Saeedreza Zeibi Shirejiniab,
Mehrdad Khamooshibc,
Damien Riskaf,
Martin Nikolove,
Marjan Azimib,
Shiyen L. Pereraabe,
Josie Carberryb,
Karen Alt
a,
Shaun D. Gregory†
*bc and
Christoph E. Hagemeyer†
*ad
aAustralian Centre for Blood Diseases (ACBD), Monash University, Melbourne, Victoria, Australia. E-mail: christoph.hagemeyer@monash.edu
bCardio-Respiratory Engineering and Technology Laboratory (CREATElab), Mechanical and Aerospace Engineering, Monash University, Clayton, Victoria, Australia
cCentre for Biomedical Technologies and School of Mechanical, Medical and Process Engineering, Queensland University of Technology, Brisbane, QLD, Australia. E-mail: shaun.gregory@qut.edu.au
dMonash Biomedical Imaging, Monash University, Clayton, Victoria, Australia
eMechanical and Aerospace Engineering, Monash University, Clayton, Victoria, Australia
fChemical and Biomedical Science, Monash University, Clayton, Victoria, Australia
First published on 30th July 2025
Aortic valve stenosis is a progressive cardiovascular disease associated with increased thrombotic risk due to abnormal blood flow patterns. Current management often culminates in valve replacement surgery, demonstrating the need for less invasive therapeutic options. This study investigates the potential of caplacizumab, a von Willebrand factor (vWF) inhibitor, in mitigating thrombosis risk in a microfluidic model of aortic valve stenosis. We employed a novel microfluidic model simulating the hemodynamics of healthy, moderate, and severe stenotic conditions, complemented by computational fluid dynamics simulations (CFD) and conventional platelet function assays. Microfluidic experiments revealed that shear gradients play a critical role in platelet aggregation, with accumulation intensifying as stenosis severity increased, even under constant peak shear rates. Caplacizumab demonstrated high specificity for vWF-mediated platelet aggregation, significantly inhibiting ristocetin-induced aggregation while not affecting ADP-induced aggregation. At an effective concentration (30 nM), caplacizumab reduced platelet coverage by up to 90% in high shear conditions (4500 s−1) and effectively mitigated shear gradient-dependent platelet aggregation across all stenotic conditions. These findings highlight caplacizumab's therapeutic potential for thrombosis prevention in patients with aortic valve stenosis, offering a foundation for personalized antithrombotic approaches that could potentially reduce thrombotic complications associated with the disease.
In aortic valve stenosis, platelet activation and aggregation, as well as von Willebrand factor (vWF) unfolding, occur due to elevated shear forces in the stenotic region.5 While high shear forces exist in both acceleration and deceleration zones, the deceleration region is particularly thrombogenic.6 As blood flows through the stenotic region, it experiences a rapid acceleration followed by a sudden deceleration. High shear stress in the stenotic region induces conformational changes in von Willebrand factor (vWF), exposing the A1 domain for direct binding to platelet GPIbα receptors. This shear-induced platelet aggregation (SIPA) occurs through shear-dependent vWF–GPIbα interactions that are rapid and independent of classical platelet activation pathways.7 In aortic valve stenosis, while high shear initiates vWF unfolding and GPIbα binding in the stenotic region, the downstream deceleration zone provides favorable conditions for platelet cluster stabilization due to lower dispersive forces and increased residence time.8
The sudden change in shear stress activates mechanosensitive receptors on the platelet surface,9 while the elongational flow causes significant stretching and unfolding of vWF multimers, exposing their hidden binding sites for platelets.10 These factors, along with the altered fluid dynamics in the deceleration region, promote platelet aggregation and thrombus formation downstream of the stenosis (Fig. 1).11 Understanding of the complex mechanisms that underlie thrombosis formation in aortic valve disorders is essential for the development of effective therapeutic and preventive strategies.
Existing interventions for aortic valve stenosis primarily centre around surgical interventions such as valve replacement or repair.12 While effective, these procedures entail inherent risks and might not be feasible for all patients, especially those with significant comorbidities or contraindications for surgery. Furthermore, pharmacological therapies aimed at mitigating platelet aggregation in this specific context remain relatively limited. The fundamental challenge of preventing shear-induced thrombosis while avoiding systemic bleeding complications in aortic stenosis highlights the clinical need for targeted antithrombotic approaches that can respond to local hemodynamic conditions.13 In recent years, significant progress has been made in the development of inhibitors targeting the GPIb-IX-V receptor complex and vWF, aiming to curtail thrombus formation. Notably, inhibitors such as caplacizumab (Cablivi), ARC1779, and anfibatide have emerged as promising agents targeting the vWF–GPIbα interaction, effectively reducing thrombus formation.14–16
Caplacizumab, a bivalent NANOBODY® molecule against vWF, helps prevent blood clots by blocking the interaction between vWF and the GPIbα receptor. This is especially effective under high-shear conditions, where it limits platelet adhesion to injured vessel walls. It was approved by the European Medicines Agency in 2018 and by the US Food and Drug Administration in 2019 for the treatment of acquired TTP (aTTP). Extensive nonclinical and clinical studies have shown that caplacizumab is highly effective in preventing platelet adhesion and aggregation, significantly reducing thrombotic events while minimising the risk of bleeding complications.17
While caplacizumab shows promising potential in addressing thrombosis, understanding its effectiveness within the complex hemodynamic environment of stenotic aortic valves requires sophisticated experimental approaches. Clinical evaluations demonstrate the drug's overall efficacy but provide limited insight into how it functions within the intricate flow environments characteristic of valve stenosis. To bridge this gap between pharmacological action and clinical outcomes, complementary technologies that can replicate and analyze the pathophysiological flow conditions of stenotic valves are essential. Conventional models, such as parallel-plate flow chambers18 and cone-and-plate shearing devices,19 often fall short in capturing the intricate flow dynamics and shear stress profiles associated with aortic valve disorders.
To overcome these limitations, microfluidic platforms offer the ability to create controlled environments with precise shear stress levels, gradients, and stenotic geometries that mimic the complex flow dynamics found in diseased valves, enabling detailed investigation of platelet aggregation under physiologically relevant conditions. When coupled with computational fluid dynamics (CFD),20,21 these experimental systems provide comprehensive insights into how local hemodynamic parameters, particularly shear gradients and flow deceleration zones, influence thrombotic processes and therapeutic interventions.
While prior studies have advanced thrombosis modeling using CFD simulations or microfluidic platforms,22–26 these approaches face critical limitations when applied to aortic stenosis. CFD-only models characterize shear environments but lack biological validation, while traditional in vitro systems such as parallel-plate and cone-and-plate devices typically apply uniform shear but fail to replicate the spatially heterogeneous and gradient-driven flow fields characteristic of stenotic valves. Moreover, existing approaches emphasize peak wall shear stress without systematically examining shear gradients or incorporating specific geometries with therapeutic validation. These limitations restrict mechanistic understanding of gradient-driven thrombus formation and hinder evaluation of targeted therapeutics such as caplacizumab under stenosis-relevant hemodynamic conditions.
This study provides novel insights into thrombosis dynamics in aortic valve stenosis through a uniquely integrated experimental–computational framework. By combining stenosis geometries with CFD modeling and real-time platelet aggregation imaging under defined shear gradients, we evaluate caplacizumab's efficacy in mitigating shear-induced thrombosis. This platform not only expands on previous peak shear-focused models but also bridges a key translational gap by capturing the interplay between hemodynamic forces and therapeutic action in stenotic cardiovascular disease.
Equipment used included an AggRam platelet aggregometer (Helena Laboratories, USA), Aurora Northern Lights flow cytometer (Cytek Biosciences, Fremont, CA), Nikon Ti-E inverted fluorescence microscope equipped with a 20× objective (NA 0.75) and sCMOS Andor Zyla 4.2 PLUS camera (Oxford Instruments), and a Legato 130 programmable syringe pump (KD Scientific, USA). Data analysis was performed using FlowJo software and GraphPad Prism v9.1 (GraphPad Software, San Diego, CA).
The PDMS microfluidic device was fabricated using standard photolithography at the Melbourne Centre for Nanofabrication (MCN). The microchannel layout, designed in AutoCAD, was transferred onto a photomask and patterned onto a silicon wafer coated with SU-8 photoresist to achieve a channel depth of 58 μm. After soft-baking to evaporate solvents, the photoresist was selectively crosslinked by UV exposure through the photomask, followed by a post-exposure bake. Unexposed photoresist was removed in a developer, leaving a patterned silicon master. The master was silanised to facilitate PDMS release. A 10:
1 mixture of PDMS prepolymer and curing agent was poured over the master and cured at 80 °C for 4 hours (Fig. 2). Once cured, the PDMS layer was peeled off, inlet and outlet ports were punched, and the device was sealed with a glass coverslip.28
CFD simulations were employed to complement experimental microfluidic analyses, providing detailed insights into fluid dynamics within the aortic valve stenosis microchannels. The simulations numerically solved the continuity and Navier–Stokes equations for laminar flow to predict velocity fields, shear rate, and shear stress profiles across the computational domain. The microchannel geometry was discretised using a poly-hexcore meshing scheme in Fluent Meshing 2023R2, resulting in a high-resolution finite volume mesh with 8103
340 cells, allowing for precise capture of complicated flow details. A mesh convergence study confirmed the adequacy of the mesh resolution, with the root mean squared error in velocity magnitude and wall shear stress remaining below 1%. Simulations were conducted in ANSYS Fluent using the semi-implicit method for pressure-linked equations (SIMPLE) algorithm to solve the coupled system of equations.
Blood was modelled as an incompressible, Newtonian fluid with a density of 1050 kg m−3 and a dynamic viscosity of 0.00345 Pa s. This assumption is valid for shear rates exceeding 500 s−1, where blood viscosity plateaus and demonstrates near-Newtonian behavior.23,30 All simulated flow conditions in this study (562–4500 s−1) fall within this validated regime. Previous studies have reported that shear rates in stenotic valves can reach over 3000 s−1 in severe cases,31,32 substantially exceeding the normal physiological range of 100–1000 s−1 in healthy valves.33 For our computational analysis, flow rates of 0.83, 1.66, 3.33, 6.66, and 13.3 μL min−1 were simulated across all stenosis models to characterize how geometrical differences affect shear rate distributions and gradients at constant input flow. These flow rates were selected to generate shear rate ranges comparable to those observed in vivo, allowing our model to capture the key hemodynamic features relevant to platelet aggregation despite the geometric simplification. Boundary conditions included a no-slip condition at channel walls and a specified pressure boundary (0 Pa) at the outlet.
The CFD model was used to establish flow rate-shear rate relationships for experimental design and does not simulate cellular-level interactions or thrombosis formation. We acknowledge that for channel dimensions below ∼100 μm, including the 48 μm severe stenosis model, discrete cellular phenomena become dominant and are not captured by continuum models.34,35
Light-transmission platelet aggregometry was conducted using AggRam systems (Helena Laboratories, USA) at 37 °C with constant stirring at 600 rpm. For calibration, 300 μL of blank PPP was placed in a cuvette under stirring conditions. To assess platelet aggregation, ristocetin (1.5 mg mL−1 final concentration; Helena Biosciences, Australia) and ADP (10 μM final concentration; Sigma-Aldrich) were added to the PRP solution in separate channels. The final volume was adjusted to 300 μL with PBS, and all runs were performed in duplicates. Before agonist addition, caplacizumab was introduced to pre-warmed (37 °C) PRP at concentrations ranging from 0 to 50 nM and incubated for 20 minutes. Platelet aggregation responses were assessed with and without caplacizumab. The amplitude of aggregation was determined 10 minutes after agonist addition as an indicator of the degree of platelet aggregation. Light transmission data were expressed as mean ± SEM percentage of platelet aggregation and plotted using HemoRam 1.3 software.
Platelet coverage area was quantified using FIJI image analysis software, and results were expressed as mean ± SEM percentage of the total surface area. Quantitative image analysis was restricted to regions characterized by elevated shear stress and steep spatial shear gradients, specifically, the stenotic throat and downstream deceleration zone, where vWF unfolding and GPIbα-mediated platelet aggregation are the primary mechanisms of interest for this study. Recessed regions, designed to replicate the anatomical sinuses of Valsalva, were excluded from analysis due to their low-shear and altered flow dynamics, which promote different platelet aggregation dynamics than the high-shear, vWF-mediated mechanisms that are the focus of this investigation.
In contrast, ADP-induced platelet aggregation remained stable across all caplacizumab concentrations, ranging from 88.3 ± 3.9% at baseline to 83.6 ± 3.6% at 50 nM (p > 0.05). Statistical analysis revealed no significant differences in ADP-induced aggregation at any caplacizumab concentration tested (multiple t-tests, p > 0.05). This selective inhibition pattern highlights caplacizumab's specificity for the VWF-mediated pathway, while preserving normal platelet responses to other physiological agonists.
Analysis of GPIb receptor availability using anti-CD42b antibody showed no significant changes in receptor expression across all conditions. Neither ristocetin treatment (0.75 ± 0.12-fold vs. control, p > 0.05) nor the addition of caplacizumab (0.92 ± 0.05-fold vs. control, p > 0.05) significantly altered CD42b levels, indicating that the vWF–platelet interactions under these conditions do not affect GPIb receptor accessibility.
The effective caplacizumab concentration used in this study falls within a clinically relevant range. Clinical dosing of caplacizumab (10 mg IV bolus followed by 10 mg SC daily) are well established for acquired thrombotic thrombocytopenic purpura.38 The selected concentration accounts for the absence of systemic clearance and plasma protein binding in vitro, while ensuring robust inhibition of vWF–platelet interactions under high shear conditions (>2250 s−1) that are characteristic of aortic valve stenosis. This concentration supports the translational potential of current findings for potential clinical applications in aortic stenosis.
The shear rate analysis revealed distinct patterns across stenosis severities (Fig. 6, top row). While all models showed elevated shear rates at the stenotic region, the pattern evolved from a moderate elevation in the healthy model to an intense, highly localised peak in the severe stenosis model. Peak shear rates were observed at the stenotic region, with values increasing dramatically with stenosis severity. The relationship between peak shear rate and flow rate, maintained linearity across all stenosis conditions (Fig. 7A), with the severe stenosis model demonstrating a significantly steeper slope (approximately 1616 s−1 min μL−1) compared to moderate (528 s−1 min μL−1) and mild stenosis geometries (327 s−1 min μL−1). At the highest tested flow rate (13.3 μL min−1), peak shear rates reached approximately 22259 s−1 in the severe stenosis model, compared to 7278 s−1 and 4500 s−1 in the moderate and healthy stenosis models, respectively. Fig. 7B presents the flow rates required to achieve equivalent peak shear rates across all three stenosis models for the experimental studies.
Detailed examination of the shear rate profiles revealed a characteristic biphasic pattern with distinct acceleration and deceleration zones (Fig. 8). In the severe stenosis model, the shear rate gradients exhibited marked asymmetry around the stenotic region (x ≈ 400 μm). At the highest flow rate (13.3 μL min−1), the acceleration phase generated positive gradients of 2 × 108 s−1 μm−1, followed by negative gradients of similar magnitude in the deceleration phase. This gradient pattern was flow rate-dependent, with proportionally lower magnitudes observed at reduced flow rates (e.g., 5 × 107 s−1 μm−1 at 3.33 μL min−1). The moderate and healthy stenosis models demonstrated qualitatively similar patterns but with substantially reduced gradient magnitude.
Wall shear stress (WSS) analysis revealed characteristic patterns that evolved with increasing stenosis severity (Fig. 6, middle row). Maximum WSS values were consistently observed at the point of maximal constriction in all models, with the severe stenosis model generating the highest local WSS. The WSS distribution showed spatial heterogeneity, particularly in the severe stenosis model, where sharp transitions were observed at the entrance and exit of the stenotic region. The highest shear rate gradients developed at these transition zones, immediately upstream and downstream of the stenosis, and these high-gradient regions corresponded to areas of platelet accumulation observed in our experimental studies, as illustrated in Fig. 9 and 10A, where platelet coverage is highest in the downstream deceleration region.
![]() | ||
Fig. 10 Comprehensive analysis of platelet accumulation and caplacizumab's inhibitory effects across stenosis conditions. (A) Regional platelet coverage in severe stenosis (4500 s−1) showing preferential downstream accumulation, effective inhibition by caplacizumab and eptifibatide, and minimal fluorescent bead deposition (****p < 0.0001 vs. control). (B) Shear rate-dependent platelet accumulation in control conditions demonstrating progressive increase with shear rate and stenosis severity. (C) Caplacizumab treatment showing maximum efficacy above 2250 s−1 across all stenosis conditions. (D) Caplacizumab effectiveness across stenosis severities at 4500 s−1 demonstrating consistent inhibition (**p < 0.01, ****p < 0.0001 vs. respective control). Panels B–D quantify downstream region coverage (blue dashed box, Fig. 9B). Note: panels B and D show results at matched peak shear rates across geometries, with differences in platelet coverage reflecting varying local shear rate gradients. Data presented as mean ± SEM (n = 3). |
Based on the CFD characterization of flow fields, which identified critical shear rate thresholds and distributions across different stenosis geometries, microfluidic experiments were designed to investigate platelet aggregation dynamics under precisely controlled shear conditions. Using the peak shear rate data from the computational analysis, specific flow rates for healthy, moderate and severe stenosis were selected to generate physiologically relevant peak shear rates (562, 1125, 1687, 2250, 3375, and 4500 s−1) at the stenotic region. These targeted shear conditions allowed exploration of platelet behaviour across the range from moderate to high pathological shear rates identified in the CFD models. Fluorescence microscopy of DiOC6-labeled platelets in severe stenosis revealed distinct shear rate-dependent aggregation patterns in control samples (Fig. 9, top row). The effective caplacizumab concentration (30 nM), determined from earlier aggregometry studies, was employed to evaluate inhibitory effects across these various shear conditions. At the lower peak shear rates (1125 s−1), moderate and diffuse platelet accumulation was observed throughout the channel. When peak shear rates were increased to 3375 s−1 and 4500 s−1, platelet aggregation became more pronounced and localized at the stenotic region, correlating with the areas of maximum shear stress identified in the CFD analysis (Fig. 6).
Time-course studies at 4500 s−1 demonstrated rapid thrombus formation in control conditions (Fig. 9, bottom panel). Substantial platelet accumulation was evident within the first 5 minutes, with continued growth observed through 10 minutes of perfusion. The spatial distribution of these aggregates corresponded to regions of high shear gradients identified in the computational analysis. Caplacizumab treatment significantly altered platelet accumulation patterns across all tested shear rates. The most pronounced inhibitory effect was observed at higher shear rates (3375 and 4500 s−1), where control samples exhibited maximum aggregate formation. Time-series analysis at 4500 s−1 revealed that caplacizumab effectively prevented the formation of large platelet aggregates throughout the 10-minute observation period (Fig. 9, bottom panels), maintaining consistently low platelet adhesion even in regions of high shear stress.
To characterise the spatial distribution of platelet aggregation under physiologically relevant shear conditions, platelet area coverage was quantified in three distinct regions (defined in Fig. 9B) of the severe stenosis model including upstream, stenosis, and downstream regions (Fig. 10A). At 4500 s−1, control samples demonstrated a distinctive pattern of platelet accumulation, with significantly enhanced coverage in the downstream region (75 ± 5%) compared to upstream (3.7 ± 2%, p < 0.001) and stenosis regions (27.3 ± 11.6%, p < 0.003). This pattern aligned with CFD-identified flow deceleration zones, consistent with enhanced post-stenotic platelet aggregation in previous studies.10
To confirm biological specificity, control experiments using 6 μm APC-conjugated fluorescent beads in PBS and platelet-poor plasma showed minimal accumulation (<2% area coverage), confirming that observed aggregation represents platelet accumulation rather than flow-induced particle entrapment (Fig. 10A shows PBS control). Therapeutic validation with eptifibatide, a clinically established GPIIb/IIIa inhibitor, demonstrated comparable inhibitory efficacy to caplacizumab in the downstream region (5 ± 0.6% vs. 6.2 ± 0.4% coverage, respectively, p = 0.15). However, these agents differ fundamentally in their mechanistic scope and clinical implications.
Caplacizumab treatment significantly altered this spatial pattern of platelet accumulation. At 4500 s−1, the drug markedly attenuated downstream platelet accumulation (6.2 ± 0.7% vs. 75 ± 5% control, p < 0.0001), while maintaining minimal effect on upstream (5.2 ± 4.4%) and stenosis regions (9 ± 7.5%). The pronounced inhibitory effect of caplacizumab specifically in the downstream (deceleration) region, suggested a mechanistic link between vWF-mediated platelet adhesion and local hemodynamic conditions. Based on these observations, subsequent analyses focused on this downstream region to further evaluate the relationship between stenosis severity and platelet accumulation.
Analysis of platelet accumulation under control conditions revealed a direct relationship between stenosis severity and shear rate-dependent platelet aggregation (Fig. 10B). In the severe stenosis model, platelet coverage increased sharply from 12.3 ± 1.4% at 562 s−1 to 66.1 ± 8.7% at 2250 s−1, ultimately reaching 75.1 ± 2.9% at 4500 s−1. The moderate stenosis condition showed a similar trend but with lower magnitudes, increasing from 6.6 ± 1% at 562 s−1 to 68.1 ± 3% at 4500 s−1. The healthy model demonstrated the most gradual increase, reaching a maximum of 42.3 ± 6.5% at 4500 s−1.
Quantification of platelet accumulation across varying shear rates demonstrated distinct response patterns in different stenosis geometries (Fig. 10C). At lower shear rates (562–1125 s−1), minimal platelet accumulation was observed, with coverage values below 15% across all conditions. As shear rates increased, platelet coverage reached maximum values at 2250 s−1, with the severe stenosis model showing the highest coverage (49.6 ± 4.9%), followed by moderate (39.3 ± 8.3%) and healthy conditions (24.9 ± 7.3%). Beyond 2250 s−1, a significant decrease in platelet accumulation was observed across all stenosis conditions. This reduction was most pronounced in the severe stenosis model, where platelet coverage decreased from 49.6 ± 4.9% at 2250 s−1 to 6.2 ± 0.39% at 4500 s−1 (p < 0.001). Similar trends were observed in moderate stenosis (39.3 ± 8.3% to 8.1 ± 3%, p < 0.05) and healthy conditions (24.9 ± 7.3% to 6.8 ± 1.5%, p > 0.05). The sharp decline in platelet accumulation above 2250 s−1 was consistent across all geometries, with minimal differences between conditions at the highest tested shear rate (4500 s−1).
The experimental design utilized matched peak shear rates across three stenosis geometries, allowing examination of how shear rate gradients influence platelet aggregation independently of peak shear magnitude. Under these controlled conditions (4500 s−1), platelet coverage increased systematically with stenosis severity: healthy (42.3 ± 6.5%), moderate (68.1 ± 3%), and severe (75.1 ± 2.9%), a 77% increase from healthy to severe conditions (Fig. 10B and D), despite identical peak shear rates.
Analysis of the downstream region at high shear rate (4500 s−1) demonstrated the inhibitory effect of caplacizumab across different stenosis severities (Fig. 10D). In healthy conditions, platelet area coverage was reduced from 42.3 ± 6.5% in control to 6.8 ± 1.5% with caplacizumab treatment (p < 0.006). Similar inhibitory effects were observed in moderate stenosis, where coverage decreased from 68.1 ± 3% to 8.1 ± 3% (p < 0.0001), and in severe stenosis, where caplacizumab reduced coverage from 75.1 ± 2.9% to 6.2 ± 0.4% (p < 0.0001).
To examine the relationship between flow velocity and platelet aggregation, we quantified area-averaged downstream velocities in the downstream deceleration zone (blue quadrilateral region, Fig. 9B) across stenosis severities under matched peak shear conditions (4500 s−1) and compared them to corresponding platelet surface coverage (Fig. 11). Downstream velocities decreased with increasing stenosis severity (healthy: 0.014 m s−1; moderate: 0.007 m s−1; severe: 0.002 m s−1), while platelet surface coverage increased (healthy: 42.3 ± 6.5%; moderate: 68.1 ± 3.0%; severe: 75.1 ± 2.9%). This inverse relationship indicates that platelet aggregation increases as downstream velocity decreases, even under matched peak shear conditions.
