Pavan Panditab,
Lingxue Kongb and
G. L. Samuel*a
aManufacturing Engineering Section, Department of Mechanical Engineering, IIT Madras, Chennai, Tamil Nadu 600036, India. E-mail: samuelgl@iitm.ac.in
bInstitute for Frontier Materials, Deakin University, Geelong, Victoria 3216, Australia
First published on 3rd January 2024
The advancement of microfluidics and the manufacturing of microdevices has led to a strategic change in the biomedical industry. The flow through narrow channels and the pillars are placed strategically, leading to the phenomenon of particle separation through deterministic lateral displacement (DLD). In such a phenomenon, the shape, size, location and orientation of the obstacles play an important role. For the first time, particle separation is achieved with DLD modules having high row shift angles of 25°, 30° and 35°, reducing the number of pillars. The significance of circular and triangular micropillars executing deterministic lateral displacement, oriented at different angles, has been investigated, and it is found that the triangular pillars oriented at 75° resulted in better separation compared to the other configurations. In this report, the fabrication, location, orientation of the micropillars and the selection of appropriate process parameters are detailed. The structures are fabricated on silicon wafers using the standard photolithography process followed by the deep reactive ion etching process. These dies are further used to fabricate the polydimethylsiloxane-based microfluidic chips. These fabricated devices are characterised by their size, structure and quality using 3D microscopy and scanning electron microscopy. Further, blood plasma separation is carried out using the devices fabricated in this work, and the particles at the inlet and outlets are evaluated using microscopy and a novel image processing technique, replacing the use of a hemocytometer. The path traced by the particles at different flow conditions is numerically evaluated and validated with experiments. The novel device is capable of separating blood cells from plasma with a recovery factor varying from 44% to 100%. PDMS–PDMS bonding experiments using oxygen and argon plasma have been carried out to evaluate the maximum bond strength and flow velocity in the devices. It is observed that the oxygen plasma results in a bond strength of 0.404 N mm−1, thus a high throughput of 135.34 μL s−1 is achieved using the fabricated device.
Continuous rapid separation of cells is crucial in the case of blood purification. The process has to be carried out so that it does not affect the functionality of the blood cells. If the process or system induces pressure gradients, biotoxicity, ionic imbalance or change in the flow behaviour, the blood gets coagulated or is rendered unusable. Hence, special care has to be taken while designing these systems. The material used in the fabrication of these devices is essentially biocompatible. The fabrication techniques should not induce biotoxicity such as corrosion, zincification or other chemical reactions. The design of the systems is such that the pressure gradients induced are minimal such that the function of the system is not compromised.
Considering these features, various systems have been established. The deterministic lateral displacement technique is one such type of continuous separation technique. DLD has been used in the applications such as emulsifying droplets in water-in-oil or oil-in-water systems,4 encapsulation of micro-organisms,5 separation of biological particles like DNA,6,7 exosomes,8,9 blood cells,10–14 mammalian cells and circulating tumour cells.15,16 It was first demonstrated by Huang et al.7 using micropillars with a circular cross-section, primarily used for rapid DNA separation. The DLD was further improved by Inglis17 and Davis14 and considering the wall effect and the parabolic velocity profile in the inter-pillar spaces. Furthermore, by changing the profile, shape and dimensions of the micropillars and the microchannels, the critical diameter of the particles being separated varies. The pillars with circular, triangular, rhomboidal and octagonal cross sections have been analysed previously. The pillars with unconventional cross-sections like streamlined, airfoil,18 and I-section19,20 were analysed to reduce the critical diameter of the separating particles. Loutherback et al.,21,22 showed that the triangular posts aided in reducing the critical diameter of the separating particles by considering various factors such as the gap-to-post size, the vertex rounding, and the width-to-height ratio of the triangle. It is seen that the pillars with a cross-section of an equilateral shape performed better, considering the critical size of the particles. Improving the Davis model, Wei et al.23 developed a numerical model to study the lateral displacement of spherical particles using the lattice Boltzmann method and immersed boundary approach. This model predicts the particle path and the critical particle diameter in DLD systems with shaped pillars. Dincau et al.18 proposed the effect of Reynolds number on the particle path. In the study, it was observed that the streamlines of the flow at different Re differed in their positions. Further, Pariset et al.24 studied the effect of the number of pillars in the channel width (wc) and anticipated the minimum number required to avoid the intermediary mode. The study also demonstrates the existence of multiple critical diameters and the techniques to obtain a single critical diameter in the entire regime.
