The proportion of full and empty capsids represents a critical quality attribute of adeno-associated virus (AAV)-based therapeutics. In this study, pH-gradient anion exchange chromatography was utilized for the separation of full and empty capsid species. The developed method allowed for applicability to multiple AAV serotypes and facilitated subsequent mass spectrometric detection of intact AAVs. This is the first study demonstrating generic applicability as well as mass spectrometric compatibility, allowing for a more sophisticated analysis of AAV-based gene therapy and paving the way for future developments in the field.
Separation of full/empty mixtures of the three serotypes after gradient optimization are shown in Fig. 1. Notably, all serotypes were analyzed using the same gradient, suggesting that the chromatographic conditions might also be applicable to other serotypes, albeit these were not investigated due to material non-availability. All three AAVs elute late in the gradient which is primarily caused by the high pH at the gradient start, equivalent to a mobile phase B concentration of 20%. Starting at lower pH was investigated but resulted in insufficient column binding or high run-to-run variability (Fig. S1, ESI†). The developed gradient facilitated the separation of full and empty capsids, represented by the two dominant peaks in each case, as verified later. While AAV5 and 6 seem base line separated, AAV8 shows partial separation, a clear distinction of two discrete peaks is however feasible. It should be mentioned that AAV6 benefitted from separation at higher temperature which significantly improved peak shape and that a comparable improvement was not observed for AAV5 and 8. Based on BLAST alignment searches we were able to conclude that sequence homology alone is not a reliable predictor for the optimum separation temperature. The mobile phases used were of low ionic strength and composed of MS-friendly components only, allowing for subsequent mass spectrometric interfacing if desired. Despite the relatively low ionic strength and limited buffering capacity, full column equilibration was achieved in 12 minutes. A flow rate of 0.2 mL min−1 was employed, which allowed for swift column equilibration as well as full desolvation of the LC effluent in the MS interface in case of mass spectrometric hyphenation. Details describing experimental procedures can be found in the ESI.† After initial method development, the quality of the separations was further investigated. Elution order was determined first by the injection of full and empty material independently, in addition to the mixtures shown in Fig. 1. The resulting chromatograms are depicted in Fig. 2. In pH-gradient AEX a starting mobile phase pH is chosen where all analytes are negatively charged and initially bind to the column. During a gradient from high to low pH, analytes successively elute off the column as the mobile phase pH equals their respective pI's, facilitating separation. In all cases, surprisingly, full capsids were eluting first, which is counterintuitive considering the apparently higher pI of the empty particle and contrary to what is reported in literature for salt-gradient mediated elution.
![]() | ||
Fig. 1 Separation of AAV5, 6 and 8. The two dominant peaks in each chromatogram represent full and empty capsid species. The mobile phase gradient is indicated in yellow. |
We propose that the acidic mobile phase pH at the end of the gradient results in a conformational change, thus altering the surface charge distribution of the particles and in consequence their pI. VP 1 was previously reported to undergo structural changes if subjected to heat stress or low pH.11,27 The N-terminal phospholipase A2 domain of the capsid protein was shown to undergo a reversible unfolding/refolding process at a pH of 5.5–4.0 which leaves the capsid intact. The authors of the study speculated that the underlying mechanism was charge repulsion between the N-terminal domain, internal capsid residues and DNA, which leads to externalization of the phospholipase A2 domain.11 This process almost certainly triggers changes in the surface charge distribution of the capsid and most likely plays out differently under a complete absence of cargo DNA. It is conceivable that sequence and size of the cargo DNA impact the magnitude of this structural change and that different transgenes could be associated with different retention behaviour in chromatography, which was however not investigated in the current study. Interestingly, full/empty samples showed high variability in their proportions of full and empty capsids, despite being all based on theoretical 1:
1 mixtures (Fig. 1 and 2, top panel). Calculations of the full-to-empty ratios revealed that samples did not perfectly represent the expected proportions. AAV6 shows the largest discrepancy with the empty particles making up a total of 77.2 instead of 50% of capsids in the sample (Table S1, ESI†). Injections of individual species shown in Fig. 2 (mid and lower panels) clearly demonstrate that the reason is insufficient purity of commercially sourced material, especially the full capsid samples. Analysis-related bias can be ruled out as results obtained for AAV5 were benchmarked against and showed excellent correlation to previously conducted CDMS experiments (Fig. S2, ESI†). In terms of chromatographic performance, it appears that the different capsid fill states of AAV5 are best separated, followed by AAV6 and AAV8. AAV8 moreover shows an increased peak width when compared to serotypes 5 and 6. Calculated chromatographic resolution for each separation supports these findings (Table 1). With an experimentally determined chromatographic resolution of 4.30, AAV5 shows the highest resolution between full and empty capsid peaks followed by AAV6 with a calculated resolution of 2.99. AAV8 is not visibly baseline separated which is reflected by a chromatographic resolution of 1.05. Calculated peak capacities follow the same trend with AAV5 showing a value of 63.5, followed by AAV6 with 56.8 and lastly AAV8 with 31.6, which was expected given the visibly broader peaks.