Flow cytometry experiments in the present study provided detailed mechanistic insights, demonstrating reduction in the median fluorescence intensity of vWF-bound platelets in the presence of caplacizumab. The observation of partially restored GPIb receptor availability, evidenced by increased CD42b MFI during caplacizumab treatment in our experiments, reveals the drug's specific mechanism in preventing vWF–GPIb interactions while preserving receptor functionality. This specificity was further supported by the platelet aggregation assays performed in this study, showing selective inhibition of ristocetin-induced but not ADP-induced aggregation. These findings are consistent with previous research38 and suggest minimal interference with normal hemostatic functions.
The CFD analysis revealed critical hemodynamic features that underlie thrombogenic potential in stenotic valve conditions. The transition from uniform flow in healthy conditions to highly heterogeneous patterns in severe stenosis was characterized by dramatic increases in local shear rates. Particularly significant was the computational identification of distinct acceleration and deceleration zones, where sharp spatial gradients in shear rate (approximately 2 × 108 s−1 μm−1) would create conditions favourable for platelet activation and aggregation (Fig. 6). The correlation between these high-gradient regions and subsequent platelet accumulation is demonstrated in the experimental results (Fig. 9).
The spatial distribution of wall shear stress, particularly the sharp transitions observed at stenosis entry and exit points, identifies potential sites for platelet activation and adhesion. These computational findings provide a mechanistic framework for understanding the enhanced platelet accumulation observed in our experimental studies. The identification of distinct high-shear activation zones and low-velocity deceleration regions demonstrate that high shear stress activates platelets via vWF unfolding, steep shear gradients amplify this activation, and downstream regions with reduced flow velocity allow activated platelets to aggregate into stable thrombi.
The microfluidic model included recessed regions designed to replicate the anatomical sinuses of the aortic valve. While platelet-associated fluorescence was observed in these areas, even with caplacizumab treatment, these regions were excluded from quantitative analysis due to their distinct hemodynamic characteristics. These recessed geometries are known to experience markedly low wall shear stress and recirculatory flow patterns,43 promoting passive platelet trapping rather than shear-induced vWF–GPIbα adhesion. Under such low-shear conditions, platelet aggregation likely occurs via alternative pathways (e.g., fibrinogen–αIIbβ3),4 explaining persistent platelet signals despite vWF inhibition. These findings highlight that multiple thrombotic mechanisms may operate in spatially distinct hemodynamic zones of the stenotic valve, suggesting that complementary therapies may be required to manage thrombosis risk beyond high-shear regions.
The experimental approach required varying inlet flow rates to achieve equivalent peak shear rates across different stenosis geometries, creating distinct velocity profiles and platelet residence times across geometries.44,45 This design reflects the physiological reality where cardiac output remains relatively constant while valve cross-sectional area decreases with disease progression. While peak shear rates were systematically matched across stenosis geometries in experimental comparisons, CFD analysis revealed that corresponding shear rate gradients varied by over an order of magnitude (Fig. 8A). This design enabled isolation of gradient-specific effects. The 77% increase in platelet coverage from healthy to severe stenosis, despite identical peak shear, highlights the mechanistic importance of shear gradients. The spatial correspondence between high-gradient downstream regions and sites of maximal platelet accumulation (Fig. 10A) provides direct evidence that local shear rate gradients, rather than peak shear stress alone, drive platelet aggregation in stenotic flow environments. This gradient-driven mechanism is consistent with prior studies demonstrating that abrupt shear transitions and steep spatial gradients can promote von Willebrand factor unfolding and platelet activation under pathological flow conditions.46,47
Importantly, multiple lines of evidence demonstrate that shear gradient-dependent mechanisms, rather than residence time effects, drive the observed platelet behavior: the preferential platelet accumulation in downstream deceleration zones correlating with CFD-predicted gradient regions (Fig. 10A), the consistent 2250 s−1 activation threshold across all geometries despite varying residence times (Fig. 10C and D), and caplacizumab's specific efficacy against shear-activated vWF pathways. These findings consistently indicate that platelet accumulation patterns correlate strongly with regions of steep shear gradients, suggesting that mechanical forces rather than residence time alone drive the enhanced thrombosis risk observed in severe stenosis.
This interpretation is further supported by the inverse relationship between local flow velocity and platelet aggregation across stenosis severities (Fig. 11), where lower velocities in severe stenosis correspond to higher aggregation despite matched peak shear rates. This confirms that flow velocity and residence time do not independently account for the observed thrombotic patterns. These results align with prior studies demonstrating that spatial shear gradients, particularly in deceleration zones, are critical determinants of platelet aggregation and represent the dominant mechanical driver of thrombus formation under high-shear conditions.24,46,47
The presented microfluidic experiments provided compelling evidence for caplacizumab's selective inhibition of shear-induced platelet aggregation, particularly above 2250 s−1. Visual inspection of platelet deposition patterns reveals preferential accumulation in the stenotic throat region where computational analysis shows maximum shear gradients (Fig. 6 and 9A). This phenomenon, attributed to higher shear gradients in severe stenosis, extends previous work by Nesbitt et al.24 on shear gradient-dependent platelet behaviour to the specific context of aortic valve stenosis.
The comprehensive validation studies using fluorescent beads and eptifibatide comparisons provide critical context for our findings. The negligible accumulation of inert particles confirms that the stenosis model captures genuine biological platelet aggregation rather than flow artifacts. While eptifibatide demonstrates comparable acute efficacy, its broad GPIIb/IIIa inhibition across all shear conditions is associated with increased bleeding complications in clinical practice.48 Caplacizumab's selective inhibition of vWF–GPIbα interactions specifically under pathological shear conditions (>2250 s−1) represents a mechanistically targeted approach that may preserve normal hemostatic function while preventing pathological thrombosis, a critical advantage in treating high-shear cardiovascular conditions.
For clinical translation, this shear-dependent mechanism enables targeted therapy. Doppler echocardiography provides peak aortic jet velocities that correlate with stenotic shear rates, while 4D flow MRI can directly quantify wall shear stress.49 Patient-specific computational modeling based on imaging data can also predict individual shear environments.50 These non-invasive approaches could guide therapeutic decision-making by identifying patients whose stenotic conditions generate the high shear rates where caplacizumab demonstrates optimal efficacy. The designed aortic valve microfluidic model has potential applications in cardiovascular medicine by providing a platform that could eventually enable patient-specific thrombosis modelling. Its ability to replicate individual stenosis geometries while testing patient blood samples under physiological conditions bridges the gap between in vitro molecular studies and in vivo outcomes.
The absence of endothelial lining, although enabling isolation of direct shear-platelet–vWF interactions, represents an area for model enhancement. Endothelial cells significantly modulate hemostasis through vasoactive mediator release and local vWF production, particularly under the dysfunctional conditions present in aortic stenosis. Future endothelialized vasculature-on-chip platforms could build upon these foundational findings to incorporate these additional biological complexities. The computational simulations operated within the laminar regime (Re < 100), which effectively captured the primary shear-activation mechanisms despite differing from potential turbulent flow conditions in severe stenoses.
The continuum-based CFD model does not account for discrete cellular phenomena that become dominant in microchannels below ∼100 μm, including cell-free layer formation, platelet margination, and spatially heterogeneous hematocrit distribution. The model likely overestimates near-wall shear stress due to the absence of plasma-rich cell-free layers. Importantly, these limitations do not compromise the primary conclusions because: (1) key findings derive from direct experimental observations in microfluidic systems using whole blood, where all relevant microscale phenomena are naturally present; and (2) the CFD simulations served solely as tools for experimental parameter selection and relative geometric comparisons. Building upon these findings, future studies employing cell-resolved modeling approaches could offer complementary insights into microscale platelet transport mechanisms in stenotic flow environments.
Additionally, the use of healthy donor blood, essential for establishing reproducible baseline drug responses, provides a foundation for future studies incorporating patient-derived samples with their inherent vWF multimer variations and altered platelet reactivity. These strategic limitations highlight clear pathways for advancing this promising therapeutic approach through pulsatile flow validation, patient-specific modeling, and combination therapy optimization.