Solid polymeric microspheres are used abundantly in the flow characterisation of the devices, but the blood components are soft and deformable.25 It is seen that the behaviour of the particles also depends on the shape of the particles being separated.20 Patel and Stark26 conducted on the spherical and biconcave particles in various clustering configurations, such as monodispersed and bi-dispersed pairs. Thus, in this study, whole blood is considered for the analysis. The blood contains erythrocytes, leukocytes27 and thrombocytes, mostly biconcave in shape.
These DLD structures and the microchannels are fabricated by various processes such as lithography and lift-off processes, electroplating, thermal oxidation, wire bonding, chemical etching, deep reactive ion etching, and laser micromachining. The most prominent technique used in the fabrication through photolithography is SU-8 lithography. However, cracking and residual stresses28 were observed in the SU-8 lithography affecting the quality of the DLD module. The other challenges associated with the above-mentioned process are that the cross-section of the slender pillars is non-uniform owing to the non-isotropic etching, the heat-affected zone, and the fusion of the adjacent micro holes. In our previous study, the P20 die steel moulds were prepared using femtosecond lasers to fabricate PDMS-based microchannels with DLD structures.29 Thus, the current structures are fabricated considering the limitations of laser micromachining in fabricating the DLD structures.
Bonding of the microfluidic chips is crucial in determining the throughput of the device. It defines the maximum flow rate of the fluid. Various bonding techniques are adopted to bind the PDMS–PDMS chips. Some prominent techniques are partial curing, adhesive bonding, solvent bonding and plasma bonding techniques. The binding of the PDMS chips occurs due to the creation of hydroxyl (–OH) groups on the surface when PDMS is exposed to oxygen plasma. The –OH groups readily react to form covalent bonds with the other plasma-exposed PDMS chip.30 The bond strength is quantified using blister, leakproof, and adhesion tests. It was observed that the bond strength of O2 and argon (Ar) plasma treat PDMS samples showed similar bond strengths when PDMS was bonded with glass coverslips.31
In this study, micropillars with circular and equilateral triangular cross-sections are considered to evaluate the effect of the process parameters, vertex rounding, placement and the orientation of the pillars in the deterministic lateral displacement zone. Numerical simulations are carried out, and validated with experiments through blood flow evaluation. The fabrication of the masters to develop the PDMS chips is explained in detail showing the variations in the pillar depth with the etching cycles. The structures are characterised by evaluating their quality and shapes. Finally, the adhesive bond strength evaluation of the PDMS–PDMS samples was carried out to determine the blood flow's maximum pressure and mass flow rate in the microchannels.
Sl no. | Configuration code | Post shape | Inclination angle | Tilt angle | Row shift fraction |
---|---|---|---|---|---|
1 | C – 25 | Circular | 25 | — | 0.4663 |
2 | C – 30 | Circular | 30 | — | 0.5774 |
3 | C – 35 | Circular | 35 | — | 0.7002 |
4 | T – 25 | Triangular | 25 | 0 | 0.4663 |
5 | T – 30 | Triangular | 30 | 0 | 0.5774 |
6 | T – 35 | Triangular | 35 | 0 | 0.7002 |
7 | T – 30–15 | Triangular | 30 | 15 | 0.4663 |
8 | T – 30–45 | Triangular | 30 | 45 | 0.4663 |
9 | T – 30–75 | Triangular | 30 | 75 | 0.4663 |
10 | T – 30–90 | Triangular | 30 | 90 | 0.4663 |
A layer of positive photoresist, namely AZ40xt, was spin coated on the HMDS coated silicon wafers. The coated wafer is subjected to pre-exposure bake at 80° on the hot plate. The baked wafers are then exposed to UV light with the mask aligned on it. This is carried out using the UV flood light. The UV exposure required to cure the AZ40xt is 12 W cm−2, i.e., the wafer is exposed to UV light at 25 W for 75 s.33 The exposed wafers are then subjected to post-exposure bake of 80 °C for 10 min. The unexposed portion of the coated wafer is removed by developing the wafer using the AZ726 MIF developed.
