Open Access Article
This Open Access Article is licensed under a
Creative Commons Attribution 3.0 Unported Licence

Non-linearity correction for variable signal analysis in mass spectrometry using discrete ion counters

Stefaan Pommé*
European Commission, Joint Research Centre (JRC), Geel, Belgium. E-mail: stefaan.pomme@ec.europa.eu; Tel: +32 (0)14 571 289

Received 20th June 2025 , Accepted 15th July 2025

First published on 18th September 2025


Abstract

The response of a discrete ion counter is not perfectly linear due to count loss caused by dead time and pulse pileup. As a result, the output rate of the counter does not scale linearly with the input rate of ions reaching the detector. The value of a stable input rate can be determined from the measured output rate by inverting the throughput formula of the ion counter. However, when the input rate varies during the measurement, a mismatch between the average input rate and the average output rate becomes apparent. The resulting bias can be particularly significant when measuring transient signals. A correction procedure is proposed to calculate a better estimate of the average input rate from the observed mean and variance of the output rate. Implementation of this refined throughput formula is recommended to improve accuracy of mass spectrometry utilising discrete ion counters.


1 Introduction

In mass spectrometry,1 the choice between using Faraday cups or discrete ion counters (such as electron multipliers, microchannel plates, or Daly detectors) for ion detection depends on various factors, including the abundance ratios of the isotopes being measured, the sensitivity required, and the dynamic range of the measurements. Faraday cups are typically used for measuring high abundance ratios, where the signal intensity is relatively high, and the signal-to-noise ratio is less critical. Faraday cups provide a direct measurement of the total ion current passing through the detector without the need for signal amplification. On the other hand, discrete ion counters are often employed for measuring low abundance ratios, where the signal intensity is low, and high sensitivity and amplification are required for accurate ion detection. These detectors have a fast response time and are advantageous in processing a wide dynamic range of signal intensities.

The linearity of the counting process in a discrete ion counter is affected by count loss due to “pulse pileup” and “non-extending dead time”.2–4 It is common practice to impose a factory-set non-extending dead time on each counted event, which exceeds the width of the electronic pulses generated by the ion detector.5–9 This dead-time setting helps suppress the effect of pulse pileup across a wide input range. However, at extremely high input rates, pulse pileup will emerge as a notable contributor to count loss.10–14 This process can be accurately predicted by an advanced EDT–NEDT throughput model that reflects the interplay between extending (EDT) and non-extending dead time (NEDT).2–4 Pommé and Boulyga recently published an overview of relevant formulas for the throughput curve of discrete ion counters, the inverted throughput function to correct for count loss, the resulting counting uncertainty, the error propagation of the characteristic dead time and pulse width, and the error made by incomplete count-loss correction when ignoring pileup.2 This approach enables extending the linearity of the ion counter across the full range of input rates, thus mitigating rate-related bias in mass spectrometry.

Existing literature provides explicit evidence that current practice in mass spectrometry encounters limitations in dynamic range of discrete ion counters at high count rates.10 Already in 2002, Nomizu et al.11 pointed to the limited linearity of the channel electron multiplier in pulse-counting mode when observing non-linear response for the analysis of airborne Zn particles. Olesik and Gray12 considered various explanations for a limited dynamic range in analysing microparticles by single-particle inductively coupled plasma mass spectrometry (spICP-MS), including incomplete ionisation of particles, local plasma cooling, and competing rates of particle vaporisation and diffusion of ions. However, they also emphasised the importance of the “dead time” near the peak of transient signals. Strenge and Engelhard13 observed nonlinear response in pulse counting mode in spICP-MS and investigated the suitability of the traditional dead-time correction method (DTC) in millisecond and microsecond time-resolved spICP-MS. By applying DTC in each time channel, they managed to extend the linear dynamic range a bit further. Lomax-Vogt et al.14 have identified pulse pileup as a limiting factor for the maximum intensity that can be accurately measured in pulse counting mode by single-particle ICRP-QMS (typically 2 million counts per second, or about 200 counts per 0.1 ms). This limits the acceptable size of nanoparticles that can be analysed, which can be extended marginally if classical DTC is applied. In a forthcoming paper, Siegmund et al.15 report on a linearity test of secondary electron multipliers (SEM) subjected to a 233U/235U beam at input fluxes from ca. 104 to about 3 × 107 ions per second. The results confirm that the diminishing response of the ion detection system at ion fluxes above 2 × 106 s−1 is compatible with the theoretical throughput formula published by Pommé and Boulyga.2

The EDT-NEDT throughput model successfully solves the non-linearity problem for measurements of a “stationary Poisson process”, where the input rate ρ of events (ions impacting the sensitive area of the ion counter) remains constant during the measurement. However, another non-linearity effect lurks around the corner when the input rate undergoes substantial changes during the measurement. Since the output rate does not vary linearly with the input rate, a conventional dead time correction over the average output rate will not result in the corresponding average input rate.16–19 This may cause a significant bias in the measurement result, particularly for transient signals. This paper explains the mechanism causing the non-linearity phenomenon and presents a correction method that reduces the error substantially.

To establish a clear understanding of this secondary non-linearity effect, the mathematical derivations are conducted using simple scenarios involving linear or exponential changes in beam intensity over time. These conditions are not limited to slowly fluctuating count rates over extended measurement durations but also encompass short-term intervals, where e.g. transient signals are sampled through time-resolved mass spectrometry at ultra-short time scales.20–26 In the latter case, the perceived count rate will depend on the dwell time.26 The significance of the non-linearity bias will be examined through numerical examples across a range of count rates and varying levels of input rate variance. While the correction formula developed in this work is broadly applicable to signal shapes of various types, including transient signals conforming approximately to normal or log-normal distributions over time,27,28 the exploration of more intricate mathematical models is best suited for a subsequent publication.

2 Throughput at fixed rate

Radionuclide metrologists29,30 have extensively researched the impact of pulse pileup, extending dead time (EDT), and non-extending dead time (NEDT) on the throughput of nuclear counters with the aim of improving the accuracy of radioactivity measurements.2–4,31–53 The resulting theories are also relevant to ion counting in mass spectrometry. A key prerequisite is that the input process conforms to a “stationary Poisson process”, implying a constant input event rate, ρ, of ions impacting the sensitive area of the ion counter, with ions arriving randomly in time following exponentially distributed time intervals.

In current practice, the primary source of dead time in mass spectrometers is the factory-set NEDT of duration τne imposed on observed events from the pulse amplifier. This dead time is triggered when the leading edge of a pulse surpasses the discriminator threshold. Any subsequent pulse arriving within the NEDT window is not counted, and prolongation of the dead-time period through pulse pileup is ignored (see Fig. 1). The throughput formula for a purely NEDT-based counter is:4,31

 
image file: d5ja00243e-t1.tif(1)
in which X(ρ) represents the rate-dependent throughput factor, ρ is the input rate, and τne is the characteristic dead time.


image file: d5ja00243e-f1.tif
Fig. 1 (Top) Simplified model commonly used to describe count loss in ion counters. NEDT is imposed on counted events (green dot) and other events falling in the NEDT period are ignored (red dot). (Bottom) More realistic EDT + NEDT model accounting for finite pulse width. A pulse is identified when it rises above a threshold level. Effective dead time can be extended through pileup of pulses, since the sum pulse is identified as a single event.

When observing a count rate R, the throughput formula must be inverted to calculate a best estimate [small rho, Greek, circumflex] of the original input rate. The following inverse throughput formula is commonly used in mass spectrometry to compensate for count loss attributed to NEDT:4–9,31

 
image file: d5ja00243e-t2.tif(2)
in which X−1(R) represents the inverse throughput factor for NEDT.

The formulas in eqn (1) and (2) are valid only by approximation, mainly in a region where the input rate is not excessively high. In reality, the pulses of the detector have a finite width, and the overlap of neighbouring pulses results in additional count loss. This count loss mechanism can be modelled as a Poisson process passing through EDT and NEDT in series2–4,31–33,43 (see Fig. 1). Mathematical formulas are available for the time-interval distribution of counted events, the expected throughput rate, and the counting uncertainty for such system.2–4,31 The appropriate throughput formula applicable to ion counters in mass spectrometry is:

 
image file: d5ja00243e-t3.tif(3)

The second version of the formula generally applies, because the NEDT is purposely chosen by the manufacturer to exceed the pulse width, i.e. τne > τe.

Inversion of eqn (3) is less straightforward than the case of NEDT (eqn (2)), and has in principle two solutions, one for ρτe ≤ 1 and another for ρτe ≥ 1. The solution at moderate input rates, ρ< < 1/τe, can be obtained in a few iterations through the following recursion formula:2–4

 
image file: d5ja00243e-t4.tif(4)

Alternatively, even faster iterations can be obtained from the Newton–Raphson method:2

 
image file: d5ja00243e-t5.tif(5)
where z = e[small rho, Greek, circumflex]τe + [small rho, Greek, circumflex](τneτe) and eqn (2) can be used as a starting point. The second solution, ρ> > 1/τe, requires another formula for the starting value.2

For a stationary Poisson process, there is a unique relationship between the (moderate) input rate and the output rate. One can be converted into the other through R = ρX(ρ) and ρ = RX−1(R) without ambiguity, provided that the dead-time parameters τne and τe are accurately known.

3 Throughput at variable rate

In actual mass spectrometry measurements, the count rate may fluctuate slightly, at the percent level, or by more significant amounts. When fluctuations are minor, inverting the average output rate, [small rho, Greek, circumflex] = [R with combining macron]X−1([R with combining macron]), can still provide a sufficiently accurate estimate of the average input rate, [small rho, Greek, macron]. This holds particularly true at low count rates, where the throughput function is quasi-linear, ensuring that [small rho, Greek, macron] and [R with combining macron] vary proportionally. However, a substantial deviation from this ideal scenario can be anticipated at high input rates and in the presence of significant fluctuations. Such conditions are common in, for example, analysis of single nanoparticles or ablated aerosol particles in inductively-coupled plasma mass spectrometry (ICPMS).11–14,27,28 Where the throughput curve is bent, there will be a mismatch between the average input and output rates, [R with combining macron]X−1([R with combining macron]) ≠ [small rho, Greek, macron] and [small rho, Greek, macron]X([small rho, Greek, macron]) ≠ [R with combining macron], as demonstrated for an extreme case in Fig. 2.
image file: d5ja00243e-f2.tif
Fig. 2 Throughput curve (full line) of an ion counter with τe = 20 ns, τne = 50 ns, covering a region where the input rate varies by 70% between 1.2 107 s−1 and 4 107 s−1. The dotted lines show the tangent (first derivative) and matching parabola (first and second derivative). There is a mismatch between the projections of [small rho, Greek, macron] and [R with combining macron] over this region.

Failing to correct for this non-linearity will introduce a bias in measured isotopic abundance ratios. This error can be significantly reduced by including a correction term within the throughput formula. It suffices to consider the simple scenario where the count rate varies linearly between two extreme values, ρmin and ρmax. While the mean input rate is simply [small rho, Greek, macron] = (ρmax + ρmin)/2, determining the mean output rate requires integrating the throughput factor. This mathematical problem can be simplified by approximating the throughput curve in the considered interval [ρmin, ρmax] by a parabola:

 
R(ρ) ≈ 2 + (6)
where the matching coefficients can be derived from a Taylor expansion:
 
image file: d5ja00243e-t6.tif(7)
and
 
image file: d5ja00243e-t7.tif(8)

Typical values of a and b at various input rates are presented in Fig. 3.


image file: d5ja00243e-f3.tif
Fig. 3 Plot of the parameters b and a, proportional to the first and second derivatives of the throughput formula, R = ρX(ρ), of an ion counter with τe = 20 ns, τne = 50 ns.

The mean output rate is then obtained from an integral:

 
image file: d5ja00243e-t8.tif(9)

The last term contains the variance of the input rate, which has a rectangular distribution. The solution in eqn (9) can be generalised to a new throughput formula applicable to any normalised statistical distribution f(ρ) of the input rate:

 
image file: d5ja00243e-t9.tif(10)

The error caused by fluctuations in the count rate is proportional to the variance of the input rate and half the second derivative of the throughput function. Since the second derivative of the throughput curve is negative (a < 0), the average count rate will be lower than expected from the stationary throughput formula in eqn (3), i.e. [R with combining macron] < [small rho, Greek, macron]X([small rho, Greek, macron]). These conclusions are compatible with observations in other fields of research, such as e.g. photon counting from a fluctuating light beam.16–19 Eqn (10) offers an elegant and broadly applicable solution to the expected change in average count rate. Other authors have also explored the impact of input fluctuations on the distribution and variance of counts.16–19

4 Inverse throughput at variable rate

From an experimental perspective, conducting measurements at variable rates will result in average output rates [R with combining macron] that align with eqn (10). Inversion of the throughput formula will yield an underestimate of the average input rate, [small rho, Greek, circumflex] = [R with combining macron]X−1([R with combining macron]) ≤ [small rho, Greek, macron], leading to a bias in isotopic abundance ratios. Since the error term in [R with combining macron] is relatively small, it can be translated into a variation in [small rho, Greek, macron] by using the inverse of the first derivative of X([small rho, Greek, macron]), which is 1/b:
 
image file: d5ja00243e-t10.tif(11)

As the experimenter cannot directly observe the variance of the input rate, σρ2, it must be estimated from the variance of the output rate, σR2. Again, a linear relationship between both standard deviations or a quadratic one between the variances, σR2b2σρ2, can serve as a suitable approximation. This leads to the following refinement of the inverse throughput formula:

 
image file: d5ja00243e-t11.tif(12)

The first term, [small rho, Greek, circumflex]< = [R with combining macron]X−1([R with combining macron]), follows from iteratively solving eqn (4) or (5). Then local values for the first and second derivatives forming a and b are calculated locally for [small rho, Greek, circumflex]< and an estimate for [small rho, Greek, macron] is calculated from eqn (12). A generally unnecessary next step would consist of another iteration of eqn (12) in which new values of a and b are determined using the obtained estimate of [small rho, Greek, macron]. Eqn (12) is applicable to virtually all mass spectrometry measurements utilising discrete ion counters. It is unnecessary for systems that are inherently linear (a = 0), such as current measurements in a Faraday cup. By incorporating this universal throughput formula into the data processing software as a standard procedure, the accuracy of abundance ratios is generally expected to improve.

