DOI:
10.1039/D5AN00225G
(Communication)
Analyst, 2025,
150, 2226-2230
Sarin simulants show limited representativeness†
Received
26th February 2025
, Accepted 21st April 2025
First published on 22nd April 2025
Abstract
In this study, we conducted colorimetric gas-phase tests on real sarin and compared the results with the most commonly used simulants under identical test conditions. Our findings indicated that reactivity extrapolation was not a reliable approach and that validation using the real toxic gas remained essential for a fair assessment of sarin sensors.
Chemical warfare agents (CWAs) represent a threat to both armed forces and civilians. Consequently, the detection of CWAs is a major area of research.1–13 Nerve agents, a particularly lethal class of chemical warfare agents, inhibit the acetylcholinesterase enzyme (AChE), leading to death by asphyxia or stroke.14,15 Sarin, also known as GB, is a G-type warfare agent. It has been used several times in attacks such as the Tokyo subway incident16 and in Syria.17 This persistent threat necessitates detection methods that are easy to deploy, rapid and unambiguous, even in complex environments. The development of fast and reliable detection methods is especially urgent, as current techniques suffer from significant limitations, such as lengthy analysis times, the need for sample preparation, limited sensitivity, and high rates of false positives.18–21 Significant resources have been allocated to the development of new sensors, resulting from advances across a range of technologies. To name but a few, notable progress has been made in optical detectors,1,2,7,9,22–38 gas chromatography-mass spectrometry,13,39,40 functionalized transistors,41–43 and biosensors.44,45 An overwhelming majority of the reported studies have been carried out using simulants to evaluate the potential of these technologies.46–49 For instance, while studying optical detection in our lab, among the publications related to the optical detection of nerve agents, only 9 out of 99 used an actual nerve agent in the trials, 68 used DCP, 16 used DFP, and 2 used DMMP (see Table S1†). Moreover, remarkably, only 3 used actual nerve agents in the gas phase.
Indeed, real warfare toxic agents, such as nerve agents, cannot be handled in most labs, as they require specific and rare authorisations, a dedicated environment to address health risks, well-defined decontamination protocols, and highly secure, regulated storage of chemicals. Moreover, the possession and use of toxic substances must be declared to the Organisation for the Prohibition of Chemical Weapons (OPCW) and, of course, be strictly for research purposes.50 The specific gas-phase applications require even more stringent authorisations and security protocols due to the higher risks of exposure. Consequently, the vast majority of the literature on nerve agent detection relies on the simulants shown in Fig. 1. DCP (diethylchlorophosphate) is commonly used for the gas-phase detection. DMMP (dimethyl methylphosphonate) is primarily used to simulate physical properties of sarin, having a similar chemical composition and vapour pressure but much lower chemical reactivity.51–53 DFP (Diisopropylfluorophosphate) contains a reactive P–F bond, similar to sarin. Although the inherent limitations of simulants have been discussed by other authors,54 we present, to the best of our knowledge, the first large-scale vapour-phase quantitative comparison of colorimetric sensor reactivity toward sarin and its most common simulants.
 |
| Fig. 1 Chemical structures of sarin, DMMP, DFP, and DCP. | |
A set of 26 different coloured organic molecules were used as probes (see Table S2†). Twenty-five molecules were purchased and another one, 4,4′-((1E,1′E)-quinoline-2,4-diylbis(ethene-2,1-diyl))bis(N,N-dimethylaniline), was obtained through a 3-step synthesis developed in our laboratory.55 The bunch of commercially available molecules was composed of nine azo compounds, twelve triarylmethane dyes, one anthraquinone derivative, two rhodamine derivatives and one cyanine derivative, all bearing reactive groups. The probes have been selected without any preconceived ideas, with the sole focus being on the colorimetric characteristics and the presence of chemically reactive functions.
