The conventional quantitative method for the analysis of inorganic elements in polymer matrices is a complex and time consuming process that presents a significant risk for error. Typically, polymers are digested in a microwave oven or other devices under high temperature and pressure for several hours while employing different mixtures of high purity acids. In many cases, particularly when high concentrations of doped elements are present, the digestion is often incomplete and therefore the reproducibility depends strongly on the type of polymer and additives used. A promising alternative technology that allows for the direct analysis of these polymers without digestion is laser ablation ICP-MS. Due to a lack of available reference materials and the presence of matrix dependent effects, a precise calibration cannot be obtained. In order to compensate for the matrix dependent effects the use of internal standardization is necessary. In this study the correlation between the carbon released during the ablation process and the 13C signal detected by ICP-MS and its use as an internal standard are investigated. For this purpose, twenty-one virgin polymer materials are ablated; the released carbon is determined and correlated with the corresponding integrated 13C signal. The correlation resulted in a direct relationship between the ablated carbon and 13C signal demonstrating the potential ability to neglect at least some of the matrix dependent and transport effects which occur during the laser ablation of virgin polymers.
The use of laser ablation (LA) as a direct sample introduction technique in ICP enables a significant reduction in the sample preparation steps required.8–10 However this technique also exhibits several disadvantages. The major problems arise from elemental fractionation9,11,12 such as matrix and transport effects.13–16 These factors restrict the routine use of this technique for the quantitative determination of elements in plastics.
Differences in the physical properties and UV absorption characteristics of varying polymer classes strongly influence the ablation efficiencies, which therefore results in different ablation rates.17 One approach in order to correct the different ablation efficiencies of varying polymers is to use the carbon 13C isotope signal as an internal standard.18–23 Carbon is primarily chosen because it is the major element present in all polymer based materials and is homogeneously distributed throughout the polymer matrix.
One of the major limitations that impacts the suitability of 13C as an internal standard arises from the formation of different products in the ablation process of different polymer types.15,24,25 The deposition of the carbon containing material onto the sample surface during the laser ablation of organic materials appears to also limit this approach.26,27 Analytical investigation indicates that the type of carbon ablation products depends on the sample as well as the ablation gas used.24,26,28 Investigation into carbon-containing gaseous species and carbon-containing particle phases elucidated a matrix dependent behavior concerning the amount of carbon-containing gaseous species.24
Despite the elemental fractionation, the suitability of 13C carbon to compensate for variations in the ablated mass and mass transport efficiencies was previously reported. Austin et al. used the 13C signal for the correction of these effects for the ablation of chicken breast tissue.19 Becker et al. corrected the water content in tobacco leaves using 13C as an internal standard.29 The determination of silver in different polymers resulted in good accordance of measured and expected values. A polypropylene sample was chosen as a standard and the silver contents of other samples consisting of polypropylene, acrylonitrile–butadiene–styrene and nitrile butadiene rubber were determined using 13C as an internal standard. Only the measurement of silver in polyoxymethylene provided results four times higher than that expected.21
The laser ablation studies performed thus far focused only on a small range of polymers and other organic materials. A wide variety of polymers with different structures, functional groups and atoms are commonly used. Due to the varying compositions the polymers possess different material properties which affect the behavior in the laser ablation process.
General investigation regarding the influence of the polymer structures on the ablation process for analytical purposes has not been conducted until now. In order to utilize 13C as an internal standard in routine applications, a correlation between the 13C-ICP-MS signals of different polymers must be present. It is necessary to assure this relationship before a possible quantification using 13C as the internal standard can be performed. Only in the case of an established coherence might a correction utilizing the 13C signal be feasible. Therefore, the influence of several parameters and the applicability of 13C correction must be studied in further detail.
In this study the correlation between ablation crater volumes of twenty-one virgin polymers and the resulting 13C-ICP-MS signals was evaluated. The ablation crater volumes were measured and the calculated ablated carbon was correlated with the corresponding 13C ICP-MS signal.
Laser parameters | |
Ablation mode | Single point measurement |
Laser pulse energy | 0.6 mJ |
Pulse repetition rate | 20 Hz |
Number of shots per point | 100/50 |
Spot size | 200 μm |
Ablation gas flow | 200 mL min−1 |
![]() |
|
ICP-MS parameters | |
Nebulizer gas flow | 1.26 L min−1 |
Auxiliary gas flow | 1 L min−1 |
Outer gas flow | 15 L min−1 |
ICP RF power | 1250 W |
Dwell time | 50 ms |
ICP-MS and laser parameters were optimized in advance by means of an in-house produced LDPE standard. The general production of these standards can be found elsewhere.30 For optimization single line scans were applied since a drop of signal intensity over time for single point measurements was observed.
For line scan measurements, the scan rate was 25 μm s−1. Nebulizer gas flow, ICP RF power, ion lens, cell path voltage, ablation gas flow and pulse repetition rate were optimized to attain the highest 13C signal intensity. The laser ablation system used limited the pulse repetition rate to 20 Hz. This pulse repetition rate resulted in the highest signal intensity. The in-house built ablation cell consists of a cylindrical cavity with a cell volume of 4.75 cm3, one gas inlet and one gas outlet. Due to the low cell volume a maximum ablation gas flow of 200 mL min−1 was applied. The nebulizer gas flow of the ICP-MS is added to the ablation gas behind the ablation cell which allows for sample exchange without influencing the plasma stability.
The laser pulse energy of 0.6 mJ enabled good sensitivities. Higher values led to a high amount of carbon containing material being deposited on the ablation cell window. The laser pulse energy was determined with a pyroelectric energy meter Orion-PE PE25BF-V2 (Ophir Optronics Solutions Ltd, Har Hitzvim, Israel).
Subsequent to LA-ICP-MS measurements, the polymer surfaces as well as the ablation crater volumes and ablation crater shapes of each ablation crater were measured with a confocal microscope μ-surf basic (NanoFocus AG, Oberhausen, Germany). The software μsoft analysis automatically calculates the ablation crater volumes as well as the maximum depth of the analyzed ablation crater. For best results the highest resolution of 50-fold magnification and height increments of 48 nm were chosen.
A total amount of twenty-one different virgin, undoped polymers were analyzed during this study. A detailed description of the polymers used in this study is provided in the ESI.†
The investigated polymers contained the classes polyolefins, polyamides, polyesters, polyethers, polyacetals, styrene containing polymers and polysulfones. In order to obtain even polymer surfaces, which are favorable to the laser ablation and confocal microscope measurements, the granulates of sixteen polymers were heated and pressed into round discs with a diameter of 40 mm using a mounting press (SimpliMed 3000, Buehler, Duesseldorf, Germany). Due to the high melting points of some investigated polymers, or the beginning of decomposition at higher temperatures, the production of plates was not possible for all polymers with the available laboratory equipment. Therefore, pellets of the polymers (polyethersulfone (PES), polyamide 6 and 66 (PA 6 and PA 66), poly(vinyl chloride) (PVC), and polyphthalamide (PPA)) were ablated. Due to the uneven surface, the precise determination of the exact ablation crater volumes for these polymers was restricted.
LA-ICP-MS experiments were initially conducted using ten polymers in order to determine the best ablation parameters for measuring the ablation crater volumes. These polymers included acrylonitrile–butadiene–styrene (ABS), polystyrene (PS), polycarbonate (PC), polyethylene terephthalate (PET), polyamide 12 (PA 12), polyamide 6/12 (PA 6/12), polyoxymethylene (POM), polyether ether ketone (PEEK), polyethersulfone (PES) and polysulfone (PSU). Five single point measurements with 50, 100 and 200 laser shots each were applied to the mentioned samples. These measurements were performed using He as the carrier gas with a flow of 200 mL min−1. In order to fit into the in-house ablation cell, the samples were cut down to a size of about 5 mm × 5 mm using a knife.
Later on five single point measurements with 100 laser shots each were carried out on defined positions of all twenty-one polymer samples. With a distance of 500 μm between the spots, five ablation spots were applied in a row at the center of the ablation cell along the x-axis.
The total mass of ablated carbon was calculated using the ablated volume, the density of the polymers and the amount of carbon within the polymers.
