In this paper, we propose a generalized concept of microfluidic polydimethylsiloxane (PDMS) on paper (POP) devices, which combines well the merits of paper chips and PDMS chips. First, we optimized the conditions for accurate PDMS spatial patterning on paper, based on screen printing and a high temperature enabled superfast curing technique, which enables PDMS patterning to an accuracy of tens of microns in less than ten seconds. This, in turn, makes it available for seamless, reversible and reliable integration of the resulting paper layer with other PDMS channel structures. The integrated POP devices allow for both porous paper and smooth channels to be spatially defined on the devices, greatly extending the flexibility for designers to be able to construct powerful functional structures. To demonstrate the versatility of this design, a prototype POP device for the colorimetric analysis of liver function markers, serum protein, alkaline phosphatase (ALP) and aspartate aminotransferase (AST), was constructed. On this POP device, quantitative sample loading, mixing and multiplex analysis have all been realized.
However, also because of this special porous structure, fabrication techniques for μPADs are quite different from those for traditional microfluidic chips.6 Paper has to be regionally impregnated with materials like wax,7–9 polydimethylsiloxane (PDMS),10,11 polystyrene12,13 and photoresists1,14 or to be treated with reactive gases15,16 to delineate hydrophilic and hydrophobic regions, so that liquid can flow only in the hydrophilic region.17,18 A channel defined in this way is filled with porous or fibrous structures, which usually generate large resistance to liquid flow and lead to inevitable nonspecific surface absorption of samples,19,20 making it difficult to transport macromolecules any length.21 In addition, many valuable functions rely on convective flow like liquid pinning, liquid mixing22 and gradient generation and thus cannot be easily realized. These all have limited the applications of μPADs.
Many efforts have been made to accelerate the flow by changing the flow mode on the paper. Jahanshahi-Anbuhi et al. first pointed out this and solved it by sandwiching a paper chip with two flexible films, and the gaps between the films greatly accelerated the speed of the liquid.20 Dou et al. placed chromatography paper pads into the PDMS reservoirs for 3D reagent immobilization, so that the merits of easy sample storage of the paper and good hydrodynamic controllability of the flows in PDMS channels are both maintained.23 More intimate combination of the paper and channels comes from the introduction of channels to the μPADs. Crooks' group proposed the concept of a hollow channel which is generated by sandwiching an engraved hollow channel layer with two other layers by wax printing and clamp sealing.24,25 Later on, Martinez' group devised two-ply channels which were formed by putting two layers of wax patterned paper together and generating channels between the two layers.19 These are all valuable contributions. However, limits still remain. Further considerations like flexibility for designing the topology and geometry of channels, reliability of interlayer bonding, throughput for tests and chip reusability should be taken into account.
PDMS is one of the most widely applied materials for microfluidics and it was first devised by Whitesides’ group, the same as μPADs.26 It is highly biocompatible, transparent, stretchable and cheap,27 and thus is the most typical material for channel based microfluidics. However, for a long time its influence was limited in the microfluidic field due to difficulties in surface modification. Interestingly, paper has complementary characteristics for easy surface modification.5 Thus, if PDMS and paper were combined properly, they might complement each other perfectly and overcome their own weak points. The combination is not only limited to the individual characteristics, but the combining of the two areas: channel based microfluidics and paper based microfluidics. PDMS patterning on paper can be an ideal glue for paper to bond with PDMS channel structures. Their combination might initiate a new hot topic in microfluidics, considering all the attractive characteristics of both paper chips and traditional channel based chips.
Fortunately, some early pioneering research studies have paved the way to patterning PDMS on paper. It was first realized using a home-made injector based printing system10 the year after the birth of μPADs. Later on, flexographic printing,11 ink printing11 and screen printing28 based PDMS patterning on paper were all proposed. However, despite the technological advancements, PDMS patterning on paper has neither gained wide popularity for μPADs, nor been successfully combined with PDMS channel structures. Basically, this is because the patterned PDMS is not delicate enough compared with existing techniques like that for wax. The final resolution of the pattern is limited by two characteristics of the uncured PDMS: high viscosity and high penetration ability. The high viscosity of the PDMS makes it difficult for it to be directly printed unless diluters like xylene or toluene are applied. Direct printing causes large pixel dots, while diluted ink printing is linked with poisonous reagents. The high penetration ability of PDMS on paper causes PDMS to quickly exude outside the predefined boundary within tens of seconds upon touching the material, leading to a deformed geometry with low linear resolution. It also results in surfaces that are too rough to allow for further sealing with other layers.
