Fabrication of paper micro-devices with wax jetting†
Received
9th December 2015
, Accepted 2nd February 2016
First published on 3rd February 2016
Abstract
Paper microfluidic devices are a promising technology in developing analytical devices for point-of-care diagnosis in the developing world. This article describes a novel method of wax jetting with a PZT (piezoelectric ceramic transducer) actuator and glass nozzle for the fabrication of paper microfluidic devices. The hydrophobic fluid pattern was formed by the permeation of filter paper with wax droplets. Results showed that the size of the wax droplet, which was determined by the voltage of the driving signal and nozzle diameter, ranged from 150 μm to 380 μm, and the coefficient of variation of the droplet diameter was under 4.0%. The smallest width of the fluid channel was 600 μm frontside and 750 μm backside. The patterned filter paper was without any leakage, and multi-assay of glucose, protein, and pH on the paper microfluidic device, and laminar diffusion flow with blue and yellow dye were realized. The wax jetting system supplied a low-cost, simple, easy-to-use and fast fabrication method for paper microfluidic devices.
1. Introduction
The paper microfluidics device is a very important kind of platform for point-of-care diagnostics and environment monitoring. It uses small volumes of sample and the paper is made of naturally abundant materials and is biodegradable. Many researchers have put lots of effort into paper-based microfluidics as it has advantages of being easy to use, inexpensive, low volume, easily adaptable, and is capable of rapid on-site detection.1–3
To realize low cost and simple paper microfluidic devices, many kinds of fabrication methods have been developed, such as photolithography,4–6 plotting,7 inkjet etching,8–12 plasma etching,13 flexographic printing,14–16 cutting,17–21 stamp22,23 and wax printing.24,25 In conditions of micro reaction or analysis, the hydrophobic patterns of on the paper need to be changed frequently or to be very complicate, and this requires that the fabrication methods are low cost and fast. The photolithography,4–6 plasma etching,13 flexographic printing,14–16 water resist ink plotting7 and stamp or punch method22,23 need metal or other kinds of templates. The processing of templates requires multi steps and not easily adaptable. Paper devices by knife, punch or laser cutting17–21 are not easy to be folded and stored. The wax printing method was very promising as the wax material is easily available and biodegradable. The commercial wax 3D (three-dimensional) printer and its suited wax were expensive. The need still remains for gentle and simple fabrication techniques of paper microfluidic devices that can be widely implemented in developing world at minimized cost.
In this paper, the paper microfluidic device was fabricated with a novel method of wax jetting which was based on PZT actuator and a home-made glass nozzle. The jetted wax droplets were linked with each other and melted into the paper to form into hydrophobic barriers. Firstly, the relationship between the size of the wax droplets and the system parameter was determined. The wax droplets were linked with each other on the glass slide using 3D printing table. Then, the relation between the wax barrier and printing width and the resolution of the wax jetting method on filter were experimentally studied. Finally the paper device were fabricated and tested. Multi bio-assays of glucose, protein, and pH, and stable micro diffusion of blue dye and yellow dye were presented.
2. Materials and method
2.1. Materials
Qualitative filter paper 102 Ø 7 cm and Ø 10 cm were from Hangzhou Xinhua Paper Industry Co., Ltd. Glass nozzle 1.0 mm × 0.6 mm was from Nanjing Lupu Chemical Co., Ltd. Paraffin wax 56–58 was supplied by Shanghai Specimen and Model Factory. Glucose oxidase (Amresco 0243), horseradish peroxidase (JS10540, Cas 9003-99-0, RZ ≥ 1.5), bovine serum albumin (BSA), tetrabromophenol blue (TBPB) (MW 691.9), glucose, erioglaucine disodium salt (BR, 85%, 3844-45-9) and tartrazine BS (85%, 1934-21-0) were purchased from Shanghai Jinsui Bio-tec Co., Ltd. Potassium iodide, ethanol AR citrate and trisodium citrate dehydrate were from Nanjing Chemical Reagent Co., Ltd.
