Valentina
Preziosi
*a,
Mario
Barra
*b,
Giovanna
Tomaiuolo
a,
Pasquale
D’Angelo
c,
Simone Luigi
Marasso
cd,
Alessio
Verna
d,
Matteo
Cocuzza
cd,
Antonio
Cassinese
be and
Stefano
Guido
afg
aDepartment of Chemical, Materials and Production Engineering - University Federico II, P.le Tecchio 80, I-80125 Naples, Italy. E-mail: valentina.preziosi@unina.it
bCNR-SPIN, c/o Department of Physics “Ettore Pancini”, P.le Tecchio, 80, I-80125 Napoli, Italy. E-mail: mario.barra@spin.cnr.it
cIMEM-CNR, Parco Area delle Scienze 37/A, I-43124 Parma, Italy
dChi-Lab, Department of Applied Science and Technology, Politecnico di Torino, C.so Duca degli Abruzzi 24, 10129 Torino, Italy
eDepartment of Physics “Ettore Pancini”, University Federico II, P.le Tecchio 80, I-80125 Naples, Italy
fNational Interuniversity Consortium for Materials Science and Technology (INSTM), 50121 Firenze, Italy
gCEINGE, Advanced Biotechnologies, 80145 Napoli, Italy
First published on 22nd November 2021
In this paper, for the first time to the best of our knowledge, organic electrochemical transistors are employed to investigate the electrical response of human blood, plasma and alternative buffer solutions that inhibit red blood cell (RBC) aggregation. Our focus is on selecting a suitable electrolytic platform and the related operating conditions, where the RBC effect on the OECT response can be observed separately from the strong ionic environment of plasma in whole blood. The transient response of whole blood to pulse experiments is characterized by two time constants, which can be related to blood viscosity and to the capacitive coupling between the ionic and electronic components of the overall system. The role of capacitive effects, likely due to enhanced double-layer formation by negatively charged RBCs, is also confirmed by the increase of transconductance which was found in RBC suspensions as compared to the suspending buffer. Overall, the complex behavior found in these experiments provides new insights for the development of innovative blood-based sensing devices for biomedical applications.
Such a highly sensitive tool and miniaturized devices could be coupled to provide robust platforms representing in vitro physio-pathologically relevant microenvironments. These could be able to mimic cell and organ level organizational structures, thanks to their user-defined design, small length scales and optimal control and sensing of mechanical and tissue environments. Taking advantage of microfluidics principles, organ-on-chip devices are integrated platforms consisting of microfluidic channels populated with several cell types to create minimal functional organ units for drug testing and modelling of human physiology and disease.14–16 Beyond specific cell lines, blood, i.e. the fluid tissue that delivers oxygen and nutrients to tissues and cells and that is responsible of the immunological response, is the most exploited biological fluid for biomedical applications. Blood is a complex fluid consisting in a suspension of cells, erythrocytes (red blood cells – RBCs), leucocytes (white blood cells – WBCs) and platelets, immersed in a low viscous fluid called plasma.17
In this scenario, OECTs can be particularly precious, unravelling the issue of blood handling and sensing in integrated microfluidic devices. In particular, RBCs, being the major constituent of blood, are deformable cells that have a great impact on the physiological flow of blood in microvasculature.18,19 Their abnormal shape and concentration, indeed, can lead to organ and tissue disfunction and death.20–23 Moreover, plasma, being a water solution of cations like Na+, K+, Ca2+ and Mg2+, anions like Cl− and HCO3−, and high molecular weight substances like proteins and lipoproteins, strongly impacts the electrical response of blood.24,25 For a long time, there has been a lot of interest about the electrical behavior of human blood, in terms, for example, of the conductivity (or its reciprocal, the resistivity), given its strict correlation with blood physiology and the repercussion of blood electrical variations on biomedical devices. In 1873, Maxwell26 published a book containing a theoretical dissertation about the conductivity of a suspension of non-conducting spheres in a conducting medium. Some years later, Hober determined for the first time the different resistivity of blood, at low and high frequencies, due to the presence of a membrane around RBCs.
Later on, first Fricke27 and then Velick and Gorin,28 extended Maxwell's work and applied it to blood, assuming RBCs as non-conducting ellipsoids and plasma as a conductive medium. Other works followed that correlated blood conductivity with the presence of erythrocytes and plasma leading to the conclusion that, even if plasma is a complex fluid full of ionic species, the presence of red cells plays a key role in the electrical response of blood.29,30 In particular, some studies addressed the change in human blood electrical properties in static conditions due to the haematocrit (Hct), i.e. the ratio of the volume fraction of the blood cells and the total volume, and to the red blood cell random orientation. Other studies instead focused on the effect of RBCs’ parallel orientation under flow confirming that, even if RBCs display a low conductivity with respect to plasma, their impact cannot be neglected.29 Despite all the studies on the electrical properties of human blood, the effect of RBCs in terms of electrical conductivity is still controversial, making the exploitation of blood-based electrical devices still an open issue.
In the present work, the problem of the response of OECTs when human blood, plasma and an alternative buffer solution containing RBCs in suspension are used, is addressed. Building on previous reports by our group on the use of OECTs as sensing devices31–37 and on the analysis of blood flow in blood-on-chip devices,38,39 the main goal of this paper is to exploit OECTs as a reliable platform to distinguish the effect of RBCs from those related to the plasma, which is characterized by a high ionic concentration.
A major challenge associated with running OECTs with blood and RBC suspensions is related to the particulate nature of blood and to the peculiar properties of RBCs, such as deformability and a cell's tendency to aggregate under static or weak flow conditions, as well as to the number of electrolytes present in plasma. Despite the key role that blood could play in integrated devices, no experimental study on the use of OECTs for blood testing has been published so far, to the best of our knowledge. Our approach is based on the exploitation of OECTs with whole blood and RBC suspensions at different values of hematocrit to separate the effects of RBC and plasma. Different responses have been found in this way.
:
100 Anticoagulant Citrate Dextrose (ACD) and Albumin solution. To this end, blood was centrifuged (Hettich Universal 32 R) at 1600 revolutions per minute for 15 min and at 20 °C to separate the RBCs from other cellular components and plasma. ACD was purchased from Haemonetics (0.6% citric acid, 1.1% anhydrous dextrose, 2.3% sodium citrate, 96% water) and used to avoid clotting. Fetal bovine serum (FBS-Albumin) was purchased from Merck and it was dissolved in ACD (4 g in 10 ml ACD) by mixing the solution in a beaker on a magnetic stirrer for about 20 min at room temperature.
Conductance measurements of ACD and ACD plus albumin solutions, performed by using a conductivity meter (PC 2700, Eutech Instruments), gave values about 14 mS for ACD and 10 mS for ACD plus albumin. These values are similar to the typical ones measured for whole blood.40
nm/100
nm) on a Si wafer (100) finished with 1
μm of thermal oxide. A 200 nm thick PEDOT:PSS layer (Clevios PH 1000, doped with 5 volume % ethylene glycol, 0.1 volume % dodecyl benzene sulfonic acid, and 1 wt% of GOPS (3-glycidyloxypropyl)trimethoxysilane) was obtained by spin coating on the source and drain electrodes and subsequently patterned via photolithography (Microchemicals AZ9260 resist) and etching with O2 plasma. A PDMS chamber, with an internal volume of 150 μl, was aligned with the PEDOT:PSS channel and irreversibly bonded on the surface (Fig. 1a). The final device geometry (see Fig. 1b) is characterized by a width to length (W/L) ratio equal to 30 (W = 6 mm, L = 200 μm). In particular, drain and source electrodes are 250 μm in width and 7 mm in length both having a contact pad of 1 mm × 2 mm.
During the overall time required to complete the OECT measurement protocol (about 30 minutes), we can assume that the RBCs were sedimented. However, our experimental findings suggest that the cell sedimentation does not play a significant role on the OECT response, as will be discussed in the Results section. After each measurement, fresh blood was replaced and all the experiments were carried out at room temperature.
