Jordan L.
Tanen
and
Kaveh
Jorabchi
*
Department of Chemistry, Georgetown University, Washington, DC 20057, USA. E-mail: kj256@georgetown.edu
First published on 15th April 2025
Recent developments in high-power nitrogen microwave plasmas have made them attractive excitation and ionization sources for elemental analysis, offering robust operation with liquid sample introduction and significant cost savings. However, nonmetal detection, particularly F analysis, remains challenging with these plasmas because of their lower excitation and ionization temperatures compared to Ar-ICP while the need for elemental F analysis continues to rise due to prevalence of fluorochemicals in pharmaceuticals and environmental contaminants. Here, we combine the advantages of a N2 microwave inductively coupled atmospheric pressure plasma (MICAP) with those of post-plasma chemical ionization and evaluate the approach for elemental F analysis. The MICAP acts as a robust plasma reactor to break down fluorochemicals into HF which is then ionized to ScFNO3(H2O)n+ and detected by a quadrupole MS as ScF+ after in-source ion activation. The approach offers a sensitivity of ∼18 cps ng−1 mL of F, superior to < 4 cps ng−1 mL of F achieved by BaF+ detection in Ar-ICP-MS/MS. An LOD of ∼40 ng mL−1 of F is obtained comparable to those of Ar-ICP-MS/MS, limited by background equivalent concentration likely dominated by F contamination. F response factors are also independent of chemical species when sample introduction biases are minimized. However, broadening of flow injection peaks over time is observed, especially with high oxygen levels in the aerosol gas, denoting future areas of improvement for better analytical performance. These studies indicate a high potential of MICAP with post-plasma chemical ionization for F quantitation.
Noting the challenges of in-plasma ionization, we have developed a post-plasma chemical ionization approach for F detection.10–13 In contrast to ICP-MS, we use the plasma to convert F in fluorospecies to HF, which then undergoes atmospheric pressure chemical ionization downstream of the plasma by reagent ions produced from a nanospray ionization source. Interestingly, BaF+ is also observed in this approach by using a Ba-containing electrolyte in nanospray, but with ∼100 fold better sensitivity (281 cps ng−1 mL) compared to that in the ICP-MS studies.10 The drastically improved ionization, however, is accompanied by susceptibility to interference from highly acidic species produced by the plasma.11,12 In particular, nitric acid, abundantly generated when high levels of oxygen is added to the Ar-ICP to accommodate organic solvents, results in suppression of BaF+ signal. Fortunately, the chemical tunability of the ionization reactions has resolved this limitation where using a Sc-containing electrolyte in nanospray results in formation of ScFNO3+ from gaseous HF in the presence of abundant nitric acid.13 Further, large changes in aerosol flow rate (1.6–2.2 L min−1) and RF power level (1100–1500 W) are found to have small effects on signal detection in this approach,14 highlighting minimal impact of plasma temperature on ion generation, thus affording more robust methods.
Moreover, thanks to the reduced effect of plasma properties on ion formation, a variety of plasma types can be readily coupled with post-plasma ionization for nonmetal detection. We have demonstrated F detection using dielectric barrier discharge coupled to chemical ionization15 as well as P and S detection by polyatomic ion formation downstream of a solution cathode glow discharge.16 Relatedly, recent years have seen a renewed interest in nitrogen plasmas for elemental analysis, partly fueled by high-power annular microwave plasmas that offer stable operation with solution introduction, high tolerance to organic solvents, and substantial savings in operating costs compared to Ar-ICP.17–23 The compatibility with organic solvents is expected to facilitate coupling of gradient liquid chromatography to the plasma-based detectors. Notably, the N2 microwave-sustained, inductively coupled, atmospheric-pressure plasma (MICAP) has also been adopted as an ionization source for elemental MS, showing promise in detection of metals and metalloids.24–30 However, nonmetal detection with this ion source is yet to be achieved. Significantly lower excitation temperatures and Mg II/Mg I ratios in N2-MICAP relative to those in Ar-ICP suggest a more difficult path for in-plasma ionization of nonmetals with this plasma.31
With the advantages noted above for N2-MICAP regarding higher tolerance to organic solvents but lower ionization temperatures compared to Ar-ICP, we envisioned this plasma as a desirable reactor for nonmetal detection when coupled with post-plasma ionization. As such, this study describes the preliminary evaluation of N2-MICAP to assess its potential for fluorine detection using post-plasma ionization and identifies areas of improvement. To our knowledge, this is the first effort in mass spectrometric elemental F analysis by using a nitrogen plasma.
The plasma sampling and downstream ionization were achieved by a homebuilt two-chamber interface schematically shown in Fig. 1. The design was similar to that reported previously.13,14 A 3 mm nickel orifice (Spectron Inc, Ventura, CA), mounted on a cooled aluminum plate and placed 3 mm downstream of the torch, provided plasma sampling port while an exhaust flow installed above the torch outlet evacuated potentially harmful gases. Threaded aluminum rods connected the cooled aluminum plate and the MS (QTRAP, Sciex, Framingham, MA) ion sampling plate, sandwiching and tightly holding the two chambers.
![]() | ||
Fig. 1 Schematic of the N2-MICAP coupled to post-plasma ionization interface. Figure reproduced with modifications from ref. 13. |
The first chamber was composed of a 4.82 cm long, 5.08 cm o.d., 4.45 cm i.d. acrylic tube pressed between two aluminum disks and housed: (1) a quartz recombination tube (50 mm long 6.35 mm o.d. × 4 mm i.d.) with its entrance 5 mm downstream of the sampling orifice and fit to the left aluminum disk using graphite ferrule, (2) two parallel plate brass electrodes (10 mm × 10 mm × 3.17 mm), and (3) the nanospray ionization source. The second chamber was composed of a 20 mm long, 5.08 cm o.d., 4.45 cm i.d aluminum tube sandwiched between the middle aluminum disk and the ion sampling plate of the MS. A 5 mm L-shaped channel through the middle aluminum disk allowed gas evacuation of the second chamber whose flow rate was monitored by a mass flow meter (Sierra Instruments, Monterey, CA). This evacuation pulled the gas from the first chamber toward the MS inlet through a steel tube (28 mm long, 4.5 mm i.d. × 6.35 mm o.d.) press fit into the middle aluminum disk. This arrangement mixed the ions generated by nanospray with neutrals generated by the plasma, enhancing ionization stability.
Plasma sampling was controlled by the flow rate of nitrogen into the first chamber (delivered by a mass flow controller, MKS instruments, Andover, MA) relative to the evacuation rate of the second chamber. The sampled plasma was cooled in the quartz tube, inducing recombination reactions to form HF from fluorochemicals introduced into the plasma.11 The bottom brass electrode was biased to +500 V and the upper one was grounded to the MS to deflect ions and electrons coming from the plasma, minimizing plasma effects on nanospray operation while allowing plasma-generated neutral species to move toward the MS.
The plasma neutral products were ionized by reagent ions generated from a nanospray ionization source prepared from borosilicate capillaries (1 mm o.d., 0.75 mm i.d., WPI, Sarasota, FL) pulled to a diameter of ∼5 μm with a micropipette puller (Sutter Instrument Company, Novato, CA, model P-87). The nanospray capillary was loaded with 1 mM scandium nitrate electrolyte and secured to a holder attached to the first chamber. The capillary protrusion was adjusted so that the spray tip was visually aligned with the top plane of the upper brass plate. A platinum electrode inserted into the capillary supplied the spray voltage (1300 to 1500 V). The nanospray-produced ions mixed with the plasma neutral products and were pulled toward the MS through the steel tube by the gas flow. The entrance of the steel tube was located 8 mm away from the nanospray tip, while its end was placed 4 mm upstream of the ion sampling plate of the MS. A fraction of the ions was sampled into the MS by the MS intake flow while the rest was swept away through the evacuation port of the 2nd chamber.
