Ana Lores-Padin
*a,
Thibaut Van Acker
a,
Niels J. de Winterb,
Martin Wiech
c,
Simon Nordstadd,
Yannic Hallierd and
Frank Vanhaecke
a
aAtomic & Mass Spectrometry-A&MS Research Group, Department of Chemistry, Ghent University, Campus Sterre, Krijgslaan 281-S12, 9000 Ghent, Belgium. E-mail: ana.lorespadin@ugent.be
bDepartement of Earth Sciences, Vrije Universiteit Amsterdam, W&N Building, De Boelelaan 1085, 1081 Amsterdam, the Netherlands
cDepartement of Foreign and Infectious Substances, Institute of Marine Research, P. O. Box 1870, Nordnes, NO-5817 Bergen, Norway
dmyStandards GmbH, Schauenburger Straße 116, 24118 Kiel, Germany
First published on 6th October 2025
This study evaluated the micro-homogeneity of seven different commercially available nanoparticulate pressed pellets based on a CaCO3 matrix and their utility for quantitative elemental mapping of biogenic carbonates using laser ablation-inductively coupled plasma-time-of-flight-mass spectrometry (LA-ICP-TOF-MS). The analytical performance of matrix-matched calibration (using the aforementioned nano-pellets) was compared against that of non-matrix-matched calibration using a silicate glass (NIST SRM 610) reference material for quantification. Calibration with nano-pellets, combined with the use of Ca as an internal standard, significantly improved the quantification accuracy, providing recoveries between 80–120% for the majority of the 18 elements selected and spread across a wide concentration range (from sub-μg g−1 to tens of wt%). However, some nano-pellets (e.g., BPLM-NP and BAM-RS3-NP) exhibited higher heterogeneity, leading to biased recoveries. Also, an inverse correlation between the mass fraction and the relative standard deviation (RSD) was observed. Throughout the work, elemental mapping was conducted with a laser beam size of 20 × 20 μm2 as a compromise between spatial resolution, sensitivity (sub-μg g−1 limits of detection required for trace elements), linear dynamic range, total analysis time and size of the region-of-interest. The quantitative mapping approach enabled the generation of high-resolution, multi-elemental 2D-maps of various CaCO3-based natural chronological archives, including fish otoliths and bivalve shells, revealing detailed elemental distribution patterns for both trace (e.g., Mn, Ba) and major elements (e.g., Sr). This LA-ICP-TOF-MS methodology provides a powerful tool for resolving intricate microstructures and thus, for chronologically tracking bioaccumulation of environmentally relevant metals, offering significant advantages over traditional 1D (line scanning) analysis as the latter may lead to misinterpretation of elemental distributions.
Laser ablation – inductively coupled plasma-mass spectrometry (LA-ICP-MS) has been commonly used in the study of biogenic carbonates4 due to its excellent limits of detection (LoDs, typically in the ng g−1 range), a linear dynamic range of up to 10 orders of magnitude, minimal sample preparation requirements and high sample throughput.5 The technique has been used for both bulk (using large beam diameters, typically ≥100 μm) and microanalysis (bulk spot analysis and 1D-line scan analysis using spots typically ranging from 20 to 80 μm diameter).6 Recent advances in LA technology, including the introduction of low-dispersion ablation cells7,8 and transfer devices with improved aerosol transport efficiency,9 have significantly enhanced the technique's capabilities. These developments now enable mapping with a spatial resolution down to approximately 1 μm and yield shorter single pulse response (SPR) profiles of ≤1 ms duration (defined as the width of the transient signal observed upon firing a single laser pulse at 10% of its maximum height). Furthermore, the integration of time-of-flight analyzers in ICP-MS instrumentation (ICP-TOF-MS), allowing quasi-simultaneous monitoring of nearly the entire periodic table (typically 23Na+–238U16O+) within just a few microseconds (<50 μs), has further expanded the applicability of LA-ICP-MS.10 Together, these advances allow for rapid data acquisition at up to 1000 pixels per second, significantly boosting the technique's potential for ultra-fast, high-resolution, multi-elemental mapping.11
Despite significant advances in LA and ICP-MS instrumentation, accurate elemental quantification in carbonate matrices remains inherently challenging. This is primarily due to the matrix-dependent nature of the laser–sample interaction: with the commonly used nanosecond pulse lasers, the occurrence of thermal effects during the ablation process results in non-stoichiometric sampling and heterogeneous particle plumes, leading to elemental fractionation during ablation, aerosol transport and plasma processes such as particle vaporization and ion generation. In this context, apart from optimally tuning the LA parameters and ICP settings to minimize elemental fractionation (238U+/232Th+ signal ratio ∼1 upon ablation of a NIST SRM 610 or 612 glass from the National Institute of Standards and Technology),12,13 previous studies on carbonates have shown that also the laser wavelength exerts an important effect on the quantification. The use of a shorter laser wavelength, such as 193 nm, instead of 213 or 266 nm, enables more effective ablation of carbonate matrices due to the higher absorption efficiency, the formation of smaller particles (which significantly reduces vaporization-induced elemental fractionation), and reduced thermal effects.14,15 This results in better defined craters and minimal sample fracturing upon ablation.13,15,16 In addition, to further enhance analytical accuracy and precision in LA-ICP-MS applications, external calibration using certified reference materials (CRMs) or well-characterized in-house standards, combined with signal normalization based on the monitoring of an internal standard (IS), remains the most effective strategy.5,17 However, the limited availability of homogeneous and compositionally representative carbonate-based reference materials continues to constrain robust quantification for such matrices.13,18
In response to these limitations, various calibration strategies have been evaluated, including both matrix-matched13,19 and non-matrix-matched approaches20 (using Ca as an internal standard in all cases). Traditionally, silicate glass reference materials, such as NIST SRM 610 and 612 have been employed owing to their excellent level of homogeneity and well-characterized composition.21 However, Jochum et al.13 demonstrated that non-matrix-matched calibration introduces a systematic bias of 10–20% in trace elemental concentrations due to differences in ablation efficiency and ion production in the plasma between silicate and carbonate matrices. While this approach remains suitable for refractory elements (e.g., Ba, Sr), it is significantly less reliable for chalcophile and siderophile elements with a low boiling point (e.g., Pb, Cu, Zn and Cd), an observation which supports the conclusions of a previous study by Kroslakova et al.22 Consequently, matrix-matched calibration using CaCO3-based RMs has become the preferred strategy.
