structure and composition characterization of the samples. X-ray diffraction (XRD) analysis was produced using an XPERT diffractometer with Cu Kα radiation (λ = 0.15450 nm) to identify the crystalline phases presented in the nanocomposites fabrication. To analyze the microstructure and surface morphology of the nanocomposite, a Thermo Scientific Prisma E Scanning Electron Microscope (SEM) was used, operating at an accelerating voltage of 10 kV. The SEM was coupled with an energy-dispersive X-ray spectroscopy (EDS) detector to enable elemental analysis, X-ray photoelectron spectroscopy (XPS) JEOL, modèl JPS-9030 was employed to investigate the surface composition and oxidation states of the sample. The wide-scan spectra were collected using an Mg Kα radiation source (1253.6 eV) with a step size of 1.0 eV and a pass energy of 50 eV, covering a binding energy range of (0–1000) eV to identify the elemental composition of the samples. Raman spectroscopy measurements for the nanocomposite were conducted at room temperature using Raman Spectroscopy (Confocal Raman microscope, WITech, alpha-300R, excitation laser 532 nm and Laser power 1 mW) for the nanomaterials characterizations. Each spectrum was recorded in the range of 100–4000 cm−1 and the sample was examined under multiple magnifications to capture detailed structural information. Electrochemical data were obtained from a Biologic EcLab Sp-300 potentiostat at the Laboratory of Sensors, Instrumentation, and Processes (LCIP), Khenchela University, with a three-electrode cell configuration. Reference electrode (Ag/AgCl electrode), Counter electrode (platinum electrode), and working electrode (the metal oxide modified paste electrode). The electrolyte solution was aqueous NaOH (0.5 M). Electrochemical impedance spectroscopy (EIS), cyclic voltammetry (CV), and chronoamperometry (CP) techniques were conducted for measurements. All stock solutions of glucose were stored at 4 °C after preparation. For CV measurements the potential range was −0.8–1.0 V at scan rate of 50 mV s−1.
2.3. Methods
2.3.1. Synthesis of Fe2O3/CuFe2O4 nanocomposite. The Fe2O3/CuFe2O4 nanocomposite was synthesized using a co-precipitation method. A mixed solution of iron chloride (0.4 M) and copper chloride (0.2 M) was prepared. Subsequently, sodium hydroxide solution (NaOH, 3 M) was added dropwise to the mixture at room temperature 2 hours under continuous magnetic stirring at 700 rpm. The pH of the solution was carefully monitored and adjusted until it reached a value between 11 and 12, as described by Kashif Ali, Ali Bahadur,1 and Abdul Jabbar.1 The resulting suspension was then heated to 80 °C and maintained at this temperature for one hour to facilitate the reaction. Afterward, the mixture was allowed to cool naturally to room temperature (∼25 °Cover ∼1 hour). The obtained precipitate was thoroughly washed with deionized water, ethanol and centrifuged for 10 minutes (3000 rpm) to separate the solid product from the supernatant. Then, the isolated precipitate was dried overnight at 80 °C and annealed at 630 °C for 7 hours, following the procedure reported by Selima and Khairy.90
2.3.2. Preparation of un/modified electrodes. The unmodified carbon paste electrode (CPE) was prepared by mixing paraffin oil with graphene nanoplates (GNPs) and graphite powder in appropriate proportions.16,100 For the modified electrodes, varying percentages of the desired modifier (Fe2O3/CuFe2O4) were incorporated into a mixture containing graphite powder, graphene, and paraffin oil. Each formulation was homogenized by grinding in a mortar for at least 20 minutes to ensure a uniform paste16 and then packed into a plastic syringe with a copper wire inserted to provide electrical contact. The surface of the electrode was polished with parchment paper before electrochemical testing.
2.3.2.1 Electrode composition optimization. A series of modified carbon paste electrodes (CPEs) were prepared to investigate the effect of graphene nanoplatelets (GNPs) content on glucose sensing performance.16,125 Bare graphite (Gt) paste served as the control (0% GNPs). Modified pastes were fabricated by incorporating GNPs at varying mass percentages relative to the total carbon content (graphite + GNPs). The tested GNPs loadings were: (0%, 5%, 10%, 15%, 20%, 25%, 30%, 35%, 40%, 45%, 50%, and 100%).99
2.3.2.2 Preparation of Fe2O3/CuFe2O4-modified graphite/graphene electrodes. To evaluate the effect of Fe2O3/CuFe2O4 loading on sensor performance, a series of modified carbon paste electrodes were prepared using a base matrix composed of graphite powder and graphene nanoplatelets (GNPs). The Fe2O3/CuFe2O4 nanocomposite was incorporated into this carbon matrix at varying mass percentages: 1%, 3%, 5%, 7%, 9%, and 10% (relative to the total mass of the carbon content). The surface of each electrode was polished with parchment paper prior to use. These electrodes were then tested for their electrochemical response to glucose oxidation.33,114
2.3.3. Electrochemical characterization. Electrochemical impedance spectroscopy (EIS) and Cyclic voltammetry (CV) were employed to study the electrochemical properties of both unmodified and modified electrodes.100 All measurements were performed in an electrolyte solution containing 0.1 M potassium chloride (KCl) and 5 mM potassium ferrocyanide/ferricyanide [Fe (CN)6]3−/4− as a redox probe.11,100 CV measurements were produced within a potential range from −0.8 V to +1.0 V and a scan rate of 50 mV s−1.16,100 The EIS data were produced at open circuit potential (OCP) with a small AC perturbation amplitude of 5 mV, across a frequency range from 100 mHz to 100 kHz (ref. 100) The resulting Nyquist plots were analyzed using an appropriate equivalent circuit model to evaluate interfacial properties and charge-transfer resistance, providing insights into the electrode surface structure and conductivity.33,100 Both unmodified and Fe2O3/CuFe2O4/GNPs-modified electrodes were characterized under identical conditions for comparative analysis.33,100,114
2.3.4. Chronoamperometric analysis of non-enzymatic glucose sensing. Chronoamperometric measurements were performed to evaluate the performance of the optimized carbon paste electrode modified with graphene nanoplatelets (GNPs) and Fe2O3/CuFe2O4 nanocomposite for non-enzymatic glucose detection.33,56 The experimental conditions were systematically optimized, including the applied accumulation potential, analyte concentrations, and electrode composition.24 A comprehensive interference study was also conducted to assess the selectivity of the sensor in the presence of common electroactive species.46 Additionally, the sensor was applied to the quantitative analysis of glucose in real human serum samples.
2.3.5. Chronoamperometric analysis of glucose in real serum samples. The practical applicability of the developed sensor was validated by measuring glucose concentrations in human serum samples obtained from both diabetic and non-diabetic individuals.11 For analysis, 100 μL of serum was mixed with 100 μL of glucose solution and diluted in 20 mL of 0.5 M sodium hydroxide.125 Chronoamperometric measurements were carried out at a fixed applied potential of +0.6 V, and the current response was recorded over time.24,56To evaluate the accuracy of the sensor, the standard addition method was employed by spiking known concentrations of glucose into the serum samples.125 The recovery rates were calculated to assess precision and reliability.11 The results obtained using the proposed sensor were compared with those from a commercial glucometer, demonstrating good agreement and confirming the sensor's potential for real-sample glucose monitoring.11
3. Results and discussion
3.1. Physical characterization
The FTIR analysis provides compelling evidence for the successful fabrication of the Gt,GNPs,Gt/GNPs and Fe2O3/CuFe2O4 /GNPs/Gt composite and its subsequent electrochemical activity towards glucose oxidation. The spectrum of the pristine electrode (the composite modified electrode before detection) confirms the composite's structure (see Fig. 1A), displaying characteristic bands of the carbon matrix (O–H at ∼3450 cm−1 and C
C at ∼1630 cm−1) alongside a distinct absorption band in the 500–700 cm−1 region,35,95 which is a fingerprint for Fe–O bond vibrations in the Fe2O3 and CuFe2O4 phases.35 Following the electrocatalytic detection of glucose in a basic medium (the modified electrode after detection), significant spectral changes are observed. The notable decrease in the O–H band intensity and, more importantly, the clear alteration in the shape and intensity of the Fe–O band indicate a direct involvement of the metal oxide components in the reaction mechanism. This suggests that the oxidation of glucose is facilitated by a redox process involving the Fe2+/Fe3+ couples at the catalyst surface, consistent with the behavior of spinel ferrites in non-enzymatic sensing5,21 Furthermore, the emergence of new features in the 1000–1200 cm−1 region may be attributed to C–O bonds from adsorbed reaction intermediates or products, such as gluconolactone or gluconic acid. These results collectively verify that the metal oxides serve as the active sites and undergo specific chemical changes during the electrocatalytic detection of glucose.
 |
| | Fig. 1 FTIR spectrum of (A) FTIR spectra of Graphite, GNPs, Gt/GNPs, Fe2O3/CuFe2O4 (nanocomposite), (B) Fe2O3/CuFe2O4 nanocomposite before and after calcination. (C) Fe2O3/CuFe2O4/GNPs/Gt before and after glucose detection. (D) XRD spectra of the Fe2O3/CuFe2O4 nanocomposite. | |
From the FTIR spectrum of the Fe2O3/CuFe2O4 heterostructure synthesized and annealed at 630 °C presented in Fig. 1B, the broad absorption bands observed at 3440 cm−1 and 1638 cm−1 correspond to the O–H stretching and bending vibrations, respectively, indicating the presence of adsorbed water molecules on the surface of the nano-ferrite material.97,104 The Two peaks observed in FTIR spectra Fe2O3/CuFe2O4 heterostructure before calcination at 1346 and 1463 cm−1 may be due to residual nitrates, acetates, or organic precursors (from synthesis reagents like metal nitrates, chlorure or citrates). These bands disappear or weaken, confirming the thermal decomposition of residual precursors.82
The characteristic metal–oxygen vibrational bands appear in the low-wavenumber region, which are indicative of the spinel ferrite phase.21 Specifically, the absorption peaks located at 584 cm−1 and 468 cm−1 are attributed to the stretching vibrations of Fe–O bonds in the tetrahedral and octahedral sites of the spinel lattice, respectively.21 The peaks in the 400–700 cm−1 range (attributed to Fe–O and Cu–O bonds) are typically broader and less distinct, indicating amorphous or poorly crystalline metal oxide phases. These become stronger and sharper, confirming the formation of the spinel crystal structure of CuFe2O4 and Fe2O3. Indicates improved crystallinity and successful phase formation.23
Furthermore, the FTIR spectroscopy was employed to investigate the structural and chemical modifications occurring in the Fe2O3/CuFe2O4/GNPs/Gt nanocomposite before and after glucose detection (Fig. 1C). Prior to glucose exposure, the FTIR spectrum exhibited a broad absorption band around 3400 cm−1, which corresponds to the O–H stretching vibrations of hydroxyl groups present in gum tragacanth and physically adsorbed water molecules. The characteristic band at 1632 cm−1 is attributed to H–O–H bending vibrations or to the stretching of carbonyl (C
O) groups within the polysaccharide matrix. The peak at 1388 cm−1 is associated with C–H and C–O deformation vibrations, while the absorption bands below 600 cm−1 correspond to Fe–O and Cu–O stretching modes, confirming the presence of ferrite phases (Fe2O3 and CuFe2O4). After interaction with glucose, distinct spectral changes were observed. The appearance of new bands at 2922 cm−1 and 2854 cm−1 can be ascribed to the asymmetric and symmetric stretching of C–H bonds in glucose. Moreover, the slight shift of the 1632 cm−1 band, accompanied by the emergence of new peaks at 1456 cm−1 and 1374 cm−1, suggests the formation of interactions between the hydroxyl (–OH) and carbonyl (C
O) groups of glucose and the active sites on the nanocomposite surface. These modifications in peak position and intensity indicate effective glucose adsorption through hydrogen bonding or coordination with Fe and Cu ions. Overall, the FTIR findings confirm the strong affinity and chemical interaction between glucose molecules and the Fe2O3–CuFe2O4/GNPs/Gt nanocomposite, highlighting its suitability for glucose sensing applications.
The X-ray diffraction (XRD) analysis (Fig. 1D) shows that a heterostructured nanocomposite made of rhombohedral hematite (α-Fe2O3) and cubic spinel copper ferrite (CuFe2O4) has been successfully formed.33 The diffraction peaks at 2θ = 24.13°, 33.12°, 39.22°, 49.42°, 54.01°, 63.97°, and 72.18° match the (012), (104), (006), (024), (116), (300), and (119) crystallographic planes of α-Fe2O3 (JCPDS card no. 01-089-0596), which shows that it has a rhombohedral structure. The peaks at 2θ = 30.17°, 35.6°, 37.18°, 43.48°, 47.30°, 53.59°, 56.13°, 62.46°, 65.98°, 74.24°, and 75.43° are linked to the (220), (311), (222), (400), (331), (422), (511), (440), (531), (533), and (622) planes of cubic spinel CuFe2O4 (JCPDS card no. 01-077-0010).119 Several peaks show overlapping reflections (for example, 35.6°/35.6°, 43.48°/43.2°, 56.13°/57.13°, 62.46°/62.75°), which means that both phases are present and that a well-integrated heterostructure has formed. The lack of impurity peaks suggests that the phase is very pure, and the sharp, well-defined peaks suggest that the crystals are very crystalline, which is important for better electrocatalytic performance in sensor applications.
The nanocomposite (Fe2O3/CuFe2O4) scanning electron microscope (SEM) analysis (Fig. 2A) revealed the presence of irregularly shaped nanoparticles forming aggregated clusters.34 The images were captured at an accelerating voltage of 15 kV with magnifications ranging from 5000× to 100
00× and horizontal field widths (HFW) between 41.4 μm and 82.9 μm, providing detailed insight into the surface morphology of the nanocomposite.26
 |
| | Fig. 2 SEM image of (A) the Fe2O3/CuFe2O4 nanocomposite and (B) EDX spectra of Fe2O3/CuFe2O4. | |
Complementary Energy-Dispersive X-ray Spectroscopy (EDX) approved the elemental composition (see Fig. 2B), indicating 62.24 wt% Fe, 15.46 wt% O, and 22.27 wt% Cu.23 The corresponding atomic percentages were 45.85% Fe, 39.74% O, and 14.41% Cu.23 The high content of Fe and O is consistent with the formation of Fe2O3, while the significant presence of Cu confirms the incorporation of CuFe2O4 (ref. 33). These results collectively verify the successful synthesis of a biphasic Fe2O3/CuFe2O4 nanocomposite, with both phases well-represented in the final material.
3.1.1. Raman spectroscopy. Raman spectroscopy was conducted to investigate the structural and compositional evolution across the synthesized materials (Fig. 3). The spectrum of graphite (Gt) showed a sharp G-band at ∼1580 cm−1 and a negligible D-band at ∼1350 cm−1, reflecting a highly crystalline sp2 carbon structure with minimal defects.75 In contrast, graphene nanoplatelets (GNPs) exhibited a pronounced D-band and a broadened G-band with a significantly higher ID/IG ratio, indicating the introduction of lattice defects and residual oxygen-containing functional groups during oxidation and the subsequent incomplete reduction process.51 The Gt/GNPs hybrid displayed intermediate characteristics, with a notably increased D-band intensity compared to graphite, suggesting interfacial strain and defect formation at the junctions between the graphitic and graphene nanoplatelets domains.59
 |
| | Fig. 3 Raman spectra of Graphite, GNPs, Gt/GNPs, Fe2O3/CuFe2O4 (before and after calcination), and Gt/GNPs-Fe2O3/CuFe2O4 nanocomposite. | |
The Fe2O3/CuFe2O4 precursor before calcination showed only broad, poorly resolved Raman features in the 200–700 cm−1 range, indicative of amorphous or poorly crystalline metal–oxygen vibrational modes.36 After calcination, these features sharpened dramatically into a well-defined set of peaks. A precise spectral assignment confirms the formation of the target crystalline phases: the peaks at 299, 412, and 613 cm−1 are unambiguous signatures of hematite (α-Fe2O3), corresponding to Eg and A1g vibrational modes.126 Concurrently, the peaks at 475, 567, and 682 cm−1 are diagnostic of the spinel structure, confirming the successful synthesis of copper ferrite (CuFe2O4).76 The coexistence of these distinct bands confirms the formation of a binary metal oxide system. Additionally, the calcined sample exhibited a significant fluorescence background in the higher wavenumber region (>3000 cm−1), attributed to charge recombination within the semiconductor oxides and possibly surface –OH groups.9 In the final Fe2O3/CuFe2O4/GNPs/Gt composite, the carbon D- and G-bands remained prominent, indicating the survival of the graphene-based network, albeit with a substantial defect density.30,51 Crucially, the Raman spectrum retained the distinct low-wavenumber bands of both crystalline α-Fe2O3 and spinel CuFe2O4, providing direct evidence for the successful integration of these metal oxide nanoparticles within the carbon matrix without phase degradation.30,109 The further increased D-band intensity and broadening of the carbon signals, relative to the Gt/GNPs hybrid, suggest strong interfacial interactions between the carbon substrate and the metal oxides. This interaction likely induces local strain and facilitates charge transfer, contributing to the enhanced disorder.13,59
3.1.2. X-ray photoelectron spectroscopy (XPS) analysis. X-ray photoelectron spectroscopy (XPS) was employed to investigate the surface composition and oxidation states of the Fe2O3/CuFe2O4 nanocomposite immobilized on graphite (Gt) sheets/graphene nanoplatelets (GNPs) (Fig. 4). The wide-scan spectra were collected using a Mg Kα radiation source (1253.6 eV) with a step size of 1.0 eV and a pass energy of 50 eV, covering a binding energy range of 0–1000 eV to identify the elemental composition of the samples. The investigated systems included S01 (Gt/GNPs), S02 (Fe2O3/CuFe2O4), S03 (Fe2O3/CuFe2O4/GNPs/Gt nanocomposite before electrochemical use), and S04 (Fe2O3/CuFe2O4/GNPs/Gt nanocomposite after glucose detection in NaOH solution). In addition to the survey scans, high-resolution spectra were recorded for the Fe 2p, Cu 2p, and O 1s regions. Peak deconvolution was carried out using Gaussian–Lorentzian fitting with Shirley background subtraction to resolve chemical states and distinguish surface oxidation environments, following established protocols in XPS data analysis12 NIST X-ray Photoelectron Spectroscopy Database. The use of XPS in nanomaterial characterization has been widely reported, with recent reviews highlighting its role in probing 2D materials for oxidation/corrosion protection,88 graphene-based sensing systems,112 and carbon nanotube/graphite hybrid structures.120 These studies reinforce the significance of XPS as a robust technique for linking chemical states with functional performance in nanostructured composites.
 |
| | Fig. 4 Wide-scan XPS spectra of (S01) Gt/GNPs, (S02) Fe2O3/CuFe2O4, (S03) Gt/GNPs/Fe2O3/CuFe2O4 nanocomposite before use, and (S04) Gt/GNPs-Fe2O3/CuFe2O4 nanocomposite after glucose detection. Overlaid comparison shows attenuation of metal signals in (S04). | |
3.1.3. C 1s high-resolution analysis. High-resolution C 1s spectra provide insights into the carbon bonding environments as shown in Fig. 5A. For sample (Fig. 5A(a)), likely corresponding to the Gt/GNPS composite, the deconvoluted peaks at 276.61 eV, 285.76 eV, and 284.97 eV are assigned to C–C/C
C (sp2 hybridized carbon),6,84 C–O (epoxy/hydroxyl groups),25,125 and adventitious carbon, respectively, with the dominant peak at ∼285 eV indicating a graphitic structure with oxygenated defects.25,125 In sample (Fig. 5A(b)), peaks at 284.95 eV and 285.83 eV suggest a shift toward higher binding energies, implying increased oxidation or interaction with metal oxides.50 Sample (Fig. 5A(c)) shows a single broad peak at 286.123 eV, which may reflect further carbon oxidation or contamination after glucose detection.7 These shifts highlight the role of graphene in stabilizing the nanocomposite50,98 while undergoing minor surface alterations during sensing.7
 |
| | Fig. 5 (A): High-resolution C 1s spectra: (a) (S01), (b) (S03), (c) (S04) showing deconvoluted components for C–C/C–H (∼284.9–285.0 eV) and C–O (∼285.7–286.1 eV), indicating increased carbon oxidation. (B): High-resolution Fe 2p spectra: (a) (S02), (b) (S03), (c) (S04) with deconvolution showing Fe3+ (2p3/2 ∼710.8 eV, 2p1/2 ∼724 eV), Fe2+, and FeOOH components. Inset: Spin–orbit splitting diagram for Fe 2p, with Fe3+/Fe2+ ratio evolution. (C): High-resolution Cu 2p spectra: (a) (S02), (b) (S03), (c) (S04) with deconvolution showing Cu2+ (2p3/2 ∼933.5 eV, 2p1/2 ∼952 eV with satellites), Cu+, and CuOOH components. Inset: Spin–orbit splitting for Cu 2p, highlighting satellite for Cu2+. (D): High-resolution O 1s spectra: (a) (S02), (b) (S03), (c) (S04) with deconvolution showing lattice O, OH, O–C, and H2O components. (E): High-resolution Na 1s spectra: (S04) with deconvolution. | |
3.1.4. Fe 2p high-resolution analysis. As shown in Fig. 5B, the Fe 2p spectrum in (S02, Fig. 5B(a)) exhibits a doublet at Fe 2p3/2 (∼710.8 eV) and Fe 2p1/2 (∼724 eV) with a satellite at ∼718 eV, indicating Fe3+ in Fe2O3 (FWHM ∼2.5 eV).15 The 2p orbital splits into 2p3/2 (higher intensity, j = 3/2 state with three spin orientations) and 2p1/2 (lower intensity, j = 1/2 state with one spin orientation) due to spin–orbit coupling, with a splitting energy of ∼13 eV typical for Fe113 Deconvolution reveals ∼68% Fe3+ and ∼32% Fe2+, consistent with spinel CuFe2O4 structures.52 In (S03, Fig. 5B(b)), intensity decreases (∼50%) due to graphite encapsulation, with a minor shift to ∼711 eV from surface OH adsorption.37 In (S04, Fig. 5Bc), further attenuation (<20% of S02) and shift to ∼711.5–712 eV (FWHM ∼10 eV) confirm FeOOH formation (∼20% contribution) and increased Fe2+ (∼35%), arising from redox cycling during glucose oxidation (Fe2O3 + OH− → FeOOH + e−).5 The 2p3/2 peak dominates (∼70% of the doublet intensity) as it has higher degeneracy, providing sensitive probing of Fe oxidation states. The Fe3+/Fe2+ ratio (∼2
:
1 in S02) decreases to ∼1.9
:
1 in (S03) and ∼1.7
:
1 in (S04, Fig. 5B(c)), reflecting progressive reduction during activation and reaction, which enhances catalytic activity by balancing oxidative power and electron transfer.52,60
3.1.5. Cu 2p high-resolution analysis. In Fig. 5C, (S02, Fig. 5C(a)) shows Cu 2p3/2 (∼933.5 eV) and Cu 2p1/2 (∼952 eV) with shake-up satellites (∼941–943 eV, ∼40% intensity), confirming Cu2+ in CuFe2O4 (∼80%) with minor Cu+ (∼20%). The 2p splitting arises from spin–orbit coupling, with 2p3/2 (j = 3/2, higher intensity) and 2p1/2 (j = 1/2, ∼50% intensity of 2p3/2) separated by ∼19 eV, characteristic of Cu's d10 configuration in Cu+ vs. d9 in Cu2+ (evident from satellite presence).14,37 (S03, Fig. 5C(b)) exhibits reduced intensity (∼60%) and a shift to ∼934 eV from graphite interaction.113(S04, Fig. 5C(c)) displays severe attenuation (<10%) and a shift to ∼934.5–935 eV, with weakened satellites and increased Cu+ (∼25%), indicating CuOOH formation (CuO + OH− → CuOOH + e−) and Cu2+/Cu3+ cycling60
3.1.6. O 1s high-resolution analysis. In Fig. 5D, S02 (Fig. 5D(a)) shows lattice O2− (∼529.5–530 eV, ∼60%) and surface-adsorbed OH groups (∼531 eV, ∼40%)37,113 S03 (Fig. 5D(b)) shifts to ∼530–531 eV (∼5000 cps) with a significant increase in the OH component (∼50%) resulting from NaOH activation.52,113 S04 (Fig. 5D(c)) broadens to ∼531–533 eV (∼6000 cps), deconvoluting to ∼25% lattice O2−, ∼55% metal-OH (FeOOH/CuOOH), ∼15% O–C (from adsorbed gluconic acid), and ∼5% H2O, confirming the role of hydroxide mediation and organic deposition during the glucose oxidation process.60
3.1.7. Na 1s high-resolution analysis. In Fig. 5E, the high-resolution XPS analysis of the Sodium (Na 1s) core level reveals four distinct spectral components (M1–M4), indicating the presence of sodium in multiple chemical states within the sample. The most prominent feature is the highly intense M3 peak centered at a binding energy of 1070.7 eV. This value is characteristic of Na+ ions in a stable configuration, strongly suggesting that this component corresponds to sodium bound to the gluconate molecule, effectively representing the sodium gluconate salt which acts as a counterion to the organic carboxylate group.The other components are attributed to surface-related species. The low binding energy component at 1065.4 eV (M2) are likely associated with highly oxidized species or surface contaminants, such as NaOH or Na2O. In contrast, the higher binding energy peaks at 1076.23 eV (M3) and 1079.42 eV (M4) are often indicative of plasmon loss features or differential charging effects, though they may also represent Na+ ions in highly electronegative environments or associated with lattice defects.