Serotype | Resolution | Peak capacity |
---|---|---|
AAV5 | 4.30 | 63.5 |
AAV6 | 2.99 | 56.8 |
AAV8 | 1.05 | 31.6 |
Also run-to-run variability was evaluated and was found negligible for retention time with relative calculated standard deviations of below 0.1% for full and empty capsid peaks of all three AAVs tested (Table S2, ESI†). Relative standard deviations of peak areas were higher, in a range of 1 to 5%. The largest deviation was observed for AAV6 (4.79%). It should be noted that AAVs are prone to non-specific adsorption to solid surfaces which can occur throughout all stages of production and characterization.28 Materials that were shown to retain AAV particles include plastics, stainless steel and importantly also glass, the material here used to contain AAV samples in the LC autosampler. Variations in peak area potentially result from adsorption to the HPLC vial, rather than being method related, which is also supported by the observed run-to-run decrease in the fluorescence response, as depicted in Fig. S3 (ESI).† Furthermore, AAVs reportedly experience diluent dependent sample degradation over time, which can be a factor contributing to the variation seen.32 AAV6 is shown as an example, the fluorescence signal can be seen to slightly decrease from run 1 to 2 and from 2 to 3. Moreover, it appears as if the extent of adsorption is not the same for full and empty species with the full particle peak showing a more pronounced signal loss over time.
After investigation of the chromatographic performance, mass spectrometric interfacing was explored by coupling the separation system to a Thermo Scientific Q Exactive UHMR mass spectrometer with a mass range adequate for the analysis of macromolecules in the megadalton range. While previous investigations show that native mass spectrometry does not reliably facilitate mass determination of AAVs, literature suggests that full and empty particles carry the same extent of positive charges after ionization.29 Consequently, their MS signals appear in different regions in the m/z domain, depending on whether they carry cargo DNA or not, allowing for their discrimination.30Fig. 3a shows the total ion current chromatograms (TICCs) of the three AAV full/empty mixtures corresponding to separations in Fig. 2. Clear MS signals were obtained in all cases with AAV5 and 6 showing good signal intensity while the intensity of AAV8 appeared lower. This was to some extent also observed in fluorescence traces in Fig. 2, likely caused by the increased peak width of that particular serotype.
![]() | ||
Fig. 3 (a) TICCs corresponding to full/empty fluorescence traces shown in Fig. 2, a magnification of the retention time region of interest is presented. Peaks corresponding to full particles are labelled in blue, peaks corresponding to empty capsids are labelled in red. (b) Associated native mass spectra obtained after integration of peaks labelled in (a). The colour coding is in line with (a). |
Other contributing factors could be a higher non-specific adsorption of AAV8 to surfaces or lower MS ionization and ion trapping efficiencies when compared to the other two serotypes investigated. Nonetheless, the MS signal strength allowed for collection of useful mass spectra also for AAV8.
Fig. 3b shows the mass spectra corresponding to peaks in A after peak integration and spectral averaging. In all cases spectra of the earlier eluting peak appeared in a m/z region of around 30000 while spectra from the later eluting peak were centered at around 22
500 m/z. Assuming a similar molecular charge after MS ionization, the earlier eluting peak must contain species of significantly higher molecular mass which can of course be explained by the presence of cargo DNA, adding an additional mass of approximately 0.8 MDa. As was already anticipated, the complexity of the samples did not allow for deconvolution and mass determination but indeed for a clear differentiation of full and empty capsids, also confirming the elution order previously determined based on injection of individual full and empty samples (Fig. 2).
The presented method allows for the pH-gradient separation of full and empty capsids of multiple AAV serotypes based on MS-friendly anion exchange chromatography. Fluorescence signals can be utilized to precisely quantify full and empty capsid proportions while subsequent mass spectrometric interfacing allows for peak assignment. Further investigations will be needed to better understand the relationship between AAV structure, including type of the transgene encapsidated and retention behaviour. Recently, interfacing of liquid chromatography to CDMS was demonstrated for the first time.31 While native MS did not allow for precise mass determination and the distinction of possibly present forms other than full and empty species, CDMS might very well be capable of doing so. Whether liquid chromatography in combination with CDMS is feasible for highly heterogeneous macromolecules such as AAVs still needs investigation, but the foundation has been laid for the development of more sophisticated and informative AAV characterization strategies which are based on liquid chromatography and mass spectrometry.