In this study, we analyzed the combination of affinity purification mass spectrometry (AP-MS) with high-field asymmetric waveform ion mobility spectrometry (FAIMS), integrated between nanoLC-MS and an Orbitrap Ascend tribrid mass spectrometer. Our primary objective was to evaluate the application of the FAIMS interface for detecting affinity purified SAP25 protein complexes with enhanced sensitivity and robustness. As a result, we observed that nanoLC-FAIMS-MS (with FAIMS) significantly improved the sensitivity and detection limits at the protein level, peptide level and significantly reduced chemical contaminants compared to nanoLC-MS alone without FAIMS (No FAIMS). This FAIMS configuration resulted in 42% and 92% increases for the total proteins and unique proteins, respectively, and 44% and 88% increases for total peptides and unique peptides compared to the No FAIMS configuration. Our in-depth comparison of FAIMS and No FAIMS shows that FAIMS outperforms by significantly reducing the missing value by <15% in datasets and plays a significant role in filtering chemical contaminants. Lastly, we searched the datasets for multiple post-translational modifications important in chromatin remodeling and found several arginine methylation sites on the bait protein SAP25. Our findings highlight the potential of FAIMS with Orbitrap Ascend tribrid mass spectrometer to enhance the depth of AP-MS analysis. The data were deposited with the MASSIVE repository with the identifier MSV000096548.
Sin3 associated protein, SAP25 is a member of the Sin3/HDAC complex, playing a vital role in transcriptional repression. Previous studies suggested that SAP25 acts as an adaptor protein and is associated with O-linked N-acetylglucosamine transferase (OGT),13 SCF E3 ubiquitin ligase complex SKP1/FBXO3/CUL1 and the ubiquitin carboxyl-terminal hydrolase 11 (USP11). These proteins play a vital role in transcription regulation, by changing post-translational modifications and DNA alterations. Notably, these enzymatic and regulatory complex proteins are uniquely associated with SAP25, among the other eleven Sin3 complex proteins.14 This makes SAP25 important to study by AP-MS. When studying protein complexes by AP-MS it is important to take into consideration the possible protein contaminants that can occur from using specific affinity purification methods.15 In addition it is also important to take into consideration the chemical contaminants that can interfere with peptide analysis by mass spectrometry.16
Nanoscale capillary liquid chromatography with mass spectrometry (nanoLC-MS) remains a cornerstone in mass spectrometry-based proteomics.17 Despite its central role, nanoLC-MS faces limitations, particularly in requiring larger sample volumes, is time-consuming, and becomes increasingly challenging when analyzing many samples.18 High-field asymmetric waveform ion mobility spectrometry (FAIMS) enhances proteomics by reducing background interference and minimizing contaminants, thus improving reliability and sensitivity.18,19 FAIMS works at atmospheric pressure and separates ions based on their various mobility in low and high electric fields.20 Ions travel between two parallel electrodes, where an asymmetric waveform is applied, causing displacement due to their differential mobility. This displacement is corrected by applying the compensation voltage, CV.21 Additionally, FAIMS offers gas-phase fractionation through multiple compensation voltages (CVs), which effectively deconvolutes overlapping peptide signals, providing significant advantages over traditional nanoLC-MS approaches.1 When combined with mass spectrometry, the FAIMS separates ions before they enter the vacuum chamber of the mass spectrometer, reducing chemical background interference and enhancing signal-to-noise ratios.22 Previous studies have highlighted the benefits of FAIMS in proteomics applications, including top-down studies and single-shot proteomics.1,18,23–27 However, there is limited comparative research on FAIMS for bottom-up proteomics with affinity-purified protein complexes.28
Technological developments in mass spectrometry (MS)-based proteomics continually enhances the detection capability and precision of proteome analyses.29 The LTQ Orbitrap, introduced by Thermo Fisher Scientific in 2005, marked a significant milestone. Over the past decade, subsequent advancements have focused on enhancing sensitivity and scan speed, addressing the evolving demands of proteomics.27 However, achieving comprehensive proteome coverage, sensitivity and speed are still major challenges.30,31 The Orbitrap Ascend, equipped with a linear ion trap, quadrupole, and Orbitrap, addresses these challenges by improving scan speed and sensitivity.31,32 Recent comparative studies by Shuken et al. and He et al. have demonstrated that the Orbitrap Ascend outperforms the Orbitrap Eclipse in both protein detection and quantification, further underscoring the potential of these technologies in proteomic research.30,31
In this study, we present the application of an advanced FAIMS device integrated between nanoLC-MS on the Orbitrap Ascend tribrid mass spectrometer for a bottom-up proteomics approach targeting affinity-purified protein complexes. First, we carried out SAP25 affinity purification from Flp-In™-293 cells stably expressing Halo-SAP25. The purified proteins were then subjected to digestion into peptides followed by mass spectrometry with FAIMS and No FAIMS configuration. (Fig. 1).33 In this study, we successfully characterized Halo-SAP25 purified proteins with both FAIMS and No FAIMS configurations. Our findings revealed that FAIMS enhances the detection of a number of proteins and peptides. Next, we tested FAIMS efficacy to filter the potential contaminants and reduce the missing values. Overall, our results demonstrate that the distinct effectiveness of FAIMS substantially improves the detection of affinity-purified protein complexes, especially the lower abundance proteins and peptides, while significantly reducing the level of contaminants and the percentage of missing values when used with the nanoLC on Orbitrap Ascend tribrid mass spectrometer.
000 × g at 4 °C for 30 minutes. The salt concentration of the resulting supernatants was reduced to 0.3 M NaCl by adding ice-cold buffer containing 10 mM HEPES (pH 7.5), 1.5 mM MgCl2, 10 mM KCl, 0.5 mM DTT, and 1X protease inhibitor. The lysates were centrifuged again at 40
000 × g at 4 °C for 30 minutes. The resulting supernatant was collected for Halo affinity purification.
000 × g for 30 minutes at 4 °C. The pellet was washed two times with acetone, air dried, and resuspended in 8 M urea dissolved in 100 mM Tris–HCl (pH 8.5). The formation of disulfide bonds was reduced and alkylated with tris(2-carboxyethyl)-phosphine hydrochloride (TCEP) and chloroacetamide (CAM) subsequently. Denatured proteins were initially treated with 0.1 μg of Lys-C for <6 hours at 37 °C. The urea concentration was lowered to 2 M by adding 100 mM Tris–HCl, pH 8.5 followed by 2 mM final concentration of CaCl2. At the end, proteins were digested with 0.5 μg of trypsin overnight at 37 °C with shaking. The reactions were terminated by the addition of formic acid to a final concentration of 5% before proceeding with mass spectrometry analysis.
All the peptides were washed and loaded with a flow rate of 350 μL min−1 with 2% mobile phase B (mobile phase A: 0.1% formic acid in water; mobile phase B: 80% acetonitrile, 0.1% formic acid). Peptide separation occurred over 100 minutes from 2–25% mobile phase B. This was followed by an increase to 40% B over the next 20 minutes. Before the next injection with 2% B, the column was cleaned and equalized for 15 min with 100% B.
The nano-LC was directly connected with the Thermo Orbitrap Ascend tribrid mass spectrometer. A silica emitter (20 μm i.d., 10 cm, CoAnn) integrated with a high field asymmetric ion mobility source (FAIMS). The data were acquired by data-dependent acquisition (DDA), and the intact peptides were detected in the Orbitrap with a resolving power of 120
000, covering the m/z range from 375 to 1500. Higher-energy collisional dissociation (HCD) at 28% normalized collision energy (NCE) was employed to fragment peptides with charge states ranging from +2 to +7 with dynamic exclusion set for 60 seconds after the initial detection of a precursor ion. Fragment ions were analyzed in the Orbitrap with resolving power of 30
000.
In terms of FAIMS, the compensation voltage (CV) were selected as −45 V (1.4 s), −60 V (1 s), and −75 V (0.6 s) for a cycle time of 3s with source ionization +1700 V, with 305 °C temperature of the transfer tube. The resulting raw data were searched in Proteome Discoverer 3.0 (PD; version 3.0; Thermo Fisher Scientific) against the human protein database of Uniprot 05-05-2023 with a common contaminants database. Protein and peptides abundances were exported to Microsoft Excel and graphs were prepared using GraphPad Prism.
The comparative analysis to identify the level of contamination between Halo-SAP25 affinity-purified protein complexes with FAIMS and No FAIMS conditions was performed by importing the raw files to the Molecular Contaminants MS1 Filtering v4.1 document with the document's default settings in Skyline-daily (64-bit) 24.1.1.284 (bc93c2813).34–36 Synchronized peak integration based on peak area was performed between replicates with wide windows of integration to account for large elution profiles and the lack of molecular identity validation since these precursors were not selected for fragmentation. Skyline34–36 results were exported and formatted with R Studio for Microsoft Excel.
To detect PTMs enriched by Halo-SAP25 purification, database searches were performed as previously described with the addition of the following dynamic modifications where a maximum of 3 modifications were permitted on a given peptide: phospho +79.966 Da (S, T, Y), acetyl +42.001 Da (K), methyl +14.016 Da (K, R), and GG +114.043 Da (K) for ubiquitination. The data was analyzed qualitatively using Proteome Discoverer 3.0 and Excel.
The Venn diagram (Fig. 2A) illustrates the overlap and uniqueness of proteins identified under both configurations. In the FAIMS-assisted analysis, a total of 385 proteins were identified, with 176 proteins uniquely detected under FAIMS configuration. In contrast, the No FAIMS configuration yielded the identification of 222 proteins, of which only 13 were unique to this setting. The usage of FAIMS resulted in a 13-fold increase in the number of unique proteins detected compared to the setup without FAIMS configured. A total of 209 proteins were commonly identified under both FAIMS and No FAIMS configurations. To further assess the reproducibility and distribution between the SAP25 runs, we performed a principal component analysis (PCA) on the proteomic data for both FAIMS and No FAIMS configurations. The PCA plot (Fig. 2B) illustrates the distribution of SAP25 runs analyzed with FAIMS and with No FAIMS configuration. The first two principal components (PC1 and PC2), accounting for 53.9% and 19.0% of the variance respectively, clearly separate the samples processed with FAIMS from those processed with No FAIMS. The distinct and tight clustering of FAIMS and No FAIMS samples suggests a differential separation between the two methods. Both the Venn diagram and PCA plot were prepared using Proteome Discoverer (PD; version 3.0; Thermo Fisher Scientific). Additionally, when comparing the total number and abundance of proteins, we observed that FAIMS detected twice as many proteins, with a particular increase in low-abundance proteins (Table S1, ESI†). Fig. 2C illustrates the distribution of the average number of proteins across six runs for both FAIMS and No FAIMS configurations, with a total of 340 proteins detected with FAIMS (depicted in blue) and 158 proteins detected with No FAIMS (depicted in red). The data demonstrate a statistically significant increase in protein detection and normalized MS peak intensity abundances when FAIMS is applied (Fig. 2D). The normalized protein abundances (log2) for individual Halo-SAP25 affinity-purified protein complex runs, as shown in Fig. 2E, indicates the reduction in normalized protein abundances (log2) with FAIMS configuration due to detection of lower abundance proteins (Table S1, ESI†). These results underscore the significant enhancement in proteomic depth afforded by the incorporation of FAIMS.
Additionally, we have also considered unique and razor peptides for quantification to check the efficiency of FAIMS. Proteins identified with at least one unique peptide were considered. Our result showed that the sum of unique peptides observed with FAIMS is higher than observed with No FAIMS, which is 949 and 745 respectively. However, in both configurations, the majority of proteins were identified by a single unique peptide while the razor peptide count was 69 for FAIMS and 51 for No FAIMS (Fig. S1 and Table S4, ESI†).