Due to the lack of commercially available CaCO3-based RMs, in-house calibration standards under the form of pressed powder pellets23 or fused glass discs,24 or obtained using co-precipitation methods25–27 were developed. Although well-characterized, limitations in terms of contamination and poor elemental micro-homogeneity constrained their wide applicability. These limitations highlighted the need for well-characterized carbonate RMs, leading to the development of certified RMs such as USGS MACS-1 and MACS-3.13,28 However, their coarse grain size (∼10–15 μm) causes inefficient and unstable ablation behaviour, often resulting in elevated analytical uncertainties, cross-contamination, and memory effects during LA analysis.16 In fact, Lazartigues et al.29 reported that using MACS-3 for obtaining quantitative high-resolution maps (i.e., 5 μm pixel size) of small otoliths led to high relative standard deviations (RSDs) of 30–80% due to trace element heterogeneity as previously discussed.
A significant advancement in the field was the development of a novel carbonate pellet fabrication method by Garbe-Schönberg and Müller,30 which enabled a reduction of the grain size to the sub-micron scale by implementing a 2-step milling procedure that uses a planetary ball mill for further grinding the material into a nanoscale powder after a freeze-drying step. The nanoparticulate powder, previously homogenized using a mixer mill to ensure uniform particle distribution, is pressed into pellets using a hydraulic press without the need of any binder. Jochum et al. demonstrated the potential of three novel “nano-pellets”: JCp-1-NP, JCt-1-NP (natural matrix source), and MACS-3NP (synthetic matrix source) showing a 2- to 3-fold higher uniformity in trace element distribution, leading to lower variability and higher accuracy in LA-ICP-MS spot analysis quantification.30 JCt-1-NP and SPLT-NP (speleothem, natural matrix) have been recently used as external standard for 1D-elemental analysis (line profiling) of speleothems with 50 μm laser beam spots19 as well as to enable accurate quantitative high spatial-resolution 2D-mapping of marine carbonates, using a laser spot size of ∼2 μm.31
Typically, LA-ICP-MS analysis of carbonate archives relies on 1D analysis (line scanning) across a specimen's cross-section.6,32 Such approach assumes homogeneous deposition of CaCO3 and the other elements within well-defined growth layers. However, recent research has shown that elemental distributions within biogenic carbonates are often more heterogeneous than previously assumed, influenced by both environmental and biological factors that introduce structural discontinuities and variations in crystallization.33 Such complexities can lead to misinterpretations, especially in areas lacking a clear chronological structure. In this context, two-dimensional (2D) elemental mapping offers a more comprehensive analytical approach, capturing the full spatial distribution of elements across the specimen. Although different 2D-mapping applications in carbonates using techniques like synchrotron X-ray fluorescence microscopy (SXFM) and wavelength-dispersive X-ray electron probe microanalysis (EPMA) have been employed for otolith analysis, they require long measurement times (often exceeding 8 hours for areas >4 mm2) and access to large-scale synchrotron facilities.34 EPMA provides high spatial resolution (<3 μm) but suffers from relatively poor detection limits (>10 μg g−1).33 Synchrotron- and lab-based micro-X-ray fluorescence analysis techniques have also aided in the 2D-characterization and speciation analysis of mollusc shell compositions.35–38 While the resolution of these techniques (typically 20–50 μm spot size) approaches that of LA-ICP-MS, quantification limits (especially of lab-based systems) tend to be higher (∼1 μg g−1). Additionally, the long attenuation length of X-rays (>100 μm for energies needed to excite and measure Sr) has a tendency to integrate results below the sample surface, which can lead to time-averaging in incremental carbonate archives which often contain curved growth increments in three-dimensional space (see, e.g., discussion in de Winter and Claeys, 2017).35
More recently, LA-ICP-TOF-MS has been applied to the study of biogenic structures, enabling fast multi-elemental 2D-mapping at high spatial resolution. For instance, otolith specimens were mapped using laser spot size of 4 μm at an acquisition rate of up to 25 pixels per s.28 Coral samples have also been analysed using similar setups, employing even smaller spot sizes (1–2 μm) and faster acquisition rates (up to 200 pixels per s).31 These advances demonstrate the capabilities of LA coupled to ICP-TOF-MS instrumentation, providing near-complete elemental coverage with high spatial resolution and high sample throughput in biogenic CaCO3 applications.
Building on recent developments, this work presents the development of a quantitative method that enables fine-scale multi-elemental 2D-mapping of microstructural features in biogenic carbonate samples. This study has evaluated a new set of commercially available nanoparticulate pressed pellets, produced adapting the protocol described by Garbe-Schönberg and Müller,30 for their micro-homogeneity and suitability as matrix-matched external calibration standards in high-resolution multi-elemental 2D-mapping. Unlike previous assessments using larger laser spot sizes (50–80 μm), we have tested their performance using significantly smaller beam diameters (5–20 μm) using a LA-unit equipped with a low-dispersion cell coupled to an ICP-TOF-MS system. The suitability of the nano-pellets was evaluated for their use as external, matrix-matched standards for quantitative mapping of two types of biogenic carbonates commonly studied in sclerochronology: otoliths and mollusc shells. We use Ca internal normalization to correct for variations in ablation yield, potential matrix effects, and instrumental drift.
LA unit | ICP-TOF-MS | ||
---|---|---|---|
Carrier gas flow, He (L min−1) | 1.05 | RF power (W) | 1550 |
Laser beam size (μm) | 20 × 20 (2D LA-mapping), 10 × 10, 5 × 5 | Sampling depth (mm) | 5.2 |
Fluence (J cm−2) | 4.0 | Neb. Flow, Ar (mL min−1) | 0.92 |
Repetition rate (Hz) | 50 (2D LA-mapping), 2 (SPR & ODG opt) | CCT-mode, H2 (mL min−1) | 2.0 |
Dosage | 1 (2D LA-mapping) | Notched masses (m/z) | 40 |
Scan speed (mm s−1) | 1.0 | Notch amplitude (V) | 1.0 |
Additionally, two types of biogenic carbonate samples were selected given their relevance in the context of microanalysis and the need for mapping at high spatial resolution. These samples serve as case studies to test the performance of the LA-ICP-TOF-MS setup on accretionary carbonate archives: (1) otoliths from two 6 year-old Atlantic cod fish (Gadus morhua), harvested from two locations in Norway—Store Lungegårdsvannet (Bergen) and Hardangerfjord. These specimens were provided by the Institute of Marine Research (IMR, Bergen, Norway). (2) Two different species of modern bivalve shells, the Pacific oyster (Crassostrea gigas) and the ocean quahog (Arctica islandica), were both harvested alive in the North Sea as part of previous studies43,44 and provided by the Vrije University of Amsterdam (VUA, The Netherlands).
For all specimens, the samples were first cleaned and disinfected and then embedded in epoxy resin. They were sectioned dorsoventrally along their axis of maximum growth in thin sections (tens of microns), these sections were polished and mounted onto microscope slides (Corning®, Merck 75 mm × 25 mm) with double-sided tape or adhesive to ensure proper fixation and prevent alteration during subsequent LA-ICP-TOF-MS analysis.
At the LA-side, it was evaluated whether for a laser beam size of 20 × 20 μm2 (square mask) different laser fluences of 2, 3, 4 and 5 J cm−2 were capable of effectively ablating the CaCO3 matrix. Using the NFHS-2-NP nano-pellet as a CaCO3 matrix reference, increasing the fluence from 2 to 4 J cm−2 resulted in improved sensitivity across all examined elements (by a factor of 1.5), including those present at low (<1 μg g−1; e.g., Co, Cr, Pr, U, Th), mid (1–20 μg g−1; e.g., Ce, Pb, Cu, Zn), and high concentrations (>20 μg g−1; e.g., Na, Mn, Mg, Al, Fe, Sr), as shown in Fig. 2. However, increasing the fluence beyond 4 J cm−2 to 5 J cm−2 resulted in a slight enhancement of <5% only, thus 4 J cm−2 was selected as the optimal value ensuring effective carbonate ablation.
Simultaneous 2D multi-element mapping of 3 × 1 mm2 regions of interest (ROIs) were performed for each nano-pellet (corresponding to 3.82% of the total nano-pellet surface) using LA-ICP-TOF-MS, generating a total of 7500 data points (i.e., pixels in the 2D-map), corresponding to 50 lines of 3 mm length per pellet (1 line = 150 data points). Visual inspection of the qualitative 44Ca normalized maps for selected target elements present in the nano-pellets reveals the degree of homogeneity. For instance, Fig. 3 allows visual comparison between NFHS-2-NP and BPLM-NP nano-pellets. While NFHS-2-NP shows a high degree of micro-homogeneity for all the target elements, even those prone to surface contamination such as Zn, in BPLM-NP elements such as Al, Mn, Fe, Cu, Zn and Pb show a less homogeneous distribution. This is also the case for other nano-pellets (Fig. S2), with BAM-RS3-NP exhibiting more heterogeneity compared to the more uniform distribution observed in CRMS-NP or ECRM-752-1-NP.
In addition, also the repeatability based on the relative standard deviation (RSD) of the average signal intensity for independent line scan measurements was evaluated (total lines, n = 50 & sub-sets n = 25, 10, 5 and 3) and compared to values obtained for NIST SRM 610 glass (RSDs <2%, Table S1). For example, while NFHS-2-NP exhibited RSD values between 1% and 7% across all elements (Table 2, top), for BPLM-NP, RSDs for elements such as Cu, Zn and Pb RSDs were as high as ∼60%, ∼80% and ∼47%, respectively (Table 2, bottom). When randomly selecting subsets of 25, 10, 5 and 3 lines (3750, 1500, 750 and 450 data points, respectively) from the full set of 50 (7500 data points), in all cases, the calculated RSDs as well as the average signal intensity remained similar to those calculated from the full dataset (see Table 2 for NFHS-2-NP and BPL M-NP, and Table S1 for an overview covering all target elements and the entire set of nano-pellets). Notably, due to the heterogeneous distribution of certain elements, the RSD could fluctuate significantly depending on the lines selected. For example, Pb in BPLM-NP showed an RSD of nearly 50% when calculated based on 10 lines, but this dropped to a value of 7% when only 3 lines were used. Similar observations were made for BAM-RS3-NP (see Table S1). This behaviour indicates high spatial heterogeneity in certain elemental distributions, as seen in the corresponding maps (Fig. 3 and S2). However, while the average signal intensities may vary depending on the number of lines (n) selected—up to 20% for Zn in BPLM—most elements showed a variation < 5% across all nano-pellets (Table S1), supporting the high micro-homogeneity of the nano-pellets.
NFHS-2-NP | 55Mn | 56Fe | 63Cu | 66Zn | 88Sr | 137Ba | 208Pb | 238U | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
n | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD |
50 | 2.14 × 102 | 1% | 6.83 × 101 | 2% | 6.58 × 100 | 4% | 6.68 × 100 | 7% | 6.26 × 103 | 1% | 1.11 × 102 | 2% | 2.15 × 101 | 2% | 2.35 × 100 | 8% |
25 | 2.13 × 102 | 1% | 6.82 × 101 | 2% | 6.66 × 100 | 4% | 6.55 × 100 | 4% | 6.26 × 103 | 1% | 1.12 × 102 | 2% | 2.14 × 101 | 2% | 2.35 × 100 | 9% |
10 | 2.14 × 102 | 1% | 6.86 × 101 | 2% | 6.72 × 100 | 5% | 6.55 × 100 | 5% | 6.28 × 103 | 1% | 1.12 × 102 | 2% | 2.14 × 101 | 2% | 2.41 × 100 | 13% |
5 | 2.15 × 102 | 1% | 6.90 × 101 | 1% | 6.64 × 100 | 7% | 6.64 × 100 | 7% | 6.28 × 103 | 1% | 1.13 × 102 | 3% | 2.12 × 101 | 3% | 2.36 × 100 | 7% |
3 | 2.15 × 102 | 1% | 6.83 × 101 | 2% | 6.75 × 100 | 6% | 6.78 × 100 | 5% | 6.28 × 103 | 1% | 1.13 × 102 | 3% | 2.15 × 101 | 3% | 2.37 × 100 | 8% |
BPLM-NP | 55Mn | 56Fe | 63Cu | 66Zn | 88Sr | 137Ba | 208Pb | 238U | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
n | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD | Mean (cts) | RSD |
50 | 7.94 × 100 | 12% | 1.12 × 101 | 10% | 2.65 × 100 | 48% | 2.72 × 100 | 84% | 8.75 × 103 | 1% | 8.09 × 100 | 4% | 9.00 × 100 | 27% | 2.62 × 100 | 7% |
25 | 7.95 × 100 | 14% | 1.16 × 101 | 11% | 3.12 × 100 | 51% | 3.34 × 100 | 77% | 8.77 × 103 | 1% | 8.19 × 100 | 4% | 9.57 × 100 | 33% | 2.59 × 100 | 7% |
10 | 8.19 × 100 | 21% | 1.12 × 101 | 8% | 3.35 × 100 | 61% | 4.36 × 100 | 85% | 8.76 × 103 | 1% | 8.26 × 100 | 3% | 1.03 × 101 | 47% | 2.62 × 100 | 9% |
5 | 7.46 × 100 | 6% | 1.06 × 101 | 4% | 2.28 × 100 | 8% | 3.35 × 100 | 83% | 8.80 × 103 | 2% | 8.15 × 100 | 2% | 8.27 × 100 | 2% | 2.63 × 100 | 6% |
3 | 8.42 × 100 | 31% | 1.07 × 101 | 5% | 2.25 × 100 | 8% | 2.25 × 100 | 8% | 8.70 × 103 | 1% | 8.14 × 100 | 2% | 8.58 × 100 | 7% | 2.55 × 100 | 3% |
Higher uncertainties tend to correlate with lower elemental concentrations, especially for elements present at or below the μg g−1 level. In Fig. 4, plotting the RSD% (based on 10-line measurements) against the element mass fraction (μg g−1) across all nano-pellets reveals a general trend: elements at higher concentrations exhibit lower RSDs. However, this trend is not consistent across all materials. In the case of NFHS-2-NP, RSDs remained below 5% for elements above 1 μg g−1 with an increase to 20% for concentrations below the μg g−1 level. However, in other nano-pellets, like CRMS-NP or Apatite-NP, elements around 5–100 μg g−1 displayed higher RSDs than some sub-μg g−1 elements, suggesting that concentration alone does not fully explain the variability. Overall, except for elements at sub-μg g−1 levels in BPLM-NP and BAM-RS3-NP (RSDs in the range of 50–100%), RSD values remained within acceptable limits—mostly below 30%, and in some pellets (e.g., NFHS-2-NP) under 10%. This result matches the RSDs found by Jochum et al.18 when evaluating the first ever nanoparticulate pellets (MACS-3NP, JCp-1NP and JCt-1-NP) using larger spot sizes (25 and 55 μm), who reported RSDs between 2–30% (n = 9 lines of 200 μm length.
Additionally, calibration curves were constructed based on the average signal intensities normalized to 44Ca+ obtained upon ablation of 50- and 3-line scans for all 7 nano-pellets and the slope and linearity were compared accordingly (Fig. 5). As an example, the calibration curves for four target elements—Mn, Zn, Sr, and Ba—are shown. Overall, for most elements, no significant differences in linearity (coefficient of determination, r2) or sensitivity (slope) were observed between the two approaches. However, for Zn, a deviation was detected: the slope of the calibration curve increased from 14.12 to 15.7 when using n = 3 lines instead of n = 50, representing an ∼11.2% increase. This variation can be attributed to the lower Zn concentrations in certain nano-pellets and the higher micro-scale heterogeneity observed. Despite this, a change of approximately 11% in the slope remains within acceptable limits for many applications, depending on the required level of accuracy and the overall uncertainty tolerance of the method.
Table 3 presents the results obtained for CRMS-NP (top) and NFHS-2-NP (bottom) using a 20 × 20 μm2 square laser beam. Decreasing the laser beam size reduces the ablation yield and thus, the relative sensitivity. As a result, signal intensities are lower, and the uncertainty becomes higher, both negatively affecting the accuracy when using a non-matrix-matched approach. In particular, low recoveries were observed for light elements such as Na, Al, and Zn, with biases ranging between 50% and 70% for both nano-pellets (Table 3, top and bottom). Significant deviations from the reference values were also found for low-concentration elements. For example, for Pr, Th, and U in CRMS-NP, and La, Ce, Pr, Pb, Th, and U in NFHS-2-NP, the results showed 40–50% bias when compared to the reference value. In contrast, with matrix-matched external calibration—particularly when applying Ca as an IS—the accuracy improved significantly (see column 4 and 5 in Table 3). Most data obtained using this method showed a bias <12%, with only a few exceptions (mainly sub-ppm elements, present at a concentration close to the corresponding LOQ) showing a bias up to 20%. These values remain within acceptable limits for trace element analysis, demonstrating the effectiveness of the matrix-matched strategy for reliable quantification.