2.2. Wax droplet generating system
A wax droplet generator on a 3D printing work table was used to prepare the paper microfluidic devices. The wax droplet generator consist of a piezoelectric actuator (Polytec PI P-844.10), a glass nozzle, a wax container and a connector as shown in the Fig. 1. The photograph of the droplet generating system was shown in the Fig. S1.† The glass nozzle was made out of a borosilicate glass tube with inside diameter of 0.6 mm and outside of 1.0 mm. The glass tube was first pulled with a micropipette puller (Sutter P2000/G) and then cut with a ceramic chip on the tip leaving an outlet diameter of 150 μm. After that, the glass nozzle was forged with a micronozzle processer of Narishige MF-900 leaving the final outlet diameter ranged from 50 μm to 100 μm as shown in the Fig. S2.† The glass nozzle was bonded to an aluminum connector which was screwed on a wax container made of aluminum. The temperature was controlled by a heating rod and a thermocouple probe.
 |
| Fig. 1 Schematic of the wax droplet generating system. | |
The drop-on-demand wax droplet generating was proposed on the inertia force of the fluid body as shown in the Fig. 2. The signal received by the PZT actuator was shown in the Fig. 1. The stack PZT was charged gently and discharged sharply. Because of the piezoelectric effect, the glass nozzle gets a speed ν in the constricting direction by the force F as shown in the Fig. 2. There exists a thin fluid boundary layer near the glass wall and the fluid outside this layer gets a relative speed ν1 cause by inertia. The jetting happened under the action of viscous force F1 and inertia force F2.
 |
| Fig. 2 Principle of microfluidic driving. | |
2.3. Jetting of wax droplet
The cylinder wax container with length of 45 mm and diameter of 7 mm was filled with paraffin wax and covered to avoid dust in the air. The wax was melted and injected to the container with a syringe and a syringe filter (filter mesh: 0.45 μm). The jetting temperature is 95 °C. This is an improvement compared with the glass nozzle26 as there is no need to control the maximum length of the wax liquid in the nozzle to avoid extruding and the fabrication process was easy to control. By the peak voltage of the driving signal ranged from 30 V to 80 V, the wax was jetted on the glass slide with 50 μm, 75 μm and 100 μm glass nozzle.
2.4. Bio-assay using paper microfluidic devices
The paper microfluidic devices were tested with 2000 mM erioglaucine (blue dye) and 1870 mM tartrazine (yellow dye) solution. 30 μL of blue dye and 30 μL of yellow dye were slowly dipped in the feeding area of the paper devices respectively. The BSA assay were realized by dipping 3 μL of a 250 mM citrate buffer solution (pH 1.8) in the reaction zone and followed by 3 μL of 3.3 mM TBPB solution in 95% ethanol. The glucose assay were realized by dipping 3 μL of potassium iodide solution and followed by 3 μL of 1
:
5 horseradish peroxidase/glucose oxidase solution (100 units per mL). For the pH detection, 3 μL of pH reagent were dipped in the reaction zone.
Devices with eight circular reaction zones were used for different concentration solution assay. For BSA assay, 3 μL of BSA solution of 0, 1, 4, 10, 25, 66, 90, 130 mg mL−1 were dipped in A–H zones separately. Devices with four circular reaction zones was used for multi-assay, 30 μL of a mixture of 50 mM of glucose solution and 7.5 mM of BSA solution with 1
:
1 ratio was dipped in the middle of the cross area.
3. Result and discussion
3.1. Preparation of wax droplets
The SEM (Scanning Electron Microscope) photo of 5 × 7 wax droplet array jetted by a micro nozzle with 75 μm outlet diameter, 45 V peak voltage and 3 Hz signal was shown in the Fig. 3. The diameter of droplet outline circle was 218 μm, the coefficient of variation (CV) of the diameter of the jetted wax droplets was under 4.0% which showed a well uniformity. In a range of 1–15 Hz, the signal frequency had little influence on the wax droplet size. The relationship between the wax droplet diameter and the peak voltage of the driving signal and outlet diameter of the glass nozzle was shown in the Fig. 4.
 |
| Fig. 3 SEM photo of wax droplet array formed on the glass slide. | |
 |
| Fig. 4 Relationship between wax droplet diameter and the peak voltage of the driving signal and outlet diameter of the glass nozzle. | |
3.2. Preparation of wax line on glass slide
The barrier “wall” in the filter paper was formed by the wax material melted into the paper, and the wax line was prepared by linking of wax droplets with each other. We defined the degree of overlapping (k) to represent the mount of wax material per unit length as shown in the Fig. 5 and k could be written as (1), where L was the length of the overlap area (μm); D was the outline circle diameter of a wax droplet (μm); v was the moving speed of the 3D printing worktable (μm s−1); f was the jetting frequency. The choosing of k was determined by the thickness of the filter paper and the size of the wax droplet. When the overlapping was less than 0, none of the wax droplets linked with each other, and k was more than 80%, wax droplets grew from the glass slide into an inclined pillar.