To further confirm the differences in missing values between the two groups, we calculated the percentage of missing values observed with FAIMS and No FAIMS configurations at the protein and peptide levels (Tables S1 and S2, ESI†). This was validated by analyzing the number of proteins identified and quantified, as well as those lacking abundance values in each configuration. In the FAIMS group, we found that on average, 85.43% of proteins were identified and quantified with only 14.57% of proteins showing missing values among the six runs of SAP25. In contrast, in the No FAIMS group, only 39.42% of proteins were identified and quantified while the remaining 60.58% of proteins had missing values (Fig. 4B). A similar comparison was carried out with peptides, demonstrating the same pattern. In the FAIMS group, 76.25% of peptides were identified and quantified, with only 23.75% showing missing values. Conversely, in the No FAIMS group, 43.01% of peptides were identified and quantified, while the remaining 56.99% had missing values (Fig. 4C). This observation further confirms the previous claims that FAIMS helps to minimize the number of missing values.6,40 Next, we sought to determine how FAIMS significantly enhances the detection of lower abundance proteins and peptides. To verify that, we selected several abundance ranges, increased 10-fold, and analyzed the number of proteins and peptides detected within each range. Our result suggests that FAIMS substantially improves the detection of lower abundance proteins (Fig. 4D) and peptides (Fig. 4E), particularly with ranges ≥1.0 × 103, ≥1.0 × 104, and ≥1.0 × 105, while at higher abundance ranges, there were minimal differences in numbers observed in FAIMS and No FAIMS configuration. Among six Halo-SAP25 affinity-purified protein complex runs on average 2, 60, and 178 proteins and 4, 110, and 352 peptides were detected with FAIMS for these abundance ranges, respectively. With No FAIMS the number was reduced to 1, 10 and 61 for proteins and 2, 22, and 140 for peptides, respectively. This observation is pointing that FAIMS notably increase the detection of lower abundance proteins and peptide levels resulting in a vast dynamic range of ion detection. Therefore, coupling FAIMS with conventional nanoLC holds promise for enhancing the sensitivity and depth of proteomic profiling.1,41,42
For example, Fig. 6 shows the chromatographic profiles of Triton X-100 without FAIMS (6A), Triton X-100 with FAIMS (6B), Tween 60 without FAIMS (6C), Tween 60 with FAIMS (6D), NP-40 without FAIMS (6E), and NP-40 with FAIMS (6F). Each of these chemical contaminants eluted late in the chromatography with Triton X-100 eluting starting at approximately 130 minutes (Fig. 6A and B), Tween60 eluting starting at approximately 116 minutes (Fig. 6C and D), and NP-40 eluting at approximately 114 minutes (Fig. 6E and F). For each chemical contaminant, the intensities were significantly higher in without FAIMS when compared to with FAIMS. The highest contaminant filtration was observed for Triton X-100, which is a reagent added to our high salt lysis buffer to lyse cells, where the level of filtration was significantly greater than 1000 times compared to the No FAIMS configuration (Fig. 6A and B). However, certain contaminants such as dipalmityldimethylammonium chloride, distearyldimethylammonium chloride, PEG, phthalate esters, SPDMA and Triton X-100 reduced showing minimal filtering by FAIMS (Fig. 7A and Tables S5, S6, ESI†).
Lastly, we decided to analyze the total grouped abundances of contaminants (log2) with FAIMS and No FAIMS configuration. We used the average of contaminant abundances across six SAP25 runs for each configuration (Fig. 7B and C). Our observation suggests that in the FAIMS configuration, a significant reduction in contaminant abundance is observed across various contaminants, many of them are having total grouped abundance less than 1 × 107 (Table S6, ESI†). On the other hand, contaminants such as dipalmityldimethylammonium chloride, PEG, SPDMA and Triton X-100 reduced having higher total grouped abundances between 1 × 108 to 5 × 108 similar to No FAIMS configuration (Fig. 7C), indicating the minimal effect of FAIMS to minimize these contaminants level. Overall, we determined that FAIMS played an important role in filtering various contaminants at significant levels, especially that of Triton X-100 (Fig. 7B and C).