Parameter | Value |
---|---|
Chamber pressure (mTorr) | 30 |
Passivation pressure (mTorr) | 19 |
Valve position | 70% |
SF6 flow rate (ccpm) | 140 |
C4F8 flow rate (ccpm) | 7 |
ICP generator forward power (W) | 2200 |
RF generator forward power (W) | 23 |
Step time (s) | 6 |
Parameter | Value |
---|---|
Chamber pressure (mTorr) | 30 |
Passivation pressure (mTorr) | 19 |
Valve position | 70% |
SF6 flow rate (ccpm) | 4 |
C4F8 flow rate (ccpm) | 140 |
ICP generator forward power (W) | 1700 |
RF generator forward power (W) | 4 |
Step time (s) | 5 |
According to the hypothesis, every cycle etches 385 nm of the silicon wafer. To determine the etch rate, three different combinations were carried out at 256, 258 and 259 cycles and it was seen that 258 cycles yielded a channel depth of 100 μm, which was the intended dimension, as discussed in Section 3.1.2.
After the DRIE of the silicon wafer, the remaining resist was washed away with acetone. The cleaned wafer was further deposited with the passivation layer such that the Teflon-like nano-coating acts as a release agent while fabricating the PDMS devices. The fabricated silicon masters are shown in Fig. 3 and detailed figures are shown in Fig S2.†
(1) |
(2) |
(3) |
Due to the low Re, in other terms due to the prominence of viscous damping, the kinetic energy term in eqn (1) becomes negligible, thus can be written as eqn (4).
(4) |
Considering the pressure gradient and the volumetric fluid flux between the pillars. The fluid resistance is deduced as shown in eqn (5) considering the circular pillars and the parabolic velocity profile between the pillars
(5) |
Further, considering the allowable shear stress (0.3 Pa in large arteries and 10 Pa in capillaries) in the blood vessels, which prevents the activation of the platelets. This is observed to be maximum at the walls
Therefore, the critical hydrodynamic diameter is twice the width of the first streamline, thus
Dc = 2ζGε | (6) |
The row shift fraction signifies the fractional distance by which the adjacent row is shifted, and this is vital in determining the separation efficiency of the device. Analytically, the least critical diameter is found to be 19.31 μm, computed using eqn (6).
Configuration | Exposure duration (mins) | Bake temperature (°C) | Post bonding bake duration (mins) | ||
---|---|---|---|---|---|
Plasma exposure | Temperature (°C) | Pressure (mTorr) | |||
O2 | 80 | 436 | 2 | 80 | 10 |
O2 | 80 | 436 | 5 | 80 | 10 |
Ar | 80 | 455 | 2 | 80 | 10 |
Ar | 80 | 455 | 5 | 80 | 10 |
(7) |
(8) |
Mass flow rate ṁ = A × v, [m3 s−1] | (9) |
Parameter | Value |
---|---|
Type of flow | Laminar flow |
Governing equation | Navier–Stokes' equation |
Inlet velocity (m s−1) | 0.5 |
Domain | 1, 2, 3 |
Domain density (kg m−3) | 1000 |
Domain dynamic viscosity (Pa) | 0.001 |
Particle density (kg m−3) | 1055 |
Particle diameter (μm) | 10, 15 |
Drag force | Governed by Stokes law |
Wall condition | No slip, bounce |
The flow experiments are conducted using the experimental setup as shown in Fig. 10a. The setup consists of a syringe pump (NE-300), a fluorescent microscope (Nikon Eclipse 80i), a Nikon DS-Ri1 camera and a computer system to analyse the images. The syringe pump is used to pump the fluid at the required mass flow rates, and the fluorescent microscope and the camera are used to capture the images and videos of the cells flowing between the pillars.
Fig. 6 3D profilometer images of the (a and b) masters and (c–e) devices, showing the channel depth, microholes and fabricated micropillars. |
Sl no. | No. of cycles | Depth of etching (μm) | |
---|---|---|---|
Channel | Hole | ||
1 | 255 | 98.64 | 69.58 |
2 | 258 | 100.02 | 75.89 |
3 | 259 | 100.64 | 76.20 |
Fig. 6a and b show the depth of the pillars and the microchannels. From the figures, it is evident that the circular and triangular pillars are 75.9 μm and 69.6 μm whereas the microchannel is 100 μm in depth. Fig. 6c shows the dimensions of the PDMS micropillars in the device. It is seen that the triangular pillars are 79.9 μm high, this variation can be attributed to the formation of defects while peeling or due to the presence of undulations on the pillar surface.
Fig. 6d and e show the variation in the depths of the micro holes of circular and triangular geometry after 255 DRIE cycles. A variation of approximately 7 μm was observed between the circular and triangular holes.
Fig. 7 The results of the T-peel test where specimen 1 and 2 corresponds to the oxygen plasma, whereas 3 and 4 to the argon plasma. |
Type of plasma | Maximum load (N) | |
---|---|---|
O2 plasma | 0.396 | 0.404 |
Ar plasma | 0.322 | 0.366 |
Configuration | Bond strength (N mm−1) | Maximum flow rate (μL s−1) | Maximum velocity (m s−1) |
---|---|---|---|
O2 plasma | 0.396 | 132.66 | 1.06 |
O2 plasma | 0.404 | 135.34 | 1.08 |
Ar plasma | 0.366 | 122.61 | 0.98 |
Ar plasma | 0.322 | 107.87 | 0.86 |
Configuration | Cell count in outlet 1 | Cell count in outlet 2 | Cell count in inlet | Cells stuck in domain 1 | Ratio of out_1 to out_2 |
---|---|---|---|---|---|
C – 25 | 9 | 10 | 27 | 7 | 0.90 |
C – 30 | 6 | 14 | 27 | 0 | 0.43 |
C – 35 | 0 | 22 | 27 | 5 | 0 |
T – 25 | 3 | 9 | 27 | 13 | 0.33 |
T – 30 | 6 | 16 | 27 | 4 | 0.38 |
T – 35 | 3 | 12 | 27 | 11 | 0.25 |
T – 30–15 | 4 | 14 | 27 | 9 | 0.29 |
T – 30–45 | 1 | 0 | 27 | 24 | — |
T – 30–75 | 2 | 11 | 27 | 14 | 0.18 |
T – 30–90 | 8 | 2 | 27 | 17 | 4.00 |
It is seen that the circular pillars oriented at 35° provides the best particle separation, but as the circular pillars result in a zero velocity zone on their surface, the possibility of the particles getting stuck in the domain is higher. The recovery factor in the devices with circular pillars is lower in comparison to the devices with triangular pillars. When the particle path in the devices with triangular pillars is observed, the orientation plays an important role. The ratio of the cells in outlet 1 to the cells in outlet 2 shows an increasing trend while the orientation angle is increased from 25° to 45°, but at 75°, the ratio is the least, thus suggesting that the triangular pillars oriented at 75° show the best separation. Meanwhile, the recovery rates in case of the devices with triangular pillars is lower in comparison to the devices with circular pillars.
From Fig. 8 and 9, it is evident that the vertices of the triangular pillars also contribute to the separation efficiency of the devices. Fig. 8 shows the count of the cells observed in the 3 chosen domains considering the sharp edged micropillars whereas the Fig. 9a represents the count of the cells in the domains considering the rounded edged micropillars as in the case of the experiments. The particle size chosen for previously discussed simulations is 10 μm whereas the simulations results shown in Fig. 9b consider a particle diameter of 15 μm. Fig. 9a and b compares behaviour of the path lines when diameter is varied from 10 μm to 15 μm. It is observed that the 15 μm particles are directed towards the outlet 2 whereas the 10 μm particles have the possibility of flowing towards the outlet 1.
Fig. 8 The particle distribution in outlet_1, outlet_2 and the domain_1 at the end of the simulation time scale of 0.1 s as observed with triangular micropillars having pointed vertex. |
Fig. 10 Experimental setup for fluid flow and analysis. (a) and (b) Flow setup used to evaluate the cells collected at the outlets. (c) Flow setup to visualize the flow paths and the cell movement. |
From the analytical measurements, the least critical particle diameter was observed to be 19.31 μm, however, lateral displacement of the blood cells (RBC – 6–8 μm and WBC – 14–16 μm) is observed in the experimental results.
The percentage recovery in the experiments was evaluated using the eqn (10) and is tabulated in Table 10. It is found that the recovery factor varies from 44% to 81% in pointed vertex micropillars, whereas it varies from 74% to 100% in the rounded vertex micropillars.
(10) |
Circular micropillars | |||
---|---|---|---|
Configuration | % recovery (%) | Configuration | % recovery (%) |
C – 25 | 70 | C – 30 | 74 |
C – 35 | 81 |
Triangular micropillars | ||||
---|---|---|---|---|
Pointed vertices | Rounded vertices | |||
T – 25 | 44 | T – 25 | 89 | 100 |
T – 30 | 81 | T – 30 | 93 | 100 |
T – 35 | 56 | T – 35 | 96 | 74 |
T – 30–15 | 67 | T – 30–15 | 89 | 93 |
T – 30–45 | 4 | T – 30–45 | 89 | 85 |
T – 30–75 | 48 | T – 30–75 | 89 | 85 |
T – 30–90 | 37 | T – 30–90 | 93 | 74 |
From the numerical simulations and the experimental studies carried out it is seen that the shape and orientation of the micropillars play an important role in evaluating the functionality of the deterministic lateral displacement-based devices. Moreover, fluid constituents can be easily manoeuvred with the triangular pillars as the pillars have faces that can be easily oriented, unlike the circular pillars. From the numerical and the experimental results is observed that the triangular micropillars with an orientation of 75° gives the best separation efficiency amidst the chosen configurations. Further, while discussing the process parameters involved in the design and fabrication of the DLD structures, it also signifies the possibility of particle separation using DLD structures at high row shift angles.
