Nanoparticles (NPs) can be found throughout our direct environment as well as in a large range of consumer products. However, there is a lack of data on abundances and properties, which requires dedicated methods to identify and characterise NP entities. Single particle inductively coupled plasma-mass spectrometry (SP ICP-MS) is becoming one of the most relevant methods for counting NPs and studying NP composition and properties. Especially time-of-flight (TOF) technology for ICP-MS holds the key to study single particle composition and to perform non-target particle screening. This is achieved through the unique detection paradigm in TOF acquiring all isotopes across the periodic table at 35 kHz and faster. Nevertheless, non-targeted screening methods face limitations due to two factors. First, large data sets challenge data processing units and require substantial time, and second, for each isotope/element, a decision limit to distinguish between random noise and a SP signal has to be established. Here, we propose a novel lognormal approximation method to rapidly model background levels and decision limits. This method was integrated in a pre-analysis tool which detected SP data signatures of all m/z recorded via SP ICP-TOF-MS. Within seconds, particulate elements could be pinpointed in large data files and selected for more dedicated characterisation steps. In a proof of concept, we used this non-target screening method to identify inorganic NPs in selected samples of whisky, vodka, gin, and liqueur. Following qualitative analyses, number concentrations, compositions, masses, and size distributions were investigated. Besides Mn, Fe and Cu as expected NP entities, Ti, Ag, Au and Sn-based particles were found in investigated samples.
The development of single particle inductively coupled plasma-mass spectrometry (SP ICP-MS) was a turning point for NP analyses. It combines high counting rates with elemental analysis and can produce models on PNCs as well as on particle masses and sizes when suitable standards are analysed concurrently. The concept of SP ICP-MS is based on the analysis of NPs at low number concentrations to ensure the introduction of individual particles into the plasma. Here, single particles are atomised and elemental cations are formed. These are subsequentially extracted in discrete ion clouds for mass spectrometry and detected as time resolved pulses. Rapid mass analysers and detectors enable the counting of several hundreds of NPs per minute whilst recording extracted ion clouds from SP events with several data points at integrations times of 100 μs or below.6,7
The maturation of time-of-flight (TOF) technology for ICP-MS initiated a paradigm shift for the characterisation of individual particles. Previous mass analysers such as quadrupoles scanned sequentially and were limited by their speed, which restricted acquisitions to solely one selected isotope or small mass windows.8,9 As such, non-target approaches were depending on procedures to detect a statistically significant number of NP signals for one element for a set period before analysing the next. Consequently, screening procedures were time and sample consuming, and methodologies could still only target one isotope per detected particle.8 However, SP ICP-TOF-MS enables the rapid and continuous recording of full mass spectra at acquisition rates exceeding 10 kHz. While collision/reaction cells (CRCs) are typically employed to remove confounding ion species, the increased mass resolution of TOF analysers can help additionally to resolve interferences from analyte signals.10
Although SP ICP-TOF-MS may not reach the size detection limits (sDL) of quadrupole and sector-field-based methods, its multi-elemental capabilities permit non-target approaches and investigations on the composition of individual particles. However, the resulting data sets are considerably large and screening efforts are hampered by the ability to carry out data processing sufficiently fast. In this work, we present a rapid non-target screening method for inorganic NPs via SP ICP-TOF-MS. We developed an algorithm for the pre-analysis of data sets before applying dedicated algorithms for signal accumulation, and calibration processes. In this proof of concept, we selected spirits as complex samples that are exposed to varying conditions and materials during fermentation, distillation and storage. We anticipated that these steps were inherent with the formation or leaching of NP entities. Whisky, gin, vodka, and liqueur samples were selected as a few of the most popular spirit-based drinks and were subjected to the screening method. Identified NP entities were subsequently analysed regarding PNCs, composition and their distributions in mass and size. While this approach may be relevant for toxicological considerations, multielement analysis of ionic and particulate elements may be useful to confirm authenticity as well as to trace adulterations. The production of spirits is often connected to considerable effort and cost from the dedicated ingredients, manufacturing steps and extended maturation or storage procedures. While key components in different spirits have been investigated in profiling efforts to confirm authenticity and to recognise adulterations,11–18 no data is available regarding particulate analytes.
Analysed spirits are listed in Table 1 and were obtained from local distributors. They were diluted 1:
10 in ultra-pure water to limit matrix effects and enable direct analysis. The effects of the low remaining ethanol content was tested by comparing sensitivity of screened elements in ultrapure water and in 4% ethanol, respectively. Values deviated for the targeted elements only slightly (in average 2.7%) rendering more dedicated matrix-matching unnecessary.
