Pan
Chang‡
ad,
Xiaoxiao
Li‡
ad,
Xing
Liu
d,
Yi
Kang
d,
Deying
Gong
d,
Wenwen
Li
b,
Zhongjun
Zhao
c,
Tao
Zhu
ad,
Jin
Liu
ad and
Wen-sheng
Zhang
*ad
aDepartment of Anesthesiology, West China Hospital, Sichuan University, Chengdu, China. E-mail: zhang_ws@scu.edu.cn
bWest China School of Public Health and West China Fourth Hospital, Sichuan University, Chengdu, China
cSchool of Mechanical Engineering, Sichuan University, Chengdu, China
dLaboratory of Anesthesia and Critical Care Medicine, National-Local Joint Engineering Research Centre of Translational Medicine of Anesthesiology, West China Hospital, Sichuan University, Chengdu, China
First published on 28th February 2025
Propofol, a fast-acting anesthetic, requires precise titration to minimize adverse effects. While plasma-based monitoring is slow, exhaled propofol offers a real-time, non-invasive alternative, though its clinical application remains limited. This study evaluates ultraviolet time-of-flight mass spectrometry (UV-TOF MS) for real-time monitoring, presenting its calibration and validation in Beagle dogs. Calibration showed excellent linearity (R2 = 0.9939) over 3.23–46.13 ppbv. The intra-day imprecision at propofol concentrations of 4.61 and 23.06 ppbv was below 5.83% and 7.75%, respectively, while the inter-day imprecision was 9.69% and 9.75%, respectively. Carry-over effects were minimal, with signal recovery within 40–60 s, measuring 8.7%, 9.1%, and 4.7% at 4.61, 9.30, and 23.06 ppbv, respectively. In Beagle dogs, Cexhaled exhibited a moderately strong linear correlation with Cplasma (R2 = 0.7950) and a moderate correlation with sedative effects, as indicated by the bispectral index (R2 = 0.5501) after a single bolus injection. Pharmacokinetic (PK) analysis revealed a delay in peak concentration (Tmax) for Cexhaled (2.00 ± 0.21 min) compared to Cplasma (1.00 ± 0.00 min). While AUC values were not directly comparable, both exhibited R_AUC > 80%, indicating reliable drug kinetic reflection. Mean residence time (MRT) and elimination rate constants (λz) showed no significant differences. These results suggest that exhaled breath analysis provides pharmacokinetic insights comparable to plasma, with a slight delay in peak concentration. UV-TOF MS proved to be an efficient method for detecting exhaled propofol, offering potential for real-time anesthesia monitoring in clinical settings.
The traditional method for monitoring propofol concentration relies on plasma measurement using liquid chromatography-mass spectrometry. However, this approach is offline, time-consuming, and requires bulky, expensive equipment, making it unsuitable for point-of-care applications.11 Recent studies have explored the detection of propofol in exhaled breath, highlighting the potential of online monitoring to significantly improve time resolution for drug detection. These studies demonstrated strong correlations between propofol concentrations in exhaled air, brain tissue, and plasma.12–14 Real-time monitoring of exhaled propofol could address the need for maintaining effective dosing, enabling more precise and personalized patient care, particularly during total intravenous anesthesia.15
There is still a lack of effective techniques for real-time monitoring of propofol. Key factors for online monitoring of exhaled drugs include sensitivity, selectivity, sampling speed, and robustness. Exhaled target molecules must be ionized and separated in the gas phase by a mass analyzer. To address this, our team has developed a novel online mass spectrometer that integrates ultraviolet (UV) ionization with time-of-flight mass spectrometry (TOF MS). UV photoionization, a soft ionization technique, excites molecules to a high-energy state where they absorb photons and become ionized, producing charged ions with minimal fragmentation.16,17 The system completes ionization and detection of all volatile organic compounds (VOCs) within 0.1 s, significantly reducing analysis time and enabling real-time monitoring. UV-TOF MS also facilitates high-throughput, comprehensive tracing of gaseous VOCs,18 effectively identifying isomers and rapidly generating visual spectra, providing anesthesiologists with quick and visualized results.
To address the limitations of plasma-based propofol monitoring, this study aims to develop and validate a real-time, non-invasive method for exhaled propofol detection using ultraviolet time-of-flight mass spectrometry (UV-TOF MS). The UV-TOF MS system (Aliben Science & Technology Co., Ltd, Chengdu, China) comprises key components, including a UV photoionization chamber, a radiofrequency multipole ion guide, electrostatic lenses, and a TOF mass analyzer. The system operates at a vacuum pressure of 500 ± 20 Pa maintained by a pump. By integrating UV photoionization with high-resolution mass spectrometry, this system enables rapid and efficient ionization of volatile organic compounds with minimal fragmentation. A key innovation is the incorporation of a fully automated reference gas generator for calibration, ensuring accurate quantification of exhaled propofol. Through a series of validation experiments and pharmacokinetic analysis in Beagle dogs, this study evaluates the feasibility, precision, and reliability of UV-TOF MS for real-time breath analysis, offering potential applications in clinical anesthesia monitoring.
V1 = nRT ÷ p = (X ÷ 18) × RT ÷ (p × 60) | (1) |
Second, the original stock solution after gasification is defined as the standard propofol gas and the concentration C0 (parts per billion by volume, ppbv) can be calculated according to eqn (2),
![]() | (2) |
Finally, the actual bulk concentration of standard gaseous propofol (Cn) was calculated using a dilution ratio according to eqn (3).
Cn = V1/VN2 × C0 | (3) |
The nominal concentration of propofol in the stock solution was calculated as 9713 ppbv using eqn (2). Standard working solutions with nominal concentrations of 3.23, 4.61, 6.92, 9.23, 23.06, and 46.13 ppbv were prepared by serially diluting the stock solution in HPLC-grade pure water, as detailed in Appendix A, Table S2.† Gradients of propofol standard gas were then generated from these working solutions following the method described in Section 2.3. Each concentration of propofol standard gas was continuously monitored for at least 10 minutes, with a signal acquisition interval of 20 s.
EEGs were recorded to evaluate anesthesia, with BIS values calculated over the temporal-frontal area every 20 s. Anesthesia induction was achieved with tiletamine–zolazepam (Zoletil®50, Virbac, France, 5 mg kg−1). The dogs were tracheally intubated and mechanically ventilated with pure oxygen, maintaining a tidal volume of 10 mL kg−1, a respiratory rate of 20 breaths per min, an inspiratory-to-expiratory ratio of 1:
2, and end-tidal CO2 at 30–40 mm Hg. A single 5 mg per kg bolus of propofol, twice the ED50 for loss of the righting reflex in Beagles, was administered over 5 s. Body temperature was maintained at 36.5 °C ± 0.5 °C. Upon recovery, the endotracheal tube and catheters were removed, and the dogs were returned to the breeding room after full recovery.
Exhaled breath was collected via the UV-TOF MS system, using a polytetrafluoroethylene tracheal tube connected by a polycarbonate three-way valve. The m/z 178 mass peak was selected for propofol detection, with signal intensity monitored every 20 s. Data collection started immediately after propofol administration and ceased once spontaneous respiration resumed. The propofol concentrations were calculated using calibration curves.
The retention times were 7.4 min for propofol and 3.9 min for thymol (internal standard). The propofol standard solution was prepared by dissolving 20 mg of propofol reference substance in methanol to a final concentration of 2.0 mg mL−1 and stored at 4 °C. Before use, the stock solution was diluted with methanol to prepare working solutions with concentrations of 48.87, 97.73, 293.19, 879.58, 2638.75, 7916.25, 10555.00, and 21
110.00 ng mL−1. The internal standard solution was made by dissolving 10 mg of thymol in methanol to a concentration of 1.0 mg mL−1, and then diluted with acetonitrile to create a working solution of 500 ng mL−1. For sample preparation, 50 μL of plasma was mixed with 200 μL of the internal standard working solution in an Eppendorf tube, vortexed for 1 min, and centrifuged at 20
000 rpm at 4 °C for 10 min. A 10 μL aliquot of the supernatant was then injected for analysis.
For the determination of propofol concentration in plasma by HPLC, an eight-point calibration curve was established by taking 45 μL blank plasma and adding 5 μL standard series working solutions to produce plasma propofol concentrations of 48.87, 97.73, 293.19, 879.58, 2638.75, 7916.25, 10555.00, and 21
110.00 ng mL−1. Then, the samples were processed according to the method described above in Section 2.3.2 and analyzed. The standard curve was established using the ratio of the peak area of propofol to the internal standard on the y-axis and the plasma concentration of propofol on the x-axis. Weighted regression (1/x2) was performed to obtain a regression equation of y = 0.000062x − 0.000104 (r = 0.999). The linearity was within the range of 48.9 and 21
110.0 ng mL−1.
To determine plasma propofol concentrations by HPLC, an eight-point calibration curve was created by mixing 45 μL of blank plasma with 5 μL of standard working solutions to achieve final concentrations of 48.87, 97.73, 293.19, 879.58, 2638.75, 7916.25, 10555.00, and 21
110.00 ng mL−1. The samples were prepared as described in Section 2.3.2 and analyzed. The calibration curve was constructed by plotting the ratio of propofol to internal standard peak areas (y-axis) against plasma propofol concentrations (x-axis). Weighted regression (1/x2) yielded the equation y = 0.000062x − 0.000104 with r = 0.999, showing linearity in the range of 48.9 to 21
110.0 ng mL−1.
When calculating PK parameters for propofol, to enhance the comparability between the exhaled concentration and plasma concentration, the exhaled concentration of propofol was converted into mass concentration based on the ideal gas law. Subsequently, PK parameters were computed using DAS 3.0 and the NCA (non-compartmental analysis) model was applied for PK modeling.
Fig. 3B illustrates the carry-over effect, assessed by transitioning from sampling propofol concentrations (4.61, 9.23, and 23.06 ppbv) to pure air (0 ppbv). The signal returned to baseline within 20–40 s (1–2 measurements) after sampling stopped. Carry-over effects were quantified as 4.7%, 9.1%, and 8.7% for concentrations of 23.06, 9.23, and 4.61 ppbv, respectively.
NC | Intra-day (n = 15) | Inter-day (n = 15) | ||
---|---|---|---|---|
MI | RSD (%) | MI | RSD (%) | |
a NC: nominal concentration of propofol standard gas. MI: measured intensity. | ||||
4.61 ppb | 42![]() |
5.83 | 43![]() |
9.69 |
51![]() |
5.31 | 42![]() |
3.62 | |
50![]() |
4.20 | 52![]() |
7.25 | |
23.06 ppb | 331![]() |
6.51 | 312![]() |
5.30 |
325![]() |
7.75 | 315![]() |
7.57 | |
308![]() |
4.98 | 313![]() |
9.75 |
Parameters | Unit | C exhaled | Unit | C plasma | P (t-test) |
---|---|---|---|---|---|
a NCA, non-compartment model; the subscript ‘z’ indicates that the analysis data results are from a non-compartment model. Data of PK parameters were expressed as mean (SD). Cexhaled, propofol concentration in exhaled air; Cplasma, propofol concentration in plasma; AUC, area under the curve; R_AUC (t/∞), corrected area under the curve; T1/2, half-life; CLz, apparent clearance; Cmax, peak concentration; MRT, mean residence time; λz, slope of the terminal phase regression line; —, solid horizontal line represents that the indicators are not comparable due to differences in units of measurement. b P < 0.05 with Tmax from Cplasma. | |||||
AUC (0–t) | ppbv min | 80.80 (21.28) | μg L−1 min−1 | 60766.58 (16457.29) | — |
AUC (0–∞) | ppbv min | 89.12 (21.83) | μg L−1 min−1 | 70240.43 (20904.37) | — |
R_AUC (t/∞) | % | 90.42 (7.36) | % | 87.07 (2.83) | 0.3029 |
AUMC (0–t) | min min ppbv | 606.22 (170.88) | min min μg L−1 | 294932.68 (118918.12) | — |
AUMC (0–∞) | min min ppbv | 909.61 (407.84) | min min μg L−1 | 629359.52 (340053.99) | — |
T 1/2z | min | 8.11 (5.43) | 9.02 (3.98) | 0.3939 | |
T max | min | 2.00 (0.21)b | 1.00 (0.00) | 0.0022b | |
CLz | L min−1 ppbv−1 | 59.33 (16.05) | L kg−1 | 0.08 (0.02) | — |
C max | ppbv | 11.09 (6.80) | μg L−1 | 11307.66 (2775.73) | — |
MRT (0–t) | min | 7.82 (2.45) | 4.73 (1.13) | 0.0931 | |
MRT (0–∞) | min | 10.55 (4.94) | 8.47 (2.92) | 0.5887 | |
λ z | min−1 | 0.15 (0.14) | min−1 | 0.09 (0.05) | 0.5128 |
Although the area under the curve (AUC) values of drug concentration over time for Cexhaled and Cplasma are not directly comparable due to dimensional differences, the R_AUC (t/∞) values exceeded 80% in both cases, indicating that both concentration forms adequately reflect the in vivo processes of absorption, distribution, metabolism, and excretion. Mean Residence Time (MRT) is an important parameter describing the residence time of a drug in the body, reflecting the total time required for the processes of absorption, distribution, and elimination of the drug. MRT (0–t) and MRT (0–∞) derived from Cexhaled (7.82 ± 2.45 min, 10.55 ± 4.94 min) were higher than those from Cplasma (4.73 ± 1.13 min, 8.47 ± 2.92 min), but the differences were not statistically significant (P = 0.0931 and P = 0.5887). Similarly, the elimination rate constants (λz) for Cexhaled (0.15 ± 0.15 min−1) and Cplasma (0.09 ± 0.05 min−1) showed no significant difference (P = 0.5128).
UV-TOF MS is a highly efficient mass spectrometry technique that eliminates the need for sample preprocessing or parameter optimization. Its key advantage lies in a high-efficiency ion analyzer, enabling high-throughput and simultaneous analysis of multiple molecules. Signal accumulation depends on acquisition time, with longer durations enhancing signal strength. In this study, a 20-second acquisition time and a gas sampling flow rate of 70 mL min−1 were used for exhaled air analysis. The method demonstrated excellent linearity, with a calibration curve spanning 3.23 to 46.13 ppbv (R2 = 0.9939), covering the reported clinically relevant propofol range by 0–20 ppbv (ref. 21) and 2–39 ppbv,22 thereby meeting clinical requirements. Accuracy and repeatability were confirmed through intra- and inter-day precision analyses, with the imprecision values remaining below 15% across different propofol concentrations.
Carry-over effects were minimized to within 10%. Unlike previous studies reporting propofol adhesion in detection systems,23 our results indicate rapid clearance within the UV-TOF MS system. This is likely due to two factors: the use of polyether ether ketone (PEEK) sampling tubes and gas pipelines with minimal surface adsorption,24 and the high-vacuum operation (500 ± 20 Pa) system, which efficiently absorbs and expels gaseous compounds, preventing accumulation.
The matrix effect was not investigated, as ultraviolet photoionization, a soft ionization technique, minimizes polarity discrimination, ion suppression, and matrix interference, enabling real-time trace analysis without pretreatment. A direct-current krypton lamp (10.6 eV) ionizes analytes in a high-vacuum environment, producing minimal fragment ions when photon energy exceeds the analyte's ionization threshold (M + hv → M+ + e−). The ionization energies of common air components—N2 (15.58 eV), CO2 (13.77 eV), O2 (12.07 eV), and H2O (12.62 eV)—exceed the UV photon energy, preventing ionization and minimizing interference. This reduces moisture impact in exhaled air, eliminating the need for matrix-matched propofol standards. UV-TOF MS enables direct gas analysis with low sample consumption, high sensitivity, and no pretreatment.
Calibrating gaseous propofol is challenging due to its low solubility and volatility. While previous studies used organic solvents like methanol or ethanol to dissolve propofol, followed by dilution in water,25,26 the low exhaled propofol concentrations (ppbv range) make its 124 mg per L water solubility sufficient. This study prepared an aqueous stock solution below the solubility limit of propofol, ensuring adequate dissolution. Calibration confirmed that aqueous solution provides accurate, reproducible measurements without organic co-solvents, validating feasibility.
After the bolus injection of propofol, a moderately strong correlation (R2 = 0.7950) was observed between Cexhaled and Cplasma. This result is consistent with findings by Lukas et al., who reported a similar correlation (R2 = 0.71) in rats.14 As an exploratory study, this correlation highlights the potential for using exhaled propofol concentrations to predict plasma levels. Overall PK profiles for both Cexhaled and Cplasma follow similar pharmacokinetic trends, adequately reflecting in vivo absorption, distribution, metabolism, and excretion processes, as indicated by (R_AUC > 80%). While AUC values are not directly comparable due to dimensional differences, both measures capture drug exposure over time. MRT values for both Cexhaled and Cplasma suggest a similar duration of drug retention in the body. Although MRT (0–t) and MRT (0–∞) were higher for Cexhaled, these differences were not statistically significant (P = 0.0931 and P = 0.5887). The λz for Cexhaled and Cplasma showed no significant difference (P = 0.5128), indicating similar elimination kinetics.
Nevertheless, a delay in Tmax for Cexhaled (2.00 ± 0.21 min) was observed compared to Cplasma (1.00 ± 0.00 min, P < 0.05), possibly due to the substantial blood/air partition coefficient of propofol and redistribution or transit time through the pulmonary system. Actually, the UV-TOF MS system could immediately detect the presence of propofol when the corresponding standard gas was introduced, indicating that these delays were not caused by adsorption on the instrument pipelines. It takes a certain time to achieve equilibrium between free propofol in the blood and volatile gaseous propofol in the breath.27,28 Additionally, conditions such as alveolar membrane diseases or hemodynamic fluctuations may exacerbate the delayed peak intensity of exhaled propofol.29 While not statistically significant, the higher MRT (0–t) and MRT (0–∞) values for Cexhaled (7.82 ± 2.45 min and 10.55 ± 4.94 min) compared to Cplasma (4.73 ± 1.13 min and 8.47 ± 2.92 min) also suggest a longer persistence of propofol in exhaled breath, possibly due to delayed pulmonary clearance or continuous release from systemic circulation. To address this, we further investigated the correlation between exhaled propofol concentration and its anesthetic effect. The relationship with the BIS value showed a moderate fit to a one-phase exponential decay model (R2 = 0.5501), indicating the potential of real-time exhaled propofol monitoring to guide titration and maintain sedation. This finding underscores its significant clinical relevance for anesthesia management.