The peptide level PTMs on key proteins in SAP25 affinity purifications are shown in Table 1. Shown are the peptides and their modifications for a methylated and acetylated peptide in ZNF556, a phosphorylation even on HDAC2, a phosphorylated, methylated and acetylated peptide on ZNF318, a phosphorylation on USP11, a phosphorylation on PARP1, and several arginine methylations found on SAP25. In Table 2 the protein level PTM information is shown where we queried the UniProt database to see if any of these PTMs have been previously reported in the literature. The PTMs on USP11, HDAC2, and PARP1 have been reported before, but the modifications seen on ZNF318 and ZNF556 have not been reported previously. In the case of SAP25, no PTMs are reported in UniProt with the multiple arginine methylations of found on this protein from this affinity purification potentially providing new mechanistic insights into the regulatory role of this protein. Further enrichment of these PTMs and their characterization may provide further functional information on the significance of these PTMs in future studies.
| Protein | Protein accession | Annotated peptide sequence | Modifications (% confidence) | Quality q-value | SVM score | Modification location in protein (% confidence) |
|---|---|---|---|---|---|---|
| Quality q-value, SVM score, and % confidence are all scoring features reported by Proteome Discoverer. | ||||||
| ZNF556 | Q9HAH1 | [K].HTGEKPVNAASVGKPSGGLCSSK.[N] | 1× carbamidomethyl [C20]; 1× methyl [K14(99.4)]; 1× acetyl [K5(100)] | 0.0036 | 0.408 | 1× methyl [K405(99.4)]; 1× acetyl [K396(100)] |
| HDAC2 | Q92769 | [R].IACDEEFSDSEDEGEGGRR.[N] | 1× carbamidomethyl [C3]; 1× phospho [S8(100)] | 0.0011 | 0.778 | 1× phospho [S422(100)] |
| SAP25 | Q8TEE9 | [R].GQRVPLYLSK.[A] | 1× methyl [R3(100)] | 0.0011 | 1.552 | 1× methyl [R94(100)] |
| SAP25 | Q8TEE9 | [R].GLRPVAPATGHWNGQQAPPDAGFPVVCCEDVFLSDPLLPR.[G] | 2× carbamidomethyl [C27; C28]; 1× methyl [R3(100)] | 0.0011 | 2.792 | 1× methyl [R54(100)] |
| SAP25 | Q8TEE9 | [R].TLCHPSFWPLYEAASGR.[G] | 1× carbamidomethyl [C3]; 1× methyl [R17(100)] | 0.0011 | 1.701 | 1× methyl [R51(100)] |
| SAP25 | Q8TEE9 | [K].AGPRPVWGANCSSGASFSGR.[T] | 1× carbamidomethyl [C11]; 1× methyl [R4(100)] | 0.0011 | 2.278 | 1× methyl [R18(100)] |
| ZNF318 | Q5VUA4 | [K].VSEKLEDQLSEGR.[N] | 1× methyl [R13(100)]; 1× phospho [S2(100)]; 1× acetyl [K4(100)] | 0.0011 | 0.771 | 1× methyl [R1241(100)]; 1× phospho [S1230(100)]; 1× acetyl [K1232(100)] |
| USP11 | P51784 | [R].RLLSPAGSSGAPASPACSSPPSSEFMDVN.[-] | 1× carbamidomethyl [C17]; 1× phospho [S18(99.5)] | 0.0011 | 1.059 | 1× phospho [S952(99.5)] |
| PARP1 | P09874 | [R].EELGFRPEYSASQLK.[G] | 1× phospho [S12(100)] | 0.0073 | 0.309 | 1× phospho [S179(100)] |
| Gene symbol | Protein accession | Exp. q-value: combined | Sum PEP score | Coverage [%] | # Peptides | # PSMs | # Protein unique peptides | # Unique peptides | Modifications in protein (% confidence) | Present in UniProt/previously detected |
|---|---|---|---|---|---|---|---|---|---|---|
| Exp. q-value: combined, sum PEP score, and % confidence are all scoring features reported by Proteome Discoverer. | ||||||||||
| USP11 | P51784 | 0 | 219.443 | 42 | 29 | 868 | 28 | 29 | Phospho [S952(99.5)] | Yes |
| HDAC2 | Q92769 | 0 | 85.678 | 26 | 8 | 452 | 1 | 4 | Phospho [S422(100)] | Yes |
| PARP1 | P09874 | 0.01 | 2.08 | 1 | 1 | 4 | 1 | 1 | Phospho [S179(100)] | Yes |
| SAP25 | Q8TEE9 | 0 | 110.312 | 56 | 6 | 1056 | 6 | 6 | Methyl [R18(100); R51(100); R54(100); R94(100)] | No PTMS listed in UniProt for SAP25 |
| ZNF318 | Q5VUA4 | 0 | 3.697 | 1 | 1 | 8 | 1 | 1 | Methyl [R1241(100)]; phospho [S1230(100)]; acetyl [K1232(100)] | Not listed in UniProt |
| ZNF556 | Q9HAH1 | 0 | 3.732 | 5 | 1 | 39 | 1 | 1 | Methyl [K405(99.4)]; acetyl [K396(100)] | Not listed in UniProt |
FAIMS expands the dynamic range of LC-MS/MS analyses.6,23,25,44 FAIMS increases signal-to-noise ratios by reducing the chemical background and acting as a separation device and generating cleaner mass spectra with higher confidence peptide detection.25 FAIMS readily separates multiply charged peptide precursors from singly charged background ions and can also separate based on charge state, conformers, and structural features.17,45 Studies have shown that FAIMS enhances sensitivity when paired with monolithic capillary columns on an orbitrap fusion lumos.1 Coupling the FAIMS interface with MS has shown significant promise in enhancing the sensitivity and depth of proteomic profiling,1,18,26 including in human serum,23 FFPE tissue, and exosomes.6,20 Previously, there were several studies conducted with FAIMS on an Orbitrap Eclipse tribrid mass spectrometer for human serum proteoforms and single-cell proteomics23,46 and FLAG-tagged Co-IP analysis,28 Orbitrap Exploris 480 for FFPE tissue and plasma-derived exosomes6 and, the Orbitrap Fusion Lumos mass spectrometer for targeted proteomics of tumor tissue.47 Beyond all these, the latest Orbitrap Ascend tribrid mass spectrometer, which features a modified Orbitrap with additional ion-routing multipole (IRM) enhances scan speed and detection sensitivity, improves mass spectra quality, expands proteome coverage, and uncovers functionally important low-abundance proteins.30,31 However, we were unable to find a study that compares the performance of the Orbitrap Ascend tribrid mass spectrometer with FAIMS and without FAIMS for affinity-purified protein complexes.
Here we studied the benefits of FAIMS for the analysis of affinity-purified protein complexes, comparing the performance of Orbitrap Ascend tribrid mass spectrometer with FAIMS and with No FAIMS. Identifying protein complexes and their subunits is a key to understanding many biological processes. Proteomic studies can identify a protein's interaction partners, shedding light on the functions and regulatory mechanisms of protein complexes.48 This method allows for the exploration of protein interaction networks within a cell, offering critical insights into the structure and function of its protein machinery.11 Therefore, an in-depth protein analysis is crucial.