This analysis confirms the successful incorporation of the gluconate salt into the composite structure.
The XPS results demonstrate successful nanocomposite formation (S03) and surface evolution post-glucose detection (S04), with metal attenuation and OH/O–C increases indicating catalytic activity. The 2p3/2 and 2p1/2 peaks provide detailed insights into oxidation states, with 2p3/2's higher intensity enabling precise deconvolution of Fe3+/Fe2+ and Cu2+/Cu+ ratios. Post-use XPS (S04) correlates with enhanced performance: OH intermediates (O 1s ∼531 eV) mediate oxidation, while passivation slightly reduces long-term stability but maintains initial efficiency.
3.2. Electrochemical characterization
Electrochemical characterizations were performed by using CV and EIS techniques for the different modified electrodes.
3.2.1. Performance of graphene/graphite ratios. Graphene nanoplatelets (GNPs) was used as a conductivity enhancer in carbon paste electrodes (CPEs), and its content significantly influenced the electrode's performance was investigated using a mixture solution containing 0.1 M KCl and 5 mM [Fe(CN)6]3−/4− and electrochemical measurements by CV and EIS techniques. The redox couple was employed to evaluate the electron transfer behavior of electrodes prepared with various graphene-to-graphite mass ratios (0, 5, 10, 15, 20, 25, 30, 35, 40, 45, 50, and 100%), as illustrated in Fig. 6.
 |
| | Fig. 6 (A) Cyclic voltammograms (B) Nyquist diagram of different percentages of graphite/GNPs (0, 5, 10, 20, 25, 30, 35; 40, 45, 50, 100 wt%) in a solution containing 5 mM Fe(CN)63–/4− and 0.1 M KCl. | |
The electrochemical performance of electrodes with varying graphene nanoplatelets (GNPs) to graphite ratios was evaluated to identify the composition with the lowest charge transfer resistance and highest redox activity. While both graphite and GNPs are conductive carbon materials, they exhibit different physical and electronic properties. Graphite provides structural robustness and moderate conductivity, whereas GNPs offer high electron mobility,2,16 and a large surface area. However, excessive GNPs content can lead to restacking and aggregation, reducing electroactive surface area and conductivity due to poor dispersion2,125
The unmodified electrode (0% GNPs) exhibited the lowest redox peak current16,17 (Fig. 6A). In contrast, all GNPs-modified electrodes showed enhanced redox responses16,17 Increasing the GNPs content resulted in a gradual rise in both anodic and cathodic peak currents, as summarized in Table 1a.125 Notably, the electrode with a 45
:
55 GNPs -to -graphite ratio116 demonstrated the highest oxidation peak current (290.8 μA) and the smallest peak potential separation (ΔEp), indicating rapid electron transfer kinetics.42 Furthermore, the peak current ratio (Ipa/Ipc) of 1.11 suggests a reversible redox process.11 The superior electrical conductivity of GNPs is likely responsible for the enhanced electrochemical performance of the modified CPEs42 Based on these results, the optimal GNPs-to- graphite ratio for electrode fabrication was determined to be 45
:
55.116 Nyquist plots for bare graphite and GNPs/graphite modified with different wt% were recorded, as shown in Fig. 5B and Table 1a. Each EIS 53 spectrum consists of two regions: a quasi-semicircular portion at high frequencies and a linear segment at low frequencies. The diameter of the semicircular region corresponds to the charge transfer resistance (Rct), with a larger diameter indicating a higher Rct.17,42 Conversely, the near-vertical linear portion reflects efficient electron transfer at the electrode interface.53 The Rct values for various GNPs to graphite mass ratios (0, 5, 10, 15, 20, 25, 30, 35, 40, 45, 50, and 100%) electrodes were recorded as shown in Table 1. The electrode with a 45
:
55 GNPs-to-graphite ratio demonstrated the lowest Rct value (151.2 Ω) which matched with the CV measurements. In contrast, the GNPs/graphite modified electrode (45
:
55%) significantly enhanced electron transfer kinetics.
Table 1 Electrochemical data from CV and EIS measurements obtained from the electrodes modified with (A) GNPs/graphite ratios and (B) Fe2O3/CuFe2O4 wt%. The data are obtained from the voltammetric and impedimetric experiments (Fig. 5 and 6)
| |
Ia (mA) |
Ic (mA) |
Eoxd (V) |
Ered (V) |
E1/2 (V) |
Rs (Ω) |
Rct (Ω) |
Q μF |
W (Ω) |
| (A) GNPs/Graphite ratio |
| 0 |
0.163 |
−0.171 |
0.380 |
−0.1712 |
0.104 |
148.6 |
1675 |
24.49 |
1453 |
| 5:95 |
0.152 |
−0.171 |
0.447 |
−0.1713 |
0.137 |
282.5 |
2110 |
21.26 |
725 |
10 : 90 |
0.143 |
−0.156 |
0.415 |
−0.1562 |
0.129 |
197.9 |
977.7 |
47.88 |
351.8 |
15 : 85 |
0.154 |
−0.164 |
0.372 |
−0.1647 |
0.103 |
106.1 |
586.1 |
67.44 |
304.7 |
20 : 80 |
0.153 |
−0.164 |
0.381 |
−0.1643 |
0.108 |
66.06 |
660.8 |
57.27 |
266.6 |
25 : 75 |
0.190 |
−0.206 |
0.384 |
−0.206 |
0.089 |
80.05 |
931.9 |
95.21 |
495.0 |
30 : 70 |
0.179 |
−0.193 |
0.375 |
−0.1936 |
0.090 |
42.13 |
2983 |
65.26 |
1784 |
35 : 65 |
0.210 |
−0.204 |
0.353 |
−0.204 |
0.074 |
71.32 |
261.9 |
287.8 |
480.6 |
40 : 60 |
0.180 |
−0.184 |
0.374 |
−0.184 |
0.095 |
146.6 |
436.7 |
253.6 |
598.7 |
45 : 55 |
0.290 |
−0.260 |
0.330 |
−0.2606 |
0.035 |
32.04 |
151.2 |
245.6 |
194.6 |
50 : 50 |
0.225 |
−0.227 |
0.318 |
−0.2272 |
0.045 |
56.7 |
259.0 |
140.9 |
432.6 |
![[thin space (1/6-em)]](https://www.rsc.org/images/entities/char_2009.gif) |
| (B) Fe2O3/CuFe2O4-wt% |
| 1 |
0.2096 |
−0.2083 |
0.3752 |
0.0656 |
0.2204 |
46.14 |
269.7 |
246.9 |
182.8 |
| 3 |
0.2204 |
−0.2281 |
0.3723 |
0.0618 |
0.2170 |
41.08 |
285.6 |
215.0 |
286.3 |
| 5 |
0.271 |
−0.2551 |
0.327 |
0.09277 |
0.2098 |
26.29 |
181.1 |
291.2 |
303.3 |
| 7 |
0.2451 |
−0.2464 |
0.3715 |
0.05369 |
0.2125 |
31.02 |
451.7 |
306.2 |
589.6 |
| 9 |
0.213 |
−0.2229 |
0.3767 |
0.05091 |
0.2138 |
39.19 |
367.4 |
213.3 |
296.4 |
| 10 |
0.2721 |
−0.2325 |
0.3202 |
0.1279 |
0.2240 |
42.37 |
77.7 |
230.5 |
266.4 |
The optimized ratio of 45
:
50 (GNPs
:
graphite) was found to balance these effects.78 This composition lies near the percolation threshold, where a continuous conductive network forms due to optimal GNPs interconnectivity while maintaining structural support from graphite.17,78,116 Compared to the 50
:
50 ratio, the 45
:
55 ratio yielded slightly lower Rct values and better peak current responses, suggesting improved electron mobility and charge transport efficiency. Ratios with higher GNPs content (>60%) led to unstable electrochemical responses, likely due to poor dispersion.83 These findings indicate a synergistic effect between graphite and GNPs, enhancing both mechanical integrity and electron transport within the composite. Moreover, the EIS interpretation (i.e. semicircle → Rct) is consistent with analogous results in composite.85
3.2.2. Electrochemical performance of Fe2O3/CuFe2O4 ratio in Fe2O3/CuFe2O4 /GNPs/Gt nanocomposites. To optimize the amount of active metal oxide phase, various loadings of Fe2O3/CuFe2O4 (1–10 wt%) were incorporated into the carbon matrix. The electrochemical behavior of a GNPs/graphite nanocomposite (mass ratio 45
:
55) modified with varying amounts (1, 3, 5, 7, 9, and 10 wt%) of Fe2O3/CuFe2O4 was investigated using CV and EIS in redox solution ferro/ferri cyanide. CV measurements at a scan rate of 50 mV s−1 for bare Gt, Gt/GNPs, andFe2O3/CuFe2O4/GNPs/Gt composites are shown in Fig. 7A and Table 1b.
 |
| | Fig. 7 (A) Cyclic voltammograms (B) Nyquist diagram of different percentages of Fe2O3/CuFe2O4 (1; 03, 05, 07; 09; 10 wt%) in a solution containing 5 mM Fe(CN)63–/4– and 0.1 M KCl. | |
The peak-to-peak separation (ΔEp, see Table 1) for GNPs/graphite electrode decreased compared with bare graphite electrode, reflecting improved electrical conductivity due to the presence of GNPs.17,117 Among the tested materials, the Fe2O3/CuFe2O4/GNPs/Gt composite exhibited the lowest ΔEp and the highest peak current (Ip), indicating enhanced electron transfer kinetics. This improvement is attributed to the excellent electrocatalytic125 properties of Fe2O3/CuFe2O4, which, when optimally incorporated into the composite, significantly boosts the electrochemical response of the electrode.4,79
Nyquist plots for GNPs/graphite modified with (1, 3, 5, 7, 9, and 10 wt%) Fe2O3/CuFe2O4 were recorded, as shown in Fig. 7B and Table 1b. The GNPs/graphite modified with (10 wt%) Fe2O3/CuFe2O4 exhibited the lowest Rct value (77.7 Ω), indicating significantly enhanced electron transfer kinetics. At lower loadings (e.g., 5% and 7%), the catalytic surface area was insufficient to generate a significant current, resulting in higher Rct values. At higher loadings (e.g., 12.5% and 15%), the metal oxide nanoparticles tended to agglomerate, obstructing the carbon network and impeding electron flow.87 This led to increased charge-transfer resistance and reduced electrochemical performance. The 10 wt% loading strikes an optimal balance between catalytic activity and conductive pathway integrity. The observed trend highlights a synergistic interaction between the metal oxide particles and the conductive carbon network, where sufficient redox-active sites are available without compromising the electron mobility across the electrode surface.