In this paper we are reporting our recent progress in preparing microfluidic PDMS on paper (POP) devices based on a screen printing technique and temperature accelerated PDMS curing. By adopting this technique, the PDMS can cure within seconds on paper, and the exuding phenomenon can be greatly inhibited. By tuning the printing times and the sitting time for pre-curing, the PDMS penetrating depth can be controlled delicately from semi-penetration to thorough penetration. The PDMS pattern on polyether sulphonyl (PES) paper with a line width of 60 μm can be realized in single layer mode, and is near 150 μm for thoroughly penetrated lines. This not only helps us to overcome the barriers to PDMS-paper hybridization, but also opens the door to a novel concept for POP devices: the integration of microfluidic PDMS channels with PDMS patterned paper. Consequently, distinctive regions for porous paper and smooth channels can be generated on the same chip, which eventually leads to the functional extension of traditional μPADs. As an example to demonstrate this general concept, POP device based multiplex colorimetric analysis was carried out for typical liver function markers: serum protein, alkaline phosphatase (ALP) and aspartate aminotransferase (AST). The three markers can give the first indications for deteriorating liver function. Based on the special structure of the POP device, we proposed a shooting star sampling mechanism that enabled quantitative and fast reagent loading and sampling for a matrix of discrete testing points that requires no equipment, only manual operation by hand.
As depicted in Scheme 1B2 and B3, two typical PDMS channel structures were adopted in our experiment: the channel array and the zigzag channel. For reagent loading, the PDMS channel array was applied in which each column needs to be loaded with different reagents, while the zigzag channel was chosen when all the testing points were to be modified with the same reagent. After that the loaded reagents were dried with a nitrogen flow. For multistep reagent loading, we just repeat the loading-drying steps. For sample loading, the PDMS channel array was also adopted for the analysis of several samples, or a zigzag channel was adopted when only one sample was to be analysed.
To measure the volume of sample trapped on each testing point, the length of the short liquid column was measured constantly. The shortened length of the liquid column was translated to volume information by multiplying with the cross-sectional area of the channel.
To generate the standard curve for serum protein, AST and ALP, the PDMS patterned PES-80 testing point array was preloaded with the corresponding colorimetric testing reagents. Before usage, the PDMS channel array layer (with 8 parallel channels) was mounted onto the surface of the PDMS patterned PES paper layer, so that the colorimetric testing could be repeated 8 times for each concentration.
To apply the chip for liver function analysis, the PDMS channel layer depicted in Scheme 1B was used to combine with the PDMS patterned PES paper. After sampling with the shooting star mode, a cell phone camera was used to monitor the colour change on the arrayed testing regions. Images were taken over 10 minutes of the reaction. The readings were extracted instantly using colour analysis.
Though their combination can be in different modes, all of them can be broken down into two unit steps: surface patterning and inter-layer bonding. Since PDMS can be reliably sealed with another piece of PDMS simply by reversible bonding (just align and put them together), we only need to focus on patterning the PDMS surface that is smooth enough. The major difficulty in their combination lies in the surface patterning step. Until now, delicate control over PDMS patterning has been challenging. Control over factors like penetration depth, linear resolution and surface smoothness all need to be considered.
In a typical heat triggered polymerization process like that for PDMS, curing temperature usually determines the curing speed, and a faster curing speed usually leads to better resolution because it reduces the time for pre-polymers to spread over the predefined boundaries. Thus we first tried to screen for better curing temperatures. Fig. 1 is an image of a series of droplets cured under different temperatures on a filter paper. The diameters of the cured droplets decrease with the increase of temperature, which implies that the spread of the PDMS precursors can be well inhibited by increasing the curing temperature. We also quantified the time for curing. As is shown in Fig. S1A,† in the tested range the curing time decreases significantly with the increase of temperature. In addition, Fig. S1B† details the geometries of the cured droplets, and these indicate that the penetration depth of the droplet also decreases with the increase of temperature. As a whole, these data prove that at higher temperatures we may get PDMS patterns with faster curing speeds and better resolution.
Fig. 1 PDMS droplet curing over different temperatures. The numbers indicate the curing temperature (°C) corresponding to each droplet. |
For condition screening, three factors were examined: the curing temperature, the number of printing times and the viscosity of PDMS. These factors would influence the final geometry of the pattern. To test their influence, the PDMS penetration depth was used as the index, which can reflect well the delicacy of the patterning technique for both semi-penetrated patterns and thoroughly penetrated patterns.