 |
| Fig. 5 Schematic diagram of the wax droplets. | |
By adjusting the moving speed of the 3D printing worktable (v), k could be changed. Thus, v could be written as v = D(f − 1)(1 − k). For our system, v was less than 50 mm s−1 and the resolution was 1 μm s−1. With the micro nozzle of 75 μm outlet diameter, 40 V peak voltage and 11 Hz signal, wax lines with k under 0 and k = 10%, 30%, 50% and 70% were prepared respectively. Results were shown in the Fig. 6. When the wax droplets did not link, they were round with diameter of 200 μm. Once k was more than 0, a wax tail formed in the moving direction and grew into the neighbor wax droplet and they link with each other to form into line shape. The top surface of wax line with k = 50% was smoother than that of k = 10%, and the number of the wax droplets could be read from the bottom surface relief. When k = 70%, the width of the wax line was nonuniform and the wax wall grew higher than that of k = 50%.
|
 | (1) |
 |
| Fig. 6 SEM photo of wax line on glass slide (a) wax droplets unlinked with each other (b) wax line prepared with k = 10%, (c) k = 30%, (d) k = 50% (e) k = 70%. | |
3.3. Relation between the wax barrier and printing width
Traditional paper diagnostic that used for the glucose, protein or urine is strip like which could only test one kind of element in the agent. Patterned paper combining multi channel and reaction zones enabled multi assay with capillary flow. Our previous work has proved that k = 50% and kept in 75 °C thermostatic well for 80 seconds is fine for the penetration of the filter paper with thickness of 90 μm.26
The relation between the final front wax barrier width and printing width was studied. Outlet diameters of glass nozzles were 50 μm, 75 μm and 100 μm, and degree of overlapping (k) was 50%. To make the wax to penetrate the filter paper sufficiently and save time, the filter paper was kept on a heating plate of 120 °C for 10 seconds. The printing width was determined by the diameter of wax droplets. The wax barriers with width of 950 μm, 1050 μm and 1200 μm are shown in the Fig. 7. As the transparency of filter paper could be improved by the penetration of wax, the color depth increased with the barrier width. The relation between the wax barrier and printing width is shown in the Fig. 8, and the wax barrier width ranged from 770 μm to 1200 μm.
 |
| Fig. 7 Wax barriers on filter paper with thickness of 90 μm, (a) wax barriers with width of 950 μm, (b) wax barriers with width of 950 μm, (c) wax barriers with width of 1200 μm. | |
 |
| Fig. 8 Relation between the wax barrier and printing width. | |
3.4. Resolution of wax jetting on filter paper
The resolution of the wax jetting method on filter paper was presented by the final width of fluid channels. With the barrier width of 770 μm, leakage may happened along the fluid channel. The jetting conditions are as follows: outlet diameter of the micro nozzle is 75 μm forged from 150 μm, the peak voltage of the driving signal is 40 V, the frequency is 11 Hz and the moving speed of the worktable is 1 mm s−1 (k = 50%). The fluid pattern is made up of 8 reservoirs with 10 mm channel sharing the same feeding pool. The designed channel width on AutoCAD increases from 800 μm to 2200 μm in 200 μm steps. Result showed that the 800 μm and 1000 μm fluid channels designed were totally blocked and the yellow dye flow into part of the 1200 μm fluid channel. The reservoirs with designed channel width of 1400 μm and more were totally wetted by capillary flow. The front and back side of the paper device are shown in the Fig. 9 and S3.†
 |
| Fig. 9 The paper microfluidic devices with designed fluid channel widths from a-800 μm to h-2200 μm in 200 μm steps on AutoCAD, (a) front side of the paper device filled with yellow dye solution, (b) back side of the paper device. | |
The smallest channel width allowing solution to wet the entire reservoir through 10 mm channel was found to be 600 μm in front side and 750 μm in back as shown in the Fig. 10. The resolution of wax jetting method is limited by the thickness, porosity, and orientation of paper fibers as well as the smallest wax droplet size jetted on the paper. Moreover, we studied the width consistency along 10 mm channel from 1400 μm to 2200 μm in 200 μm steps. The channel width was measured 15 times with step of 600 μm along the back side, and 10 times with step of 900 μm for the frontside channel. Result showed that the relative standard deviation was less than 11.0% for all front and back channel widths as shown in the Tables S1 and S2.†
 |
| Fig. 10 The channel width of front side and backside formed by wax barriers with center-to-center distance from 1400 μm to 2200 μm in 200 μm steps. | |
3.5. Test of paper device used for multi-assay
The paraffin wax was jetted on the filter paper to form the pattern designed for bio-assay. Wax patterns for multi-assay with eight circular reaction zones, eight square reaction zones and five polygon zones were prepared respectively. With capillarity of filter paper, the solution reached the reaction zones for multi-assay in 3 minutes without any leakage. The wax patterns dried after 15 minutes are shown in the Fig. 11.