The protocol to prepare different samples is as follows: first, each chemical probe was solubilized in DMSO in order to obtain a 10 mM solution. Then, a volume of 3.5 μL of each solution was dropped onto a fiberglass paper. The paper was dried at room temperature for 7 h. The supported probes were then placed inside a homemade sample holder containing 26 holes to allow the gas to pass freely through the paper. Four identical sets containing the 26 papers were prepared. Each of them was exposed independently to vapours of the 3 simulants (DMMP, DFP and DCP) and sarin. Each test set was exposed using the same procedure. The chemicals to be detected were deposited in the form of liquids in excess onto a fiberglass paper at the bottom of a Petri dish. The fiberglass paper was used to enhance the gas diffusion within the test chamber to reach saturated vapour pressure. The sample holder containing the 26 samples was then placed above the fiberglass paper to avoid any contact between the toxicants and the supported probes. The latter were then exposed to the various chemical vapours during 1 h at room temperature under ambient atmosphere in a closed Petri dish to avoid any leakage, particularly of the toxic compounds (Fig. 2). The estimated saturated vapor pressures obtained under the experimental conditions were 2533 ppmv (14.75 mg L−1) for sarin, 128 ppmv (0.92 mg L−1) for DCP, 751 ppmv (5.75 mg L−1) for DFP and 775 ppmv (3.99 mg L−1) for DMMP. Due to high restrictions for real sarin-based experiments, the possible parameters to be changed were unfortunately very limited (just one exposure time and no direct measurement of the toxic concentration).
 |
| Fig. 2 (a) Image of the test group with 26 colorimetric sensors arranged in the sample holder, placed inside the Petri dish. (b) Schematic of the sample holder within the Petri dish. | |
Pictures were acquired under visible light and scans of the supported probes were recorded before and after exposure using a desktop scanner. During the scan, a black and white sight was used in order to adjust the colours of the scan to be uniform between all the experiments. The colorimetric results are displayed in Table 1.
Table 1 Colorimetric results of the test group exposed to vapours of sarin, DCP, DFP, and DMMP. Green boxes highlight responses with ΔE1994 > 10%
The colorimetric results were subjected to mathematical calculations in order to eliminate the potential for reliance on the subjective interpretation of colour by the human eye. RGB coordinates were measured at the centre of each spot. They were then transformed into XYZ coordinates and then into Lab coordinates. The coordinates before and after exposure were then mathematically compared in order to obtain a percentage from the ΔE1994 value defined by the International Commission on Illumination (details are presented in ESI†). The measurements performed before each exposure were used as reference for comparison. The ΔE1994 value corresponds to the difference in colour before and after exposure.
A limit value of 10% was chosen to differentiate a positive and a negative response, thus if ΔE1994 > 10%, the change was considered significant and the response was positive to the presence of a toxicant. If ΔE1994 was less than 10%, the change was considered insufficiently significant and the response was deemed negative. The main limitation of the method using ΔE1994 would be the results close to the limit value of 10%, but out of the 104 calculated values of ΔE1994, only twelve displayed a result between 8% and 12%. Table 2 shows the results before and after exposure to sarin, DCP, DFP and DMMP (see Table S3† for all ΔE1994 values).
Table 2 Positive results from exposure to sarin and its simulants
|
Sarin |
DCP |
DFP |
DMMP |
Number of positive responses |
5/26 |
13/26 |
0/26 |
1/26 |
In percentage |
19% |
50% |
0% |
4% |
It was observed that 19% of the molecules change colours after exposure to sarin, 50% after exposure to DCP, only 4% reacted to DMMP and none reacted to DFP. These data point out DCP as the best simulant of sarin for the tested molecules. However, only 31% of the molecules reacting to DCP showed positive sensing to sarin. This is indicative of a high difference in reactivity. This demonstrates that the observation of a positive detection with DCP must not be used to extrapolate a positive result for real sarin. Molecule 20 is the only one that reacted to DMMP, but it did not react with the other simulants or sarin. Molecule 25 showed no reactivity towards any simulant but reacted to sarin, proving again the absolute difference of reactivity between sarin and its simulants.
