DOI:
10.1039/D5RA06396E
(Paper)
RSC Adv., 2025,
15, 38612-38623
Portable smartphone-integrated ratiometric fluorescence probe for visual detection of mercury ions in environmental water with greenness evaluation
Received
27th August 2025
, Accepted 7th October 2025
First published on 15th October 2025
Abstract
Mercury is an extremely hazardous environmental pollutant with strong bioaccumulation potential, underscoring the need for rapid and sensitive detection of mercury ions (Hg2+) to protect both the environment and public health. In this work, a ratiometric fluorescence sensor was developed for the selective detection of Hg2+. The use of ratiometric fluorescence probes minimizes background interference, thereby enhancing accuracy compared to single-emission intensity-based nanoprobes. The sensor was fabricated by combining red carbon dots (r-CDs) with green carbon dots (g-CDs). The developed ratiometric probe exhibited high sensitivity, achieving a remarkably limit of detection (LOD) of 60 nM. Furthermore, the sensor's capability for visual detection was validated through a smartphone-based solution assay, with a LOD of 26.68 μM. To assess its environmental sustainability, three greenness assessment metrics AGREE, BAGI, and RAPI were employed, all confirming the method's outstanding eco-friendliness. Overall, the synthesized probe demonstrates strong potential for practical applications in the detection of Hg2+ in real samples.
1 Introduction
Heavy metals have played a crucial role in the advancement of human civilization through their widespread industrial applications. However, their excessive use and the discharge of metal-contaminated wastewater have raised serious concerns regarding both human health and environmental safety.1,2 In particular, toxic heavy metal ions have attracted significant attention due to their harmful impact on ecosystems and the well-being of living organisms.3 The detection and monitoring of heavy metals in living cells and the environment are therefore essential to mitigate their adverse effects.
Mercury ions (Hg2+) are recognized as a highly toxic contaminant commonly found in soil, water, and food.4,5 Owing to their tendency to bioaccumulate, they can build up in living organisms and bind to thiol groups in proteins, leading to serious harm to the central nervous system and presenting a major risk to human health and the environment.6–11 Therefore, creating highly sensitive sensing approaches for the identification and detection of Hg2+ in environmental systems is essential. A range of methods, including ICP-MS,12 electrochemistry,13 SERS14,15 and AAS,16,17 have been proposed for Hg2+ detection. However, these traditional methods often suffer from drawbacks such as high cost, complex instrumentation, time-consuming sample preparation, and limited applicability in remote areas.18,19 In recent years, fluorescent sensors have emerged as promising candidates for heavy metal ion detection due to their high sensitivity, rapid response, affordability, and ease of signal detection.20,21 Therefore, fluorescence remains particularly significant for the rapid determination of Hg2+ in real samples.
To date, many fluorescence-based probes for the detection of Hg2+ have been developed, including polymers,22,23 DNA enzymes,24 metal nanoclusters,25 sulfur dots,26,27 covalent organic frameworks,28 metal organic frameworks,29–31 and semiconductor quantum dots.32 Most fluorescence-based sensors, however, depend solely on a single emission intensity as the responsive signal, which can be significantly affected by factors such as instrumental deviations, solvent effects, and more. In contrast to single-emission fluorescence methods, ratiometric fluorescence techniques can minimize these interferences and achieve higher analytical accuracy through the self-calibration of two fluorescence intensities. Typically, ratiometric fluorescence methods are accompanied by noticeable color changes, enabling rapid visual identification.33–36
Carbon dots (CDs) offer numerous benefits, such as low toxicity, cost-effectiveness, strong FL emission, excellent biocompatibility, chemical stability, and ease of synthesis.37–40 They have been widely employed as sensing probes for detecting a variety of targets, including metal ions,41,42 anions,43 and biological molecules.44,45 With the advancement of information technology, smartphones have become widely available. Recent research has explored their application as portable detectors for both colorimetric and fluorescence-based analyses. This strategy not only minimizes analysis time and expenses but also supports on-site and real-time testing.46,47
Here, as depicted in Scheme 1, a novel probe has been developed using green and red carbon dots (CDs) for the precise detection of Hg2+. With increasing Hg2+ concentrations, the green fluorescence of g-CDs at 520 nm was gradually quenched, while the red fluorescence of CDs at 600 nm remained unchanged. Additionally, the presence of Hg2+ could be visually detected through a distinct color change from green to red under UV light. This probe enables rapid, accurate, and efficient detection of mercury ions without the need for expensive or complex equipment, making it well-suited for environmental monitoring applications.
 |
| | Scheme 1 Illustration of the synthesis of CDs and the visual detection of mercury ions. | |
2 Experimental part
2.1 Synthesis of green CDs
The g-CDs were synthesized. Briefly, 0.2 g of 2,3-dihydroxyquinoxaline was dissolved in a mixture of ammonia, DMF, and deionized water in specific proportions (1.5
:
1.5
:
20 mL, respectively). The resulting solution was then transferred to a 50 mL autoclave and heated at 200 °C for 12 hours. After the heating process, the autoclave was allowed to cool naturally. The obtained solution was passed through a 0.22 μm microporous membrane to eliminate larger particles, resulting in a transparent filtrate. This filtrate was further purified via dialysis. After purification, the solution was stored at 4 °C for subsequent testing.
2.2 Synthesis of red CDs
The r-CDs were prepared following this procedure.48 0.5 g of cellulose was dissolved in 40 mL of 2.5 M H3PO4 and heated at 180 °C for 1.5 hours. After that, the solution was purified.
2.3 Instruments
The optical properties of both CDs were characterized using a Cary Eclipse Fluorescence Spectrophotometer and a Cary 60 UV-Vis Spectrophotometer. Functional groups were analyzed by FTIR with a diamond ATR attachment (Nicolet iS50, Thermo Scientific). Morphology and structure were examined using high-resolution TEM (FEI Tecnai G2 F30), and surface composition was investigated via X-ray photoelectron spectroscopy (XPS, Thermo Fisher ESCALAB 250Xi).
2.4 The detection of Hg2+ with the probe
The ratiometric fluorescent probe was synthesized by combining g-CDs and r-CDs in a 2
:
1 FL intensity ratio in 2 mL of solution at pH 4. Hg2+ ions at varying concentrations were then introduced into the ratiometric fluorescent probe solution, and the resulting fluorescence spectra were recorded.
2.5 Visual detection
The solution containing a fluorescent probe (g-CDs and r-CDs in a 2
:
1 ratio) underwent the addition of different concentrations of Hg2+. Using a smartphone-based colorimetric device with a 365 nm UV lamp, we captured images of the fluorescence colors. The RGB values corresponding to the FL color information were then acquired utilizing the Color Grab Android application.49
2.6 Selectivity
To evaluate the specificity of the probe for Hg2+, 1 mM of the cations Co2+, Zn2+, Cr3+, Ni2+, Mg2+, Mn2+, Fe2+, Cd2+, and Cu2+, along with 1 mM of molecules such as glucose, ascorbic acid, phenylalanine, arginine, glycine, histidine, tartaric acid, and cysteine, were added to the system as potential interferences.
2.7 Real application
The fluorescence method was applied to detect Hg2+ in real samples, including laboratory tap water and commercially available bottled water from a local supermarket.
3 Result and characterization
3.1 Characterization of CDs
The surface functional groups of g-CDs were further analyzed using XPS. As shown in Fig. 1A, the full-range XPS spectrum reveals three distinct peaks at 284.79, 398.81, and 532.49 eV, corresponding to the characteristic signals of C1s, N1s, and O1s, respectively. This suggests that the surface of g-CDs is primarily composed of nitrogen (N), carbon (C), and oxygen (O). Fig. 1B presents the high-resolution XPS spectrum of C1s, which shows three peaks associated with C
O, C–C, C–OH, and C–N functional groups. Similarly, the high-resolution XPS spectrum of N1s, shown in Fig. 1C, displays three absorption peaks corresponding to the characteristic signals of C–N–C and C–N groups.50 Additionally, as illustrated in Fig. 1D, the O1s spectrum features two peaks, which can be assigned to C–OH/C–O–C groups and C
O.51
 |
| | Fig. 1 (A) XPS survey scan of g-CDs; (B)–(D) high-resolution XPS scans of the elemental composition of g-CDs. | |
The FTIR spectrum of the g-CDs is presented in Fig. 2A. A broad absorption band at 3385 and 2923 cm−1 is attributed to O–H stretching and methylene or methyl (C–H) groups, confirming the presence of aliphatic hydrocarbons. The peak at approximately 1605 cm−1 corresponds to C
C stretching vibrations, while the bands at 1406 and 1116 cm−1 are assigned to C–N and C–O functional groups, respectively52–56 In addition, the spectrum of r-CDs is shown in Fig. 2B. The broad and intense band in the range of 3600–2500 cm−1 is attributed to O–H groups. The absorption peak at 2359 cm−1 corresponds to P–H groups. The broad peak at 1900–1635 cm−1 corresponds to C
O and C
C groups. Moreover, the band at 1139 cm−1 is related to C–O stretching vibrations.48,57–59
 |
| | Fig. 2 FTIR spectra of (A) g-CDs, and (B) r-CDs. | |
The shape, size, and morphology of the green and red CDs (g-/r-CDs) were assessed by HRTEM. Fig. 3A and B show the TEM images and particle size distributions of both green and red CDs, which exhibit an approximately spherical shape. As illustrated in the HRTEM images, both g-CDs and r-CDs demonstrated excellent dispersion and uniform particle size. The average particle sizes were approximately 1.1 nm for g-CDs and 1.2 nm for r-CDs.
 |
| | Fig. 3 HRTEM images, and histogram distribution plot (A and C) g-CDs and (B and D) r-CDs. | |
The XRD patterns of g-CDs and r-CDs exhibit broad peaks, with the g-CDs pattern indicating the presence of highly disordered carbon atoms in the synthesized material (Fig. 4A). The g-CDs show broad 2θ peaks at approximately 27.6° and 42°, corresponding to disordered carbon atoms and confirming the predominance of the C(002) and C(100) planes associated with the hexagonal graphite structure of the g-CD particles.60 The broad XRD pattern suggests that the crystalline particles are small in size. Similarly, the XRD pattern of r-CDs (Fig. 4B) features a broad peak around 22.75°, indicating the presence of graphitic carbon (001) crystal planes within the r-CDs.48
 |
| | Fig. 4 The XRD spectra of the prepared (A) g-CDs and (B) r-CDs. | |
3.2 Optical properties
The UV-Vis spectra of green and red CDs were examined. The UV-Vis spectrum of g-CDs, shown in Fig. 5A, displays a broad absorption band ranging from 275 to 360 nm. Notably, absorption peaks are observed at 276, 315, and 325 nm, corresponding to the π–π transition and the n–π transition, which are associated with aromatic sp2 hybridization and C
O double bonds, respectively.60 In contrast, the UV-Vis spectrum of r-CDs reveals an absorption peak at 275 nm and two additional peaks within the 330–500 nm range. The first peak is attributed to π–π* transitions, while the absorption between 330 and 500 nm is linked to n–π* transitions, as shown in Fig. 5B.48 Peaks observed beyond 300 nm are attributed to the presence of carbonyl-based groups on the surface.
 |
| | Fig. 5 UV-Visible spectra of (A) g-CDs and (B) r-CDs. | |
The dual-emission g/r-CDs fluorescence probe was easily created by combining green and reed CDs. Typically, dual-emission fluorescence probes exhibit two distinct emission peaks when excited by a single wavelength.61 As shown in Fig. 6A, the fluorescence emission intensity of g-CDs and r-CDs was optimal around the 390–400 nm excitation range. Based on Fig. 6A, 400 nm was chosen as the optimal excitation wavelength for the dual-emission g/r-CDs throughout the experiments.
 |
| | Fig. 6 (A) Emission spectra of the ratiometric probe; (B) examination of IS; (C) effect of temperature; and (D) stability of the probe. | |
3.3 Effect of ionic strength
The effect of ionic strength on the fluorescence intensity of the ratiometric system was investigated. As shown in Fig. 6B, the addition of NaCl had no impact on the fluorescence intensity of the system.
3.4 Effect of temperature
As depicted in Fig. 6C, the ratio of I520 to I600 remained relatively stable within the temperature range of 35–75 °C. Consequently, 35 °C was chosen as the optimal temperature for this study to facilitate ease of operation.
3.5 Stability test
The stability of the g/r-CDs was evaluated by monitoring their fluorescence intensity over a three-month storage period. As shown in Fig. 6D, no significant changes in fluorescence intensity were observed, indicating that the CDs maintained their structural integrity and optical properties with minimal variation, even after extended storage. This demonstrates the excellent long-term stability of the synthesized g/r-CDs, making them suitable for practical applications in environmental sensing.
3.6 Fluorescence detection
Fluorescence responses were recorded after the addition of varying concentrations of Hg2+ to assess the sensitivity of the ratiometric probe (Fig. 7A and B). As the concentration of Hg2+ increased, the fluorescence intensity of the green-emitting QDs gradually decreased, while the red-emitting CDs showed only a slight decline and served as reference signals. This variation in emission intensity between the two peaks resulted in a noticeable fluorescence color transition from green to orange and eventually red, enabling visual detection of Hg2+ with the naked eye. As shown in Fig. 7A, the fluorescence intensity ratio (F520/F600) exhibited a strong correlation with Hg2+ concentration. A linear relationship was established in the range of 2–16 μM, described by the equation F520/F600 = 2.1489 − 0.0376 × [Hg2+], with a high coefficient of determination (R2 = 0.9976). The LOD was determined to be as low as 0.06 μM (Fig. 7B).
 |
| | Fig. 7 (A) FL spectra of green and red CDs with mercury ions (g-CDs and r-CDs in a 2 : 1 ratio). (B) Plot of F520/F600 versus mercury ion concentrations (μM). | |
As listed in Table 1, when compared with previous works on Hg2+ detection, this probe showed comparable performance in terms of fluorescence.
Table 1 Comparison of fluorescent probes used to detect Hg2+
| Materials used |
Sample |
Linear range |
LOD |
Sensor type |
Ref. |
| N-GQDs |
Water |
0–4.31 μM |
23 nM |
Turn-off |
62 |
| R6GH |
Seafoods |
0–5 μM |
2.5 × 10−2 μM |
Turn-on |
63 |
| BA-Eu-MOF |
Water |
1–60 μM |
220 nM |
Turn-on |
64 |
| SCDs |
Water |
0.05 to 5.8 μmol L−1 |
33.3 nM |
Turn-off |
65 |
| NCQDs |
Water |
0–5.0 μM |
0.017 μM |
Turn-off |
66 |
| g/r-CDs |
Water |
0–16 μM |
60 nM |
Turn-off |
This work |
3.7 Selectivity investigation
Selectivity is a key factor in assessing the detection capability of the ratiometric probe. To evaluate its specificity, the sensor's response was tested in the presence of potentially interfering ions. These interfering substances, including various metal ions and molecules, were introduced at concentrations of 100 μM ten times higher than that of Hg2+. As illustrated in Fig. 8, their presence caused only minimal changes in fluorescence, indicating no significant interference with Hg2+ detection. These findings confirm that the dual-emission ratiometric probe exhibits excellent selectivity for Hg2+. Furthermore, the sensor's performance remained unaffected even when Hg2+ coexisted with other ions.
 |
| | Fig. 8 The selectivity study of the probe. | |
3.8 Sensing mechanism
The FL intensity of g-CDs at 520 nm was significantly quenched in the presence of Hg2+. This observation indicates that the synthesized g-CDs can function as a novel “on–off” fluorescent sensor for the sensitive detection of Hg2+. To investigate the sensing mechanism, the UV-Vis spectra of the sensing platform were analyzed. As depicted in Fig. 9A, no noticeable changes were observed when comparing the UV-Vis spectra of g-CDs alone and g-CDs with Hg2+. Additionally, no new peaks appeared in the sensing system, ruling out the possibility of static quenching (SQ). SQ is a common fluorescence quenching mechanism that alters the absorption spectrum through the formation of a ground-state complex. These results suggest that the quenching effect is likely caused by efficient electron transfer between g-CDs and Hg2+,67 To further elucidate the quenching mechanism of Hg2+ on the fluorescence of g-CDs, it is proposed that Hg2+ suppresses fluorescence through electron transfer between the surface ligands of the g-CDs and the Hg2+ ions.55 The remarkable selectivity of g-CDs toward mercury can be attributed to the preferential chelation of Hg2+ with oxygen and/or nitrogen atoms on the g-CD surface, a property not commonly observed with other metal ions. Experimental evidence from FTIR and XPS analyses indicates that hydroxyl and amine functional groups on the g-CDs readily coordinate with Hg2+, thereby providing high specificity. This strong interaction leads to the formation of a stable Hg-CD complex, which in turn results in fluorescence quenching,68 as illustrated in Fig. 9B.
