The quantification of inorganic ions such as sodium and phosphate ions in pharmaceutical formulations is crucial for ensuring the consistency and safety of the drug product. This study presents a robust and selective HPLC method coupled with an Evaporative Light Scattering Detector (ELSD) for the simultaneous determination of sodium and phosphate ions in aripiprazole extended release injectable suspensions. A trimodal stationary phase column, offering a combination of reversed-phase/cation-exchange/anion-exchange mechanisms, was employed to enhance selectivity and retention of the highly polar analytes. The use of ELSD enabled the detection of non-chromophoric inorganic ions with satisfactory sensitivity and reproducibility. Validation of the method was performed in accordance with ICH guidelines, demonstrating acceptable linearity (R2 > 0.99), precision (RSD < 10%), accuracy (recoveries between 95–105%) and LOD values suitable for routine quality control. The method showed adequate robustness in the studied formulations. This novel application of a trimodal column with HPLC-ELSD offers a powerful alternative to ion chromatography or ICP-MS, providing a simpler and cost-effective approach for inorganic ion analysis in complex pharmaceutical matrices.
Separation techniques and particularly high-performance liquid chromatography (HPLC) are fundamental in pharmaceutical analysis. Among them, reversed-phase HPLC (RP-HPLC) using hydrophobic stationary phases (e.g., C8, C18, phenyl) with UV detection is considered the gold standard.2 However, RP-HPLC is less effective for polar or charged species (including inorganic anions or cations) often yielding poor retention and peak shapes. To overcome these limitations, ion-exchange and mixed-mode stationary phases are employed offering enhanced selectivity and orthogonality.2 Despite its usefulness, ion-exchange chromatography suffers from long run and equilibration times, limited detector compatibility, and high maintenance requirements; factors that reduce its suitability for routine pharmaceutical applications, especially when simultaneous cation and anion separation is needed.3
Mixed-mode stationary phases integrate multiple retention mechanisms such as reversed-phase, ion-exchange, and hydrophilic interaction chromatography (HILIC) within a single ligand on the stationary surface.4,5 By adjusting the mobile phase pH, the ionization of functional groups can be controlled, enabling the retention of weakly ionizable compounds. The presence of polar and ionizable groups also supports HILIC-like retention at lower acetonitrile concentrations, enhancing column capacity and analyte loading.6,7 Retention can be finely tuned by modifying the organic solvent content, pH, or buffer concentration. These columns offer high flexibility, orthogonality, and broad applicability for simultaneous separation of polar and non-polar pharmaceuticals, making them suitable for API, impurity, and degradation product analysis.5,8 They are also compatible with mass spectrometric detection.9
Since most pharmaceutical counterions and excipients lack UV-absorbing chromophores, detectors such as conductivity and refractive index (RI) have commonly been employed. However, these detection modes often suffer from low sensitivity or face compatibility challenges with gradient elution. ELSD and charged aerosol detection (CAD) have emerged as novel HPLC detectors.10 Both detection techniques are universal and suitable for non-volatile and semi-volatile compounds but the exhibits inherent non-linearity in response. Generally, ELSD provides higher limit of detection compared to CAD or conductivity detectors. The characteristics of each detection system are summarized in Table S1 (SI). The principle of operation is based on the evaporation of the column effluent using a stream of nitrogen gas introduced into a nebulizing chamber. Large droplets are removed, while the remaining aerosol is transported through a drying tube, where it is further desolvated into particles ranging from a few nanometers to several hundred nanometers in size. Detection is subsequently performed by measuring either the charged particles or the intensity of the scattered light.10 Coupling ion-exchange chromatography with the aforementioned detection systems presents challenges, primarily due to the low volatility of the eluents. Salt precipitation can occur, potentially damaging the nebulizer and optical cell, and leading to the need for frequent maintenance.11
The purpose of this research work is to develop a HPLC-ELSD method that can simultaneously separate positively and negatively charged inorganic ions (sodium and phosphate ions) in aripiprazole extended-release injectable suspensions. Compared to conventional ion-exchange chromatography, ICP-MS, and mixed-mode HILIC techniques, the proposed approach provides a balanced combination of sensitivity, robustness, and ease of implementation. It operates on standard HPLC instrumentation without requiring suppressors, high-purity argon plasma, costly equipment, or mobile phases rich in organic solvents. Although ICP-MS offers excellent elemental sensitivity, its application for sodium and phosphate determination is limited by matrix effects, space-charge interferences, high ionization potential of sodium, memory effects which can lead to unstable signals and poor quantitation.12 These techniques also demand rigorous sample preparation and frequent maintenance, making them less practical for routine quality control of complex pharmaceutical matrices.13 To the best of our knowledge, no other analytical method has been reported for this purpose up to date. The complexity of samples containing aripiprazole API and other species supports the use of trimodal-based stationary phase (HILIC, cation- and anion-exchange properties), which offers significant flexibility in adjusting various parameters to enhance analyte selectivity (Fig. S1). The developed analytical scheme was validated according to ICH guidelines and finally applied to both generic and reference products.
