Open Access Article
This Open Access Article is licensed under a
Creative Commons Attribution 3.0 Unported Licence

First attempt to determine oxygen isotopes in oxygen by MC-ICP-MS

Zhenyi Liu a, Jie Lin *a, Xin Jiang a, Xi Zhu a, Wengui Liu a, Yongsheng Liu ab, Wen Zhang a and Zhaochu Hu ac
aState Key Laboratory of Geological Processes and Mineral Resources, School of Earth Sciences, China University of Geosciences, Wuhan 430074, China. E-mail: linjie@cug.edu.cn
bYangtze University, Jingzhou, 434023, China
cFaculty of Materials Science and Chemistry, China University of Geosciences, Wuhan 430074, China

Received 21st January 2025 , Accepted 3rd February 2025

First published on 11th March 2025


Abstract

Oxygen is the key component of crustal and mantle rocks and fluids. The oxygen isotopic composition is a key tool to understand Earth's geological history and processes, such as continental formation, magmatic-hydrothermal processes, and crust–mantle interactions. The oxygen isotopic analysis is commonly implemented by Isotope Ratio Mass Spectrometry (IRMS) and Secondary Ion Mass Spectrometry (SIMS); however, its wide application is limited by the high cost and serious matrix effect. LA-MC-ICP-MS has been the method of choice for in situ isotopic analysis due to its relatively low cost, high analysis speed, high spatial resolution, and the low matrix effect. The determination of oxygen isotope using Ar-ICP has two limitations. Firstly, the exposure of Ar-ICP to the atmosphere may result in atmospheric interference, leading to an increase in the blank of oxygen isotopes and a reduction in the signal-to-blank (S/B) ratio. Secondly, the presence of doubly-charged Ar ions introduces interference that affects the accuracy of oxygen isotopic analysis. In order to investigate whether MC-ICP-MS can be used in the determination of oxygen isotopes, we attempt to use three approaches (18O16O/16O16O, 18O/16O and 18O1H2/16O1H2) to determine oxygen isotopes in oxygen, and the applicability of three approaches is assessed based on interference, peak width, sensitivity, and stability. With our built methods, the obtained long-term productivity of δ18O measured by 16O18O/16O16O was greater than 0.16‰ (2 SD), and the measured results for oxygen were consistent with those obtained by IRMS and MC-MIP-MS within the uncertainty limit. This demonstrates the feasibility of our method and also lays the foundation for the realization of in situ oxygen isotope analysis using LA-MC-ICP-MS.


1. Introduction

Oxygen is a major component of crustal and mantle rocks and fluids and is the most abundant element in the silicate Earth. Oxygen isotopes are considered an effective means to study geological processes1 and have been widely used in frontier scientific research fields, such as continental formation and evolution, early Earth history, magmatic-hydrothermal processes and metallogeny, crust–mantle composition and interactions, paleo-oceanic and paleoclimatic changes, and planetary geology and cosmogony.2–14 Traditionally, two primary methods are commonly applied to analyze the oxygen isotopes: Isotope Ratio Mass Spectrometry (IRMS) and Secondary Ion Mass Spectrometry (SIMS).15–22

The IRMS analysis method is a bulk analysis method for oxygen isotopes and includes the conventional BrF5 method23–25 and laser microprobe BrF5 method.26–29 The conventional BrF5 method is a reliable technique for accurate oxygen isotope analysis of whole rocks and individual minerals, including silicates, phosphates, and sulfates. It typically requires a sample volume of 5–15 mg to react with excess BrF5 in nickel reaction tubes at specified temperatures and durations. Although the analytical accuracy of this method can reach ±0.05–0.1‰, certain refractory minerals demand higher temperatures and a longer duration for sample melting. For instance, garnet and olivine necessitate a high temperature of 690 °C maintained for 12 hours.23,25 In contrast, the laser microprobe BrF5 method replaces external heating of nickel tubes with direct laser heating, which reaches a high temperature of approximately 2000 K. This innovation enables oxygen isotopic analysis of some refractory minerals, significantly reduces sample volume requirements to less than 100 μg and also maintains good analytical precision (±0.05–0.1‰).29–31 However, the thermal effects occurring during the laser heating process can lead to significant isotopic fractionation, thereby influencing the precision and accuracy of the analysis.32–36 Furthermore, both methods require the use of strong corrosive reagents, posing potential safety hazards. Compared with the bulk analysis method by IRMS, SIMS can be capable of performing in situ microanalysis of oxygen isotopes, which can offer the high spatial resolution of 5–20 μm beam spot and 1–2 μm depth with high analytical precision (better than 0.4‰).37 However, the analytical results obtained by SIMS are susceptible to a severe matrix effect, i.e., the difference in chemical compositions and morphological characteristics of the standard and sample.21,22,37–39 Additionally, the high maintenance and operation cost has also constrained the widespread use of SIMS for in situ analysis of oxygen isotopes,15,18,20–22,40 and it has only been applied to the in situ oxygen isotopic analysis of the simple single mineral (e.g., pyroxene, zircon barite and olivine).22,40–43 In addition, the high sample preparation requirements for SIMS, which necessitate that samples be compatible with high vacuum conditions, can significantly limit its applicability, especially when dealing with loosely structured and volatile samples.44 In particular, with the development of modern geosciences, more and more studies have shown that in situ oxygen isotope analysis plays a key role in revealing some major geoscientific issues. For example, in situ microanalysis of oxygen isotopes from ancient zircons, targeting pristine domains within individual crystals to avoid later alteration, indicated the existence of a hydrosphere and water–rock interaction on Earth before >4.1 Ga;45,46 oxygen isotopic composition in diamonds and their inclusions, specifically the pristine domains within the crystals, illustrated the presence of cryogenically altered oceanic crust in deep continental areas.5,6,47 High-resolution paleoclimate records obtained through in situ microanalysis of oxygen isotopes can elucidate the relationships between paleoclimate changes, biological explosions, and mass extinctions.3,7,48 Therefore, it is essential to develop new techniques for in situ oxygen isotope analysis with high spatial resolution, high precision, and high accuracy.

Multi-Collector Inductively Coupled Plasma Mass Spectrometry (MC-ICP-MS) has become the method of choice for analyzing the isotopic composition by virtue of its advantages such as high precision and accuracy, fast analysis speed, a weak matrix effect and low operating cost.49–54 In particular, since the ICP operates under atmospheric pressure conditions, it can be flexibly switched among multiple sample introduction methods (traditional nebulizers and spray chambers, membrane desolvation,55,56 laser ablation system49,57–60 as well as the direct introduction of gaseous samples.61,62). Among them, LA-MC-ICP-MS, which combines the advantages of in situ sample introduction by a laser ablation system and the high-precision isotopic determination by MC-ICP-MS, has already accurately analyzed more than 23 isotopes.51 In particular, the successful determination of high-ionization-energy isotopes (e.g., C (10.4 eV),49 S (10.0 eV)63 and Cl (12.97 eV)50) using LA-MC-ICP-MS has encouraged us to explore the in situ microanalysis of oxygen isotopes. Compared with C, S and Cl, O has a higher ionization energy of 13.6 eV, although it is lower than that of Ar (15.6 eV). Oxygen can be ionized in argon inductively coupled plasma (Ar-ICP), but its ionization efficiency is only about 0.1% based on the Saha equation.64 Moreover, two additional challenges exist. First of all, the Ar-ICP is exposed to the atmosphere, and the determination of oxygen isotopes can be interfered with by the oxygen components in the air. Second, as Ar is used as the plasma gas, the precision and accuracy of oxygen isotopic analysis can be severely impacted by the interference from 36Ar2+. Consequently, in view of the inherent characteristics of Ar-ICP, no attempts have been made to employ MC-ICP-MS for oxygen isotopic analysis. Therefore, it is necessary to first explore whether the determination of the oxygen isotopic composition can be achieved by MC-ICP-MS, which is the prerequisite for discussing the applicability of LA-MC-ICP-MS in analyzing the oxygen isotope.

In this study, with the application of the simplest oxygen gas as the analyzing sample, we attempted to determine the oxygen isotope composition of oxygen using MC-ICP-MS by employing three methods, 18O/16O, 16O18O/16O16O, and 18O1H2/16O1H2. The ionization efficiency of oxygen in Ar-ICP was assessed by evaluating the sensitivity of all three analytes. Meanwhile, interference and the peak width were analyzed and quantified to assess the impacts of atmospheric oxygen components and argon. From the perspective of isotopic analysis, we evaluated the isotopic analytical reproducibility, the stability of oxygen isotope ratios, as well as the measurement accuracy. Through systematic investigation, 18O/16O and 16O18O/16O16O were selected for oxygen isotopic analysis. Furthermore, the accuracy of the established method was evaluated by comparing the obtained results with those acquired from MAT 253 and MC-MIP-MS. This demonstrated the feasibility of our method and laid the foundation for the realization of in situ oxygen isotope analysis using LA-MC-ICP-MS.

2. Experimental design

A double-focusing Neptune Plus MC-ICP-MS system (Thermo Fisher Scientific, Germany), equipped with seven ion counters and nine Faraday cups, was used at the State Key Laboratory of Geological Processes and Mineral Resources, China University of Geosciences in Wuhan (GPMR). In our experiment, 18O/16O, 18O16O/16O16O and 18O1H2/16O1H2 were used for oxygen isotopic analysis, in which the high-mass particles (18O, 18O16O, and 18O1H2) were measured on an H4 cup and low-mass particles (16O, 16O16O, and 16O1H2) were measured on an L4 cup, and “dummy” masses (m/z = 16.990, 33.030, 19.042) were measured on the center cup. The H4 Faraday cup was equipped with a 1011 Ω resistor, whereas the L4 Faraday cup was equipped with a 3 × 109 Ω resistor due to the high blank intensity of 16O+ and (16O16O)+. To eliminate the interference of 36Ar2+ in the determination of 18O+ and because two adjacent Faraday cups cannot receive both signals simultaneously (the mass difference between 18O+ and 36Ar2+ is 0.0154 amu), we built a sub-cup configuration for the determination of 36Ar2+, wherein a Faraday cup H4 was used, and a “dummy” mass (m/z = 16.975) was measured in the center cup. In addition to 36Ar2+, there are also interferences from 40Ar2+,32S+ and 34S+ in oxygen isotope measurements, thus the high mass resolution (mm = 5500) was applied to eliminate related polyatomic ion interference. Two tanks of oxygen (>99.999%) were used as the bracketing standard and sample, respectively. The O2 (0.1–2 mL min−1) and He sweeping gas flows were controlled using a mass flow meter (MFC). High-purity He (99.999%) was used as a sweeping gas to clean residual oxygen from pipelines. The schematic of the MC-ICP-MS inlet system in the oxygen isotopic analysis can be seen in Fig. 1.
image file: d5ja00025d-f1.tif
Fig. 1 Schematic of the MC-ICP-MS inlet system in the oxygen isotopic analysis.

The oxygen isotope composition of the oxygen was determined using the sample-standard bracketing method (SSB), a technique designed to correct for mass bias and instrumental drift during isotopic analysis. To minimize the influence of blank on the isotopic measurements,65 the on-peak blank was analyzed prior to each sample and standard measurement, utilizing a block of 15 cycles with an integration time of 4.194 seconds. Analysis of each sample and standard was conducted by one block of 30 cycles, with the integration time of 4.194 s. A repeated analytical sequence of “blank, O2-std, blank, O2-sample, blank, O2-std, blank…” was conducted. And offline calculations of the blank and analyte signals, time-drift correction, and isotopic analysis calibration were performed using Iso-Compass software.66 To verify the accuracy of the proposed method, an IRMS (Thermo Scientific™ MAT253 Plus™) at GPMR was used as a comparative method to analyze oxygen isotopes, and the instrumental operating parameters for MC-ICP-MS and IRMS are summarized in Table 1. The oxygen isotopes of the sample are expressed as δ18O, which can be calculated using eqn (1)–(4).

 
r = (16O18O)+/(16O16O)+(1)
 
r′ = 18O/16O = r/(2 + r)(2)
 
r′ = 18O/16O = (18O1H1H)+/(16O1H1H)+(3)
 
image file: d5ja00025d-t1.tif(4)
where r is the measured 16O18O/16O16O ratio. r′ is the measured 18O/16O ratio (when 18O/16O was measured) or the 18O/16O ratio when converted from 16O18O/16O16O (when 16O18O/16O16O was measured) or the 18O/16O ratio when converted from 18O1H1H/16O1H1H (when 18O1H1H/16O1H1H was measured), image file: d5ja00025d-t2.tif is the measured 18O/16O ratio of sample oxygen, and image file: d5ja00025d-t3.tif is the measured 18O/16O ratio of standard oxygen. δ18O is the oxygen composition of the analyzed sample.

Table 1 Summary of the operating parameters for MC-ICP-MS
Instruments Analyzed method Operating conditions
Neptune Plus MC-MS system
Cup configuration 18O16O/16O16O L4 C H4
16O16O 33.03 16O18O
18O1H1H/16O1H1H 16O1H1H 19.042 18O1H1H
18O/16O 16O 16.990 18O
Sub-cup configuration 36Ar2+ 16.975 36Ar2+
Resistor of the Faraday cup 3 × 109 Ω 1011 Ω 1011 Ω
Cool gas flow rate 15 L min−1
Aux gas flow rate 0.95 L min−1
Sample gas flow rate 1.12–1.45 L min−1
RF power 1050 W
Guard electrode (GE) On
Extraction −2000 V
Focus −676 V
X-Defl 0.27 V
Y-Defl −3.48 V
Shape 248 V
Rot quad 1 0.01 V
Source offset −1.00 V
Foc quad 1 −19.89 V
Rot quad 2 0.00 V
Focus offset 50.00 V
Matsuda plate 0.01 V
Focus quad −8.00 V
Dispersion quad −1.00 V
Interface cones Jet sample cone + X skimmer cone
Mass resolution High (mm = 5500)
Block × cycle 1 × 30 (sample); 1 × 15 (blank)
Integration time 4.194 s
[thin space (1/6-em)]
MAT253 Plus IRMS
Cup configuration 18O16O/16O16O C1 C2 C3
16O16O 16O17O 16O18O
Resistor of the Faraday cup 3 × 108 Ω 3 × 1011 Ω 1 × 1011 Ω
High voltage 9.450 KV
Emission 1.2 mA
Electron energy 108.926 V
Trap 22.6 V


In our experiment, we integrated the existing oxygen isotopic analysis methods of SIMS, which determines negative ions such as 18O/16O, as well as IRMS, which analyzes positive oxygen molecules like 16O18O+/16O16O+, and combined the determination methods of other isotopes by MC-ICP-MS, including positive ions like 13C+/12C+, 32S+/34S+, and 37Cl+/35Cl+. Based on these, a comprehensive evaluation of three possible methods (18O/16O, 18O16O/16O16O and 18O1H2/16O1H2) was conducted to explore the feasibility of using MC-ICP-MS for oxygen isotopic analysis.

3. Results and discussion

Given that oxygen isotopic analysis via MC-ICP-MS has never been explored before, we referred to the determination methods of other isotopes using MC-ICP-MS (determination of positive ions, e.g., 13C+/12C+,4932S+/34S+,63,67 and 37Cl+/35Cl+)50 as well as SIMS (determination of negative ions, 18O/16O)22 and IRMS (determination of positive oxygen molecules, 16O18O+/16O16O+).68 Taking into account the high ionization capacity of ICP as well as the lens parameter in our instrument (with the extraction lens set at −2000 V), we opted to select 18O+16O+ and 16O18O+16O16O+ for oxygen isotopic analysis. Additionally, during the process of scanning the blank, we also detected H2O+. Hence, currently, there are three analytes available for oxygen isotopic analysis. To explore the feasibility of MC-ICP-MS to determine oxygen isotopes, a comprehensive evaluation of three possible methods was conducted, including the interference, sensitivity, dynamic linear range, stability of oxygen isotope ratios, analytical precision and accuracy of oxygen isotope ratios.

3.1 Interference on oxygen ions

The peak shapes of 18O–16O, 16O18O–16O16O and 18O1H216O1H2 as well as the related interferences are scanned at an O2 flow rate of 0 mL min−1 with a high mass resolution (mm = 5000) (Fig. 2). For the measurement of 18O/16O, 16O18O/16O16O, and 18O1H2/16O1H2, the main interferences are listed in Table 2. It can be seen that the determination of oxygen isotopes is susceptible to the interference of the plasma gas (Ar2+) (Fig. 2b and e), S blank (Fig. 2c) and the atmosphere (O–H related interference) (Fig. 2a and f). Therefore, it is necessary to select the appropriate peak center to avoid the interference and obtain the true oxygen signal intensity for oxygen isotope determination.
image file: d5ja00025d-f2.tif
Fig. 2 Peak shapes of 16O+18O+ (a), (16O16O)+–(16O18O)+ (c) and (16O1H1H)+–(18O1H1H)+ (e) with high mass resolution in MC-ICP-MS. In the legend of “18Mass × 420”, “34Mass × 200’’ and “20Mass × 500”, 420-, 200- and 500-folds were used to obtain the signal intensities of 18Mass, 34Mass and 20Mass, consistent with those of 16Mass, 18Mass and 32Mass to check the simultaneous collection of the two peaks. Part figures (a, c and e) in the dashed box are enlarged to locate more suitable peak centers (mass = 16.990, 33.030 and 19.042) (b, d and f).
Table 2 Data for analyzed isotopes, related interfering ions, and mass resolutions
Analyzed isotopes Mass Interfering ions Mass Required mass resolution Peak width (amu) Blank (V) Signal-to-blank ratio Sensitivity (V ml−1 min−1)
16O 15.9949 14N1H2+ 16.0187 672 0.008 65.80 (16O) 22.49 2430
18O 17.9992 36Ar2+ 17.9838 1170
16O1H2+ 18.0106 1578
14N1H4+ 18.0344 511
16O16O 31.9898 32S+ 31.9721 1807 0.015 0.80 (16O16O) 1585 719
14N18O 32.0022 2579
14N16O1H2+ 32.0136 1344
16O18O 33.9941 34S+ 33.9679 1297
16O16O1H2+ 34.0055 2981
16O1H2 18.0107 36Ar2+ 17.9838 670 0.008 0.43 (16O1H2) 32.39 7
18O+ 17.9992 1566
14N1H4+ 18.0344 760
18O1H2 20.0132 40Ar2+ 19.9812 625


In terms of 18O–16O, 36Ar2+ and 16O1H2+ are two main interferences. The doubly-charged interference of 36Ar2+ is located on the low-mass side and the polyatomic ionic interference of 16O1H2+ is located on the high-mass side. The interference of 16O1H2+ on 18O+ could be isolated as the mass center of the peak was chosen at the mass side of 16.990 with the high mass resolution (Fig. 2a and b). And for 36Ar2+ (the mass difference between 18O+ and 36Ar2+ is 0.0154 amu, and two nearby Faraday cups cannot receive both signals at the same time), a sub-cup configuration was built, and the 36Ar2+ signal can be directly acquired through peak jumping and deducted accurately. As for 16O18O–16O16O, 32S+ and 34S+ are two main interferences, and fortunately both interferences are located on the low-mass side; thus, a mass center of 33.03 can be selected to avoid the interferences of 32S+ and 34S+ (Fig. 2c and d). For the determination of 18O1H216O1H2, the main interferences are 36Ar2+, 18O+ and 40Ar2+. Three interferences are all located on the low mass side, and a mass center of 19.042 was selected to avoid the interferences. Notably, the applied peak width of 16O18O/16O16O was approximately 0.015 amu (Fig. 2f), whereas those of 18O/16O and 18O1H2/16O1H2 were 0.008 amu (Fig. 2b and d). Therefore, 16O18O/16O16O was estimated to be the preferred method in view of the peak width of the mass shoulder, which is crucial for accurate isotope measurement as it allows for better resolution and less interference from adjacent peaks.69,70

3.2 Sensitivity and dynamic linear range of oxygen ions

The Saha equation predicts a low ionization rate (approximately 0.1%) for oxygen in an Ar-ICP,64 making high sensitivity a critical factor for achieving precise and accurate oxygen isotopic analysis. Besides, as the Ar-ICP is exposed to the atmosphere, the oxygen components in air generate a significantly high oxygen blank during isotopic analysis, which affects the accuracy of oxygen isotope measurements. Consequently, a comprehensive assessment of the sensitivity, blank levels, and signal-to-blank (S/B) ratios of the two methods is essential to ascertain their influence on the reliability of isotopic analysis.

The blank and sensitivity (sensitivity was obtained by the ratio of net signal intensity and oxygen gas flow rate) of 16O and 16O16O were examined as the O2 flow rate increased from 0 to 2 mL min−1 (Fig. 3). For the measurement of 18O/16O, the 16O blank can be as high as 64 V, which may be due to the exposure of the ICP in the atmosphere.71 At the sample gas flow rate of 1.29 L min−1 and RF power of 1050 W, the 16O intensity increased as the O2 flow rate increased from 0 to 0.5 mL min−1, as demonstrated by the linear function of intensity versus the O2 flow rate. The slope of the linear regression relationship was 2430, which can be demonstrated as sensitivity (i.e., ∼2430 V mL−1 min−1). However, beyond this flow rate, the 16O signal intensity will exceed the Faraday cup's threshold of 1666 V when using a 3 × 109 Ω resistor. Thus, the applied O2 flow rate was 0.5 mL min−1, and the S/B ratio of this method was 22.49. For the analyzed method of 16O18O/16O16O, at the sample gas flow rate of 1.254 L min−1 and RF power of 1050 W, the 16O16O intensity increased linearly with the O2 flow rate of up to 1.8 mL min−1, beyond which a nonlinear relationship was observed. The slopes of the linear regression relationship were 718, which can be demonstrated as the sensitivity (i.e., ∼680 V mL−1 min−1). This nonlinearity may stem from incomplete ionization of the O2 sample under the current conditions.72,73 To reduce the isotopic fractionation and obtain the best precision, the O2 flow rate of 1.6 mL min−1 was selected for the measurement of 18O16O/16O16O with the S/B ratio of this method being 1585.25.


image file: d5ja00025d-f3.tif
Fig. 3 Relationship between the O2 flow rate and the signal intensity (a) as well as the sensitivity (b) of 16O and (16O16O).

In order to select an appropriate analysis method, we proceeded with two perspectives: sensitivity and S/B ratio. Based on our experiment, 18O/16O is the preferred option in terms of sensitivity (2430 V ml−1 min−1). This observation is somewhat surprising, the first ionization energy of an O atom (13.62 eV) is higher than that of 16O16O (12.07 eV), which suggests that the sensitivity of 16O16O should be greater than that of 16O theoretically. The high sensitivity of 16O may be attributed to the high gas temperature of the ICP ion source, which can lead to the dissociation of most O2 into O atoms, leaving a small fraction of undissociated O2 molecules. However, from the perspective of S/B ratio, the S/N ratio of 16O (22.49) was 750 times lower than that of 16O16O (1585.25). Therefore, when considering the two aspects of sensitivity and S/N ratio, the choices of optimal methods to be tested are contradictory. To explore the interaction between the S/B ratio and signal intensity in isotopic analysis, a simulation experiment was performed. Two simulated datasets were created: one with low signal intensity but high S/B ratios and the other with high signal intensity but low S/B ratios, while keeping the isotope composition of blank (−12‰) and sample (0.5‰) constant. For each dataset, the blank intensity was fixed (0.005 V vs. 0.2 V), and the signal intensity (1–8 V vs. 21–28 V) was varied to achieve different S/B ratios (200–1600 vs. 105–140). Isotopic compositions were calculated without blank correction, and the relative error (RE) between the calculated and true sample compositions was assessed. Fig. 4a and b shows that for the group with a high signal intensity but low S/B ratio, the blank intensity had a more significant impact on isotopic measurements. For example, even at a signal intensity of 28 V with a low S/B ratio (140), the effect of blank on isotopic composition was considerable, with an RE as high as −29.94%. Conversely, for the high S/B ratio group, where the S/B ratio reached 1000 at a low signal intensity of 5 V, the effect of blank on isotopic composition (RE) was less than 2.5%. Here, despite an 82.1% reduction in signal intensity (from 28 V to 5 V) compared to the high signal intensity group, the effect of the blank can be reduced by 91.6%. Fig. 4c and d further examines the effect of 1% fluctuation in blank intensity. The results indicated that such fluctuations affected the high signal intensity but a low S/B ratio group more than the low signal intensity but a high S/B ratio group. At an S/B ratio of 1000, a 1% blank fluctuation caused only a 0.03% change in the calculated isotopic composition at a signal intensity of 5 V. However, for the high signal intensity group, the same fluctuation led to a 1.01% change in isotopic composition at 28 V due to the low S/B ratio (140).


image file: d5ja00025d-f4.tif
Fig. 4 (a) and (b) Relationship between the signal intensity, S/B ratio and relative error of the two simulated datasets; the relative error was calculated between the calculated and true sample compositions. (c) and (d) Relationship between the signal intensity, S/B ratio and relative deviation of the two simulated datasets; the relative deviation was assessed by the sample composition before and after 1% blank fluctuation.