These EIS findings are in good agreement with the cyclic voltammetry results, further validating the improved electrochemical performance of the modified electrodes.42
To enhance the understanding of the electrochemical performance, the effective electrochemically active surface area (EASA) was determined for unmodified graphite electrodes, graphite/GNPs electrodes, and those modified with Fe2O3/CuFe2O4/GNPs/Gt nanocomposites.103
This evaluation was conducted by studying the influence of scan rates ranging from 10 to 250 mV s−1, as illustrated in Fig. 8(A–C).
 |
| | Fig. 8 Plots of peak current against the square root of scan rate and inset figures is cyclic voltammograms of (A) graphite, (B) graphite/GNPs, (C) Fe2O3/CuFe2O4/GNPs/Gt in a 5 mM Fe(CN)63−/4− solution containing 0.1 M KCl at varying scan rates. | |
The relationship between the peak current and the square root of the scan rate was analyzed to assess the diffusion-controlled electrochemical behavior.96 The active surface area for each electrode was calculated using the Randles–Ševčík eqn (1).108 Which is commonly applied for reversible redox systems:32
| |
 | (1) |
where
Ip is the voltammetric peak current,
n is the number of electron transfer,
A is the electrochemical active surface area (cm
2),
D is the diffusion coefficient (cm
2 s
−1),
C is the concentration of the electrochemical redox species [Fe (CN)
6]
3–/4– (mol l
−1),and
ν means the scan velocity (V s
−1).
By plotting Ip versus the square root of the scan rate (ν1/2), the electroactive surface areas were calculated for each electrode type. The unmodified graphite electrode had an active area of 0.249 cm2, while the graphite/GNPs and Fe2O3/CuFe2O4/GNPs/Gt modified electrodes showed larger areas of 0.453 cm2 and 0.511 cm2, respectively.96,103 This indicates that modification with the nanocomposite significantly enhances the effective surface area available for electron transfer.79
Moreover, the nearly linear relationship between Ip and ν1/2, particularly in the nanocomposite-modified electrode, suggests a diffusion-controlled process.32 The observed peak separation also supports a quasi-reversible redox behavior.4
3.3. Optimization of NaOH concentration as electrolyte
The electrochemical activity of carbon paste sensors based on Fe2O3/CuFe2O4/GNPs/Gt composites was evaluated using cyclic voltammetry (CV) Fig. 9 in varying concentrations of NaOH (0.01, 0.02, 0.05, 0.1, 0.2, and 0.5 M), both in the presence of 2 mM glucose.79 Measurements were conducted at a scan rate of 50 mV s−1 using an Ag/AgCl reference electrode, following the method reported by Yanwu Zhang80,119
 |
| | Fig. 9 Cyclic voltammograms ofFe2O3/CuFe2O4/GNPs/Gt indifferent concentrations of NaOH solutions with 2.0 mM glucose at 50 mV s−1 Ag/AgCl. | |
As shown in Fig. 9, the anodic current increased with higher NaOH concentrations, which is attributed to the generation of more high-potential hydroxyl radicals that facilitate glucose oxidation.8,111 Among the tested concentrations, 0.5 M NaOH produced the highest anodic response, a trend consistent with literature findings on the critical role of alkaline media in non-enzymatic glucose sensing.124 Therefore, selected as the optimal alkaline electrolyte for further electrochemical studies.
In All experiments were performed in alkaline solution 50 mM with and without the addition of 2 mM glucose. In the potential range −0.8 to 1 V at a scan rate of 50 mV s−1.
The cyclic voltammograms of bare graphite, graphite/graphene, and Fe2O3/CuFe2O4/GNPs/Gt electrodes, recorded in the absence and presence of 2 mM glucose, are presented in Fig. 10(A–C). The bare graphite electrode exhibited a very low oxidation current for glucose, indicating limited electrocatalytic activity.81,110
 |
| | Fig. 10 Cyclic voltammograms of the (A) graphite, (B) graphite/graphene and (C) Fe2O3/CuFe2O4/GNPs/Gt in the (a) presence and (b) absence of 2 mM glucose. The electrolyte is 0.5 M NaOH and the scan rate is 50 mV s−1. | |
In contrast, the graphite/GNPs-modified electrode demonstrated a notably higher oxidation current, which can be attributed to the enhanced electrical conductivity of graphene.2,93,118 Furthermore, the Fe2O3/CuFe2O4/GNPs/Gt electrode showed a significant increase in the oxidation peak current, along with the appearance of a distinct shoulder peak around 0.6 V. This suggests improved electrocatalytic behavior due to the incorporation of the Fe2O3/CuFe2O4 nanocomposite86,92 Additionally, the electrocatalytic oxidation of 2 mM glucose on the Fe2O3/CuFe2O4/GNPs/Gt modified electrode in 0.5 M NaOH was investigated across scan rates ranging from 50 to 500 mV s−1 (Fig. 11A). The resulting linear relationship between the peak current and scan rate (see Fig. 11B) confirms that the glucose electrooxidation process is surface-controlled under these conditions.81 Overall, the results confirm that the presence of graphene is essential for improving the conductivity and sensitivity of the electrode, while the Fe2O3/CuFe2O4 nanocomposite contributes to enhanced electrocatalytic activity. The Fe2O3/CuFe2O4/GNPs/Gt electrode thus benefits from a synergistic effect between graphene and the metal oxide nanocomposite.45
 |
| | Fig. 11 (A) Cyclic voltammograms of Fe2O3/CuFe2O4/GNPs/Gt in 0.5 mM NaOH solution with 02 mM Glucose at different scan rates; from inner to outer: 10, 20, 40, 50, 60, 80, 100, 120, 140, 160, 180, 200, 300, 400 and 500 mV s−1. Inset shows the plots of (B) peak current versus the square root of scan rate. | |
3.3.1. Electron transfer kinetics and proposed catalytic mechanism. To gain deeper insight into the kinetics of glucose oxidation on the Fe2O3/CuFe2O4/GNPs/Gt nanocomposite, cyclic voltammetry (CV) was conducted at varying scan rates (10–100 mV s−1) in 0.5 M NaOH containing 2.0 mM glucose. As shown in Fig. 11C, the anodic peak current (Ip) increased linearly with the log square root of the scan rate (ν), indicating that the glucose oxidation process is primarily diffusion-controlled. However, the shift in peak potential with increasing scan rate suggests quasi-reversible kinetics, which are appropriately described by Laviron's equation for quasi-reversible redox systems.54To further analyze the electron transfer process at the Fe2O3/CuFe2O4/GNPs/Gt electrode, Laviron's eqn (2) for quasi-reversible redox systems was applied:40
| |
Ep = E0 + RT/(1 − α) nF ln(RTk0/(1 − α)nF) + RT/(1 − α)nF ln ѵ
| (2) |
where
Ep is the peak potential (V),
ν is the scan rate (V s
−1),
n is the number of electrons transferred,
α is the charge transfer coefficient (assumed to be 0.5),
k0 is the standard rate constant, and
R,
T, and
F are universal gas constant, absolute temperature, and Faraday constant respectively. A plot of
Ep versus lg
ν (
Fig. 11D) showed a linear trend with a slope of approximately 30 mV per decade. Using the slope expression from Laviron's formalism, Slope =
RT(1 −
α)
nF, and substituting
R = 8.314 J mol
−1 K
−1 R = 8.314 J mol
−1 K
−1,
T = 298 K,
F = 96485 C mol
−1, and the observed slope of 0.032 V, the number of electrons transferred was calculated as approximately
n ≈ 2.05.
102 This value indicates that the glucose oxidation process at the electrode surface involves the transfer of two electrons, consistent with the conversion of glucose to gluconolactone.
81,101
The efficient electrocatalytic performance of the Fe2O3/CuFe2O4/GNPs/Gt composite electrode can be ascribed to the synergistic interaction among its redox-active constituents. Copper ions play a central catalytic role, wherein Cu2+ is electrochemically oxidized to Cu3+, which then chemically oxidizes glucose to gluconolactone, following an EC (Electrochemical–Chemical) mechanism:81,102
| Cu3+ + glucose → Cu2+ + gluconolactone |
This redox cycling facilitates sustained catalytic turnover and electron transfer. Additionally, iron centers from both Fe2O3 and CuFe2O4 may contribute to the redox process either through direct participation or by assisting in the regeneration of Cu2+ species via internal electron shuttling.19 The inclusion of Graphene Nanoplatelets (GNPs) further enhances performance by providing a highly conductive network that accelerates charge transport, minimizes charge transfer resistance (Rct), and prevents nanoparticle agglomeration, thereby maximizing the exposure of active sites.123 Collectively, these factors result in enhanced electron transfer kinetics, reduced overpotential, and a heightened anodic response during glucose oxidation. In Fig. 11A, a distinct shoulder peak appears at approximately 0.6 V prior to the main oxidation peak (∼0.75 V). This feature is attributed to the initial adsorption and partial dehydrogenation of glucose molecules on the Fe2O3/CuFe2O4/GNPs/Gt surface. In alkaline media, non-enzymatic glucose oxidation proceeds through multiple steps, beginning with the formation of a glucose-derived intermediate, such as gluconolactone or enediol species. This step is facilitated by the electrochemically generated Cu3+ or Fe3+ species, which act as active catalytic centers. The shoulder peak therefore corresponds to the oxidation of glucose to an intermediate, while the subsequent peak represents further oxidation to the final product. Similar multi-step oxidation behavior with shoulder peaks has been reported in literature for Cu- and Fe-based glucose sensors, confirming the complex surface-controlled reaction mechanism.19,38
3.3.2. Effects of applied potential. The applied potential significantly affects the chronoamperometric electrochemical sensor's signal response.56 Therefore, the effect of different applied potential on the glucose response of the Fe2O3/CuFe2O4/GNPs/Gt modified electrode was investigated. The chronoamperometric measurements were conducted in 0.5 M NaOH by successive injections of 10 μL glucose at 120 seconds intervals, under various applied potentials ranging 0.40–0.65 V, as illustrated in Fig. 12A and B.
 |
| | Fig. 12 (A) Chronoamperometric response and (B) the of linearity plot of the Fe2O3/CuFe2O4/GNPs/Gt electrode at different potentials 0.4 V, 0.45 V, 0.5 V, 0.55 V, 0.6 V and 0.65 V with a fixed addition of 10 μL glucose into gently stirred 0.5 M NaOH solution. | |
The current response increased progressively with rising detection potential, reaching a maximum at 0.60V.19 However, further increasing the potential to 0.65 V resulted in a decline in current response, likely due to the non-specific oxidation of interfering species or by-products at higher potentials.38,81 Therefore, +0.60 V was selected as the optimal detection potential for subsequent glucose sensing due to its balance between sensitivity and selectivity.
