D.
Godfrey
,
J. H.
Bannock
,
O.
Kuzmina
,
T.
Welton
and
T.
Albrecht
*
Department of Chemistry, Imperial College London, London, UK SW7 2AZ. E-mail: t.albrecht@imperial.ac.uk; Tel: +44 (0)20 759 43704
First published on 7th December 2015
Cu extraction from chalcopyrite ores is typically a slow process that involves aggressive chemical reagents with significant environmental impact. Ionic liquids (IL) have been proposed as a potentially more benign solution, but the sheer number of IL variants complicates the search for the most efficient solvent systems. Here, we present an automated electrochemical platform that allows for screening of 180 and more leaching samples in parallel with minimal solvent consumption. In a proof-of-concept study, we screen 25 samples with different IL and water contents, and find two orders of magnitude difference in leaching performance within this array. The best performing system is then applied in a tank leaching configuration, with real-time electrochemical monitoring of Cu evolution in solution. All electrochemical data is found to be in excellent agreement with off-line ICP-AES data.
Efforts to decrease reliance on traditional, high-consumption pyrometallurgy from flotation concentrates (∼30% Cu) have focused on enhancing hydrometallurgic beneficiation of sulfide ores,4 with the principal copper ore chalcopyrite (CuFeS2(s), ∼70% global reserves5) receiving primary attention.
The industry is also increasingly challenged with processing lower-grade chalcopyrite ores,6 driving the development of new solution-based processes, applicable to both (bio-)heap and tank leaching.7 Conventional acid-oxidant chemical leaching systems, including the commonly employed Fe2(SO4)3–H2SO4(aq) system, are established and cheap (see recent reviews6,8). However, the overall leaching process is slow at moderate temperatures, whilst also ultimately limited by surface passivation9 or other kinetic effects.10,11 Hence, there is still significant potential towards improving the leaching performance.6
Accordingly, improved Cu extraction from CuFeS2(s) has been achieved, e.g. by using alternative acid-oxidant combinations, microorganisms, ultra-fine grinding, elevated temperatures and pressures.12 However, the benefit of such methods is offset by increased cost and energy consumption, incompatibility with existing workflows or more generally higher environmental impact.
In the search for alternative lixiviant systems, ionic liquids (ILs) have emerged as an interesting alternative. ILs have well-documented benefits as solvents in synthetic chemistry,13 and it is conceivable that their chemical structure could be tailored in such a way to avoid the formation of kinetic barriers and maintain a high Cu extraction efficiency over time. To this end, notable dissolution enhancement, compared to ferric H2SO4-based lixiviants, has been reported for first generation alkylimidazolium hydrogen sulfate ionic liquids towards CuFeS2(s). Maximum Cu recovery of 86.8% from a Cu concentrate (∼70% CuFeS2(s)) by 100% [C4C1im][HSO4] (bmim·HSO4; pH ∼ −1 [ESI†]) presented a ∼60% enhancement compared to 1 M H2SO4(aq) benchmark solutions (pH −0.3), both containing excess Fe2(SO4)3 oxidant.14 However, as noted above, these cited enhancements have not been rigorously normalised for medium pH, such that there may be no IL-related Cu extraction enhancement. A subsequent kinetic study of [C4C1im][HSO4](aq) leaching applied to chalcopyrite ore (∼20% Cu), focused on the effect of temperature and agitation, measuring an Arrhenius-type activation energy from chalcopyrite dissolution of 69.4 kJ mol−1.15
While such studies are a promising first step, further improvement is highly likely, given the structural diversity of ILs. However, the complex nature of the dissolution process renders ‘ab initio’ rational design of optimal lixiviant systems out of reach. On the other hand, screening a large number of ILs or lixiviant compositions simultaneously and at a small scale appears to be a more realistic option. Such a combinatorial approach would furthermore allow for the optimisation of the leaching performance over several generations of leaches, where the best performing lixiviants could then be upscaled for more in-depth studies. Similar strategies have yielded excellent results in protein design and other areas.16
Here, we demonstrate that such a combinatorial methodology is indeed a powerful tool for screening and improving the Cu extraction performance of IL-based lixiviants, Fig. 1. Similar robotic electrochemical workstations have been applied in other contexts before, namely for automated combinatorial electrochemistry in large sample arrays for microelectrode studies17,18 and in endothelial cell NOx excretion screening.19
Our platform allows for the screening of up to 180‡ samples at the same time, with electrochemical, in situ monitoring of Cu extraction from CuFeS2(s). We found that the detection capabilities of cupric ion-selective electrodes (ISE) were insufficient to detect Cu2+ in solution in the presence of ILs at the initial stages of the leach ([Cu] < ∼1 mM – see ESI Fig. S4†), but that Cu electrodeposition/anodic stripping improved the detection limit by approximately one order of magnitude, based on the conditions used here, see below. Comparison with ICP-AES as a standard ex situ method, yielded a 1:
1 correspondence with the electrodeposition/anodic stripping results. After calibration and testing with model samples, we have applied the robotic screening platform to an array of 25 leaching samples in a proof-of-concept study. We find a 100-fold variation of the leaching performance between the best and worst performing lixiviants within the array. Subsequently, we examined the best-performing sample via electrochemical monitoring in a 120 mL scale tank leaching reactor over approximately 6 days.