5 Exponential variation

Whereas the correction factor in eqn (9) pertains to a linear change of the input rate, it is also a very good approximation for an exponential variation of the input rate, for example due to evaporation of the sample at a rate −λ. The mean input rate is slightly lower than for a rectangular distribution:
 
image file: d5ja00243e-t12.tif(13)
in which λT = ln(ρmax/ρmin) ≈ (ρmaxρmin)/[small rho, Greek, macron]. The variance divided by the squared mean of the exponential distribution is
 
image file: d5ja00243e-t13.tif(14)

The relative variation of the exponentially-distributed input rate is well approximated by the same formula as for the linear distribution, albeit that the mean value [small rho, Greek, macron] from eqn (13) is slightly lower than (ρmax + ρmin)/2. Consequently, (ρmaxρmin)2/12 can be used as variance in eqn (10) for an accurate prediction of the output rate (with [small rho, Greek, macron] from eqn (13)), and in eqn (11) and (12) to improve the input rate estimation.

6 Numerical examples

Numerical examples are provided in Table 1 and visualised in Fig. 4. An ion counter is modelled with a characteristic pulse width of τe = 20 ns and an imposed NEDT of τne = 50 ns. Table 1 shows a series of input rates that vary linearly during measurement, their rectangular distribution being characterised by the average value, [small rho, Greek, macron], and relative width, (ρmaxρmin)/[small rho, Greek, macron]. For each hypothetical measurement, the relative error on the estimated count rate, [small rho, Greek, circumflex]/[small rho, Greek, macron] − 1, is presented. The errors can be effectively reproduced through eqn (12), even when approximating the output rate variance with that of a rectangular distribution:
 
image file: d5ja00243e-t14.tif(15)
Table 1 Evaluation of relative error in estimated average input rate for linearly varying input rates between ρmax and ρmin. The values represent the relative error, [small rho, Greek, circumflex]/[small rho, Greek, macron] − 1, by inversion of the throughput formula, [small rho, Greek, circumflex] = [R with combining macron]X−1([R with combining macron]), for an ion counter with pulse width τe = 20 ns and NEDT τne = 50 ns, as a function of the average input rate and the relative width of the interval in which the input rate varies
[small rho, Greek, macron] [small rho, Greek, circumflex]/[small rho, Greek, macron] − 1 versus [small rho, Greek, macron] and (ρmaxρmin)/[small rho, Greek, macron]
10% 20% 30% 40% 50%
1.00 × 105 −0.00042% −0.0017% −0.0037% −0.0066% −0.0104%
3.00 × 105 −0.0012% −0.0049% −0.011% −0.020% −0.031%
1.00 × 106 −0.0040% −0.016% −0.036% −0.064% −0.100%
3.00 × 106 −0.0113% −0.045% −0.102% −0.18% −0.28%
7.00 × 106 −0.0237% −0.095% −0.213% −0.38% −0.59%
1.00 × 107 −0.032% −0.128% −0.29% −0.51% −0.80%
1.50 × 107 −0.045% −0.180% −0.40% −0.72% −1.12%
2.00 × 107 −0.059% −0.235% −0.53% −0.94% −1.47%
2.50 × 107 −0.075% −0.302% −0.68% −1.20% −1.88%
3.00 × 107 −0.098% −0.39% −0.88% −1.6% −2.4%



image file: d5ja00243e-f4.tif
Fig. 4 Relative error on the calculated average input rate by inverting the throughput curve for different widths of the input rate distribution during the measurement in an ion counter with τe = 20 ns, τne = 50 ns. Note that the width of the output rate distribution will appear smaller by a factor of b.

The statistical significance of the calculated errors should be evaluated by comparing them with the claimed accuracy of measurement results. By remarkably good approximation (relative difference <2%), the same relative errors apply to a scenario with an exponential variation of the input rates.