3.4. Calibration curve
To evaluate the sensitivity and detection limit of the non-enzymatic glucose sensor based on the Fe2O3/CuFe2O4/GNPs/Gt modified electrode, chronoamperometric measurements were performed under optimized conditions in 0.5 M NaOH solution.19 Successive injections of varying glucose concentrations were introduced over a broad concentration range. The corresponding chronoamperometric responses are shown in Fig. 13A.
 |
| | Fig. 13 (A) Chronoamperometric responses and (B) the linear relationships between the catalytic current and the concentration of Fe2O3/CuFe2O4/GNPs/Gt electrode upon the successive addition of glucose into gently stirred 0.5 M NaOH at 0.6 V. | |
The resulting calibration curves (Fig. 13B) demonstrate excellent linearity and sensitivity. The sensor exhibited high sensitivities of 0.062 μA μM−1 cm−2 and 0.001329 μA μM−1 cm−2 within the respective concentration ranges.81 The proposed sensor reveals two distinct linear response ranges: from 5.0 to 75 μM and from 75 to 1000 μM μM with R2 = 0.9983. The limit of detection (LOD) was calculated using the IUPAC-recommended formula LOD = 3σ/S, where σ is the standard deviation of the blank (n = 7) and S is the slope of the calibration curve. The resulting LOD was 0.049 μM, indicating very high analytical sensitivity under laboratory conditions. It is important to note, however, that this value reflects performance in buffered standard solutions and may differ in complex biological matrices due to sample interference or electrode fouling. Comparatively, recent reports on non-enzymatic glucose sensors have achieved LODs in the range of 0.055–10 μM. Our result is among the lowest reported, which we attribute to the synergistic conductivity and catalytic activity of the Fe2O3/CuFe2O4/GNPs/Gt nanocomposite.
Moreover, the rapid and consistent responses—observed for the modified sensor every 120 seconds following each glucose addition—highlight the strong electrocatalytic synergy provided by the nanocomposite materials.
A comparison with previously reported electrochemical glucose sensors is provided in Table 2. Notably, the current system demonstrates superior sensitivity relative to many other non-enzymatic detection approaches.
Table 2 Evaluation of the proposed glucose sensor and the others modified electrodes in the literaturea
| Modification type of sensor |
Linear range |
Limit of detection (μM) |
Sensitivity (μA mM−1 cm−2) |
Ref. |
| GCE glassy carbon electrode, MWCNT multi-walled carbon nanotubes CuFe2O4 copper ferrite,rGO reduced graphene oxide nanocomposite, Au gold nanoparticles, PPy core–shell nanoparticle polypyrrole core–shell nanoparticle, GRE graphite rod electrode. |
| Nickel foam electrode modified with CuFe2O4 nanotubes |
20 μM to 5 mM |
0.22 mM, 22 μM |
1239 |
111 |
| 500 μM to 25 mM |
219 for H2O2 |
| Cu-Doped ZnO nanoparticles |
1 nM to 100 μM |
0.7 nM |
— |
74 |
| RGOs-Cu2O nanocomposites |
0.01 to 6 mM |
0.055 μM |
185 |
121 |
| CuFe2O4/PPyCore-shell nanoparticles |
0.1–7.5 mM |
1 μM |
1824 |
93 |
| Ni/NiO/graphene |
00.299–6.44 mM |
1.8 μM |
1997 |
118 |
| Hexagonal Co 3O 4/reduced graphene oxide sheets |
0.001–0.5 mM |
0.4 μM |
1315 |
105 |
| Co3O4/CuO nanorod array supported on carbon cloth |
Fast response time of 1.9 s |
0.38 μM |
405 |
20 |
| Cu(OH)2/3D porous graphene foam (PGF) |
0.0012–6 mM |
1.2 μM |
3360 |
91 |
| g-C3N4/α-Fe2O3 |
0.002–2.4 mM |
Not specified |
— |
57 |
| CoNi2Se4/rGO |
1–4 mM |
0.65 μM |
18 890 |
3 |
| ZnO/Co3O4/reduced graphene oxide (rGO) nanocomposite |
0.015–10 mM |
0.093 μM |
1551.38 |
44 |
| NiO/Fe2O3 |
50 μM to 2.867 mM |
3.9 μM |
230.5 |
58 |
| Ni-NPs/Ppy/GRE-based |
1–1000 μM |
0.4 μM |
2873 |
29 |
| Fe2O3/CuFe2O4/GNPs nanocomposite |
5–75 μM |
0.049 μM |
624 |
This work |
75–13 000 μM |
25 μM |
3.5. Selectivity
To evaluate the real-world applicability of the sensor, selectivity tests were repeated using clinically relevant concentrations: 5 mM glucose and 0.1 mM interferents (AA, UA, urea, fructose, mannose, lactose, galactose,maltose, and sucrose).33,43 As shown in the updated Fig. 14, the sensor response to glucose remained dominant, while the signal change due to interferents was negligible (<5%), confirming excellent selectivity in a physiologically relevant matrix. Additionally, NaCl was included not as an electroactive interferent but to simulate physiological ionic strength, as its presence can influence the electrochemical environment.123 The stable current response in the presence of NaCl further confirms the robustness of the electrode under biological conditions.44,80
 |
| | Fig. 14 Response to sequential injections of 5 mM glucose, 0.1 mM of ascorbic acid (AA), uric acid (UA), urea, fructose, mannose, lactose, galactose, maltose, saccharose, NaCl, and 5 mM glucose. | |
3.6. Stability, repeatability, and reproducibility
The stability of the non-enzymatic Fe2O3/CuFe2O4/GNPs/Gt sensor was evaluated via cyclic voltammetry in 0.5 M NaOH containing 2 mM glucose over 20 continuous cycles. As shown in Fig. 15a, the oxidation current in the 20th cycle retained 86.87% of its initial value, indicating good electrochemical stability during repeated use.81 To assess repeatability,38 the same electrode was tested six consecutive times under identical conditions (Fig. 15b), yielding a relative standard deviation (RSD) of 1.09%, which demonstrates excellent short-term consistency.122 Reproducibility was further examined using six independently fabricated electrodes tested with 2 mM glucose under the same conditions (Fig. 15c). The resulting RSD was 2.36%, confirming the reliability44 and consistency of the sensor fabrication process.
 |
| | Fig. 15 (a) CV curves of graphite/Fe3O4/CuFe2O4/GNPs/Gt in 2 mM glucose at NaOH 0.5 M for 20 cycles. (b) Repeatability of the voltammetric responses of 2 mM glucose in NaOH 0.5 M (repeatability n = 6). (c) reproducibility (n = 6 sensors). | |
3.7. Real sample examination
The evaluation of human serum samples using the newly developed non-enzymatic glucose sensor is summarized in Table 3. The relative standard deviation (RSD) values ranged from 0.92% to 2.96%, with an average of 1.92%, indicating strong measurement precision and reproducibility.107 These results closely matched those obtained from a commercial glucose biosensor used in a local hospital, demonstrating that the sensor's performance is not limited to standard glucose solutions. Instead, it is reliably applicable to real human serum samples, confirming its potential for practical clinical use. Similar studies have also reported high precision and accurate detection of glucose in human serum using non-enzymatic sensors based on metal oxide/graphene composites: Haghparas114 et al. demonstrated effective detection with RSD values below 3% in human serum samples using CuO/ZnO hollow microstructures,38,114 and Hussein, B,44 et al. confirmed good recovery and stability in serum and practical real sample analysis using ZnO/Co3O4/rGO/GCE.
Table 3 Validation of the Fe2O3/CuFe2O4/GNPs/Gt sensor against a standard commercial glucose meter
| Glucose concentration mM |
RSD (%) |
Recovery (%) |
| The proposed sensor |
The commercial instrument |
| 2.26 |
2.31 |
2.42 |
97.83 |
| 4.84 |
4.82 |
0.92 |
100.41 |
| 7.74 |
7.64 |
1.38 |
101.31 |
| 9.36 |
9.49 |
2.96 |
98.63 |
4. Conclusion
In this study, a simple synthesized methods for Fe2O3/CuFe2O4 nanocomposite was evaluated. Then, a novel non-enzymatic glucose sensor based on a Fe2O3/CuFe2O4/GNPs nanocomposite-modified electrode was successfully developed and characterized. The optimized composition (graphene
:
graphite ratio of 45
:
55 with 10% Fe2O3/CuFe2O4) demonstrated excellent electrocatalytic activity, high conductivity, and enhanced electron transfer kinetics, as confirmed by cyclic voltammetry and electrochemical impedance spectroscopy. The sensor exhibited high sensitivity, a low detection limit, and a wide linear range for glucose detection under optimal conditions in 0.5 M NaOH.The electrode showed outstanding stability, repeatability, and reproducibility, with minimal signal loss over repeated use and consistent performance across multiple fabricated sensors. Selectivity tests confirmed the sensor's ability to distinguish glucose in the presence of common interfering species. Furthermore, the sensor demonstrated strong agreement with commercial glucose biosensors in the analysis of real human serum samples, highlighting its potential for clinical and point-of-care applications. Overall, the developed sensor offers a promising, low-cost, and reliable platform for non-enzymatic glucose detection, with practical applicability in biomedical diagnostics and glucose monitoring.
Conflicts of interest
The authors declare that there is no conflict of interest.
Data availability
The authors confirm that the data supporting the findings of this study are available within the article.
References
- K. Ali, B. Ali, A. Jabbar, S. Iqbal, I. Ahmad and M. I. Bashir, Synthesis, Structural, Dielectric and Magnetic Properties of Cufe2o4/Mno2 Nanocomposites, J. Magn. Magn. Mater., 2017, 434, 30–36 CrossRef CAS.
- A. Ambrosi, C. K. Chua, A. Bonanni and M. Pumera, Electrochemistry of Graphene and Related Materials, Chem. Rev., 2014, 114, 7150–7188 CrossRef CAS PubMed.
- B. G. Amin, J. Masud and M. Nath, A Non-Enzymatic Glucose Sensor Based on a Coni 2 Se 4/Rgo Nanocomposite with Ultrahigh Sensitivity at Low Working Potential, J. Mater. Chem. B, 2019, 7, 2338–2348 RSC.
- A. Ali, K. Karim, O. Hakami, W. M. Alamier, S. K. Ali, T. Zelai, M. S. Rashid, A. K. Tareen and E. A. Al-Harthi, Recent Progress in Metal Oxide-Based Electrode Materials for Safe and Sustainable Variants of Supercapacitors, Front. Chem., 2024, 12, 1402563 CrossRef PubMed.
- K. Atacan, Cufe2o4/Reduced Graphene Oxide Nanocomposite Decorated with Gold Nanoparticles as a New Electrochemical Sensor Material for ʟ-Cysteine Detection, J. Alloys Compd., 2019, 791, 391–401 CrossRef CAS.
- A. Aykaç, H. Gergeroglu, B. Beşli, E. Ö. Akkaş, A. Yavaş, S. Güler, F. Güneş and M. Erol, An Overview on Recent Progress of Metal Oxide/Graphene/Cnts-Based Nanobiosensors, Nanoscale Res. Lett., 2021, 16, 65 CrossRef.
