Annemarijn S. M.
Steijlen
*a,
Jeroen
Bastemeijer
a,
Pim
Groen
b,
Kaspar M. B.
Jansen
c,
Patrick J.
French
a and
Andre
Bossche
a
aDelft University of Technology, Faculty of Electrical Engineering, Mathematics & Computer Science, Mekelweg 4, Delft, 2628 CD, The Netherlands. E-mail: a.s.m.steijlen@tudelft.nll
bDelft University of Technology, Faculty of Aerospace Engineering, Kluyverweg 1, 2629 HS Delft, The Netherlands
cDelft University of Technology, Faculty Industrial Design Engineering, Landbergstraat, 15, Delft, 2628 CE, The Netherlands
First published on 17th November 2020
This paper presents a method to continuously collect and reliably measure sweat analyte concentrations during exercise. The method can be used to validate newly developed sweat sensors and to obtain insight into intraindividual variations of sweat analytes in athletes. First, a novel design of a sweat collection system is created. The sweat collection patch, that is made from hydrophilized foil and a double-sided acrylate adhesive, consists of a reservoir array that collects samples consecutively in time. During a physiological experiment, sweat can be collected from the back of a participant and the filling speed of the collector is monitored by using a camera. After the experiment, Na+, Cl− and K+ levels are measured with ion chromatography. Sweat analyte variations are measured during exercise for an hour at three different locations on the back. The Na+ and Cl− variations show a similar trend and the absolute concentrations vary with the patch location. Na+ and Cl− concentrations increase and K+ concentrations seem to decrease during this exercise. With this new sweat collection system, sweat Na+, Cl− and K+ concentrations can be collected over time during exercise at medium to high intensity, to analyse the trend in electrolyte variations per individual.
Until now, little is known about how sweat analytes change over time and how they can be used to monitor athletes. The physiological literature about this topic is scattered and papers often show contradicting results. The application of sweat analytes in monitoring the physiological status of athletes is currently limited, because little is known about how sweat analyte concentrations relate to blood values.1,2
An important reason for this is the lack of suitable sweat collection methods for physiologists, who are studying the physiological mechanisms behind sweating. Local methods include collection with absorbent patches, pouches and the Macroduct sweat collection system.3–6 All these methods require repeated placement and removal of these devices when you want to measure electrolyte variations during an exercise. Several challenges of sweat measurement can be identified. Below, the most important challenges are reported, based on the literature and our own experiments.3,7
(1) Sweat rates vary over time and influence the sampled ion concentrations.
(2) Due to the low sweat rates per gland (nl min−1 mm−2), sample volumes are small.
(3) Skin and sweat gland metabolism influence the concentration levels in sweat.
(4) Contamination of new sweat with old sweat or by skin contaminants can occur.
(5) Sample collection is difficult due to evaporation from the highly distributed sweat glands, and the irregular skin surface.
(6) Several sweat analytes, such as glucose and proteins, are present in very low concentrations.
Eccrine sweat mainly consists of water and NaCl. Na+ and Cl− are secreted in the secretory coil of the sweat gland and a part is reabsorbed in the duct. When the sweat rate becomes higher, less ions are reabsorbed in the duct and the concentrations of Na+ and Cl− are higher.2,10 Since the concentration varies with the sweat rate, measuring Na+ and Cl− levels may be useful in determining the sweat rate of an athlete. Several researchers state that Na+ and Cl− increase when an athlete dehydrates and lower levels of Na+ and Cl− can indicate electrolyte loss.1,11 However, papers that present Na+ and Cl− measurements together with changes in the hydration status, show mixed results.12 One of the main reasons for this is the lack of standardized collection and analysis methods.
For K+, part of the secretion mechanism is known. However, contradictory results about K+ variations in sweat and the relation of K+ levels to the sweat rate are found in the literature.13,14 It is also expected that K+ levels are often overestimated due to skin contamination.15 More research is needed to find out if this parameter turns out to be useful in monitoring the physical status of an athlete.
Although the literature shows contradictory results for NH4+ measurements as well, some researchers state that NH4+ levels are related to lactate levels in the blood, which makes this analyte potentially interesting for tracking muscle fatigue.16–19
Secondly, researchers use microfluidics systems for sweat sensing purposes. Ma et al. designed a patch that contains a wick and a microfluidic channel that collects the sweat and brings the sweat to the electrode of the sensor.28 Koh et al. created microfluidic channels from PDMS that enabled colorimetric measurement of sweat and Choi et al. created a microfluidic network with small bursting valves for chrono-sampling of sweat.29,30 Sample volumes of the collected sweat in these systems are small (e.g. 6 μl). Aranyosi et al. created a fluidic system for single but larger samples.31
Physiologists lack suitable collection and analysis systems for continuous monitoring to find useful sweat biomarkers and how they change over time. In the technological literature, validation of novel sweat sensor systems in human trials appears difficult, because there are no standardized methods to perform reference measurements by chrono sampling sweat and analysing it in the lab afterwards. To solve these problems, we propose a new sweat collection device that can automatically collect a sequence of sweat samples (>100 μl). The system is easier to fabricate and the collection surface is larger than in previously presented solutions, which facilitates the use of the system by both engineers and physiologists and increases reliability of the measurements. Physiological experiments and chromatography measurements are executed to prove that this method can be used to analyse electrolyte variations over time.
The sweat collector device can in principle be connected to an electrochemical sensing system for continuous measurements. The testing of our new sweat collectors with an in situ analysis system will be discussed in a subsequent paper.
![]() | ||
Fig. 1 Design of the sweat patch: (a) left: exploded view, right: front view of the patch. (b) The filling process of the reservoirs. |
Due to capillary forces, the sweat flows from the inlet (Fig. 1, no. 7.) to the reservoirs (Fig. 1, no. 10.). The walls of the capillaries are highly hydrophilic to ensure that the flow rate of sweat is not negatively influenced by the channels. The sweat passes a T-junction (Fig. 1, no. 8) at a certain moment. At this junction the sweat needs to flow down into the reservoir. By creating sharp edges at the junction and by directing the left branch of the junction upwards, it is ensured that the sweat goes straight into the reservoir. When the reservoir is filled, the sweat will take the left turn and will flow to the next reservoir by capillary forces. Entrapment of air is prevented by placing air outlets (Fig. 1, no. 9) at the right corner of each reservoir.
The materials were cut into the desired shapes with a CO2 laser system (Merlin Lasers, Lion Laser Systems, The Netherlands). The film is rinsed with demineralized water (to remove the extra ionic surfactant that was applied to the film by the manufacturer, because it influenced the sodium measurements in preliminary experiments). Samples of the foil and adhesive were placed in vials with demineralized water (3 ml, 24 hours), and ion concentrations (Na+, Cl− and K+) in these vials were analysed to ensure the absence of background contaminants from the materials. No significant Na+, Cl− and K+ peaks were detected in these samples. Contact angles of the materials were measured with optical tensiometry (KSV Instruments Ltd). The spacer tape has an average contact angle of 50.5° and the hydrophilic foil has an average contact angle of 92.2° (Fig. S1, ESI†).
![]() | (1) |
Several designs were made and CFD simulations with COMSOL Multiphysics software were performed to test the functioning of the devices. For the simulations it is assumed that water can flow freely into the channel (ideal case). The transport of the fluid interface is given by a level set function.36
This function is coupled with the Navier–Stokes equations to describe mass and momentum transport of the fluid. The velocity of the fluid is dependent on convection, pressure, diffusion, the surface tension and gravity in this model.
Initially, a small reservoir at the beginning of the microchannels is filled with water and the rest of the channels and the big reservoir are filled with air. The initial velocity is 0. A hydrostatic pressure is applied at the inlet of the channels (P = ρ × g × y) and the pressure at the outlet is 0. Atmospheric pressure can be omitted since it is acting on both the inlet and the outlet. The gravity is added to the model as a volume force. The wetted wall feature is used to identify hydrophobic and hydrophilic walls.