CRMS-NP | Reference value | Against NIST SRM 610/Ca (IS) | Against nano-pellets CAL | Against nano-pellets CAL/Ca (IS) | Against NFHS-2-NP/Ca (IS) | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Element | Conc. (μg g−1) | sd | Conc. (μg g−1) | sd | RSD % | Recovery % | Conc. (μg g−1) | sd | RSD % | Recovery % | Conc. (μg g−1) | sd | RSD % | Recovery % | Conc. (μg g−1) | sd | RSD % | Recovery % | LOD (μg g−1) | LOQ (μg g−1) |
Na | 4270 | 160 | 1398 | 16 | 1% | 33% | 55![]() |
63 | 1% | 131% | 4640 | 52 | 1% | 109% | 4920 | 230 | 5% | 115% | 55 | 180 |
Mg | 23![]() |
690 | 21![]() |
200 | 1% | 90% | 25![]() |
230 | 1% | 105% | 23![]() |
220 | 1% | 100% | 26![]() |
920 | 3% | 111% | 1.7 | 5.8 |
Al | 1207 | 38 | 441 | 20 | 5% | 37% | 2550 | 120 | 5% | 211% | 1140 | 52 | 5% | 94% | 1080 | 140 | 13% | 89% | 2.9 | 9.8 |
P | 521 | 26 | 673 | 16 | 2% | 129% | 580 | 13 | 2% | 111% | 590 | 14 | 2% | 112% | 550 | 13 | 2% | 106% | 100 | 340 |
Mn | 254 | 17 | 247 | 13 | 5% | 97% | 415 | 21 | 5% | 163% | 240 | 13 | 5% | 96% | 270 | 14 | 5% | 108% | 0.20 | 0.65 |
Fe | 2479 | 88 | 2076 | 56 | 3% | 84% | 3368 | 91 | 3% | 136% | 2340 | 63 | 3% | 94% | 2150 | 160 | 8% | 87% | 0.22 | 0.72 |
Cu | 86.8 | 5.7 | 92.6 | 8.2 | 9% | 107% | 88.4 | 7.8 | 9% | 102% | 87.0 | 7.7 | 9% | 100% | 109 | 11 | 10% | 126% | 0.19 | 0.63 |
Zn | 42.9 | 3.2 | 34.7 | 7.5 | 21% | 81% | 50.7 | 13.6 | 27% | 118% | 44.2 | 5.9 | 13% | 103% | 41.0 | 6.7 | 16% | 96% | 0.15 | 0.52 |
Sr | 2663 | 59 | 2618 | 31 | 1% | 98% | 3620 | 220 | 6% | 136% | 2770 | 120 | 4% | 104% | 2770 | 110 | 4% | 104% | 0.021 | 0.071 |
Cd | 0.30 | * | 0.335 | 0.044 | 13% | 112% | 0.568 | 0.074 | 13% | 189% | 0.35 | 0.10 | 29% | 115% | 0.31 | 0.10 | 32% | 102% | 0.28 | 0.92 |
Ba | 38.5 | 2.3 | 34.6 | 4.4 | 13% | 90% | 60.4 | 7.8 | 13% | 157% | 39.7 | 5.1 | 13% | 103% | 38.5 | 2.3 | 6% | 100% | 0.16 | 0.53 |
La | 6.62 | 1.40 | 5.21 | 1.72 | 33% | 79% | 11.7 | 3.2 | 27% | 176% | 7.1 | 2.1 | 30% | 107% | 6.8 | 2.3 | 34% | 103% | 0.0070 | 0.023 |
Ce | 19.1 | 5.0 | 15.7 | 3.6 | 23% | 82% | 29.0 | 6.7 | 23% | 152% | 17.6 | 2.0 | 12% | 92% | 20.3 | 1.9 | 9% | 106% | 0.0046 | 0.015 |
Pr | 0.281 | 0.041 | 0.231 | 0.043 | 19% | 82% | 0.506 | 0.094 | 19% | 180% | 0.310 | 0.057 | 19% | 110% | 0.310 | 0.069 | 22% | 110% | 0.0054 | 0.018 |
Pb | 391 | * | 332.5 | 4.9 | 1% | 85% | 391.2 | 6.2 | 2% | 100% | 391.0 | 6.2 | 2% | 100% | 397.2 | 2.6 | 1% | 102% | 0.032 | 0.11 |
Th | 0.1700 | 0.03 | 0.112 | 0.012 | 11% | 66% | 0.327 | 0.035 | 11% | 192% | 0.200 | 0.021 | 11% | 117% | 0.1648 | 0.0012 | 1% | 97% | 0.010 | 0.034 |
U | 0.4160 | 0.037 | 0.2861 | 0.00051 | <1% | 69% | 0.6119 | 0.0011 | <1% | 147% | 0.37190 | 0.00067 | <1% | 89% | 0.523 | 0.012 | 2% | 126% | 0.0090 | 0.030 |
NFHS-2-NP | Certified value | Against NIST SRM 610/Ca (IS) | Against nano-pellets CAL | Against nano-pellets CAL/Ca (IS) | Against CRMS-NP | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Element | Conc. (μg g−1) | sd | Conc. (μg g−1) | sd | RSD % | Recovery % | Conc. (μg g−1) | sd | RSD % | Recovery % | Conc. (μg g−1) | sd | RSD % | Recovery % | Conc. (μg g−1) | sd | RSD % | Recovery % | LOD (μg g−1) | LOQ (μg g−1) |
Na | 1195 | 52 | 340 | 32 | 9% | 28% | 880 | 47 | 5% | 74% | 1140 | 36 | 3% | 95% | 1037 | 35 | 3% | 87% | 55 | 180 |
Mg | 637 | 28 | 520 | 16 | 3% | 82% | 387.0 | 5.8 | 2% | 61% | 574.0 | 8.7 | 2% | 90% | 575 | 11 | 2% | 90% | 1.7 | 5.8 |
Al | 641 | 45 | 260 | 26 | 10% | 41% | 940 | 20 | 2% | 147% | 660 | 14 | 2% | 102% | 718 | 24 | 3% | 112% | 2.9 | 9.8 |
P | 91.0 | 8.0 | <LOQ | * | * | * | <LOQ | * | * | * | <LOQ | * | * | * | <LOQ | * | * | * | 100 | 340 |
Mn | 87.00 | 0.35 | 79.8 | 1.1 | 1% | 92% | 87.0 | 1.2 | 1% | 100% | 80.0 | 1.1 | 1% | 91% | 80.8 | 1.8 | 2% | 93% | 0.20 | 0.65 |
Fe | 532 | 36 | 513 | 11 | 2% | 96% | 530 | 15 | 3% | 100% | 580 | 16 | 3% | 108% | 612 | 25 | 4% | 115% | 0.22 | 0.72 |
Cu | 6.40 | 0.30 | 6.10 | 0.25 | 4% | 95% | 3.80 | 0.25 | 6% | 60% | 5.80 | 0.38 | 6% | 91% | 5.09 | 0.13 | 3% | 80% | 0.19 | 0.63 |
Zn | 15.9 | 1.5 | 15.0 | 1.1 | 8% | 93% | 13.80 | 0.72 | 5% | 87% | 18.62 | 0.98 | 5% | 117% | 16.63 | 0.87 | 5% | 105% | 0.15 | 0.52 |
Sr | 1190 | 45 | 1120 | 10 | 1% | 94% | 1000 | 12 | 1% | 84% | 1186 | 14 | 1% | 100% | 1147 | 11 | 1% | 96% | 0.021 | 0.071 |
Cd | 0.1630 | 0.0090 | <LOD | * | * | * | <LOD | * | * | * | <LOD | * | * | * | <LOD | * | * | * | 0.28 | 0.92 |
Ba | 76.0 | 1.0 | 65.0 | 1.1 | 2% | 85% | 74.0 | 2.5 | 3% | 97% | 75.3 | 2.6 | 3% | 99% | 76.0 | 1.3 | 2% | 100% | 0.16 | 0.53 |
La | 2.43 | 0.11 | 1.860 | 0.052 | 3% | 77% | 2.710 | 0.076 | 3% | 111% | 2.554 | 0.072 | 3% | 105% | 2.300 | 0.040 | 2% | 95% | 0.0070 | 0.023 |
Ce | 1.50 | 0.10 | 1.160 | 0.073 | 6% | 77% | 1.390 | 0.087 | 6% | 93% | 1.310 | 0.082 | 6% | 87% | 1.410 | 0.090 | 6% | 94% | 0.0046 | 0.015 |
Pr | 0.460 | 0.040 | 0.340 | 0.014 | 4% | 73% | 0.480 | 0.014 | 3% | 104% | 0.452 | 0.013 | 3% | 98% | 0.410 | 0.050 | 12% | 89% | 0.0054 | 0.018 |
Pb | 1.650 | 0.050 | 1.370 | 0.030 | 2% | 83% | 1.030 | 0.027 | 3% | 63% | 1.600 | 0.042 | 3% | 97% | 1.624 | 0.030 | 2% | 98% | 0.032 | 0.11 |
Th | 0.140 | 0.010 | 0.100 | 0.015 | 16% | 69% | 0.170 | 0.013 | 8% | 123% | 0.163 | 0.012 | 8% | 116% | 0.144 | 0.020 | 14% | 103% | 0.010 | 0.034 |
U | 0.120 | 0.030 | 0.0660 | 0.0050 | 8% | 55% | 0.0910 | 0.0074 | 8% | 76% | 0.0855 | 0.0070 | 8% | 71% | 0.10 | 0.012 | 13% | 79% | 0.0090 | 0.030 |
It is important to emphasize the importance of using Ca as an IS in this type of application—not only when the matrix of the calibration material differs from that of the sample, but also when using nano-pellets composed of similar yet non-identical carbonate matrices for calibration. As shown in Table 3, recoveries were significantly improved when Ca was used as an IS in external calibration against nano-pellets. This is illustrated by the calibration curves (Fig. S3), for which omission of Ca as IS resulted in poor linearity. For example, R2 values of 0.5274 (Al), 0.7803 (Sr), and 0.8642 (Mn) improved to 0.9969, 0.9965, and 0.9990, respectively, when Ca was applied as an IS. These results highlight the variability in ablation behaviour even among carbonate-based materials, particularly in the case of Apatite, which has a distinct Ca5(PO4)3(F, Cl, OH) crystal structure, which deviates significantly from that of typical carbonate matrices. Using an IS significantly improved the linearity of all calibration curves, resulting in an average R2 value of 0.9958 ± 0.0048 across all target elements. Also, small variations in laser beam focus during ablation could negatively affect the linearity which can be overcome when using internal standardization.
The recoveries obtained for the target elements in the remaining nano-pellets (Apatite-NP, SPLT-NP, BAM-RS3-NP, ECRM-752-1-NP and BPLM-NP) are shown in the SI (Table S2). The results highlight a clear relationship between the micro-homogeneity of nano-pellets and the accuracy of elemental quantification. Nano-pellets with poorer homogeneity—evidenced by a more heterogenous elemental distribution as also evidenced by higher relative standard deviations (RSDs) on the signal intensities/element concentration values as discussed in the previous section—tend to yield less favourable recoveries for the elements affected. This is particularly evident in the case of BAM-RS3 and BPLM-NPs. For example, in BPLM-NP, the recovery for Th was <30%, likely due to its sub-ppm concentration close to the LOQ combined with the pellet's higher heterogeneity. For BAM-RS3-NP, poor accuracies were obtained for most of the elements (e.g., Fe ∼30%, Mn ∼138% and Zn ∼229%). A similar observation was made for the ECRM-752-1 nano-pellet: while the nano-pellet-based calibration improved overall accuracy, limitations remain for specific elements like Zn, (recovery of ∼190%). In other cases, like for Apatite-NP, quantification using the non-matrix-matched approach (NIST SRM 610 glass) was accurate for certain elements, such as Sr, Ba, La or Ce, only, all showing less than 10% bias. However, for the majority of elements, recoveries were highly inaccurate (Table S2). In contrast, the matrix-matched approach using nano-pellets significantly improved performance, yielding recoveries between 86% and 119% across all target elements. Overall, elements present at sub-μg g−1 concentration (especially those close to the LOQ), as well as elements prone to contamination such as Zn show poorer accuracies even when using a matrix-matched calibration and Ca normalization.