Since Cu ion-selective electrodes (ISE) appeared to be a facile and straightforward real-time detection approach for Cu2+ in solution, we first tested Cu ISE suitability for the task at hand (see ESI for full experimental details inc. Fig. S4†). Using a commercial ISE sensor (Cole-Palmer, Cupric Combination ISE), flat-line indistinguishable sensor response was obtained for [CuSO4](aq) < 10−6 mol dm−3 and ∼10−4 mol dm−3 in H2SO4(aq) and IL(aq) media, respectively. Above the respective lower [Cu2+] detection limit, all studied calibration plots exhibit near ideal Nernstian potential dependence of 29.6 mV per [Cu2+] decade (75 mM H2SO4 = 26.9 mV; 450 mM [C4Him][HSO4] = 30.6 mV; 450 mM NH4·HSO4 = 31.7 mV). Thus, in strongly Cu-coordinating IL(aq) media lower cupric detection limits are deemed unsuitable for the present purpose of monitoring ambient IL(aq) leaching on timescales of <2 days. However, as we show below, electrodeposition combined with anodic stripping of copper (ASV) indeed enables real-time quantification of [Cu2+] in solution, with sufficient sensitivity even in the presence of ILs.
In light of our aforementioned automation objectives, a powerful robotic electrochemical platform has been built (Fig. 1). For our ASV studies, the instrument is fitted with a Pt-disc working electrode, assembled into a glass fused probe construct (dWE = 1 mm, Pt CE, Ag/AgCl RE). The fabricated probe is docked at the labelled ‘electrode mount’. Motorised probe positioning and potentiostat functions are programmatically controlled via USB 2.0 serial port connectivity. Further details can be found below and in ESI.†
Initial testing of the platform setup included determining the geometric factor for several fabricated electrodes in certified KCl(aq) conductivity standards (Sigma Aldrich), followed by accurate measurement of CuSO4(aq) solution conductivities with <6% error (1–50 mmol dm−3).22 Automated data acquisition for various IL(aq) ASV calibration plots (see ESI – Fig. S8†) provided some ASV specific platform validation, however more complex electrochemical study was desirable, as described next.
Fe3+ ions are a common additive oxidant for acid-sulfate chemical leaching (1) and are also regenerated through the oxidation of leached Fe2+(3).6,8 In a similar fashion to previous wastewater studies,23 the influence of [Fe3+] on [Cu2+] ASV was investigated. A conventional ASV approach was used, in which the stripping process occurs in fresh electrolytic media (0.5 mol dm−3 H2SO4(aq)), avoiding potential interference from electroactive species within the analyte-bearing solution (i.e. a ‘2-vial’ ASV configuration). An array of 50 samples containing 1–10 mmol dm−3 [CuSO4](aq) and 0–45 mmol dm−3 [Fe2(SO4)3](aq) (combinations indicated in Fig. 2), were prepared in 75 mmol dm−3 H2SO4(aq) electrolyte (pH 1.3 ± 0.05) and sequentially subjected to the described ‘2-vial’ ASV procedure. ICP-AES was later employed to confirm [Cu] and [Fe] (Fig. 2a and ESI†).
CuFeS2 + 4Fe3+ → Cu2+ + 5Fe2+ + 2S0 | (1) |
CuFeS2 + 4H+ + O2 → Cu2+ + Fe2+ + 2S0 + 2H2O | (2) |
Fe2+ + [ox] → Fe3+ + [ox]− | (3) |
![]() | ||
Fig. 2 (a) ICP-AES confirmation of [CuSO4](aq) across a 50 sample array. Samples also contain the indicated concentrations of [Fe2(SO4)3](aq) (see ESI†) in background 75 mM H2SO4(aq). (b) ASV Cu(s) stripping charge data obtained as a result of electrodeposition within standard aqueous solution containing indicated [Cu2+] and [Fe3+], with constant 75 mM H2SO4(aq) background electrolyte. |
In the absence of ferric ions, Cu(s) stripping data (Fig. 2b) indicates that Cu(s) electrodeposition is increased by 517 nC mM−1 [Cu2+] s−1. The addition of reducible ferric ions impedes cupric electrodeposition by 83 nC mM−1 [Fe3+] s−1 at constant [Cu2+]. In cases where [Fe3+] is comparable to or greater than [Cu2+], parasitic ferric-ferrous reduction accounts for 60–100% of total reduction currents (see ESI†), leading to gross under-estimation of [Cu2+] via ASV when compared to ICP-AES reference values. Independent Cu ICP-AES sampling is essential to highlight any such ASV [Cu] measurement deviations – however, accepted reaction dynamics (1)–(3) should regulate [Fe3+] to negligible levels for CuFeS2(s) dissolution in the absence of Fe2(SO4)3 oxidant addition.
For our lixiviant systems of interest, adaptation to a simplified one pot (‘1-vial’) ASV procedure proved beneficial from numerous perspectives. Cupric electrodeposition and Cu(s) stripping can be performed back-to-back within the sample vial, whilst significantly reducing the standard deviation of ASV repeats leading to 30–70% reduction in fitting standard errors (Table 2 and ESI Fig. S8b†). Additionally, ‘1-vial’ ASV requires significantly fewer probe positioning steps, thereby minimising the combined duration of probe motion to <7% of the overall automation cycle, in turn maximising sample throughput. Hereafter, ASV experiments have been performed with the aforementioned in situ simplifications, unless otherwise stated.
Having established that ASV is capable of monitoring [Cu2+] with sufficient sensitivity in the presence of IL(aq), we then moved on to demonstrate array-based monitoring of leaching performance, as a precursor to large-scale IL(aq) screening experiments.
In order to investigate the sample-to-sample variability, 10 identical samples (4 mL; 0.45 mol dm−3 NH4·HSO4(aq); cf. Experimental) were leached for a total of 216 h. ASV and ICP-AES [Cu] measurements were obtained after 72, 120 and 216 h (Table 1, Fig. 3a). A linear fit of these two independent [Cu] measures yields a strong correlation showing <10% discrepancy (m = 1.10 ± 0.14; R2 = 0.994 – Fig. 3b). For rigour, each Cu(s) stripping charge datapoint was normalised by its corresponding ICP-determined [Cu], producing an average ‘molar’ Cu(s) stripping charge of 59 ± 5 mC mol dm−3 with low variance (rel. σ2 = 0.04%; see ESI Fig. S16†), in excellent agreement with expectations from Table 2 (cf. 56.6 ± 0.6 mC mol dm−3). A [Cu]:
[Fe] extraction ratio of unity is found, with all average ICP-AES values falling inside (low) measurement uncertainty ranges. Ultimately, sample-to-sample variation was found to be small, with relative variances in the range of 0.7–2.1%, providing adequate scope to distinguish between lixiviant performance and inter-sample variability within a screening assay, as exploited further below.