The reactivity with sarin and simulants of each family of dye is presented in the form of Venn diagrams in Fig. 3. Azo compounds display a good reactivity towards DCP with 67% of reacting probes, but no reactivity with sarin or other simulants was observed. Moreover, 50% of the triarylmethane dyes reacted with DCP and 25% reacted positively with sarin. The triarylmethane dyes reacting with sarin also showed positive results with DCP. The anthraquinone derivative did not show any reactivity. One of the rhodamine derivatives reacted with DMMP, while the other induced a colour change with sarin. The cyanine derivative displayed no reactivity. The synthesized probe reacted with both DCP and sarin.
 |
| Fig. 3 Venn diagrams showing positive results of (a) the azo compounds only, (b) triarylmethane dyes only, (c) all other molecules, and (d) the entire set of colorimetric sensors. The ratios inside the bubbles represent positive sensing for sarin or simulants, while the values in the overlapping areas between the bubbles indicate compounds that detect both organophosphorus molecules. A detailed table of all the results is presented as Table S4.† | |
Conclusions
The main output of this study, based on twenty-six coloured molecules and four different organophosphorus reactants in the vapour phase including real sarin, is that extrapolating results for a real toxicant based on the response of simulants are not relevant and must be avoided. Nevertheless, the use of simulants remains the only expedient and straightforward method for obtaining a proof of concept, given the challenging and complex accessibility to authentic nerve agents, in particular in the gas phase.
In conclusion, it is important to keep in mind that the results obtained with simulants present very significant limitations. It is therefore imperative that trials be conducted with a genuine toxic substance in order to validate a gas-phase nerve agent sensing technology.
Author contributions
Valentin A. Bureau: Conceptualization, data curation, formal analysis, investigation, methodology, writing – original draft. Sébastien Penlou: Formal analysis, methodology, supervision, validation. Sonia Sousa Nobre: Formal analysis, methodology, funding acquisition, supervision, validation. Alexandre Carella: Formal analysis, methodology, supervision, validation. Jean-Pierre Simonato: Formal analysis, supervision, validation, writing – review and editing.
Data availability
The data supporting this article have been included as part of the ESI.†
Conflicts of interest
There are no conflicts to declare.
Acknowledgements
The authors thank DGA CBRN Defence for its contribution. This work was supported by the Agence de l'Innovation de Defense (AID, France) through a PhD grant awarded to Valentin Bureau.
References
- C. E. Davidson, M. M. Dixon, B. R. Williams, G. K. Kilper, S. H. Lim, R. A. Martino, P. Rhodes, M. S. Hulet, R. W. Miles, A. C. Samuels, P. A. Emanuel and A. E. Miklos, ACS Sens., 2020, 5, 1102–1109 CrossRef CAS PubMed.
- S. H. Lim, L. Feng, J. W. Kemling, C. J. Musto and K. S. Suslick, Nat. Chem., 2009, 1, 562–567 CrossRef CAS PubMed.
- M. Grabka, K. Jasek and Z. Witkiewicz, Sensors, 2024, 24, 2477 CrossRef CAS PubMed.
- D. Sarkar, A. Das and S. Naha, Monatsh. Chem., 2023, 154, 673–682 CrossRef CAS.
- Z. Witkiewicz, K. Jasek and M. Grabka, Sensors, 2023, 23, 3272 CrossRef CAS PubMed.
- L. Yang, M. Qin, G. Zhang, J. Yang, J. Yang and J. Zhao, Rev. Anal. Chem., 2023, 42 DOI:10.1515/revac-2022-0052.
- B. Zhu, R. Sheng, T. Chen, J. Rodrigues, Q.-H. Song, X. Hu and L. Zeng, Coord. Chem. Rev., 2022, 463, 214527 CrossRef CAS.
- S. Balasubramanian, A. J. Kulandaisamy, K. J. Babu, A. Das and J. B. B. Rayappan, Ind. Eng. Chem. Res., 2021, 60, 4218–4239 CrossRef CAS.
- G. H. Dennison and M. R. Johnston, Chem. – Eur. J., 2015, 21, 6328–6338 CrossRef CAS PubMed.
- D. Pragney and U. V. R. V. Saradhi, Trends Environ. Anal. Chem., 2012, 37, 73–82 CrossRef CAS.
- X. Zhang, J. Zhang, C. Li, X. Zhang, J. Yun and D. Cao, Adv. Compos. Hybrid Mater., 2024, 7, 108 CrossRef CAS.
- S. Giannoukos, B. Brkić, S. Taylor, A. Marshall and G. F. Verbeck, Chem. Rev., 2016, 116, 8146–8172 CrossRef CAS PubMed.