First, the influences of the spacer tape and corner flow effects were simulated with a straight channel. Second, parts of the collection patch were simulated to test if the reservoirs are filled in the right way and to prevent entrapment of air.
The syringe pump (KDScientific 200, USA) is set at a rate of 48 μl min−1. Via the funnel-shaped structure, the sweat will drop down towards the inlet of the reservoirs.
The collector is designed to be placed on the back of an athlete. The back is chosen, because of the high sweat rate and the presence of eccrine sweat glands at this location. Smith and Havenith measured a sweat rate at the back of 1.2 mg cm−2 min−1 during a running exercise at around 155 bpm of 30 min (25.6 ± 0.4 °C, 43.4 ± 7.6% relative humidity).37 The patch will have a collection surface of 40 cm2, which means that the sweat rate will be around 48 μl min−1.
Collector filling is recorded with a camera. After the experiment, stills are taken from the movie. Due to the blue colorant that was added to the fluid, all blue pixels in the image can be counted to derive the volume that was filled with fluid. This process was automated with a MATLAB program.
![]() | ||
Fig. 2 (a) Setup of the physiological experiment (reproduced from ref. 38). (b) Placement of the patches in the final experiment. The patches A (location 1, 2 and 3) are placed at the start. Patches B (1, 2 and 3) are placed when patches A start filling. Patches of C (1–3) are placed when A is filled. |
In the first two series of experiments the collectors were replaced at the same location, when a collector was filled completely. Unfortunately, a replaced collector does not start filling immediately (it can take approximately 10 minutes). To improve the continuity of the sweat measurements, the patches were placed as shown in Fig. 2b in the final experiment. Patch B can now be placed 10 minutes before patch A is filled and more sweat can be collected. It is assumed that the sweat rates at locations left and right of the spine are identical.
To analyse the cations, a Metrosep C6 – 150/4.0 column is used for separation with a solution of 3 mM HNO3 as the eluent. The flow rate is set at 0.9 ml min−1. For the anions, a Metrosep A supp 5 – 150/4.0 column is used, and chemical suppression is performed by using the Metrohm Suppressor Module, which is regenerated with 150 mM H3PO4. The eluent of the anion system contains 1 mM NaHCO3 and 3.2 mM Na2CO3. The flow rate in this system is 0.7 ml min−1. A sample loop of 20 μl is used in both systems.
Before each analysis, standards need to be made for all the ions of interest. Solutions with 0.1 till 100 ppm are made. The measurements of the standards are used for calibration. After assigning manually the right retention times to the peaks, the MagIC Net software, automatically creates a calibration graph based on the peak areas which can be applied to the measurements.
To quantify the accuracy of our method, standard deviation experiments are executed. Standard solutions with 0.1 to 100 ppm Na+, NH4+ and K+ ions and separate solutions with Cl− ions are prepared. The solutions are divided into 5 samples and measured one by one to obtain an idea of the standard deviation of the method at different concentration levels.
To study the effect of corner flow, channels with rounded corners and the same dimensions as the previous channel are simulated as well. One channel has hydrophobic rounded sidewalls and the other has hydrophilic rounded sidewalls. Fig. 3a shows the influence of the rounded and sharp, hydrophilic, and hydrophobic sidewalls on the contact point position of the fluid–air interface over time. The volumetric flow rate decreases significantly when the hydrophobic sidewalls have rounded corners. Since the laser cut spacer tape is placed on top of another layer of foil, the corners will be relatively sharp.
Simulations are also performed to test whether the T-junction works suitably. In Fig. 3b it can be seen that the fluid first flows down and once the water reaches the bottom (t = 5 ms), it goes left to the next reservoir.
Lastly, a simulation was performed to check if entrapment of air would be a problem in the reservoirs. Fig. 3c shows that the fluid does not block the air outlet during the filling process, so that air bubbles can be released. The filling time of a reservoir with a height of 10 mm, is now very small. This is due to the infinite supply of water at the inlet. The sweat rate will limit the fluid supply in reality. However, the simulation shows that even with an infinite supply of water, the reservoir is filled in the right way. This accounts for the T-junction as well.