7 Time-resolved transient

Another numerical example pertains to conducting a time-resolved analysis of a Gaussian-shaped transient signal. It bares resemblance to a measurement result published by Strenge and Engelhard,20 who recorded 0.2 ms transient signals with 5 μs time resolution. The output rate reached values near 60 counts per 5 μs (equivalent to 1.2 × 107 s−1), which corresponded to estimated input rates as high as 140 counts per 5 μs (or 2.8 × 107 s−1) after dead-time correction. The authors applied the default dead-time correction method outlined in eqn (2) for each time interval to mitigate count loss in the spike's count integral. At such elevated count rates, the throughput formula significantly diverges from the NEDT formula in eqn (1), necessitating the adoption of eqn (3) for a cascade of EDT-NEDT instead.

For the sake of illustration, the dwell time in the numerical example is chosen to be about 1/10th of the width of the Gaussian spike, as shown in Fig. 5. Consequently, the average count rate exhibits variability within each time interval, which can be approximated by comparing the average count rates in neighbouring time intervals:

 
image file: d5ja00243e-t15.tif(16)
in which i is the index number of the time interval. Alternative formulas for the variance have been derived, e.g.
 
image file: d5ja00243e-t16.tif(17)
or, based on a parabolic fit to three data:
 
image file: d5ja00243e-t17.tif(18)
but in numerical tests of the integral error, eqn (16) appeared the most successful in combination with eqn (15).


image file: d5ja00243e-f5.tif
Fig. 5 (Top) Transient signal with Gaussian shape and corresponding output rate in a discrete counter with τe = 20 ns, τne = 50 ns. (Bottom) Time-resolved input and output count rates averaged over finite time intervals indicated by vertical dashed lines, and the dead-time corrected count rate based on the NEDT model in eqn (2).

In Table 2, the input and output rates of the Gaussian spike are averaged in each interval. Starting from the output rates Ri, estimates of the input rates ρi are calculated using three approaches: (1) [small rho, Greek, circumflex]i(1) = the NEDT inverse throughput formula in eqn (2), (2) [small rho, Greek, circumflex]i(2) = the static EDT-NEDT formula in eqn (5), and (3) [small rho, Greek, circumflex]i(3) = the variance-corrected EDT–NEDT formula in eqn (12). For a Gaussian with peak input rate of 3 × 107 s−1 while τe = 20 ns and NEDT τne = 50 ns, the error in reproducing the count integral associated with the three methods are (1) −12%, (2) −1.6%, and (3) −0.1%, respectively. The biggest gain in accuracy is achieved by accounting for the pileup effect. Accounting for the count rate variance within each channel can further reduce the remaining error by another order of magnitude, depending on the dwell time of the spectrometer.

Table 2 Time-resolved analysis of a Gaussian-shaped transient signal using a discrete ion counter with τe = 20 ns and NEDT τne = 50 ns. The data refer to the (i) hypothetical time intervals, (ii) the average input rate and (iii) output rate in each time range, (iv) the estimated output-rate variance from eqn (16), the estimated input rate based on the inverse throughput formulas in (v) eqn (2), (vii) eqn (5) and (ix) eqn (12), and (vi) (viii) (x) the corresponding relative differences with the real input rate
Time range ρi Ri σRi2 [small rho, Greek, circumflex]i(1) Error [small rho, Greek, circumflex]i(2) Error [small rho, Greek, circumflex]i(3) Error
0–9 1.04 × 105 1.03 × 105 1.35 × 1010 1.04 × 105 −0.2% 1.04 × 105 −0.2% 1.04 × 105 0.4%
10–19 8.56 × 105 8.12 × 105 2.34 × 1011 8.46 × 105 −1.2% 8.46 × 105 −1.1% 8.60 × 105 0.4%
20–29 4.31 × 106 3.45 × 106 9.79 × 1011 4.18 × 106 −3.2% 4.19 × 106 −2.9% 4.28 × 106 −0.7%
30–39 1.33 × 107 7.67 × 106 1.00 × 1012 1.24 × 107 −6.5% 1.29 × 107 −3.2% 1.32 × 107 −0.8%
40–49 2.51 × 107 1.04 × 107 2.21 × 1011 2.16 × 107 −13.8% 2.48 × 107 −1.2% 2.51 × 107 0.1%
50–59 2.91 × 107 1.09 × 107 7.13 × 1010 2.41 × 107 −17.3% 2.91 × 107 −0.1% 2.92 × 107 0.5%
60–69 2.07 × 107 9.60 × 106 5.07 × 1011 1.85 × 107 −10.6% 2.02 × 107 −2.2% 2.06 × 107 −0.2%
70–79 8.99 × 106 5.99 × 106 1.17 × 1012 8.55 × 106 −4.9% 8.69 × 106 −3.4% 8.90 × 106 −1.0%
80–89 2.40 × 106 2.10 × 106 6.56 × 1011 2.34 × 106 −2.3% 2.34 × 106 −2.2% 2.39 × 106 −0.2%
90–99 3.91 × 105 3.81 × 105 8.92 × 1010 1.42 × 105 −0.3% 1.42 × 105 −0.3% 1.43 × 105 0.6%
Integral         −12%   −1.6%   −0.1%