- D. R. Ifa, A. U. Jackson, G. Paglia and R. G. Cooks, Anal. Bioanal. Chem., 2009, 394, 1995–2008 CrossRef CAS PubMed.
- R. T. Delfino, T. S. Ribeiro and J. D. Figueroa-Villar, J. Braz. Chem. Soc., 2009, 20, 407–428 CrossRef CAS.
- M. Pohanka, Biomed. Pap., 2011, 155, 219–223 CrossRef CAS PubMed.
- The Sarin Gas Attack in Japan and the Related Forensic Investigation, https://www.opcw.org/media-centre/news/2001/06/sarin-gas-attack-japan-and-related-forensic-investigation, (accessed March 1, 2023).
- First Report of the IIT, https://www.opcw.org/iit/first-report-iit, (accessed August 7, 2024).
- Y. Seto, M. Kanamori-Kataoka, K. Tsuge, I. Ohsawa, H. Maruko, H. Sekiguchi, Y. Sano, S. Yamashiro, K. Matsushita, H. Sekiguchi, T. Itoi and K. Iura, Toxin Rev., 2007, 26, 299–312 CrossRef CAS.
- V. Kumar, H. Rana, G. Raviraju, P. Garg, A. Baghel and A. K. Gupta, RSC Adv., 2016, 6, 59648–59656 RSC.
- C. A. Valdez, R. N. Leif, S. Hok and B. R. Hart, Rev. Anal. Chem., 2018, 37 DOI:10.1515/revac-2017-0007.
- B. C. Hauck, B. S. Ince and P. C. Riley, ACS Sens., 2023, 8, 2945–2951 CrossRef CAS PubMed.
- W.-Q. Meng, A. C. Sedgwick, N. Kwon, M. Sun, K. Xiao, X.-P. He, E. V. Anslyn, T. D. James and J. Yoon, Chem. Soc. Rev., 2023, 52, 601–662 RSC.
- L. Ordronneau, A. Carella, M. Pohanka and J.-P. Simonato, Chem. Commun., 2013, 49, 8946–8948 RSC.
- M. Pohanka, J. Z. Karasova, K. Kuca, J. Pikula, O. Holas, J. Korabecny and J. Cabal, Talanta, 2010, 81, 621–624 CrossRef CAS PubMed.
- L. Termeau, S. Penlou and A. Carella, ACS Sens., 2023, 8, 1510–1517 CrossRef CAS PubMed.
- O. R. Keisar, A. Pevzner, G. Fridkin, O. Shelef, D. Shabat and N. Ashkenazi, Anal. Methods, 2024, 16, 1736–1740 RSC.
- R. Puglisi, R. Santonocito, A. Pappalardo and S. G. Trusso, ChemPlusChem, 2024, 89, e202400098 CrossRef CAS PubMed.
- M. Safarkhani, H. Kim, S. Han, F. Taghavimandi, Y. Park, R. Umapathi, Y.-S. Jeong, K. Shin and Y. S. Huh, Coord. Chem. Rev., 2024, 509, 215804 CrossRef CAS.
- E. D. Fussell, N. D. Kline, E. Bennin, S. S. Hirschbeck and A. Darko, ACS Sens., 2024, 9, 2325–2333 CrossRef CAS PubMed.
- K. Gupta and A. K. Patra, ACS Sens., 2020, 5, 1268–1272 CrossRef CAS PubMed.
- R. AbhijnaKrishna, S.-P. Wu and S. Velmathi, Microchem. J., 2024, 205, 111369 CrossRef CAS.
- P. Sarkar, N. Tohora, M. Mahato, S. Ahamed, T. Sultana and S. K. Das, J. Mol. Struct., 2024, 1306, 137846 CrossRef CAS.
- S. Ahamed, N. Tohora, R. Sahoo, M. Mahato, T. Sultana, A. Maiti, S. Ghanta and S. Kumar Das, J. Mol. Liq., 2024, 409, 125411 CrossRef CAS.
- S. Royo, R. Martínez-Máñez, F. Sancenón, A. M. Costero, M. Parra and S. Gil, Chem. Commun., 2007, 4839–4847 RSC.
