Correction for ‘A UV digital micromirror spectrometer for dispersive AFS: spectral interference in simultaneous determination of Se and Pb’ by Chen Tao et al., J. Anal. At. Spectrom., 2018, 33, 2098–2106.
Parameters | Setting value (by UV digital micromirror spectrometer method) | Setting value (by non-dispersive method) |
---|---|---|
Flow rate of carrier gas | 400 mL min−1 | 400 mL min−1 |
Flow rate of shield gas | 850 mL min−1 | 850 mL min−1 |
KBH4 concentration (w/v) | 1.6% | 1.6% |
HCl concentration (v/v) | 3% | 3% |
K3Fe(CN)6 concentration (w/v) | 2.5% | 2.5% |
Sample flow rate | 0.1 mL s−1 | 0.1 mL s−1 |
Reductant flow rate | 0.05 mL s−1 | 0.05 mL s−1 |
Lamp primary/boost current of Se | 60 mA/30 mA | 60 mA/30 mA |
Lamp primary/boost current of Pb | 60 mA/30 mA | 60 mA/30 mA |
Negative high voltage of PMT | 450 V | 280 V |
The sentence beginning “The K3Fe(CN)6 (Sinopharm Group Chemical Reagent Co., Ltd.)” and the following sentence on page 2100 should read as follows: “The K3Fe(CN)6 (Sinopharm Group Chemical Reagent Co., Ltd.) and KBH4 (Tianjin Institute of Chemical Reagents, Tianjin, China) solution were prepared by dissolving KBH4 in a 0.5% (w/v) NaOH solution (Beijing Chemicals Co., China). Argon (99.99%) (Praxair Inc., Beijing, China) was used as the carrier gas and the shield gas.”
The sentence beginning “As can be seen in Fig. 12” and the following sentence on page 2103 should read as follows: “As can be seen in Fig. 12, the signal increased with increasing HCl concentration from 2% to 7% (v/v) for Se. Higher concentration was unfavourable for Pb excitation. Using 3% (v/v) HCl is suitable for simultaneous determination of Se and Pb, and the reaction system is stable.”
The sentence beginning “Because the grating is designed to divide” on page 2100 should read as follows: “Because the grating is designed to divide the wavelengths nearly equally and project them onto the DMD, the relationship between the lines of the DMD and wavelength satisfies the approximate formula λ = −3 × 10−5C2 + 0.195C + 174.25, where λ is the wavelength and C is the number of lines of the DMD.”
The relative fluorescence intensity of Se at 196.0 nm and 203.9 nm in Table 3 has been written incorrectly in the original article. The corrected version of Table 3 is shown below.
Element | Wavelength (nm) | Relative fluorescence intensity (%) | Linear range (μg L−1) | Relative coefficient | Detection limit (μg L−1) |
---|---|---|---|---|---|
Se | 196.0 | 42.83 | 4–200 | 0.9984 | 2.05 |
203.9 | 100 | 2–200 | 0.9963 | 0.89 | |
Pb | 217.0 | 29.35 | 4–200 | 0.9978 | 1.46 |
261.3 | 52.84 | 2–200 | 0.9987 | 0.94 | |
283.3 | 100 | 2–200 | 0.9991 | 0.42 |
The Royal Society of Chemistry apologises for these errors and any consequent inconvenience to authors and readers.