8 Conclusion

The throughput factor in eqn (3) quantifies the relationship between a constant input rate of ions entering an ion counter and the resulting output rate, which is affected by count loss through dead time and pulse pileup. However, when the input rates exhibit significant variations during measurement, the throughput formula has to be refined as outlined in eqn (9) and (10), with a correction factor to account for the non-linearity between the average input and output rates. Table 1 provides examples of anticipated relative errors, applicable to scenarios where the input rates vary linearly or exponentially by 10% to 50%. Table 2 shows similar errors in time-resolved analysis of a Gaussian-shaped transient signal. In such instances, it suffices to calculate the variance of the output rate from its extreme values using eqn (15) and to propagate this variance to derive an unbiased input rate using eqn (12). As a result, the linearity of discrete ion counters can be improved for varying beam intensities, and bias in mass spectrometry measurement with discrete ion counters can be mitigated.

Data availability

All data in Tables and graphs have been generated using the formulas presented in this paper.

Conflicts of interest

The author has no competing interests to declare. The author ensures that the work is entirely original.

References

  1. N. S. Lloyd, J. Schwieters, M. S. A. Horstwood and R. R. Parrish, Particle detectors Used in Isotope Ratio Mass Spectrometry, with Applications in Geology, Environmental Science and Nuclear Forensics, in Handbook of Particle Detection and Imaging, ed. C. Grupen and I. Buvat, Springer-Verlag, Berlin Heidelberg, 2012 Search PubMed.
  2. S. Pommé and S. Boulyga, A throughput model explaining non-linearity in discrete ion counters used in mass spectrometry, J. Anal. At. Spectrom., 2025 10.1039/D5JA00085H.
  3. S. Pommé, Cascades of pile-up and dead time, Appl. Radiat. Isot., 2008, 66, 941–947 CrossRef PubMed.
  4. S. Pommé, R. Fitzgerald and J. Keightley, Uncertainty of nuclear counting, Metrologia, 2015, 52, S3–S17 CrossRef.
  5. S. Richter, S. A. Goldberg, P. B. Mason, A. J. Traina and J. B. Schwieters, Linearity tests for secondary electron multipliers used in isotope ratio mass spectrometry, Int. J. Mass Spectrom., 2001, 206, 105–127 CrossRef CAS.
  6. U. Nygren, H. Ramebäck, M. Berglund and D. C. Baxter, The importance of a correct dead time setting in isotope ratio mass spectrometry: Implementation of an electronically determined dead time to reduce measurement uncertainty, Int. J. Mass Spectrom., 2006, 257, 12–15 CrossRef CAS.
  7. S. Richter, A. Alonso, Y. Aregbe, R. Eykens, F. Kehoe, H. Kühn, N. Kivel, A. Verbruggen, R. Wellum and P. D. P. Taylor, A new series of uranium isotope reference materials for investigating the linearity of secondary electron multipliers in isotope mass spectrometry, Int. J. Mass Spectrom., 2009, 281, 115–125 CrossRef CAS.
  8. U. Nygren, H. Ramebäck, A. Vesterlund and M. Berglund, Consequences of and potential reasons for inadequate dead time measurements in isotope ratio mass spectrometry, Int. J. Mass Spectrom., 2011, 300, 21–25 CrossRef CAS.
  9. S. Richter, S. Konegger-Kappel, S. F. Boulyga, G. Stadelmann, A. Koepf and H. Siegmund, Linearity testing and dead-time determination for MC-ICP-MS ion counters using the IRMM-072 series of uranium isotope reference materials, J. Anal. At. Spectrom., 2016, 31, 1647–1657 RSC.
  10. M. Tanner and D. Günther, Short transient signals, a challenge for inductively coupled plasma mass spectrometry, a review, Anal. Chim. Acta, 2009, 633, 19–28 CrossRef CAS PubMed.
  11. T. Nomizu, H. Hayashi, N. Hoshino, T. Tanaka, H. Kawaguchi, K. Kitagawa and S. Kaneco, Determination of zinc in individual airborne particles by inductively coupled plasma mass spectrometry with digital signal processing, J. Anal. At. Spectrom., 2002, 17, 592–595 RSC.
  12. J. W. Olesik and P. J. Gray, Considerations for measurement of individual nanoparticles or microparticles by ICP-MS: determination of the number of particles and the analyte mass in each particle, J. Anal. At. Spectrom., 2012, 27, 1143–1155 RSC.