The operating parameters were optimized for best signal intensity via one factor at a time approach using flow injections of 1860 ng F mL−1 from fluconazole. Anticipating a minimal effect of plasma power based on our previous studies with Ar-ICP,14 the gas flow optimizations were conducted at a middle power of 1300 W. Aerosol N2 makeup gas of 0.4 L min−1 (yielding 1.5 L min−1 total aerosol gas flow rate) and torch coolant (outer) and auxiliary gas flow rates of 16 and 1.1 L min−1, respectively, provided the optimized conditions for plasma operation. Although oxygen addition to aerosol gas is not required for preventing soot formation upon organic solvent introduction into N2-MICAP, we noticed that small amounts of oxygen in aerosol gas improved the plasma stability (e.g. preventing sudden extinguishment) particularly at lower power levels. Given the minimal effect of power on signal generation, plasma operation at lower powers is preferred as it reduces the burden on interface cooling and minimizes energy usage. Therefore, 5 mL min−1 oxygen (controlled by a mass flowmeter, MKS instruments, Andover, MA) was added to 0.4 L min−1 N2 makeup gas via a tee for all experiments, except for peak broadening studies where higher oxygen level was intentionally used to accelerate the broadening for more effective investigation. The interface gas evacuation flow rate was optimized at 4 L min−1 while 3.4–3.5 L min−1 interface N2 input gas flow rate provided optimal F detection. The effect of power was checked in these conditions and in line with anticipation, no significant effect was observed in 1100–1500 W. The experiments were conducted at 1300 W unless otherwise noted.
Sc(NO3)2(H2O)+n(g) + HF(g) ⇄ ScNO3F(H2O)+n(g) + HNO3(g) | (R1) |
Accordingly, we selected Sc-based ionization for post-MICAP ionization of HF. To evaluate major MICAP products relevant to Sc-base ionization, we examined the ions generated from nanospray of a 1 mM Sc nitrate solution before and after interaction with plasma afterglow.
Fig. S1† depicts the effect of plasma sampling on nanospray ions. To monitor the original nanospray ions prior to any interaction with plasma afterglow, the plasma was kept off and the interface input nitrogen flow rate was set to 4.5 L min−1. The higher input flow rate compared to the evacuation flow rate (4 L min−1) ensures that the interface is flush with nitrogen (with extra gas pushing out of the sampling orifice) and that the lab air is prevented from entering the interface. Hydrated Sc(OH)2+, Sc(OH)(NO3)+, and Sc(NO3)2+ are detected from nanospray (Fig. S1A†) in these conditions. These ions form during the electrospray ionization process, where nitrate ions bind to Sc3+ and hydroxide species are produced from hydrolysis by Sc(III).
When the plasma is turned on and the interface input nitrogen flow rate is lowered to 3.5 L min−1 to sample the plasma into the interface, all Sc-containing ions convert into Sc(NO3)2(H2O)n+, as depicted in Fig. S1B.† This conversion confirms abundant nitric acid formation in the plasma afterglow. No other major species are evident in Fig. S1B,† leaving Sc(NO3)2(H2O)n+ as the main reagent ion for ionization of HF. With this confirmation, we proceeded to evaluate analytical performance for F detection via(R1) as discussed in the next section.
Analytical potential for F detection using ScF+ was evaluated by constructing a calibration curve in the concentration range of 207–1881 ng mL−1 F using flow injection peak heights from injections of flucytosine as shown in Fig. 2. A linear regression model with weights of 1/σsignal resulted in r2 of 0.9988, indicating good linearity with a F sensitivity of 18.4 cps ng−1 mL. No detectable peaks were observed upon injections of blanks. A detection limit of 38 ng F mL−1 was determined based on 3 × σbaseline/sensitivity where σbaseline was obtained from the standard deviation of the baseline prior to flow injections.