 |
| | Fig. 9 (A) The UV-Vis absorption spectra of g-CDs and g-CDs + Hg2+. (B) A schematic diagram illustrating the formation of blocked emissions in g-CDs. | |
3.9 Real application
The analytical performance of the ratiometric fluorescence probe for Hg2+ detection was evaluated using tap and bottled water samples, following the pretreatment procedures reported in the literature.69 As shown in Table 2, the recoveries of Hg2+ in real samples ranged from 96.84% to 105.72%, indicating that the probe is suitable for trace-level analysis of Hg2+ in environmental water.
Table 2 Determination of mercury in real samples (n = 3)
| Sample |
Spiked (μM) |
Found (μM) |
Recovery % |
RSD (n = 3) % |
| Tap water |
25 |
24.22 |
96.84 |
1.12 |
| 30 |
30.09 |
100.27 |
2.03 |
| Bottled water |
25 |
26.44 |
105.72 |
2.45 |
| 30 |
30.31 |
101 |
1.37 |
3.10 Visual based detection
As the mercury concentration increased from 150 to 330 μM, a dramatic change in the visual color of the green and red CDs was observed, transitioning from green to orange and eventually to red under ultraviolet light (Fig. 10A). To extract the sample's color information (RGB values), the fluorescent image was separated into three channels using ImageJ. The results revealed that the fluorescence intensity in the red channel increased, while that in the blue channel decreased, with each channel displaying distinct linear ranges and sensitivities. By utilizing the ratio of red to blue channel intensities (R/B) as the signal, a linear regression curve was established to correlate Hg2+ concentration with the R/B values. In the concentration range of 150–330 μM, a strong correlation (R2) was observed between the R/B values and Hg2+ concentration, with a LOD of 26.68 μM and a LOQ of 88.93 μM (Fig. 10B).
 |
| | Fig. 10 (A) Color changes observed in the solutions. (B) Calibration curve for the visual assay. | |
3.11 Greenness profile
Environmental considerations are playing an increasingly important role in the development of analytical methods. To evaluate the eco-friendliness of both the sample preparation and the overall analytical process, three distinct assessment tools were applied.60,70,71 Specifically, two green evaluation tools AGREE and BAGI were used to assess the environmental impact of the sample preparation step as well as the complete analytical procedure.
The AGREE tool was used to assess the environmental sustainability of the two developed methods. As shown in Fig. 11A and B, both approaches demonstrated environmentally friendly attributes, with the visual based method achieving a higher score (0.78) compared to the fluorimetry method (0.73). This enhanced performance can be attributed to factors such as reduced sample volume, the ability to analyze multiple samples simultaneously, system miniaturization, and lower energy consumption. These findings highlight the safety, eco-friendliness, and environmental sustainability of the proposed techniques.
 |
| | Fig. 11 Evaluation of the environmental sustainability of the proposed fluorimetry (A) and smartphone-based (B) methods using the AGREE metric. Evaluation of the environmental sustainability of the proposed fluorimetry (C) and smartphone-based (D) methods using the BAGI metric. | |
Furthermore, the BAGI tool was employed to assess the environmental impact of the methods involving fluorescent materials. As shown in Fig. 11C and D, the pictograms indicate BAGI scores of 75 and 70 for the visual-based and fluorimetry methods, respectively again underscoring the greener profile of the visual-based approach. Both methods also demonstrate strong potential for mercury ion detection.
Another novel tool introduced by Nowak et al., the Red Analytical Performance Index (RAPI), was used to evaluate the proposed method.72 This tool is applied here for the first time in a CD-based sensing approach and considers 10 parameters to evaluate analytical performance. Based on these 10 criteria, the fluorimetry method achieved a score of 80 (Fig. 12A), while the smartphone-based approach scored 72.5 (Fig. 12B). The lower score of the smartphone-based approach is primarily due to its working range, which is not as strong as that of the fluorimetry method.
 |
| | Fig. 12 RAPI analysis: (A) fluorimetry, (B) smartphone-based approach. | |
4 Conclusions
In conclusion, the dual-wavelength fluorescent probe features an inherent self-calibration capability, which significantly enhances both the sensitivity and accuracy of detection. This strategy capitalizes on the advantages of ratiometric fluorescence techniques, enabling semi-quantitative visual detection of Hg2+ and accurate quantification through dual-channel fluorescence signals. The sensor demonstrated excellent sensitivity toward Hg2+, achieving low detection limits. Moreover, it was successfully applied to the rapid analysis of Hg2+ in real samples, yielding high accuracy and satisfactory recovery rates. These findings suggest that the proposed fluorescent probe has strong potential for use in environmental monitoring and analytical sensing applications.
Conflicts of interest
There are no conflicts to declare.
Data availability
The authors confirm that the data supporting the findings of this study are available within the article.
References
- T. Rasheed, F. Nabeel, M. Adeel, M. Bilal and H. M. N. Iqbal, “Turn-on” fluorescent sensor-based probing of toxic Hg(II) and Cu(II) with potential intracellular monitoring, Biocatal. Agric. Biotechnol., 2019, 17, 696–701 CrossRef.
- P. Das, A. Ghosh, H. Bhatt and A. Das, A highly selective and dual responsive test paper sensor of Hg 2+/Cr 3+ for naked eye detection in neutral water, RSC Adv., 2012, 2(9), 3714–3721 RSC.
- T. Rasheed, M. Bilal, F. Nabeel, H. M. N. Iqbal, C. Li and Y. Zhou, Fluorescent sensor based models for the detection of environmentally-related toxic heavy metals, Sci. Total Environ., 2018, 615, 476–485 CrossRef CAS PubMed.
- K. F. Kayani and S. J. Mohammed, Mercury in aquatic environments: toxicity and advances in remediation using covalent organic frameworks, Mater. Adv., 2025, 6, 3371–3385, 10.1039/D5MA00208G.
- H. Park, E.-T. Oh, J. Park, S. Subedi, H. J. Park and K.-H. Lee, Real-Time Detection of Methylmercury and Hg (II) Using a Reversible Ratiometric Fluorescent Probe in Cellular and Aqueous Environments, Anal. Chem., 2025, 97(11), 5982–5991 CrossRef CAS PubMed.
- L. Chen, L. Yang, H. Li, Y. Gao, D. Deng and Y. Wu, et al., Tridentate lysine-based fluorescent sensor for Hg(II) in aqueous solution, Inorg. Chem., 2011, 50(20), 10028–10032 CrossRef CAS PubMed.
- H. Guo, X. Wang, N. Wu, M. Xu, M. Wang and L. Zhang, et al., In-situ synthesis of carbon dots-embedded europium metal-organic frameworks for ratiometric fluorescence detection of Hg2+ in aqueous environment, Anal. Chim. Acta, 2021, 1141, 13–20, DOI:10.1016/j.aca.2020.10.028.
- J. Li, Y. Wu, F. Song, G. Wei, Y. Cheng and C. Zhu, A highly selective and sensitive polymer-based OFF-ON fluorescent sensor for Hg 2+ detection incorporating salen and perylenyl moieties, J. Mater. Chem., 2012, 22(2), 478–482 RSC.
- Y. Liu, X. Su, H. Liu, G. Zhu, G. Ge and Y. Wang, et al., Construction of eco-friendly dual carbon dots ratiometric fluorescence probe for highly selective and efficient sensing mercury ion, J. Environ. Sci., 2025, 148, 1–12, DOI:10.1016/j.jes.2024.01.013.
- J. Wang, J. Fu, M. Chen and J. Zhang, Red-emitting carbon dots for fluorescent and smartphone-assisted dual-mode detection of Cu(II), Hg(II), and Fe(III) and investigation of the sensing mechanism, Mater Today Chem, 2024, 41, 102331 CrossRef CAS . Available from: https://www.sciencedirect.com/science/article/pii/S2468519424004373.
- J. Zeng, L. Liao, X. Lin, G. Liu, X. Luo and M. Luo, et al., Red-emissive sulfur-doped carbon dots for selective and sensitive detection of mercury (II) ion and glutathione, Int. J. Mol. Sci., 2022, 23(16), 9213 CrossRef CAS PubMed.
- X. Jia, Y. Han, C. Wei, T. Duan and H. Chen, Speciation of mercury in liquid cosmetic samples by ionic liquid based dispersive liquid-liquid microextraction combined with high-performance liquid chromatography-inductively coupled plasma mass spectrometry, J. Anal. At. Spectrom., 2011, 26(7), 1380–1386 RSC.
- X. Niu, Y. Ding, C. Chen, H. Zhao and M. Lan, A novel electrochemical biosensor for Hg2+ determination based on Hg2+-induced DNA hybridization, Sensor. Actuator. B Chem., 2011, 158(1), 383–387, DOI:10.1016/j.snb.2011.06.040.
- J. Duan, M. Yang, Y. Lai, J. Yuan and J. Zhan, A colorimetric and surface-enhanced Raman scattering dual-signal sensor for Hg 2+ based on Bismuthiol II-capped gold nanoparticles, Anal. Chim. Acta, 2012, 723, 88–93, DOI:10.1016/j.aca.2012.02.031.
- W. Tang, C. D. Bruce, D. L. Sparks and J. F. Rabolt, Selective and quantitative detection of trace amounts of mercury(II) Ion (Hg2+) and Copper(II) Ion (Cu2+) using surface-enhanced raman scattering (SERS), Appl. Spectrosc., 2015, 69(7), 843–849 CrossRef CAS PubMed.
- J. Sardans, F. Montes and J. Peñuelas, Electrothermal atomic absorption spectrometry to determine as, Cd, Cr, Cu, Hg, and Pb in soils and sediments: A review and perspectives, Soil Sediment Contam, 2011, 20(4), 447–491 CrossRef CAS.
- M. Shamsipur, A. Shokrollahi, H. Sharghi and M. M. Eskandari, Solid phase extraction and determination of sub-ppb levels of hazardous Hg 2+ ions, J. Hazard. Mater., 2005, 117(2–3), 129–133 CrossRef CAS PubMed.
- K. F. Kayani and K. M. Omer, A red luminescent europium metal organic framework (Eu-MOF) integrated with a paper strip using smartphone visual detection for determination of folic acid in pharmaceutical formulations, New J. Chem., 2022, 46(17), 8152–8161, 10.1039/D2NJ00601D.
- A. E. Kandjani, Y. M. Sabri, M. Mohammad-Taheri, V. Bansal and S. K. Bhargava, Detect, remove and reuse: A new paradigm in sensing and removal of Hg (II) from wastewater via SERS-active ZnO/Ag nanoarrays, Environ. Sci. Technol., 2015, 49(3), 1578–1584 CrossRef PubMed.
- S. Chowdhury, B. Rooj, A. Dutta and U. Mandal, Review on Recent Advances in Metal Ions Sensing Using Different Fluorescent Probes, J. Fluoresc., 2018, 28(4), 999–1021 CrossRef CAS PubMed.
