A. Tommasiniab,
G. Suarato*b,
S. Fiocchib,
E. Chiaramellob,
A. Marrellab,
M. Lenzunib,
M. Parazzinib,
B. Cortesecde and
P. Ravazzanib
aDipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Piazza Leonardo da Vinci 32, Milano, 20133, Italy
bCnr-Istituto di Elettronica e di Ingegneria Dell’Informazione e Delle Telecomunicazioni, Piazza Leonardo da Vinci 32, Milano, 20133, Italy. E-mail: giulia.suarato@cnr.it
cInterventional Coronary Center, University Hospitals Harrington Heart & Vascular Institute, Cleveland, Ohio, USA
dFondazione Ricerca e Innovazione Cardiovascolare, Via E. Ponti, 49, Milan, 20136, Italy
eDCB Academy, Via E. Ponti, 49, Milan, 20136, Italy
First published on 22nd August 2025
Leveraging their ability to induce an intense and highly localized electric polarization upon application of an external magnetic field, core–shell magnetoelectric nanoparticles (MENPs) constitute powerful nanotools to achieve wireless nano-electroporation and targeted drug delivery in critical applications such as cardiovascular medicine, where frequent interventions may pose general yet high risks to the patients. In this study, the magnetoelectric feature is exploited to thoroughly map in silico the behavior of a model MENP under a set of either static or time-varying external magnetic fields, with the dual aim of (i) providing a series of operational parameters to elicit at the nanoscale a reversible cell membrane poration and (ii) correlating the electric potential difference developed at the activated MENP surface with the charge displacement potentially induced in a hypothetical drug-MENP bond. The finite element analysis framework considers a single cobalt ferrite core–barium titanate shell MENP, either immersed in culture medium or engulfed in a blood vessel wall. Simulation results obtained under static magnetic field conditions show that maximizing magnetoelectric efficiency (79.82 mV Oe−1 cm−1) generates electric fields in line with reversible nano-electroporation values (3.7 × 104 to 4.7 × 104 V m−1). Furthermore, a correlation between the profiles of the electric potential registered at the MENP outer border (2.86–2.92 mV) and the charge displacement hypothetically experienced by a drug-MENP bond is highlighted, when an alternating, low-intensity external magnetic field is used. Our computational study sets a preliminary base to gain a deeper understanding of the interface phenomena of MENP-mediated cell poration and drug delivery, and it serves as a preliminary step towards complex future analyses.
To tackle this medical issue, in the late 1970s and mid-1980s Dotter and Judkins performed the first coronary angioplasty in a patient with a bare metal stent.5 Later, bare metal devices were replaced by drug-eluting stents, approved by the Food and Drug Administration in 2003,6 which improved physiological revascularization and lowered the mortality rate. Despite optimization in stent design, the use of thinner struts and lightweight materials, or the research on more biocompatible coatings7 to further sustain the therapeutics release, persistent inflammation around the implanted stent and delayed endothelial repair remain critical in ISR.8 Later, a new class of devices has been developed, namely drug-coated balloons, which allow drug delivery without prosthesis implantation.9,10 This novel medical apparatus was designed to initially tackle ISR and later showed promising results also for the treatment of native coronary artery disease.11
To this end, nanotechnology can offer a variety of curative approaches, especially when it comes to engineering platforms for targeted drug delivery. For example, micro- and nanoparticles can find applications in a plethora of different medical fields (i.e., tissue engineering, wound healing, and anticancer therapy, to name a few) due to their variety in terms of sizes, organic and inorganic components,12,13 morphologies, and fabrication and functionalization techniques.14,15 By tuning the surface chemistry of these nanosystems, site-specific and cellular-targeted delivery can be achieved16 and the drug loading content optimized,17 while challenges such as poor pharmacokinetics, drug instability under physiological conditions, or enzymatic degradation can be limited by properly combining cargo molecules with nanoparticle carriers.18 Leveraging their therapeutic potential, micro- and nano-polymeric particles or lipid-based micelles have been recently exploited for cardiovascular applications.7,19–21 To achieve a more precise spatial directionality towards a stenotic vessel region, magnetic nanoparticles, properly decorated with polymeric coatings and active drugs, have been locally and systemically administered to mice and subjected to an external static magnetic field. For example, Fellows et al.22 delivered heparin-coated magnetic nanoparticles for drug therapy of neointimal hyperplasia, while Polyak and co-workers23 injected endothelial cells previously loaded with superparamagnetic nanoparticles, to allow for their blood navigation and targeted positioning near a stented area to inhibit ISR.
In addition to these more conventional systems, a novel class of multi-functional nanostructures is represented by magnetoelectric materials in which magnetic and electric fields can be strongly coupled at body temperature.24 More specifically, magnetoelectric nanoparticles (MENPs) have recently emerged as a promising frontier in biomedical research due to their unique ability to act as nanoconverters.25 In fact, these multi-phase structures are based on the combination of a magnetostrictive core and a piezoelectric shell, with different sizes and shapes, whose interface coupling can be optimized26 to achieve desired features and performance.27 Thanks to their ability to polarize and generate an intense electric field at their surface (order of magnitude of 105 V m−1) and in their immediate nearby, MENPs have been investigated for (i) wireless brain and nervous tissue stimulation,28–31 (ii) cell manipulation,32 (iii) cancer treatment, through hyperthermia or targeting of cancer cells,33,34 (iv) tissue engineering,35 and (v) nano-electroporation and drug delivery.36–38
Classical electroporation methods use traditional electrodes to generate high-intensity electric fields for the permeabilization of cell membranes. However, such electrodes are often invasive and difficult to apply in in vivo settings.39,40 Despite its efficacy, the application of a high intensity, non-localized electric field can lead to difficult and non-specific control of the process and even to loss of function or cell death (irreversible nano-electroporation). Greater spatial control of the generated electric field and high versatility in nano-electrode design can be achieved by using MENPs, which could act, when activated with a direct current magnetic field (HDC), as electric field nano-generators within safe ranges (2 × 104 to 5 × 104 V m−1),41 able to provide reversible and wireless cell membrane dielectric breakdown.42 Similarly, conventional approaches to delivering drugs must contend with poor specificity and loss of drug concentration, as well as difficulties in crossing biological barriers. MENPs can be easily and highly functionalized, as reported in extensive in vitro studies by Guduru et al.,43 where nanoparticles were loaded with paclitaxel as a mitotic inhibitor for ovarian cancer treatment and labelled with a biomarker-specific antibody, and by Nair and coworkers,44 who loaded MENPs with an anti-HIV compound. MENPs can achieve drug release following remote activation via alternating current, low-intensity magnetic fields (HAC),44 thus providing a field-controlled and highly specific delivery to the diseased tissue. Experimentally, the MENP-mediated drug release has been achieved with the application of an external HAC of small magnitude (approximately below 50 Oe, with a frequency ranging from 10 to 100 Hz36,37), which was sufficient to liberate a significant drug concentration into the cells.
By combining their ability to electroporate biological membranes with nanometric precision (when subjected to a HDC field) alongside the possibility of remotely modulating their cargo molecule unloading (in the presence of a HAC field), MENPs appear as a fascinating alternative to target pathological issues also in a yet unexplored field, such as interventional cardiology. In fact, the risk of restenosis and the need for further interventions could be tremendously reduced by employing MENPs, properly decorated with anti-inflammatory or anti-proliferative drugs.
With the aim of assessing the feasibility and functionality of such advanced nano-systems for cardiovascular drug release applications, the present study uses an in silico computational approach. A preliminary assessment of the MENP baseline behavior when subjected to an external magnetic field above its core magnetization saturation (Ms) was carried out to validate the soundness of our computational model and confirm the presence of the magnetoelectric effect (Fig. 1, step 1). Next, a series of simulations were carried out considering a single nanoparticle immersed in two different biological surroundings, such as the culture medium and the blood vessel wall tissue, exposed to magnetic fields of varying waveforms and intensities. A first parametric-sweep study under a static magnetic field (HDC) (Fig. 1, step 2) was envisioned to define the operational ranges of MENP stimulation able to subsequently induce reversible membrane permeability. Afterwards, the external magnetic field assumes an alternating pattern (HAC) (Fig. 1, step 3), which causes an oscillation of the generated potential difference across the nanoparticle, a phenomenon that has been hypothesized to be related to drug bond weakening and, consequently, drug release. The workflow reported in Fig. 1 summarizes the computational steps, highlighting the operational parameters considered in the analysis.
![]() | ||
Fig. 2 Computational model of the core–shell magnetoelectric nanoparticle, immersed in the surrounding material, with sizes of the geometry. |
Property | Symbol | Value | Reference |
---|---|---|---|
Density (kg m−3) | ρ | 5.2 × 103 | 45, COMSOL library |
Relative permittivity | εr | 10 | COMSOL library |
Electrical conductivity (S m−1) | σ | 5.2 × 106 | COMSOL library |
Poisson's ratio | ν | 0.48 | 46, COMSOL library |
Young's modulus (Pa) | E | 230 × 109 | 46 and 47 |
Saturation magnetization (A m−1) | Ms | 3.69 × 105 | 48 |
Initial magnetic susceptibility | χ0 | 3 | 49, COMSOL library |
Saturation magnetostriction (ppm) | λ | −200 | 42 and 46 |
Two different scenarios were modeled, based on the environment in which the nanoparticle was embedded: (i) single MENP floating in the culture medium in a Petri dish; (ii) single MENP enclosed in the blood vessel wall. In both simulations, the surrounding environments were considered as purely resistive materials, with a relative permittivity (εr) of 1; the culture medium was modelled as a liquid environment, while the blood vessel was considered a solid material (properties reported in Table 3).
At this step of the investigation, particular attention was paid to the electric field generated by the particle within its surroundings. To do so, three representative values of the external magnetic field (Hext) were considered for the triggering of the magnetoelectric phenomenon and the following data analysis: (1) Hext(1) identified in the range of linear dependence between αME and the magnetic field; (2) Hext(2) corresponding to the maximum αME with the aim of taking advantage of the optimal MENP performance; (3) Hext(3) corresponding to the minimum magnetic field yielding the highest electric potential. In particular, Hext(3) was identified as the value of the magnetic field at which the increment between two consecutive potential differences was less than 0.15 mV. A MATLAB code was implemented to examine the near surroundings of the nanoparticle starting from the electric field data distribution exported from COMSOL software. The regions of interest were modeled as concentric, circular semi-coronas around the nanoparticle, identifying areas within three different, increasing distances from the MENP surface. More specifically, the radii of the semi-coronas considered for the analysis were 75, 80 and 90 nm to sample electric field values located at distances of 5, 10 and 20 nm, respectively, from the shell external border (rMENP = 70 nm). For each configuration, the electric field distribution was visualized, and its statistical metrics (in terms of median value and the 1st and the 99th percentiles) were extracted to allow for data comparison. More specifically, three regions were considered, depending on how likely the generated electric fields could cause Irreversible Nano-electroporation (Ir-Nep), Reversible Nano-electroporation (Rev-Nep) or none of the two. Such a tendency was assessed by checking the 1st and 99th percentiles of each distribution. The analysis was conducted for the two case scenarios presented in paragraph 2.1 (i.e., culture medium and blood vessel wall). The outputs were processed via OriginPro 2023 (OriginLab, United States) and presented as box plots, highlighting median values and percentiles. As summarized in Fig. 1, the aim of this intermediate step was to identify the optimal magnetic field (HthresholdRev-Nep) upon activation of which an electric field comparable to those used in conventional reversible electroporation (EthresholdRev-Nep) could be developed in the immediate vicinity of the MENP. Such Hfield and Efield and their spatial distributions will be essential for the last step of the investigation.
To properly model the MENP behavior within a time-dependent study, the Jiles–Atherton (J–A) hysteresis model was implemented in the Solid Mechanics section in the software interface model builder, for the magnetostrictive core. The COMSOL entries in the core material settings for the J–A model parameters have been modified according to Table 4, considering our group's previous work.27 The mathematical equations governing the J–A model are summarized in the SI. In the Study section of the COMSOL interface, a Time Dependent Study was added, and a parametric sweep was imposed. The Parameter Value List section was filled in with specific values (such as 1, 2, 2.5, 3, 4.5, 5, and 6.5 mT). For the nano-electroporation phase, the magnetization profile was studied, paying particular attention to the value of magnetization obtained after ΔtRev-Nep (nano-electroporation time), i.e. when the constant static magnetic field was switched off. For the drug release phase, the core magnetization values, the electrical potential differences (ΔV) and the αME, observed/calculated at a given time, were plotted in relation to the increasing amplitude of the sinusoidal wave. Ultimately, the concept of charge displacement was considered, derived as a function of the alternating magnetic field amplitudes and the efficiency factor (αME). Such physical quantity, referred to as Qionic, gives an indication of the stability involved in a hypothetical drug-MENP ionic bond (see the SI and Stimphil et al.55). All the outputs were processed via OriginPro 2023 (OriginLab, United States).
The steps that a spherical magnetoelectric nanoparticle of our dimensions goes through, when ideally immersed in the blood vessel wall and exposed to a strong magnetic field of 2T (a field above its magnetization saturation, homogeneous and directed along the z-axis, considered the “easy axis” for the MENP magnetization) are shown in Fig. 4. Upon switching on the high intensity magnetic field, indicated by the black arrows pointing upwards, the CFO core becomes completely magnetized (Fig. 4 step 1, dark core, herein represented by the distribution of the magnetization, expressed in A m−1). Through deformation coupling of a magnetostrictive and piezoelectric material, magnetoelectricity can be exploited directly. Therefore, strain at the core level and stress at the shell level are reported (Fig. 4, step 2). The inhomogeneous stress distribution is the highest (but always within several tenths of ppm) along the sides of the nanoparticle. The magnetoelectric coupling causes a movement of charges, resulting in a maximum potential difference at a shell outer border of 6.07 mV (Fig. 4, step 3), and the consequent generation of an electric field in the MENP surrounding (Fig. 4, step 4), with values up to 7 × 104 V m−1, sharply decreasing downwards with the distance from the outer surface of the shell. A similar electric behavior, with highly comparable values, is observed when the MENP is immersed in the culture medium, and the resulting distributions are shown in Fig. S2 and Table S1 of the SI.
The MENP behaves as an electric dipole when exposed to an external magnetic field. To evaluate the magnetoelectric coefficient (αME), according to eqn (1), the electric potential difference generated on the surface of the MENP was considered (Fig. 5b). The coefficient αME rapidly increases in the first part of the curve (Hfield in the range of 2–50 mT), until it reaches a peak of about 79.82 mV Oe−1·cm−1, corresponding to a magnetic field of 300 mT. Beyond this value, the curve gradually decreases asymptotically, dropping down to 11.15 mV Oe−1·cm−1 at high-intensity Hfield values. Fig. 5b also reports on the evolution of the electric potential (ΔV, calculated as the difference between the maximum and minimum values of the two opposite poles generated on the surface of the MENP, and expressed in mV), which shows a similar profile to that previously described for the magnetization: an initial linear trend at low-intensity Hfield (reaching a value of 0.11 mV at 40 mT), followed by the attainment of a plateau (around 6.22 mV). Furthermore, the trend of this parameter in the two different environments under study appears to perfectly overlap.
The left side of Fig. 6 shows the electric field distributions at distances of 5, 10 and 20 nm from the outer border of the MENP shell (rMENP = 70 nm), as extrapolated from COMSOL software and processed with MATLAB (see paragraph 2.2 in the Materials and methods section). On the right side of Fig. 6, the statistical metrics of the electric field distributions (in terms of median values and the 1st and the 99th percentiles) are reported, for each of the three progressively larger areas considered. Furthermore, the statistical metrics of the electric field distributions generated by the external magnetic fields under study are presented in Table 5, to ease the data comparison. While the data presented in Fig. 6 and Table 5 refer to the case of a MENP engulfed in the blood vessel wall, a similar analysis was also carried out for the case of the culture medium as the surrounding environment (Fig. S3 and Table S1). According to the data, stimulation with a magnetic field of 150 mT appears to be below the threshold that triggers a biological membrane response (above 2 × 104 V m−1), while at 650 mT, Efields distributions falling in the range of irreversible nano-electroporation are obtained (above 5 × 104 V m−1). Therefore, maintaining conservative conditions, the following analyses have been carried out considering an external HDC of 300 mT, whose generated Efields distributions completely remain in the range of reversible nano-electroporation (2 × 104 to 5 × 104 V m−1).