  13. I. Strenge and C. Engelhard, Single particle inductively coupled plasma mass spectrometry: investigating nonlinear response observed in pulse counting mode and extending the linear dynamic range by compensating for dead time related count losses on a microsecond timescale, J. Anal. At. Spectrom., 2020, 35, 84–99 RSC.
  14. M. Lomax-Vogt, L. M. Carter, J. Wielinski, S. Kutuzov, G. V. Lowry, R. Sullivan, P. Gabrielli and J. W. Olesik, Challenges in measuring nanoparticles and microparticles by single particle ICP-QMS and ICP-TOFMS: size-dependent transport efficiency and limited linear dynamic range, J. Anal. At. Spectrom., 2025, 40, 848–859 RSC.
  15. H. Siegmund, S. Konegger-Kappel, S. Boulyga, M. Kilburn and S. Pommé, Recording high-intensity transient signals with secondary electron multipliers in MC-ICPMS: dead-time correction and non-linearity effects, 2025, in preparation.
  16. G. Bédard, Dead-time corrections to the statistical distribution of photoelectrons, Proc. Phys. Soc., 1967, 90, 131–141 CrossRef.
  17. M. C. Teich and W. J. McGill, Neural Counting and Photon Counting in the Presence of Dead Time, Phys. Rev. Lett., 1976, 36, 754–758 CrossRef.
  18. G. Vanucci and M. C. Teich, Effects of rate variation on the counting statistics of dead-time-modified Poisson processes, Opt. Commun., 1978, 25, 267–272 CrossRef.
  19. M. L. Larsen and A. B. Kostinski, Simple dead-time corrections for discrete time series of non-Poisson data, Meas. Sci. Technol., 2009, 20, 095101 CrossRef.
  20. I. Strenge and C. Engelhard, Capabilities of fast data acquisition with microsecond time resolution in inductively coupled plasma mass spectrometry and identification of signal artifacts from millisecond dwell times during detection of single gold nanoparticles, J. Anal. At. Spectrom., 2016, 31, 135–144 RSC.
  21. D. Mozhayeva and C. Engelhard, A critical review of single particle inductively coupled plasma mass spectrometry – A step towards an ideal method for nanomaterial characterization, J. Anal. At. Spectrom., 2020, 35, 1740–1783 RSC.
  22. T. Hirata, S. Yamashita, M. Ishida and T. Suzuki, Analytical Capability of High-Time Resolution-Multiple Collector-Inductively Coupled Plasma-Mass Spectrometry for the Elemental and Isotopic Analysis of Metal Nanoparticles, Mass Spectrom., 2020, 9, A0085 CrossRef CAS PubMed.
  23. S. Yamashita, M. Ishida, T. Suzuki, M. Nakazato and T. Hirata, Isotopic analysis of platinum from single nanoparticles using a high-time resolution multiple collector Inductively Coupled Plasma - Mass Spectroscopy, Spectrochim. Acta, Part B, 2020, 169, 105881 CrossRef CAS.
  24. T. Hirata, S. Niki, S. Yamashita, H. Asanuma and H. Iwano, Uranium-lead isotopic analysis from transient signals using high-time resolution-multiple collector-ICP-MS (HTR-MC-ICP-MS), J. Anal. At. Spectrom., 2021, 36, 70–74 RSC.
  25. A. M. Duffin, E. D. Hoegg, R. I. Sumner, T. Cell, G. C. Eiden and L. S. Wood, Temporal analysis of ion arrival for particle quantification, J. Anal. At. Spectrom., 2021, 36, 133–141 RSC.
  26. A. Hineman and C. Stephan, Effect of dwell time on single particle inductively coupled plasma mass spectrometry data acquisition quality, J. Anal. At. Spectrom., 2014, 29, 1252–1257 RSC.
  27. P. Shaw and A. Donard, Nano-particle analysis using dwell times between 10 μs and 70 μs with an upper counting limit of greater than 3 × 107 cps and a gold nanoparticle detection limit of less than 10 nm diameter, J. Anal. At. Spectrom., 2016, 31, 1234–1242 RSC.
  28. A. Schardt, J. Schmitt and C. Engelhard, Single particle inductively coupled plasma mass spectrometry with nanosecond time resolution, J. Anal. At. Spectrom., 2024, 39, 389–400 RSC.
  29. S. Pommé, Radionuclide metrology: confidence in radioactivity measurements, J. Radioanal. Nucl. Chem., 2022, 331, 4771–4798 CrossRef.
  30. S. Pommé, When the model doesn't cover reality: examples from radionuclide metrology, Metrologia, 2016, 53, S55–S64 CrossRef.