- M. Mahato, T. Sultana, A. Maiti, S. Ahamed, N. Tohora, S. Ghanta and S. K. Das, Anal. Methods, 2024, 16, 1371–1382 RSC.
- A. Maiti, T. Sultana, B. Rajbanshi, B. Bhaumik, N. Roy and M. N. Roy, Microchem. J., 2024, 199, 109977 CrossRef CAS.
- N. Tohora, S. Ahamed, M. Mahato, T. Sultana, J. Chourasia, A. Maiti and S. K. Das, Photochem. Photobiol. Sci., 2024, 23, 763–780 CrossRef CAS PubMed.
- J. Chourasia, N. Tohora, T. Sultana, M. Mahato, A. Maiti, S. Ahamed and S. K. Das, Luminescence, 2024, 39, e4731 CrossRef CAS PubMed.
- C. A. Valdez and R. N. Leif, Molecules, 2021, 26, 4631 CrossRef CAS PubMed.
- G. Mishra, P. Gupta, A. Kumar, A. Purohit, V. Tak and D. Pardasani, J. Chromatogr., A, 2024, 1716, 464645 CrossRef CAS PubMed.
- S. Clavaguera, A. Carella, L. Caillier, C. Celle, J. Pécaut, S. Lenfant, D. Vuillaume and J.-P. Simonato, Angew. Chem., Int. Ed., 2010, 49, 4063–4066 CrossRef CAS PubMed.
- V. Passi, F. Ravaux, E. Dubois, S. Clavaguera, A. Carella, C. Celle, J.-P. Simonato, L. Silvestri, S. Reggiani, D. Vuillaume and J.-P. Raskin, IEEE Electron Device Lett., 2011, 32, 976–978 CAS.
- X. Huo, X. Shan, Q. Pan, Z. Cao, Z. Cong, J. Song, J. Jiang, Q. Mia, L. Jia and J. Gao, Sens. Actuators, B, 2024, 403, 135184 CrossRef CAS.
- M. Pohanka, P. Dobes, L. Drtinova and K. Kuča, Electroanalysis, 2009, 21, 1177–1182 CrossRef CAS.
- D. Zappi, M. M. Ramma, V. Scognamiglio, A. Antonacci, G. Varani and M. T. Giardi, Biosensors, 2020, 10, 176 CrossRef CAS PubMed.
- G. Paimard, M. Nejatian, Z. Sarlak, R. Mohammadi and M. Rouhi, Microchem. J., 2024, 198, 110173 CrossRef CAS.
- J. Disley, G. Gil-Ramírez and J. Gonzalez-Rodriguez, TrAC, Trends Anal. Chem., 2023, 168, 117282 CrossRef CAS.
- A. Emelianova, A. Reed, E. A. Basharova, A. L. Kolesnikov and G. Y. Gor, ACS Appl. Mater. Interfaces, 2023, 15, 18559–18567 CrossRef CAS PubMed.
- M. Agrawal, D. F. Sava Gallis, J. A. Greathouse and D. S. Sholl, J. Phys. Chem. C, 2018, 122, 26061–26069 CrossRef CAS.
-
Chemical Weapons Convention, https://www.opcw.org/chemical-weapons-convention/download-convention, (accessed August 7, 2024).
- S. L. Bartelt-Hunt, D. R. U. Knappe and M. A. Barlaz, Crit. Rev. Environ. Sci. Technol., 2008, 38, 112–136 CrossRef CAS.
- R. Puglisi, A. Pappalardo, A. Gulino and G. T. Sfrazzetto, Chem. Commun., 2018, 54, 11156–11159 RSC.
- R. Puglisi, A. Pappalardo, A. Gulino and S. G. Trusso, ACS Omega, 2019, 4, 7550–7555 CrossRef CAS.
- B. Díaz De Greñu, D. Moreno, T. Torroba, A. Berg, J. Gunnars, T. Nilsson, R. Nyman, M. Persson, J. Pettersson, I. Eklind and P. Wästerby, J. Am. Chem. Soc., 2014, 136, 4125–4128 CrossRef PubMed.
-
L. Termeau, S. Penlou and A. Carella, Use of a specific compound for the detection of toxic chemical compounds, WO2023111471A1, 2022 Search PubMed.
|