To get an idea of the standard deviation that is introduced by the IC measurement method, standard deviations of the method are included. These are determined by analysing five samples of each standard (0.1–100 ppm) and plotting the average relative standard deviations (RSDs) in a graph (Fig. 4a). The fitting curve is used to determine the SDs in the actual measurements. For Cl− and Na+ the SDs are relatively small, for K+, the concentrations in sweat are around ten times lower so the RSDs of the measurement method become important to consider. The measured concentration levels of Na+, Cl− and K+ of patch location 2 are shown in Fig. 4b. The Na+ and Cl− concentrations show an initial steady increase which levels off. However, Cl− levels are systematically lower by an average of 16.4 mM (SD = 2.5 mM) than the Na+ levels. This deviation can possibly be explained by the presence of other negative ions like lactate and bicarbonate (C3H5O3− and HCO3−) in sweat.
In Fig. 4c, the Na+, Cl− and K+ levels of patch location 1 (between the shoulder blades) and 3 (lower back) are added. The Na+ and Cl− levels are very similar for collector 2 and 3 (Cl−: M (mean) = 1.6, SD= 2.9, Na+: M = 0.4, SD = 3.4 mM) while the Na+ and Cl− levels of collector location 1 are significantly higher (Cl−: M = 10.3, SD = 2.4, Na+: M = 10.4, SD = 2.9 mM). One of the reasons for this difference between the absolute concentrations of collector 1 and collector 2 and 3 is the variation in the sweat rate across the different locations. Smith and Havenith measured at the sides of the upper back a median sweat rate of 0.84 mg cm−2 min−1, while at the sides of the lower back this sweat rate was 0.75 mg cm−2 min−1.37 Higher sweat rates decrease ion reabsorption per unit volume and therefore higher concentrations can be measured at the upper back.
Fig. 4d shows the K+ levels at the three locations over time. A decrease is measured over time and no difference between the three locations can be detected within the error margin. To test whether the trend is still a decreasing line, when taking the large SDs into account, random samples and their linear fits (n = 1000) were created, assuming a normal distribution.
Within the 95% confidence interval, the fits for location 1 and 2 were always a decreasing line. However, from the literature it was assumed that only minimal changes in K+ levels occur in sweat.9 Therefore, it is expected that elevated values after the start of the exercise are due to the presence of old sweat or residue in the channels or other physiological effects that need further research.
Because an increased sweat rate leads to less ion reabsorption in the reabsorptive duct, sweat rate measurements are important as well. The sweat collection rate is measured at location 2 during the experiments with the help of a camera. The average sweat rate is 1.19 mg cm−2 min−1. The sweat collection rate varied between 0.74 and 1.54 mg cm−2 min−1 (Fig. 5a).
![]() | ||
Fig. 5 (a) Sweat rate and Cl− variation over time (b) Relation between the sweat rate and Na+ and Cl− concentration. |
The sweat rate was plotted against the concentration of Na+ and Cl−, to check whether this relation was visible. The data allows us in principle to verify a possible relation between the sweat rate and Na+ and Cl− concentrations. Although the plot of Fig. 5b suggests a mild increase in ion concentrations with increasing sweat rate, the scatter in the current data is too large to draw a solid conclusion. In a next phase of this project, other sweat rate measurement techniques will be explored by, for example, conductive or capacitive measurements. It would also be interesting to compare this new method of sweat rate measurement with conventional methods like the ventilated capsule measurement and the absorbent patch method in the same physiological experiment.40
In the current study, data from only one physiological experiment of a single individual are compared, because it is well known that sweat analyte concentrations can differ considerably between individuals and that the day-to-day variability in sweat analyte concentrations can be large. Therefore, comparing, and interpreting data between individuals in small scale experiments would be less relevant.
However, the presently developed device will facilitate more systematic larger studies on the inter-individual variability as well as the daily variations of sweat electrolytes in a single individual. In this way, the new device can contribute to a better understanding of sweat mechanisms and their relation to blood values, to find useful biomarkers in sweat to monitor the status of an athlete.