Gold gate electrodes were basically considered to study the OECT response. After gate immersion into the reservoir, a positive voltage between the gate and source, VGS, was applied in order to drive the PEDOT:PSS-based OECT operation in the depletion-mode regime.10
In particular, in the first set of experiments the OECT response was visualized by applying sequences of VGS pulses, with a duration of 60 s and progressively-increasing amplitudes going from 0 to 0.6 V (step 0.1 V) at a fixed drain–source voltage, VDS = −0.1 V (Fig. 3a). The drain–source current (IDS) and the corresponding gate–source current (IGS) were thus recorded as a function of time. Hence, the OECT current modulation capability could be expressed by the ratio, ΔIDS/I0, derived from the pulsed measurements and described in detail elsewhere31 (Fig. 3b).
Measurements in Fig. 3 confirm that, even in this complex media, OECTs operate correctly as depletion-mode transistors under the application of positive VGS (i.e., IDS is lowered in absolute value). It should also be mentioned that, in this case, the measurements were exploited up to VGS = 0.6 V, due to the occurrence of device response degradation at larger VGS (namely, the lack of the correct depletion-mode operation). While comparing the OECT responses in blood and plasma with gold electrodes, we notice that only small differences are appreciable mainly for VGS between 0.2 and 0.5 V, with slightly enhanced current modulation values in blood. The strong similarity between the responses in these two fluids can be likely ascribed to the very large ionic concentrations in plasma.
In order to get additional insights about the OECT response in plasma and blood, the IDS(t) transient pulses, achieved with the gold electrode and for the different VGS pulses, were further analyzed. Hence, these experimental curves were normalized with respect to the initial value and shifted at t = 0 s. The obtained datasets are reported in Fig. 4a for whole blood and in Fig. 4b for plasma. Then, following the approach introduced in ref. 31, these data were fitted by using the double-exponential equation:
![]() | (1) |
![]() | ||
| Fig. 4 (a) Set of IDSvs. time curves (normalized to the initial value and time shifted at t = 0 s) recorded at different VGS for blood and (b) plasma using a gold gate. (c) τ1 and (d) τ2 time constants as a function of VGS extracted by fitting experimental IDS(t) curves with eqn (1) for whole blood and plasma. (e) OECT modulation ratio values estimated by the fitting procedure in the steady state conditions as a function of VGS for plasma and whole blood. | ||
For better clarity, a single IDS(t) pulse, compared with the related fitting curve, is shown (Fig. S2a) in the ESI,† where the overall behavior of A1 and A2 achieved for the whole set of transient measurements is also presented (Fig. S2b and c, ESI†).
Appreciable differences between blood and plasma can be observed in Fig. S2b and c (ESI†) for the A1 pre-factor with respect to the A2 one. In the case of τ1 and τ2, for both time constants, blood experimental data provide higher values in the VGS range between 0.2 and 0.5 V (Fig. 4c and d). As a whole, these data suggest that the OECT response in blood is able to have larger IDS modulation but its dynamics as a function of time is slightly slower if compared with the case of plasma.
As a further confirmation of this interpretation, in Fig. 4e we report the estimated current modulation at the end of the 60 s pulse conditions (i.e. derived from the ISS fitting parameter in eqn (1)). Here, the enhanced current modulation capability in blood is still more evident than what was observed in Fig. 3b.
As discussed in ref. 31, the larger values of τ1 should be ascribed to the enhanced viscous character of whole blood.43 This type of observation is in qualitative agreement with a recently-proposed model where the diffusion of cations throughout the electrolytes was correlated to the solution viscosity, in analogy with the Einstein–Smoluchowski theory.44 On the other hand, the short time constant τ2 should be more related to the capacitive coupling between the ionic and electronic components of the overall system. In our experiments, in particular, by considering the size of the gate immersed area and the overall dimensions of the PEDOT:PSS channel with the related volumetric capacitance (CV ∼ 40 F cm−3), we can infer that the OECT response is basically ruled by the electric double-layer (EDL) capacitance (CG) at the gate-electrolyte interface (i.e. CG ≪ CV).45 In this scenario, the increased values of τ2 (with VGS between 0.2 and 0.5 V) for blood may be associated to larger values of CG as a possible effect of the presence of RBCs and related motion toward the gate surface.