These simulations highlight the crucial balance between the signal intensity and S/B ratio in isotope analysis. The higher S/B ratios enhance the accuracy of isotope ratio measurements by reducing blank, which is especially vital in laser-based in situ microanalysis of samples with low elemental concentrations and low signal intensities. Achieving high S/B ratios is essential for reliable measurements in such scenarios. In summary, accurate determination of isotopic compositions necessitates not only sufficient signal intensity but also optimized S/B ratios to ensure high precision and accuracy across a variety of analytical conditions. In this context, 18O16O/16O16O, which exhibits a higher S/B ratio, is more suitable as the analyte of analysis. Additionally, the peak width of 18O16O/16O16O (0.015 amu) is wider than that of 18O/16O (0.008 amu) and the blank intensity of 16O16O (0.8 V) is lower than that of 16O (65.8 V). Therefore, considering the S/B ratio, peak width and blank, the use of 18O16O/16O16O is anticipated to be a more favorable choice for isotopic analysis.

3.3 Effect of sample gas and sampling depth on oxygen isotopic analysis

Accurate calibration of isotopic mass discrimination is essential for achieving precise isotope ratios with MC-ICP-MS.53 For the oxygen isotope, the lighter weight and lack of stable isotope pairs as internal standards determine that the standard-sample bracketing (SSB) method can only be used. This method is simple and convenient to operate; however, a higher level of stability for the instruments is required because the fractionation factor of standard is assumed to be that of sample with this method.74 The region of stable O isotopic composition can be explored by studying the radial and axial distributions of O ions in the ICP, which can be confirmed by determining the variation of O signal intensity and 18O/16O ratio with respect to the position of the torch tube (radial and axial directions) and the sample gas flow rate. It has been established that the radial position of the torch that provides the maximum analyte intensity also results in the most stable isotopic ratios.75,76 In contrast, the axial position of the torch and sample gas flow rate may not be optimal for both the maximum signal intensity and stable isotopic ratios.52 The axial distribution of ions in the ICP can be achieved by changing the position of the Z-axis of the torch tube (Fig. 5a and c) or the sample gas flow rate (Fig. 5b and d). The Z-axis position is the distance from the outer end of the torch tube to the hole of the sampling cone. A more negative Z-axis value indicates the plasma is closer to the sample cone. The sample gas flow rate was also manipulated to investigate the effect on ion distribution at the axial position, as a higher flow rate means that the plasma is closer to the sample cone.
image file: d5ja00025d-f5.tif
Fig. 5 Relationship between the Z position (a and c) and sample gas flow rate (b and d) of (16O)+, (16O16O)+ signal intensity and oxygen isotope ratio (18O/16O).

For investigation of the Z-axis position, as the Z-axis position varied from −1.7 mm to 2.0 mm, with the sample gas flow rate remaining constant, the signal intensity of the two analytes initially increased and then decreased, following a similar trend. Comparing the stability of the obtained 18O/16O ratio within ±10% of the highest signal intensity, the stability of the 18O/16O ratio obtained by measuring 18O16O/16O16O (0.002291 ± 0.000001, 2SD) is about 20 times higher than that of 18O/16O (0.002222 ± 0.000023, 2SD). To further explore the effect of the sample gas flow rate, the Z-position was kept constant, and as the sample gas flow rate increased from 1.0 L min−1 to 1.25 L min−1, more ions were extracted into the mass spectrometer, leading to an increase in O intensity. However, the amount of ions extracted is limited by the sample cone diameter77 and the temperature of the ICP;78,79 thus, the O signal will not continue to rise indefinitely as the carrier gas flow rate increases. In terms of isotope ratio, the obtained 18O/16O by measuring 18O16O/16O16O is more stable (0.002289 ± 0.000002, 2SD) than that obtained by measuring 18O/16O (0.002251 ± 0.000019, 2SD). Additionally, both the optimum sampling depth and sample gas flow rate for the maximum signal intensity of 16O16O can be aligned with the stable area of the isotope ratio, which is superior for tuning and isotopic analysis. Therefore, in terms of isotope ratio stability, compared with the 18O/16O ratio, the measured 18O16O/16O16O ratio is the preferred choice. And, in practical oxygen isotopic analysis, it is advisable to adjust the sample gas flow rate and the Z-axis position to where the maximum signal intensity is achieved.

3.4 Precision and accuracy of oxygen isotopes in oxygen

Analytical precision is dependent on the signal intensity,52,80 as demonstrated by the exponential function of the precision for the 18O/16O isotopic ratios (1 SE) vs. the 16O+ and (16O16O)+intensities (Fig. 6a). The internal precision of 18O/16O was low when the (16O16O)+ and 16O+ signal intensities were less than 100 V. As the signal intensity increases from 100 to 1000 V, there is a rapid improvement in the within-run precision. Therefore, using (16O16O)+ and 16O+ signal intensities of more than 1000 V, where the internal precision of 18O/16O was better than 1 × 107, a 3 × 109 Ω resistor was recommended. To obtain long-term reproducibility, the δ18O value was repeatedly measured by analyzing 16O18O/16O16O and 18O/16O with an optimum O2 flow rate, respectively. The δ18O value was repeatedly measured by analyzing 16O18O/16O16O and 18O/16O with O2 flow rates of 1.6 and 0.5 mL min−1, sample gas flow rates of 1.254 and 1.285 L min−1, and Z-axis positions of 0.1 and −0.6 mm, respectively, and an RF power of 1050 W. The obtained δ18O values were 0.05 ± 0.14‰ (2 SD, n = 50) and 0.07 ± 0.17‰ (2 SD, n = 50) for 16O18O/16O16O and 18O/16O, respectively (Fig. 6b), and 18O16O/16O16O was preferred.
image file: d5ja00025d-f6.tif
Fig. 6 (a) Relationship between the signal intensity of 16O, (16O16O) and the internal precision (1 SE) of 18O/16O for oxygen isotopic analysis. (b) Long-term reproducibility of δ18O by measuring 16O18O/16O16O and 18O/16O.

For the assessment of measurement accuracy of δ18O using the proposed method, the oxygen isotopes of two different oxygen tanks were analyzed. One oxygen tank was used as the bracketing standard and the other one as the sample. The δ18O values measured by 18O and 16O were determined at a sample gas flow rate of 1.285 L min−1, an oxygen flow rate of 0.5 ml min−1, and a Z-axis position of −0.6 mm. And the δ18O values measured by 18O16O and 16O16O were determined at a sample gas flow rate of 1.254 L min−1, an oxygen flow rate of 1.6 ml min−1, and a Z-axis position of 0.1 mm. The δ18O values of sample-1 were −0.03 ± 0.22‰ (2 SD, n = 10, measured by 18O and 16O) and 0.16 ± 0.16‰ (2 SD, n = 10, measured by 18O16O and 16O16O). In terms of test accuracy, the accuracy of using 18O16O/18O16O to determine the δ18O value in oxygen was preferred. The obtained values were consistent with those measured according to MAT 253 (0.13 ± 0.1‰; 2 SD, n = 10) and MC-MIP-MS (0.14 ± 0.22‰; 2 SD, n = 6),81 as shown in Fig. 7, which substantiated the validity of our method.


image file: d5ja00025d-f7.tif
Fig. 7 Measured δ18O values of oxygen samples using MC-ICP-MS, IR-MS and MC-MIP-MS.

4. Conclusions

Due to the interference from Ar plasma gas, high blank intensity of O from the atmosphere, and the high first ionization energy, no attempt has been made to determine oxygen isotopes by MC-ICP-MS. However, high mass resolution and sub-cup configuration can be used to eliminate the interference of 36Ar2+ on 18O+; the application of a 3 × 109 Ω amplifier can be used to obtain the true signal intensity of O blank. Additionally, the applied 18O16O/16O16O method can be used to overcome the interference of Ar2+ and the ultra-high blank intensity of 16O. Based on the above improvement measures, the obtained long-term reproducibility of δ18O measured by 18O16O/16O16O was better than 0.16‰ (2 SD). The results obtained by MC-ICP-MS were consistent with those measured by IRMS and MC-MIP-MS, which demonstrated that MC-ICP-MS can be used for the accurate determination of oxygen isotopes. It is worth noting that the obtained precision by MC-ICP-MS was worse than that obtained by IRMS, which may be attributed to the high blank. Fortunately, with the use of a shielding gas to maintain positive pressure, preventing atmospheric air from entering the ICP area has been proven to reduce interferences caused by gases such as H, C, N, and O.82 Using the Atmospheric Interface for Reduced Dust (AIRD) device, optimized through simulations, significantly enhances the analytical capabilities of ICP-MS and effectively manages atmospheric interference. Another reason may be the low ionization rate of O in the Ar-ICP, so the stable He-ICP with high ionization ability may be promising. Therefore, there is still a long way to go for in situ microanalysis of oxygen isotopes using LA-MC-ICP-MS. However, we have already taken the first step to evaluate the capability of MC-ICP-MS to analyze oxygen. Further work will be continued to achieve this goal.

Data availability

The authors declare that the data supporting the findings of this study are available within the paper. Should any raw data files be needed in another format they are available from the corresponding author upon reasonable request. Source data are provided with this paper.

Conflicts of interest

There are no conflicts to declare.

Acknowledgements

This work was funded by the National Natural Science Foundation of China (41927803 and 42473038), MOST (Ministry of Science and Technology) Special Fund from the State Key Laboratory of Geological Processes and Mineral Resources (MSFGPMR01), and the Fundamental Research Funds for the Central Universities, China University of Geosciences (Wuhan).

References

  1. J. Hoefs, Stable Isotope Geochemistry, Springer International Publishing, Cham, 2015 Search PubMed .
  2. B. Dhuime, C. J. Hawkesworth, P. A. Cawood and C. D. Storey, Science, 2012, 335, 1334–1336 CrossRef CAS PubMed .
  3. J. A. Trotter, I. S. Williams, C. R. Barnes, C. Lécuyer and R. S. Nicoll, Science, 2008, 321, 550–554 CrossRef CAS PubMed .
  4. C. J. Hawkesworth and A. I. S. Kemp, Nature, 2006, 443, 811–817 CrossRef CAS PubMed .
  5. R. B. Ickert, T. Stachel, R. A. Stern and J. W. Harris, Geochem. Perspect. Lett., 2015, 65–74 Search PubMed .
  6. D. J. Schulze, B. Harte, J. W. Valley, J. M. Brenan and D. M. D. R. Channer, Nature, 2003, 423, 68–70 CrossRef CAS PubMed .
  7. J. A. Trotter, I. S. Williams, A. Nicora, M. Mazza and M. Rigo, Earth Planet. Sci. Lett., 2015, 415, 165–174 CrossRef CAS .
  8. U. Ryb and J. M. Eiler, Proc. Natl. Acad. Sci. U. S. A., 2018, 115, 6602–6607 CrossRef CAS PubMed .
  9. I. N. Bindeman, A. Bekker and D. O. Zakharov, Earth Planet. Sci. Lett., 2016, 437, 101–113 CrossRef CAS .
  10. U. Wiechert, A. N. Halliday, D.-C. Lee, G. A. Snyder, L. A. Taylor and D. Rumble, Science, 2001, 294, 345–348 CrossRef CAS PubMed .
  11. R. C. Greenwood, J.-A. Barrat, M. F. Miller, M. Anand, N. Dauphas, I. A. Franchi, P. Sillard and N. A. Starkey, Sci. Adv., 2018, 4, eaao5928 CrossRef PubMed .
  12. I. N. Bindeman, D. O. Zakharov, J. Palandri, N. D. Greber, N. Dauphas, G. J. Retallack, A. Hofmann, J. S. Lackey and A. Bekker, Nature, 2018, 557, 545–548 CrossRef CAS PubMed .
  13. Y.-F. Zheng, B. Fu, B. Gong and L. Li, Earth-Sci. Rev., 2003, 62, 105–161 CrossRef CAS .
  14. G. A. Henkes, B. H. Passey, E. L. Grossman, B. J. Shenton, T. E. Yancey and A. Pérez-Huerta, Earth Planet. Sci. Lett., 2018, 490, 40–50 CrossRef CAS .
  15. Q.-L. Li, X.-H. Li, Y. Liu, G.-Q. Tang, J.-H. Yang and W.-G. Zhu, J. Anal. At. Spectrom., 2010, 25, 1107–1113 RSC .
  16. D. Rubatto, B. Putlitz, L. Gauthiez-Putallaz, C. Crépisson, I. S. Buick and Y.-F. Zheng, Chem. Geol., 2014, 380, 84–96 CrossRef CAS .
  17. D. Liu, S. A. Wilde, Y. Wan, S. Wang, J. W. Valley, N. Kita, C. Dong, H. Xie, C. Yang, Y. Zhang and L. Gao, Chem. Geol., 2009, 261, 140–154 CrossRef .
  18. Y. Sun, M. Wiedenbeck, M. M. Joachimski, C. Beier, F. Kemner and C. Weinzierl, Chem. Geol., 2016, 440, 164–178 Search PubMed .
  19. A. K. Schmitt, M.-C. Liu and I. E. Kohl, J. Anal. At. Spectrom., 2019, 34, 561–569 Search PubMed .
  20. N. M. Ellis and B. H. Passey, Chem. Geol., 2023, 635, 121616 CrossRef CAS .
  21. S. Schmidt, A. Hertwig, A. K. Schmitt, K. Cionoiu, K. D. McKeegan, I. Bindeman, T. D. Rocco and A. Pack, J. Anal. At. Spectrom., 2024, 39, 439–446 RSC .
  22. B. Li, M. Wiedenbeck, F. Couffignal, A. M. Álvarez-Valero, H.-M. Bao, C.-F. Fan, J. Han, G.-S. Jin, Y.-B. Peng, M. D. Syczewski, K. T. Tait, F. D. H. Wilke and U. G. Wortmann, Geostand. Geoanal. Res., 2024, 48, 179–205 CrossRef CAS .
  23. R. N. Clayton and T. K. Mayeda, Geochim. Cosmochim. Acta, 1963, 27, 43–52 CrossRef CAS .
  24. G. D. Garlick and S. Epstein, Geochim. Cosmochim. Acta, 1967, 31, 181–214 CrossRef CAS .
  25. H. P. Taylor Jr and S. Epstein, Geol. Soc. Am. Bull., 1962, 73, 461–480 CrossRef .
  26. D. Elsenheimer and J. W. Valley, Chem. Geol. Isot. Geosci. Sect., 1992, 101, 21–42 CAS .
  27. U. Wiechert and J. Hoefs, Geochim. Cosmochim. Acta, 1995, 59, 4093–4101 CrossRef CAS .
  28. Z. D. Sharp, Geochim. Cosmochim. Acta, 1990, 54, 1353–1357 CrossRef CAS .
  29. D. Elsenheimer and J. W. Valley, Geochim. Cosmochim. Acta, 1993, 57, 3669–3676 CrossRef CAS .
  30. D. Mattey, D. Lowry and C. Macpherson, Earth Planet. Sci. Lett., 1994, 128, 231–241 CrossRef CAS .
  31. J. W. Valley, J. R. Chiarenzelli and J. M. McLelland, Earth Planet. Sci. Lett., 1994, 126, 187–206 CrossRef CAS .
  32. S. M. Eggins, R. L. Rudnick and W. F. McDonough, Earth Planet. Sci. Lett., 1998, 154, 53–71 CrossRef CAS .
  33. Z. Li, Z. Hu, Y. Liu, S. Gao, M. Li, K. Zong, H. Chen and S. Hu, Chem. Geol., 2015, 400, 11–23 CrossRef CAS .
  34. H.-R. Kuhn and D. Günther, J. Anal. At. Spectrom., 2004, 19, 1158–1164 RSC .
  35. M. Guillong, H.-R. Kuhn and D. Günther, Spectrochim. Acta, Part B, 2003, 58, 211–220 Search PubMed .
  36. Z. Chen, J. Anal. At. Spectrom., 1999, 14, 1823–1828 CAS .
  37. N. T. Kita, T. Ushikubo, B. Fu and J. W. Valley, Chem. Geol., 2009, 264, 43–57 CAS .
  38. G.-Q. Tang, X.-H. Li, Q.-L. Li, Y. Liu, X.-X. Ling and Q.-Z. Yin, J. Anal. At. Spectrom., 2015, 30, 950–956 Search PubMed .
  39. M. J. Whitehouse and A. A. Nemchin, Chem. Geol., 2009, 261, 32–42 Search PubMed .
  40. Q. Yang, X. Xia, W. Zhang, Y. Zhang, B. Xiong, Y. Xu, Q. Wang and G. Wei, Solid Earth Sci., 2018, 3, 81–86 Search PubMed .
  41. J. Isa, I. E. Kohl, M.-C. Liu, J. T. Wasson, E. D. Young and K. D. McKeegan, Chem. Geol., 2017, 458, 14–21 CAS .