To examine the influence of curing temperature, a single layer of screen printed PDMS was applied on each paper. As shown in Fig. 2A, for all four types of paper, the PDMS penetration depth decreases with the increase of temperature. However, compared with the results for PDMS droplets in Fig. S1B,† the inhibition of the penetration is not significant. This is mainly because of the use of screen printing, where the amount of PDMS applied to a unit area is constant and is much less than in the droplet case. Thus the decrease of the penetration depth here is totally attributed to the increase of the temperature. Besides the general tendency, there are also differences between the four types of paper. At 160 °C, the percentages of penetration are 40% for PES-80, 33% for PES-22, 80% for filter paper and 80% for printing paper. The results imply that the penetration depth is much smaller for the PES papers than for filter paper and printing paper. The differences come from the porosity which defines the amount of PDMS required to fill up a unit volume of paper. For PES papers, though the pores are small, they are highly dense, while for filter paper and printing paper they are the opposite. In addition, the corresponding error is also much smaller for PES papers due to their better structure uniformity. Thus patterning on PES papers is more controllable.
Besides temperature, Fig. 2B shows the relationship between the number of printing times and penetration depth. The results imply that by increasing the number of printing times we can effectively increase the penetration depth of the resulting PDMS pattern. Penetrated patterning throughout the paper can be achieved for filter paper, printing paper and PES-80 paper for 5 times of repeated screen printing. However, penetrated patterning can hardly be achieved for the PES-22 paper after 5 times.
To achieve better control over the penetration depth we further explored the relationship between penetration depth and PDMS viscosity. To generate PDMS with different viscosities, PDMS samples have been precured at room temperature for different predefined times. It is well known that at room temperature, the degree of polymerization for PDMS will gradually increase with time, until it ultimately forms a uniform piece of polymerized PDMS. As expected, the experiment data in Fig. 2C show that PDMS with a longer precuring time can lead to a lower penetration depth after curing. It is also noticeable that the errors associated with the depth increase with precuring time, which become prominent over 6 h. Thus there's a balance between better controllability of penetration depth and the error. PDMS pre-curing over less than 6 h is recommended. The resulting thinner layer of patterned PDMS is especially valuable for defining 3D microchannels on a single layer mode. The better we can control the penetration depth, the more delicate 3D structures on a single layer of paper can be constructed in the future. This is one of the most valuable developing directions for 3D paper chips, and is thus a good measurement for the delicacy of the proposed patterning techniques. It can also contribute to increasing the flexibility of chip design and thus more powerful functions.
First, we tried to generate standard curves for the three markers. As depicted in Fig. 4A1, a typical POP device was composed of an arrayed PDMS channel layer and a piece of PDMS patterned paper preloaded with testing reagents was applied. A series of standard sample solutions was tested using a shooting star mode. It is so named because in this mode an isolated short column of liquid is quickly flowed throughout the channel like a star shooting across the sky. Because PES is hydrophilic and PDMS is hydrophobic, after liquid was flowed through the region, discrete testing points were loaded with sample solutions through one injection. More interestingly, the amount of liquid trapped on the testing points is generally the same because it is defined by the geometry and surface tension of the testing point region. In our case, the amount of liquid trapped on each testing point reads as 0.15 μL. This result also implies that we can do serial quantitative sampling in the submicroliter range on this platform, simply by hand. This is a big advantage for in situ applications, where equipment and apparatus are limited. In addition, since the liquid is propelled by a hand controlled syringe, the whole process for one injection takes only seconds to accomplish, and the whole process for sample injections that are needed for a standard curve takes less than two minutes, which greatly reduced the time error for colour development between each testing point.
The standard curves and the original colorimetric data visualized in Fig. 4A2–4 supported the above conclusions. The inserted images, which depict the resulting colours of the testing point array of the papers, were taken after ten minutes of colour development. They showed that in each column the colours for each testing points are homogeneous. Meanwhile in each row, the colour changed gradually. Quantitative analysis of the colours in the figures further implies that the data in each column are in good consistency from point to point, with errors in an acceptable range. The effective testing ranges sufficiently covered the range for healthy people as indicated in the shadowed bars, which proved that this test can distinguish abnormal samples from healthy ones.
To check possible crosstalk between the sample and the reagents on the testing points, the selectivity was examined. As shown in Fig. 4B, columns 1–2 were pre-loaded with AST and the ALP testing reagent, and after contact with the solution of protein, no colour change was observed, indicating no crosstalk between them. The results are the same for ALP and AST, so it can be deduced that the colour changes for the reagent are highly selective for their targets.
The analysis results for mixtures of the three markers were tested, as shown in Fig. 4C. In this case, a test with 8 repeat units for the three markers was carried out over one shooting star process. The results from each repeat unit show good consistency with each other. Although only three markers were tested, the system with 8 × 8 testing points actually can support tests for 8 samples with 8 multiplex markers. The above data, in general, indicate that combining the PDMS channels and the PDMS patterned paper has provided the resulting POP device with many advantages. Besides easy reagent immobilization, which is typical for μPADs, the POP device is accustomed to the shooting star liquid loading mechanism, which makes fast, quantitative, high-throughput and homogeneous sampling and analysis possible.