As reported in the literature,46,47 a RBC's surface is negatively charged and hence, in an electrolytic solution, an electrical double layer is present also around the RBC. Under the effect of an external field (i.e. under positive VG during the OECT operation) RBCs, with the surrounding charge layers and the related diffusive components, are attracted toward the gate electrode. As a whole, this effect should produce a local increase of the ionic concentrations close to the gate surface, providing the final increase of CG. Although this is a tentative explanation requiring additional and more specific experiments to be definitely proved, it is to be highlighted how the hypothesized CG capacitance increase in blood vs. plasma is compatible also with the observation of the corresponding larger gate current IGATE values (see the data below and refer to Fig. S3b, ESI†). Given the polarizable nature of the gold electrode, in this case, a direct relation between the gate current and electrolyte-gate capacitive couplings is indeed expected.
Complementary experiments were also performed by using a platinum wire as the gate electrode. In particular, Fig. S3 (ESI†) shows IDS(t) pulsed measurements achieved in this case for whole blood and plasma. As shown, the OECT responses with the platinum gate electrode are even more similar and no significant differences can be observed in terms of ΔIDS/I0 values which are, however, larger (by a factor close to 2) than those measured with the gold electrode. These findings highlight the less polarizable nature of platinum when compared with gold, in agreement with other recent works where the occurrence of faradaic reactions at the platinum surface was indeed exploited to develop specific sensing devices.48–50 In particular, the excellent electrocatalytic activity of Pt (even in the form of nanoparticles incorporated in the PEDOT:PSS channel) toward H2O2 oxidation was exploited to demonstrate the efficient detection of neurotransmitters.51 Overall, it is essential to remember that the OECT response can be strongly affected by the physiochemical nature of the gate.52,53
ACD is commonly employed as an anticoagulant to prevent the formation of cell aggregates and albumin, being the most abundant protein in blood plasma, operates, at an appropriate concentration, as a nutrient for RBCs. ACD and albumin supplemented ACD solutions have been used by the authors in other studies18,19,55 and have been preliminarily tested here as electrolytes in the OECT (Fig. S4, ESI†). As shown, ACD and ACD plus albumin solutions show a very similar OECT response in terms of the current modulation capability, with the measured ΔIDS/I0 values remaining remarkably smaller than those observed in plasma and blood (by about a factor of 5 at VGS = 0.6 V). Fig. S4b (ESI†) also shows the IGATE values achieved for ACD and ACD plus albumin solutions in comparison again with blood and plasma. This trend reproduces very well that observed for the ΔIDS/I0 values with the IGATE being strongly reduced for the ACD and ACD plus albumin solutions.
Once characterized in terms of their baseline OECT modulation signal, ACD plus albumin solutions were used as a buffer to suspend the RBCs. Two different RBC concentrations, at 25% and 45% (the latter corresponding to the hematocrit in whole blood), were considered here.
In this regard, Fig. S5a (ESI†) shows the ΔIDS/I0 modulation curves achieved for the whole blood, the basic ACD plus albumin solution and for the same type of solution containing 25% and 45% of RBCs. These data outline unambiguously how the OECT response is affected by the presence of RBCs. In particular, the responses recorded in the solutions containing the two RBC concentrations, 25 and 45% (the latter corresponding to whole blood Hct), exhibit an intermediate character in comparison with the whole blood and the basic suspending medium. The IDS modulation capability was somewhat enhanced for the larger RBC concentration, at increasing VGS (Fig. S5b, ESI†).
In order to further analyze the OECT response in these fluids, transfer curves at VDS = 0.1 V were also recorded. In this case, the investigated VGS range was extended up to 0.8 V without observing any degradation signature in the depletion-mode OECT response.