 |
| Fig. 11 (a and b) Paper microfluidic devices with eight square reaction zones for multi-assay filled with yellow dye and blue dye (c and d) devices with eight circular reaction zones filled with yellow dye and blue dye (e and f) devices with five polygon reaction zones filled with yellow dye and blue dye. | |
3.6. Multi-assay using paper micro devices
Results of the multi-assay were shown in the Fig. 12. After 10 minutes of dipping the BSA solution to zones A to H respectively, the color of reaction zones from A to H changed from aquamarine blue to cobalt blue as the concentration increased as shown in the Fig. 12(a). For the multi-assay, after 5 minutes of dipping the sample solution, zone A appeared a yellow coffee ring showing the presentation of glucose and inlet area of zone B turned blue as the existence of BSA. Zone C turned light yellow in the whole area indicating that the pH was 7.0 as shown in the Fig. 12(b).
 |
| Fig. 12 (a) Protein assay, 10 minutes after respectively dipping 3 μL of 0, 1, 4, 10, 25, 66, 90, 130 mg mL−1 of BSA solution on zone A–H, (b) multi-assay for 30 μL mixture of 25 mM glucose and 3.75 mM of BSA solution, 5 minutes after dipping of the test agent. | |
3.7. Laminar flow diffusion using paper micro devices
Traditional transverse diffusion across adjacent flow stream is usually realized with Y shape micro channels with two or more inlet to study inter diffusion of one or multiple species, to evaluate transport processes, and to sense analytes ranging from pH to immunoassay targets.27 The capillary flow in paper was extremely stable which enabled a very stable diffusion interface and thus need not extremely stable pumps. Micro mixers with two-dimensional paper networks were designed with two and three inlet areas. The diffusion area was laid with a source pad made from lens paper to keep the dye solution.
The wax pattern of paper device with two inlet area is shown in the Fig. 9a. Assuming the inlet area is an infinite source and the absorbent area is an infinite sink, the flow in a fully wetted paper device is approximated by Darcy's law:
|
 | (2) |
where
Q is volumetric flow rate,
κ is the permeability of the paper,
wh is the area of the channel cross-section with width,
w, and height,
h,
μ is the dynamic viscosity, and Δ
P is the pressure drop occurring over the length,
L, of the paper channel.
For the designed filter paper microfluidic device in the Fig. 9b, k and μ are constant. For the inlet channels, w1 = w2 and L1 = L2. When the same amount of sample solutions are dipped in the left and right source area, ΔP1 = ΔP2. Deduced from Q1 = Q3 and Q2 = Q4, we could get w3 = w4. The diffusion coefficient γ was defined by ratio of width of the diffusion area and the flow channel in the outlet area. γ is written as:
|
 | (3) |
For the device with two inlets, 100 μL of yellow dye was dipped in the inlet zone A, and 100 μL of blue dye into the inlet zone B in the same time. The laminar flow of blue branch and yellow branch had a clear boundary in the beginning. The blue and yellow dye diffused into each other along with the fluid channel and the interface turned green as shown in the Fig. 13. Calculated by device dipped of solutions after 10 minutes, the diffusion coefficient γ was 46.9%.
 |
| Fig. 13 Micro diffusion of blue and yellow dye in the device with two inlet areas. (a) Schematic of the paper device, the diffusion coefficient γ was defined by ratio of width of the diffusion area and the flow channel in the outlet area. (b) Before dipping (c) after 40 seconds, (d) after 2 minutes, (e) after 10 minutes. | |
For the device three inlets, 100 μL of yellow dye was dipped in the inlet zone A and C, and 100 μL of blue dye into the inlet zone B in the same time. The yellow and blue dye met in the reaction channel and divided into three branches with yellow, blue and green colors. The interface between the yellow and blue branches turned green by diffusion, and the area of green color increased along with the capillary flow as shown in the Fig. 14. Calculated by device dipped of solutions after 10 minutes, the diffusion coefficient γ as defined before was 57.1%.
 |
| Fig. 14 Micro diffusion of blue and yellow dye in the device with three inlet areas (a) before dipping (b) after 20 seconds, (c) after 2 minutes, (d) after 5 minutes, (e) after 10 minutes. | |
4. Conclusions
The paper microfluidic devices were fabricated with wax jetting by a PZT stack actuator, glass nozzle and 3D printing worktable. The wax jetting was a promising fast and easy method for the fabrication of paper microfluidic devices for the point of care or other experimental study. The wax material (0.0019 $ per g) and the glass capillary tube (0.0078 $ per piece) was cheap and easy available. The paper microfluidic devices need not any templates or multiple printing steps, whereas the pattern is formed directly on the paper and only need to be designed on the computer, which saved the time and cost. The patterned filter paper by the wax droplets was without any leakage, and multi-assay of glucose, protein, and pH on paper microfluidic device, and laminar diffusion flow with blue and yellow dye could be realized. This method also supplied a fast and easy way for the scientific researchers to develop new types of paper microfluidics.
Acknowledgements
This paper is supported by the National Natural Science Foundation of China (no. 61273243, no. 51407095), Program of Natural Science Research of Jiangsu Higher Education Institutions of China (Grant No. 14KJB480004), Program of Natural Science Foundation of Jiangsu Province (BK20150973), the Fundamental Research Funds for the Central Universities.
References
- K. Yamada, T. G. Henares, K. Suzuki and D. Citterio, Paper-Based Inkjet-Printed Microfluidic Analytical Devices, Angew. Chem., Int. Ed., 2015, 54(18), 5294–5310 CrossRef CAS PubMed.
- Y. He, Y. Wu, J. Z. Fu and W. B. Wu, Fabrication of paper-based microfluidic analysis devices: a review, RSC Adv., 2015, 5(95), 78109–78127 RSC.
- D. M. Cate, J. A. Adkins, J. Mettakoonpitak and C. S. Henry, Recent Developments in Paper-Based Microfluidic Devices, Anal. Chem., 2014, 87(1), 19–41 CrossRef PubMed.
- L. Yu and Z. Z. Shi, Microfluidic paper-based analytical devices fabricated by low-cost photolithography and embossing of Parafilm, Lab Chip, 2015, 15(7), 1642–1645 RSC.
- T. San Park, C. Baynes, S. I. Cho and J. Y. Yoon, Paper microfluidics for red wine tasting, RSC Adv., 2014, 4(46), 24356–24362 RSC.
- A. W. Martinez, S. T. Phillips, B. J. Wiley, M. Gupta and G. M. Whitesides, FLASH: a rapid method for prototyping paper-based microfluidic devices, Lab Chip, 2008, 8(12), 2146–2150 RSC.
- J. Nie, Y. Zhang, L. Lin, C. Zhou, S. Li, L. Zhang and J. Li, Low-cost fabrication of paper-based microfluidic devices by one-step plotting, Anal. Chem., 2012, 84(15), 6331–6335 CrossRef CAS PubMed . Z. Nie; F. Deiss; X. Liu.
- O. Akbulut and G. M. Whitesides, Integration of paper-based microfluidic devices with commercial electrochemical readers, Lab Chip, 2010, 10(22), 3163–3169 RSC.
- J. Wang, M. R. N. Monton, X. Zhang, C. D. Filipe, R. Pelton and J. D. Brennan, Hydrophobic sol–gel channel patterning strategies for paper-based microfluidics, Lab Chip, 2014, 14(4), 691–695 RSC.
- M. Elsharkawy, T. M. Schutzius and C. M. Megaridis, Inkjet patterned super hydrophobic paper for open-air surface microfluidic devices, Lab Chip, 2014, 14(6), 1168–1175 RSC.
- K. Yamada, S. Takaki, N. Komuro, K. Suzuki and D. Citterio, An antibody-free microfluidic paper-based analytical device for the determination of tear fluid lactoferrin by fluorescence sensitization of Tb3+, Analyst, 2014, 139(7), 1637–1643 RSC.
- K. Maejima, S. Tomikawa, K. Suzuki and D. Citterio, Inkjet printing: an integrated and green chemical approach to microfluidic paper-based analytical devices, RSC Adv., 2013, 3(24), 9258–9263 RSC.
- X. Li, J. Tian and W. Shen, Quantitative biomarker assay with microfluidic paper-based analytical devices, Anal. Bioanal. Chem., 2010, 396(1), 495–501 CrossRef CAS PubMed.
- A. Määttänen, D. Fors, S. Wang, D. Valtakari, P. Ihalainen and J. Peltonen, Paper-based planar reaction arrays for printed diagnostics, Sens. Actuators, B, 2011, 160(1), 1404–1412 CrossRef.
- Y. Sameenoi, P. N. Nongkai, S. Nouanthavong, C. S. Henry and D. Nacapricha, One-step polymer screen-printing for microfluidic paper-based analytical device (μPAD) fabrication, Analyst, 2014, 139(24), 6580–6588 RSC.
- W. Dungchai, O. Chailapakul and C. S. Henry, A low-cost, simple, and rapid fabrication method for paper-based microfluidics using wax screen-printing, Analyst, 2011, 136(1), 77–82 RSC.
- E. M. Fenton, M. R. Mascarenas, G. P. López and S. S. Sibbett, Multiplex lateral-flow test strips fabricated by two-dimensional shaping, ACS Appl. Mater. Interfaces, 2009, 1(1), 124–129 CAS.
- J. Nie, Y. Liang, Y. Zhang, S. Le, D. Li and S. Zhang, One-step patterning of hollow microstructures in paper by laser cutting to create microfluidic analytical devices, Analyst, 2013, 138(2), 671–676 RSC.
- X. Fang, S. Wei and J. Kong, Paper-based microfluidics with high resolution, cut on a glass fiber membrane for bioassays, Lab Chip, 2014, 14(5), 911–915 RSC.
- X. Mu, L. Zhang, S. Chang, W. Cui and Z. Zheng, Multiplex microfluidic paper-based immunoassay for the diagnosis of hepatitis C virus infection, Anal. Chem., 2014, 86(11), 5338–5344 CrossRef CAS PubMed.
- M. Sun and M. A. Johnson, Measurement of total antioxidant capacity in sub-μL blood samples using craft paper-based analytical devices, RSC Adv., 2015, 5(69), 55633–55639 RSC.
- G. Perozziello, P. Candeloro, F. Gentile, A. Nicastri, A. Perri, M. L. Coluccio and E. Di Fabrizio, Microfluidics & nanotechnology: towards fully integrated analytical devices for the detection of cancer biomarkers, RSC Adv., 2014, 4(98), 55590–55598 RSC.
- M. M. Thuo, R. V. Martinez, W. J. Lan, X. Liu, J. Barber, M. B. Atkinson and G. M. Whitesides, Fabrication of Low-Cost Paper-Based Microfluidic Devices by Embossing or Cut-and-Stack Methods, Chem. Mater., 2014, 26(14), 4230–4237 CrossRef CAS.
- C. Renault, J. Koehne, A. J. Ricco and R. M. Crooks, Three-dimensional wax patterning of paper fluidic devices, Langmuir, 2014, 30(23), 7030–7036 CrossRef CAS PubMed.
- S. Chaiyo, W. Siangproh, A. Apilux and O. Chailapakul, Highly selective and sensitive paper-based colorimetric sensor using thiosulfate catalytic etching of silver nanoplates for trace determination of copper ions, Anal. Chim. Acta, 2015, 866, 75–83 CrossRef CAS PubMed.
- Z. Li, L. Hou, W. Zhang and L. Zhu, Preparation of paper micro-fluidic devices used in bio-assay based on drop-on-demand wax droplet generation, Anal. Methods, 2014, 6(3), 878–885 RSC.
- J. P. Esquivel, F. J. Del Campo, J. G. de la Fuente, S. Rojas and N. Sabate, Microfluidic fuel cells on paper: meeting the power needs of next generation lateral flow devices, Energy Environ. Sci., 2014, 7(5), 1744–1749 CAS.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/c5ra26255k |
|
This journal is © The Royal Society of Chemistry 2016 |
Click here to see how this site uses Cookies. View our privacy policy here.