t, h | [M]av ± σ (rel. σ2), mmol dm−3 | ||
---|---|---|---|
ICP-AES [Fe] | ICP-AES [Cu] | ASV [Cu] | |
72 | 0.89 ± 0.08 (0.78%) | 0.85 ± 0.08 (0.78%) | 0.99 ± 0.08 (0.71%) |
120 | 1.07 ± 0.09 (0.70%) | 1.05 ± 0.09 (0.77%) | 1.15 ± 0.16 (2.10%) |
216 | 1.91 ± 0.20 (1.99%) | 1.91 ± 0.19 (1.92%) | 2.03 ± 0.21 (2.11%) |
Medium | pH (±0.05) | ASV Mode | m, μC mM−1 (σ) | c, μC (σ) | R 2 |
---|---|---|---|---|---|
75 mM H2SO4(aq) | 1.3 | 2-vial | 120 (±2.0) | 49.3 (±8.2) | 0.997 |
1-vial | 70.6 (±1.4) | −11.3 (±9.5) | 0.996 | ||
450 mM [C4Him][HSO4](aq) | 1.2 | 2-vial | 59.0 (±2.0) | 71.9 (±9.2) | 0.990 |
1-vial | 64.9 (±1.1) | −4.5 (±5.3) | 0.997 | ||
450 mM NH4·HSO4(aq) | 0.9 | 2-vial | 87.6 (±2.2) | 84.0 (±12.9) | 0.994 |
1-vial | 56.6 (±0.6) | −4.5 (±3.1) | 0.999 |
Subsequently, a two-variable IL(aq) lixiviant screening assay was undertaken, as a proof-of-concept experiment towards larger arrays. 25 [NH4·HSO4]-based lixiviant samples were leached at room temperature for 264 h. [NH4·HSO4](aq) (3 mL; 100–1800 mM) and CuFeS2(s) mass (37.5–600 mg) were varied logically across the two-dimensions of a square 5 × 5 sample array with ICP-AES (Fig. 4a) and ASV (Fig. 4b) [Cu] sampling after 264 h of ambient chemical leaching. ICP-determined [Fe] measurements fall within ±5% of ICP-[Cu] values, averaging 101.3 ± 1.7% of extracted [Cu] levels.
Strictly speaking, ASV Cu(s) stripping charge calibration parameters apply only at one single [IL](aq). However, for the concentration range used in this experiment, we found that the variation is in fact relatively small (Table 2). For simplicity, we used a single set of calibration parameters for all samples, namely those obtained for 450 mM [NH4·HSO4](aq) – Table 2. This decision is justified by retaining a strong correlation between [Cu] measures (m = 1.02 ± 0.04; R2 = 0.963 – see ESI Fig. S21†).
Crucially, equivalent regions of darkened ‘hotspot’ lixiviant performance are highlighted in each panel of Fig. 4. Within the 25 sample array, 2 orders of magnitude difference in leaching performance are observed between the best performing (450 mmol dm−3; 600 mg) and poorest performing combinations (1800 mmol dm−3; 75 mg). Broad variation in leaching performance is also reflected by relative variances of 70–320%; a minimum of 35-fold larger than rel. σ2 values for 10 equivalently leached [NH4·HSO4] samples (cf. Table 1). Thus, we have established confidence limits for distinguishing lixiviant performance from intersample variability, which operate on different magnitude scales.
Notably, these results suggest a non-trivial optimal [NH4·HSO4](aq) in the vicinity of 450 mmol dm−3. Since the pH is lowered as [IL](aq) increases, factors other than proton consumption (cf. eqn (2)) must play an important role during the leaching process in [NH4·HSO4](aq). Dutrizac24 found that high [SO4]2−(aq) contributed to reduced CuFeS2(s) dissolution rates. Therefore dissociation of the acidic [HSO4]− anion (pKa ∼ 1.99) to form high quantities of [SO4]2− may potentially impose limits to [IL](aq) for such lixiviant systems. We further explore this 450 mmol dm−3 [NH4·HSO4](aq) as a potential lead system in larger scale tank leaching experiments below. Overall, whilst the array size is limited in this proof-of-concept experiment, results do suggest that further, potentially significant performance enhancement may be achieved with a more comprehensive screening effort.
Proceeding to scale-up this ‘hotspot’ performance system, a two-neck round bottomed flask was used for a 120 mL scale, 6 day leaching study with automated [Cu] sensing. Freshly milled CuFeS2(s) (3 g; 32 ≤ x ≤ 75 μm) was leached at room temperature in [NH4·HSO4](aq) (450 mmol dm−3; 120 mL; 40 mL g−1), while stirred at a constant rate of ∼120 rpm. Stirring was intermittently stopped (marked * – Fig. 5a), providing extended periods of unstirred ASV for ease of calibration. A second equivalent experiment was conducted using 450 mmol dm−3 [C4Him][HSO4](aq). Our electrode system was pre-conditioned§ and inserted as a static probe, with a programmed electrochemical schedule set to ascertain ASV response at 2 h intervals, for a total leach duration of 140 h.