The densities of the polymer discs were calculated using the heights, diameters and masses of the samples. The density determinations of the polymer pellets were conducted using a pycnometer (BRAND GmbH & Co KG, Wertheim, Germany). The amount of carbon was determined using VARIO Macro (Elementar Analysensysteme GmbH, Hanau, Germany) at Federal Institute for Materials Research and Testing (BAM, Berlin, Germany).
The measured 13C-ICP-MS signals of all polymers in each experiment were integrated over time. The mean value and the corresponding standard deviation of five single measurements of the 13C signal were calculated. The gas blank was subtracted from the mean. Afterwards, the integrated 13C signals were correlated with the total mass of ablated carbon of the corresponding polymers.
Microtome sections of 10 μm thickness were prepared (Leitz 1400 sledge microtome, Leica Microsystems, Wetzlar, Germany) and UV/VIS spectra were recorded from PA 12, LDPE, HDPE, ABS and HIPS in the range from 190 to 400 nm with an UV/VIS spectrometer (Specord210, Analytik Jena AG, Jena, Germany) to specify the ablation characteristics of different polymers. For other polymers, the preparation of microtome sections with 10 μm thickness was not possible due to their brittleness. UV-VIS measurements were not possible for microtome sections with thicknesses exceeding 10 μm.
These measurements resulted in figures showing elevation profiles. The crater bottoms, which are the deepest parts of the samples, are depicted in a dark blue color, while the polymer surfaces, the highest parts of the samples, are depicted in a light rose color as can be seen in Fig. 1. Colors between dark blue and light rose reflect the walls of the ablation crater and give information about the shape (cylindrical or conical) of the ablation craters.
The conducted confocal microscope measurements showed a dependency of the topology of the ablation crater on the type of polymer applied. Most pronounced conical ablation craters were observed for LDPE, HDPE, POM and PP, while most pronounced cylindrical ablation craters were observed for the polyamides. The other polymers exhibited primarily conical shaped ablation craters without flat bottoms (Fig. 1). Non-cylindrical ablation craters have been found before for glass and polymer samples.31–33
The attained results regarding the ablation crater appearances of different virgin polymers lead to the conclusion that the polymeric backbone has an important influence on the ablation behavior. Those polymers whose topologies diverge most from the cylindrical shaped ablation craters are the polyolefins.
The structures and related properties of polyolefins differ from those of most other polymers used in this study. Polyolefins are composed only of carbon and hydrogen, while POM also contains oxygen in the backbone.
The structures of polymers have an important influence on the absorption of UV light due to their different functional groups.17,34–36 The difference in the structures leads to a less pronounced absorption of UV light by the polyolefins when compared to other polymers with strongly absorbing functional groups in their structures like ABS or the polyamides. In Fig. 2 the UV/VIS spectra show stronger absorption of virgin ABS and virgin PA 12 compared to virgin LDPE. The absorption of ABS and PA 12 especially in the range from 190 to 225 nm is too strong to perform reliable measurements. Nevertheless it can be stated that the UV absorption in this range is stronger for these two polymers compared to LDPE. The UV/VIS spectrum for HIPS was equal to that for ABS, and the UV/VIS spectrum of HDPE was equal to that for LDPE.
In general, samples with stronger UV absorption show smaller penetration depth and therefore smaller ablation craters.33,37 However, in the case of polymers not only the absorption of UV light but also their thermal properties and the ablation products' influence on the ablation crater volume have to be considered. Pham et al.38 have shown a decrease in the ablation threshold with decreased glass transition temperatures. Kunz et al. raised the UV absorbance of polyestercarbonate samples through increased ester parts. The stronger UV absorbance resulted in higher ablation rates due to the increased release of gaseous decomposition products.39
The more pronounced absorption of UV light by polymers with aromatic groups in their structure resulted with the exception of PPA and PET in lower ablation crater volumes when compared to polymers without aromatic groups in their structure (Fig. 3). The largest ablation crater volume (POM) was four times larger than the least ablation crater volume (PEEK). Although PA 12 showed stronger UV absorbance the ablation craters of the polyamides were comparable or even larger than those of the polyolefins.