- J. Wang, X. Wang, X. Pan, W. Pan, Y. Li and X. Liang, et al., Dual-emission carbon dots achieved by luminescence center modulation within one-pot synthesis for a fluorescent ratiometric probe of pH, Hg2+, and glutathione, Microchim. Acta, 2020, 187(6), 1–11 Search PubMed.
- H. Wang, P. Zhang, J. Chen, Y. Li, M. Yu and Y. Long, et al., Polymer nanoparticle-based ratiometric fluorescent probe for imaging Hg2+ ions in living cells, Sensors Actuators, B Chem., 2017, 242, 818–824 CrossRef CAS.
- N. Choudhury, B. Saha, B. Ruidas and P. De, Dual-Action Polymeric Probe: Turn-On Sensing and Removal of Hg2+Chemosensor for HSO4-, ACS Appl. Polym. Mater., 2019, 1(3), 461–471 CrossRef CAS.
- S. Wang, X. Li, J. Xie, B. Jiang, R. Yuan and Y. Xiang, A metal ion-triggered and DNA-fueled molecular machine for amplified and sensitive fluorescent detection of Hg2+, Sensor. Actuator. B Chem., 2018, 259, 730–735, DOI:10.1016/j.snb.2017.12.081.
- L. Yu, L. Zhang, G. Ren, S. Li, B. Zhu and F. Chai, et al., Multicolorful fluorescent-nanoprobe composed of Au nanocluster and carbon dots for colorimetric and fluorescent sensing Hg2+ and Cr6+, Sensor. Actuator. B Chem., 2018, 262, 678–686, DOI:10.1016/j.snb.2018.01.192.
- K. F. Kayani, O. B. A. Shatery, S. J. Mohammed, S. B. Aziz, N. N. Mohammad and G. H. Abdullah, et al., Fluorescent sulfur quantum dots for environmental monitoring, Nanotechnol. Rev., 2025, 14(1), 20240138 CrossRef CAS.
- K. F. Kayani, S. J. Mohammed, N. N. Mohammad, A. M. Abdullah, D. I. Tofiq and M. S. Mustafa, et al., Sulfur quantum dots for fluorescence sensing in biological and pharmaceutical samples: a review, Mater. Adv., 2024, 5, 6351–6367, 10.1039/D4MA00502C.
- K. F. Kayani, Tetracycline in the environment: Toxicity, uses, occurrence, detection, and removal by covalent organic frameworks − Recent advances and challenges, Sep. Purif. Technol., 2025, 6, 132418, DOI:10.1016/j.seppur.2025.132418.
- K. F. Kayani, O. B. A. Shatery, S. J. Mohammed, H. R. Ahmed, R. F. Hamarawf and M. S. Mustafa, Synthesis and applications of luminescent metal organic frameworks (MOFs) for sensing dipicolinic acid in biological and water samples: a review, Nanoscale Adv., 2025, 7(1), 13–41, 10.1039/D4NA00652F.
- K. F. Kayani, Nanozyme based on bimetallic metal–organic frameworks and their applications: A review, Microchem. J., 2025, 208, 112363 CrossRef CAS . Available from: https://www.sciencedirect.com/science/article/pii/S0026265X24024767.
- K. F. Kayani, Bimetallic metal–organic frameworks (BMOFs) for dye removal: a review, RSC Adv., 2024, 14(43), 31777–31796, 10.1039/D4RA06626J.
- Z. Q. Zhou, R. Yan, J. Zhao, L. Y. Yang, J. L. Chen and Y. J. Hu, et al., Highly selective and sensitive detection of Hg2+ based on fluorescence enhancement of Mn-doped ZnSe QDs by Hg2+-Mn2+ replacement, Sensor. Actuator. B Chem., 2018, 254, 8–15, DOI:10.1016/j.snb.2017.07.033.
- K. Yi, X. Zhang and L. Zhang, Smartphone-based ratiometric fluorescent definable system for phosphate by merged metal−organic frameworks, Sci. Total Environ., 2021, 772, 144952, DOI:10.1016/j.scitotenv.2021.144952.
- T. Wang, Q. Mei, Z. Tao, H. Wu, M. Zhao and S. Wang, et al., A smartphone-integrated ratiometric fluorescence sensing platform for visual and quantitative point-of-care testing of tetracycline, Biosens. Bioelectron., 2020, 148, 111791, DOI:10.1016/j.bios.2019.111791.
- X. Chen, Q. Mei, L. Yu, H. Ge, J. Yue and K. Zhang, Rapid and On-site Detection of Uranyl Ions via Ratiometric Fluorescence Signals Based on a Smartphone Platform, ACS Appl. Mater. Interfaces, 2018, 10(49), 42225–42232 CrossRef CAS PubMed.
- W. Li, X. Zhang, X. Hu, Y. Shi, Z. Li and X. Huang, et al., A smartphone-integrated ratiometric fluorescence sensor for visual detection of cadmium ions, J. Hazard. Mater., 2021, 408, 124872, DOI:10.1016/j.jhazmat.2020.124872.
- X. Sun and Y. Lei, Fluorescent carbon dots and their sensing applications, TrAC–Trends Anal. Chem., 2017, 89, 163–180, DOI:10.1016/j.trac.2017.02.001.
- K. F. Kayani, Phosphorus-doped carbon dots and their analytical and bioanalytical applications: a review, Talanta, 2026, 297, 128768 CrossRef CAS PubMed . Available from: https://www.sciencedirect.com/science/article/pii/S0039914025012597.
- K. F. Kayani, Metal-doped carbon dots for sensing applications in food analysis: A critical review, Microchem. J., 2025, 215, 114296 CrossRef CAS . Available from: https://www.sciencedirect.com/science/article/pii/S0026265X25016509.
- K. F. Kayani, Carbon Dots: Synthesis, Characterization, and Applications in the Detection of Bilirubin – Recent Advances and Challenges, J. Fluoresc., 2025 DOI:10.1007/s10895-025-04366-z.
- M. Zheng, Z. Xie, D. Qu, D. Li, P. Du and X. Jing, et al., On-off-on fluorescent carbon dot nanosensor for recognition of chromium(VI) and ascorbic acid based on the inner filter effect, ACS Appl. Mater. Interfaces, 2013, 5(24), 13242–13247 CrossRef CAS PubMed.
- J. Du, Y. Yang, T. Shao, S. Qi, P. Zhang and S. Zhuo, et al., Yellow emission carbon dots for highly selective and sensitive OFF-ON sensing of ferric and pyrophosphate ions in living cells, J. Colloid Interface Sci., 2021, 587, 376–384, DOI:10.1016/j.jcis.2020.11.108.
- Z. Liu, C. Zhu, G. Chen, Z. Cai, A. Hu and T. Yang, et al., Construction of Cyan Blue Fluorescence Probe based on Nitrogen-Doped Carbon Dots for Detecting Nitrite Ion in Ham Sausage, J. Fluoresc., 2024, 1–10 CrossRef PubMed.
- C. Zhao, Y. Jiao, F. Hu and Y. Yang, Green synthesis of carbon dots from pork and application as nanosensors for uric acid detection, Spectrochim. Acta, Part A, 2018, 190, 360–367, DOI:10.1016/j.saa.2017.09.037.
- X. Luo, W. Zhang, Y. Han, X. Chen, L. Zhu and W. Tang, et al., N,S co-doped carbon dots based fluorescent “on-off-on” sensor for determination of ascorbic acid in common fruits, Food Chem., 2018, 258, 214–221 CrossRef CAS PubMed . Available from: https://www.sciencedirect.com/science/article/pii/S0308814618304606.
- L. Hou, Y. Qin, J. Li, S. Qin, Y. Huang and T. Lin, et al., A ratiometric multicolor fluorescence biosensor for visual detection of alkaline phosphatase activity via a smartphone, Biosens. Bioelectron., 2019, 143, 111605, DOI:10.1016/j.bios.2019.111605.
- B. Lin, Y. Yu, Y. Cao, M. Guo, D. Zhu and J. Dai, et al., Point-of-care testing for streptomycin based on aptamer recognizing and digital image colorimetry by smartphone, Biosens. Bioelectron., 2018, 100(378), 482–489, DOI:10.1016/j.bios.2017.09.028.
- K. F. Kayani, S. J. Mohammed, N. N. Mohammad, G. H. Abdullah, D. A. Kader and N. S. Hamad Mustafa, Ratiometric fluorescence detection of tetracycline in milk and tap water with smartphone assistance for visual pH sensing using innovative dual-emissive phosphorus-doped carbon dots, Food Control., 2024, 164, 110611 CrossRef CAS.
- M. Fan, Z. Pan, C. Wang, Y. Guo, J. Sun and M. Liu, et al., Quantitative Visual Detection of Mercury Ions With Ratiometric Fluorescent Test Paper Sensor, Front Chem, 2022, 10, 859379 CrossRef CAS PubMed.
- H. Liu, H. Li, K. Du and H. Xu, Yellow fluorescent carbon dots sensitive detection of Hg2+ and its detection mechanism, Mater. Today Commun., 2022, 33, 104880 CrossRef CAS . Available from: https://www.sciencedirect.com/science/article/pii/S2352492822017214.
- H. Wang, P. Sun, S. Cong, J. Wu, L. Gao and Y. Wang, et al., Nitrogen-doped carbon dots for “green” quantum dot solar cells, Nanoscale Res. Lett., 2016, 11, 1–6 CrossRef CAS PubMed.
- V. Manikandan and N. Y. Lee, Green synthesis of carbon quantum dots and their environmental applications, Environ. Res., 2022, 212, 113283 CrossRef CAS PubMed . Available from: https://www.sciencedirect.com/science/article/pii/S0013935122006107.
- M. Li, P. Zhang, J. Mao, J. Li, Y. Zhang and B. Xu, et al., Construction of cellulose-based hybrid hydrogel beads containing carbon dots and their high performance in the adsorption and detection of mercury ions in water, J. Environ. Manage., 2024, 359, 121076, DOI:10.1016/j.jenvman.2024.121076.
- S. Xie, D. Yang, Q. Li, J. Li, D. Zhong and Y. Yang, et al., Effects of functional groups on carbon dots-based AuNPs for the specific detection of mercury ions, Carbon, 2025, 243, 120481 CrossRef CAS . Available from: https://www.sciencedirect.com/science/article/pii/S000862232500497X.
- X. Guo, L. Li, L. Wang, M. Yang, Y. Wang and M. Zhang, et al., A paper-based ratiometric sensor for mercury (II) detection using fluorescent carbon dots, Talanta, 2025, 293, 128036 CrossRef CAS PubMed . Available from: https://www.sciencedirect.com/science/article/pii/S0039914025005260.
- H. Xu, X. Yang, G. Li, C. Zhao and X. Liao, Green Synthesis of Fluorescent Carbon Dots for Selective Detection of Tartrazine in Food Samples, J. Agric. Food Chem., 2015, 63(30), 6707–6714 CrossRef CAS PubMed.
- G. Dong, K. Lang, H. Ouyang, W. Zhang, L. Bai and S. Chen, et al., Facile synthesis of N, P-doped carbon dots from maize starch via a solvothermal approach for the highly sensitive detection of Fe 3+, RSC Adv., 2020, 10(55), 33483–33489 RSC.
- S. Mohandoss, S. Ganesan, S. Palanisamy, S. You, K. Velsankar and S. Sudhahar, et al., Nitrogen, sulfur, and phosphorus Co-doped carbon dots-based ratiometric chemosensor for highly selective sequential detection of Al3+ and Fe3+ ions in logic gate, cell imaging, and real sample analysis, Chemosphere, 2023, 313, 137444 CrossRef CAS PubMed.
- R. Gao, X. Yi, X. Liu, H. Wang, L. Wang and B. Zeng, et al., Phosphorus-doped carbon dots as an effective flame retardant for transparent PVA composite films with enhanced UV shielding property, React. Funct. Polym., 2024, 197, 105877 CrossRef CAS . Available from: https://www.sciencedirect.com/science/article/pii/S138151482400052X.
- K. F. Kayani, O. B. A. Shatery, M. S. Mustafa, A. H. Alshatteri, S. J. Mohammed and S. B. Aziz, Environmentally sustainable synthesis of whey-based carbon dots for ferric ion detection in human serum and water samples: evaluating the greenness of the method, RSC Adv., 2024, 14(8), 5012–5021 RSC.
- X.-X. Chen, L.-Y. Niu, N. Shao and Q.-Z. Yang, BODIPY-based fluorescent probe for dual-channel detection of nitric oxide and glutathione: visualization of cross-talk in living cells, Anal. Chem., 2019, 91(7), 4301–4306 CrossRef CAS PubMed.
- F. Du, L. Sun, Q. Zen, W. Tan, Z. Cheng and G. Ruan, et al., A highly sensitive and selective “on-off-on” fluorescent sensor based on nitrogen doped graphene quantum dots for the detection of Hg2+ and paraquat, Sensor. Actuator. B Chem., 2019, 288, 96–103, DOI:10.1016/j.snb.2019.02.109.
- Z. Zhang, R. Han, S. Chen, F. Zheng, X. Ma and M. Pan, et al., Fluorescent and Colorimetric Dual-Mode Strategy Based on Rhodamine 6G Hydrazide for Qualitative and Quantitative Detection of Hg2+ in Seafoods, Foods., 2023, 12(5), 1085 CrossRef CAS PubMed.
- H. Wang, X. Wang, M. Liang, G. Chen, R. M. Kong and L. Xia, et al., A Boric Acid-Functionalized Lanthanide Metal-Organic Framework as a Fluorescence “turn-on” Probe for Selective Monitoring of Hg2+ and CH3Hg+, Anal. Chem., 2020, 92(4), 3366–3372 CrossRef CAS PubMed.
- S. Wang, H. Chen, H. Xie, L. Wei, L. Xu and L. Zhang, et al., A novel thioctic acid-carbon dots fluorescence sensor for the detection of Hg2+ and thiophanate methyl via S-Hg affinity, Food Chem., 2021, 346, 128923 CrossRef CAS PubMed.
- J. Zhu, H. Chu, T. Wang, C. Wang and Y. Wei, Fluorescent probe based nitrogen doped carbon quantum dots with solid-state fluorescence for the detection of Hg2+ and Fe3+ in aqueous solution, Microchem. J., 2020, 158, 105142, DOI:10.1016/j.microc.2020.105142.
- Y. Liu, P. Zhou, Y. Wu, X. Su, H. Liu and G. Zhu, et al., Fast and efficient “on-off-on” fluorescent sensor from N-doped carbon dots for detection of mercury and iodine ions in environmental water, Sci. Total Environ., 2022, 827, 154357 CrossRef CAS PubMed . Available from: https://www.sciencedirect.com/science/article/pii/S0048969722014504.
- K. F. Kayani and C. N. Abdullah, A Dual-Mode Detection Sensor Based on Nitrogen-Doped Carbon Dots for Visual Detection of Fe(III) and Ascorbic Acid via a Smartphone, J. Fluoresc., 2025, 35(2), 1125–1137, DOI:10.1007/s10895-024-03604-0.
- Z. Li, S. Pang, M. Wang, M. Wu, P. Li and J. Bai, et al., Dual-emission carbon dots-copper nanoclusters ratiometric photoluminescent nano-composites for highly sensitive and selective detection of Hg2+, Ceram. Int., 2021, 47(13), 18238–18245, DOI:10.1016/j.ceramint.2021.03.143.
- M. S. Mustafa, N. N. Mohammad, F. H. Radha, K. F. Kayani, H. O. Ghareeb and S. J. Mohammed, Eco-friendly spectrophotometric methods for concurrent analysis of phenol, 2-aminophenol, and 4-aminophenol in ternary mixtures and water samples: assessment of environmental sustainability, RSC Adv., 2024, 14(23), 16045–16055 RSC.
- K. F. Kayani, S. J. Mohammed, N. N. Mohammad, M. K. Rahim, M. S. Mustafa and H. R. Ahmed, et al., Exploring Green Practices: a Review of Carbon Dot-Based Sustainable Sensing Approaches, J. Fluoresc., 2025 DOI:10.1007/s10895-025-04254-6.
- P. M. Nowak, W. Wojnowski, N. Manousi, V. Samanidou and J. Płotka-Wasylka, Red analytical performance index (RAPI) and software: the missing tool for assessing methods in terms of analytical performance, Green Chem., 2025, 27(19), 5546–5553 RSC.
|
| This journal is © The Royal Society of Chemistry 2025 |