Normalized transfer curves, reporting the change in the device current IDS normalized to the value measured at VGS = 0 V, are shown in Fig. 5a. Coherently with the features observed in the ΔIDS/I0 curves (Fig. S5a, ESI†), the responses recorded in the ACD plus albumin solutions containing the two RBC concentrations exhibit again an intermediate character. A more careful inspection of these measurements highlights also a different VGS dependence. This specific behavior was better emphasized by analyzing the OECT transconductance
which is a very important parameter simply defined as the derivative of the channel current IDS with respect to the gate voltage VGS. In particular, such a parameter describes the OECT signal amplification performances.56,57 In Fig. 5b, transconductance (gm) is plotted as a function of VGS for all the solutions investigated and very different trends can be observed. Indeed, gm rises at increasing VGS for all fluids, but the specific VGS value where this behavior starts and the presence of a peak, with the related magnitude, are characteristic of any solution.
In particular, three different regions can be identified:
– The low VGS range (approximately up to VGS = 0.3 V) where all gm curves assume values of few hundreds of μS, with a slight VGS dependence. Here, the solutions containing RBCs seem to display smaller gm even in comparison with the basic ACD plus albumin suspending medium;
– The intermediate VGS region (approximately between 0.3 and 0.6 V) where the gm rising behavior is much more evident. In this case, in comparison with the ACD plus albumin suspending medium, the presence of a major concentration of RBCs tends to reduce the VGS threshold above which this trend is more pronounced.
– The high VGS range (between 0.6 and 0.8 V), where, except for the basic ACD plus albumin suspending medium, gm exhibits a peak for all the solutions. Here, the ACD plus albumin solutions with 25% and 45% RBC concentrations differ for both the peak amplitude and the VGS value where this occurs.
Thus, the gm behavior in the intermediate VGS region, with higher trans-conductance for the ACD plus albumin solution at 45% RBC concentration with respect to that at 25% RBC, seems to suggest again an increase of the capacitive coupling (gm ∝ CG since CG ≪ CV) between the gold gate electrode and the electrolytic medium. This occurrence agrees well with the abovementioned considerations about the OECT transient response in blood (i.e. larger values of the short time constant τ2), pointing again at the role of negatively-charged RBC in enhancing the CG value in the same VG range.
On the other hand, when VGS values increase (between 0.6 V and 0.8 V), the OECT response reaches a different regime where the major RBC concentration in ACD plus albumin solutions provides a reduced IDS modulation capability (i.e. the gm peak amplitude is lowered and occurs at lower VGS). This condition underlines a weaker penetration of cationic species into the PEDOT:PSS film and is compatible with a field effect-like operation mode localized at the PEDOT:PSS surface.58
The physical interpretation of this phenomenon is not straightforward but it could be tentatively explained in terms of an increasing repulsion effect among the negatively-charged RBCs close to the gold surface or to the detrimental occurrence of RBC sticking effects at the same surface. Both occurrences, indeed, could degrade the electric double layer formation and make the progressive enhancement of the gate signal ineffective in promoting a further cationic uptake by the active channel. It is worth mentioning, in any case, that while it was previously shown that the position and magnitude of the gm peak in the OECT response can be tailored as a function of the device geometrical layout and/or the biasing conditions, here we found that these features can be clearly associated also with the nature of the analyzed electrolyte.59,60
As a final comment suggested by the data summarized in Fig. 5, we want to highlight how the transconductance curves recorded for blood and ACD plus albumin solution at 45% RBC concentration display a very different behavior, even in the presence of the same RBC concentration.
As discussed above, this finding is obviously related to the different suspending medium, but it is also a demonstration that the RBC sedimentation does not affect ion diffusion towards the PEDOT:PSS channel. On the other hand, if RBC sedimentation on the PEDOT:PS surface was the dominant feature affecting the operation of OECT here investigated, higher IDS modulation values would be expected for plasma in comparison with the whole blood. This occurrence was clearly not observed during our experiments.