Due to the weak absorption of UV light by the polyolefins, the energy from the laser penetrates deeper into the polymer leading to a higher heating of the material. The higher temperature within the polymer leads to a higher degree of bond breakage. This results in the ejection of more material, and in deeper but less smooth ablation craters. The less smooth ablation craters were also detected for other weakly absorbing samples.33,37
Besides the difference in the polymeric structure, resulting in different UV absorption, also the thermal properties change. Especially the glass transition temperature is a critical factor. The glass transition temperatures of the polyolefins and POM show values of −100 °C (LDPE) to −10 °C (PP), by far the lowest compared to those of the other polymers used in this study. The glass transition temperatures of polymers with aromatic structures are between 47 °C (PBT) and 225 °C (PES) and those for polyamides are between 37 °C (PA 12) and 50–60 °C (PA 6/PA 66). Samples with low glass transition temperatures show better flexibility in the polymer chains when compared to samples with higher glass transition temperatures. Due to the better flexibility less energy is required to break the bonds and to start the ablation.
Hence, the formation of comparable ablation crater volumes requires less energy for polymers with lower glass transition temperatures. For ABS and PA 12 equal UV absorption was found but the ablation crater volume of PA 12 was about 3.5 times greater than the ablation crater volume of ABS. The glass transition temperature of PA 12 is 37 °C and that for ABS is 110 °C. Since less energy is required for the ablation of PA 12 compared to ABS the ablation under the same conditions results in greater ablation crater volumes for PA 12. Due to the low glass transition temperature of LDPE even less energy is required for the ablation. Although this should result in an even deeper ablation crater for LDPE the ablation crater of PA 12 is about 2 times greater than that of LDPE. This probably can be attributed to the increased formation of gaseous decomposition products. The decomposition of polyamides at higher temperatures results in the release of gaseous components like CO, CO2 and H2O.40,41 The detailed investigation of the kind of ablation product was not the aim of this work and has to be analyzed in greater detail in further studies.
The difference in the glass transition temperature not only influences the ablation crater volume but also the ablation crater shape. In the cases of LDPE, HDPE, PP and POM, which have the lowest glass transition temperatures, the material, especially on the walls of the ablation crater, melts to a higher degree than that in the cases of other polymers. Thus, the ablation performance is different for polymers with low glass transition temperatures since the higher flexibility of the polymer chains of these polymers leads to an easier set-back of the molten material into the ablation crater. This results in elongated shaped ablation craters as observed for LDPE for example (Fig. 4 and 6). The profiles of the resulting ablation craters are influenced by the ablation gas flow. In the direction of the ablation gas and horizontal to it the ablation crater shows a conical shape. However, in the Y-direction (direction of ablation gas flow) it is less pronounced compared to that in the X-direction (horizontal to the ablation gas flow). The ablation gas transports the molten material out of the ablation crater. Horizontally to the ablation gas flow the molten material can more easily set back into the ablation crater forming a more pronounced conical structure. Polymers with higher glass transition temperatures do not tend to melt at a high degree. The profiles in the Y-direction (direction of ablation gas flow) are comparable to the profiles in the X-direction (horizontal to ablation gas flow) as observed for PA 12 for example (Fig. 5). The upper parts of the PA 12 ablation crater are cylindrical in both profiles. The lower parts also show differences. The profile horizontal to the ablation gas flow shows a conical shape while a cylindrical shape is more pronounced in the direction of the ablation gas flow. Also polymers with higher glass transition temperatures tend to melt but to a much lower degree compared to polymers with very low glass transition temperatures.
![]() | ||
Fig. 4 Profile of a LDPE ablation crater; left: Y-direction in ablation gas flow; right: horizontal to ablation gas flow. |
![]() | ||
Fig. 5 Profile of a PA 12 ablation crater; left: Y-direction in ablation gas flow; right: horizontal to ablation gas flow. |
The different ablation behaviors of LDPE, HDPE, PP and POM compared to those of other polymers were also observed through consideration of the surrounding surface of the ablation craters. The survey with a confocal microscope not only enables the measurements of ablation crater volume and form, but furthermore obtains information about the volume of deposited materials on the surfaces of the polymers.