Notably, the sensitivity for F detection using MICAP with post-plasma nanospray ionization is substantially greater than those of ICP-MS methods utilizing in-plasma ion formation as summarized in Table 1. The differences between sensitivities of various post-plasma ionization methods in Table 1 largely originate from variations in MS ion sampling and transmission efficiencies within different MS instruments as well as from differences in fragmentation schemes. Nonetheless, the sensitivity in the current work shows the potential of MICAP post-plasma ionization MS for F detection, particularly considering the use of a dated Qtrap MS platform.
Method | Ion or MS/MS transition | Integration time (ms) | LOD (ng mL−1) | BEC (ng mL−1) | Sensitivity (cps ng−1 mL) | Sample flow rate and solvent |
---|---|---|---|---|---|---|
a Recalculated from original raw data obtained from author. | ||||||
ICP-HRMS5 | F+ | N/A | 5070 | 2050 | 0.026 | 100–150 μL min−1 water |
ICP-MS/MS7 | BaF+ → BaF+ | 1000 | 43 | 607 | 3.2 | 330 μL min−1 water |
ICP-MS/MS6 | BaF+ → BaF(NH3)3+ | 1000 | 22 | 21 | 1.6 | 300 μL min−1 water |
ICP-MS/MS8 | BaF+ → BaF+ | 1000 | 60 | 400 | 2 | Water (flow not reported) |
ICP-MS32 | BaF+ | 9000 | 32 | 30 | 0.4 | Water (flow not reported) |
Ar-ICP-nanospray-MS13 | ScNO3F(H2O)4+ | 500 | 19 | 357 | 360 | 50 μL min−1 1![]() ![]() ![]() ![]() |
Ar-ICP-nanospray-MS/MS13 | ScNO3F(H2O)4+ → ScFO2+ | 500 | 16 | 120 | 14 | 50 μL min−1 1![]() ![]() ![]() ![]() |
Ar-ICP-nanospray-MS/MS14 | ScNO3F(H2O)4+ → ScFO2+ | 500 | 2.2 | 25a | 134a | 50 μL min−1 0.15![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() |
N2-MICAP-nanospray-MS (this work) | ScF+ (broad ion activation) | 500 | 38 | 437 | 18.4 | 50 μL min−1 1![]() ![]() ![]() ![]() |
The LOD for fluorine detection in the current work is within the range of LODs reported using ICP-MS/MS methods (20–60 ng F mL−1). However, a higher LOD is obtained here compared to our previous works using post-ICP chemical ionization as noted in Table 1. This higher LOD is largely a result of higher background equivalent concentration (BEC) rather than major changes in signal stability. The BEC has contributions from isobaric interferences and F contamination. One potential isobaric interference is from Sc18OH+, as ScOH+ is generated by in-source fragmentation of reagent ions, as shown in Fig. 3. Based on the isotopic distribution of ScOH+, Sc18OH+ comprises about 2.9% of BEC, ruling out this ion as a major contributor. Further studies will be needed to confirm the components of BEC, however, based on our experience with Ar-ICPs coupled to high-resolution Orbitrap MS that readily separate isobaric interferences, we hypothesize F contamination from various sources is likely the major contributing factor to the BEC.