Sample | Details |
---|---|
W1 | Mackmyra Bjorks, Swedish single malt whisky |
W2 | Talisker, single malt Scotch whisky, aged 10 years |
W3 | Longrow, peated single malt Scotch whisky |
W4 | Mackmyra Destination, Swedish single malt whisky |
W5 | Suntory Whisky Toki, blended Japanese whisky |
W6 | Speyside single malt Scotch whisky, aged 14 years |
W7 | Dewar's, blended Scotch whisky, aged 15 years |
V1 | Zubrowka, Bison Grass Vodka, Poland |
G1 | Bombay Sapphire, dry gin, England |
L1 | Amaretto, Il Barone, Italy |
L2 | Kruska, Kravica, Croatia |
To decide whether a particle has been detected or not requires statistical measures and is subject to two types of errors: false positive decisions (type I error, rate α) and false negative decision (type II error, rate β). Here, α and β are used to represent the probabilities of data being misinterpreted as a NP or a NP detection being missed. These errors enable to set underlying analytical limits: a decision limit based on an a posteriori assessment whether a NP has been detected, a detection limit to indicate the reliability of an a priori threshold as well as the quantification limit to provide a quantitative estimate.20 At low background levels, the decision limit is solely defined over the α-error and the mean of data set and can be used to count NP detection events over the critical value with a certainty of 1 − α. This approach requires a minimum of processing power and can be applied to TOF data sets efficiently.
The rapid mass spectra acquisition of m/z in SP ICP-TOF-MS requires the application of a sub-nanosecond analogue-to-digital conversion ion detector which somewhat complicates the modelling of background and decision limit. For large background values, a Gaussian-like distribution can be assumed and therefore, Gaussian statistics are used for thresholding. However, low background values do not follow a normal distribution. Unlike for quadrupole-based SP ICP-MS with secondary electron multiplier detectors, data produced by a SP ICP-TOF-MS is also not Poisson distributed and background thresholding cannot rely on Poisson statistics.21 Different elemental ions have varying sensitivities and produce a signal distribution which is subsequently calibrated into “counts”. This leads to non-integer intensity values and further, the rapid sub-nanosecond acquisition involves accumulation of multiple spectra before writing data to the disk. For low backgrounds, this can be modelled using a compound Poisson distribution that considers the single ion signal area (SIA).22–24
To enable non-target NP analysis, we propose an algorithm, which first decides whether to use Gaussian or compound Poisson statistics. The former was used when all data points were above 5 cts and the decision limit was selected as μ + 5σ. When mean background levels were low, compound Poisson statistics were approximated using a lognormal (LN) approximation methods as follows: a Poisson distribution with an expected rate (λ) equal to the signal mean was created. Only Poisson events k in the Poisson distribution with k > 0 and P(k) > 0.0001% were considered and approximated with the LN approach. For each value of k a LN distribution with a mean of k was then approximated from the sum of k LN distributions, each with a mean of 1 and a shape parameter determined from a LN fit of the SIA.25 This approximates the distribution produced by sampling the SIA k times, as happens in multi-ion events. The cumulative density function of each LN was then weighted by the Poisson probability of k and summed. The decision limit was then determined as the first value at which the summed cumulative distribution was greater than the desired zero-truncated quantile. This LN approximation method matched results produced by Poisson sampling the SIA as shown in Fig. 1, being significantly faster and only requiring knowledge of the shape parameter of a LN fit of the SIA (in this case 0.47). Given that files can reach the size of several gigabytes, screenings were further accelerated by limiting the number of evaluated data to the first 500000 data points for each element. Elements within the mass window of 45–210 amu with at least 25 ppm of data points above the respective decision limits were reported and were then selected for in-depth characterisation. The interested reader will find the underlying python script for this non-target approach in the ESI† and https://github.com/djdt/djdt.github.io/blob/feece3a08b1ff90131bbd9d618ba778509941467/scripts/non_target_screening_script.py.