  42. B. Peng, M. He, M. Yang and Y. Shi, Crystals, 2023, 13, 987 CAS .
  43. F. M. Deegan, M. J. Whitehouse, V. R. Troll, D. A. Budd, C. Harris, H. Geiger and U. Hålenius, Chem. Geol., 2016, 447, 1–10 CAS .
  44. B. De Samber, R. De Rycke, M. De Bruyne, M. Kienhuis, L. Sandblad, S. Bohic, P. Cloetens, C. Urban, L. Polerecky and L. Vincze, Anal. Chim. Acta, 2020, 1106, 22–32 CAS .
  45. S. J. Mojzsis, T. M. Harrison and R. T. Pidgeon, Nature, 2001, 409, 178–181 CAS .
  46. D. Trail, P. Boehnke, P. S. Savage, M.-C. Liu, M. L. Miller and I. Bindeman, Proc. Natl. Acad. Sci. U. S. A., 2018, 115, 10287–10292 CAS .
  47. A. D. Burnham, A. R. Thomson, G. P. Bulanova, S. C. Kohn, C. B. Smith and M. J. Walter, Earth Planet. Sci. Lett., 2015, 432, 374–380 CAS .
  48. C. Huang, M. M. Joachimski and Y. Gong, Earth Planet. Sci. Lett., 2018, 495, 174–184 Search PubMed .
  49. W. Chen, J. Lu, S.-Y. Jiang, K.-D. Zhao and D.-F. Duan, Anal. Chem., 2017, 89, 13415–13421 CAS .
  50. C. Toyama, J.-I. Kimura, Q. Chang, B. S. Vaglarov and J. Kuroda, J. Anal. At. Spectrom., 2015, 30, 2194–2207 CAS .
  51. J. Lin, A. Yang, R. Lin, J. Mao, Z. Hu and Y. Liu, J. Earth Sci., 2023, 34, 1663–1691 CAS .
  52. J. Lin, Y. Liu, X. Tong, L. Zhu, W. Zhang and Z. Hu, J. Anal. At. Spectrom., 2017, 32, 834–842 CAS .
  53. L. Yang, Mass Spectrom. Rev., 2009, 28, 990–1011 CAS .
  54. V. Balaram, W. Rahaman and P. Roy, Geosystems Geoenvironment, 2022, 1, 100019 Search PubMed .
  55. M.-Y. He, L. Deng, J. Liu, Z. dong Jin and T. Ren, RSC Adv., 2023, 13, 32104–32109 CAS .
  56. T.-L. Yu, B.-S. Wang, C.-C. Shen, P.-L. Wang, T. F. Yang, G. S. Burr and Y.-G. Chen, Anal. Chim. Acta, 2017, 988, 34–40 CrossRef CAS PubMed .
  57. Z. Bao, C. Zong, L. Chen, D. Lei, K. Chen and H. Yuan, J. Anal. At. Spectrom., 2018, 33, 2143–2152 RSC .
  58. W. Zhang, Z. Hu, Y. Liu, L. Feng and H. Jiang, Chem. Geol., 2019, 522, 16–25 CrossRef CAS .
  59. J. Lin, Y. Liu, Z. Hu, W. Chen, C. Zhang, K. Zhao and X. Jin, J. Anal. At. Spectrom., 2019, 34, 1145–1153 RSC .
  60. W. Zhang, Z. Wang, F. Moynier, E. Inglis, S. Tian, M. Li, Y. Liu and Z. Hu, J. Anal. At. Spectrom., 2019, 34, 1800–1809 RSC .
  61. I. Gunther-Leopold, B. Wernli, Z. Kopajtic and D. Gunther, Anal. Bioanal. Chem., 2004, 378, 241–249 CrossRef PubMed .
  62. E. M. Krupp and O. F. X. Donard, Int. J. Mass Spectrom., 2005, 242, 233–242 CrossRef CAS .
  63. Q. Li, W. Zhang, Z. Hu, Y. Feng, H. Hu, T. Luo, J. Huang and X. Zeng, Chem. Geol., 2023, 636, 121651 CAS .
  64. H. Niu and R. S. Houk, Spectrochim. Acta, Part B, 1996, 51, 779–815 Search PubMed .
  65. C. T. Gerritzen, S. Goderis, H. F. James and C. Snoeck, Spectrochim. Acta, Part B, 2024, 217, 106955 CAS .
  66. W. Zhang, Z. Hu and Y. Liu, J. Anal. At. Spectrom., 2020, 35, 1087–1096 Search PubMed .
  67. J. Fu, Z. Hu, W. Zhang, L. Yang, Y. Liu, M. Li, K. Zong, S. Gao and S. Hu, Anal. Chim. Acta, 2016, 911, 14–26 CrossRef CAS PubMed .
  68. N. M. Ellis and B. H. Passey, Chem. Geol., 2023, 635, 121616 CrossRef CAS .
  69. N. Dauphas, A. Pourmand and F.-Z. Teng, Chem. Geol., 2009, 267, 175–184 CrossRef CAS .
  70. Y. He, S. Ke, F. Teng, T. Wang, H. Wu, Y. Lu and S. Li, Geostand. Geoanal. Res., 2015, 39, 341–356 CrossRef CAS .
  71. X. Yan, T. Tanaka and H. Kawaguchi, Appl. Spectrosc., 1996, 50, 182–187 CrossRef CAS .
  72. M. Wu, Y. Duan, Q. Jin and G. M. Hieftje, Spectrochim. Acta, Part B, 1994, 49, 137–148 CrossRef .
  73. E. V. Polyakova, Yu. N. Nomerotskaya and A. I. Saprykin, J. Anal. Chem., 2020, 75, 474–478 CrossRef CAS .
  74. N. J. Pearson, W. L. Griffin and S. Y. O'Reilly, in Laser Ablation-ICP-MS In The Earth Sciences, ed. P. Sylvester, Mineralogical Association of Canada, Québec, 2008, pp. 93–116 Search PubMed .
  75. H. Andrén, I. Rodushkin, A. Stenberg, D. Malinovsky and D. C. Baxter, J. Anal. At. Spectrom., 2004, 19, 1217–1224 RSC .
  76. J. Barling and D. Weis, J. Anal. At. Spectrom., 2012, 27, 653 RSC .
  77. M. M. Fraser and D. Beauchemin, Spectrochim. Acta, Part B, 2000, 55, 1705–1731 CrossRef .
  78. C. Agatemor and D. Beauchemin, Anal. Chim. Acta, 2011, 706, 66–83 CrossRef CAS PubMed .
  79. A. A. Mills, J. H. Macedone and P. B. Farnsworth, Spectrochim. Acta, Part B, 2006, 61, 1039–1049 CrossRef .
  80. Z. Hu, Y. Liu, S. Gao, W. Liu, W. Zhang, X. Tong, L. Lin, K. Zong, M. Li, H. Chen, L. Zhou and L. Yang, J. Anal. At. Spectrom., 2012, 27, 1391 Search PubMed .
  81. J. Lin, Z. Liu, Y. Liu, W. Liu, X. Jiang, G. Zhao, L. Chen and Z. Hu, Anal. Chem., 2023, 95, 16877–16884 Search PubMed .
  82. X. Jiang, Y. Liu, W. Liu, J. Lin, Z. Liu, L. Chen, X. Zhu, W. Zhang and Z. Hu, J. Anal. At. Spectrom., 2024, 39, 2452–2460 RSC .

This journal is © The Royal Society of Chemistry 2025
Click here to see how this site uses Cookies. View our privacy policy here.