  31. International Commission on Radiation Units and Measurements (ICRU), Particle counting in radioactivity measurements, ICRU Report, Maryland, USA, 1994, vol. 52 Search PubMed.
  32. I. De Lotto, P. F. Manfredi and P. Principi, Counting statistics and dead-time losses, Energ. Nucl., 1964, 11, 557–564 Search PubMed.
  33. J. W. Müller, Sur l’arrangement en série de deux temps morts de types différents, Internal Report BIPM, Sèvres, France, BIPM-73/9, 1973 Search PubMed.
  34. J. W. Müller, Dead-time problems, Nucl. Instrum. Methods, 1973, 112, 47–57 CrossRef.
  35. J. W. Müller, Stastiques de comptage, Internal Report BIPM, Sèvres, France, BIPM-82/13, 1982 Search PubMed.
  36. J. Libert, Détermination du taux de comptage en amont de deux temps morts classiques en série et de types différents, Nucl. Instrum. Meth. A, 1989, 274, 319–323 CrossRef.
  37. S. Pommé, B. Denecke and J.-P. Alzetta, Influence of pileup rejection on nuclear counting, viewed from the time-domain perspective, Nucl. Instrum. Meth. A, 1999, 426, 564–582 CrossRef.
  38. S. Pommé, Time-interval distributions and counting statistics with a non-paralysable spectrometer, Nucl. Instrum. Meth. A, 1999, 437, 481–489 CrossRef.
  39. S. Pommé and G. Kennedy, Pulse loss and counting statistics with a digital spectrometer, Appl. Radiat. Isot., 2000, 52, 377–380 CrossRef PubMed.
  40. S. Pommé and J. Uyttenhove, Statistical precision of high-rate spectrometry with a Wilkinson ADC, J. Radioanal. Nucl. Chem., 2001, 248, 263–266 CrossRef.
  41. S. Pommé, Dead time, pile-up and counting statistics, in Applied Modeling and Computations in Nuclear Science, ACS Symposium Series, 2007, vol. 945, pp. 218–233 Search PubMed.
  42. S. Pommé and J. Keightley, Count rate estimation of a Poisson process - Unbiassed fit versus central moment analysis of time interval spectra. In: Applied Modeling and Computations in Nuclear Science, ACS Symp. Ser., 2007, 945, 316–334 CrossRef.
  43. C. Michotte and M. Nonis, Experimental comparison of different dead-time correction techniques in single-channel counting experiments, Nucl. Instrum. Meth. A, 2008, 608, 163–168 CrossRef.
  44. S. Cheng, B. Pierson, S. Pommé and M. Flaska, Time interval distributions of nuclear events in a digital spectrometer, Nucl. Instrum. Meth. A, 2024, 1063, 169218 CrossRef CAS.
  45. S. Pommé and K. Pelczar, A simplified approach to counting statistics with an imperfect pileup rejector, Nucl. Instrum. Meth. A, 2025, 1070, 170067 CrossRef.
  46. S. Cheng, B. Pierson, S. Pommé, B. Archambault and M. Flaska, Time-interval distributions in a digital gamma multi-channel analyzer at extreme input rates, Nucl. Instrum. Meth. A, 2025, 1073, 170260 CrossRef CAS.
  47. R. Fitzgerald, Corrections for the combined effects of decay and dead time in live-timed counting of short-lived radionuclides, Appl. Radiat. Isot., 2016, 109, 335–340 CrossRef CAS PubMed.
  48. G. P. Westphal, Review of loss-free counting in nuclear spectroscopy, J. Radioanal. Nucl. Chem., 2008, 275, 677–685 CrossRef CAS.
  49. S. Pommé, J.-P. Alzetta, J. Uyttenhove, B. Denecke, G. Arana and P. Robouch, Accuracy and precision of loss-free counting in γ-ray spectrometry, Nucl. Instrum. Meth. A, 1999, 422, 388–394 CrossRef.
  50. S. Pommé, How pileup rejection affects the precision of loss-free counting, Nucl. Instrum. Meth. A, 1999, 432, 456–470 CrossRef.
  51. S. Pommé, Experimental test of the 'zero dead time' count-loss correction method on a digital gamma-ray spectrometer, Nucl. Instrum. Meth. A, 2001, 474, 245–252 CrossRef.
  52. S. Pommé, A plausible mathematical interpretation of the 'variance' spectra obtained with the DSPECPLUS TM digital spectrometer, Nucl. Instrum. Meth. A, 2002, 482, 565–566 CrossRef.
  53. S. Pommé, On the statistical control of 'loss-free counting' and 'zero dead time' spectrometry, J. Radioanal. Nucl. Chem., 2003, 257, 463–466 CrossRef.

This journal is © The Royal Society of Chemistry 2025
Click here to see how this site uses Cookies. View our privacy policy here.