Hext(1) = 150 mT | Hext(3) = 300 mT | Hext(3) = 650 mT | |||||||
---|---|---|---|---|---|---|---|---|---|
(V m−1) | 5 nm | 10 nm | 20 nm | 5 nm | 10 nm | 20 nm | 5 nm | 10 nm | 20 nm |
Median | 1.50 × 104 | 1.45 × 104 | 1.43 × 104 | 3.90 × 104 | 3.77 × 104 | 3.70 × 104 | 6.01 × 104 | 5.81 × 104 | 5.71 × 104 |
99th percentile | 1.80 × 104 | 1.80 × 104 | 1.79 × 104 | 4.67 × 104 | 4.66 × 104 | 4.66 × 104 | 7.21 × 104 | 7.19 × 104 | 7.19 × 104 |
1st percentile | 1.21 × 104 | 1.02 × 104 | 0.76 × 104 | 3.14 × 104 | 2.63 × 104 | 1.97 × 104 | 4.85 × 104 | 4.06 × 104 | 3.04 × 104 |
The analysis comprises two phases: (i) induction of reversible nano-electroporation on the cell membrane of a hypothetical cell in the vicinity of the MENP, triggered by a static magnetic field (HDC); (ii) promotion of the dissociation of drug molecules hypothetically bonded at the nanoparticle surface, by applying a HAC stimulation, thus leveraging the generation of an alternating electric potential at the MENP outer shell. To achieve the reversible nano-electroporation, a HDC of 300 mT (as obtained from the analysis reported in Fig. 6 and Table 1) is initially applied for Δt = tRev-Nep (Fig. 7a, pink line) and then switched off. As this study is not intended to propose the definition of a precise protocol for nano-electroporation but is primarily interested in assessing the MENP-mediated process feasibility, specific time intervals are not considered. The magnetization profiles of the MENP core over time generated by such static magnetic field stimulation are reported in Fig. 7a for the case of both environments (blood vessel – red line, and culture medium – dashed blue line), with the aim of understanding how the nanoparticle core behaves during and after the removal of the stimulation. According to Fig. 7a, the core magnetization obtained at 300 mT is roughly 5.65 × 104 A m−1 and remains constant over ΔtRev-Nep. When the HDC is switched off, the core retains a residual magnetization of approximately 5.2 × 104 A m−1 (∼10 emu g−1) for both simulation environments, except for a negligible delta. Such residual magnetization is retained for t > tRev-Nep, thus leaving the particle in a “magnetized” state.
Next, the drug release step is considered to investigate how the alternating magnetic field may affect the MENP specific behavior. In Fig. 7c, sinusoidal waves with the different amplitude values considered in the study are graphed. The core magnetization and the electrical potential difference profiles are plotted in Fig. 7e and in Fig. 7f, respectively, for a single MENP engulfed in the blood vessel well. For the case of the CM scenario, the corresponding plots and values are reported in SI, Fig. S4 and Table S2. As noticeable from Fig. 7e and f and Table 6, higher values of magnetization and ΔV are obtained when the amplitude of the sinusoidal stimulation is greater (HAC ranging from 4.5 mT to 6 mT), with more pronounced oscillations, and values reaching 5.27 × 104 A m−1 and 2.92 mV, respectively. Similarly to the previous analysis carried out for the trend assessment of MENP performance under steady state conditions (paragraph 3.2), the values of core magnetization, electrical potential difference and magnetoelectric coefficient related to the increase of the alternating field amplitude at a given time point (arbitrarily chosen) are reported in Table 6.
HAC (mT) | Magnetization (A m−1) | ΔV (mV) | αME (mV cm−1 Oe−1) |
---|---|---|---|
1 | 5.23 × 104 | 2.86 | 87.36 |
2 | 5.24 × 104 | 2.87 | 84.00 |
2.5 | 5.24 × 104 | 2.88 | 83.43 |
3 | 5.24 × 104 | 2.88 | 83.88 |
4.5 | 5.25 × 104 | 2.90 | 81.70 |
5 | 5.26 × 104 | 2.91 | 82.07 |
6.5 | 5.27 × 104 | 2.92 | 79.85 |
Fig. 8 reports simulation data comparison involving the concept of average charge displacement (referred to as Qionic; the detailed explanation is available in the SI) to which hypothetical drug-MENP ionic bonds may be subjected when the system is immersed in a homogenous, low-intensity HAC field at 100 Hz. The Qionic profiles at different HAC amplitudes over the studied period reveal a trend comparable to that of the core magnetization and potential difference (Fig. 8a): an increase in the stimulation sine wave amplitude is associated with more visible oscillations. The susceptibility to rupture of hypothetical drug-MENP bonds is related to the alternating magnetic field values, which, in turn, cause oscillation of the electric potential difference. Therefore, the trend of Qionic versus ΔV at a given time is also shown (Fig. 8b), highlighting a linear relationship, with charge shift values ranging from 1.29 to 1.32 × 10−17 coulomb (C), which are in line with Nair et al.44 Lastly, Fig. 8c compares the evolution over time of the electric potential difference and the average charge displacement of the hypothetical ionic drug-MENP bonds from their equilibrium positions. As a representative example, the HAC = 6.5 mT stimulation case was considered for the BV scenario: as expected, Qionic and ΔV oscillate in a reciprocal manner. For the CM case, equivalent results are shown in the SI (Fig. S5).