This study has limitations. First, inaccuracies in gaseous propofol concentrations may arise from its semi-volatile nature, deviating from ideal volatile compound behavior. Second, BIS monitoring, used as a surrogate for sedation depth, showed only moderate correlation with exhaled propofol, limiting its clinical applicability. Previous studies30 caution against using BIS for anesthetic depth in dogs, as its algorithm was designed for humans. In Beagles, this moderate correlation may result from the lack of a species-specific BIS algorithm, with factors like posture and movement further affecting reliability.
To overcome these limitations, further studies should incorporate advanced EEG-derived measures, such as the patient state index, burst suppression ratio, spectral edge frequency (95%), and density spectral array,31,32 for more precise differentiation of propofol-induced unconsciousness in dogs. Combining these methods with BIS monitoring could enhance sedation assessment and support real-time exhaled propofol monitoring. Given interspecies differences, clinical trials are needed for validation beyond animal studies.
A key challenge in continuous exhaled breath collection is the inability to distinguish between exhalation and inhalation phases and to eliminate anatomical dead space effects. While we have investigated dead space influence,33 inhalation introduces gases (e.g., room air or oxygen) into the sampling system, diluting target compounds like propofol and leading to lower measured concentrations than in exhaled breath.
In this study, canine respiratory parameters were maintained at a tidal volume of 10 mL kg−1, respiratory rate of 20 breaths per minute, and an inhalation-to-exhalation ratio of 1:
2. This ratio indicates that two-thirds of the respiratory cycle is exhalation, suggesting that approximately two-thirds of the collected gas originates from exhaled breath, while one-third is influenced by inhaled air dilution. Since gas composition differs between phases, continuous sampling causes signal fluctuations, with higher target compound concentrations during exhalation and lower or near-zero levels during inhalation. These fluctuations affect data stability and repeatability, complicating analysis.
However, as a direct mass spectrometry technique, UV-TOF MS used in our study integrates all detected components over a sampling period, where longer collection times yield higher signal intensities. The UV-TOF MS system can complete ionization and detection within 0.1–60 s. In this work, we used a 20 s acquisition time with about 70 mL min−1 gas to perform exhaled breath sample detection. This rapid response time may partially mitigate signal interference and no significant signal fluctuations were observed. But our future research will focus on further quantifying the impact of the dilution effect and exploring more advanced signal processing techniques to achieve more precise measurements of drug components in exhaled breath.
To mitigate these effects, scholars have made different attempts.34,35 First is ensuring that the collected gas primarily originates from the exhalation phase. This can be achieved by using a respiratory valve or specialized sensors (e.g., CO2 monitoring sensors) or integrating respiratory monitoring devices (e.g., flow meters or pressure sensors) to continuously track the respiratory cycle, ensuring that gas collection and analysis occur exclusively during exhalation. Second is employing algorithms or signal processing techniques to identify and separate exhalation and inhalation signals, allowing for the quantification and correction of dilution effects and background interference introduced during the inhalation phase. Perl et al. used a CO2-controlled sample acquisition technique to provide more precise sampling and therefore more accurate breath analysis results, but there was still a delay of more than 150 ms in CO2 data acquisition for mainstream sensors and may lead to invalid sample acquisition.13 A study reported that human breath contains isoprene, a volatile organic compound with low inspiratory and plateau expiratory concentrations. C. Hornuss et al.36,37 identified the real-time expiratory propofol signal using breath CO2 and isoprene levels, presenting data as 30-second medians via ion–molecule reaction/electron impact mass spectrometry (IMR-MS). They found that isoprene measurement enables expiratory propofol detection in real-time breath monitoring, potentially simplifying CO2 measurement. However, isoprene was measured every 1.2 seconds, about three times less frequently than CO2, potentially reducing its precision for identifying expiratory propofol. Jiang et al. coupled the humidity correction in real-time with a unidirectional anisole-assisted photoionization IMS, and achieved direct breath-by-breath measurement of intraoperative propofol for the first time, which provided more clinical information for studying the anesthetic pharmacokinetics.21
Footnotes |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d4ay02131b |
‡ Pan Chang and Xiaoxiao Li contributed equally to this study. |