Sample | Elements found at a NP detection score ≥25 ppm |
---|---|
Tap water | Fe, Cu, Zn |
W1 | Ti, Fe, Ag, Au |
W2 | Ti, Fe, Ag, Sn, Au |
W3 | Ti, Mn, Fe, Ag, Sn, Au |
W4 | Ti, Mn, Fe, Ag, Au |
W5 | Ti, Mn, Fe, Ag, Au |
W6 | Ti, Fe, Ag, Au |
W7 | Ti, Fe, Ag, Au |
V1 | Fe, Sn |
G1 | Fe, Cu, Sn |
L1 | Ti, Mn, Fe, Ag, Sn, Au |
L2 | Ti, Mn, Fe, Ag |
The signals of detected particulate elements were subsequently evaluated for potential interferences. However, the application of a CRC and He/H2 cell gas flows enabled the attenuation of most polyatomic interferences and significant levels of naturally abundant isobars were ruled out following the interrogation of isotopic abundances, as demonstrated for Ti in Fig. 3. Fig. 3A shows the transient analysis of 4 selected isotopes 48Ti, 107Ag, 120Sn and 197Au in W3 as a representative sample. 48Ti (isotopic abundance: 73.7%) is often interfered by the isobaric 48Ca interference and sequentially operating mass analysers are usually set to target the less abundant 47Ti (7.4%) to avoid confounding Ca isotopes and to ensure accurate analysis, although sensitivity decreases drastically. The use of a TOF mass analyser enabled the simultaneous analysis of all isotopes across the selected mass range and allowed the interrogation of isotope ratios to indicate the presence of spectral interferences. Fig. 3B shows the analysis of all Ti isotopes in a single NP signal event and was used to determine relative isotopic abundances. Fig. 3C shows the resulting 47Ti/48Ti ratios for all detected Ti NPs and data was fitted with a linear function to return a slope of 0.105 as experimental isotope ratio, which was in line with the natural ratio of 0.103. As such, no 48Ca interference was evident and 48Ti was targeted to benefit from the higher sensitivity. This approach was applied for all analysed polyisotopic elements and isotope ratios were consistent with theoretical values except for Fe, for which the experimental abundance of 56Fe was significantly biased. Large particles especially deviated regarding isotope ratios, which was a result from a limited dynamic range. However, other Fe isotopic abundance were in line with theoretical values as shown in Fig. S1 in the ESI.† Therefore, the 56Fe isotope was solely considered to determine PNCs and 54Fe was used to determine size and mass distributions.
The non-target screening found Sn NPs in multiple samples and calculations considered Sn4+ as most common oxidation state in an oxide phase with a mass fraction of 0.788. Considering each particulate event allowed to model size and mass distributions in histograms. An example is shown for sample W2, which contained the highest number concentrations of SnO2 NPs (39500 ± 1800 mL−1). The SnO2 mass and size distributions in this sample are shown in Fig. 4 and data on other elements is available in Table S1.† For this sample, the mean SnO2 NP mass was 3.5 ± 4.2 fg, the mean size was calculated to be 89 ± 29 nm and the size detection limit was 39 nm.
Across all samples, ionic background levels for Cu, Fe and Mn were relatively high and were assumed to arise from contact with distillation and container materials or from drinking water sources. Detectable Fe2O3 sizes were relatively variable and were smallest in tap water (58.6 ± 24.3 nm) ranging up to 300 ± 112 nm for sample W3. Mean TiO2 sizes varied between 80.4 ± 45.5 nm (W4) and 148 ± 54 nm (L1). Ag NPs were detectable in samples where Au NP were detectable as well. Only sample L2 contained Ag NPs but no Au NPs. Au and Ag NPs and exhibited relatively consistent sizes between 30.7 and 44.4 nm and, 27.4 and 36.0 nm, respectively. Mean SnO2 NPs sizes were between 133 and 169 nm. The mean sizes and standard deviations of Au, Ag, MnO, SnO2, Fe2O3 and TiO2 across all samples are illustrated in Fig. 5. More details on absolute PNCs, ionic background, mean masses, sizes, standard deviations, and figures of merit are listed in Table S1.†
![]() | ||
Fig. 5 Overview of selected NP identified in analysed samples. Shown are the mean size of NP entities and the standard deviations. |
It is unlikely that detected NPs have toxicological implications as absolute PNCs are in a comparably low range when for example compared against drinking (tap) water. Particles based on elements such as Sn, Ag, Fe and Cu have previously been described in tap water and have likely a natural or incidental origin (e.g., plumbing or previous treatment processes).26 It is possible that particles found in spirits originate from different drinking water sources, are a result from leaching processes (e.g., from the distillation process, or during storage), or are intentionally included as additive to improve organoleptic properties.27 Knowledge on the levels and properties of ionic and particulate elements may be useful to confirm authenticity of expensive and elaborated spirits. While element profiles have been investigated for this purpose in the past,16,17 distinguishing between particulate and ionic entities as well as comparing numbers, compositions and sizes may add additional dimensions which facilitate discrimination between different spirits as well as to carry out provenance analysis. However, this needs to be evaluated by targeting a larger sample size consisting of different brands from various areas and countries.
In a proof of concept, NPs were qualitatively and quantitatively interrogated in spirit samples. Samples were subjected to the developed screening approach and found particulate elements were calibrated to determine PNCs, figures of merit, mass and size distributions. The simultaneous acquisition capabilities of SP ICP-TOF-MS enabled the determination of isotope ratios to confirm the absence of spectral interferences and to ensure accurate calibrations. The elements Sn, Fe, Ti, Ag, Au, Cu, Zn and Mn were found in particles in spirits and tap water and the analysis of PNCs, sizes/masses and ionic background may have utility for profiling efforts.