In this study, we found that mass spectrometry analyses with No FAIMS of affinity-purified SAP25 protein complex isolated from Flp-In™-293 cells resulted in fewer unique proteins and peptides, which is 13 and 169, respectively. Additionally, the average number of proteins and peptides detected among six SAP25 runs were less than 159 and 390, respectively. However, the number of proteins and peptides substantially increased when FAIMS was coupled to the Orbitrap Ascend tribrid mass spectrometer, resulting in 92% and 77% gain in numbers for unique peptides and 53% and 43% for a total number of proteins and peptides respectively (Tables S1 and S2, ESI†). The significant rise in the number of proteins and peptides for AP-MS samples with FAIMS is due to the ability of FAIMS to filter ions of interest at the entrance of the mass spectrometer.6,23,26,40 In addition, the normalized total abundances for proteins and peptides observed with FAIMS and No FAIMS have a significant difference among the total six SAP25 runs. Despite that, when we consider the abundance of individual SAP25 runs, the increase in total normalized protein and peptide abundances at the log2 scale with No FAIMS was higher mostly due to the presence of high abundance proteins.6,21 This suggests that using FAIMS with nanoLC-MS will strongly enhance the detection of proteins and peptides and is superior to traditional nanoLC-MS, as observed previously with U937 myeloid leukemia cells,17 HeLa cells26 and K562 cells.18
One of the noteworthy results of the study was related to the effects of FAIMS on missing values, considered as a major issue in mass spectrometry data analysis37–39 by hindering the statistics and reproducibility of data.49 Previously it was determined that using FAIMS with mass spectrometry significantly reduces the number of missing values.6,40 The same observation occurs with our AP-MS study. We have observed that FAIMS significantly reduced the number of missing values compared to the No FAIMS configuration, therefore, significantly increasing the number of peptides and proteins identified and quantified (Fig. 4A–C). The missing values in No FAIMS group were mostly related to low abundances at the peptide and protein levels. Importantly, the proteins and peptides with low abundances were identified and quantified exclusively under FAIMS configurations (Fig. 4D and E), which may be lost by conventional LC-MS analyses. This observation suggests that FAIMS significantly improves signal-to-noise (S/N) ratio and filtering ions based on distinct mobility resulted in the identification of low abundance proteins and peptides.25,44
Finally, a potential limitation of the study is the work was carried out with three CVs and the default FAIMS configuration. We found that FAIMS with three CVs with DDA acquisition provides the highest protein and peptide-level detection. Further, we determined that FAIMS minimizes missing values and filtered the contaminants.37 However, some SAP25-associated proteins (Fig. 5) detected under the No FAIMS configuration were lost with FAIMS. This may be attributed to FAIMS self-cleaning mechanism, field heating effects or limited ion transmission.25,50 Therefore, further experiments may be required to optimize FAIMS parameters and the right selection of CVs value, which could lead to a better outcome by enhancing detection of proteins and peptides.17,24,45 Furthermore, optimizing both chromatography and mass spectrometry parameters may enhance the separation and detection at protein and peptide levels.
The FAIMS source, when coupled between nanoLC and Orbitrap Ascend tribrid mass spectrometer (nanoLC-FAIMS-MS) improves the dynamic for bottom-up proteomics, including affinity-purified protein complexes. Here, we demonstrated that the application of FAIMS with three CVs on an Orbitrap Ascend tribrid mass spectrometer in DDA increased the sensitivity for protein and peptide detection by 42% and 44% respectively compared to No FAIMS configuration. More specifically, FAIMS enhances the sensitivity of nanoLC by reducing the level of chemical contaminants and reducing the number of missing values from 60.17% (No FAIMS) to 14.57% (FAIMS). Therefore, our study suggests FAIMS is a valuable component to proteomics analyses. Overall, our result indicates that FAIMS addresses several limitations of AP-MS based proteomics analysis at chemical contaminant and peptide levels.
One potential approach for researchers to use when they do not have FAIMS on their system would be to use sample preparation approaches or digested sample clean up approaches that would reduce the chemical contaminants of their sample of interest prior to loading it onto an LC-MS/MS system. At the isolated protein complex level this could include gel-based fractionation or size exclusion chromatography, for example. In addition, sample clean up methods could be used during the digestion process like the S-trap or other similar methods.
Footnote |
| † Electronic supplementary information (ESI) available: Detailed information for protein level detection observed with SAP25 affinity purification (Table S1), detailed information for peptide level detection observed with SAP25 affinity purification (Table S2), detailed information about log2 value for Protein level detection (Table S3), detailed information about unique and razor peptides observed with SAP25 affinity purification (Table S4), detailed information about the contaminants derived from skyline (Table S5), detailed information about the contaminants observed with SAP25 affinity purification (Table S6), and the results from the post-translational modification search of the datasets (Table S7). See DOI: https://doi.org/10.1039/d5mo00038f |