In a next phase, the comparison between the outcomes using the full set of nano-pellets for calibration and using a simplified one-point calibration approach, were compared. In this case, only NFHS-2-NP and CRMS-NP, selected based on their superior micro-homogeneity and covering the majority of target elements (above their LOD), were used as one-point calibration standards. For all target elements (except for Cu and U with uncertainties reaching 30%) the average offset from the reference value was below 20% for CRMS-NP, NFHS-NP and SPLT-NP, highlighting the potential of one-point calibration using well-characterized, homogeneous standards (Fig. 6A–C, respectively). However, a trend towards higher uncertainties was observed when using one-point calibration compared to the full calibration curve for the entire set of nano-pellets (Fig. 6). Particularly for Apatite-NP, offsets for most elements increased from 10% (dark green area) to nearly 30%, and for ECRM-752-1-NP, deviations from the reference values reached up to 50%. These results suggest that the main limitations in calibration approaches arise from the heterogeneity of certain standards (e.g., ECRM, BPLM) and from mismatched concentration ranges between standards and samples, rather than from the calibration model itself. Thus, while one-point calibration with a homogeneous, well-characterized standard can yield accurate results under suitable conditions, multi-point calibration remains advantageous for maintaining robustness across a broader range of analytes, sample types and compensating for potential heterogeneities in the standard materials.
Although laser spot sizes of 5 × 5 and 10 × 10 μm2 (square masks) were tested and yielded good linearity for major elements such as Ba, Mn, Sr, and Pb, the reduced sensitivity at smaller spot sizes led to higher limits of detection (LODs) and quantification (LOQs) in the μg g−1-range (Table S2), hindering the monitoring of trace elements (sub- μg g−1 levels), especially in the low-mass range (due to the inherent lower sensitivity exhibited by TOF-based instrumentation). Nevertheless, Fig. 7 shows that for major and minor elements (above μg g−1 levels) the bias from the reference value remained <10% (darker green) for most of the elements (i.e., Na, Mg, Al, Mn, Fe, Sr and Ba), or <20% (light green) for some (i.e., Cu and Pb) even when decreasing the spot size to 5 × 5 μm2. These results confirm that while smaller spot sizes reduce the overall sensitivity, especially critical when working with ICP-TOF-MS instrumentation, the micro-homogeneity of the nano-pellets is sufficient to provide reliable and accurate data for minor (particularly in the high-mass range) and major elements, though a full elemental overview is restricted under such conditions (e.g., for Th or U, which were present at sub-ppm levels, the recoveries were less than 80% when using a 10 × 10 or 5 × 5 μm2 laser beam).
The data processing workflow for generating quantitative elemental maps involved: (1) selection of the qualitative map of the nuclide of interest (e.g., 88Sr+) and of the internal standard (44Ca+) (Fig. 8A and B, respectively); (2) normalization of the target element's signal intensity at each pixel to that for 44Ca+ to account for differences in ablation yield; (3) conversion of the normalized qualitative map into a quantitative one using matrix-matched external calibration against the nano-pellets (n = 3 lines per nano-pellet) and assuming a concentration of Ca of 40%.
Sr is typically used as a proxy for fish age. The distribution of Sr in the otolith displays concentric bands interpreted as chronological markers, as can be seen in the quantitative Sr map in Fig. 8C. However, it is also studied as a proxy for reconstructing environmental histories such as habitat shifts and waterbody changes.48,49 In fact, Sr incorporation in otoliths reflects the ambient water chemistry—mainly salinity and temperature variations.50,51 While these patterns may capture seasonal or episodic events, high lateral resolution mapping can reveal more Sr bands than actual annual growth increments. In this case, 15–17 Sr bands were identified in an otolith originating from a 6 year old fish, highlighting the enhanced time resolution offered by spatially resolved LA-ICP-TOF-MS mapping.2 When examining the Mn distribution (Fig. S4A), it is still possible to distinguish an enrichment in the core regions corresponding to juvenile life (Mn concentration close to the LOQ of the method). Elevated Mn/Ca ratios were described to be related with hypoxic conditions (i.e., reduced oxygen levels) typically in estuaries or sediment-rich areas.51–54 The combination of high Mn concentration with low Sr levels in the rings near to the core suggests a juvenile phase where the animal was exposed to low oxygen (high concentration of Mn) and low salinity (low concentration rings of Sr) water environments like hypoxic bottom waters or estuaries. Peaks in Mn can also indicate metabolic activity or stress during key life events (e.g., hatching, settlement).48 The findings were compared to those for a second otolith from a cod sampled in the same waterbody (Lungegårdsvannet) that shows the same patterns: highly defined layers are observed based on the Sr distribution along the entire life of the fish, while Mn layers (higher Mn exposure) overlapping with low Sr concentration layers are only seen for early-life periods (Fig. S4B and C, respectively).
Notably, the diagonal lines—likely polishing artifacts—observed in the elemental maps of 88Sr+ (Fig. 8A) and 44Ca+ (Fig. 8B) are no longer visible after normalization (Fig. 8C). This demonstrates the effectiveness of IS normalization in correcting not only for matrix-related ablation yield variations (e.g., due to crystallization or porosity) or instrumental drifts, but also for surface artifacts introduced during sample preparation.
For the other otolith from the cod caught in the Hardanger fjord in Norway, the Mn distribution was not revealed, but in the juvenile lines, an enrichment in Ba (Fig. 9B) coinciding with low-concentration rings of Sr (Fig. 9A) was observed. Ba is typically used as marker of specific habitat transitions.50 While a high concentration of Sr is related to high salinity, a high concentration of Ba corresponds to low salinity environments (i.e., fresh waters).48 The co-localization of a low Sr and a high Ba concentration in the early growth rings suggests a juvenile phase where the cod inhabited shallower or more brackish environments (low salinity environments). A later migration to deeper, more saline waters (no Ba exposure) for foraging could explain the corresponding changes in element ratios recorded in the otolith. Taken together, these data underscore the value of quasi-simultaneous quantitative multi-elemental mapping in enhancing the interpretation of otolith chemistry covering the fish's lifespan. While strontium (Sr) alone may not reliably indicate chronological age, integrating multiple elemental ring distributions (i.e. Mn or Ba) enables a more nuanced reconstruction of the fish's environmental history. Noted that, fully interpreting the readings of each otolith and tracking the corresponding fish's life history requires extensive analysis, which is beyond the scope of this work.
Traditional 1D-analysis (line scanning) was compared to 2D-elemental mapping in all cases to emphasize the advantages of 2D-mapping in sclerochronological studies. Although a well-chosen 1D transect (e.g., core-to-edge along the longest growth axis) can capture chronological information, it may not always represent the full spatial variability of elemental incorporation. Recent work by de Pontual et al.55 demonstrated that otoliths can exhibit subtle asymmetries and heterogeneous distributions of certain elements, underscoring the limitations of relying solely on single transects. Our results are consistent with this view. As shown in Fig. 9B and C, the layer patterns for Sr and Ba differ depending on the direction of the line scan—whether from the core to the left (Fig. 9B), or from the core to the top (Fig. 9C) of the otolith. Notably, scans from the centre (juvenile layers) to the top show a loss of resolution, failing to capture the fine scale elemental distribution. In addition, as can be observed from the 2D-map (Fig. 9A and C), the layer arrangement on the left & right sides is not perfectly symmetrical. Consequently, 1D analysis can yield different patterns depending on scan orientation, potentially leading to misinterpretation of elemental (Fig. S5). By contrast, 2D-mapping provides a comprehensive view of the otolith structure, allowing potential spatial heterogeneities to be identified and interpreted more reliably.
To validate the quantitative results obtained via LA-ICP-TOF-MS mapping, a complementary bulk analysis was performed. From each fish, two otoliths were collected. One was embedded in epoxy resin for direct solid sampling by laser ablation, while the second was acid-digested and used for validation through conventional solution ICP-MS, as described in the Experimental Section of the SI. The results were in the same range as the values obtained by solid sampling for Sr, Ba and Mn (Table S4).
Notably, the ability to visualize strontium (Sr) growth lines within the shells allows for sub-annual, even monthly, climate reconstruction, highlighting the method's value in paleoclimate studies.26,60 In this context, the Sr distribution shows rings with higher and lower concentration than can used as proxy for environmental conditions. The observation that these Sr concentration bands are continuous throughout the oyster hinge region (Fig. 10A) is of particular interest. The shell of C. gigas consists of alternating layers of the foliated calcite microstructure interspersed with lenses of the more porous chalky calcite microstructure, which often exhibit marked variations in chemical composition, complicating the observation of isochronous growth structures in the shells.44,56 In the A. Islandica shells, localized enrichments of barium (Ba) were observed in specific growth bands both in the outer and inner shell layers of the ventral margin (Fig. 10C) and in corresponding growth bands in the shell hinge (Fig. 10E). These likely correspond to seasonal phytoplankton bloom events in spring and/or autumn, in accordance with observations in the literature.44 Clear seasonal banding in Sr/Ca observed in A. islandica (Fig. 10C and E) corroborates the timing of these Ba/Ca peaks in this specimen. An important observation in these mollusc maps is that the concentration of Sr relative to Ca is not uniform in contemporary portions of the shell (see Fig. 10A for C. gigas and Fig. 10B for A. islandica). This corroborates previous studies by Freitas et al.61 who demonstrated such heterogeneity and like for the otoliths, this highlights the value of 2D-mapping for the interpretation of trace element profiles in mollusc shell material in terms of climate and environmental change.
The micro-homogeneity assessments showed differences among set of the nano-pellets studied. While NFHS-2-NP, CRMS-NP, and Apatite-NP nano-pellets exhibited the highest micro-homogeneity for nearly all elements, higher RSDs and consequently poorer elemental recoveries were found for BAM-RS3-NP and BPLM-NP. It was noted that sub-μg g−1 elemental concentrations in the nano-pellets yielded higher RSDs. Additionally, when decreasing laser beam sizes to reach very high lateral resolution (down to 5 × 5 μm2), LODs and LOQs deteriorated, hampering detection and quantification of trace elements at concentrations below sub-μg g−1, as well as increasing uncertainties for elements close to the LOD. This reduction in sensitivity associated with the use of a smaller laser beam spot and the inherently lower sensitivity of ICP-TOF-MS instrumentation compared to quadrupole-based systems jeopardize the detection of some minor and trace elements (often in the low μg g−1 or even sub-μg g−1 range). Nevertheless, the loss of sensitivity at ultrahigh lateral resolution could in principle be compensated for by increasing both the number of laser shots per pixel (i.e., using a dosage >1) and the repetition rate accordingly. While this approach could boost signal intensity, it also exponentially increases the number of required laser pulses and introduces potential challenges with detector saturation. Therefore, dosage >1 approaches may be better suited for specific regions of interest (ROIs) where very high spatial detail is required.
Overall, this quantitative multi-elemental LA-ICP-TOF-MS 2D elemental mapping approach was successfully applied to two types of biogenic carbonate samples: otoliths and bivalve shells. This methodology enabled the quantitative visualization of growth bands and internal structures, allowing distinction of seasonal and sub-seasonal growth patterns. By combining Sr and Mn maps or Sr and Ba maps, the 2D reconstructions provide additional spatial context beyond that obtained from 1D transects, supporting a more robust interpretation of carbonate archives for tracking historical events within an animal's lifespan. While Sr reflects seasonal-scale environmental variations in both archives, Mn and Ba serve as sensitive indicators of specific habitat conditions such as water oxygenation and phytoplankton blooms. Together, these tracers underline the utility of biogenic samples such as otoliths and bivalve shells as chronological environmental archives. Furthermore, comparison of 2D-maps with traditional 1D (line scanning) analyses revealed that, while 1D transects along the main growth axis can capture temporal Sr and Ba patterns, they may not consistently reflect finer-scale heterogeneities or asymmetries that occur orthogonal to the line scanning direction. This highlights the need for caution when interpreting 1D profiles and when assuming homogeneity of elemental patterns in isochronous parts of biominerals, since non-symmetrical layer structures and chemical variability within contemporary growth layers can occur. In contrast, 2D-mapping provides a more comprehensive and spatially representative view, which is essential for robust sclerochronological and environmental reconstructions.
Supplementary information is available. See DOI: https://doi.org/10.1039/d5ja00280j.