Individual solutions of carboxymethyl cellulose sodium, mannitol, and sodium hydrogen phosphate were prepared by weighing the appropriate amounts of each solid, transferring them to separate volumetric flasks, and diluting with water to obtain final concentrations corresponding to the product composition.
According to the column manufacturer, optimal separations are typically achieved using a mobile phase composed of aqueous HCOONH4 buffer solution and ACN. The pH of the buffer solution was evaluated within the range of 2.5 to 4.0. Fig. 1A indicates the retention time of each ion and the resolution dependence of the buffer pH. All other experimental conditions were maintained consistently at different pH values. The retention time of phosphate ion remains almost unaffected in the tested pH range due to its persistent monovalent negative charge (pKa1 = 1.96, pKa2 = 7.12, and pKa3 = 12.32) and the constant positive charge of the amino groups of stationary phase. On the other hand, longer retention times were observed for sodium ion due to the dissociation of weak anion-exchange groups (–COO−) of the stationary phase which resulted in stronger coulombic attraction. At lower pH value the resolution (Rs) between ions was less than 1.5 but it increased linearly at higher pH values. The pH value of 3.2 was selected as optimum since it provides lower retention time of the analytes and adequate resolution (Rs = 7.4). Additionally, this value is withing the buffering range of the HCOONH4 solution and the recommended pH range of the analytical column.
![]() | ||
| Fig. 1 (A) Effect of the buffer pH and (B) % ACN content of the mobile phase on the retention time and resolution of phosphate and sodium ions. | ||
The ACN content was investigated between 20–40% v/v. As can be observed in Fig. 1B, the phosphate ion exhibited stronger retention than the sodium ion. This is likely due to the weak anion-exchange nature of the anion-exchange domains in the Amaze TH stationary phase. The ionization of these binding sites decreases with increasing ACN content, thereby weakening the electrostatic interactions with anions. In contrast, the cation-exchange domains of the column as strong cation exchangers, with their ionization remaining stable despite changes in the organic solvent content. Similar findings have been previously reported by Liu et al.7
In ion-exchange mode, the retention of ionic species will reduce with the increase of buffer ionic strength. The capacity factor (k) of an analyte typically exhibits a relationship with the mobile phase concentration described by the following equation:
log k = log ko − n log C |
Fig. 2 presents chromatograms illustrating the separation of the analytes under varying concentrations of buffer solution. As the HCOONH4 concentration increased from 10 to 30 mM, a decrease in the retention times of the ions was observed. At higher salt levels, reduced separation efficiency and partial peak splitting of the phosphate ion were observed. Fundamentally, the retention time of an ion reflects the change in the standard Gibbs free energy of transfer of the analyte between the mobile phase and the stationary phase, as influenced by variations in the salt concentration of the mobile phase.8 At moderate buffer strength (20 mM, pH 3.2), sufficient ionization of both analytes and stable electrostatic interactions with the mixed-mode sites of the stationary phase were achieved, resulting in symmetrical peaks and reproducible retention. Therefore, 20 mM HCOONH4 at pH 3.2 was adopted for further experiments.
![]() | ||
| Fig. 2 Effect of HCOONH4 concentration, (A) 10 mM, (B) 20 mM and (C) 30 mM on the separation of the analytes. | ||
The effect of the column temperature on the peak symmetry of the analytes peaks was studied in the range of 25–40 °C. As is illustrated in Fig. 3, the retention time of both analytes were not changed over the examined temperature range. However, improved peak symmetry was observed for the phosphate ion at higher temperature, while the peak symmetry of the sodium ion remained almost unaffected. Thus, the value of 40 °C was selected as optimum column temperature.
![]() | ||
| Fig. 3 Effect of column temperature, (A) 25 °C, (B) 30 °C and (C) 40 °C on the separation of the analytes. | ||
The selectivity of the method was investigated by examining the chromatographic potential interferences by analysing a standard solution, sample solution, placebo and individual excipients such as carboxymethyl cellulose, mannitol, and dihydrogen phosphate ions. The analysis showed no co-elution of peaks originating from the sample components and excipients with the peaks of interest, so it can be concluded that the developed method is selective.