From the 3D profilometer images shown in Fig. 6d and e, it is evident that the shapes of the holes also play an important role in defining the etch rate and the aspect ratio of the holes. The etching rate is higher in circular than triangular holes. And since the area of exposure on the site beside the channels is higher, the rate of etching is much higher than the micro-holes. This can be attributed to the reach of the etchant gases into the etching region. The narrower the etching region, the lower the etching depth in comparison to the wider regions.
Bond strength between the chips in the microfluidic devices depends on the plasma exposure time and the post-exposure baking temperature. Moreover, the bond strength is also considered to be critical in designing the maximum permissible pressure difference in the microfluidic chip. Thus, using the Hagen–Poiseuille equation (eqn (8)), the maximum flow rate of the fluid flowing in the chip is evaluated and shown in Table 8. Further, the maximum flow velocity is calculated using the continuity equation (eqn (9)). From the T-peel tests, the results shown in Table 7 show that the oxygen plasma bonding bond strength is 10% higher than the argon plasma with a maximum bond strength of 0.404 N mm−1. It is observed that the bonding in the chips exposed to oxygen plasma for 5 min resulted in the highest bond strength in combinations carried out. This can be attributed to the formation of a large number of free sites for bonding thus increasing the availability of the bonding sites. The maximum flow rate and maximum fluid velocity achievable are 135 μL s−1 and 1.08 m s−1. It is also observed that the hydrophilic nature of the Ar plasma exposed surface does not vanish with time, as seen in O2 plasma bonding. As observed previously by Malecha et al.,31 the surface exposed to Ar plasma showed an irreversible change in the hydrophobicity. Thus, Ar plasma is not preferred in device fabrication as it induces a small amount of hydrophilicity. This affects the functioning of the microfluidic device by inducing unwanted drag forces. This suggests the application of oxygen plasma over argon plasma. Hence, in the development of the current device, oxygen plasma is used. The flow velocity of 0.5 m s−1 is used for the experiments as the available volume for the flow is constricted in the DLD zone, thereby increasing the fluid pressure.
With the flow parameters and the wall conditions, the devices were tested for blood particle separation. The effect of the placement, orientation and shape of the micropillars on the particle separation was studied. It is evident from the particle tracing simulations that every configuration results in a unique flow pattern. From the results obtained in the configuration C-25, C-30 and C-35, it can be concluded that the particles are moved towards the outlet 2. However, the flow pattern in the devices with the triangular pillars is different suggesting that the shape of the pillars is significant in directing the particles. It is seen that increasing the row shift fraction in the circular DLD increases the possibility of the particles moving towards the outlet 2 whilst the triangular pillars present a completely different pattern, the number of particles reaching the outlet 2 depends on the orientation of the pillars. This can be attributed to the inter-pillar gap and the velocity profile between the pillars. As observed by Loutherback et al.,22 the velocity profile of the fluid flowing between the triangular pillars is skewed in comparison to the fluid flowing between the circular pillars, thus affecting the streamlines. A similar phenomenon occurs when the orientation of the pillars is considered. The inter-pillar gap changes significantly; thus, the flow pattern obtained is skewed, and an orientation of 75° gives the best outlet 1 to outlet 2 ratio. The other important factor can be the first critical angle at the instant when the particle leaves the pillar that determines the path taken by the particles.4
The possibility of the particles getting stuck in the domain increases with the zero velocity zones offered to the flow on the walls of the microchannels and the walls of the pillars. The zero velocity zones are observed at the walls of the circular pillars.29
The recovery factor in these devices is also dependent on the configurations, it is seen that the recovery rate varied from 44% to 81%, and C-35 and T-30 result in the highest recovery factor when the pillars with the pointed vertices are considered. The recovery factors for the devices with pillars having rounded vertices were found to vary between 74% and 100%, this suggests that the rounded triangular micropillars provided a higher recovery in comparison to the pointed micropillars. The recovery factor of the configurations can be further improved using surface treatment with hydrophobic agents or other surface modification techniques.