Beginning with 300 s, electrodeposition duration was adjusted to maintain Cu(s) stripping charges within a calibrated linear range (<1.5 mC). Calibration irregularities (plateaus and high standard deviations) have previously been observed above 2 mC, the origins of which are unclear and are under investigation (see ESI – Fig. S9†).
Fig. 5a displays the acquired Cu(s) stripping data from fast leaching 450 mmol dm−3 [NH4·HSO4](aq), when normalised for 30 s electro-deposition – see ESI for analogous [C4Him][HSO4](aq) data. A regular slope (∼10 μC h−1) indicates a near-linear Cu extraction profile (/●, Fig. 5b), with excellent [Cu] leaching sensitivity below 5 h leaching time. Ambient cupric ion leaching in 450 mmol dm−3 [NH4·HSO4](aq), occurs at a rate of 1.37 mM day−1 (1.01% Cu extraction per day), exhibiting no indication of kinetic retardation over 6 days. In one key comparative study, Ahmadi et al.25 compare unaided Fe2(SO4)3–H2SO4 chemical leaching (pH 1.8, 35 °C, 300 rpm) to enhancements attained through ORP control and microbial action. They discuss parabolic Cu extraction profiles found for unaided chemical leaching, with clear plateaus forming inside 1–5 days and resulting in extraction plateaus at <15% Cu, which persist to over 30 days. In our studies, leaching in 450 mM NH4·HSO4(aq) reaches 5.9% in 6 days at 25 °C with 120 rpm stirring. The continuation of linear Cu extraction in promising IL(aq) systems is the subject of future studies. [Fe] extraction was determined as 98.9 ± 2.5% of extracted [Cu] at all ICP-AES sampling points for NH4·HSO4 leaching, in agreement with generally accepted acid-sulfate CuFeS2(s) dissolution schemes (1)–(3), producing a Cu
:
Fe extraction ratio of unity.6,8
Despite using CuFeS2(s) from the same batch and equal [IL](aq), comparative leaching in 450 mmol dm−3 [C4Him][HSO4](aq) (□/■, Fig. 5b) produces a parabolic, kinetically slow, Cu extraction profile. Interestingly, and in stark contrast to NH4·HSO4, an initial period with very little leaching is observed below 50 h, after which point, familiar electrochemical response and [Cu] tracking is resumed (see ESI – Fig. S30a†). This further exemplifies the value of our continuous automated approach to leached [Cu] monitoring and the leach-specific insights that can be extracted from reconstruction of a time-dependent extraction profile. Extracted [Fe] levels were found to be significantly higher than that of [Cu], averaging 132.6 ± 1.4% of corresponding [Cu] (see ESI – Fig. S30b†).
Aqua regia-based digestion of the milled, unleached CuFeS2(s) starting material, confirmed the expected Cu:
Fe metal ratio of unity – see below. Differing solution pH of 450 mmol dm−3 [NH4·HSO4](aq) (0.9 ± 0.05) and [C4Him][HSO4](aq) (1.2 ± 0.05) may go some way in explaining the difference in leaching performance. However, pH alone cannot explain the presence/absence of induction periods or linear/parabolic Cu extraction behaviour for equivalent CuFeS2(s) starting material – further detailed study is required.
Notably, there are currently few published IL-CuFeS2 studies. Existing studies are disparate and difficult to compare; varying in CuFeS2(s) source, leach temperature and focused mainly on [C4C1im][HSO4].14,15 Therefore, there is a significant lack of available data for ambient IL(aq) leaching of CuFeS2(s), for comparison. However, our Cu leaching studies consistently reveal that [C4Him][HSO4](aq) outperforms [C4C1im][HSO4](aq) by up to 200% (e.g. see ESI – Fig. S31a†), earmarking the former as the superior Cu lixiviant despite bearing a higher pH at equivalent [IL](aq) (see ESI – Fig. S31b†). Once again, more detailed insight is needed regarding the pH dependence of IL(aq) leaching, as discussed above (cf. also Fig. 4). Furthermore the focal IL(aq) system presented herein, NH4·HSO4(aq), vastly outperforms [C4Him][HSO4](aq) by up to 400% over equivalent leach durations (Fig. 5b).
Overall, our automated platform for data acquisition has proven effective in addressing several challenges existing within the field of acid-sulfate hydrometallurgy. Indication of promising IL(aq) systems amongst wide-ranging leaching performances within a modest-scale screening experiment has paved the way for large array screening of unstudied IL(aq) systems, which can utilise assessed sample-to-sample variability to define confidence limits. Furthermore, we have presented a new in situ approach to automated CuFeS2(s) leach monitoring. The tool is applicable across diverse [IL](aq) systems, minimising reliance on laborious ex situ ICP-AES sampling, and allows full reconstruction of Cu extraction profiles – where leaching dynamics can be clearly observed. There are thus significant prospects in employing this approach to even larger scale studies in pursuit of next generation Cu lixiviant systems.
Screening has been characterized through parallel leaching of ten equivalent samples in 450 mmol dm−3 [NH4·HSO4](aq), revealing sample-to-sample variation of 0.7–2.1% (rel. σ2). Subsequent screening within a modest 5 × 5 sample array returned lixiviant performances ranging over 2 orders of magnitude, at least 35-fold larger than measured sample-to-sample variability. The presence of an ‘optimum’ leaching performance at [NH4·HSO4](aq) = 450 mmol dm−3 was unexpected and is incompatible with purely pH dependent leach dynamics. This lixiviant composition was then further explored in up-scaled ambient leach experiments and displayed linear extraction dynamics over 6 days of continuous [Cu] sensing. This in situ electrochemical monitoring of leached [Cu] proved effective for reconstructing full extraction profiles for two IL(aq) systems, with high time resolution. Differentiation of the two 450 mmol dm−3 [IL] systems was straightforward, through clear differences in extraction rates, the shape of the extraction profiles (i.e. linear/parabolic). Additional, potentially mechanistically relevant features were uncovered, including a 50 h dormant period for CuFeS2(s) leaching in [C4Him][HSO4](aq).
Work to-date suggests that some promising IL(aq) lixiviant systems, such as NH4·HSO4(aq), may not suffer from the same surface passivation effects as conventional ferric-acid-sulfate media,6,8 although longer duration studies with focus on other key variables (Eh, constant T etc.) are required. Moving forward, we will employ the tools introduced herein, in an iterative approach to large scale IL(aq) screening and extended electrochemical monitoring of lead systems for up-scaled studies. A broad unexplored IL chemical space awaits.
Full potentiostatic functionality is accessed through a manufacturer designed dynamic link library (.dll – Compactstat, Ivium Technologies) interfaced with the graphical programming suite. All operations are sequenced back-to-back for custom automation design, with phase completion and triggering managed by monitoring appropriately constructed instrument status signals.
Calibration plots for unstirred ASV can be rapidly generated utilising the electrochemical platform and [CuSO4](aq) solutions of known concentration. Consequently, stirring was switched off intermittently during in situ Cu sensing experiments, providing sections of reference data for rapid production of electrochemical [Cu] measurements. ICP-AES sampling provided regular [Cu] and [Fe] reference points for comparison of ASV [Cu] measurements.
In an effort to quantify the detection limit for Cu stripping under the present conditions, we divide the standard error of the intercept in Table 2 by the sensitivity, and obtain values between 0.11 and 0.31 mM, depending on the solution medium. We take this is as an estimate for the minimum stripping charge that we can detect in the present experimental configuration.
Footnotes |
† Electronic supplementary information (ESI) available: General information, detailed overview of ASV protocol, extended platform validation, raw ASV electrodeposition and stripping data, in situ Cu sensor setup, ICP-AES [Fe] extraction data. See DOI: 10.1039/c5gc02306h |
‡ Maximum of 180 samples is limited by our vial holder design and could be increased to ∼250. |
§ Prior to insertion, the electrode is pre-conditioned using five back-to-back electrodeposition/stripping cycles in 0.45 M IL(aq) containing 10 mM CuSO4(aq). |
This journal is © The Royal Society of Chemistry 2016 |