The ablation of the aliphatic polymers LDPE, HDPE, PP and POM led to deposits of molten materials on the surfaces (Fig. 6). This also indicates that these polymers melt at a higher degree than that in the case of the other polymers during laser ablation. The formation of conical ablation craters caused by a set-back of the molten material furthermore is confirmed by the deposited material on the surfaces of aliphatic polymers. Under the applied conditions deposits on the surfaces of approx. 2% of the ablation crater volume were found.
Polyolefins and POM exhibited a different behavior compared to the other polymers ablated. The products formed during the ablation of polyolefins can condense more easily and redeposit on the polymer surface.25 During the ablation of POM the formation of larger particles, which can easily condense and redeposit on the polymer surface, is very likely.
For polymers offering measurable ablation crater bottoms for 50, 100 and 200 laser shots, either a linear or a polynomial increase of the integrated 13C signal curve was observed. These findings correlated well with the determined crater volumes. In the case of a linear increase of the integrated 13C signal the ablation crater volume also showed a linear behavior (Fig. 7). For a polynomial increase of the integrated 13C signal, a polynomial increase of the ablation crater volume was observed as well (Fig. 8).
![]() | ||
Fig. 7 Linear correlation for number of laser shots and 13C integral (primary Y-axis) and for number of laser shots and ablation crater volume (secondary Y-axis) observed for PA 6/12. |
![]() | ||
Fig. 8 Polynomial correlation for number of laser shots and 13C integral (primary Y-axis) and for number of laser shots and ablation crater volume (secondary Y-axis) observed for PET. |
An influence of the polymer structure on either a linear or polynomial increase of integrated 13C signal and crater volume was observed. The structure of the polymeric backbone not only influences the ablation crater structure but also the increase of the 13C signal intensity. Therefore a first direct coherence between the ablation crater volume and hence the ablated carbon and the measured 13C-ICP-MS signal was determined.
It was found that the increase of the integrated 13C signal as well as the increase in crater volume were related to the structure of the ablation crater. For ablation craters with a pronounced cylindrical shape, the increase was linear. The less pronounced cylindrical shaped ablation craters resulted in a polynomial increase in 13C signal and crater volume.
The calculated ablated carbon and integrated 13C signal intensities of seven different polymers were correlated choosing only those polymers with measurable ablation craters for 50, 100 and 200 laser shots. Satisfactory coefficients of determination of R2 = 0.896 (50 laser shots per point), R2 = 0.833 (100 laser shots per point) and R2 = 0.930 (200 laser shots per point) were obtained.
Since crater volumes for 100 laser shots were measureable for the first ten chosen polymers, all twenty-one polymers were ablated. The ablation crater volumes were measured followed by a correlation of the mean of the integrated 13C signal intensities with the calculated total amount of carbon ablated which resulted in a coefficient of determination of R2 = 0.885 (Fig. 9). The measurements have shown an existing correlation between ablated carbon and integrated 13C signal intensity. This correlation is independent of the chemical compositions of the different polymers.
![]() | ||
Fig. 9 Linear correlation of 13C signal integral and ablated carbon for 21 polymers ablated in a He atmosphere applying 100 laser shots with a coefficient of determination of R2 = 0.885. |
The correlation between ablated carbon by LA and the 13C signal measured by ICP-MS was investigated to correct for changes in the ablation rates of different polymers.
Some effects caused by ablation, such as interaction of the laser with the sample and the resulting crater shape can be corrected using the 13C signal.
Nonetheless, the best means for quantification is a matrix matched standard reference material. Only if such a material is unavailable should a correction using the 13C signal be applied for different polymers. Some effects caused by ablation, such as interaction of the laser with the sample and the resulting crater shape, are visible in the ion intensity as well. Differences in the ablated mass can be corrected. However, the influence of inorganic additives on the ablation crater shapes and volumes of different polymers must be investigated in further studies. Due to differences in the absorption properties of UV light of inorganic additives, an influence on the ablation craters is very likely.