The in silico work starts with a validation of the ME effect of a MENP engulfed in the blood vessel wall or immersed in the culture medium, as modeling environments, to observe the generation of an electric potential on the MENP surface and an electric field in the surroundings, in response to a 2T static magnetic field (above the core magnetization saturation). The MENP baseline behavior is analyzed in Fig. 4: the magnetization involves a realignment of the magnetic domains within the core, a compression along the direction of the applied magnetic field (i.e., the z-axis), and an expansion in the other two dimensions, leading to a core dimensional change (lattice deformation) on the order of a hundred ppm. This deformation is imparted to the piezoelectric shell, transforming the stress exerted by the core into a change in the charge position within the atoms of the piezoelectric lattice, thus generating a surface potential and an electric field distribution on the MENP and in its vicinity, which are consistent with those found in previous studies conducted by our group,30 where MENPs have been proposed for peripheral nerve stimulation29,60 and sensing,61 and nerve regeneration,62 highlighting the versatility of the adopted computational framework. After having confirmed the basic response of the MENP, a wider range of DC magnetic field values is assessed, and the CFO-BTO nanoparticle parameters are thoroughly characterized. The results of the calculated magnetization recorded at the core of the MENP (Fig. 5a) describe a curve whose trend agrees with the one of Ma et al.63 In fact, when the applied magnetic field gradually increases, the domains become progressively aligned with its direction and the magnetization values rise. At a certain point, the magnetization profile reaches a plateau, because of the complete orientation of the magnetic domains in the direction of the applied field, and any further increase in the external stimulus does not translate into a stronger magnetization. The trend of the magnetoelectric coefficient as a function of the applied field (Fig. 5b) is in accordance with the work of Kumari et al.28 Considering that MENPs of different sizes and material properties are modelled, our maximum value of αME is about 0.1 V Oe−1·cm−1, registered at 300 mT, while they reported a peak of 0.4 V Oe−1·cm−1 corresponding to an external HDC of 600 mT. It is therefore possible to conclude that the MENP performance does not necessarily improve as the potential difference increases, but it rather diminishes at the expense of a progressively higher magnetic field.
By tuning the magnetoelectric coupling, a wide range of diagnostic and therapeutic applications based on MENPs can be explored. In this context, ME properties can be exploited to wirelessly and selectively manipulate cells and even reversibly permeabilize their membrane to accomplish targeted drug delivery. Efficient permeabilization of the membrane is conventionally achieved by following protocols based on the use of electric fields (i.e. electroporation), which generally rely on bulk electrodes placed in the proximity of cells and/or tissues. Plate electrodes for non-invasive stimulation or needle electrodes inserted into the tissue for invasive stimulation are two types of traditional devices.39,40 Invasive stimulation is usually the most effective strategy for electroporation, but it suffers from drawbacks common to all invasive procedures (complications can include bleeding, inflammatory responses and infections or systemic issues). These latter hinder the use of such a technique for broad-spectrum treatments or when the localization of the diseased area might be difficult to reach (e.g., cardiovascular applications). Recently, some studies have focused on the use of inorganic nanoparticles, such as gold nanoparticles or MENPs, as a completely contactless method to achieve electroporation with enhanced spatial specificity at the nanoscale.42,64 In particular, in the very recent work of Bryant et al.,65 MENPs have been explored as nanotheranostic agents for specific targeting of cancer cells and damaging via irreversible electroporation (IRE) by means of a high amplitude DC magnetic field (on the order of magnitude of 7T). From the simulations carried out in our study (Fig. 6 and Table 5), the electric field values considered suitable for reversible nano-electroporation (i.e., ranging from 2 × 104 to 5 × 104 V m−1, according to Lv et al.41 which we have taken as reference despite referring to bulky electrode-mediated pulsed stimulation) are reached when a HDC of 300 mT is used to excite the MENP. This corresponds to the optimal performance scenario where the maximum efficiency (αME) is observed. In contrast, when the induced electric field exceeds 5 × 104 V m−1 (corresponding to a HDC of 650 mT), cell death caused by irreversible nano-electroporation has been ascribed to several reasons,66 such as permanent membrane lysis, loss of cell homeostasis, flush in/out of molecules, and even cytotoxic effects as a consequence of electrolysis and by-products generated around the electrodes.66
Having verified the occurrence of reversible nano-electroporation in the presence of an external 300 mT HDC, we proceeded with a time-dependent static simulation. The MENP, which exhibits its non-superparamagnetic properties at room temperature,24 presents a residual magnetic flux density48 when a HDC of 300 mT is applied for Δt = tRev-Nep and subsequently removed (Fig. 7a). To further investigate the potential of MENPs not only as wireless permeabilizing nanodevices of biological membranes but also as emerging tools for the controlled release of therapeutic agents into specific tissues, we focused our investigation on the alternating magnetic field simulation. Ideally, it has been previously postulated that the application of a HAC onto a magnetized nanoparticle decorated with drug molecules would create conditions for an efficient drug-MENP bond breaking process, thus enhancing the drug release efficacy.44 Such phenomenon is influenced by several factors, either related to the physico-chemical features of the system (i.e., organic coating on the MENPs, nature of the drug-MENP bond, and type and amount of drug loaded), or the conditions of magnetic stimulation (duration, amplitude and frequency of HAC and possible temperature increase during the stimulation).43,67 In this preliminary computational study, the main parameter considered was the variation of the HAC amplitudes (values in the range of the experimental published literature43,67), from which the core magnetization and the subsequent electric potential difference at the MENP surface are dependent. In line with the work of Kozielski et al.,68 where it is affirmed that applying a sinusoidal magnetic field to magnetoelectric materials produces a sinusoidal electric field with a frequency that matches the input, Fig. 7e and f display comparable trends for the core magnetization and the electric potential difference profiles. As a consequence of the ME effect, the electric dipole generated at the MENP outer border induces a charge surface density (defined as σMENP) of the order of magnitude of ±αH (following the theoretical dissertation of Stimphil et al.,55 eqn (18) of the SI). Assuming a hypothetical drug molecule attached at the MENP outer shell via an ionic bond, its release would occur if such σMENP equals and/or exceeds the charge density involved in the drug-MENP ionic bond. Moreover, fundamental to the dissolution of such a hypothetical ionic bond is the effective alternation of stretching and contracting of the ionic bond length, which ultimately results from the movement of positive and negative charges that constitute the ionic bond itself from their equilibrium positions. Considering that multiple drug molecules may be bound to a single MENP surface, an average positive/negative charge displacement can be reasonably defined (labelled as Qionic, according to the theoretical modelling of nanoparticle-bound drug release by Stimphil et al.,55 eqn (19)–(22) of the SI). Our simulations conducted under varying HAC conditions (Fig. 8b) highlight that such an average charge displacement in the hypothetical ionic bond is indeed governed by a linear dependence on the electrical potential difference (ΔV) developed at the MENP surface. ΔV oscillations over time would promote Qionic variations, thus perturbing the equilibrium positions of the positive and negative charges. We speculate that the length of the hypothetical drug-MENP bonds would be far from their equilibrium when such oscillations become more readily detectable (i.e., HAC equal to or greater of 4.5 mT, Fig. 8c). Higher alternating magnetic field values would result in larger oscillation amplitudes, thereby spanning over a wider range of electrical potential differences at the MENP outer shell, which, in turn, would be responsible for a stronger variation in the average bond charge displacement. In other terms, such stimulation conditions would cause a strong ‘shaking effect’ on the system, favoring the delivery of the therapeutic agent.43
Supplementary information include the mathematical expressions for (a) the COMSOL multiphysics model of the magnetoelectric phenomenon; (b) the COMSOL multiphysics model of the magnetic hysteresis phenomenon; (c) the theoretical model of the drug delivery phenomenon; (d) magnetization loop of the cobalt ferrite core; (e) static study results to validate the MENP baseline behavior in the culture medium environment; (f) distributions of the electric field produced by various external HDC fields within culture medium surroundings; (g) time dependent study for the drug release step in the CM environment; (h) MENP core magnetization, electric potential difference and magnetoelectric coefficient values in relation to the increase in amplitude of the applied HAC for the CM case scenario. See DOI: https://doi.org/10.1039/d5na00438a.