The experiments were carried out to validate the simulations. The experimental flow study was carried out in two different ways, one keeping the device upright and the other with a reversed position, to understand the effect of gravitational forces. In the upright position, the outlets are facing upwards, and thus, there's a possibility of backward flow due to the requirement of high-pressure head. Thus, the collection tube was placed at a lower position such that the pressure head didn't affect the flow. The flow in the upright and reversed positions yields the same results. However, since the placement of the tubing in the devices is easier in the upright position, the devices in the upright position with an inverted microscope give an optimal configuration to capture the flow through the microchannels. Moreover, the strategy of collecting the samples at the outlets and enumerating the cells from the microscopic images helps reduce the error caused by visual counting using a haemocytometer.
In the manufacturing aspect, the increase in the row shift angle results in a reduction the number of pillars required for the particles to separate. The previous DLD chips consider a low inclination angle, which makes the manufacturing process difficult and also leads to the breakage of the micropillars, creating voids in the chips. In this study, an row shift angle of 25°, 30° and 35° is considered, thus reducing the pillars in the DLD zone. Increasing the inclination angle increases the inter-row and inter-column spacing, thereby increasing the number of ways the pillars can be fabricated. As observed in our previous study, laser micromachining was used to fabricate the DLD structures, but it was observed that the micro holes fused with each other when the inter-array gap was reduced to 25 μm.
Particle separation through a deterministic lateral displacement is a hydrodynamic phenomenon. Hence, it depends on the factors causing changes in the hydrodynamic forces. The laminar flow in the microchannels is affected by the roughness elements. The hydrodynamic drag and the pressure drop across the microchannels tend to change, thus aiding in particle separation. Literature suggests the application of sheath layers to avoid wall friction and considers the particle separation only between the pillars. In this study, the effect of wall friction is also considered as it is a significant factor in determining the separation efficiency of the device. Here, whole blood or diluted blood is flown through the microchannel, where the blood constituents experience the drag from the walls of the device until they reach the DLD zone and further experience deviation in the flow path due to the pillars placed in their flow direction. It is also seen that the separation efficiency depends on the first angle of incidence.
Thus, with this study, it is observed that efficient particle separation is possible even at higher row shift angles. The numerical studies can be further extended to 3 dimensional analysis to consider the friction and intercellular interactions in the particle movement. This work considers the stationary and rigid micropillars for its analysis, whereas at higher flow rates, the bending of the pillars can be observed, thus in the further studies the bending in the micropillars can be considered. The effect of the bending on the interpillar spaces, thus their effect on the particle separation can be analysed. The current work doesn't consider the effect of hydrodynamic nature of the pillars and the walls caused by the exposure to the plasma, in further studies, there is scope of introduction of these forces in determining the particle flow paths. As, the bond strength experiments show that the maximum achievable flow rates in these microfluidic devices are in the range of 100–135 μL s−1, this strategy can be utilised in the development of closed microfluidic devices with higher throughput.
In the further studies, the simulations can be improvised for 3-dimensional study as the drag forces on the walls play a vital role in determining the number of particles getting separated. The role of the plasma exposure and the surface quality of the devices can be studied, thus providing more insights on the effect of fabrication technique on the actual particle separation. The micropillars in the current study are considered to be stationary and rigid, but at higher velocities, the micropillars are deemed to bend, and as a result, there is a variation in the interpillar gap and thus might influence the particle flow path.
Ar | Argon |
p | Pressure |
Dc | Critical diameter |
RF | Radio frequency |
Ac | Cross-sectional area |
Re | Reynolds number |
DRIE | Deep reactive ion etching |
λ | Row shift fraction |
ρ | Density |
SSP | Single side polished |
DLD | Deterministic lateral displacement |
UV | Ultraviolet |
Γ | Dynamic viscosity |
v | Velocity |
EDTA | Edetic acid |
μ | Viscosity |
G | Gap size |
vx | x component for velocity |
DH | Hydraulic diameter |
vy | y component for velocity |
ICP | Inductively coupled plasma |
E | Young's modulus |
PDMS | Polydimethylsiloxane |
SU-8 | Negative photoresist |
PEEK | Polyether ether ketone |
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d3ra06431j |
This journal is © The Royal Society of Chemistry 2024 |