- A. Aykaç, and İ. Ok, Graphene/Metal Oxide-Based Nanocomposite for Electrochemical Sensors, in Optical Properties of Metal Oxide Nanostructures, Springer, 2023, pp. 331–70 Search PubMed.
- T. G. Satheesh Babu and T. Ramachandran, Development of Highly Sensitive Non-Enzymatic Sensor for the Selective Determination of Glucose and Fabrication of a Working Model, Electrochim. Acta, 2010, 55, 1612–1618 CrossRef.
- A. Bagheri, A. Fallah, J. Karczewski, A. Eslami, A. M. S. Asadi and G. Boczkaj, Investigating Covid-19 Active Pharmaceutical Ingredients (Apis) Degradation Using Peroxydisulfate/Femnox Binary Metal Oxide/Ultrasound System, Water Resour. Ind., 2024, 31, 100232 CrossRef CAS.
- G. Balkourani, T. Damartzis, A. Brouzgou and P. Tsiakaras, Cost Effective Synthesis of Graphene Nanomaterials for Non-Enzymatic Electrochemical Sensors for Glucose: A Comprehensive Review, Sensors, 2022, 22, 355 CrossRef CAS PubMed.
- A. J. Bard, L. R. Faulkner, and H. S. White, Electrochemical Methods: Fundamentals and Applications, John Wiley & Sons, 2022 Search PubMed.
- G. Beamson, High Relution Xps of Organic Polymers. The Scienta Esca 300 Database, ICIplc, 1993, 5, 10 Search PubMed.
- S. Berciaud, S. Ryu, L. E. Brus and T. F. Heinz, Probing the Intrinsic Properties of Exfoliated Graphene: Raman Spectroscopy of Free-Standing Monolayers, Nano Lett., 2009, 9, 346–352 CrossRef CAS PubMed.
- M. C. Biesinger, Advanced Analysis of Copper X-Ray Photoelectron Spectra, Surf. Interface Anal., 2017, 49, 1325–1334 CrossRef CAS.
- M. C. Biesinger, B. P. Payne, A. P. Grosvenor, L. W. M. Lau, A. R. Gerson and R. St C. Smart, Resolving Surface Chemical States in Xps Analysis of First Row Transition Metals, Oxides and Hydroxides: Cr, Mn, Fe, Co and Ni, Appl. Surf. Sci., 2011, 257, 2717–2730 CrossRef CAS.
- D. A. C. Brownson and C. E. Banks, Graphene Electrochemistry: An Overview of Potential Applications, Analyst, 2010, 135, 2768–2778 RSC.
- D. A. C. Brownson, C. W. Foster and C. E. Banks, The Electrochemical Performance of Graphene Modified Electrodes: An Analytical Perspective, Analyst, 2012, 137, 1815–1823 RSC.
- R. A. Sobh and H. S. Magar, Innovative formulation of a functional nano-copolymer derived from glycidyl methacrylate and acrylonitrile as an exceptionally sensitive and selective electrochemical sensor for folic acid detection in pharmaceutical and food samples, Nanoscale, 2025, 17, 18359–18376 RSC.
- S. Carinelli, P. A. Salazar-Carballo, J. E. Rosa Melián and F. García-García, Porous Copper Oxide Thin Film Electrodes
for Non-Enzymatic Glucose Detection, Chemosensors, 2023, 11, 260 CrossRef CAS.
- S. Cheng, S. DelaCruz, C. Chen, Z. Tang, T. Shi, C. Carraro and R. Maboudian, Hierarchical Co3o4/Cuo Nanorod Array Supported on Carbon Cloth for Highly Sensitive Non-Enzymatic Glucose Biosensing, Sens. Actuators, B, 2019, 298, 126860 CrossRef CAS.
- R. M. Cornell, and U. Schwertmann, The Iron Oxides: Structure, Properties, Reactions, Occurrences and Uses, John Wiley & Sons, 2003 Search PubMed.
- E. Cuara, U. Sierra, A. Mercado, E. D. Barriga-Castro, A. Cortés, C. Gallardo-Vega, M. Valle-Orta and S. Fernández, Synthesis of Copper Oxides-Graphene Composites for Glucose Sensing, Carbon Trends, 2021, 4, 100050 CrossRef CAS.
- B. D. Cullity, and S. R. Stock, Elements of X-Ray Diffraction, Prentice Hall, Upper Saddle River, NJ, 2001, p. 388 Search PubMed.
- Q. Dong, H. Ryu and Y. Lei, Metal Oxide Based Non-Enzymatic Electrochemical Sensors for Glucose Detection, Electrochim. Acta, 2021, 370, 137744 CrossRef CAS.
- D. R. Dreyer, S. Park, C. W. Bielawski and R. S. Ruoff, The Chemistry of Graphene Oxide, Chem. Soc. Rev., 2010, 39, 228–240 RSC.
- R. F. Egerton, Physical Principles of Electron Microscopy, Springer, 2016 Search PubMed.
- G. A. El-Fatah, H. S. Magar, R. Y. A. Hassan, R. Mahmoud, A. A. Farghali and M. E. M. Hassouna, A Novel Gallium Oxide Nanoparticles-Based Sensor for the Simultaneous Electrochemical Detection of Pb2+, Cd2+ and Hg2+ Ions in Real Water Samples, Sci. Rep., 2022, 12, 20181 CrossRef CAS PubMed.
- A. M. El-Kosasy, M. H. A. Rahman and S. H. Abdelaal, Graphene Nanoplatelets in Potentiometry: A Nanocomposite Carbon Paste and Pvc Based Membrane Sensors for Analysis of Vilazodone Hcl in Plasma and Milk Samples, Talanta, 2019, 193, 9–14 CrossRef CAS PubMed.
- G. Emir, Y. Dilgin, A. Ramanaviciene and A. Ramanavicius, Amperometric Nonenzymatic Glucose Biosensor Based on Graphite Rod Electrode Modified by Ni-Nanoparticle/Polypyrrole Composite, Microchem. J., 2021, 161, 105751 CrossRef CAS.
- A. C. Ferrari, F. Bonaccorso, V. Fal'Ko, K. S. Novoselov, S. Roche, P. Bøggild, S. Borini, F. H. L. Koppens, V. Palermo and N. Pugno, Science and Technology Roadmap for Graphene, Related Two-Dimensional Crystals, and Hybrid Systems, Nanoscale, 2015, 7, 4598–4810 RSC.
- F. Foroughi, M. Rahsepar, M. J. Hadianfard and H. Kim, Facile Synthesis and Electrochemical Performance of Graphene-Modified Cu2o Nanocomposite for Use in Enzyme-Free Glucose Biosensor, Iran. J. Chem. Chem. Eng., 2020, 39, 1–10 Search PubMed.
- A. García-Miranda Ferrari, C. W. Foster, P. J. Kelly, D. A. C. Brownson and C. E. Banks, Determination of the Electrochemical Area of Screen-Printed Electrochemical Sensing Platforms, Biosensors, 2018, 8, 53 CrossRef PubMed.
- R. Ghosh, L. Xiao and M. Z. Yates, Nonenzymatic Glucose Sensor Using Bimetallic Catalysts, ACS Appl. Mater. Interfaces, 2023, 16, 17–29 CrossRef.
- J. I. Goldstein, D. E. Newbury, J. R. Michael, N. W. M. Ritchie, J. H. J. Scott, and D. C. Joy, Scanning Electron Microscopy and X-Ray Microanalysis, Springer, 2017 Search PubMed.
- J. M. Gonçalves, L. V. de Faria, A. B. Nascimento, R. L. Germscheidt, S. Patra, L. P. Hernández-Saravia, J. A. Bonacin, R. A. A. Munoz and L. Angnes, Sensing Performances of Spinel Ferrites Mfe2o4 (M= Mg, Ni, Co, Mn, Cu and Zn) Based Electrochemical Sensors: A Review, Anal. Chim. Acta, 2022, 1233, 340362 CrossRef.
- M. Gotić and S. Musić, Mössbauer, Ft-Ir and Fe Sem Investigation of Iron Oxides Precipitated from Feso4 Solutions, J. Mol. Struct., 2007, 834, 445–453 CrossRef.
- A. P. Grosvenor, B. A. Kobe, M. C. Biesinger and N. S. McIntyre, Investigation of Multiplet Splitting of Fe 2p Xps Spectra and Bonding in Iron Compounds, Surf. Interface Anal., 2004, 36, 1564–1574 CrossRef CAS.
- Z. Haghparas, Z. Kordrostami, M. Sorouri, M. Rajabzadeh and R. Khalifeh, Highly Sensitive Non-Enzymatic Electrochemical Glucose Sensor Based on Dumbbell-Shaped Double-Shelled Hollow Nanoporous Cuo/Zno Microstructures, Sci. Rep., 2021, 11, 344 CrossRef CAS.
- M. S. Hashem and H. S. Magar, Creative Synthesis of Ph-Dependent Nanoporous Pectic Acid Grafted with Acrylamide and Acrylic Acid Copolymer as an Ultrasensitive and Selective Riboflavin Electrochemical Sensor in Real Samples, Int. J. Biol. Macromol., 2024, 280, 136022 CrossRef CAS.
- A. Heller and B. Feldman, Electrochemical Glucose Sensors and Their Applications in Diabetes Management, Chem. Rev., 2008, 108, 2482–2505 CrossRef CAS PubMed.
- D. Hernández-Ramírez, L. H. Mendoza-Huizar, C. A. Galán-Vidal, G. Y. Aguilar-Lira and G. A. Álvarez-Romero, Development of a Non-Enzymatic Glucose Sensor Based on Fe2o3 Nanoparticles-Carbon Paste Electrodes, J. Electrochem. Soc., 2022, 169, 067507 CrossRef.
- Z. Hsine, R. Mlika, N. Jaffrezic-Renault and H. Korri-Youssoufi, Review—Recent Progress in Graphene Based Modified Electrodes for Electrochemical Detection of Dopamine, Chemosensors, 2022, 10, 249 CrossRef CAS.
- Yu-W. Hsu, T.-K. Hsu, C.-L. Sun, Y.-T. Nien, N.-W. Pu and M.-D. Ger, Synthesis of Cuo/Graphene Nanocomposites for Nonenzymatic Electrochemical Glucose Biosensor Applications, Electrochim. Acta, 2012, 82, 152–157 CrossRef CAS.
- B. A. Hussein, A. A. Tsegaye, G. Shifera and A. M. Taddesse, A Sensitive Non-Enzymatic Electrochemical Glucose Sensor Based on a Zno/Co 3 O 4/Reduced Graphene Oxide Nanocomposite, Sens. Diagn., 2023, 2, 347–360 RSC.
- H. Jeong, L. Ku Kwac, C. G. Hong and H. G. Kim, Direct Growth of Flower Like-Structured Cufe Oxide on Graphene Supported Nickel Foam as an Effective Sensor for Glucose Determination, Mater. Sci. Eng., C, 2021, 118, 111510 CrossRef CAS PubMed.
- D. Jiang, Q. Liu, K. Wang, J. Qian, X. Dong, Z. Yang, X. Du and B. Qiu, Enhanced Non-Enzymatic Glucose Sensing Based on Copper Nanoparticles Decorated Nitrogen-Doped Graphene, Biosens. Bioelectron., 2014, 54, 273–278 CrossRef CAS PubMed.
- L. Jothi, N. Jayakumar, S. K. Jaganathan and G. Nageswaran, Ultrasensitive and Selective Non-Enzymatic Electrochemical Glucose Sensor Based on Hybrid Material of Graphene Nanosheets/Graphene Nanoribbons/Nickel Nanoparticle, Mater. Res. Bull., 2018, 98, 300–307 CrossRef CAS.
- L. Kabir, Y. J. Joo, K. Y. Cho and W.-C. Oh, Electrochemical Approach for Nonenzymatic Glucose Sensing with Noble Metal-Free 2d Graphene-Based Ternary Nanocomposite, Sens. Actuators, A, 2024, 372, 115367 CrossRef CAS.
- P. K. Kalambate, N. Larpant, R. P. Kalambate, W. Niamsi, V. Primpray, C. Karuwan and W. Laiwattanapaisal, A Portable Smartphone-Compatible Ratiometric Electrochemical Sensor with Ultrahigh Sensitivity for Anticancer Drug Mitoxantrone Sensing, Sens. Actuators, B, 2023, 378, 133103 CrossRef CAS.
- P. V. Kamat, Graphene-Based Nanoarchitectures. Anchoring Semiconductor and Metal Nanoparticles on a Two-Dimensional Carbon Support, J. Phys. Chem. Lett., 2010, 1, 520–527 CrossRef CAS.
- A. Kaniyoor and S. Ramaprabhu, A Raman Spectroscopic Investigation of Graphite Oxide Derived Graphene, AIP Adv., 2012, 2, 032183 CrossRef.
- A. Kumar, P. Choudhary, A. Kumar, H. C. C. Pedro and V. Krishnan, Recent Advances in Plasmonic Photocatalysis Based on Tio2 and Noble Metal Nanoparticles for Energy Conversion, Environmental Remediation, and Organic Synthesis, Small, 2022, 18, 2101638 CrossRef CAS.
- A. Lasia, Electrochemical Impedance Spectroscopy and Its Applications, in Modern Aspects of Electrochemistry, Springer, 2002, pp. 143–248 Search PubMed.
- E. J. J. Laviron, General Expression of the Linear Potential Sweep Voltammogram in the Case of Diffusionless Electrochemical Systems, J. Electroanal. Chem. Interfacial Electrochem., 1979, 101, 19–28 CrossRef CAS.
- R. Li, X. Deng and L. Xia, Non-Enzymatic Sensor for Determination of Glucose Based on Ptni Nanoparticles Decorated Graphene, Sci. Rep., 2020, 10, 16788 CrossRef CAS PubMed.
- J. Liu, J. Shen, S. Ji, Q. Zhang and W. Zhao, Research Progress of Electrode Materials for Non-Enzymatic Glucose Electrochemical Sensors, Sens. Diagn., 2023, 2, 36–45 RSC.
- L. Liu, J. Wang, C. Wang and G. Wang, Facile Synthesis of Graphitic Carbon Nitride/Nanostructured Α-Fe2o3 Composites and Their Excellent Electrochemical Performance for Supercapacitor and Enzyme-Free Glucose Detection Applications, Appl. Surf. Sci., 2016, 390, 303–310 CrossRef CAS.
- L. Luo, J. Cui, Y. Wang, Y. Wang, H. Zheng, Y. Qin, X. Shu, D. Yu, Y. Zhang and Y. Wu, Synthesis of Nio/Fe2o3 Nanocomposites as Substrate for the Construction of Electrochemical Biosensors, J. Solid State Electrochem., 2018, 22, 1763–1770 CrossRef CAS.
- R. Lv, E. Cruz-Silva and M. Terrones, Building Complex Hybrid Carbon Architectures by Covalent Interconnections: Graphene–Nanotube Hybrids and More, ACS Nano, 2014, 8, 4061–4069 CrossRef CAS PubMed.
- M. E. G. Lyons and M. P. Brandon, The Oxygen Evolution Reaction on Passive Oxide Covered Transition Metal Electrodes in Aqueous Alkaline Solution. Part 1-Nickel, Int. J. Electrochem. Sci., 2008, 3, 1386–1424 CrossRef CAS.
- H. S. Magar, H. Abdelghany, M. N. Abbas, U. Bilitewski and R. Y. A. Hassan, Fast Analysis of Staphylococcus Aureus in Food Products Using Disposable Label-Free Nano-Electrochemical Immunosensor Chips, Microchem. J., 2023, 193, 109097 CrossRef CAS.
- H. S. Magar, P. K. Brahman and R. Y. A. Hassan, Disposable Impedimetric Nano-Immunochips for the Early and Rapid Diagnosis of Vitamin-D Deficiency, Biosens. Bioelectron.:X, 2022, 10, 100124 CAS.
- H. S. Magar, E. S. M. Duraia and R. Y. A. Hassan, Dopamine Fast Determination in Pharmaceutical Products Using Disposable Printed Electrodes Modified with Bimetal Oxides Carbon Nanotubes Nanocomposite, Sci. Rep., 2025, 15, 11229 CrossRef CAS PubMed.
- H. S. Magar, A. M. El Nahrawy, R. Y. A. Hassan and A. B. Abou Hammad, Nanohexagonal Iron Barium Titanate Nanoparticles Surface-Modified Nife2o4 Composite Screen-Printed Electrode for Enzymatic Glucose Monitoring, RSC Adv., 2024, 14, 34948–34963 RSC.
- H. S. Magar and A. M. Fahim, Sensing Platform Based on Rgo/Cellulose-Triazole Composite for the Electrochemical Detection of Mercury (Ii) Ions in Food Samples, Microchem. J., 2025, 215, 114243 CrossRef CAS.
- H. S. Magar, A. M. Fahim and M. S. Hashem, Accurate, Affordable, and Easy Electrochemical Detection of Ascorbic Acid in Fresh Fruit Juices and Pharmaceutical Samples Using an Electroactive Gelatin Sulfonamide, RSC Adv., 2024, 14, 39820–39832 RSC.
- H. S. Magar, M. Fayez, F. Febbraio and R. Y. A. Hassan, Esterase-2 Mutant-Based Nanostructured Amperometric Biosensors for the Selective Determination of Paraoxon (Neurotoxin), Anal. Biochem., 2025, 698, 115751 CrossRef CAS PubMed.
- H. S. Magar, M. S. Hashem and R. A. Sobh, Design of Metal Oxide Nanoparticles-Embedded Polymeric Nanocomposites for Hydrogen Peroxide Chronoamperometric Sensor, Polym. Compos., 2024, 45, 3653–3665 CrossRef CAS.
- H. S. Magar, R. Y. A. Hassan and M. N. Abbas, Non-Enzymatic Disposable Electrochemical Sensors Based on Cuo/Co3o4@Mwcnts Nanocomposite Modified Screen-Printed Electrode for the Direct Determination of Urea, Sci. Rep., 2023, 13, 2034 CrossRef CAS PubMed.
- H. S. Magar, B. A. Hemdan, H. R. M. Rashdan and R. Y. A. Hassan, Rapid and Selective Detection of Foodborne Pathogens Using a Disposable Bio-Sensing System Designed by Stepwise Antibody Immobilization on Aunps@Cu-Mof Nanocomposite, J. Anal. Test., 2024, 8, 478–492 CrossRef.
- H. S. Magar, E. E. A. E. Magd, R. Y. A. Hassan and A. M. Fahim, Rapid Impedimetric Detection of Cadmium Ions Using Nanocellulose/Ligand/Nanocomposite (Cnt/Co3o4), Microchem. J., 2022, 182, 107885 CrossRef CAS.
- H. M. Mahdy, Y. M. Abbas, H. S. Magar, R. Y. A. Hassan and E. S. M. Duraia, Silver Vanadate and Cerium Oxide Decorated Graphene Oxide Ternary Heterostructure Nanocomposites: Structural, Electrochemical, and Optoelectrical Properties, J. Mater. Sci.: Mater. Electron., 2024, 35, 1891 CrossRef CAS.
- A. R. Mahfouz, H. G. Nada, H. S. Magar, K. Z. ElBaghdady and R. Y. A. Hassan, Nano-Biosensors for Rapid Detection of Antibiotic Resistance Genes Blactx-M in Escherichia Coli and Blakpc in Klebsiella Pneumoniae, Int. J. Biol. Macromol., 2025, 310, 143216 CrossRef CAS.
- A. Mahmoud, M. Echabaane, O. Karim, J. Boudon, L. Saviot, N. Millot and R. Ben Chaabane, Cu-Doped Zno Nanoparticles for Non-Enzymatic Glucose Sensing, Molecules, 2021, 26, 929 CrossRef CAS PubMed.
- L. M. Malard, M. Assunção Pimenta, G. Dresselhaus and M. S. Dresselhaus, Raman Spectroscopy in Graphene, Phys. Rep., 2009, 473, 51–87 CrossRef CAS.
- S. Mallesh, G. Minji and Ki H. Kim, Cubic to Tetragonal Phase Transition in Cufe2o4 Nanoparticles, J. Magn., 2021, 26, 7 CrossRef.
- A. M. Mansour, H. S. Magar, A. Elzwawy, A. B. Abou Hammad and A. M. El Nahrawy, Structural, Optical, and Electrochemical Properties of Tungsten-Doped Cadmium Zinc Phosphate Nanoporous Materials for Energy Storage and Peroxide Detection, RSC Adv., 2025, 15, 15670–15693 RSC.
- A. James Marsden, D. G. Papageorgiou, C. Valles, A. Liscio, V. Palermo, M. A. Bissett, R. J. Young and I. A. Kinloch, Electrical Percolation in Graphene–Polymer Composites, 2D Materials, 2018, 5, 032003 CrossRef.
- M. Mbebou, S. Polat and H. Zengin, Sustainable Cauliflower-Patterned Cufe2o4 Electrode Production from Chalcopyrite for Supercapacitor Applications, Nanomaterials, 2023, 13, 1105 CrossRef CAS PubMed.
- A. Mohammadpour-Haratbar, S. Mohammadpour-Haratbar, Y. Zare, K. Y. Rhee and S.-J. Park, A Review on Non-Enzymatic Electrochemical Biosensors of Glucose Using Carbon Nanofiber Nanocomposites, Biosensors, 2022, 12, 1004 CrossRef CAS PubMed.
- G. A. Naikoo, H. Salim, I. U. Hassan, T. Awan, F. Arshad, M. Z. Pedram, W. Ahmed and A. Qurashi, Recent Advances in Non-Enzymatic Glucose Sensors Based on Metal and Metal Oxide Nanostructures for Diabetes Management-a Review, Front. Chem., 2021, 9, 748957 CrossRef CAS PubMed.
- K. Nakamoto, Infrared and Raman Spectra of Inorganic and Coordination Compounds, Part B: Applications in Coordination, Organometallic, and Bioinorganic Chemistry, John Wiley & Sons, 2009 Search PubMed.
- X. Nan, Y. Zhang, J. Shen, R. Liang, J. Wang, L. Jia, X. Yang, W. Yu and Z. Zhang, A Review of the Establishment of Effective Conductive Pathways of Conductive Polymer Composites and Advances in Electromagnetic Shielding, Polymers, 2024, 16, 2539 CrossRef CAS PubMed.
- A. V. Naumkin, A. Kraut-Vass, S. W. Gaarenstroom, and C. J. Powell, Nist X-Ray Photoelectron Spectroscopy Database, Nist Standard Reference Database 20, Version 4.1, US Department of Commerce, Washington, 10 ( 2012), p. T4T88K Search PubMed.
- J. Navarro-Laboulais, J. J. García-Jareño, J. Agrisuelas and F. Vicente, Eis Behavior of Polyethylene + Graphite Composite Considered as an Approximation to an Ensemble of Microelectrodes, J. Compos. Sci., 2024, 8, 378 CrossRef CAS.
- D. Neravathu, A. R. Paloly and M. J. Bushiri, Hybrid Nanomaterial of Znfe2o4/Α-Fe2o3 Implanted Graphene for Electrochemical Glucose Sensing Application, Diamond Relat. Mater., 2020, 106, 107852 CrossRef CAS.
- P. Pazhamalai, V. Krishnan, M. S. M. Saleem, S.-J. Kim and H.-W. Seo, Investigating Composite Electrode Materials of Metal Oxides for Advanced Energy Storage Applications, Nano Convergence, 2024, 11, 30 CrossRef CAS PubMed.
- S. M. Ramteke, M. Walczak, M. De Stefano, A. Ruggiero, A. Rosenkranz and M. Marian, 2d Materials for Tribo-Corrosion and-Oxidation Protection: A Review, Adv. Colloid Interface Sci., 2024, 331, 103243 CrossRef CAS PubMed.
- S. Rodsud and W. Limbut, A Simple Electrochemical Sensor Based on Graphene Nanoplatelets Modified Glassy
Carbon Electrode (Grnps/Gce) for Highly Sensitive Detection of Yohimbine (Yoh), J. Electrochem. Soc., 2019, 166, B771 CrossRef CAS.
- S. S. Selima, M. Khairy and M. A. Mousa, Comparative Studies on the Impact of Synthesis Methods on Structural, Optical, Magnetic and Catalytic Properties of Cufe2o4, Ceram. Int., 2019, 45, 6535–6540 CrossRef CAS.
- I. Shackery, U. Patil, A. Pezeshki, N. M. Shinde, S. Kang, S. Im and S. Chan Jun, Copper Hydroxide Nanorods Decorated Porous Graphene Foam Electrodes for Non-Enzymatic Glucose Sensing, Electrochim. Acta, 2016, 191, 954–961 CrossRef CAS.
- Z. Shaghaghi, A. Saghebi Shalghuni and S. Jafari, Fe2o3/Cuo Hybrid Nanostructures as Electrode Materials for Glucose Detection, Anal. Bioanal. Chem. Res., 2024, 11, 343–360 CAS.
- Z. Shahnavaz, P. M. Woi and Y. Alias, A Hydrothermally Prepared Reduced Graphene Oxide-Supported Copper Ferrite Hybrid for Glucose Sensing, Ceram. Int., 2015, 41, 12710–12716 CrossRef CAS.
- M. S. Hashem and H. S. Magar, Inventive pectic acid grafted with polyacrylamide for a highly sensitive and selective non-enzymatic dopamine sensor in pharmaceutical samples, RSC Adv., 2025, 15, 33758–33772 RSC.
- M. D. P. Silva, F. C. Silva, F. S. M. Sinfrônio, A. R. Paschoal, E. N. Silva and C. W. A. Paschoal, The Effect of Cobalt Substitution in Crystal Structure and Vibrational Modes of Cufe2o4 Powders Obtained by Polymeric Precursor Method, J. Alloys Compd., 2014, 584, 573–580 CrossRef CAS.
- A. J. Slate, D. A. C. Brownson, S. A. D. Ahmed, G. C. Smith, K. A. Whitehead and C. E. Banks, Exploring the Electrochemical Performance of Graphite and Graphene Paste Electrodes Composed of Varying Lateral Flake Sizes, Phys. Chem. Chem. Phys., 2018, 20, 20010–20022 RSC.
- G. Socrates, Infrared and Raman Characteristic Group Frequencies: Tables and Charts, John Wiley & Sons, 2004 Search PubMed.
- S. Stankovich, D. A. Dikin, G. H. B. Dommett, K. M. Kohlhaas, E. J. Zimney, E. A. Stach, R. D. Piner, S. T. Nguyen and R. S. Ruoff, Graphene-Based Composite Materials, Nature, 2006, 442, 282 CrossRef CAS PubMed.
- D. Stauffer, and A. Aharony, Introduction to Percolation Theory, Taylor & Francis, 2018 Search PubMed.
- I. Švancara, K. Vytřas, K. Kalcher, A. Walcarius and J. Wang, Carbon Paste Electrodes in Facts, Numbers, and Notes: A Review on the Occasion of the 50-Years Jubilee of Carbon Paste in Electrochemistry and Electroanalysis, Electroanalysis, 2009, 21, 7–28 CrossRef.
- Y. Tian, Q. Wang, J. Niu, J. Liang, Y. Geng, X. Liu, J. Wang and M. Ma, Cuco-Based Carbonate Hydrosulfide Nanoarray Supported on Carbon Cloth as Electrocatalysts for Glucose Sensing, ACS Appl. Nano Mater., 2024, 7, 25249–25258 CrossRef CAS.
- K. E. Toghill and R. G. Compton, Metal Nanoparticle Modified Boron Doped Diamond Electrodes for Use in Electroanalysis, Electroanalysis, 2010, 22, 1947–1956 CrossRef CAS.
- M. G. Trachioti, A. Ch Lazanas and M. I. Prodromidis, Shedding Light on the Calculation of Electrode Electroactive Area and Heterogeneous Electron Transfer Rate Constants at Graphite Screen-Printed Electrodes, Microchim. Acta, 2023, 190, 251 CrossRef CAS PubMed.
- T. U. Tran Thi, Van H. Phan, H. T. P. Nguyen, T. L. Nguyen, An N. Vu and K. Le Tien, Synthesis of Magnetic Cufe2o4/Fe2o3 Core-Shell Materials and Their Application in Photo-Fenton-Like Process with Oxalic Acid as a Radical-Producing Source, Asian Ceram. Soc., 2021, 9, 1091–1102 CrossRef.
- A. T. E. Vilian, B. Dinesh, M. Rethinasabapathy, S.-K. Hwang, C.-S. Jin, Y. S. Huh and Y.-K. Han, Hexagonal Co 3 O 4 Anchored Reduced Graphene Oxide Sheets for High-Performance Supercapacitors and Non-Enzymatic Glucose Sensing, J. Mater. Chem. A, 2018, 6, 14367–14379 RSC.
- M. Wang, M. Wang, T. Shen, M. Wang, D. Zhang, Z. Tong and J. Chen, One-Pot Synthesis of Α-Fe2o3 Nanoparticles-Decorated Reduced Graphene Oxide for Efficient Nonenzymatic H2o2 Biosensor, Sens. Actuators, B, 2014, 190, 645–650 CrossRef CAS.
- M. Wang, M. Shi, E. Meng, F. Gong and F. Li, Non-Enzymatic Glucose Sensor Based on Three-Dimensional Hierarchical Co3o4 Nanobooks, Micro Nano Lett., 2020, 15, 191–195 CrossRef CAS.
- M. Wei, Y. Qiao, H. Zhao, J. Liang, T. Li, Y. Luo, S. Lu, X. Shi, W. Lu and X. Sun, Electrochemical Non-Enzymatic Glucose Sensors: Recent Progress and Perspectives, Chem. Commun., 2020, 56, 14553–14569 RSC.
- J.-B. Wu, M.-L. Lin, X. Cong, He-N. Liu and P.-H. Tan, Raman Spectroscopy of Graphene-Based Materials and Its Applications in Related Devices, Chem. Soc. Rev., 2018, 47, 1822–1873 RSC.
- P. Wu, Q. Shao, Y. Hu, J. Jin, Y. Yin, H. Zhang and C. Cai, Direct Electrochemistry of Glucose Oxidase Assembled on Graphene and Application to Glucose Detection, Electrochim. Acta, 2010, 55, 8606–8614 CrossRef CAS.
- H. Xia, J. Li, L. Ma, Q. Liu and J. Wang, Electrospun Porous Cufe2o4 Nanotubes on Nickel Foam for Nonenzymatic Voltammetric Determination of Glucose and Hydrogen Peroxide, J. Alloys Compd., 2018, 739, 764–770 CrossRef CAS.
- K. Xu and W. Zheng, Fabrication of Graphene-Based Ammonia Sensors: A Review, Curr. Nanosci., 2024, 20, 578–598 CrossRef CAS.
- T. Yamashita and P. Hayes, Analysis of Xps Spectra of Fe2+ and Fe3+ Ions in Oxide Materials, Appl. Surf. Sci., 2008, 254, 2441–2449 CrossRef CAS.
- D. Yang, Y. Chen, S. Che and K. Wang, Recent Advances in Non-Enzymatic Glucose Sensors Based on Nanomaterials, Coatings, 2025, 15, 892 CrossRef CAS.
- J. Yu, X. Shan, D. Zhou, X. Zhao and W. Sheng, Non-Enzymatic Glucose Detection by Fe2o3 Nanorods-Reduced Graphene under Physiological Ph, Curr. Pharm. Anal., 2024, 20, 275–282 CrossRef.
- B. Zhang, Y. Yang, Y. Liu, Z.-D. Huang, Y.-B. He and J.-K. Kim, Percolation Threshold of Graphene Nanosheets as Conductive Additives in Li 4 Ti 5 O 12 Anodes of Li-Ion Batteries, Nanoscale, 2013, 5, 2100–2106 RSC.
- L. Zhang, F. Zhang, X. Yang, G. K. Long, Y. P. Wu, T. F. Zhang, K. Leng, Y. Huang, Y. F. Ma and A. Yu, Porous 3d Graphene-Based Bulk Materials with Exceptional High Surface Area and Excellent Conductivity for Supercapacitors, Sci. Rep., 2013, 3, 1408 CrossRef PubMed.
- X. Zhang, Z. Zhang, Q. Liao, S. Liu, Z. Kang and Y. Zhang, Nonenzymatic Glucose Sensor Based on in Situ Reduction of Ni/Nio-Graphene Nanocomposite, Sensors, 2016, 16, 1791 CrossRef PubMed.
- Y. Zhang, E. Zhou, Y. Li and X. He, A Novel Nonenzymatic Glucose Sensor Based on Magnetic Copper Ferrite Immobilized on Multiwalled Carbon Nanotubes, Anal. Methods, 2015, 7, 2360–2366 RSC.
- X. Zhao, D. Chu, X. Zhang, C. Gao, Y. He and W. Bai, Growth and Characterization of Carbon Nanotubes and Study of Modified Carbon Fiber—a Review, Diamond Relat. Mater., 2024, 147, 111308 CrossRef CAS.
- D.-L. Zhou, J.-J. Feng, L.-Y. Cai, Q.-X. Fang, J.-R. Chen and A.-J. Wang, Facile Synthesis of Monodisperse Porous Cu2o Nanospheres on Reduced Graphene Oxide for Non-Enzymatic Amperometric Glucose Sensing, Electrochim. Acta, 2014, 115, 103–108 CrossRef CAS.
- C. Zhu, G. Yang, H. Li, D. Du and Y. Lin, Electrochemical Sensors and Biosensors Based on Nanomaterials and Nanostructures, Anal. Chem., 2015, 87, 230–249 CrossRef CAS PubMed.
- H. Zhu, F. Shi, M. Peng, Y. Zhang, S. Long, R. Liu, J. Li and Z. Yang, Non-Enzymatic Electrochemical Glucose Sensors Based on Metal Oxides and Sulfides: Recent Progress and Perspectives, Chemosensors, 2025, 13, 19 CrossRef CAS.
- H. Zhu, L. Li, W. Zhou, Z. Shao and X. Chen, Advances in Non-Enzymatic Glucose Sensors Based on Metal Oxides, J. Mater. Chem. B, 2016, 4, 7333–7349 RSC.
- Y. Zhu, S. Murali, W. Cai, X. Li, Ji W. Suk, J. R. Potts and R. S. Ruoff, Graphene and Graphene Oxide: Synthesis, Properties, and Applications, Adv. Mater., 2010, 22, 3906–3924 CrossRef CAS PubMed.
- A. Zoppi, C. Lofrumento, E. M. Castellucci and Ph Sciau, Al-for-Fe Substitution in Hematite: The Effect of Low Al Concentrations in the Raman Spectrum of Fe2o3, J. Raman Spectrosc., 2008, 39, 40–46 CrossRef CAS.
|