![]() | ||
Fig. 3 Ions observed from in-source fragmentation (175 V DP) of reagent ions with plasma sampling (see Fig. S1B† with soft ion sampling for comparison). Experiments were performed with interface input nitrogen flow rate of 3.5 L min−1. |
To investigate the causes of this deviation, we considered potential sources of compound dependent response. Species-specific variation in response can arise from two processes: (1) differences in degradation of compounds inside the plasma, leading to HF production efficiency shifts, and (2) biases in sample introduction. While volatility is known to affect the analyte transport efficiency in sample introduction,14 it is not clear if this effect is a major source for the deviation observed in Fig. 4A. The PFAS compounds are expected to have higher volatility but they do not show deviations consistent with their boiling points or vapor pressures (Table S1†). Nonetheless, the sample introduction biases can be minimized by using a high-transport efficiency setup (e.g. a single pass spray chamber), thus allowing the differentiation of the two causes noted above. To this end, we utilized a custom-built single pass spray chamber (∼20 mL volume, Precision Glassblowing, Centennial, CO) with an orthogonal tangential makeup gas introduction port. The same HEN nebulizer was used but at a nebulizer N2 gas flow rate of 0.4 L min−1 while makeup gases of 1.1 L min−1 nitrogen and 5 mL min−1 oxygen were added to maintain the same total aerosol flow rate and gas composition between the two spray chamber configurations. PFOA response factor was then measured along with that of p-fluoro-DL-phenylalanine. As depicted in Fig. 4B, about 2-fold higher response factor for the two compounds was measured using single pass spray chamber relative to those using cyclonic spray chamber, suggesting improved analyte transport efficiency. Further, the response factors for the two compounds were similar in Fig. 4B, denoting sample introduction bias as the main reason for deviation from species-independent response using cyclonic spray chamber. These studies show the importance of establishing species-independent response with new sample introduction configurations particularly when nebulizer gas is switched from typical argon gas.
Notably, a major acceleration in peak broadening was observed when larger amounts of oxygen were introduced into the aerosol flow. On the other hand, using higher oxygen levels offered two advantages: (1) a more stable plasma operation at lower power levels, reducing burden on interface cooling and energy consumption without a compromise in ion generation efficiency; (2) maintaining an oxidative plasma chemistry and minimizing impact of solvent composition on end plasma products. The latter provides more robust post-plasma ionization and promotes full oxidation of carbon to CO2, minimizing formation of toxic species such as HCN. With these considerations, the follow-up peak broadening investigations were conducted at 1100 W in accelerated broadening conditions by using 0.4 L min−1 oxygen (instead of nitrogen) as makeup gas merged with 1.1 L min−1 nitrogen aerosol gas from a cyclonic spray chamber.
Fig. 5A depicts the flow injection peaks in above conditions. A fresh quartz tube was used for these experiments. Peak broadening is visually observable in Fig. 5A starting at 20 minutes of plasma sampling time, defined as the amount of time elapsed since the interface nitrogen gas flow rate was lowered from 4.5 to 3.4 L min−1 to sample plasma into the interface. The area-to-height ratios of these peaks are shown by the red trace (Tube 1) in Fig. 5B, quantifying the broadening. Peak areas are plotted in Fig. 5C (red trace corresponding to Tube 1), showing constant signal response, confirming that no analyte was lost over time and that the change in area-to-height ratio was solely due to broadening.
Interestingly, when the post-plasma quartz tube was replaced with a fresh quartz tube, the behavior was reproduced with an initial narrow peak shape followed by broadening as shown in Fig. 5B (Tubes 2 and 3). The pace of broadening was variable between different tubes, however, they all produced similar peak areas (Fig. 5C). These observations suggest that plasma sampling modifies the surface of the quartz tube, leading to enhanced interactions of HF with the tube surface during passage to the ionization region. Interestingly, such high levels of peak broadening were not observed in our experiments using post-plasma chemical ionization with Ar-ICP,10,13,14 denoting critical effect of plasma chemistry and plasma properties on recombination tube surface modification whose details remain to be elucidated. It is of note that eliminating high oxygen levels in the aerosol gas allows operation for several hours without significant peak broadening, enabling analytical measurements as reported in the earlier sections of this work.
Although quantitative potential is evident, peak broadening was observed throughout this study, particularly when high oxygen levels are used in the aerosol gas, degrading the performance over time. This effect is associated with adsorption/desorption of HF during passage through the quartz recombination tube. A redesign of the ionization interface to remove the quartz tube while maintaining plasma cooling for HF formation and effective chemical ionization would be a logical solution to eliminate the issue of peak broadening. The work in this area is underway in our laboratory.