ELSD exhibits a non-linear response, as the amount of analyte detected is related to the intensity of the scattered light through an exponential relationship.15 However, a linear calibration curve can be constructed after double logarithmic transformation of the exponential equation. In our case, second-order polynomial regression (area = Ax2 + Bx + C) was utilized to describe the relationship between peak area and analyte concentration as it yielded slightly higher correlation coefficient compared to logarithmic one. Calibration curves were constructed in the range of 25–75 μg mL−1 and 50–150 μg mL−1 for phosphate and sodium ion, respectively. The correlation coefficients (R2) were greater than 0.998 for both ions, indicating strong correlation within the specified concentration range (see Fig. 4). Furthermore, residuals analysis revealed no evidence of outliers or influential points across the entire concentration range, indicating homoscedasticity and supporting the reliability of the model. The normality of the residuals for the analytes was confirmed by the Shapiro–Wilk test, with p-values of 0.9337 and 0.9502 for phosphate and sodium ions, respectively, thereby validating the assumptions underlying the polynomial model. Other validation parameters are tabulated in Table 1. The recovery (R%) values for both ions ranged between 95% and 105%, with % RSD not exceeding 10%, meeting the acceptance criteria and confirming the method's accuracy.16 Representative HPLC-ELSD chromatograms of the analysis of samples spiked at concentration levels of 50, 100 and 150% of the specification level of each analyte are shown in Fig. S2 (SI). Intraday precision was assessed by a single analyst on the same instrument, whereas inter-day precision involved different analysts and instruments across separate days. Both repeatability and interday precision (expressed as % RSD), were below 4.1%, demonstrating the method's satisfactory precision. The limits of detection (LOD) and quantification (LOQ) were established based on signal-to-noise ratios of 3 and 10, respectively. The LODs were found to be 0.61 and 0.13 μg mL−1 for phosphate and sodium ion, respectively.
![]() | ||
| Fig. 4 Calibration curves of (A) phosphate ion and (B) sodium ion. The inserts display the residual plots corresponding to each calibration curve. | ||
| Compound | Linear range (μg mL−1) | Calibration curve Y = AX2 + BX + C | R2 | Precisiona (% RSD) | Accuracyb (% RR) | LOD (μg mL−1) | ||
|---|---|---|---|---|---|---|---|---|
| Repeatability (n = 6) | Intermediate precision | Concentration level (μg mL−1) | Recovery (R%) | |||||
| a RSD: acceptance value ≤ 10%.b Recovery: acceptance range 95–105%.c Aripiprazole lyophilized powder.d Aripiprazole suspension. | ||||||||
| Phosphate | 25–75 | Y = 74.49X2 + 55996X − 533951 | 0.9988 | 2.8c/3.0d | 4.1c/2.8d | 25 | 100.3 | 0.61 |
| 50 | 97.5 | |||||||
| 75 | 98.3 | |||||||
| Sodium | 50–150 | Y = 569.3X2 + 204653X − 2.21 × 10−6 | 0.9999 | 2.2c/2.0d | 4.1c/2.9d | 50 | 97.6 | 0.13 |
| 100 | 97.3 | |||||||
| 150 | 95.8 | |||||||
The stability of the SST solution (containing 50 μg mL−1 of phosphate and 100 μg mL−1 of sodium) as well as the product solutions were evaluated over a 7-day period. Multiple replicate tests were performed within 48 hours and after 7 days from the solution preparation. During this time, all solutions were stored at 25 °C and protected from light. The % difference between stability intervals for each analyte in the sample solution was calculated using the following formula:
The experimental robustness testing is a required step of proper method validation as it is a measure of method capacity to remain unaffected by small but deliberate variations of its parameters and provides an indication of its reliability during normal usage.17 In our case, the robustness of the method was assessed by analyzing sample solutions. Small changes in flow rate (±0.1 mL min−1), column temperature (±2 °C), % buffer content in the mobile phase (±2%), buffer concentration (±10%) and pH value (±0.2) (Table S3, SI). The results showed that the % difference of the % content of both analytes in the tested sample between the modified and unmodified conditions was less than 4.8% which is within the acceptance criterion of 5%. Among the tested parameters, column temperature and buffer concentration exhibited the greatest influence while the buffer pH is sensitive for both ions. In contrast, flow rate had moderate effect, indicating good method robustness. These results confirm that the method is reliable for routine quality control.