A capacitive DNA sensor for sensitive detection of Escherichia coli O157:H7 in potable water based on the z3276 genetic marker: fabrication and analytical performance

Abstract

We report a label-free biosensor for the detection of Escherichia coli O157:H7 ATCC 43895 in potable water using a newly designed DNA sensing probe targeting the z3276 genetic marker. The surface of indium tin oxide (ITO) was functionalized with the novel sensing probe by covalent coupling using APTES as a crosslinker to fabricate the DNA sensor (dubbed ZEC [[z with combining low line]3276 gene of [E with combining low line]. [c with combining low line]oli O157:H7 ATCC 43895]). The electrochemical characterization of the fabricated ZEC sensor was performed using cyclic voltammetry and electrochemical impedance spectroscopy. Atomic force microscopy and scanning electron microscopy revealed significant changes in the surface topographies of the fabricated ZEC sensor chip. Equivalent circuit analyses suggested the capacitive nature of the ZEC sensor chip, which demonstrated a declining trend of the capacitance value from 1.568 μF (Bare ITO) to 1.221 μF (after DNA hybridization). Non-faradaic sensing measurements revealed systematically declining capacitance values upon DNA hybridization, with a 10 min response time at 10 Hz frequency and 10 mV applied potential. The ZEC sensor chip exhibited linearity in the range of 0.5 to 25 pg per 10 mL for E. coli O157:H7, with ubiquitous cross-validation of each DNA concentration using quantitative PCR prior to the analyses of real water samples. The limit of detection (LOD) at 95% confidence estimated by logistic regression was 0.1 pg DNA per 10 mL of E. coli O157:H7 (equivalent to 13.67 CFU per 10 mL) with a p-value of 0.0237. Consequently, the obtained results demonstrate the possible application of the developed ZEC sensor chip for E. coli O157:H7 detection in real water samples.

Graphical abstract: A capacitive DNA sensor for sensitive detection of Escherichia coli O157:H7 in potable water based on the z3276 genetic marker: fabrication and analytical performance

Supplementary files

Article information

Article type
Paper
Submitted
14 Nov 2019
Accepted
06 Jan 2020
First published
30 Jan 2020

Analyst, 2020,145, 2267-2278

A capacitive DNA sensor for sensitive detection of Escherichia coli O157:H7 in potable water based on the z3276 genetic marker: fabrication and analytical performance

R. Deshmukh, A. K. Prusty, U. Roy and S. Bhand, Analyst, 2020, 145, 2267 DOI: 10.1039/C9AN02291K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements