Needle tip-enhanced laser-induced breakdown spectroscopy for nebulized aqueous solution analysis

Abstract

In the detection of aqueous solutions through Laser-induced Breakdown Spectroscopy (LIBS), the implementation of nebulization is instrumental in mitigating liquid splashing and preventing plasma quenching. However, gas dilution during nebulization would greatly undermine the detection limits. This study proposes a needle tip-enhancement method that excites the metal tip to produce a high-temperature plasma before ablating the nebulized solution. The optimization results for needle tip materials reveal that among the seven materials Zn, Ag, Cu, Mo, Ni, Ti, and W, Zn, Ag, and Cu exhibit minimal interference with the target elements Mn and Cr, with Zn showing the highest spectral enhancement effect. The average Pearson correlation between the enhancement effect and the melting and boiling points of the materials exceeds 0.90. Further optimization of the needle tip diameter revealed that a diameter that is too thin can cause material deformation, while a diameter that is too thick would lead to liquid adsorption and deep pit formation. Therefore, a diameter of 0.5 mm was chosen as the optimal under experimental conditions. Additionally, the optimization results of the relative position between the aerosol plume and the needle tip indicate that the best enhancement effect occurs when the aerosol plume passes through the core of the plasma. Using a needle tip-enhancement method, we detected Cu, Mn, and Cr in water samples, applying machine learning for quantitative analysis. The Gaussian Process Regression (GPR) model yielded the highest accuracy at 0.99. The spectral intensities of Cu, Mn, and Cr were enhanced by 7.20, 8.57, and 11.49 times, respectively. Detection limits of Cu, Mn, and Cr with needle tip-enhancement reached 42.80 mg L−1, 51.25 mg L−1, and 58.40 mg L−1. This study has the potential to facilitate real-time online monitoring of water solutions.

Graphical abstract: Needle tip-enhanced laser-induced breakdown spectroscopy for nebulized aqueous solution analysis

Article information

Article type
Paper
Submitted
22 Oct 2024
Accepted
16 Dec 2024
First published
28 Dec 2024

J. Anal. At. Spectrom., 2025,40, 503-512

Needle tip-enhanced laser-induced breakdown spectroscopy for nebulized aqueous solution analysis

Y. Xiong, W. Huang, J. Yang, J. Nie and L. Guo, J. Anal. At. Spectrom., 2025, 40, 503 DOI: 10.1039/D4JA00381K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements