RBC deformability measurement based on variations of pressure in multiple micropillar channels during blood delivery using a disposable air-compressed pump

Abstract

Subpopulations of red blood cells (RBCs) are attributed to symptoms associated with hematological disorders. To detect differences in subpopulations of RBCs, RBC deformability has been considered as a promising label-free biomarker. To realize high throughput and a label-free approach, previous methods have required bulky syringe pumps for blood delivery and time-resolved micro-PIV techniques to obtain blood velocity fields. In this study, blood (∼0.2 mL) is delivered using a disposable air-compressed pump. Subsequently, RBC deformability is quantified by analyzing the cell-to-liquid interface in the pressure channel (β) caused by successive clogging of RBCs in multiple micropillar channels. By averaging temporal variations in the intensity of captured images of the blood within the pressure channel for a specific duration, a normalized deformability index (DIβ) is defined to quantify RBC deformability. Through the operation of the disposable air-compressed pump, blood velocity is gradually reduced over time. As observed, RBC clogging in the micropillar channel led to an increase in the value of β accompanied by fluctuations with the passage of time. A quantitative comparison of the proposed method (DIβ) against the previous method (ΔV) is performed using blood samples composed of homogeneous and heterogeneous hardened RBCs. Upon performing linear regression analysis between DIβ and ΔV, the coefficient of regression provides values as high as R2 = 0.96–0.99 (i.e., P < 0.001). Thus, DIβ could be effectively used to detect variations in the deformability of RBCs. Finally, the variation in RBC deformability is quantified over a storage period of 1 week. As a result, the values of these properties remained constant for a storage time of 1 week. The indices exhibited a significant relationship because regression analysis yielded a high value of R2 = 0.7916 (i.e., P < 0.001). In view of these experimental demonstrations, the proposed method of the normalized deformability index can be used to quantify RBC deformability, especially using a disposable air-compressed pump.

Graphical abstract: RBC deformability measurement based on variations of pressure in multiple micropillar channels during blood delivery using a disposable air-compressed pump

Article information

Article type
Paper
Submitted
05 Jul 2018
Accepted
27 Aug 2018
First published
28 Aug 2018

Anal. Methods, 2018,10, 4549-4561

RBC deformability measurement based on variations of pressure in multiple micropillar channels during blood delivery using a disposable air-compressed pump

Y. J. Kang, Anal. Methods, 2018, 10, 4549 DOI: 10.1039/C8AY01486H

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements